510(k) Data Aggregation

    K Number
    K252402

    Validate with FDA (Live)

    Device Name
    Midline Catheter
    Date Cleared
    2025-12-18

    (140 days)

    Product Code
    Regulation Number
    880.5200
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Midline Catheter is indicated for insertion into a patient's vascular system for short-term use (<30 days) to sample blood or administer fluids intravenously. It may be used for any patient population with consideration given to adequacy of vascular anatomy and appropriateness of the procedure. The Midline Catheter is suitable for use with power injectors.

    Device Description

    The Midline Catheter is a peripherally inserted catheter composed of a radiopaque polyurethane catheter, a catheter slider, a catheter hub, needle hub, and 3D-Shin™ needle featuring an engineered three-dimensional geometry of the needle tip, with a passive safety mechanism. After observing the blood return, the 3D-Shin™ needle is withdrawn, and the needle tip is fully covered by the safety mechanism, reducing the risk of accidental needlestick injury.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K213855

    Validate with FDA (Live)

    Date Cleared
    2022-09-02

    (266 days)

    Product Code
    Regulation Number
    880.5200
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Arrow(R) Pressure Injectable Midline Catheter is indicated for short-term (≤ 30 days) peripheral access to the venous system for intravenous therapy, blood sampling, infusion, and pressure injection of contrast media. The maximum pressure of pressure injector equipment used with the Arrow Pressure Injectable Midline Catheter may not exceed 300 psi (2068.4 kPa). The maximum pressure injection flow rate for the specific lumen being used for pressure injection is printed on the extension line hub.

    Device Description

    The Arrow Pressure Injectable Midline Catheter is a non-coated, single use catheter designed to provide short-term peripheral access to the venous system. The midline catheter is a peripherally inserted intravenous catheter manufactured with medical grade, flexible polyurethane. It consists of a non-tapered, radiopaque polyurethane extruded catheter body with a single opening or double opening at the distal end and centimeter markings placed along its length to facilitate its positioning. The catheter is available in 4 Fr. Single lumen and 5 Fr. Double lumen configurations with a usable catheter length of 20 cm. Pinch clamps are an integral part of the catheter and are provided on the extension lines to occlude flow through the lumens, as needed. The catheters can be used for the injection of contrast media. The maximum recommended infusion rate is 5 mL/sec.

    The catheters will be packaged sterile in kits that will include components to facilitate insertion.

    AI/ML Overview

    The provided text is a 510(k) summary for a medical device (Arrow® Pressure Injectable Midline Catheter) and does not describe acceptance criteria for an AI/ML powered device or a study proving that an AI/ML device meets said criteria. It focuses on demonstrating substantial equivalence to a predicate device through non-clinical performance testing, biocompatibility testing, and sterilization/packaging testing.

    Therefore, most of the requested information regarding AI/ML device acceptance criteria, sample sizes for test/training sets, data provenance, expert ground truth, adjudication methods, MRMC studies, or standalone algorithm performance simply does not apply to this document.

    However, I can extract the information that is present about the device's characteristics and the types of non-clinical tests performed:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't provide specific numerical acceptance criteria (e.g., "Pass if flow rate >= X mL/s") or specific reported performance values resulting from the tests. Instead, it indicates that "Non-clinical testing related to the device changes has been completed to support the substantial equivalence." This implies that the device met the internal and standard-based requirements for these tests to demonstrate equivalence, but the precise thresholds and results are not detailed.

    Test CategoryStandard / DescriptionImplied Performance (Met)
    Luer Hub TestingBS EN ISO 80369-7: 2016 (Sub-atmospheric pressure air leakage, Stress Cracking/Fluid Leakage, Resistance to separation (axial load, unscrewing, overriding))Met requirements of standard
    LabelingBS EN ISO 10555-1: 2013 (Catheter Nominal Effective Length, Catheter outside diameter, Pressure Injection Gravity flow Rate, Visual Inspection, Surface Quality)Met requirements of standard
    Mechanical TestingBS EN ISO 10555-1: 2013 (Force at Break – Juncture Hub & Catheter Body, Force at Break - Luer Hub & Extension Line, Catheter Body Elongation, Liquid Leakage under Pressure, Air Leakage during Aspiration, Priming Volume Pump Flow Rate, Pressure Injection Flow rate & Repeat Injection, Static Burst under Pressure Injection)Met requirements of standard
    RadiopacityASTM F640-20Met requirements of standard
    Catheter Body KinkBS EN ISO 13868: 2002Met requirements of standard
    Collapse ResistanceInternal Test MethodMet internal requirements
    Blood DrawInternal Test MethodMet internal requirements
    Extension Line Clamp Closure EfficacyInternal Test MethodMet internal requirements
    Clamp ForceInternal Test MethodMet internal requirements
    BiocompatibilityBS EN ISO 10993-1: 2020 (Cytotoxicity, Sensitization, Irritation, Material Mediated Pyrogenicity, Acute/Subacute Systemic Toxicity, Hemocompatibility, Genotoxicity, Chemical Characterization)Met requirements of standard
    Sterilization EvaluationANSI/AAMI/ST72:2019 (Bacterial Endotoxin), BS EN ISO 10993-7: 2008 (Ethylene Oxide Residuals)Met requirements of standard
    PreconditioningISTA 3A, ASTM D4169 (Shipping, Distribution Simulation)Met requirements of standard
    Package Protection TestingBS EN ISO 11607-1: 2020Met requirements of standard

    2. Sample size used for the test set and the data provenance: Not applicable. This document describes non-clinical engineering and biological performance testing, not AI/ML model testing with patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for AI/ML models is not relevant here. The "ground truth" for these tests would be defined by the technical specifications of the standards (e.g., a specific pressure, flow rate, or biological response).

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI/ML device.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an AI/ML device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable to an AI/ML device. For this device, the "ground truth" for testing is established by recognized international and national standards (e.g., ISO, ASTM, ANSI/AAMI) and internal company test methods, which define acceptable parameters for physical and chemical properties, functionality, and biocompatibility.

