Search Filters

Search Results

Found 25 results

510(k) Data Aggregation

    K Number
    K253002

    Validate with FDA (Live)

    Date Cleared
    2026-02-19

    (153 days)

    Product Code
    Regulation Number
    878.4430
    Age Range
    34 - 120
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SkinStylus SteriLock® MicroSystem is a microneedling device and accessories intended to be used as a treatment to improve the appearance of periorbital wrinkles in Fitzpatrick skin types I, II, III, IV, V, and VI in adults aged 34 years or older.

    Device Description

    The SkinStylus SteriLock® MicroSystem (SSM) is a handheld microneedling device that uses an adjustable speed 5V 1A DC motor to rapidly reciprocate an array of 32-gauge microneedles that are no longer than 2.5 mm to create channels as well as microinjuries into the skin. The device consists of a power source, a motor body with depth adjustment, a removable nosecone interface, and a disposable, single-use, gamma ray sterilized cartridge, containing an array of microneedles.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K252752

    Validate with FDA (Live)

    Date Cleared
    2025-10-30

    (62 days)

    Product Code
    Regulation Number
    878.4430
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ellacor System with Micro-Coring Technology is indicated for use by medical professionals for the treatment of moderate and severe wrinkles in the mid and lower face in adults aged 22 years or older with Fitzpatrick skin types I-IV. It is further indicated for Fractional Skin Resurfacing.

    Device Description

    The ellacor System with Micro-Coring Technology removes unwanted skin by rapidly excising full thickness micro-columns of the dermal and epidermal tissue via specially designed hollow coring needles. The needles are controlled to precise coring depths and positions by the device console and handpiece. The system provides a range of skin removal percentages (treatment densities) and needle depths settings to customize the user needs that is controlled by the LCD touch display on the console. The treatment is applied in 1cm² areas of the skin at a time by the handpiece, which has a vacuum assisted disposable in order to help with alignment between treatment areas. The ellacor System with Micro-Coring Technology mechanically removes full thickness sections of skin without evidence of scarring or need for surgical procedure. As much as 7.9% of the surface area of the skin is removed in columns that are the size of the inner diameter of the needle which is too small to cause visible scarring. The holes quickly close because of the natural elastin of the skin as they start to heal. This wound response, combined with the mechanical removal of the skin's surface area, leads to a reduction in the appearance of wrinkles.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K253153

    Validate with FDA (Live)

    Date Cleared
    2025-10-29

    (33 days)

    Product Code
    Regulation Number
    878.4430
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SER Pen Carain MicroSystem is intended for use as a treatment to improve the appearance of facial acne scars in adults with Fitzpatrick Skin Types I - III, aged 22 years and older and as a treatment to improve the appearance of surgical or traumatic hypertrophic scars on the abdomen in adults aged 22 years or older.

    Device Description

    The SER Pen Carain MicroSystem is a handheld powered microneedling device consisting of a rechargeable motorized handpiece and disposable, single-use needle cartridges (12-pin and 36-pin). Microneedles are made of surgical-grade stainless steel (304) and are gamma sterilized.

    Design modifications include:

    • Addition of a powered 7-speed control head (6,200–9,000 RPM) with direct corded power option.
    • Modified disposable cartridge (Model T2.0), shortened by 7 mm, with protective cap and compensatory spring.
    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K243800

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-08-25

    (257 days)

    Product Code
    Regulation Number
    878.4430
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PRO Pen Microneedling System is a microneedling device and accessories intended to be used as a treatment to improve the appearance of facial acne scars in adults aged 22 years and older.

    Device Description

    The PRO Pen microneedling system is a microneedling device and disposable 14-pin cartridges intended for use in adults. The device can be powered with a rechargeable battery or a provided power adapter. The device uses disposable single-use, sterile 14-pin (14-needle) cartridges made of polymer and stainless steel.

    AI/ML Overview

    The provided FDA 510(k) Clearance Letter for the PRO Pen Microneedling System (K243800) does not describe specific acceptance criteria and the detailed study that proves the device meets those criteria in the way typically expected for an AI/CAD device. The document is a clearance letter for a physical medical device, a microneedling system, not an AI software.

    Therefore, many of the requested items (like ground truth establishment, expert qualifications, adjudication methods, MRMC studies, standalone performance of an algorithm, and specifics about training/test sets for an AI) are not applicable to this type of device clearance document.

    However, I can extract the information relevant to the performance tests conducted and the conclusions made regarding the device's adherence to regulatory requirements.

    Here's a breakdown based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document describes performance tests in accordance with special controls for microneedling devices. While explicit numerical acceptance criteria are not provided in this summary, the conclusion of the testing is that the device "successfully passed all performance tests" and "complies with FDA requirements."

