Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K242545

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-05-23

    (269 days)

    Product Code
    Regulation Number
    864.3700
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    RadiForce MX317W-PA is intended for in vitro diagnostic use to display digital images of histopathology slides acquired from IVD-labeled whole-slide imaging scanners and viewed using IVD-labeled digital pathology image viewing software that have been validated for use with this device.

    RadiForce MX317W-PA is an aid to the pathologist and is used for review and interpretation of histopathology slides for the purposes of primary diagnosis. It is the responsibility of the pathologist to employ appropriate procedures and safeguards to assure the validity of the interpretation of images using this product. The display is not intended for use with digital images from frozen section, cytology, or non- formalin-fixed, paraffin embedded (non-FFPE) hematopathology specimens.

    Device Description

    RadiForce MX317W-PA is a color LCD monitor for viewing digital images of histopathology slides. The color LCD panel employs in-plane switching (IPS) technology allowing wide viewing angles and the matrix size is 4,096 x 2,160 pixels (8MP) with a pixel pitch of 0.1674 mm.

    Since factory calibrated display modes, each of which is characterized by a specific tone curve, a specific luminance range and a specific color temperature, are stored in lookup tables within the monitor. This helps ensure tone curves even if a display controller or workstation must be replaced or serviced.

    "Patho" is for intended digital pathology use mode.

    AI/ML Overview

    The provided FDA 510(k) clearance letter for the RadiForce MX317W-PA describes a display device for digital histopathology. It does not contain information about an AI/ML medical device. Therefore, a study proving the device meets acceptance criteria related to AI/ML performance (such as accuracy, sensitivity, specificity, MRMC studies, and ground truth establishment methods for large datasets) is not present in this document.

    The document primarily focuses on the technical performance and equivalence of a display monitor to a predicate device. The "performance testing" section refers to bench tests validating display characteristics like spatial resolution, luminance, and color, not the clinical performance of an AI algorithm interpreting medical images.

    Given the information provided, here's an analysis based on the actual content:


    Based on the provided document, the RadiForce MX317W-PA is a display monitor, not an AI/ML medical device designed for image interpretation. Therefore, the acceptance criteria and study detailed below pertain to the display's technical performance and its equivalence to a predicate display, not to an AI algorithm's diagnostic accuracy.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document states that "the display characteristics of the RadiForce MX317W-PA meet the pre-defined criteria when criteria are set." However, the exact numerical acceptance criteria for each bench test (e.g., minimum luminance, pixel defect limits) are not explicitly listed in the provided text. The document only lists the types of tests performed and states that the device "has display characteristics equivalent to those of the predicate device" and "meet the pre-defined criteria."

    Acceptance Criteria CategoryReported Device Performance Summary (as per document)
    User controls (Modes & settings)Performed, assumed met
    Spatial resolutionPerformed, assumed met, equivalent to predicate
    Pixel defectsPerformed, assumed met, equivalent to predicate
    ArtifactsPerformed, assumed met, equivalent to predicate
    Temporal responsePerformed, assumed met, equivalent to predicate
    Maximum and minimum luminancePerformed, assumed met, equivalent to predicate
    GrayscalePerformed, assumed met, equivalent to predicate
    Luminance uniformity and Mura testPerformed, assumed met, equivalent to predicate
    Stability of luminance and chromaticity responsePerformed, assumed met, equivalent to predicate
    Bidirectional reflection distribution functionPerformed, assumed met, equivalent to predicate
    Gray TrackingPerformed, assumed met, equivalent to predicate
    Color scalePerformed, assumed met, equivalent to predicate
    Color gamut volumePerformed, assumed met, equivalent to predicate

    Note: The document only states that these tests were performed and that the results show equivalence to the predicate device and that the device meets pre-defined criteria. It does not provide the specific numerical results or the exact numerical acceptance criteria for each test.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: The document describes bench tests performed on a single device, the RadiForce MX317W-PA (it's a physical monitor, not a software algorithm processing a dataset). There is no mention of a "test set" in the context of a dataset of medical images.
    • Data Provenance: Not applicable. The "data" here refers to the measured performance characteristics of the physical display device itself during bench testing, not patient data.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • Not applicable. The ground truth for a display monitor's technical performance is established by standardized measurement equipment and protocols, not by expert interpretation of images. The device itself is the object under test for its physical characteristics.