    8. The sample size for the training set: Not applicable. This is not an AI/ML device.

    9. How the ground truth for the training set was established: Not applicable. This is not an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K213203

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-05-13

    (226 days)

    Product Code
    Regulation Number
    880.5200
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Provena™ Midline Catheters are indicated for short term access to the peripheral venous system for selected intravenous therapies, blood sampling, and power injection of contrast media. These catheters may be used for any patient population with consideration given to adequacy of vascular anatomy and appropriateness of the procedure. The Provena™ Midline Catheters are suitable for use with power injectors.

    Device Description

    Provena™ Midline Catheters are a family of peripherally placed catheters made from radiopaque body-softening polyurethane materials. Each Provena™ Midline Catheter is designed with kink-resistant, reverse taper design. Catheters are packaged in a tray with accessories for reliable short term (less than 30 days) vascular access. The Provena™ Midline Catheters are suitable for use with power-injectors.

    AI/ML Overview

    The provided FDA 510(k) summary for the Provena™ Midline Catheter does not describe a study involving an AI/Machine Learning device or an analysis of clinical images. Instead, it details a traditional medical device submission focused on demonstrating substantial equivalence to a predicate device through material, dimensional, and performance characteristic comparisons.

    Therefore, many of the requested details regarding acceptance criteria and study design for an AI/ML device (e.g., sample size for test set, number of experts, MRMC studies, ground truth establishment) are not applicable to the content provided.

    However, I can extract and present the device performance characteristics and the type of testing performed to demonstrate substantial equivalence, which serves as a form of "acceptance criteria" for a non-AI/ML device in this context.

    Here's a breakdown based on the provided document, addressing the applicable points:


    Device: Provena™ Midline Catheter (Non-AI/ML Device)

    1. A Table of Acceptance Criteria and the Reported Device Performance (as demonstrated by testing against the predicate):

    This table compares the characteristics of the subject device (Provena™ Midline Catheter) to the predicate device (PowerMidline™ Catheter) to demonstrate substantial equivalence, rather than defining specific statistical performance metrics against a clinical ground truth as would be done for an AI/ML device. The "reported device performance" is the equivalency shown through these characteristics and the results of specified performance tests.

    AttributeSubject Devices: Provena™ Midline CathetersPredicate Devices: Dual Lumen PowerMidline™ CathetersAcceptance Criterion (Implicit: Substantial Equivalence)Reported Device Performance (as Compared to Predicate)
    Indications for UseShort term access to peripheral venous system for selected IV therapies, blood sampling, and power injection of contrast media. Suitable for power injectors.Same as subject.SameEquivalent
    Duration of UseShort term (<30 days)Short term (<30 days)SameEquivalent
    Means of InsertionPercutaneous, Modified Seldinger Technique and guidewireSame as subject.SameEquivalent
    Insertion SitePeripheralSame as subject.SameEquivalent
    Tip Placement LocationPeripheral venous system, tip prior to axillaSame as subject.SameEquivalent
    Number of Lumens3F Single Lumen: 1; 4F Dual Lumen: 24F Dual Lumen: 2-Different (new 3F SL option) but deemed not to raise new safety/effectiveness questions. 4F DL is same.
    Lumen Size3F SL: 18 Ga; 4F DL: 18 Ga (both lumens)4F DL: 19 Ga (large), 21 Ga (small)-Different but deemed not to raise new safety/effectiveness questions.
    Lumen Geometry3F SL: Circular; 4F DL: Two identical D shaped lumens4F DL: One large D shaped and one small D shaped lumen-Different but deemed not to raise new safety/effectiveness questions.
    Catheter Base MaterialsPolyurethane (modifications made)Polyurethane-Similar material; specific modifications evaluated.
    Power Injection Max Flow Rate3F SL: 6 mL/sec; 4F DL: 6 mL/sec4F DL: 3 mL/sec-Different (higher for subject device) but evaluated through performance testing.
    SterilityProvided Sterile (EO)Same as subject.SameEquivalent
    Key Performance TestsPriming Volume, Particulate Matter, Assembly Burst, Catheter Assembly Leak, Catheter Assembly Tensile, Gravity Flow, Device Dimensional Characterization, Power Injection Testing, Biocompatibility EvaluationSimilar tests previously conducted on predicate device.Passed specified tests and met predetermined acceptance criteria.All tests passed and demonstrated substantial equivalence.

    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not applicable in the context of clinical data for AI/ML. The "test set" here refers to the physical devices for engineering performance testing. Numbers of units tested for each engineering test (e.g., burst pressure, tensile strength) are not specified in this summary but would be part of the underlying design verification documentation.
    • Data Provenance: The 'data' for this submission is primarily engineering test results on newly manufactured devices. There is no mention of patient data (e.g., country of origin, retrospective/prospective).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable as this is not an AI/ML device requiring clinical image interpretation or ground truth establishment by experts. The "ground truth" for this device's performance is established by engineering specifications and industry standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable as there is no human interpretation of data requiring adjudication for this device type.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, an MRMC study was not done as this is not an AI/ML device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable as this is not an AI/ML device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • The "ground truth" for this device's substantial equivalence is based on:
      • Engineering Specifications: Defined physical and performance characteristics.
      • Industry Standards: Compliance with standards like ISO 10555-1 (Sterile, single-use intravascular catheters) and USP 788 (Particulate Matter Testing).
      • Risk Analysis: Performed in accordance with BS EN ISO 14971 (Risk Management for Medical Devices).

    8. The sample size for the training set:

    • Not applicable as this is not an AI/ML device that requires a training set.

    9. How the ground truth for the training set was established:

    • Not applicable as this is not an AI/ML device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K210047

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-08-27

    (231 days)

    Product Code
    Regulation Number
    880.5200
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Stiletto Extended Dwell Catheter is inserted into a patient's vascular system for short-term use (<28 days) to sample blood or administer fluids intravenously. These catheters may be used for any patient population with consideration given to adequacy of vascular anatomy and appropriateness of the procedure. The Stiletto Extended Dwell Catheter is suitable for use with power injectors.

