Found 34 results

510(k) Data Aggregation

    K Number
    K251585

    Validate with FDA (Live)

    Date Cleared
    2025-07-18

    (56 days)

    Product Code
    Regulation Number
    876.5980
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is intended for use as a dispenser, a measuring device and an oral fluid transfer device. It is intended to be used to deliver fluids into the body orally or enterally. It is intended to be used in clinical or home care settings by operators ranging from qualified medical practitioners to laypersons (under the supervision of a qualified medical practitioners) in all age groups.

    Device Description

    Profoject™ Enteral Feeding Syringe provided in a variety of sizes from 0.5mL to 60mL. It consists of plunger, plunger stopper, barrel, and used to deliver fluids into the body orally or connected to an enteral access device with male ENFit connector.

    The proposed syringe is sterile or non-sterile. Sterile device was sterilized by Ethylene Oxide Gas to achieve a SAL of 10⁻⁶ and supplied sterility maintenance package. The shelf life of both sterile and non-sterile proposed devices has been validated as 5 years.

    AI/ML Overview

    This looks like a 510(k) clearance letter for a medical device that is NOT an AI/ML device. Therefore, the questions related to AI/ML device evaluations (such as test set, ground truth, expert qualifications, MRMC studies, and training set) are not applicable.

    The document describes the non-clinical test conclusions for the Profoject™ Enteral Feeding Syringe.

    Here's a breakdown of the requested information based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document states that the device was evaluated against several ISO and ASTM standards, as well as USP-NF monographs. The acceptance criteria are implicitly those defined by these standards, and the reported performance is that the device "Complied with" these standards.

    Acceptance Criteria (Implicit from Standards)Reported Device Performance
    Compliance with ISO 20695 (Enteral feeding systems - Design and testing)Complied with: ISO 20695
    Compliance with ISO 7886-1 (Sterile hypodermic syringes for single use - Part 1)Complied with: ISO 7886-1
    Compliance with ISO 80369-1 (Small-bore connectors for liquids and gases - Part 1)Complied with: ISO 80369-1
    Compliance with ISO 80369-3 (Connectors for enteral applications)Complied with: ISO 80369-3
    Compliance with ISO 80369-20 (Common test methods for small-bore connectors)Complied with: ISO 80369-20
    Compliance with ISO 11737-1 (Microbiological methods - Population determination)Complied with: ISO 11737-1
    Compliance with USP - NF <62> (Microbiological Examination of Nonsterile Products)Complied with: USP - NF <62>
    Compliance with ASTM F1886/F1886M-16 (Seal integrity of flexible packaging)Complied with: ASTM F1886/F1886M-16
    Compliance with ASTM F1140/F1140M-13 (Internal pressurization failure resistance)Complied with: ASTM F1140/F1140M-13
    Compliance with ASTM F88/F88M-23 (Seal Strength of flexible barrier materials)Complied with: ASTM F88/F88M-23
    Compliance with ASTM F1929-23 (Detecting seal leaks by dye penetration)Complied with: ASTM F1929-23
    Compliance with USP - NF <71> (Sterility Tests)Complied with: USP - NF <71>
    Compliance with ISO 10993-7 (Ethylene oxide sterilization residuals)Complied with: ISO 10993-7
    Compliance with ISO 11135 (Ethylene oxide sterilization processes)Complied with: ISO 11135
    Compliance with ISTA 3A: 2018 (Packaged-Products for Parcel Delivery System Shipment)Complied with: ISTA 3A: 2018
    Compliance with Safety Considerations to Mitigate the Risks of Misconnections with Small-bore Connectors Intended for Enteral Applications - Guidance for Industry and FDA StaffComplied with: (Implicitly, as it's listed under performance testing)
    Dose accuracy: ±10% when filled with a minimum of 20% of overall syringe capacityDose accuracy testing conducted, results implicitly met criteria
    Biocompatibility: No Cytotoxicity (per ISO 10993-5)No Cytotoxicity
    Biocompatibility: No Sensitivity (per ISO 10993-10)No Sensitization
    Biocompatibility: No Irritation (per ISO 10993-23)No Irritation
    Sterility Assurance Level (SAL) of 10⁻⁶ for sterile devicesEO Sterilization to SAL=10⁻⁶
    Shelf life validationValidated as 5 years

    2. Sample size used for the test set and the data provenance

    The document indicates that non-clinical verification was performed. This typically involves laboratory testing on device prototypes or production samples. The specific sample sizes for each test are not provided in this summary. Data provenance would be from laboratory testing conducted by or for CMT Health PTE. Ltd. (Singapore), likely in a controlled, prospective manner as part of the device development and validation process.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This question is not applicable as the device is a physical medical device (enteral feeding syringe), not an AI/ML device requiring expert-established ground truth from medical images or clinical data. The "ground truth" for non-clinical performance and safety relates to adherence to recognized standards and specifications, which are typically evaluated by engineers and quality assurance professionals, not clinical experts for ground truth establishment.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies or for establishing ground truth in diagnostic accuracy studies for AI/ML algorithms. This document describes non-clinical performance and biocompatibility testing.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI/ML device, and no MRMC study was conducted or is relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI/ML device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For non-clinical testing of a physical device, the "ground truth" is defined by the technical specifications, performance limits, and safety requirements outlined in the referenced ISO, ASTM, and USP standards. For example, for dose accuracy, the ground truth is the specified ±10% accuracy. For sterility, it's the SAL of 10⁻⁶. For biocompatibility, it's the absence of cytotoxicity, irritation, or sensitization as defined by the ISO 10993 series. These are objective, measurable criteria, not subjective expert consensus or clinical outcomes data in the usual sense.

    8. The sample size for the training set

    Not applicable. This is not an AI/ML device.

    9. How the ground truth for the training set was established

    Not applicable. This is not an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Monoject™ Enteral Syringe with ENFit Connector is indicated for use as a dispenser, a measuring device, and a fluid transfer device. It is used to deliver fluids, either orally or enterally, into the gastrointestinal system of a patient. The orallenteral syringes are intended to be used in clinical or home care settings by users ranging from laypersons (under the supervision of a clinician) to clinicians, in all age groups.

    Device Description

    Not Found

    AI/ML Overview

    This document is an FDA 510(k) clearance letter for the Monoject™ Enteral Syringe with ENFit Connector. It does not contain any information regarding acceptance criteria or a study proving that a device meets those criteria for software or AI/ML-enabled devices.

    The letter is a regulatory document indicating that the FDA has determined the device to be substantially equivalent to a legally marketed predicate device. This process primarily focuses on comparing the new device's intended use, technological characteristics, and safety and effectiveness to an existing, legally marketed device.

    Therefore, I cannot provide the requested information from this document. The document discusses regulatory aspects like:

    • Trade/Device Name
    • Regulation Number and Name
    • Regulatory Class
    • Product Code
    • Indications for Use
    • General controls provisions of the Act (e.g., annual registration, listing of devices, good manufacturing practice, labeling, misbranding, adulteration)
    • Quality System (QS) regulation requirements
    • Unique Device Identification System (UDI Rule)
    • Medical Device Reporting (MDR)
    • Contact information for the FDA.

