Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K250925

    Validate with FDA (Live)

    Date Cleared
    2025-12-16

    (264 days)

    Product Code
    Regulation Number
    866.6010
    Age Range
    22 - 87
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ADVIA Centaur Cytokeratin Fragment 21-1 (CYFRA) assay is for in vitro diagnostic use in the quantitative measurement of cytokeratin 19 fragments in human serum and plasma (EDTA and lithium heparin) using the ADVIA Centaur XPT system.

    The measurement of cytokeratin 19 is used as an aid in monitoring disease progression during the course of disease and treatment in lung cancer patients. Serial testing for patient CYFRA 21-1 assay values should be used in conjunction with other clinical information used for monitoring lung cancer.

    Device Description

    The ADVIA Centaur Cytokeratin Fragment 21-1 assay includes 1 ReadyPack primary reagent pack containing ADVIA Centaur CYFRA Lite Reagent and Solid Phase and ADVIA Centaur CYFRA master curve card.

    The ReadyPack consists of the following:

    ADVIA Centaur Cytokeratin Fragment 21-1 ReadyPack primary reagent pack; Liquid Lite Reagent 10.0 mL/reagent pack KS19.1 monoclonal antibody acridinium conjugate in buffer containing bovine serum albumin; surfactant; and preservatives.

    ADVIA Centaur Cytokeratin Fragment 21-1 ReadyPack primary reagent pack; Liquid Solid Phase Reagent 17.5 mL/reagent pack BM19.21 monoclonal antibody coupled to magnetic microparticle bead in buffer containing bovine serum albumin; surfactant; and preservatives.

    ADVIA Centaur Multi-Diluent13 ReadyPack ancillary reagent pack; 2×10.0mL/reagent pack; Buffer; surfactant; sodium azide (< 0.1%)

    This assay is a fully automated 1-step sandwich immunoassay using acridinium ester chemiluminescent technology. The Solid Phase contains magnetic microparticles coated with anti-CYFRA 21-1 BM19.21 mouse monoclonal antibody. The Lite Reagent consists of acridinium ester-labeled anti-CYFRA 21-1 KS19.1 mouse monoclonal antibody. A direct relationship exists between the amount of CYFRA 21-1 present in the patient sample and the amount of relative light units (RLUs) detected by the system.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K160915

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-12-14

    (257 days)

    Product Code
    Regulation Number
    866.6010
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Elecsys CYFRE 21-1:

    Immunoassay for the in vitro quantitative determination of tragments of cytokeratin 19 in human serum and plasma (Li-Heparin. K2-EDTA and K3-EDTA). The assay is to be used as an aid in monitoring disease progression during the course of disease and treatment in lung cancer patients. Serial testing for patient CYFRA 21-1 assay values should be used in conjunction with other clinical methods used for monitoring lung cancer.

    The electrochemiluminescence immunoassay "ECLIA" is intended for use on Elecsys and cobas e immunoassay analyzers.

    Elecsys CYFRA 21-1 CalSet:

    CYFRA 21-1 is used for calibrating the quantitative Elecsys CYFRA 21-1 assay on the Elecsys and cobas e immunoassay analyzers.

    Elecsys PreciControl Tumor Marker:

    PreciControl Tumor Marker is used for quality control of Elecsys immunoassays on Elecsys and cobas e immunoassay analyzers.

    Device Description

    (1) The Elecsys CYFRA 21-1 Immunoassay is a two-step sandwich immunoassay with streptavidin microparticles, a biotinylated monoclonal cytokeratin 19-specific antibody, and a monoclonal cytokeratin 19-specific antibody labeled with a ruthenium complex and electrochemiluminescence detection. Results are determined using a calibration curve that is generated specifically on each instrument by a 2 point calibration master curve (5-pointcalibration) provided with the reagent bar code.

    (2) The Elecsys CYFRA 21-1 CalSet is a lyophilized product consisting of Cytokeratin from cell culture of the cell line MCF-7 in two concentration ranges (approximately 0 ng/mL and 50 ng/mL) in a cytokeratin free human serum matrix with preservative. During manufacture, the analyte is spiked ito the matrix at the desired concentration levels.