    Device Description

    The Stiletto Extended Dwell Catheter (EDC) is a sterile, single use device designed to allow users to sample blood and/or administer fluids over a short duration (<28 days) through a peripherally inserted catheter. The Stiletto EDC utilizes an introducer needle and sheathing cannula to support advancement of a single lumen, power-injectable, catheter into peripheral vasculature. Unique catheter tip geometry is employed that works in conjunction with a sheathing cannula to provide support as the catheter advances into the vessel, providing a reliable method of insertion without need of a guidewire. The catheter, hub and insertion components are integrated into a single housing to enhance ease of use.

    The Stiletto EDC will be available in 8 cm and 10 cm lengths with a 20 gauge catheter, and 10 cm and 12 cm lengths with an 18 gauge catheter.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the "Stiletto Extended Dwell Catheter" based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Standard / RequirementTest NameAcceptance Criteria (Implied)Reported Device Performance
    ISO 7864Needle Surface and SharpnessPassedPassed
    ISO 9626Needle Surface and SharpnessPassedPassed
    ISO 10555-1Needle Surface and SharpnessPassedPassed
    Internal RequirementDimensions – Needle Effective LengthPassedPassed
    ISO 10555-5Catheter to Needle Bevel Heel Lie DistancePassedPassed
    ISO 10555-1RadiopacityPassedPassed
    ASTM F640Needle Cap Removal ForcePassedPassed
    Internal RequirementNeedle Cap Removal ForcePassedPassed
    ISO 10555-5Flashback DetectionPassedPassed
    ANSI/AAMI HE75Assembly Force/Safety Activation ForcePassedPassed
    ISO 23908Assembly Force/Safety Activation ForcePassedPassed
    Internal RequirementInsertion Tool Withdrawal ForcePassedPassed
    Internal RequirementCatheter Kink DiameterPassedPassed
    FDA Guidance DocumentDimensions – Catheter Effective LengthPassedPassed
    ISO 10555-5Dimensions – Catheter ODPassedPassed
    ISO 10555-1Ink PermanencePassedPassed
    ISO 10555-1LeakPassedPassed
    Internal RequirementPump FlowPassedPassed
    Internal RequirementAssembly Tensile – Cannula – PEEK to Stainless SteelPassedPassed
    ISO 23908Needle Safety Override ForcePassedPassed
    ANSI/AAMI HE75Needle Safety Override ForcePassedPassed
    ISO 7864Assembly Tensile – Needle to Top CasePassedPassed
    ISO 9626Needle Tubing StiffnessPassedPassed
    Internal RequirementAssembly Tensile – Cannula – Stainless Steel to Thumb SlidePassedPassed
    ISO 9626Cannula Resistance to BreakagePassedPassed
    ISO 9626Needle Resistance to BreakagePassedPassed
    ISO 10555-5Needle LooseningPassedPassed
    FDA Guidance DocumentPriming VolumePassedPassed
    Internal RequirementGravity FlowPassedPassed
    Internal RequirementCatheter CollapsePassedPassed
    Internal RequirementDimensions – Catheter Outer Diameter (Optical Micrometer)PassedPassed
    ISO 10555-1Assembly Tensile – Catheter Shaft to Core HubPassedPassed
    ISO 10555-1Catheter Shaft TensilePassedPassed
    ISO 10555-1Catheter BurstPassedPassed
    ISO 80369Luer/Hub Evaluations Per ISO 80369PassedPassed
    Biocompatibility
    CytotoxicityMEM ElutionNon-cytotoxicNon-cytotoxic
    SensitizationMagnusson-Kligman MaximizationNon-sensitizerNon-sensitizer
    IrritationIntracutaneous ReactivityNon-irritantNon-irritant
    Systemic Toxicity (Acute)Systemic InjectionNon-toxicNon-toxic
    PyrigenicityMaterial Mediated PyrogenNon-pyrogenicNon-pyrogenic
    HemolysisHemolysis (indirect)Non-hemolyticNon-hemolytic
    HemolysisHemolysis (direct)Non-hemolyticNon-hemolytic
    HemocompatibilityPartial Thromboplastin TimeMinimal ActivatorMinimal Activator
    HemocompatibilityComplement ActivationSimilar to Comparison DeviceSimilar to Comparison Device
    HemocompatibilityDog ThrombogenicityEquivocalEquivocal
    ImplantationImplantationNon-irritantNon-irritant (with rationalization from previous testing)
    Toxicity (various) and CarcinogenicityChemical evaluation of leachates and Toxicological Risk AssessmentMOS > 1 for all compounds and populationsMOS > 1 for all compounds and populations
    Particulate MatterSizing and Counting Particulate Matter: Light Obscuration MethodAcceptable per USP <788> and Legally Marketed Comparative DeviceAcceptable per USP <788> and Legally Marketed Comparative Device

    2. Sample Size Used for the Test Set and the Data Provenance

    The provided text details bench testing and biocompatibility testing. It does not describe a clinical test set in the traditional sense of patient data.

    • Bench Testing: The specific sample sizes for each bench test are not provided in the document. The provenance is implied to be laboratory testing of the device itself.
    • Biocompatibility Testing: The sample sizes for these tests are not explicitly stated, but they would typically involve material samples of the device components. The provenance is implied to be laboratory testing in accordance with ISO 10993 standards.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    Not applicable. The study relies on objective bench test measurements and established biological safety protocols, not subjective expert assessment of a test set in the context of diagnostic or interpretive performance.

    4. Adjudication Method for the Test Set

    Not applicable. There is no expert adjudication method described as the performance is based on objective measurements against standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    Not applicable. This is a medical device (catheter) and the provided documentation focuses on its physical and biological performance, not an AI or imaging diagnostic tool. Therefore, an MRMC study is not relevant.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    Not applicable. This document is about a physical medical device (catheter), not an algorithm or AI system.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's performance is established by:

    • International Standards: Adherence to various ISO and ASTM standards (e.g., ISO 10555-1, ISO 10555-5, ISO 9626, ISO 7864, ISO 80369, ASTM F640, ANSI/AAMI HE75, ISO 23908).
    • FDA Guidance Documents: Compliance with specific FDA guidance.
    • Internal Requirements: Company-defined performance specifications.
    • Biocompatibility Standards: Adherence to ISO 10993 series and USP <788> for particulate matter.
    • Predicate Device Comparison: Performance is often benchmarked against the "PowerGlide Pro™ Midline Catheter" (K162377) to demonstrate substantial equivalence, particularly in terms of flow rates and other functional characteristics.