    However, none of these sections detail acceptance criteria, performance studies, sample sizes, ground truth establishment, or expert qualifications as typically provided for AI/ML device submissions.

    Ask a Question

    Ask a specific question about this device

    K Number
    K231300

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-10-06

    (155 days)

    Product Code
    Regulation Number
    876.5980
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    NUTRICAIR™ enteral feeding syringes with ENFit® are disposable, sterile medical devices that are intended for the measurement, the preparation and enteral administration of solutions (hydration, feeding) or medications for neonatal, pediatric and adult patient populations. These devices are intended to be used in clinics and home care settings by: - health care professionals - adult laypersons (instructed by/or under the supervision of health care professionals).

    Device Description

    The Nutricair™ enteral syringe with ENFit is sterile, single use device. It is provided in size 10 ml. It consists in a body with ENfit connector in polypropylene, a plunger in polypropylene and a piston seal in isoprene. This device incorporates a female ENFit connection to enteral access device with a male ENFit connector that is compliant to ISO 80369-3. Graduations from 1 to 10 ml are printed on the body. The 10 mL syringe model number is NCE10SE.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device called "Nutricair Enteral Syringe with ENFit," which functions as an enteral feeding syringe. This type of submission relies on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving novel clinical effectiveness or conducting clinical trials.

    As such, the information you're requesting regarding acceptance criteria and studies demonstrating a device meets acceptance criteria, an AI-driven system, ground truth establishment, sample sizes for training/test sets, expert adjudication, MRMC studies, or standalone algorithm performance, is not applicable to this document.

    This submission focuses on engineering and performance criteria to show the new device performs as safely and effectively as an existing one, usually through non-clinical bench testing and comparative analysis of design features.

    Here's a breakdown of why your specific questions generally don't apply to a 510(k) for a physical medical device like a syringe, based on the provided text:

    • 1. A table of acceptance criteria and the reported device performance:

      • The document largely relies on meeting established international standards (e.g., ISO) for medical devices, particularly those related to syringes and enteral connectors. The "acceptance criteria" are these standard requirements, and the "reported device performance" is that the device met these standards.
      • The table on pages 6-7 ("Table 5-1: Device comparison table") highlights comparisons to the predicate device and states "Equivalent" or "Similar" and "No impact on safety or performance" as the outcome, indicating that existing standards and comparative performance are the basis for acceptance.
      • Section VIII, "Performance Testing," lists the types of tests performed (biocompatibility, visual inspections, enteral device performance tests, enteral connector performance tests, risk analysis, usability analysis) and the standards they comply with (e.g., ISO 10993, ISO 20695, ISO 7886-1, ISO 80369-20, ISO 80369-3, ISO 14971, ISO 62366-1). The acceptance criteria for each of these would be the specific quantitative or qualitative limits defined within those ISO standards. The document states that the device "met updated standard ISO 20695:2020" or "met the standards of 80369-3." However, specific numeric performance results are not typically included in these public 510(k) summaries unless a deviation from a standard is justified or an equivalence needs quantitative proof for a specific parameter.
    • 2. Sample sizes used for the test set and the data provenance:

      • For physical product testing as described (e.g., fluid leakage, stress cracking, barrel dimensions), sample sizes are typically defined by the test protocols within the relevant ISO standards (e.g., a certain number of syringes tested per batch). This information is generally not included in the 510(k) summary itself, as it's part of the detailed test reports provided to the FDA.
      • Data provenance for such physical device testing is typically the manufacturer's internal testing labs or third-party accredited labs. Country of origin for testing is not specified, but the manufacturer is based in France. The testing is retrospective in the sense that it was completed before the submission to demonstrate compliance.
    • 3. Number of experts used to establish the ground truth... and their qualifications:

      • This is not applicable. There is no "ground truth" in the diagnostic AI or clinical outcome sense for a basic physical device like a syringe. The "truth" is whether it meets engineering specifications and safety standards defined by regulatory bodies and consensus standards. These are established through engineering principles, laboratory testing, and risk assessment by qualified engineers and scientists.
    • 4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not applicable. This concept pertains to human review of diagnostic imaging or clinical findings for establishing ground truth, which is not relevant here.
    • 5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not applicable. This is not an AI/software device that would involve human readers or diagnostic interpretation.
    • 6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Not applicable. This is not an AI/software device.
    • 7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • As explained in #3, "ground truth" in the diagnostic sense is not applicable. The "truth" is whether the physical device meets its specified engineering performance parameters and safety requirements, as verified through standard laboratory tests and adherence to ISO standards.
    • 8. The sample size for the training set:

      • Not applicable. This device does not involve machine learning or a "training set."
    • 9. How the ground truth for the training set was established:

      • Not applicable. No training set is involved.

    In summary: The provided document is an FDA 510(k) clearance summary for a disposable medical syringe. The "acceptance criteria" are the relevant ISO standards for medical device safety and performance, and the "study" is a series of non-clinical, bench-top engineering tests demonstrating compliance with these standards and substantial equivalence to a predicate device. Concepts related to AI/ML performance, diagnostic ground truth, or multi-reader studies are entirely outside the scope of this particular device and its regulatory submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K222772

    Validate with FDA (Live)

    Date Cleared
    2023-08-17

    (337 days)

    Product Code
    Regulation Number
    876.5980
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is intended for use as a dispenser, a measuring device and an oral fluid transfer device. It is intended to be used to deliver fluids into the body orally or enterally. It is intended to be used in clinical or home care settings by operators ranging from qualified medical practitioners to laypersons (under the supervision of a qualified medical practitioners) in all age groups.

    Device Description

    The Oral/Enteral Syringe with ENFit connector is a disposable enteral feeding syringe provided in a variety of sizes from 0.5ml to 60ml. The Oral/Enteral Syringe with ENFit connector consists of plunger, gasket/piston, barrel with ENFit connector, and used to deliver fluids into the body orally or connected to an enteral access device with male ENFit connector. The proposed syringe is sterilized by Ethylene Oxide Gas to achieve a SAL of 10-6 and supplied sterility maintenance package which could maintain the sterility of the device during the shelf life of three years.

    AI/ML Overview

    This document is a 510(k) premarket notification for an Oral/Enteral Syringe with ENFit connector, seeking substantial equivalence to a predicate device. As such, it does not describe a study to prove a device meets acceptance criteria in the way a clinical trial or performance study for a novel diagnostic algorithm might. Instead, it demonstrates substantial equivalence by showing that the device meets relevant performance standards and has similar technological characteristics to a predicate device.

    Therefore, many of the requested categories (sample size for test set, data provenance, number of experts, adjudication method, MRMC study, standalone performance, ground truth for test/training set, training set sample size) are not applicable in the context of this 510(k) submission for a non-active medical device like a syringe.

    However, I can extract the acceptance criteria where applicable (i.e., performance standards and biocompatibility) and report the device's performance against those standards as described in the document.


