The X-ray system creates data for digital exposures in the maxillofacial area and in sub-areas for dentistry and pediatric dentistry and for hard tissue diagnostics within ENT medicine.
Primevision 3D is a digital panoramic and cone beam computed tomography (CBCT) x-ray device. A Computed tomography x-ray system (21 CFR 892.1750) is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. The subject Primevision 3D uses just one imaging detector to collect data for both 2D and 3D modes of operation. Additionally, the Primevision 3D does not offer a cephalometric mode.
N/A
Ask a specific question about this device
(166 days)
RCT700 is CBCT and panoramic x-ray imaging system with cephalometric. Which is intended to radiographic examination of the dento-maxillofacial, sinus, TMJ, Airway and ENT structure for diagnostic support for adult and pediatric patients. And a model scan is included as an option. Cephalometric image also includes wrist to obtain carpus images for growth and maturity assessment for orthodontic treatment.
The device is to be operated and used by dentists or other legally qualified heath care professionals
RCT700 provides 3D computed tomography for scanning hard tissues such as bone and teeth. By rotating the C-arm, which houses a high-voltage generator, an X-ray tube and a detector on each end, CBCT images of dental maxillofacial structures are obtained by recombining data scanned from the same level at different angles. Functionalities include panoramic image scanning for obtaining images of whole teeth, and a cephalometric option for obtaining cephalometric images.
The software of RCT700 saves the patient and image data and offers an inquiry function, in addition, supports the image generate function intended to obtain images using the RCT700 equipment and various sensors for diagnosis.
N/A
Ask a specific question about this device
(150 days)
The product is intended to produce X-ray Cone-Beam Computed Tomography, Panoramic tomography and Cephalometric (optional) images. The medical institutions can use the images for diagnostic purposes in oral and maxillofacial regions. The product is intended for use in hospitals and clinics, operated and used by trained professionals under the guidance of a physician.
The device is used for X-ray image diagnosis of oral and maxillofacial region in medical institutions through X-ray cone-beam computed tomography, panoramic and cephalometric photography. The device is intended for use in hospitals and clinics, operated and used by trained professionals under the guidance of a physician.
This device is divided into two models: Matrix 7000(Rubik X1), Matrix 7800(Rubik X3).
The device is composed of X-ray tube head, plate detector, control device, positioning aid, frame, cephalometric positioning shooting aid, workstation software, etc.
N/A
Ask a specific question about this device
(271 days)
It is intended to produce 2D (panoramic, cephalometric) or 3D (Cone Beam Computed Tomography) images. It provides diagnostic details of the dental-maxillofacial, TMJ and SINUS for adult and pediatric patients. The system also utilizes carpal images for orthodontic treatment.
GT300 & GT300-C is the Diagnostic computed tomography limited view field X-ray System which consists of image acquisition modes; panorama, cephalometric, and computed tomography. It is used to capture scanned image for obtaining diagnostic information for craniofacial surgery or other treatments.
This equipment is a device that generates and controls X-rays and consists of an X-ray generator, an X-ray control unit, an X-ray support unit, and some accessories (Chinrest, Exposure switch, Temple support, Carpus Plate).
GT300 and GT300-C have same intended use, essential design and manufacturing. The GT300 can acquire panoramic and CT images, while the GT300-C can acquire panoramic, CT, and Cephalo images. The only difference is the presence or absence of the Cephalo feature.
N/A
Ask a specific question about this device
(29 days)
Hasla is intended to produce two-dimensional digital x-ray images including panoramic and cephalometric image, and three-dimensional digital x-ray images of the dental, oral, maxillofacial region, at the direction of healthcare professionals as diagnostic support for adult and pediatric patients.
This device is not intended for use on patients less than approximately 21 kg (46 lb) in weight and 113 cm (44.5 in) in height; these height and weight measurements approximately correspond to that of an average 5 year old.
The product (Dental CBCT X-ray System, Model: Hasla) uses cone-beam computed tomography (CBCT) through X-ray cone-beam, panoramic radiography, cephalometric radiography to produce images of the dental, oral and maxillofacial areas to provides diagnostic details for the medical facilities. This product consists of Frame, X-ray generator (including Integrated X-ray source tube head, X-ray tube, Collimator), Image receptor and Image processing system (including Computer, Monitor, and Software workstation).
The provided FDA 510(k) clearance letter and summary for the Yian Medical Technology (Haining) Co., Ltd. Dental CBCT X-ray System (Model: Hasla) indicate that this is a Special 510(k) for a device modification. This means the company is asserting that the modified device is substantially equivalent to their own previously cleared predicate device (Dental Cone-beam Computed Tomography, Model: iDT901X1, K232710) and that the modifications do not raise any new safety or effectiveness concerns.