    8. The Sample Size for the Training Set

    Not applicable. There is no mention of a training set as this is a physical medical device, not a machine learning model.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set involved. The evaluation is based on direct physical and biological testing of the device itself.

    Ask a Question

    Ask a specific question about this device

    K Number
    K192802

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-10-28

    (28 days)

    Product Code
    Regulation Number
    880.5200
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The primeMIDLINE™ Catheters are indicated for short term access to the peripheral venous system for selected intravenous therapies, blood sampling and power injection of contrast media. These catheters are indicated for patients older than 28 days and that weigh more than 10 kg, with consideration given to adequacy of vascular anatomy and appropriateness of the procedure. The primeMIDLINE™ Catheters are suitable for use with power injectors.

    Device Description

    The 4F DL primeMIDLINE™ Catheter is a peripherally placed midline catheter designed to provide access to the vascular system. The catheter lumen is an open-ended design comprised of radiopaque polyurethane. The 4F DL primeMIDLINE ™ Catheter is packaged sterile in a tray with the accessories necessary to establish short-term (less than 30 days) vascular access. The 4F DL primeMIDLINE Catheter is suitable for use with power injectors.

    AI/ML Overview

    The provided text describes a 510(k) submission for a medical device (4F Dual Lumen primeMIDLINE™ Catheter) and focuses on demonstrating its substantial equivalence to a predicate device, rather than proving the device meets specific acceptance criteria based on a study with a test set, expert ground truth, and statistical performance metrics commonly seen in studies for AI/ML-enabled devices.

    Therefore, the requested information regarding a table of acceptance criteria and reported device performance, sample size for test set, data provenance, number of experts for ground truth, adjudication method, MRMC study, standalone performance, type of ground truth, training set size, and training set ground truth establishment is not present in the provided document.

    This document details the regulatory pathway for a conventional medical device, emphasizing its similarity to a previously cleared predicate device through comparisons of intended use, technological characteristics, and performance testing (e.g., tensile strength, flow rate). These tests are typically engineering/bench tests rather than clinical studies with human "ground truth" and reader performance metrics.

    Here's a breakdown of what is provided and why it doesn't fit the requested AI/ML study format:

    1. Acceptance Criteria and Reported Device Performance: Instead of performance in terms of diagnostic accuracy or reader improvement, the document lists various non-clinical performance tests conducted on the device (e.g., Visual Inspection, Tensile Strength, Catheter Elongation, Radiopacity, Lifecycle Performance, Static Burst Pressure, Power Injection Flowrate, Biocompatibility tests). The acceptance criteria for these tests are stated as having been "met all predetermined acceptance criteria as defined in the referenced Standards, FDA Guidance Document or PFM Medical, Inc. Internal Standards." Specific numerical results are not provided in this summary but would be in the full submission.

    2. Sample Size for Test Set and Data Provenance: Not applicable in the context of the reported performance testing. These are bench tests on manufactured units, not a dataset of patient cases.

    3. Number of Experts & Qualifications for Ground Truth: Not applicable. Ground truth for these tests would be objective measurements (e.g., a caliper for dimensional characteristics, a flow meter for flow rate).

    4. Adjudication Method: Not applicable.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study: Not applicable. This is for evaluating changes in human reader performance, typically with AI assistance.

    6. Standalone Performance: Not applicable. This device is a physical catheter, not an algorithm.

    7. Type of Ground Truth Used: For non-clinical tests, "ground truth" refers to the objectively measured values compared against specifications derived from standards or internal requirements.

    8. Sample Size for Training Set & How Ground Truth for Training Set was Established: Not applicable. This device does not involve machine learning or training a model.

    In summary, the provided document describes the regulatory clearance of a physical medical device (a catheter) based on its substantial equivalence to an existing device and a series of engineering and biocompatibility tests. It does not contain information about an AI/ML-based device study with a "test set" in the diagnostic sense, expert ground truth, or human reader performance metrics.

    Ask a Question

    Ask a specific question about this device

    K Number
    K181208

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-11-02

    (179 days)

    Product Code
    Regulation Number
    880.5200
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The catheters are indicated for short term access to the peripheral venous system intravenous therapies, blood sampling, and power injection of contrast media. These catheters may be used for any patient with consideration given to adequacy of vascular anatomy and appropriateness of the catheters are suitable for use with power injectors.

    Device Description

    The Vygon Midline Catheters are EO sterilized, single-use, polyurethane, reverse taper, short-term, peripheral venous system access catheters. They include materials containing barium sulfate for radiopacity.

    AI/ML Overview

    The provided text is a 510(k) Summary for the Vygon Midline Catheters. It details the device's characteristics, intended use, and a comparison to a predicate device to demonstrate substantial equivalence. However, it does not describe a study involving an AI/Machine Learning device.

    Therefore, I cannot extract information regarding:

    • Sample sizes for test or training sets
    • Data provenance (country of origin, retrospective/prospective)
    • Number of experts or their qualifications for ground truth establishment
    • Adjudication methods
    • MRMC comparative effectiveness studies or effect sizes
    • Standalone algorithm performance
    • Type of ground truth (expert consensus, pathology, outcomes)
    • How ground truth for the training set was established

    The document explicitly states under "Clinical Testing": "No Clinical Testing was performed." This further confirms that the type of study you are asking about (related to AI performance, human reader improvement, and ground truth establishment) was not conducted for this medical device.

    The study described is primarily bench testing and biocompatibility testing to demonstrate that the Vygon Midline Catheters are substantially equivalent to a previously cleared predicate device.

    Here's the relevant information that can be extracted, pertaining to acceptance criteria and performance for this non-AI medical device:

    1. A table of acceptance criteria and the reported device performance

    The document lists various bench tests and biocompatibility tests along with their results. The "Acceptance Criteria" are implied by the "Result" column consistently stating "Passed" or specific equivalent findings.