    Acceptance Criteria and Device Performance (Oral/Enteral Syringe with ENFit connector)

    Device Under Review: Oral/Enteral Syringe with ENFit connector (K222772)
    Predicate Device: Oral/Enteral Syringe with ENFit connector (K211025)

    1. Table of Acceptance Criteria and Reported Device Performance

    Category / Test TypeAcceptance Criteria (Standard Reference)Reported Device Performance / Conclusion
    BiocompatibilityISO 10993-1:2018 (Surface medical device-mucosal membrane, contact < 24 hours)Biocompatibility was evaluated and "does not show any adverse effect." Specific tests:
    In Vitro Cytotoxicity TestISO 10993-5:2009No Cytotoxicity
    Intracutaneous Reactivity TestISO 10993-10:2010No Irritation
    Skin Sensitization TestISO 10993-10:2010No Sensitization
    SterilizationISO 11135 (Sterilization of health care products — Ethylene oxide)Sterilization method validated to ISO11135, routine control, and monitoring parameters determined. Device is EO Sterilized to SAL 10-6.
    EO ResidueISO 10993-7:2008Complied
    ECH ResidueISO 10993-7:2008Complied
    Shelf LifeASTM F1980-16 (Standard Guide for Accelerated Aging of Sterile Medical Device Packages)Shelf life of three years determined based on stability studies, including accelerated aging test.
    Package IntegrityASTM F88/F88M-15 (Seal strength)All packaging deemed acceptable for protection of product and sterility maintenance after environmental conditioning and simulated transportation.
    ASTM F1929-2015 (Blue Dye Penetration)All packaging deemed acceptable for protection of product and sterility maintenance.
    ASTM F1886/F1886M-16 (Seal Integrity (Visual Inspection))All packaging deemed acceptable for protection of product and sterility maintenance.
    Product PerformanceASTM D638-22, ASTM D790-17 Procedure A (Mechanical properties of plastics)Report of PP 5090T provided, implying compliance with these material property standards.
    Dose AccuracyISO 7886-1:2017 (Sterile hypodermic syringes for single use Part 1: Syringes for manual use)Testing performed, implying compliance with dose accuracy requirements.
    Enteral ConnectorsISO 80369-3:2016 (Small-bore connectors for liquids and gases in healthcare applications — Part 3: Connectors for enteral applications)Performance testing performed, implying compliance with enteral connector standards. The Gasket difference from the predicate was determined not to raise new safety/effectiveness issues as the device "passed the verification of ISO 80369-3, ISO 80369-20 and ISO 7886-1 standard tests, and the results meet the requirements."
    General ConnectorsISO 80369-1:2018 (Small-bore connectors for liquids and gases in healthcare applications - Part 1: General requirements)Performance testing performed, implying compliance.
    Common Test MethodsISO 80369-20:2015 (Small-bore connectors for liquids and gases in healthcare applications - Part 20: Common test methods)Performance testing performed, implying compliance.

    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not explicitly stated for each test, but standard test methods were followed. The number of devices required for each standard (e.g., ISO 7886-1, ISO 80369-3) would dictate the sample sizes.
    • Data Provenance: The tests are non-clinical (bench testing, material testing, biocompatibility testing) performed on the manufactured device. The location of testing labs is not specified, but the applicant is based in China. The data is prospective for the purpose of this submission (i.e., new tests performed on the specific device for this submission).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not Applicable. This is a non-clinical device submission. "Ground truth" in the diagnostic/AI sense is not relevant. Performance is measured against established international and harmonized standards.

    4. Adjudication method for the test set:

    • Not Applicable. This refers to clinical assessments or expert review, which is not part of this type of submission. Testing involves laboratory measurements against predefined criteria in standards.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not Applicable. This is a non-clinical syringe, not an AI-enabled diagnostic device.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Not Applicable. This is a non-clinical syringe, not an AI-enabled diagnostic device.

    7. The type of ground truth used:

    • Not Applicable. For this type of device, performance is evaluated against objective, measurable criteria defined in international scientific and engineering standards (e.g., ISO, ASTM). There isn't a "ground truth" derived from expert consensus, pathology, or outcomes data in the context of a diagnostic product.

    8. The sample size for the training set:

    • Not Applicable. This is a non-AI device submission; there is no "training set."

    9. How the ground truth for the training set was established:

    • Not Applicable. This is a non-AI device submission; there is no "training set" or ground truth for it.

    Summary of Study (Based on Submitted Information):

    The "study" in this context is a collection of non-clinical performance evaluations and testing conducted to demonstrate that the Anhui Tiankang Medical Technology Co., Ltd. Oral/Enteral Syringe with ENFit connector is substantially equivalent to its predicate device. This involves:

    • Biocompatibility Testing: Evaluating the device's materials for cytotoxicity, irritation, and sensitization according to ISO 10993 series. The device reported "No Cytotoxicity; No Irritation; No Sensitization."
    • Sterilization Validation: Confirming the Ethylene Oxide (EO) sterilization process meets ISO 11135 standards and an SAL of 10-6, along with EO and ECH residue testing per ISO 10993-7. The device was found to comply.
    • Shelf Life Determination: Performing accelerated aging tests according to ASTM F1980-16 to support a 3-year shelf life.
    • Package Integrity Testing: Conducting tests like seal strength (ASTM F88/F88M-15), blue dye penetration (ASTM F1929-2015), and visual inspection (ASTM F1886/F1886M-16) to ensure packaging maintains sterility. All packaging was deemed acceptable.
    • Product Performance Testing: Evaluating the syringe against specific functional standards:
      • Dose accuracy according to ISO 7886-1:2017.
      • Enteral connector compatibility and performance according to ISO 80369-3:2016.
      • General connector requirements and common test methods from ISO 80369-1:2018 and ISO 80369-20:2015.
      • Material properties referenced by ASTM D638-22 and ASTM D790-17.

    The submission concludes that the non-clinical testing demonstrates the device's safety, effectiveness, and performance are comparable to the legally marketed predicate device, thereby establishing substantial equivalence. No clinical studies were included or deemed necessary for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K222155

    Validate with FDA (Live)

    Date Cleared
    2022-09-16

    (58 days)

    Product Code
    Regulation Number
    876.5980
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Single use O-ring gasket syringe with ENFit connector (provided sterile and non-sterile):
    The device is indicated for use as a dispenser, a measuring device and a fluid transfer device. It is used to deliver fluids into the body orally or enterally. It is intended to be used in clinical settings by users ranging from clinicians to laypersons in all age groups.

    Reusable O-ring gasket syringe with ENFit connector-(provided non-sterile):
    The device is indicated for use as a dispenser, a measuring device. It is used to deliver fluids into the body orally or enterally. It is intended to be used multiple times in non-clinical settings by users ranging from clinicians to laypersons in all age groups. The device is indicated for single patient use only.

    Device Description

    The proposed device is manual feeding syringe, used as a dispenser, a measuring device, and a fluid transfer device. And it is used to deliver fluids into the body orally or enterally in clinical and non-clinical settings.