    Acceptance Criteria Category (based on special controls in 21 CFR 878.4430)Reported Device Performance
    Needle Penetration Depth AccuracySuccessfully passed. (Details not specified, but implies meeting the established accuracy requirements for microneedling devices). The device can be adjusted in 0.2mm increments up to a maximum penetration depth of 1.5mm.
    Puncture rate characterizationSuccessfully passed. (Details not specified, but implies meeting the established rate charactersitics).
    Vacuum and fluid ingress prevention evaluationSuccessfully passed. Demonstrated reliable performance across all critical tests, including vacuum prevention and fluid ingress.
    Sterility of patient contacting componentsSuccessfully passed.
    Aging distribution and shelf-life testingSuccessfully passed. 14-pin cartridges have a shelf life of 3 years.
    Electromagnetic compatibility and electrical safetySuccessfully passed. Confirmed safe operation in healthcare environments.
    In vitro cytotoxicitySuccessfully passed.
    Material characterization through extractables and leachablesSuccessfully passed. (Toxicological risk assessment of findings also passed).
    Human patch testing for irritationSuccessfully passed. No irritation reactions observed. Classified as non-irritating.
    Human repeat insult patch testing (HRIPT)Successfully passed. No sensitization reactions observed. Classified as non-sensitizing.
    Cleaning and disinfection validationSuccessfully passed. Procedures were validated, confirming the device can be safely cleaned and disinfected to reduce contamination risks.
    Cross contamination protectionComplies with FDA requirements (Implied by Vacuum and fluid ingress prevention evaluation and Cleaning/disinfection validation).
    Packaging durabilityComplies with FDA requirements (Implied by aging/shelf-life testing and overall safety). Implied by successful vacuum and fluid ingress prevention which often relates to packaging integrity.
    Needle retentionDemonstrated reliable performance.

    2. Sample size used for the test set and the data provenance

    The document does not specify general "test set" sample sizes or data provenance (e.g., country of origin, retrospective/prospective) because this clearance is for a physical device, not an AI model. The "tests" mentioned are engineering and biocompatibility evaluations. For human patch testing (HRIPT), the sample size would typically be small (e.g., 50-200 subjects) for such tests, but no specific number is provided in this summary.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This is not an AI/CAD system requiring expert ground truth for image interpretation or diagnosis. The "ground truth" for the device's performance is established by objective engineering standards, biological assays (cytotoxicity, extractables/leachables), and clinical observation (human patch testing for irritation/sensitization).

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. As this is not an AI/CAD system evaluating diagnostic images, there would be no adjudication method involving multiple human readers.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI system.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI system.

    7. The type of ground truth used

    The "ground truth" for the PRO Pen Microneedling System's performance is based on:

    • Engineering Standards: For needle penetration depth, puncture rate, electrical safety, EMC.
    • Biological/Chemical Testing Standards: For sterility, vacuum/fluid ingress, cytotoxicity, extractables/leachables, material characterization.
    • Clinical Observation/Standard Dermatological Assessments: For human patch testing (irritation and sensitization reactions).
    • Microbiological Standards: For cleaning and disinfection validation.
    • Stability/Accelerated Aging Protocols: For shelf-life testing.

    8. The sample size for the training set

    Not applicable. This is not an AI system.

    9. How the ground truth for the training set was established

    Not applicable. This is not an AI system.

    Ask a Question

    Ask a specific question about this device

    K Number
    K243143

    Validate with FDA (Live)

    Device Name
    E-PEN (E-PEN)
    Date Cleared
    2025-04-07

    (189 days)

    Product Code
    Regulation Number
    878.4430
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The E-PEN is a microneedling device and accessories intended to be used as a treatment to improve the appearance of facial acne scars in adults aged 22 years and older.

    Device Description

    The E-PEN is a microneedling device that mechanically creates microscopic punctures in the epidermal and dermal layers of the skin by means of micro-needles in a reciprocating cartridge head. The E-PEN is comprised of a reusable pen body, a sterile, single use microneedling cartridge, and a power adapter. The microneedling cartridge is attached to the pen body and activated with an On/Off button. The depth of needle penetration can be adjusted by the user depending on the condition of the skin being treated. Charging is accomplished by attaching the E-PEN pen body to the power adapter.

    AI/ML Overview

    This document is a 510(k) clearance letter for the E-pen microneedling device. It contains regulatory information and a summary of performance data. However, it does not contain information about acceptance criteria or a study proving the device meets those criteria in the context of clinical effectiveness or performance as an AI/software device. This type of information is typically found in clinical study reports, scientific publications, or more detailed regulatory submissions for AI-powered devices, which are not present in this 510(k) clearance letter for a microneedling device.