    Table of Acceptance Criteria (Implied) and Reported Device Performance

    StandardSterile Device Test / Test NameReported Device Performance (Result)
    ISO 9626:2016Dimensional InspectionsPassed
    ISO 10555-1:2014Catheter Conditioning (Saline)Passed
    ISO 10555-1:2014Leak TestPassed
    Internal RequirementPriming Volume TestPassed
    ISO 10555-1:2014Gravity Flow TestPassed
    ISO 10555-1:2014Pump Flow TestPassed
    ISO 10555-1:2014Tensile TestPassed
    FDA Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters Section II. D. b. (Mechanical Specifications)Catheter StiffnessPassed
    ISO 10555-1:2014Burst TestPassed
    ISO 10555-1:2014Catheter Flexural FatiguePassed
    ISO 10555-1:2014Power Injection ConditioningPassed
    Internal RequirementPrinting Font SizePassed
    ASTM F1842-15Ink PermanencePassed
    ISO 594-1: 1986 Via ISO 80369-7:2016Luer Taper TestPassed
    ASTM F1842-15Ink Integrity TestPassed
    Internal RequirementAspiration Flow TestPassed
    Internal RequirementLuer Color/Orientation TestPassed
    Internal RequirementClamp Closure MaintenancePassed
    Internal RequirementClamp and ID Tag Fit TestPassed
    Internal RequirementClamp Actuation TestPassed
    ISO 10555-1:2014 (Aged)Power Injection ConditioningPassed
    ISO 10555-1:2014 (Aged)Burst TestPassed
    ISO 10993-5:2009/(R)2014CytotoxicityNon-cytotoxic
    AAMI/ANSI/ISO 10993-10:2010/(R)2014SensitizationNon-sensitizer
    AAMI/ANSI/ISO 10993-10:2010/(R)2014IrritationNon-irritant
    ISO 10993-11:2006Acute Systemic InjectionNon-toxic
    USP37Material Mediated PyrogenNon-pyrogenic
    ISO 10993-4: 2002/(R) 2013 & A1: 2006/ (R) 2013Hemolysis (extract)Non-hemolytic
    ISO 10993-4: 2002/(R) 2013 & A1: 2006/ (R) 2013Hemolysis (direct)Non-hemolytic
    ISO 10993-4: 2002/(R) 2013 & A1: 2006/ (R) 2013Complement Activation (Predicate: PowerPICC® Catheter)Similar when compared to predicate
    ISO 10993-4: 2002/(R) 2013 & A1: 2006/ (R) 2013Partial Thromboplastin Time (PTT) (Predicate: PowerPICC® Catheter)Minimal activator (same as predicate)
    ISO 10993-4: 2002/(R) 2013 & A1: 2006/ (R) 2013Dog ThrombogenicityEquivocal similar
    AAMI/ANSI/ISO 10993-6:2007/(R)2014ImplantationNon-irritant
    ISO 10993-18 (Nelson Labs Report does not specify year.)Extractable/Leachable AnalysisSummarized in project #MJ16357-BIO01
    ISO 10993-17:2008Subacute/Subchronic Toxicity, Genotoxicity, Chronic Toxicity and CarcinogenicityEvaluated in Toxicological Risk Assessment

    2. Sample sizes used for the test set and the data provenance
    The document does not explicitly state the sample sizes (number of devices tested) for each bench or biocompatibility test, nor does it specify data provenance as this is not a study involving human data or image data as typically seen with AI devices. The testing is laboratory-based "bench testing" of the physical devices.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
    Not applicable, as this is a physical medical device clearance, not an AI/ML device where expert ground truth is established. Performance is determined by standardized laboratory tests adhering to ISO standards and internal requirements.

    4. Adjudication method
    Not applicable.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done
    No, this type of study was not performed. The document explicitly states "No Clinical Testing was performed."

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
    Not applicable, as this is a physical medical device, not an algorithm.

    7. The type of ground truth used
    Not applicable for the reasons stated above. The "ground truth" for this device's performance is compliance with international and national standards for medical devices (e.g., ISO, ASTM, FDA guidance) and internal requirements, demonstrated through bench and biocompatibility testing.

    8. The sample size for the training set
    Not applicable. This is not an AI/ML device.

    9. How the ground truth for the training set was established
    Not applicable. This is not an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K173114

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-03-07

    (159 days)

    Product Code
    Regulation Number
    880.5200
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The primeMIDLINE™ Catheters are indicated for short term access to the peripheral venous system for selected intravenous therapies, blood sampling and power injection of contrast media. These catheters are indicated for patients older than 28 days and that weigh more than 10kg, with consideration given to adequacy of vascular anatomy and appropriateness of the procedure. The primeMIDLINE™ Catheters are suitable for use with power injectors.

    Device Description

    The primeMIDLINE™ Catheters are a family of peripherally placed midline catheters designed to provide access to the vascular system. These catheters are offered in 3F, 4F and 5F Single Lumen and 5F Dual Lumen configurations with 20cm lengths. The catheter lumen is an open-ended design comprised of radiopaque polyurethane. Each primeMIDLINE™ Catheter is packaged sterile in a tray with the accessories necessary to establish short-term (less than 30 days) vascular access. The primeMIDLINE™ Catheters are suitable for use with power injectors.

    AI/ML Overview

    The document provided describes the primeMIDLINE™ Catheters and their substantial equivalence to a predicate device, PowerMidline Catheters. It outlines verification testing performed to demonstrate that the device performs as intended and meets established acceptance criteria.

    However, the document does not contain information about a study comparing the device's performance against specific acceptance criteria in the format requested. While it lists various performance testing categories, it does not provide:

    • A table of specific quantitative acceptance criteria for each test (e.g., "Tensile Strength should be > X N") and the reported device performance against those criteria.
    • Details on sample sizes for a "test set" in the context of device performance, data provenance, or the number/qualifications of experts for ground truth.
    • Information on adjudication methods, MRMC comparative effectiveness studies, or standalone algorithm performance, as these are typically relevant to AI/diagnostic imaging devices, not direct hardware performance.
    • The type of ground truth used for performance validation.
    • Training set sample size or how ground truth was established for a training set (again, this points more towards AI/ML studies).