    There are 2 types of connector for proposed device: Low dose tip ENFit connector(1ml to 6ml) and Standard ENFit connector(12ml to 60ml). The proposed device is standard piston syringe consisting of a plastic hollow barrel with female ENFit connector of ISO 80369-3, plunger, 0-ring piston, Connector of 1ml and 3ml syringe situated centrally, connector of other syringe situated eccentrically. All connectors are compatible only with enteral access or accessories having male ENFit connector of ISO 80369-3. The proposed syringes are available in transparent barrel and amber barrel.

    The proposed device is provided sterile or non-sterile, single use or reusable. The sterile device is sterilized by Ethylene Oxide Gas (EtO) to achieve a SAL of 10-6 and supplied sterility maintenance package which could maintain the sterility of the device during the shelf life. The reusable device is for single patient use only.

    The proposed device is made with polypropylene and Silica gel. No DEHP, BPA and Natural Rubber Latex are added in the proposed device. The proposed device also has an optional syringe tip cap.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification for a medical device (O-ring gasket syringe with ENFit connector). This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than proving a device meets specific acceptance criteria through a large-scale clinical study or AI performance evaluation.

    Therefore, the information typically requested in your prompt regarding AI/algorithm performance, ground truth establishment for training and test sets, multi-reader multi-case (MRMC) studies, expert opinions, and sample sizes for AI model development is not present in this document.

    The document demonstrates compliance with performance standards for a physical medical device. Here's what can be extracted and inferred based on the document's content:

    1. A table of acceptance criteria and the reported device performance:

    The document doesn't present a table of "acceptance criteria" in a quantitative sense as might be found for algorithm performance (e.g., minimum accuracy). Instead, it lists the standards the device was tested against and states that the test results "demonstrated that the proposed device complies with the following standards." Compliance with these standards is the acceptance criterion for a physical device in this context.

    Acceptance Criteria (Standards Complied With)Reported Device Performance
    ISO 10993-7:2008 Biological evaluation of medical devices Part 7: Ethylene oxide sterilization residuals, AMENDMENT 1: Applicability of allowable limits for neonates and infants (2019)Complies
    ISO 10993-5:2009 Biological evaluation of medical devices- Part 5: Test for in vitro cytotoxicityComplies
    ISO 10993-6:2016 Biological evaluation of medical devices -- Part 6: Tests for local effects after implantationComplies
    ISO 10993-10:2010 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitizationComplies
    ISO 10993-11:2017 Biological evaluation of medical devices -- Part 11: Tests for systemic toxicityComplies
    USP-NF <151> Pyrogen TestComplies
    ASTM F1929-15 Standard test method for detecting seal leaks in porous medical packaging by dye penetration.Complies
    ASTM F88/F88M-21 standard method for seal strength of flexible barrier materialsComplies
    ASTM F1886/F1886M-16 Standard Test Method for Determining Integrity of Seals for Flexible Packaging by Visual InspectionComplies
    USP-NF <85> Bacterial Endotoxins TestComplies
    ISO 7886-1:2017 Sterile hypodermic syringes for single use - Part 1: Syringes for manual useComplies
    ISO 80369-3:2016 Small-bore connectors for liquids and gases in healthcare applications - Part 3: Connectors for enteral applicationsComplies
    ISO 80369-20:2015 Small-bore connectors for liquids and gases in healthcare applications - Part 20: Common test methodsComplies
    Microbial Enumeration and Specified Microorganisms Tests reportComplies
    Cleaning instructions verificationComplies
    Re-use cycle parameters studyComplies
    Microbial study in simulation usingComplies
    Risk assessment in accordance to ISO 14971Complies
    Dosing Accuracy TestingComplies
    Additional Acceptance Criteria for Reusable Devices (implied):
    Ability of immediate package to maintain sterility for 5 yearsValidated (in accordance with ASTM F1886/F1886M-16, ASTM F88/F88M-21, ASTM F1929-15)
    Performance after 5 years accelerated agingComplies with ISO7886-1 and ISO 80369-3

    2. Sample sizes used for the test set and the data provenance:

    • Sample Size: Not explicitly stated for each test. These would be determined by the specific requirements of each ISO/ASTM standard (e.g., how many units are tested for seal integrity or biocompatibility).
    • Data Provenance: The tests were "Non-clinical tests" conducted to verify compliance with international standards. The country of origin for the testing itself is not specified, but the applicant (Jiangsu Caina Medical Co., Ltd.) is from China. These are laboratory-based compliance tests, not data collected from human subjects.
    • Retrospective or Prospective: Not applicable, as these are technical performance tests, not clinical data collection.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This is not applicable. The "ground truth" for the device's physical performance is established by the definitions and methodologies within the cited international standards (ISO, ASTM, USP-NF). The tests are designed to objectively measure properties and assess compliance according to these universally accepted procedures, not through expert consensus in the way a medical image diagnosis might be.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable. Adjudication methods are typically used for subjective assessments (like medical image interpretation). For compliance testing against engineering and biological standards, the results are typically quantitative measurements that either meet or fail a predefined specification.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This type of study (MRMC, human readers, AI assistance) is relevant for AI-powered diagnostic devices. The device in question is a physical syringe, not an AI diagnostic tool.
    • The document explicitly states: "No clinical study is included in this submission."

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • No. This is not an AI algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For this device, the "ground truth" for proving performance is adherence to established international and national engineering and biological standards. This involves:
      • Validated test methods: e.g., the specific procedures for pyrogen testing (USP-NF <151>), seal integrity (ASTM F88/F88M-21), or material biocompatibility (ISO 10993 series).
      • Defined acceptance limits/specifications: e.g., allowable residuals of ethylene oxide, specified ranges for dosing accuracy, or criteria for passing cytotoxicity tests.

    8. The sample size for the training set:

    • Not applicable. This is not an AI model requiring a training set.

    9. How the ground truth for the training set was established:

    • Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K211661

    Validate with FDA (Live)

    Date Cleared
    2022-02-28

    (272 days)

    Product Code
    Regulation Number
    876.5980
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ENFit Nutrifit™ enteral syringes deliver enteral fluids to the gastrointestinal system of a patient who is physically unable to eat and swallow. The syringes are intended to be used in clinical or home care settings by users ranging from laypersons (under the supervision of a clinician), to clinicians to administer enteral fluids. ENFit Nutrifit™ enteral syringes are single use devices. ENFit Nutrifit™ enteral syringes are intended for pediatric and adult use.

    Device Description

    An enteral syringe consists of a syringe barrel with integral tip (ENFit), plunger, gasket, barrel lubricant and can be supplied with or without a syringe tip cap. Only the 5mL single use syringe can be connected with compatible enteral syringe-driver pumps; the other syringes are for manual use.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification for ENFit enteral syringes, which are medical devices used for delivering enteral fluids. It outlines the regulatory process for clearance, not a study proving the device meets acceptance criteria for an AI/ML powered medical device. The document primarily focuses on establishing substantial equivalence to a predicate device through bench testing and conformance to established standards, rather than clinical studies with human participants.