    4. Adjudication Method for the Test Set

    • Not applicable. This concept applies to human or AI interpretation of medical images, where discrepancies among readers or algorithms might need resolution. For physical device performance, measurements are generally objective.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • Not performed/Applicable. An MRMC study is designed to assess the performance of a diagnostic aid (like AI) on image interpretation by human readers. This device is a display monitor, not an AI algorithm. Its function is to display images, not to interpret them or assist human interpreters in a diagnostic decision-making process that would warrant an MRMC study.

    6. Standalone (i.e., Algorithm Only Without Human-in-the-Loop Performance) Study

    • Not applicable. As stated, this is a display monitor, not an algorithm.

    7. Type of Ground Truth Used:

    • The ground truth for the display's performance tests would be metrology-based standards and calibration references (e.g., standard luminance values, colorimetry standards) against which the display's output is measured. It is not expert consensus, pathology, or outcomes data, as these relate to diagnostic accuracy studies.

    8. The Sample Size for the Training Set

    • Not applicable. This device is hardware; it does not involve training data or machine learning algorithms.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable. No training set exists for this device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K233119

    Validate with FDA (Live)

    Date Cleared
    2024-12-28

    (458 days)

    Product Code
    Regulation Number
    864.3700
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    8MP Color LCD Displays C811W, C811WT, PA27 and PA27T are intended for in vitro diagnostic use to display digital images of histopathology slides acquired from IVD-labeled whole-slide imaging scanners and viewed using IVD-labeled digital pathology image viewing software that have been validated for use with this device. They are an aid to the pathologist to review and interpret digital images of histopathology slides for primary diagnosis. It is the responsibility of the pathologist to employ appropriate procedures and safeguards to assure the validity of the interpretation of images using 8MP Color LCD Displays C811W, C811WT, PA27 and PA27T. The displays are not intended for use with digital images from frozen section, cytology, or non-formalin-fixed, paraffin embedded (non-FFPE) hematopathology specimens.

    Device Description

    The C811WT. PA27T. C811W. PA27 are 8MP Color LCD Displays, specifically intended for review and interpretation of surgical pathology slides from IVD-labeled whole-slide imaging scanners that have been validated for use with the display.

    The displays are equipped with a 27-inch color LCD panel with a fine pixel pitch. They use the latest generation of LED backlight panels. The built-in brightness stabilization control circuits make sure the brightness of these displays is stable in their life, so the products meet the demands of high precision medical imaging.

    For C811WT. PA27T. C811W. PA27. the only difference is the capacitive touch screen.

    C811WT, PA27T have the same capacitive touch screen. C811W, PA27 don't have a capacitive touch screen.

    The difference between C811WT and PA27T is only different customers which results in the two models. The difference between C811W and PA27 is only different customers which results in the two models.

    AI/ML Overview

    The provided text describes the acceptance criteria and a study proving the device meets these criteria for the Shenzhen Beacon Display Technology Co., Ltd.'s 8MP Color LCD Displays (C811W, C811WT, PA27, PA27T).

    Here's the breakdown of the information requested:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implicitly derived from the "Test" column, and the reported device performance is presented in the "C811W", "C811WT", "PA27", and "PA27T" columns. The predicate device (Barco MDPC-8127) is also listed for comparison of technological characteristics.

    Acceptance Criteria (Derived from "Test")Reported Device Performance (Example for C811W)
    User Controls (Luminance, White Point, Color Space, Warm-up time)Luminance target, Maximum: 500 cd/m2; Display function: sRGB; White point: D65 (6500K); Color space: sRGB; 30 minutes of warm-up time
    Spatial resolution (MTF at Nyquist frequency)Vertical and Horizontal MTFs are 0.861 and 0.862 at Nyquist frequency
    Pixel defects (count and map)Total number of bright and dark pixels <= 5
    Artifacts (Deviation)Deviation <0.54%
    Temporal response (Tmax, Tmin)Tmax: 20.87ms, Tmin: 9.88ms
    Maximum, minimum, and calibrated luminanceMaximum: 523.6 cd/m², Minimum: 0.501 cd/m²
    Grayscale (Maximum deviation)Maximum deviation: 5.94 %
    Luminance uniformity and Mura test (Non-uniformity)11.16% non-uniformity on 100% video level
    Stability of luminance and chromaticity response with temperature and lifetime (Deviation from target luminance, Variations for chromaticity)Deviation from target luminance (500 cd/m²): 0.48%; Variations for chromaticity: < 0.55%
    Bidirectional reflection function (Specular and Diffuse reflection coefficients)Specular reflection coefficient: 4.4%, Diffuse reflection coefficient: 0.92%
    Gray Tracking (Δu'v')0.0006 Δu'v'
    Color scale (Maximum color error, Average color error)Maximum color error: <2ΔE
    Color gamut volume (2D color gamut overlapped with sRGB)2D color gamut overlapped with sRGB: 99.5%

    2. Sample size used for the test set and the data provenance

    The document does not explicitly state a sample size in terms of a number of distinct display units tested within each model group (C811W, C811WT, PA27, PA27T). The results are presented as single values for each model, implying that these are representative measurements from the tested devices.

