Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K242302

    Validate with FDA (Live)

    Date Cleared
    2024-10-02

    (58 days)

    Product Code
    Regulation Number
    878.3300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    RejuvaKnee is intended for use in surgical procedures for the reinforcement and repair of the medial meniscus. In repairing and reinforcing medial meniscal defects, the patient must have an intact meniscal rim and anterior and posterior horns for attachment of the mesh. In addition, the surgically prepared site for the RejuvaKnee must extend at least into the red/white zone of the meniscus to provide sufficient vascularization.

    RejuvaKnee reinforces soft tissue and provides a reasonable scaffold that is replaced by the patient's own soft tissue. The RejuvaKnee is not a prosthetic device and is not intended to replace normal body structure.

    Device Description

    The RejuvaKnee™ Collagen Meniscus Implant is comprised primarily of bovine Type I collagen (nominally 99%) derived from meniscus. The device is provided in a semi-lunar shape with a triangular cross section to be used to reinforce weakened soft tissue and provide a resorbable scaffold that is replaced by the patient's own tissue.

    RejuvaKnee is supplied sterile and is intended for single use.

    AI/ML Overview

    This is a 510(k) summary for the RejuvaKnee™ Collagen Meniscus Implant.
    The device is a resorbable scaffold made of bovine Type I collagen intended for the reinforcement and repair of medial meniscal defects. It is compared to the Ivy Sports Medicine Collagen Meniscus Implant (K170364) as a predicate device.

    Here's an analysis of the provided information concerning acceptance criteria and supporting studies:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document provides a comparison of features between the RejuvaKnee™ Collagen Meniscus Implant and its predicate. Acceptance criteria are implied by the reported values and the claim of substantial equivalence to the predicate.

    FeatureAcceptance Criteria (Implied by Predicate & Device Performance)RejuvaKnee™ Collagen Meniscus Implant Performance
    Indications for UseSubstantially equivalent to predicateSame as predicate
    Material CompositionType I bovine collagenPrimarily Type I bovine collagen
    Material OriginBovineBovine
    Tissue SourceNot explicitly defined as an acceptance criterion; comparison shownMeniscus
    Physical FormSemi-lunar shape with triangular cross-sectionSemi-lunar shape with a triangular cross-section
    Product Sizes (cm)Comparable to predicateSmall Medial - 7.5 mm; Large Medial - 9 mm
    Porosity - SEMSufficient to allow for cellular integration from host tissue (>5 micron for RejuvaKnee)Provides sufficient porosity to allow for cellular integration from host tissue (>5 micron)
    pHComparable to predicate (predicate: 7.12)6 - 9.5
    Tensile Strength (kg/cm²)Comparable to predicate (predicate: 62.5 kg/cm²)≥ 150
    Suturability (kg)Comparable to predicate (predicate: 0.86 ±0.08 kg)≥ 0.5
    Thermal Stability (°C)Comparable to predicate (predicate: 61±1 °C)60 ± 3
    BiocompatibilityMeets ISO 10993 series of testingMeets ISO 10993 biocompatibility series of testing
    SterilitySterile, SAL 10-6 using Gamma irradiationSterile, SAL 10-6 using Gamma irradiation
    PyrogenicityNon-pyrogenic - <0.5EU/ml and ≤20.0EU/deviceNon-pyrogenic - <0.5EU/ml and ≤20.0EU/device
    Single Use/ReuseSingle use onlySingle use only

    2. Sample Size Used for the Test Set and Data Provenance:

    • Biocompatibility Testing: Conducted in accordance with ISO 10993-1. Specific sample sizes for each test within the series are not provided but would be standard for such evaluations. Data provenance is not specified but is typically from laboratory testing in a controlled environment.
    • Bench Testing: Testing included dimensions, porosity, pH, tensile strength, suturability, pyrogenicity, and thermal stability. "Each lot of finished devices is tested for bacterial endotoxin for lot release." Specific sample sizes per test are not provided but are routine for product characterization. Data provenance is from laboratory testing.
    • Animal Testing: The performance was evaluated in goat and canine models. The exact number of animals is not specified. Data provenance is from animal studies.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:

    This information is not applicable as the studies presented are non-clinical (bench and animal testing). There is no "ground truth" to be established by human experts in the context of diagnostic performance or clinical outcomes for this type of device.