    Therefore, I cannot provide the information requested in the prompt's structured format (acceptance criteria table, sample sizes, expert qualifications, adjudication methods, MRMC studies, standalone AI performance, ground truth types, training set details) because this document does not describe such a study for an AI/ML powered device.

    The document details the following about the device's testing and intended performance:

    • Device Type: ENFit enteral syringes (NUTRIFIT™)
    • Intended Use: Delivering enteral fluids to the gastrointestinal system of patients unable to eat and swallow. Intended for use in clinical or home care settings by laypersons (under supervision) and clinicians, for pediatric and adult use.
    • Regulatory Class: Class II (Product Code: PNR, Regulation Number: 21 CFR 876.5980 - Gastrointestinal tube and accessories).
    • Predicate Device: ENFit enteral pump syringes PENTA™/ENFit enteral pump syringes NUTRIFIT™ (K161141).
    • Testing and Evaluation:
      • Performance test: According to ISO 7886-1/-2 & ISO 80369-3/-20, including testing after simulated clinical use and cleaning.
      • Biocompatibility tests: According to ISO 10993 series requirements.
      • Sterility validation and tests: According to ISO 11135.
      • Packaging validation and tests: According to ISO 11607-1/-2.
      • Labeling requirements: According to ISO 15223-1 and FDA Guidance "Safety Considerations to Mitigate the Risks of Misconnections with Small-bore Connectors Intended for Enteral Applications."
    • Conclusion: The tests demonstrated that the ENFit enteral syringes NUTRIFIT™ are safe and effective, meet their intended use, and are similar to the predicate device in terms of intended use, indications for use, and medical technique.

    The document emphasizes compliance with international standards (ISO) and FDA guidance for medical devices, specifically for physical characteristics, sterility, biocompatibility, and packaging, rather than performance metrics related to diagnostic accuracy, image interpretation, or other tasks that would involve AI/ML algorithms and require the types of studies outlined in your request.

    Ask a Question

    Ask a specific question about this device

    K Number
    K211593

    Validate with FDA (Live)

    Date Cleared
    2021-09-30

    (129 days)

    Product Code
    Regulation Number
    876.5980
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Enteral Pump Syringe delivers nutritional formula to the gastrointestinal system of a patient who is physically unable to eat and swallow.

    Enteral Pump Syringe is intended to be used in clinical or home care settings by users ranging from laypersons (under the supervision of a clinician) to health care professionals, to administer nutritional formula. In particular, these syringes can be connected with enteral infusion pump.

    Enteral Pump Syringe is intended for neonatal, pediatric and adult use.

    Enteral Pump Syringe is single use device.

    Device Description

    The proposed device is modified device of K190502(ENFit Oral/Enteral Syringe,cleared July 2, 2019) according to 21CFR 820.30. The change is a operation mode change from manual use to use with infusion pump.

    The proposed device is a disposable enteral feeding syringe provided in a variety of sizes from 1ml to 60ml. The ENFit enteral feeding syringe consists of moveable plunger, piston, calibrated hollow barrel and tip cap(not necessary). At the distal end of the barrel is a female connector, designed according to ISO 80369-3 (often referred to as an ENFit connection). The tip is designed to mate only with enteral administration devices and is incompatible with Luer connectors. The device is used to deliver fluids into the body enterally by connected to an enteral access device with male ENFit connector. The syringes (size 6ml to 60ml) incorporate a female Standard ENFit connector; the syringes (size 1ml and 6ml) incorporate a female low dosing ENFit connector.

    There are 2 types of ENFit connector ENFit syringe and central connector ENFit syringe. The sizes of the Side connector ENFit syringe range from 12ml to 60ml; and The sizes of the central connector ENFit syringe range from 1ml to 60ml. The syringes are provided with transparent or amber color barrel.

    The material used of deviec are Polypropylene. Polyisoprene rubber, Polydimethysiloxane(inside surface coating of barrel ). Amber additive(for used amber color barrel only), which are also patient-contacting materials for the proposed devices. The time of contact duration is less than 24 hours. The contact level is mucosal membrane, indirect.

    The proposed device is sterile or non-sterile, single use. The sterilized by Ethylene Oxide Gas to achieve a SAL of 10-6 and supplied sterility maintenance package which could maintain the sterility of the device during the shelf life of five years.

    The proposed device intended to be used in clinical or home care settings by users ranging from laypersons (under the supervision of a clinician) to health care professionals. The device has no energy in itself. But the syringe works with infusion pump and extension sets.The syringe is aspirated nutritional formula liquid, connected the extension set. Then user installed it on the infusion pump, and pump fluids into a body enterally in a controlled manner for administer enteral nutrition. The accuracy of graduations for enteral syringes is +/- 4% of expelled volume.

    AI/ML Overview

    This document is a 510(k) Premarket Notification for an "Enteral Pump Syringe." It focuses on demonstrating substantial equivalence to a predicate device, rather than proving a medical device meets specific performance acceptance criteria through the kind of study described in the prompt. This device is a syringe, not an AI or imaging device, so the requested information regarding AI performance metrics, expert adjudication, MRMC studies, and ground truth establishment is not applicable.

    Therefore, I cannot provide the requested information for an AI/imaging device. However, I can extract the information relevant to the performance and testing of this specific medical device, the Enteral Pump Syringe, as presented in the document:

    Device: Enteral Pump Syringe

    Context: This document is an FDA 510(k) submission, aiming to demonstrate "substantial equivalence" of the proposed device to an already legally marketed predicate device. It is not a clinical study report for an AI or imaging device.


    Information that can be extracted from the provided text regarding this device's performance and testing:

    1. A table of acceptance criteria and the reported device performance:

    The document primarily refers to compliance with established international standards rather than specific quantitative acceptance criteria and reported performance values in a table format. The "acceptance criteria" here are the requirements of these standards.