    The data provenance is from bench testing conducted on the newly submitted devices and compared against the specifications of a predicate device. Regarding country of origin, the manufacturer is Shenzhen, Guangdong, China. The testing itself would likely have been performed in a lab setting, possibly in China or by a certified testing facility collaborating with the manufacturer. The document does not specify if the data is retrospective or prospective, but as it's for a 510(k) submission, it would be considered prospective with respect to the regulatory approval process.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable as the study described is a technical performance assessment of display characteristics, not a clinical study involving human interpretation of medical images or ground truth established by medical experts. The "ground truth" for these tests are objective, measurable physical properties of the displays, governed by established test methods and standards (e.g., IDMS 1.03).

    4. Adjudication method for the test set

    This information is not applicable for the same reasons as #3. Adjudication methods (e.g., 2+1, 3+1) are typically used in clinical studies where multiple human readers assess cases, and a process is needed to resolve discrepancies. This document describes objective bench testing.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, an MRMC comparative effectiveness study was not done. This study focuses on the technical performance of a medical display, not a clinical diagnostic aid involving AI or human reader performance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable. The device is an LCD display, not an algorithm. Its "performance" is its ability to accurately render visual information, which is evaluated through the technical tests detailed.

    7. The type of ground truth used

    The ground truth used for this technical performance assessment is based on objective measurements of physical display characteristics as defined by industry standards, specifically the "TPA Guidance" and "IDMS 1.03" (International Display Measurement Standard). Examples include:

    • Pre-defined target luminance (e.g., 500 cd/m²)
    • Reference sRGB color space
    • Calibrated measurement equipment to determine values like MTF, pixel defect counts, deviation percentages, and color errors against established benchmarks.

    8. The sample size for the training set

    This information is not applicable. The device is a hardware display, not a machine learning model that requires a training set.

    9. How the ground truth for the training set was established

    This information is not applicable for the same reason as #8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K172922

    Validate with FDA (Live)

    Device Name
    MMPC-4127F1
    Manufacturer
    Date Cleared
    2017-12-21

    (87 days)

    Product Code
    Regulation Number
    864.3700
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Barco MMPC-4127F1 (PP27QHD) device is intended for in vitro diagnostic use to display digital images of histopathology slides acquired from IVD-labeled whole-slide imaging scanners that have been validated for use with this device, for review and interpretation by pathologists. The display is not intended for use with digital images from frozen section, cytology, or non-formalin-fixed, paraffin embedded (non-FFPE) hematopathology specimens.

    Device Description

    The MMPC-4127F1, also marketed as the PP27QHD, is a stabilized, color-calibrated display system designed to review and interpret digital images of surgical pathology slides.

    The display is equipped with a 27-inch color LCD panel with a fine pixel pitch that can be calibrated to the sRGB gamut.

    Accessories:

    • QAWeb quality assurance software
    • DisplayPort cable
    • USB cable
    • AC power cord cables
    AI/ML Overview

    The provided text describes a 510(k) submission for a medical device, the Barco MMPC-4127F1 (PP27QHD) display, which is intended for in vitro diagnostic use to display digital images of histopathology slides. The submission aims to demonstrate substantial equivalence to a predicate device.

    The study presented is not for an AI/algorithm-only device, but for a display system. Therefore, many of the requested fields related to AI/algorithm performance (e.g., effect size, standalone performance, training set details, adjudication methods, multi-reader multi-case studies) are not applicable to this type of device. The acceptance criteria and performance testing focus on the display's technical characteristics and visual performance as they compare to the predicate device.

    Here's an analysis based on the provided document:

    1. A table of acceptance criteria and the reported device performance

    The document doesn't explicitly state "acceptance criteria" in a pass/fail quantifiable manner for each parameter, but rather compares the new device's specifications directly to the predicate device's specifications. The implicit acceptance criterion is that the new device's performance should be similar to or better than the predicate device without introducing new issues of safety or performance. The "reported device performance" is essentially the listed "New Device MMPC-4127F1 (PP27QHD)" specifications.