    (3) The Elecsys PreciControl Tumor Marker (CYFRA 21-1) is a lyophilized control serum in two concentration ranges (approximately 3.29 ng/mL and 27.2 ng/mL).

    Note: The reagent, calibrator, and the quality control materials are all packaged separately.

    AI/ML Overview

    The provided document describes the Elecsys CYFRA 21-1 Immunoassay, a device intended for quantitative determination of cytokeratin 19 fragments in human serum and plasma, used as an aid in monitoring disease progression and treatment for lung cancer patients.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    The FDA clearance for this in-vitro diagnostic device does not typically specify "acceptance criteria" in the same way a medical image analysis AI might have specific sensitivity/specificity thresholds. Instead, it demonstrates substantial equivalence to a predicate device by comparing various performance characteristics. The acceptance is implied by demonstrating performance that is comparable to or better than the predicate device across critical analytical and clinical metrics.

    Performance CharacteristicAcceptance Criteria (Implied by Predicate/Regulatory Standards)Reported Device Performance (Elecsys CYFRA 21-1)
    Intended Use/IndicationsAid in monitoring disease progression during course of disease and treatment in lung cancer patients (for serum)Aid in monitoring disease progression during course of disease and treatment in lung cancer patients (for human serum and plasma, Li-Heparin, K2-EDTA and K3-EDTA)
    Measuring Range0.5 - 50 ng/mL0.5 – 100 ng/mL
    Precision (Within-run CV)Predicate: 2.40% - 5.10% CV0.7% - 4.8% CV
    Precision (Total/Intermediate CV)Predicate: 4.90% - 8.40% CV0.7% - 9.6% CV
    Limit of Blank (LoB)Not determined for predicate0.0946 ng/mL
    Limit of Detection (LoD)0.12 ng/mL0.1813 ng/mL
    Limit of Quantitation (LoQ)0.21 ng/mL with 17.78% imprecision0.286 ng/mL with 30% total allowable error
    Hook EffectNo high-dose hook effect up to 1100 ng/mLNo high-dose hook effect up to 2427 ng/mL
    Cross-Reactivity/InterferenceLimited information on specific thresholds for predicate; generally, minimal interference expected.Largely unaffected by common interferents (e.g., Lipemia, Biotin, Bilirubin, Hemoglobin, Rheumatoid Factor, HAMA). Recovery of spiked HAMA serum: 98.51%.
    Method Comparison (vs. Predicate)High correlation (implied by good slope and correlation coefficient)Slope 0.911 (95% CI: 0.86, 0.96); Intercept -1.4 (95% CI: -2.86, 0.06); Pearson r 0.955
    Clinical Sensitivity (at 50% change)45.9%44.1%
    Clinical Specificity (at 50% change)87.3%91.0%
    Expected Values (Normal Population)Predicate: Primarily ≤ 2.37 ng/mLElecsys: 228/240 (95%) of apparently healthy individuals had values ≤ 2.37 ng/mL
    Normal Range (Benign Conditions)Predicate: Primarily ≤ 2.37 ng/mLElecsys: Varies by condition, but generally low for benign lung disease, higher for CHF & benign kidney/liver disease.

    2. Sample Size Used for the Test Set and Data Provenance

    The document details numerous analytical validation studies. Here's a breakdown for key performance metrics:

    • Precision (Human Sera):
      • Internal: Seven-member panel (5 patient samples, 2 controls). Tested in single determination, 4 aliquots, over 21 operating days. Total measurements: 7 samples * 4 aliquots/day * 21 days = 588 measurements (at least, assuming single determination each aliquot).
      • External: Two replicates of each control (PC TM 1 & 2) and six human serum samples per run, two runs per day for 20 days at three external sites. Total measurements: (2 controls + 6 HS) * 2 replicates * 2 runs/day * 20 days * 3 sites = 1,920 measurements.
      • Provenance: Not explicitly stated, but common for such studies to use mixtures of human serum/controls which could be sourced from various regions due to ethical and availability considerations. Prospective data collection for these specific tests.
    • Limit of Blank (LoB): 5 analyte-free serum samples. 60 measuring points collected (implies 5 samples * 2-fold determination/run * 6 runs = 60).
    • Limit of Detection (LoD): 5 low-level human serum samples. 60 measuring points collected (implies 5 samples * 2-fold determination/run * 6 runs = 60).
    • Limit of Quantitation (LoQ): 8 spiked human serum samples. 200 measuring points collected (implies 8 samples * 5 runs * single determination = 40, assuming some dilutions).
    • Linearity: Six dilution series (3 serum, 3 plasma), each with at least 11 dilutions.
    • Endogenous Interferences: Three human serum samples (low, mid, high CYFRA 21-1) for each of six interfering substances.
    • HAMA Effect: One HAMA serum and one control serum, both spiked to two different analyte concentrations. Each concentration had 11 dilutions, measured in duplicate.
    • High-Dose Hook Effect: Human serum pools spiked up to 2,470ng/mL.
    • Exogenous Interferences - Drugs: Two human serum samples, spiked with 27 pharmaceutical compounds. Samples tested in three-fold determination.
    • Serum/Plasma Comparison: At least 44 serum/plasma pairs per sample type. For Serum/Plasma Separation Tube comparison, 5 samples in duplicate.
    • Clinical Performance (Monitoring): 83 patients with lung cancer, 398 samples measured (86 baseline, 315 monitoring values). Patients had ≥ 3 blood draws over time.
      • Provenance: Not explicitly stated, but implies clinical samples from lung cancer patients, likely retrospective from a patient cohort or collected prospectively for the study.
    • Expected Values (Normal Population):
      • 240 apparently healthy men and women (equally divided into smokers and non-smokers).
      • 195 benign disease conditions.
      • 440 cancer cases (including 120 lung cancer, 40 bladder, breast, cervical, ESCC, GI tract, head and neck, prostate, and ovarian cancer each).
      • Provenance: This data represents cohort studies, likely retrospective collection from various institutions.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    For this type of immunoassay, "ground truth" is typically defined by the actual analyte concentration in the samples or by established clinical diagnoses/disease progression based on a combination of clinical assessment, imaging, and potentially pathology.

    • Clinical Performance (Monitoring): The document states that clinical effectiveness was assessed by "comparing changes in CYFRA 21-1 levels in serial serum samples from 83 patients compared to changes in disease status." The "disease status" would be the ground truth for clinical progression. However, the number and qualifications of experts (e.g., oncologists, radiologists) establishing this disease status ground truth are NOT specified in the provided text.

    4. Adjudication Method for the Test Set

    Not applicable in the typical sense for an immunoassay. The clinical performance study relies on a comparison of assay values to "disease status," which is a clinical endpoint rather than an adjudicated radiological finding or similar interpretation. The judgment of disease progression would be based on standard clinical practice for lung cancer, likely involving follow-up imaging, biopsy results, and physician assessment, but no specific adjudication panel or method (e.g., 2+1) is described for the clinical endpoint data.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    Not applicable. This is an in-vitro diagnostic device for quantitative measurement of a biomarker, not an AI for image interpretation that augments human readers. Therefore, an MRMC study comparing human readers with and without AI assistance is not relevant to this device's evaluation.

    6. Standalone Performance Study (Algorithm Only Without Human-in-the-Loop Performance)

    Yes, the entire submission is a standalone performance study of the Elecsys CYFRA 21-1 immunoassay. The device generates quantitative results (ng/mL) without human interpretation as part of its primary function. The clinical performance section assesses how these quantitative results_correlate_ with disease status, but the assay itself doesn't have a "human-in-the-loop" component in its operation.

    7. Type of Ground Truth Used

    • Analytical Studies (Precision, LoB, LoD, LoQ, Linearity, Interference, Calibration, Stability): The ground truth is the actual concentration of the analyte (CYFRA 21-1) in spiked samples, known concentrations in controls, or the absence of the analyte in blank/interferent-free samples. This is established through highly controlled laboratory procedures and reference methods.
    • Clinical Performance (Monitoring): The ground truth is the "disease status" (progression or no progression) of lung cancer patients. The method for establishing this status is not detailed beyond "changes in disease status," but it inherently relies on a combination of medical evidence and clinical assessment over time.
    • Expected Values (Normal Population & Cancer/Benign Cohorts): The ground truth is the clinical diagnosis/health status of the individuals in those cohorts (e.g., "apparently healthy," "benign lung disease," "lung cancer").