    Standard / CharacteristicAcceptance Criteria (implied by standard compliance)Reported Device Performance (as stated in document)
    ISO 80369-3:2016 (Small-Bore Connectors for Enteral Applications)Compliance with specifications for enteral connectors"The proposed device complies with the following standards: ISO 80369-3:2016" and "These product models were tested according to ISO 80369-3."
    ISO 80369-20:2015 (Common Test Methods for Small-Bore Connectors)Compliance with common test methods"The proposed device complies with the following standards: ISO 80369-20:2015"
    ISO 7886-1:2017 (Syringes for Manual Use)Compliance with specifications for sterile hypodermic syringes for manual use"The proposed device complies with the following standards: ISO 7886-1:2017" and "These product models were tested according to ISO 7886-1."
    ISO 7886-2:2020 (Syringes for use with power-driven syringe pumps)Compliance with specifications for sterile hypodermic syringes for use with power-driven syringe pumps"The proposed device complies with the following standards: ISO 7886-2:2020" and "These product models were tested according to ISO 7886-2."
    ISO 10993-5:2009 (Biocompatibility - in Vitro Cytotoxicity)No cytotoxicity"Cytotoxicity: No Cytotoxicity"
    ISO 10993-10:2010 (Biocompatibility - Irritation and Skin Sensitization)No irritation, no sensitization"Skin Irritation: No Irritation," "Sensitization: No Sensitization"
    ISO 10993-7:2008 (Biocompatibility - Ethylene oxide sterilization residuals)Compliance with residuals limits"ISO 10993-7:2008" listed as a complied standard.
    ISO 14971:2007 (Medical Devices - Application of Risk Management)Application of risk management principles"ISO 14971:2007" listed as a complied standard.
    ASTM F88/F88M-15 (Seal Strength of Flexible Barrier Materials)Compliance with seal strength requirements"ASTM F88/F88M-15" listed as a complied standard.
    ASTM F1929-15 (Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration)Compliance with seal leak detection"ASTM F1929-15" listed as a complied standard.
    ASTM F1886F1886M-16 (Determining Integrity of Seals for Flexible Packaging by Visual Inspection)Compliance with visual inspection for seal integrity"ASTM F1886F1886M-16" listed as a complied standard.
    Graduation Accuracy (for enteral syringes)+/- 4% of expelled volume"The accuracy of graduations for enteral syringes is +/- 4% of expelled volume." (This is stated as a characteristic, implying it meets this.)
    SAL (Sterility Assurance Level)10^-6"EO Sterilized to achieve a SAL of 10^-6"
    Shelf Life5 years"maintain the sterility of the device during the shelf life of five years."

    2. Sample sized used for the test set and the data provenance:

    • Sample Size: Not explicitly stated in the provided text. The document refers to "non-clinical tests" and "bench tests" to verify compliance with standards, but not the number of units tested.
    • Data Provenance: Not specified, but likely from the manufacturer's internal testing facilities in China (Jiangsu Caina Medical Co., Ltd.). The tests are "non-clinical" (bench tests) and therefore not from patients.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not applicable as this is a medical device (syringe) and not an AI or imaging device that requires expert-established ground truth from images or clinical data. The "ground truth" for this device's performance is adherence to engineering and biocompatibility standards measured through bench tests.

    4. Adjudication method for the test set:

    Not applicable. As above, this is not an AI/imaging device study requiring human reader adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This is not an AI-assisted device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Not applicable. This device does not involve an algorithm.

    7. The type of ground truth used:

    The "ground truth" for this device's performance is established by the international and national standards it is required to meet (e.g., ISO and ASTM standards for sterility, biocompatibility, connector compatibility, and physical performance). Compliance is shown through bench tests and material characterization, not expert consensus, pathology, or outcomes data in the traditional sense for diagnostic or prognostic devices.

    8. The sample size for the training set:

    Not applicable. This device does not involve machine learning or a "training set."

    9. How the ground truth for the training set was established:

    Not applicable. This device does not involve machine learning.

    Ask a Question

    Ask a specific question about this device

    K Number
    K203613

    Validate with FDA (Live)

    Date Cleared
    2021-08-27

    (260 days)

    Product Code
    Regulation Number
    876.5980
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ENFit® Reusable Enteral Syringe is indicated for use as a dispenser, a measuring device and a fluid transfer device. It is used to deliver fluids into the body orally or enterally. It is intended to be used multiple times in non-clinical settings by users ranging from clinicians to laypersons in all age groups. The device is indicated for single patient use only.

    Device Description

    The proposed device with Enfit® connector (1ml-60ml) are standard piston style syringes consisting of syringe barrel with integral EnFit syringe tip, syringe plunger and ring type piston. The proposed device are supplied non-sterile, and reusable. The sizes rang from 1ml to 60ml nominal capacity. The integral syringe tip is a female ENFit connector which is compatible only with enteral access devices or accessories having ENFit compliant or compatible male connectors to form a dedicated system that prevents wrong-route administration of fluids.

    AI/ML Overview

    This is a 510(k) summary for a medical device (ENFit Reusable Enteral Syringe). 510(k) submissions typically demonstrate substantial equivalence to a predicate device through non-clinical testing, rather than extensive clinical efficacy studies with acceptance criteria in the manner of AI/software devices. Therefore, the information requested in the prompt, especially related to AI-specific metrics, reader studies, and ground truth establishment, is not applicable or present in this document.

    However, I can extract the non-clinical tests performed and mention that these tests serve as the "acceptance criteria" for demonstrating substantial equivalence for this type of device.

    1. A table of acceptance criteria and the reported device performance

    The document lists various non-clinical tests performed to verify that the proposed device met design specifications and was substantially equivalent to the predicate device. These tests, and the implication that the device met their requirements, serve as the "reported device performance" and "acceptance criteria" for a 510(k) submission. No specific numerical thresholds are provided in this summary, but the successful completion of these tests indicates compliance.

    Acceptance Criteria (Non-Clinical Tests Conducted)Reported Device Performance
    A. Finished Device (Performance Testing per ISO 7886-2:2020)
    Risk management report in accordance with ISO 14971:2007Met (implied by conclusion of SE)
    Usability testing for ENFit Reusable Enteral Syringe (human factor and usability validation)Met (implied by conclusion of SE)
    AppearanceMet (implied by conclusion of SE)
    Pump ForceMet (implied by conclusion of SE)
    Short-term Flow Rate ErrorMet (implied by conclusion of SE)
    Syringe ComplianceMet (implied by conclusion of SE)
    Syringe Design critical dimensionMet (implied by conclusion of SE)
    B. Reusability
    Cleaning Instructions Validation and Use Cycle Parameters StudyMet (implied by conclusion of SE)
    C. Biocompatibility
    ISO 10993-5:2009 (Tests for Vitro Cytotoxicity)No Cytotoxicity (explicitly stated in comparison table, and implied by conclusion of SE)
    ISO 10993-10:2010 (Tests for irritation and skin sensitization)No Irritation and Sensitization (explicitly stated in comparison table, and implied by conclusion of SE)
    ISO 10993-11:2017 (Tests for systemic toxicity)No Acute Toxicity (explicitly stated in comparison table, and implied by conclusion of SE)
    D. Shelf Life
    Package verification test reportMet (implied by conclusion of SE)
    Bench test after real-time aging (ISO 80369-3:2016)
    Fluid LeakageMet (implied by conclusion of SE)
    Sub-atmospheric Pressure Air LeakageMet (implied by conclusion of SE)
    Stress CrackingMet (implied by conclusion of SE)
    Resistance to Separation from Axial LoadMet (implied by conclusion of SE)
    Resistance to Separation from UnscrewingMet (implied by conclusion of SE)
    Resistance to OverridingMet (implied by conclusion of SE)
    Disconnection by UnscrewingMet (implied by conclusion of SE)
    Bench test after real-time aging (ISO 7886-1:2017)
    AppearanceMet (implied by conclusion of SE)
    Overall Length of Scale to Nominal Capacity LineMet (implied by conclusion of SE)
    Push-button DistanceMet (implied by conclusion of SE)
    Force Required to Operate PlungerMet (implied by conclusion of SE)
    Freedom from Air and Liquid Leakage past PistonMet (implied by conclusion of SE)
    Limits for Acidity or AlkalinityMet (implied by conclusion of SE)
    Limits for Extractable MetalsMet (implied by conclusion of SE)
    Tolerance on Graduated CapacityMet (implied by conclusion of SE)
    Maximum Dead SpaceMet (implied by conclusion of SE)
    E. Finished Device Performance Test (not related to aging)
    Critical Dimension verificationMet (implied by conclusion of SE)
    Ink AdhesionMet (implied by conclusion of SE)
    Bench test (ISO 80369-3:2016)(These are repeated from the "after real-time aging" section, but without the "after real-time aging" qualifier, suggesting they are also tested in the non-aged state.)
    Fluid LeakageMet (implied by conclusion of SE)
    Sub-atmospheric Pressure Air LeakageMet (implied by conclusion of SE)
    Stress CrackingMet (implied by conclusion of SE)
    Resistance to Separation from Axial LoadMet (implied by conclusion of SE)
    Resistance to Separation from UnscrewingMet (implied by conclusion of SE)
    Resistance to OverridingMet (implied by conclusion of SE)
    Disconnection by UnscrewingMet (implied by conclusion of SE)
    Bench test (ISO 7886-1:2017)(These are repeated from the "after real-time aging" section, but without the "after real-time aging" qualifier, suggesting they are also tested in the non-aged state.)
    AppearanceMet (implied by conclusion of SE)
    Overall Length of Scale to Nominal Capacity LineMet (implied by conclusion of SE)
    Push-button DistanceMet (implied by conclusion of SE)
    Force Required to Operate PlungerMet (implied by conclusion of SE)
    Freedom from Air and Liquid Leakage past PistonMet (implied by conclusion of SE)
    Limits for Acidity or AlkalinityMet (implied by conclusion of SE)
    Limits for Extractable MetalsMet (implied by conclusion of SE)
    Tolerance on Graduated CapacityMet (implied by conclusion of SE)
    Maximum Dead SpaceMet (implied by conclusion of SE)