    The provided information focuses on:

    • Regulatory Classification: The device is a Class II microneedling device for aesthetic use (21 CFR 878.4430).
    • Intended Use: To improve the appearance of facial acne scars in adults aged 22 years and older.
    • Technical Specifications: Comparison with a predicate device (SkinPen Precision System, DEN160029) showing similar design, materials, and operating parameters (e.g., number of needles, needle gauge, penetration depth, speed).
    • Non-Clinical Performance Data: Biocompatibility, sterilization validation, sterility/shipping/shelf-life, EMC (Electromagnetic Compatibility), electrical safety, software validation (moderate level of concern), and bench testing (needle length, penetration depth, speed, prevention of re-use).

    Therefore, I cannot provide the requested information regarding acceptance criteria and a study proving clinical effectiveness or AI performance based on this document. The document primarily demonstrates substantial equivalence to a predicate device based on technical and safety specifications, not a new clinical performance study against specific acceptance criteria for a health outcome based on AI.

    If this were a submission for an AI-powered device, the information requested (AI performance, ground truth, experts, etc.) would be crucial and would typically be provided in a different section or supplementary documentation.

    Ask a Question

    Ask a specific question about this device

    K Number
    K243472

    Validate with FDA (Live)

    Device Name
    1NEED Pro
    Manufacturer
    Date Cleared
    2025-03-19

    (131 days)

    Product Code
    Regulation Number
    878.4430
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The 1NEED Pro is a microneedling device and accessories intended to be used as a treatment to improve the appearance of facial acne scars in adults aged 22 years or older.

    Device Description

    The 1NEED Pro is a minimally invasive microneedling device that microscopic punctures in the epidermal and dermal layers of the skin by means of micro-needles in a reciprocating cartridge head. The 1NEED Pro is comprised of a reusable pen body, a sterile, single use microneedling cartridge and a power adapter. The microneedling cartridge is attached to the pen body and activated with an On/Off button. The depth of needle penetration can be adjusted by the user depending of the skin being treated. Charging is accomplished by attaching the 1NEED Pro pen body to the power adapter.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a microneedling device named "1NEED Pro." While it discusses the device's intended use, technological characteristics, and results from non-clinical bench testing, it does not contain information about acceptance criteria or a study proving the device meets said criteria in the context of clinical performance (e.g., efficacy in improving acne scars).

    The document focuses on demonstrating substantial equivalence to a predicate device (K230420) based on similar indications for use and technological features, and the results of various engineering and safety bench tests.

    Therefore, I cannot provide a table of acceptance criteria and reported device performance related to efficacy (e.g., reduction of acne scars) or details about a clinical study. The document explicitly mentions "Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)" but then only lists "Bench testing."

    Here's a breakdown of what can be extracted based on the provided text, and what is explicitly missing:

    1. Table of Acceptance Criteria and Reported Device Performance (Related to Efficacy - Missing)

    • There is no information on acceptance criteria for clinical effectiveness (e.g., a specific percentage improvement in acne scar appearance) or reported device performance against such criteria. The document focuses on demonstrating that the device performs similarly to its predicate and passes engineering/safety tests.

    2. Sample Size Used for the Test Set and Data Provenance (Missing for Clinical Efficacy)

    • The document lists various bench tests (e.g., Motor Speed Puncture Rate Testing, Needle Penetration Depth and Extension Accuracy Testing, etc.). For these engineering tests, the sample sizes and data provenance are not specified.
    • For clinical efficacy, this information is entirely missing, as no clinical study is described.

    3. Number of Experts Used to Establish Ground Truth and Qualifications (Missing)

    • This information is typically relevant for studies involving human assessment (e.g., assessing the severity of acne scars before and after treatment). Since no clinical efficacy study is described, this information is missing.

    4. Adjudication Method (Missing)

    • Similar to point 3, adjudication methods are used in clinical studies to resolve discrepancies between expert assessments. Since no clinical efficacy study is described, this information is missing.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study (Not Conducted/Not Described)

    • The document does not mention any MRMC study. It focuses on the device performing as well as the predicate through technological comparison and non-clinical tests.

    6. Standalone (Algorithm Only) Performance (Not Applicable/Missing)

    • The "1NEED Pro" is a physical microneedling device, not an algorithm or AI software for diagnosis or image analysis. Therefore, a standalone performance study in that context is not applicable.

    7. Type of Ground Truth Used (Missing for Clinical Efficacy)

    • For the described bench tests, the "ground truth" would be engineering specifications or expected physical properties, which were then measured against.
    • For clinical efficacy, this information is missing as no clinical study defining a ground truth for acne scar improvement is described.

    8. Sample Size for the Training Set (Not Applicable/Missing)

    • The device is a physical microneedling pen, not a machine learning algorithm that requires a training set in the conventional sense. Therefore, this information is not applicable.

    9. How the Ground Truth for the Training Set was Established (Not Applicable/Missing)

    • As above, this is not applicable for a physical device.