    Instead, the document states: "The subject devices met all predetermined acceptance criteria as defined in the referenced Standards, FDA Guidance Documents or PFM Medical, Inc. Internal Standards." This is a general statement that the device passed all tests, but without quantifying the criteria or the results.

    Therefore, based on the provided text, I cannot complete the table or answer most of the requested questions because the specific data is not available.

    Summary of available information:

    • Device: primeMIDLINE™ Catheters
    • Predicate Device: PowerMidline Catheters (K153393)

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria CategoryReported Device Performance
    Visual InspectionMet (Statement: "The subject devices met all predetermined acceptance criteria")
    Dimensional CharacteristicsMet (Statement: "...met all predetermined acceptance criteria")
    RadiopacityMet (Statement: "...met all predetermined acceptance criteria")
    Catheter CollapseMet (Statement: "...met all predetermined acceptance criteria")
    LeakMet (Statement: "...met all predetermined acceptance criteria")
    Priming VolumeMet (Statement: "...met all predetermined acceptance criteria")
    Flowrate: GravityMet (Statement: "...met all predetermined acceptance criteria")
    Flowrate: Power InjectionMet (Statement: "...met all predetermined acceptance criteria")
    Tensile StrengthMet (Statement: "...met all predetermined acceptance criteria")
    Catheter ElongationMet (Statement: "...met all predetermined acceptance criteria")
    Lifecycle PerformanceMet (Statement: "...met all predetermined acceptance criteria")
    Static Burst PressureMet (Statement: "...met all predetermined acceptance criteria")
    Suture Wing IntegrityMet (Statement: "...met all predetermined acceptance criteria")
    Luer TestingMet (Statement: "...met all predetermined acceptance criteria")
    Biocompatibility Tests:
    CytotoxicityMet (Statement: "...met all predetermined acceptance criteria")
    SensitizationMet (Statement: "...met all predetermined acceptance criteria")
    Irritation/Intracutaneous ReactivityMet (Statement: "...met all predetermined acceptance criteria")
    Acute Systemic ToxicityMet (Statement: "...met all predetermined acceptance criteria")
    Material Mediated PyrogenicityMet (Statement: "...met all predetermined acceptance criteria")
    Subacute/Subchronic ToxicityMet (Statement: "...met all predetermined acceptance criteria")
    GenotoxicityMet (Statement: "...met all predetermined acceptance criteria")
    CarcinogenicityMet (Statement: "...met all predetermined acceptance criteria")
    Chronic ToxicityMet (Statement: "...met all predetermined acceptance criteria")
    HemocompatibilityMet (Statement: "...met all predetermined acceptance criteria")
    ImplantationMet (Statement: "...met all predetermined acceptance criteria")
    Sterilization Validation:
    Sterilization Validation – Overkill MethodMet (Statement: "...met all predetermined acceptance criteria")
    EO/ECH ResidualsMet (Statement: "...met all predetermined acceptance criteria")
    Direct Product SterilityMet (Statement: "...met all predetermined acceptance criteria")
    Bioburden/Organism CharacterizationMet (Statement: "...met all predetermined acceptance criteria")
    Bacteriostasis Fungistasis TestingMet (Statement: "...met all predetermined acceptance criteria")
    Bacterial Endotoxins/Limulus Amebocyte LysateMet (Statement: "...met all predetermined acceptance criteria")

    Study Information (based on available text):

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    • Not specified. The document states "design verification testing" was conducted but does not provide sample sizes for individual tests or data provenance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    • Not applicable/Not specified. The testing described is performance testing of a physical device against standards, not expert-adjudicated diagnostic performance.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable/Not specified.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This type of study is for diagnostic tools, typically AI-powered. The device is a physical catheter.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • No. This is not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • "Predetermined acceptance criteria as defined in the referenced Standards, FDA Guidance Documents or PFM Medical, Inc. Internal Standards." This refers to objective measurements against established engineering and safety standards.

    8. The sample size for the training set:

    • Not applicable. This is not an AI/ML device that requires a training set.

    9. How the ground truth for the training set was established:

    • Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K171483

    Validate with FDA (Live)

    Date Cleared
    2017-12-15

    (207 days)

    Product Code
    Regulation Number
    880.5200
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Arch-Flo CT Midline is indicated for Short-Term peripheral venous system for selected intravenous therapies, blood sampling, and power injection of contrast media. This catheter may be inserted via basilic, cephalic, or median cubital vein.

    Device Description

    The Arch-Flo'™ CT Midline is available in a 4F Single Lumen configuration. The catheter lumen terminates through an extension to a female luer-lock connector. The extension has an in-line clamp to control fluid flow and is marked POWER INJECTABLE MIDLINE along with the lumen gauge size. The transition between lumen and extension is housed within a molded hub. The hub is marked MIDLINE to identify that the catheter is not centrally placed. The tip of the lumen is notched to increase the exposed area of the lumen. The outside diameter of the lumen increases gradually near the hub to aid in kink resistance and to provide a mechanical obstruction to bleeding from the venotomy. The lumen is marked with depth marks every centimeter.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device called "Arch-Flo CT Midline." This document primarily focuses on establishing substantial equivalence to a predicate device through comparison and performance testing. It does not contain information about acceptance criteria and studies related to the performance of an AI/algorithm-based device.

    The device in question is an "Intravascular Catheter," which is a traditional medical device, not an AI or software-based application. Therefore, the specific questions regarding AI device performance (like the number of experts for ground truth, MRMC studies, standalone performance with AI, training set details, etc.) are not applicable to the content provided.

    However, I can extract the general "acceptance criteria" (or performance data) for the physical device based on the bench/performance data and comparison to the predicate device.

    Here's the information that can be extracted, interpreted based on the context of a traditional medical device submission:

    1. A table of acceptance criteria and the reported device performance:

    The document describes performance data for the Arch-Flo CT Midline and compares it to the predicate device, CT Midline (K141151). The "acceptance criteria" are implied by the standards the device was tested against and by demonstrating similar or superior performance to the predicate device.