    8. Sample Size for the Training Set

    For an immunoassay, the concept of a "training set" as understood in machine learning (where an algorithm learns from labeled data) doesn't directly apply. However, the development and optimization of the assay (e.g., antibody selection, reagent formulation, instrument parameters) would involve extensive internal testing. The submission here focuses on validation, demonstrating performance on independent (test) samples. The assay itself has pre-defined chemical and electrical principles, not an adaptive learning algorithm.

    9. How the Ground Truth for the Training Set Was Established

    As explained above, "training set" doesn't directly apply. However, during the initial development and optimization phase, the ground truth would be established through:

    • Reference materials: Samples with precisely known concentrations of CYFRA 21-1 (traceable to internal or international standards).
    • Clinical samples: Use of various clinical samples (e.g., from healthy individuals, various disease states) to understand the assay's behavior and define preliminary measuring ranges and expected values. This preliminary testing would inform the design of the formal validation studies presented in the 510(k). The "Enzymun-Test CYFRA 21-1 method" is mentioned as a method against which the Elecsys CYFRA 21-1 has been standardized, indicating a reference for comparison during development.
    Ask a Question

    Ask a specific question about this device

    K Number
    K100831

    Validate with FDA (Live)

    Date Cleared
    2011-05-26

    (428 days)

    Product Code
    Regulation Number
    866.6010
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CYFRA 21-1 EIA kit is intended for the quantitative determination of soluble cytokeratin 19 fragments in human serum. The assay is to be used as an aid in monitoring disease progression during the course of disease and treatment in lung cancer patients. Serial testing for patient CYFRA 21-1 assay values should be used in conjunction with other clinical methods used for monitoring lung cancer.

    Device Description

    The CYFRA 21-1 EIA is a solid phase, non-competitive immunoassay based on two monoclonal antibodies (derived from mice) directed against two separate antigenic determinants of soluble fragments of cytokeratin 19. Calibrators, controls and patient samples are incubated together with biotinylated Anti-CYFRA 21-1 MAb and horseradish peroxidase (HRP) labeled Anti-CYFRA 21-1 MAb in streptavidin coated micro strips. After washing, buffered Substrate/Chromogen reagent (hydrogen peroxide and 3, 3', 5, 5' tetramethylbenzidine) is added to each well and the enzyme reaction is allowed to proceed. During the enzyme reaction a blue color will develop if antigen is present. The intensity of the color development is proportional to the amount of CYFRA 21-1 present in the samples. The color intensity is determined in a microplate spectrophotometer at 620 nm (or optionally at 405 nm after addition of Stop Solution). Calibration curves are constructed for each assay by plotting absorbance value versus the concentration for each calibrator. The CYFRA 21-1 concentrations of patient samples are then read from the calibration curve.

    AI/ML Overview

    Here's an analysis of the provided text regarding the CYFRA 21-1 EIA Kit, focusing on acceptance criteria and the supporting study:

    Acceptance Criteria and Device Performance

    The document does not present explicit, pre-defined "acceptance criteria" in a typical numerical format (e.g., "Sensitivity must be >X%"). Instead, it describes performance characteristics, and the "acceptance" is implied by the FDA's clearance based on demonstrated substantial equivalence to a predicate device and adherence to relevant guidelines.

    However, based on the studies presented, we can infer performance metrics that were deemed acceptable for clearance. The key study provided is a clinical study for monitoring disease progression in lung cancer.