    ParameterNew Device MMPC-4127F1 (PP27QHD)Predicate Device MMPC-4127 (PP27QHDCR)Implicit Acceptance Criteria (based on comparison)
    Display Optical Performance
    Display technologyIPS LCDIPS LCDSame
    Native Resolution2560 x14402560 x1440Same
    Display LUT≥ 10 bit / sub-pixel≥ 10 bit / sub-pixelSame or equivalent
    Pixel Pitch (Horizontal) (mm)0.23310.2331Same
    Pixel pitch (Vertical) (mm)0.23310.2331Same
    Active screen (Diagonal) (mm / inch)685.8 / 27.0685.8 / 27.0Same
    Active screen (Width) (mm)596.74596.74Same
    Active screen (Height) (mm)335.66335.66Same
    Contrast Ratio - (typical)1000:11000:1Same or equivalent
    Luminance stabilization+2% / -2% front sensor+2% / -2% backlight sensorSimilar (different sensor type, but same stability range)
    Maximum Luminance (typical) (cd/m²)500500Same or equivalent
    DICOM Calibrated Luminance (Typical) (cd/m²)350350Same or equivalent
    Luminance non-uniformity (per DIN6868-157), diagnostic room class<25% on 9 points, 10%/80% luminance<25% on 9 points, 10%/80% luminanceSame or equivalent
    Response Time (typical) (ms)1212Same or equivalent
    Response Time (max) (ms)2424Same or equivalent
    Calibration Performance
    Supported color spacessRGB, DICOM, nativesRGBNew device supports more color spaces without affecting safety/effectiveness
    Average dE2000 deviation from sRGB<2 on 6x6x6 grid<2 on 6x6x6 gridSame or equivalent
    Maximum dE2000 deviation from sRGB<5 on 6x6x6 grid<5 on 6x6x6 gridSame or equivalent
    White point chromaticity stabilization+/-0.01 u'v'+/-0.01 u'v'Same or equivalent
    Gray tracking (per IEC 62563)+/-0.01 u'v'+/-0.01 u'v'Same or equivalent
    Grayscale transfer function deviation<10% on 18 points<10% on 18 pointsSame or equivalent
    Calibration SoftwareMediCal QAWebNucleus SoftwareDifferent software, but performance judged similar
    Environmental Specifications
    Operational Temperature (min/max) (°C)10 / 4010 / 40Same
    Within specification temperature (min/max) (°C)10 / 3510 / 35Same
    Power Consumption (@ 100-240VAC typical) (W)12085New device has higher power consumption, deemed not to affect safety/effectiveness

    The conclusion states: "The tests showed that the device has similar characteristics compared to the predicate device and did not reveal new issues of safety and performance." This broadly represents the "acceptance criteria" and "reported device performance" for the studied parameters.

    2. Sample size used for the test set and the data provenance

    This is a technical device (display) study, not a clinical study on diagnostic accuracy with patient data. The "test set" refers to the device itself and its components undergoing various bench tests. There is no mention of "sample size" in terms of number of patient cases or images, as the testing focuses on the display's physical and technical performance.

    • Test Set: The device itself (Barco MMPC-4127F1).
    • Data Provenance: The tests are bench tests performed on the display device, not on clinical data. The submitter is Barco N.V. based in Kortrijk, Belgium.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. Ground truth for a display device's performance is established through objective measurements using specialized equipment and established standards (e.g., DIN6868-157, IEC 62563), not through expert consensus or clinical interpretation.

    4. Adjudication method for the test set

    Not applicable. This is not a study involving human interpretation of images or adjudication of discrepant readings.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a display device, not an AI or algorithm. There were no human reader studies described.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a display device, not an AI or algorithm.

    7. The type of ground truth used

    The "ground truth" for this type of device effectively refers to the measurement standards and expected performance specifications for a medical display. These are objective measurements of optical and electrical properties, such as luminance, contrast, resolution, color accuracy, and temperature stability, validated against industry standards. It is not expert consensus, pathology, or outcomes data related to patient diagnosis.

    8. The sample size for the training set

    Not applicable. This is a display device, not an AI or algorithm. Therefore, there is no "training set" in the context of machine learning.

    9. How the ground truth for the training set was established

    Not applicable. As there is no training set for an AI/algorithm, this question is not relevant.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1