    Summary of available information from the document related to "acceptance criteria" and "proof":

    The document argues for substantial equivalence primarily based on:

    • Same intended use: "The 1NEED Pro is a microneedling device intended to be used as a treatment to improve the appearance of facial acne scars in adults aged 22 years or older." This is identical to the predicate device.
    • Same technological characteristics and mode of action: Both the 1NEED Pro and the predicate device use microneedling for the stated indication. Specific characteristics like power source, control mechanism, speed, puncture rate, needle details (number, gauge, material, shape, spacing), penetration depth settings, and sterility of cartridges are compared and deemed similar.
    • Non-clinical (bench) testing: These tests demonstrate that the device meets engineering specifications and safety requirements. The conclusion states: "Nonclinical tests demonstrate that the device is as safe, as effective, and performs as well as the legally marketed predicate device."

    The listed bench tests are:

    • Motor Speed Puncture Rate Testing
    • Needle Penetration Depth and Extension Accuracy Testing
    • Needle Bonding Strength Test
    • Use Life Testing
    • Cartridge Life Testing
    • Anti-suction Testing
    • Microbial Ingress Testing

    The document does not provide specific numerical acceptance criteria or results for these bench tests in a detailed table or report. It only lists that these tests were conducted and concluded the device is "as safe, as effective, and performs as well as the legally marketed predicate device." This implies that the device met the internal specifications and performance levels established as acceptable for these bench tests, leading to the conclusion of substantial equivalence. However, the exact criteria and results are not detailed in this 510(k) summary.

    Ask a Question

    Ask a specific question about this device

    K Number
    K241790

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-03-06

    (258 days)

    Product Code
    Regulation Number
    878.4430
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The XCELLARISPRO TWIST microneedling device and accessories is intended to be used as a treatment to improve the appearance of facial acne scars in Fitzpatrick skin types I. II, III and IV in adults aged 22 years or older and for treatment of wrinkles in Fitzpatrick skin types I, II and/or III in the following facial areas: glabellar frown lines, periorbital lines and cheek folds in adults aged 22 years or older.

    Device Description

    The XCELLARISPRO TWIST is a microneedling device and intended to create many, very tiny, microscopic punctures in the epidermal and dermal layers of the skin using sterile stainlesssteel needles. The XCELLARISPRO TWIST has 4 component parts and 2 consumables. Component parts: (1) handpiece, (2) control unit, (3) power supply and (4) foot switch. Consumables: (1) sterile, single-use needling module (Ethylene oxide sterilized) and (2) disposable barrier sleeves that cover the handpiece during each patient use.

    AI/ML Overview

    The provided text describes the acceptance criteria and supporting studies for the XCELLARISPRO TWIST microneedling device.

    It's important to note that this submission does not include clinical performance testing for the XCELLARISPRO TWIST device itself. Instead, it relies on non-clinical performance testing and substantial equivalence to predicate devices, which did undergo clinical studies.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are derived from the "Special Controls" for microneedling devices (21 C.F.R. § 878.4430(b)(1)-(10)) and the non-clinical performance data provided. The document primarily focuses on verifying technical specifications and safety aspects.

    Acceptance Criterion (Special Control)Reported Device Performance
    1. Technical specifications and needle characteristics identified (needle length, geometry, puncture rate, maximum protrusion depth).Needle Length: 2.5 mm (each of 6 stainless steel microneedles) Needle Geometry: Not explicitly detailed beyond "6 stainless steel microneedles in squared arrangement." Puncture Rate: 50-150 per second (adjustable, ±10% variation) Maximum Protrusion Depth: 2.5 mm (recommended max 1.5 mm)
    2. Non-clinical performance data demonstrate that the device performs as intended under anticipated conditions of use.i. Accuracy of needle protrusion depth and puncture rate: Tested in pig skin with high-speed camera measurements. ii. Safety features: Fluid ingress protection via bellows confirmed. iii. Maximum safe needle protrusion depth: Identified in pig skin with high-speed camera measurement.
    3. Performance data demonstrates sterility of patient-contacting components (needling module) according to ISO 11737-2, ISO 11135.Needling module provided sterile, Ethylene oxide sterilized. (Implicitly, performance data supports this claim.)
    4. Performance data supports shelf life (continued sterility, package integrity, device functionality) according to ISO 11607-1.(Implicitly, performance data supports this claim as it is a required special control.)
    5. Performance data demonstrates electrical safety and EMC of all electrical components according to IEC 60601-1 and IEC 60601-1-2.(Implicitly, performance data supports this claim as it is a required special control.)
    6. Software verification, validation, and hazard analysis.(Implicitly, performed as it is a required special control.)
    7. Biocompatibility of patient-contacting components (cytotoxicity, irritation, sensitization, acute systemic toxicity, material-mediated pyrogenicity) per ISO 10993-1.(Implicitly, tested and found biocompatible as it is a required special control.)
    8. Cleaning and disinfection validations performed for reusable components.Surfaces of control unit and handpiece are made of materials which can be cleaned and disinfected (Implicitly, validations performed.)
    9. Labeling includes: device operation, typical course of treatment, technical parameters, validated reprocessing methods, disposal instructions, shelf life.(Implicitly, confirmed that labeling includes these items as per special control.)
    10. Patient labeling includes: device operation, typical course of treatment, probable risks/benefits, post-operative care.(Implicitly, confirmed that patient labeling includes these items as per special control.)