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not specify the sample sizes for each non-clinical test conducted. It states that "Non-clinical tests were conducted to verify that the proposed device met all design specifications." The data provenance (country of origin, retrospective/prospective) is not mentioned, as these are material performance tests rather than clinical data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This is a physical medical device (syringe), not an AI/software device that requires expert ground truth for classification or diagnosis. Therefore, this information is not applicable and not present in the document.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable, as this is not an AI/software device or a study involving human interpretation with adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a physical medical device. No MRMC study was conducted or reported.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a physical medical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For these physical device tests, the "ground truth" is defined by the established industry standards (e.g., ISO 14971, ISO 7886-2, ISO 10993 series, ISO 80369-3) and pre-defined specifications for device performance (e.g., freedom from leakage, specific dimensions, resistance to force). The device's performance against these verifiable physical and chemical metrics serves as the ground truth.

    8. The sample size for the training set

    Not applicable. This is a physical medical device, not an AI/software device requiring a training set.

    9. How the ground truth for the training set was established

    Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K211025

    Validate with FDA (Live)

    Date Cleared
    2021-06-30

    (85 days)

    Product Code
    Regulation Number
    876.5980
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is intended for use as a dispenser, a measuring device and an oral fluid transfer device. It is intended to be used to deliver fluids into the body orally or enterally. It is intended to be used in clinical or home care settings by operators ranging from qualified medical practitioners to laypersons (under the supervision of a qualified medical practitioners) in all age groups.

    Device Description

    The proposed device is a disposable enteral feeding syringe provided in a variety of sizes from 0.5ml to 60ml. It consists of plunger, piston, barrel and tip cap, and used to deliver fluids into the body orally or connected to an enteral access device with male ENFit connector. The syringes (size 12ml, 35ml and 60ml) incorporate a female Standard ENFit connector; the syringes (size 0.5ml, 1ml, 3ml, 6ml) incorporate a female low dosing ENFit connector. There are 2 types of the syringe: Side connector ENFit syringe and central connector ENFit syringe. The sizes of the central connector tapered syringe range from 0.5ml and 1ml; and side connector ENFit syringe range from 3ml to 60ml. The proposed syringe is sterile or non-sterile. Sterile device was sterilized by Ethylene Oxide Gas to achieve a SAL of 10th and supplied sterility maintenance package which could maintain the sterility of the device during the shelf life of three years.

    AI/ML Overview

    This document is a 510(k) Summary for the Oral/Enteral Syringe with ENFit connector (K211025). It does not contain information about a study proving the device meets acceptance criteria in the manner you describe for AI/imaging devices. Instead, it demonstrates substantial equivalence to a predicate device (K161039) through non-clinical testing.

    Here's an analysis of the information provided in the context of your questions, explaining why certain sections are not applicable:

    1. A table of acceptance criteria and the reported device performance

    The document provides a list of standards the device complies with, which implicitly serve as acceptance criteria for different aspects of the device's performance. It also states that "dose accuracy testing is conducted to demonstrate the enteral syringes are accurate to ±10% when the syringe is filled with a minimum dose of 20% of the overall syringe capacity." This is a key performance metric and a stated acceptance criterion.

    Acceptance Criteria (Standards and Specific Performance)Reported Device Performance
    ISO 7886-1:2017 (Sterile Hypodermic Syringes for Single Use-Part 1: Syringes for Manual Use)Complied
    ISO 10993-5:2009 (Biological Evaluation of Medical Devices-Part 5: Tests for in Vitro Cytotoxicity)Complied (No Cytotoxicity)
    ISO 10993-10:2010 (Biological Evaluation of Medical Devices-Part 10: Tests for Irritation and Skin Sensitization)Complied (No Irritation, No Sensitization)
    ISO 10993-7:2008 (Biological evaluation of medical devices- Part 7: Ethylene oxide sterilization residuals)Complied
    ISO 14971:2007 (Medical Devices-Application of Risk Management to Medical Devices)Complied
    ASTM F88/F88M-15 (Standard Test Method for Seal Strength of Flexible Barrier Materials)Complied
    ASTM F1929-15 (Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration)Complied
    ISO 80369-3:2016 (Small-Bore Connectors for Liquids and Gases in Healthcare Applications-Part 3: Connectors for Enteral Applications)Complied
    ISO 80369-20:2015 (Small-Bore Connectors for Liquids and Gases in Healthcare Applications-Part 20: Common Test Methods)Complied
    Dose accuracy: accurate to ±10% when filled with a minimum dose of 20% of overall syringe capacityDemonstrated (See "Non-Clinical Test Conclusion")
    Sterility Assurance Level (SAL)10^-6