    Performance CharacteristicAcceptance Criteria (Implied by Standards/Predicate)Proposed Device (Arch-Flo CT Midline) PerformancePredicate Device (CT Midline) Performance
    Max Power Injection Flow5cc/sec (matching predicate)5cc/sec5cc/sec
    Priming VolumeComparable to predicate0.36cc0.41cc
    Gravity FlowComparable to predicate56 ml/min31.5 ml/min
    Air LeakageMeets ISO 10555-1 and ISO 10555-3 requirementsTested and metImplied met as predicate
    Liquid LeakageMeets ISO 10555-1 and ISO 10555-3 requirementsTested and metImplied met as predicate
    Force at BreakMeets ISO 10555-1 requirementsTested and metNot specified
    ElongationMeets ISO 10555-1 requirementsTested and metNot specified
    Chemical ExposureMeets ISO 10555-1 requirementsTested and metNot specified
    Maximum Burst PressureMeets ISO 10555-1 requirementsTested and metNot specified
    Cyclic FlexureMeets ISO 10555-1 requirementsTested and metNot specified
    Sterilization (SAL)10^-6 (ANSI/AAMI/ISO 11135-1:2014)SAL is 10^-6Not specified
    ETO Residual Level< sterilant residual limits per ISO 10993-7<= sterilant residual limitsNot specified
    Pyrogenicity (LAL)Non-pyrogenic (ANSI/AAMI ST72:2011)Non-pyrogenicNot specified
    EndotoxinMax 20 EU/device (ANSI/AAMI ST72:2011)Max 20 EU/deviceNot specified
    BiocompatibilityMeets ISO 10993-1, -10, -11, -4, -3, -5, -6, ASTM F 756-08, ASTM F 619-03Tested and met for all relevant endpointsNot specified

    2. Sample sized used for the test set and the data provenance:

    • Test Set Sample Size: Not explicitly stated in terms of a specific number of devices for each test. The document mentions "performance testing" was performed according to various ISO and ASTM standards. These standards typically define the sample sizes required for validation studies.
    • Data Provenance: The testing was "bench / performance data / non-clinical testing" conducted by the manufacturer, Medcomp®. There is no mention of country of origin for data (as it's bench testing, not patient data) or whether it was retrospective or prospective (which applies to patient studies).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This question is not applicable. The device is a physical intravascular catheter. Its performance is evaluated through objective physical and chemical tests against established international standards (e.g., ISO, ASTM) and comparison to a predicate device, not by expert interpretation of data points in the way an AI model would be validated.

    4. Adjudication method for the test set:

    Not applicable, as the evaluation is based on objective measurements and comparison to standards, not human interpretation requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This is a traditional medical device, not an AI device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Not applicable. This is a traditional medical device.

    7. The type of ground truth used:

    The "ground truth" used for this device is based on:

    • International Standards: e.g., ISO 10555-1, ISO 11607-1, ISO 10555-3, ANSI/AAMI/ISO 11135-1, AAMI TIR 28, ANSI/AAMI ST72, ISO 10993 series, ASTM F 756-08, ASTM F 619-03.
    • Predicate Device Performance: The performance of the legally marketed predicate device (CT Midline K141151) serves as a benchmark for substantial equivalence.
    • Objective Measurements: Physical and chemical measurements (e.g., flow rates, volumes, material properties, sterility levels) are the "ground truth" for compliance.

    8. The sample size for the training set:

    Not applicable. This is a traditional medical device, not an AI device that requires a training set.

    9. How the ground truth for the training set was established:

    Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K170770

    Validate with FDA (Live)

    Device Name
    CT Midline
    Date Cleared
    2017-07-25

    (133 days)

    Product Code
    Regulation Number
    880.5200
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CT Midlines are indicated for Short-Term peripheral access to the peripheral venous system for selected intravenous therapies, blood sampling, and power injection of contrast media. The maximum recommended infusion rate varies by catheter French size and is printed on the catheter.

    Device Description

    The CT Midline Catheter is designed for peripheral vein catheterization and power injection of contrast media. The lumen is an open-ended design comprised of a soft radiopaque polyurethane material with barium sulfate for radiopacity. The lumen is connected to the extensions via a soft pliable hub with suture wing for secure placement. Clamps are provided on the extension tubes to prevent air/fluid contamination. Female luer connectors provide the connection for intravenous administration.

    The CT Midline Catheter is available in a 4Fx20cm single-lumen, or 5Fx20cm double-lumen configuration. The outside diameter of the lumen has a reverse taper increasing gradually near the hub to aid in kink resistance and to provide a mechanical obstruction to bleeding from the venotomy. The lumen has depth marks every centimeter and numerical marks every fifth centimeter. The CT Midline is packaged sterile with the necessary accessories to facilitate insertion.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a medical device (CT Midline) and primarily discusses its substantial equivalence to a predicate device, rather than presenting a study of the device's performance against specific acceptance criteria.

    The submission states that the purpose of this particular 510(k) is to "remove a contraindication that was once cited in the Journal of Infusion Nursing but has since been revised." This is a change to the labeling, not a new performance claim or a new device design that would necessitate a comprehensive performance study with acceptance criteria and reported results.

    Therefore, the document does not contain the detailed information requested regarding acceptance criteria and a study proving the device meets those criteria.

    Here's a breakdown of why each point cannot be thoroughly answered from the provided text:

    1. A table of acceptance criteria and the reported device performance:

      • The document implies that the device's performance (e.g., flow rates for power injection) is identical to the predicate device K141151.
      • Table 5.1 "Design Comparison Matrix" lists "Performance Testing" with "Power Injection: 5cc/sec (4F), 7cc/sec (5F)" for both the proposed and predicate devices, stating "Identical, no change." This indicates the expected performance, but not explicit acceptance criteria, nor the results of a new study designed to meet those criteria for this specific submission. It's a statement of equivalence.
    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

      • No test set is described for evaluating new performance. The comparison is based on the predicate device's existing performance data.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

      • Not applicable as no new performance test set requiring expert ground truth is described.
    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not applicable.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not applicable. This device is an intravascular catheter, not an AI-powered diagnostic tool.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Not applicable.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

      • Not applicable. The "ground truth" here is the established performance of the predicate device, against which the proposed device is declared equivalent.
    8. The sample size for the training set:

      • Not applicable as this is a physical medical device, not an algorithm requiring a training set.
    9. How the ground truth for the training set was established:

      • Not applicable.