    2. Sample size used for the test set and the data provenance

    • Test Set Sample Size:
      • For testing "Accuracy of needle protrusion depth and puncture rate" and "Maximum safe needle protrusion depth," pig skin was used. The specific number of pig skin samples or measurements is not provided in the document.
      • For other non-clinical tests (sterility, shelf life, electrical safety, biocompatibility, cleaning/disinfection), the sample sizes are not specified in this summary.
    • Data Provenance: The tests are stated as "non-clinical performance testing." The location or specific origin of the data/testing facility is not specified. It is retrospective in the sense that the tests were completed before the 510(k) submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • This information is not provided in the document. The non-clinical tests described (e.g., high-speed camera measurements on pig skin, electrical safety, sterility) typically rely on standardized measurement techniques and validated protocols rather than expert consensus on a "ground truth" derived from human observation.

    4. Adjudication method for the test set

    • An adjudication method is typically relevant for studies involving human interpretation or subjective assessment. Since the described tests are non-clinical and objective measurements (e.g., physical parameters, microbiological tests, electrical tests), an adjudication method is not applicable and therefore not mentioned.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No MRMC comparative effectiveness study was done. This device is a microneedling device, not an AI software or diagnostic tool that would involve human readers interpreting cases.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • No standalone algorithm performance study was done. This device is a hardware microneedling device, not an algorithm.

    7. The type of ground truth used

    • For the non-clinical performance tests:
      • Technical specifications: Measured physical parameters (e.g., needle length, protrusion depth, puncture rate) against engineering specifications.
      • Sterility: Absence of microorganisms (based on microbiological testing).
      • Biocompatibility: Response of biological systems to materials (based on standardized in-vitro/in-vivo tests).
      • Electrical Safety/EMC: Compliance with international safety standards.
      • Cleaning/Disinfection Validation: Reduction of microbial load after reprocessing.
      • The "ground truth" for these tests is defined by established scientific principles, recognized standards, and specific pass/fail criteria outlined in those standards.

    8. The sample size for the training set

    • Not applicable. This device is a microneedling device and does not involve machine learning or a "training set" in the context of an AI/algorithm.

    9. How the ground truth for the training set was established

    • Not applicable. As above, there is no training set for this device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K241355

    Validate with FDA (Live)

    Date Cleared
    2024-10-09

    (148 days)

    Product Code
    Regulation Number
    878.4430
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SER Pen Carain MicroSystem is intended for use as a treatment to improve the appearance of facial acne scars in adults with Fitzpatrick Skin Types I - III, aged 22 years and older and as a treatment to improve the appearance of surgical or traumatic hypertrophic scars on the abdomen in adults aged 22 years or older.

    Device Description

    The SER Pen Carain MicroSystem (MP1209SP) is a handheld device that creates channels as well as microinjuries into the skin, by virtue of a DC motor drive system that rapidly reciprocates an array of 12 needles and 36 needles that are no longer than 2.5mm. The device consists of a power source, a motor body with depth adjustment, and a disposable, single use cartridge containing an array of needles. The power source consists of a rechargeable lithium-ion battery that delivers no more than 5 volts DC and 1 amp of current to power the motor. A power cord connects the wall adaptor to the device via a USB connector and a standard 1/8" headphone plug on the device side. The motor body is comprised of anodized aluminum with a dial mechanism that controls the depth of penetration of the microneedles from 0.0 mm to a maximum of 2.5mm. The SER Pen Carain MicroSystem's disposable cartridge is designed in two configurations, a 12-needle array and a 36-needle array. The needle array is housed in a specially designed cartridge housing with a silicon spring that prevents liquids from entering the motor body via the inside lumen of the cartridge. The SER Pen Carain MicroSystem biolock sleeve is a one end elastic opening PE bag to protect the whole system from cross contamination from the outside during the operation procedure. The SER Pen Carain MicroSystem microneedles are composed of 304 18/8 surgical steel tested and certified as biocompatible under GLP testing conditions that conform to ISO 10993-5, 10993-10, and 10993-11. A metallurgical analysis under GLP testing conditions that conform to ASTM E1019-11(Method A)(C/S Analyzer) was also completed and is attached. Each lot of cartridges are individually packaged and then gamma ray sterilized by TUV Sud accredited facility under procedure code P1501314 with a minimum dose of 25kGy. Certifications on file with the Sponsor and available for review.