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document mentions "non clinical tests were conducted," but does not specify the sample sizes for these tests for each standard or the dose accuracy testing. It also does not mention data provenance in terms of country of origin or retrospective/prospective nature, as these types of details are typically associated with clinical studies involving patient data, not non-clinical device testing.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This question is not applicable to this type of device submission. The "ground truth" concept, especially involving expert consensus and specific qualifications of medical professionals, is relevant for AI/imaging devices interpreting medical images or data. For this physical medical device (an enteral syringe), performance is assessed through objective engineering and biological tests against established international standards, not against expert human interpretations of data.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This question is not applicable. Adjudication methods like 2+1 or 3+1 are used in clinical trials or reader studies where multiple experts evaluate ambiguous cases. For non-clinical tests on a physical device, the results are typically objectively measured and do not require expert adjudication in this manner.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This question is not applicable. MRMC studies are specifically designed for assessing the performance of diagnostic devices, particularly AI-powered ones, and their impact on human reader performance. This submission is for a physical medical device (a syringe), not a diagnostic AI tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not applicable. This refers to the standalone performance of an algorithm. The device here is a physical syringe, not an algorithm.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    For this device, the "ground truth" is established by the requirements of the referenced international standards (e.g., ISO for sterility, biocompatibility, connector features, and ASTM for material properties) and specific performance specifications (e.g., dose accuracy ±10%). There is no "expert consensus," "pathology," or "outcomes data" in the typical sense for proving the performance of a syringe.

    8. The sample size for the training set

    This question is not applicable. This refers to the training data for an AI algorithm. This submission is for a physical medical device.

    9. How the ground truth for the training set was established

    This question is not applicable. This refers to establishing ground truth for AI model training. This submission is for a physical medical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K203410

    Validate with FDA (Live)

    Date Cleared
    2021-06-28

    (221 days)

    Product Code
    Regulation Number
    876.5980
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Wealy ENFit® Disposable Enteral Syringe is indicated for use as a dispenser, a measuring device, and a fluid transfer device. It is used to deliver fluids into the gastrointestinal system of a patient who is physically unable to eat and swallow. The enteral syringes are intended to be used in clinical or home care setting by users ranging from laypersons (under the supervision of a clinician) to clinicians, in all age groups.

    Device Description

    The proposed device is a disposable enteral feeding syringe provided in a variety of sizes from 1ml~100ml. This device incorporates a female ENFit® connector for connection to an enteral access device with male ENFit® connector. The proposed syringe is sterilized by Ethylene Oxide Gas to achieve a SAL of 10 ° and supplied sterility maintenance package which could maintain the sterility of the device during the shelf life of three years.

    AI/ML Overview

    The provided document is a 510(k) summary for the ENFit® Disposable Enteral Syringe. It focuses on demonstrating substantial equivalence to a predicate device rather than presenting a de novo study with specific acceptance criteria and detailed performance of an AI-powered diagnostic device. Therefore, many of the requested categories (expert ground truth, MRMC study, training set details) are not applicable as this is a medical device for fluid delivery, not an AI diagnostic.

    However, I can extract the relevant information regarding acceptance criteria and performance based on the non-clinical tests described.

    Here's a breakdown of the information that can be extracted from the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document lists various tests based on international standards (ISO, ASTM, USP). The reported performance is generally stated as "The testing demonstrates the proposed devices conform to the requirements" or "Conforms to the requirement of [standard]". Specific quantitative acceptance criteria are given by referencing clauses within these standards, implying that the device met those quantitative values.

    Acceptance Criteria Category (Referenced Standard/Clause)Reported Device Performance
    ISO 80369-3:2016 (Connectors for enteral applications) & ISO 80369-20:2015 (Common test methods):The testing demonstrates the proposed devices conform to the requirements of ISO80369-3:2016.
    Fluid Leakage (Annex B)Conforms
    Sub-atmospheric pressure Air Leakage (Annex D)Conforms
    Stress Cracking (Annex E)Conforms
    Resistance to separation from axial load (Annex F)Conforms
    Resistance to separation from unscrewing (Annex G)Conforms
    Resistance to overriding (Annex H)Conforms
    Disconnection by unscrewing (Annex I)Conforms
    ISO 7886-1:2017 (Sterile hypodermic syringe for single use):The testing demonstrates the proposed devices conform to the requirements of ISO 7886-1:2017.
    Appearance (Clause 5)Conforms
    Overall Length of Scale to Nominal Capacity Line (Clause 9.3)Conforms
    Push-button Distance (Clause 11)Conforms
    Force to Operate the Plunger (Clause 13.3)Conforms
    Freedom from air and liquid leakage past piston (Clause 13.2)Conforms
    Limits for Acidity or Alkalinity (Clause 6.2)Conforms
    Limits for Extractable Metals (Clause 6.3)Conforms
    Tolerance on graduated capacity (Clause 8)Conforms
    Dead Space (Clause 13.1)Conforms
    Lubricant (Clause 7)Conforms
    Cleanliness (Clause 5)Conforms
    Graduated Scale (Clause 9)Conforms (Implied, listed under ISO 7886-1:2017 tests)
    Barrel (Clause 11)Conforms (Implied, listed under ISO 7886-1:2017 tests)
    Piston/Plunger Assembly (Clause 13)Conforms (Implied, listed under ISO 7886-1:2017 tests)
    Biocompatibility Standards (ISO 10993 series):Conforms to the requirement of ISO 10993 series Standards. No Cytotoxicity, No Irritation to Skin, No significant evidence of sensitization. Biocompatibility testing has demonstrated the biological safety.
    ISO 10993-5:2009 (Vitro Cytotoxicity)Conforms
    ISO 10993-10:2010 (Irritation and skin sensitization)Conforms
    ISO 10993-7:2008 (Ethylene oxide sterilization residuals)Conforms
    ISO 10993-11:2017 (Systemic toxicity)Conforms
    Other Tests:
    ASTM F 88/F88M-09 (Seal strength of flexible barrier materials)Conforms
    USP38-NF33 <85> (Bacterial Endotoxins Limit)Conforms
    Product performance after Real-time Aging testEvaluated properties after aging in support of labeling.
    Package Integrity after Accelerated Aging testEvaluated properties after aging in support of labeling.
    Sterilization (Ethylene Oxide Gas)Achieves SAL of 10^-6^ (for 3-year shelf life)

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not explicitly state the sample sizes used for each specific test in the non-clinical performance testing. It also does not specify the country of origin of the data (beyond the manufacturer being in China) or whether it was retrospective or prospective, as these are non-clinical hardware tests.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. This device is an enteral syringe for fluid delivery, not a diagnostic device requiring expert interpretation for ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This is not a diagnostic device involving expert adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is not an AI diagnostic or image analysis tool, so an MRMC study is irrelevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device does not involve an algorithm or AI.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    For a medical device like a syringe, "ground truth" is established by adherence to recognized international performance standards (ISO, ASTM, USP). For example, the "ground truth" for fluid leakage would be the acceptable limits defined in ISO 80369-3, and the device passing means it met those objective, standardized measures. For biocompatibility, the ground truth is defined by the toxicological profiles and testing methods in the ISO 10993 series.

    8. The sample size for the training set

    Not applicable. This device does not use machine learning or AI, and therefore does not have a "training set."

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 4