    Table of Performance Metrics

    Performance CharacteristicReported Device PerformanceComments / Context
    Precision< 8.6 %Total precision based on multi-site, multi-lot testing.
    Linearity100 ± 20% mean dilution linearity; ≤ 10% non-linearity across 0.5-50.0 ng/mLBased on CLSI guideline EP6-A.
    Limit of Detection (LoD)0.12 ng/mLDetermined using CLSI guideline EP17-A.
    Limit of Quantitation (LoQ)0.21 ng/mLCorresponds to highest allowable imprecision of 17.78%.
    Interference (Endogenous)Mean assay specificity 100 ± 15% (for tested substances)No interference found for Triglycerides (30 mg/mL), Bilirubin (0.2 mg/mL), Hemoglobin (5 mg/mL), Total Protein (120 mg/mL).
    Interference (Chemotherapeutic drugs)No interference (for tested substances)No interference found for Carboplatin (500 µg/mL), Cisplatin (165 µg/mL), Dexamethasone (10 µg/mL), Doxorubicin (1.16 µg/mL), Leucovorin (2.68 µg/mL), Methotrexate (45 µg/mL), Paclitaxel (3.5 ng/mL).
    Interference (Clinical Conditions)Mean % recovery: HAMA 98%, RF 101%For CYFRA 21-1 spiked at ~5 and 25 ng/mL.
    Clinical Study Performance (Monitoring Lung Cancer Progression)(See details below for specific percentages at a >50% change threshold)
    - Concordance76% (239/314)Total concordance between CYFRA 21-1 change and disease status.
    - Sensitivity (>50% increase)45.9%Proportion of patients with disease progression that had elevated CYFRA 21-1.
    - Specificity (>50% increase)87.3%Proportion of patients without disease progression that had no elevation in CYFRA 21-1.
    - Positive Predictive Value (PPV) (>50% increase)57.4%Proportion of patients with elevated CYFRA 21-1 that had disease progression.
    - Negative Predictive Value (NPV) (>50% increase)81.3%Proportion of patients with no elevation in CYFRA 21-1 that did not have disease progression.

    Study Details

    The primary study mentioned proving the device meets its intended use is a retrospective clinical study focused on monitoring the course of disease in lung cancer patients.

    1. Sample size used for the test set and the data provenance:

      • Sample Size: 100 lung cancer patients, yielding a total of 314 observation pairs of serial serum samples. An average of 4.1 observations per patient.
      • Data Provenance: Retrospective clinical study. Patient samples were collected from a tertiary cancer center. The country of origin is not explicitly stated, but the submission is to the US FDA, implying the study was likely conducted in the US or under comparable standards.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • The document does not specify the number of experts or their qualifications used to establish the ground truth for disease progression/no progression. It states that "Changes in CYFRA 21-1 levels in serial serum samples... were compared to changes in disease status." This implies clinical assessment of disease status was the ground truth, likely determined by treating physicians or a review board at the tertiary cancer center, but no details are provided.
    3. Adjudication method for the test set:

      • The document does not describe an explicit adjudication method (like 2+1 or 3+1) for establishing disease status. The "changes in disease status" used for ground truth would typically be based on standard clinical practices (imaging, biopsy, clinical examination), but the specific process for confirming or adjudicating these statuses for the study is not outlined.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No, an MRMC comparative effectiveness study was not performed. This device is an in vitro diagnostic (IVD) assay (a blood test), not an imaging-based AI system that would typically involve human readers. Therefore, the concept of "human readers improve with AI vs without AI assistance" does not apply here.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Yes, a standalone performance was done in the sense that the assay results (CYFRA 21-1 concentrations) were directly compared to the established disease status without an intermediate "human-in-the-loop" interpretation of the assay values themselves for the purpose of the study. The device provides a quantitative measurement, and thresholds were applied to this measurement to determine its correlation with disease progression. However, the intended use states it is an "aid" in monitoring and "should be used in conjunction with other clinical methods," implying human interpretation within a broader clinical context. The performance metrics (sensitivity, specificity etc.) presented in the tables are for the assay's output as a standalone indicator of change.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The ground truth for the clinical study was based on "changes in disease status." This likely encompasses a combination of clinical assessments, imaging results, and possibly pathology reports over time, interpreted by clinicians at the tertiary cancer center. The specific methods used to define "disease progression" and "no progression" within the study are not detailed beyond "changes in disease status."
    7. The sample size for the training set:

      • The document does not provide information on a specific training set size. The clinical study described is the test set for assessing the final device's performance for monitoring. For an immunoassay, the "training" analogous to machine learning would be the development and optimization of the assay itself (antibody selection, reagent concentrations, calibration methods), which doesn't typically rely on a "training set" of patient data in the same way an AI algorithm does.
    8. How the ground truth for the training set was established:

      • As there's no explicitly defined "training set" in the context of an AI/ML algorithm development, this question is not directly applicable. The assay's analytical performance (precision, linearity, LoD, LoQ, interference) and the clinical utility (monitoring study) are reported, which are evaluated against established analytical standards and clinical outcomes, respectively.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1