    AI/ML Overview

    This document describes the SER Pen Carain MicroSystem (MP1209SP), a microneedling device cleared via a 510(k) submission (K241355) by demonstrating substantial equivalence to predicate devices, rather than through a direct study proving specified acceptance criteria.

    The submission specifically states: "The Sponsor intends to rely on the same performance data submitted in Collagen P.I.N. (K222199) and SkinStylus (K200044 and K231073) devices and not include any clinical data." This means that acceptance criteria and performance data are inferred from the predicate devices through a comparison of technological characteristics, rather than direct testing of the SER Pen Carain MicroSystem against pre-defined acceptance criteria with a dedicated study.

    Therefore, the following information is based on the provided document and the nature of a 510(k) submission relying on predicate devices.

    1. Table of Acceptance Criteria and Reported Device Performance

    Since this is a 510(k) submission relying on predicate devices, explicit "acceptance criteria" and "reported device performance" in the context of a de novo clinical study are not directly presented for the SER Pen Carain MicroSystem. Instead, the "acceptance criteria" are implied by meeting the same performance and safety standards demonstrated by the predicate devices, and "device performance" is established by showing identical or highly similar technological characteristics and performance to those predicates.

    Based on the document, the "performance" of the SER Pen Carain MicroSystem is derived from its substantial equivalence to the predicate devices across various technological characteristics.

    Feature/TestAcceptance Criteria (Implied by Predicate Devices)SER Pen Carain MicroSystem Performance (Reported as "Identical" or similar to Predicates)
    Indications for UseImprove appearance of facial acne scars (Fitzpatrick Skin Types I-III, 22+ years); Improve appearance of surgical/traumatic hypertrophic scars on abdomen (22+ years).Intended for use to improve appearance of facial acne scars in adults with Fitzpatrick Skin Types I - III, and to improve appearance of surgical or traumatic hypertrophic scars on the abdomen in adults aged 22 years and older. (Matches combined indications of predicates)
    Mode of ActionMicroneedling (mechanically puncturing skin for aesthetic use)Identical (Microneedling)
    Power Source5 Volt DC/1 amp rechargeable lithium-ion batterySame: 5Volt DC/ 1 amp rechargeable lithium-ion battery
    Length/Width/shape/weightIdentical to predicate devicesIdentical to predicate device(s)
    Range of Needle Length0.0-3.0mm (Collagen P.I.N.) or 0.0-2.5mm (SkinStylus SteriLock)0.0-2.5mm
    Maximum Penetration3.0mm (Collagen P.I.N.) or 2.5mm (SkinStylus SteriLock)2.5mm
    Needle Geometry36 solid needles (Collagen P.I.N.) or 12 solid needles (SkinStylus SteriLock)Identical, 12 or 36 solid needles
    Speed7000RPM-9000RPM (Collagen P.I.N.) or 6200RPM-9000RPM (SkinStylus SteriLock)Identical, 6200RPM-9000RPM
    Sterility & CleaningDisposable cartridge, gamma ray or ethylene oxide sterilized prior to packagingSame. Disposable cartridge Gamma Ray Sterilized prior to packaging
    Cross Contamination FeatureCartridge design with sealing spring or other features to prevent cross-contaminationIdentical: Cartridge design with a sealing spring inside the cartridge housing, and a biolock sleeve to cover from the cartridge outshell and the whole handpiece to stop cross contamination.
    BiocompatibilityConforms to ISO 10993-5, 10993-10, 10993-11 standardsMet (through reliance on predicate data and direct testing of microneedles per ISO 10993-5, -10, -11, and ASTM E1019-11 for metallurgical analysis, as listed in "Summary of Non-Clinical Test Reports" for the needles themselves).
    Sterilization ValidationConforms to ISO 17665-1, ISO 11737-1, ISO 11737-2, USP 71Met (through reliance on predicate data and direct testing per ISO 17665-1, ISO 11737-1, ISO 11737-2, and USP 71).
    Reprocessing ValidationConforms to ISO 11737-1, AAMI TIR-30, AAMI TIR-12Met (through reliance on predicate data and direct testing per ISO 11737-1, AAMI TIR-30, AAMI TIR-12).
    Fluid Ingress ValidationDemonstrated to prevent fluid ingress into motor/housingMet (through reliance on predicate data and similarity of design).
    Shelf-life TestingConforms to ASTM F88/F88M-15 and ASTM F1929-15 for seal strength and leak detectionMet (through reliance on predicate data and direct testing per ASTM F88/F88M-15, ASTM F1929-15).