    4. Adjudication Method for the Test Set:

    This information is not applicable as the studies presented are non-clinical.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No MRMC comparative effectiveness study was mentioned or performed. The studies are non-clinical and focus on demonstrating substantial equivalence of the material and physical properties of the device.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

    This information is not applicable as the device is a physical medical implant, not a software algorithm or AI-driven system.

    7. The Type of Ground Truth Used:

    • Bench Testing: The "ground truth" is established by direct measurement against quantitative specifications and comparison to the predicate device's reported characteristics.
    • Biocompatibility Testing: The "ground truth" is adherence to the requirements of the ISO 10993-1 standard.
    • Animal Testing: The "ground truth" is the observed biological response in the animal models, including safety, biocompatibility, tissue repair/remodeling, and biomechanical function, compared against expected outcomes and published performance of the predicate device.

    8. The Sample Size for the Training Set:

    This information is not applicable as the studies involve physical device testing, not machine learning or algorithm training.

    9. How the Ground Truth for the Training Set Was Established:

    This information is not applicable for the same reason as above.

    Ask a Question

    Ask a specific question about this device

    K Number
    K170364

    Validate with FDA (Live)

    Date Cleared
    2017-05-02

    (85 days)

    Product Code
    Regulation Number
    878.3300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Ivy Sports Medicine Collagen Meniscus Implant (CMI) is intended for use in surgical procedures for the reinforcement and repair of soft tissue injuries of the medial meniscus. In repairing medial meniscal defects, the patient must have an intact meniscal rim and anterior and posterior horns for attachment of the mesh. In addition, the surgically prepared site for the CMI must extend at least into the red/white zone of the meniscus to provide sufficient vascularization.

    The CMI reinforces soft tissue and provides a resorbable scaffold that is replaced by the patient's own soft tissue. The CMI is not a prosthetic device and is not intended to replace normal body structure.

    Device Description

    The Ivy Sports Medicine Collagen Meniscus Implant XL (CMI XL) is a resorbable collagen matrix comprised primarily of bovine type I collagen. The CMI is provided in a semi-lunar shape with a triangular cross section to be used to reinforce weakened soft tissue and provide a resorbable scaffold that is replaced by the patient's own tissue. The surgeon trims the device to the size necessary for repair of the damaged or weakened soft tissue. The CMI XL is provided sterile for single use only.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device called the Ivy Sports Medicine Collagen Meniscus Implant XL (CMI XL). It asserts substantial equivalence to a predicate device, the ReGen Collagen Scaffold (K082079).

    Based on the provided document, here's a breakdown of the requested information:

    1. Table of acceptance criteria and the reported device performance

    The document states: "Testing performed on the Collagen Meniscus Implant XL demonstrated that all of the predetermined acceptance criteria were met with passing results." However, it does not explicitly list the specific acceptance criteria or the numerical performance results against those criteria. It only makes a general statement about meeting them.

    2. Sample size used for the test set and the data provenance

    The document does not provide details on the sample size used for the test set or the data provenance. It mentions "Design verification testing" and "Bench performance test data" but no specifics on how many devices were tested or where the components/data originated.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable as the document indicates "Clinical evaluation is not required for this device." The testing performed appears to be bench testing rather than clinical studies requiring expert ground truth for interpretation.

    4. Adjudication method for the test set

    This information is not applicable as the document indicates "Clinical evaluation is not required for this device" and the testing described is bench performance testing.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable. The device is a physical collagen implant, not an AI-assisted diagnostic or interpretative tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable. The device is a physical collagen implant, and there is no algorithm or human-in-the-loop component described.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Since clinical evaluation was not required and the testing was described as "Design verification testing" and "Bench performance test data," the ground truth would likely be established by engineering specifications, material science standards, and comparison to the predicate device's established performance. The document implicitly suggests that the ground truth for performance relied on the predicate device's characteristics.