    Summary from the document:

    The submission explicitly states under "Bench / Performance Data / Non-Clinical Testing":
    "No performance testing was required for the revision to the instructions in conjunction with the substantial equivalence claims, effectively demonstrate the proposed device, CT Midline, is equivalent to the predicate device, CT Midline (K141151)."

    The basis for equivalence is that the "Proposed CT Midline is substantially equivalent to the predicate device. CT Midline (K141151), in all aspects, including but not limited to, indications for use, intended use, anatomical location, basic design, dimensions, lengths performance, biocompatibility, materials, manufacturing process and method of sterilization." The only difference is a "revision to the instructions for use (labeling)" to remove a contraindication.

    Therefore, for this specific 510(k) submission (K170770), no new study was conducted to prove the device meets performance acceptance criteria, because the device itself and its performance characteristics are deemed identical to a previously cleared predicate device. The change is solely in the labeling information, specifically removing a contraindication.

    Ask a Question

    Ask a specific question about this device

    K Number
    K170158

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-06-01

    (134 days)

    Product Code
    Regulation Number
    880.5200
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PowerGlide ST™ Midline Catheter is inserted into a patient's vascular system for short term use to sample blood or administer fluids intravenously. These catheters may be used for any patient population with consideration given to adequacy of vascular anatomy and appropriateness of the procedure. The PowerGlide ST™ Midline Catheter is suitable for use with power injectors.

    Device Description

    Bard Access Systems, Inc.'s PowerGlide ST™ Midline Catheter is a sterile, single use device designed to provide access to the patient's vascular system. The device is intended for short term use (<30 days) to sample blood and administer fluids intravenously. The subject device consists of a single lumen, radiopaque, body softening polyurethane catheter rated for power injection, and a dilator (when applicable). Additional procedural kit accessories (e.g. introducer needle and guidewire) are included to facilitate catheter placement. The PowerGlide ST™ Midline Catheter is offered in 18, 20, and 22 gauge sizes. The 18 and 20 gauge devices are offered in 8 cm or 10 cm lengths with an integrated dilator. The 22 gauge device is offered in only an 8 cm length and does not include a dilator.

    AI/ML Overview

    This document describes the premarket notification for a medical device, the PowerGlide ST™ Midline Catheter. It focuses on demonstrating substantial equivalence to a predicate device, not on proving clinical effectiveness through novel AI or diagnostic algorithms. Therefore, many of the requested criteria related to AI performance, ground truth establishment, expert readers, and MRMC studies are not applicable to this type of regulatory submission.

    Here's an analysis based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document lists performance tests and states that the "subject device met all predetermined acceptance criteria". However, the specific quantitative acceptance criteria values and the exact reported performance results for each test are not disclosed in this summary. It only indicates that the device passed.

    Performance TestAcceptance Criteria (Not Explicitly Stated)Reported Performance
    Dilator Outer DiameterWithin design specifications (Bard internal stds)Met criteria
    Dilator Effective LengthWithin design specifications (BS EN ISO 11070:2014)Met criteria
    Dilator Luer ConnectorComply with ISO 594-1 and 594-2 (6% luer taper)Met criteria
    Catheter/Dilator Tip Adhesion Break ForceForce required to remove dilator (Bard internal stds)Met criteria
    Dilator Assembly TensileDoes not fail in tension (BS EN ISO 11070:2014)Met criteria
    Catheter/Dilator Assembly Burst PressureGreater than maximum use pressure (ISO 10555-1:2013)Met criteria

    Biocompatibility Tests:

    Biological TestAcceptance Criteria (Not Explicitly Stated)Reported Performance
    CytotoxicityPer ISO 10993-5 (2009)Met criteria
    SensitizationPer ISO 10993-10 (2010)Met criteria
    Intracutaneous ReactivityPer ISO 10993-10 (2010)Met criteria
    Acute Systemic ToxicityPer ISO 10993-11 (2006)Met criteria
    PyrogenicityPer ISO 10993-11 (2006)Met criteria
    HemocompatibilityPer ISO 10993-4 (2002, amended 2006)Met criteria
    GenotoxicityPer ISO 10993-3 (2014), ISO/TR 10993-33 (2015)Met criteria
    Sterilization ResidualsPer ISO 10993-7Met criteria

    2. Sample size used for the test set and the data provenance
    The document does not specify the sample sizes used for the described physical and biological performance tests. It also does not discuss data provenance (e.g., country of origin, retrospective/prospective) as these are non-clinical, in-vitro/bench tests on device components, not clinical data sets.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
    This question is not applicable. The "ground truth" for this medical device submission is based on engineering specifications, physical/chemical properties, and biological safety standards (ISO standards). It does not involve expert interpretations of medical images or diagnoses that would require human expert consensus.

    4. Adjudication method for the test set
    This question is not applicable. Adjudication methods like 2+1 or 3+1 are used in clinical studies, particularly for diagnostic accuracy, where there's a need to resolve discrepancies between human readers or between AI and human reads. This document describes bench testing and biocompatibility assessments which do not involve this type of adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
    This question is not applicable. This submission is for a physical medical device (catheter) and does not involve AI or human readers for diagnostic purposes. Therefore, an MRMC study and effects on human reader performance are irrelevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
    This question is not applicable. This device is a physical medical catheter, not an algorithm.

    7. The type of ground truth used
    The "ground truth" for this device's acceptance is based on:

    • Engineering design specifications: The physical dimensions and properties of the dilator and catheter.
    • Industry Standards (ISO/BS EN ISO): Established international and national standards for medical devices, specifically for intravascular catheters, luer connectors, and biocompatibility.
    • Bard internal standards and procedures: Proprietary testing methods and acceptance criteria used by the manufacturer.

    8. The sample size for the training set
    This question is not applicable. This is not an AI/machine learning device, so there is no concept of a "training set."

    9. How the ground truth for the training set was established
    This question is not applicable for the same reason as point 8.

    Ask a Question