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: This information is not provided as the submission relies on existing data from predicate devices. There was no specific "test set" for the SER Pen Carain MicroSystem in the context of a new clinical study.
    • Data Provenance: The document explicitly states the "Sponsor intends to rely on the same performance data submitted in Collagen P.I.N. (K222199) and SkinStylus (K200044 and K231073) devices and not include any clinical data." Therefore, the data provenance would be from the studies conducted for these predicate devices, which are not detailed in this document. Information on country of origin or whether the data was retrospective or prospective is not available in this document.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • This information is not applicable as no new clinical study with a "test set" requiring expert ground truth was conducted for the SER Pen Carain MicroSystem. The data used for establishing substantial equivalence comes from pre-existing regulatory submissions for the predicate devices.

    4. Adjudication Method for the Test Set

    • This information is not applicable as no new clinical study with a "test set" requiring adjudication was conducted for the SER Pen Carain MicroSystem.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

    • No, an MRMC comparative effectiveness study was not done. The submission explicitly states "not include any clinical data," indicating a reliance on non-clinical data and comparison to predicate devices for substantial equivalence.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    • No, this is a microneedling device, not an AI algorithm. "Standalone performance" in the context of an algorithm is not applicable.

    7. The Type of Ground Truth Used

    • As no new clinical studies were conducted for the SER Pen Carain MicroSystem, the concept of "ground truth" in this context is not directly applicable. The basis for clearance is established through demonstrating similar technological characteristics and performance to predicate devices that were previously cleared by the FDA based on their own evidence (which may have included clinical data, expert consensus, etc., but that information is not present here). For the non-clinical tests (biocompatibility, sterilization, etc.), the "ground truth" is adherence to established international and national standards (ISO, ASTM, USP).

    8. The Sample Size for the Training Set

    • This information is not applicable as there was no AI algorithm or "training set" for this device.

    9. How the Ground Truth for the Training Set was Established

    • This information is not applicable as there was no AI algorithm or "training set" for this device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K241400

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-08-29

    (104 days)

    Product Code
    Regulation Number
    878.4430
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SkinPen® Precision Elite system is a microneedling device and accessories intended to be used as a treatment to improve the appearance of wrinkles of the neck for Fitzpatrick skin types II IV and to improve the appearance of facial acne scars in adults with all Fitzpatrick skin types aged 22 vears and older.

    Device Description

    The SkinPen® Precision Elite System consists of a microneedling pen handpiece (SkinPen® Precision Elite) and a sterile needle cartridge (SkinPen® Precision Elite Cartridge). The accessories are a charging base with power adaptor and a BioSheath to cover the handpiece.

    AI/ML Overview

    The provided text is a 510(k) summary for the SkinPen® Precision Elite System. It details the device, its intended use, and comparison to a predicate device. However, it explicitly states "No Clinical testing was conducted as part of this submission" (Section 12, page 6).

    Therefore, there is no information within this document to address the requested points regarding acceptance criteria and a study proving the device meets those criteria, as no clinical study was performed for this submission. The device's substantial equivalence was based on non-clinical performance testing (reciprocating rate, puncture depth), electrical safety, risk management, sterility, software verification and validation, and biocompatibility, along with a direct comparison to a predicate device.

    To answer your request, if a clinical study had been performed and described in the document, here's what the response would look like if the information were present:


    Based on the provided 510(k) summary, no clinical performance study was conducted to demonstrate the device meets specific acceptance criteria based on human subject data. The submission relies on substantial equivalence to a predicate device and non-clinical testing.

    Therefore, I cannot provide the detailed information requested in points 1 through 9.

    Explanation from the document:

    • Section 12 "Clinical Testing Summary" (Page 6): "No Clinical testing was conducted as part of this submission."

    The acceptance criteria and study detailed in your request typically refer to clinical performance studies, which were not performed for this device's submission as stated. The changes made to the device (user interface, cartridge connection points, NFC chip for verification) were supported by:

    • Non-clinical performance testing (reciprocating rate, puncture depth).
    • Software verification and validation.
    • Electrical safety standards (IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6).
    • Biocompatibility testing (ISO 10993-1).
    • Sterility standards (ISO 11135).
    • Risk Management (IEC 62304).

    The submission argues that these changes "do not impact the device effectiveness, and safety is increased with the use of the NFC chip," and that "There are no changes to the general technology, mechanism of action, or indications for use." This forms the basis for demonstrating substantial equivalence without a new clinical study.

    Ask a Question

    Ask a specific question about this device

    K Number
    K233709

    Validate with FDA (Live)

    Device Name
    Exceed Unlimited
    Manufacturer
    Date Cleared
    2024-02-23

    (95 days)

    Product Code
    Regulation Number
    878.4430
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Exceed Unlimited is a microneedling device and accessories intended for the treatment of wrinkles in Fitzpatrick skin types I. Il and/or III in the following facial areas: qlabellar frown lines, periorbital lines and cheek folds in adults aged 22 years or older.