    8. The sample size for the training set

    This information is not applicable. The device is a physical implant, not an AI model that requires a training set.

    9. How the ground truth for the training set was established

    This information is not applicable. The device is a physical implant, not an AI model that requires a training set.

    In summary, the document focuses on demonstrating substantial equivalence through bench testing, asserting that the new device meets predefined, unlisted, acceptance criteria based on the predicate device's performance, and explicitly stating that clinical evaluation was not required.

    Ask a Question

    Ask a specific question about this device

    K Number
    K082079

    Validate with FDA (Live)

    Date Cleared
    2008-12-18

    (148 days)

    Product Code
    Regulation Number
    878.3300
    Age Range
    All
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ReGen Collagen Scaffold (CS) is intended for use in surgical procedures for the reinforcement and repair of soft tissue injuries of the medial meniscus. In repairing and reinforcing medial meniscal defects, the patient must have an intact meniscal rim and anterior and posterior horns for attachment of the mesh. In addition, the surgically prepared site for the CS must extend at least into the red/white zone of the meniscus to provide sufficient vascularization.

    The CS reinforces soft tissue and provides a resorbable scaffold that is replaced by the patient's own soft tissue. The CS is not a prosthetic device and is not intended to replace normal body structure.

    Device Description

    The ReGen Collagen Scaffold (CS) is a resorbable collagen matrix comprised primarily of bovine type I collagen. The CS is provided in a semi-lunar shape with a triangular cross section to be used to reinforce weakened soft tissue and provide a resorbable scaffold that is replaced by the patient's own tissue. The surgeon trims the device to the size necessary for repair of the damaged or weakened soft tissue.

    AI/ML Overview

    This document is a 510(k) summary for a medical device (ReGen Collagen Scaffold (CS)), not a study report detailing acceptance criteria and performance. Therefore, most of the requested information regarding study details, sample sizes, and expert qualifications cannot be extracted directly from this document. This document primarily focuses on establishing substantial equivalence to predicate devices.

    However, based on the text provided, I can infer and state the following:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state quantitative acceptance criteria or a direct table of reported device performance against such criteria. The "performance" is implicitly demonstrated through the claim of substantial equivalence to predicate devices based on biomechanical, biocompatibility, animal testing, and clinical studies.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    • Test Set Sample Size: The document mentions a prospective, randomized multicenter clinical trial that had two separately controlled and randomized arms:
      • Acute arm: 157 patients with no prior surgery to the involved meniscus.
      • Chronic arm: 154 patients with one to three prior treatments to the involved meniscus.
      • Total patients: 311.
    • Data Provenance: The study was a "multicenter clinical trial." The document does not explicitly state the country of origin, but given the submission to the FDA in the US, it is likely that the data originated from the US or included US sites. The trial was prospective and randomized.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    This information is not available in the provided text. The document refers to a clinical trial but does not detail how the "ground truth" (e.g., success or failure of the repair) was established by experts, nor does it specify the number or qualifications of such experts.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    This information is not available in the provided text.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable. The device is a "Collagen Scaffold" (surgical mesh), not an AI or imaging diagnostic device that would involve human readers or AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable for a physical surgical mesh device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    Given that it was a clinical trial for a surgical mesh, the "ground truth" would likely be based on clinical outcomes data related to the reinforcement and repair of soft tissue injuries of the medial meniscus. This would typically involve measures of healing, function, pain, and potentially re-injury rates over the follow-up period.

    8. The sample size for the training set

    This is not applicable. For a physical medical device like a surgical mesh, there isn't typically a "training set" in the context of machine learning. The clinical trial data served as the primary evidence.

    9. How the ground truth for the training set was established

    This is not applicable for a physical medical device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1