    The Exceed Unlimited is a microneedling device and accessories intended to be used as a treatment to improve the appearance of facial acne scars in Fitzpatrick skin types I, II, III and IV in adults aged 22 years or older.

    Device Description

    The Exceed Unlimited device is intended to create many, very tiny, microscopic punctures in the epidermal and dermal layers of the skin using sterile stainless steel needles.

    The Exceed Unlimited device consists of 7 component parts; Handpiece, sterile single use Safety needle cartridge, Handpiece holder, Barrier Sleeve, Battery, Charger incl. USB cable, power supply. The handpiece contains a motor that moves the needles. The handpiece receives power via a Li-Ion battery with an output of 3,7 V. The battery charger receives power from a standard 100-240V, 50-60Hz, 0.45A wall socket transformer with an USB output of 5 V, 3.2A.

    The frequency of the handpiece can be adjusted from 100-150Hz and is indicated on a display. The needle penetration is adjusted by the needle protrusion setting dial, which can be adjusted between 0.0 (min.) and 1.9 mm (max.).

    The standard safety needle cartridge (6-needle plate) contains 6 stainless steel microneedles of 1.5 mm length (0.35 mm gauge) and an internal safety membrane. The safety needle cartridge is screwed into the handpiece. The safety needle cartridge is sterile and for single use ONLY.

    AI/ML Overview

    The provided text is a 510(k) Summary for the Exceed Unlimited microneedling device. It describes the device, its intended use, and compares it to a predicate device (Exceed Microneedling device) to demonstrate substantial equivalence.

    However, the document does not contain specific acceptance criteria for a study demonstrating device performance, nor does it present the results of such a study in a quantitative manner (e.g., a table of performance metrics). Instead, it primarily focuses on non-clinical performance testing and risk mitigation measures to establish substantial equivalence.

    Based on the provided text, I can infer some aspects and highlight what is explicitly stated versus what is absent:

    1. A table of acceptance criteria and the reported device performance:

    • Absent. The document does not provide a table of acceptance criteria nor reported quantitative device performance metrics from a study. It only lists performance characteristics that were tested during non-clinical performance testing.

    2. Sample size used for the test set and the data provenance:

    • Absent directly. The document mentions "non-clinical performance data" and "accuracy of needle penetration depth and puncture rate in pig skin." While this indicates the type of test, it does not specify the sample size used (e.g., number of pigs, number of punctures, number of test runs).
    • Data Provenance: The pig skin tests would be considered experimental/laboratory data, not human patient data. The document does not mention the country of origin for these tests. It implicitly would be from the manufacturer's testing facilities or a contracted lab. The tests are not from retrospective or prospective human clinical studies for performance metrics.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Absent. This information is relevant for human-in-the-loop studies or studies relying on expert-verified data, typically clinical studies. Since the described performance testing is non-clinical (on pig skin), the concept of "experts establishing ground truth" in the diagnostic sense is not applicable here. The "ground truth" for the pig skin tests would be the measured physical properties (e.g., actual penetration depth, puncture rate) using high-speed cameras or other measurement tools.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Absent. Adjudication methods are typically used in clinical trials where multiple readers or experts assess outcomes. As the described performance testing is non-clinical/technical, this concept does not apply.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Absent. This device is a microneedling device, not an AI or imaging diagnostic tool. Therefore, an MRMC study comparing human readers with and without AI assistance is not relevant to its function and was not performed or described.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable in the typical sense of AI/software. This is a hardware microneedling device. Its "performance" is evaluated by its physical characteristics (e.g., needle penetration accuracy, frequency, safety features). There isn't an "algorithm only" performance metric in the way it would apply to a diagnostic AI. The performance data mentioned (e.g., accuracy of needle penetration) are inherent to the device's mechanical operation.

    7. The type of ground truth used:

    • For the non-clinical performance testing mentioned ("Accuracy of needle penetration depth and puncture rate in pig skin with high-speed camera measurement"), the ground truth would be physical measurements (e.g., actual depth measured from camera footage or other instruments). It is not expert consensus, pathology, or outcomes data in the clinical sense.

    8. The sample size for the training set:

    • Not applicable. This section is relevant for machine learning models. The Exceed Unlimited is a physical medical device, not a software AI model that requires a training set.

    9. How the ground truth for the training set was established:

    • Not applicable. As per point 8, there is no training set for this type of device.

    In summary, based solely on the provided text, the document focuses on demonstrating substantial equivalence through non-clinical performance and safety testing, rather than presenting a performance study with acceptance criteria and quantitative results. The provided information is insufficient to complete the requested table and fully answer all questions, as many are relevant to AI/diagnostic software devices, not a microneedling device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 3