Therefore, the submission relies on the previous clearance and asserts that the current device meets the same acceptance criteria as the predicate without requiring new comprehensive studies to prove that from scratch. Instead, it demonstrates that the modifications themselves do not negatively impact the performance, safety, or efficacy previously established.
Given this context, here's a breakdown of the information requested. It's important to understand that the document does not report on a new study for the current device to explicitly prove it meets acceptance criteria with new data. Instead, it leverages the prior clearance through a demonstration of equivalence.
Acceptance Criteria and Device Performance (as related to the predicate device and the claim of equivalence)
The document states: "The subject device and the predicate device are equivalent in the model name, same in the indications for use, patient population, use environment, software functions, and software level of documentation, performance characteristics and technical specification." It further specifies that "Any difference between the predicate device and the subject device have no impact on safety or efficacy of the subject device and do not raise any new potential or increased safety risks."
Therefore, the "acceptance criteria" for the current device are implicitly met by demonstrating that its performance and safety characteristics are unchanged from the predicate device, which had previously met its own acceptance criteria during its clearance process (K232710).
The summary provides a detailed comparison table (Table 1) between the subject device (Hasla) and the predicate device (iDT901X1). Essentially, all listed specifications and performance characteristics are presented as identical between the two.
Table 1: Acceptance Criteria (Implied from Predicate Equivalence) and Reported Device Performance
| Feature/Acceptance Criteria Category (Implied) | Reported Performance (Subject Device is equivalent to Predicate) |
|---|---|
| Indications for Use | "Hasla is intended to produce two-dimensional digital x-ray images including panoramic and cephalometric image, and three-dimensional digital x-ray images of the dental, oral, maxillofacial region, at the direction of healthcare professionals as diagnostic support for adult and pediatric patients. This device is not intended for use on patients less than approximately 21 kg (46 lb) in weight and 113 cm (44.5 in) in height..." |
| Target Population | For patients 6 years and older |
| Anatomical Site | The dental, oral, maxillofacial region |
| Users | Health care professionals |
| Patient Contact Material | Meet ISO 10993 series standard (Non-sterile) |
| Tube Voltage | 60kV~120kV |
| Tube Current | 1 mA~20 mA |
| Nominal Focal Spot Size | 0.5 |
| Pulse Exposure function | Yes |
| Power | Frequency: 50/60Hz, Voltage: 110-120VAC, Power: 3.0KVA |
| Detector | FPD(TFT), material: CsI |
| Pixel Size | 200µm x 200µm (With binning) (CT, Cephalometric); 100µm x 100µm(Without binning)(CT, Panoramic) |
| Pixel Number | 15041248 (With binning) (CT, Cephalometric); 8002496 (Without binning) (CT); 300865 (Without binning) (Panoramic); 15041248 (With binning) (Cephalometric) |
| Size of Area Receiving X-ray | 300.8mm x 249.6mm (CT, CEPH); 80mm x 249.6mm (CT); 300.8mm x 6.5mm (PANO) |
| Spatial Resolution | 5lp/mm |
| Number of Bits | 16bits |
| SID/SOD | SID: 750mm/490mm |
| Dimension (WxDxH) | Max.overall dimension 1300mm1272mm2365mm |
| Weight | 233kg±5kg |
| Imaging Mode | CBCT, PANO, CEPH |
| Panoramic Scan Performance (Scan Time) | 15sec |
| Cephalometric Radiography (Scan Time) | 2.3sec |
| CT Scan Time | 22sec |
| CT FOV (Voxel Size) | Diameter 230 mm x 180mm (0.45mm); Diameter 150 mm x 120mm (0.29mm); Diameter 80 mm x 80mm (0.15mm); Diameter 50 mm x 50mm (0.097mm); Diameter 50 mm x 50mm (0.071mm) |
| Software functions | Image acquisition, data management, image display, image processing, System Settings |
| Software level of documentation | Basic Documentation Level |
| Compliance with Electrical Safety Standards | ANSI/AAMI ES60601-1, IEC 60601-1, IEC60601-1 Edition 3.2 2020-08, ANSI/AAMI ES: 60601-1:2005/A2:2010/AMD2:2021 |
| Compliance with Electromagnetic Compatibility Standards | IEC 60601-1-2 |
| Compliance with Radiation Safety Standards | IEC 60601-1-3 |
| Compliance with Electrical Equipment Usability Safety | IEC 60601-1-6 |
| Compliance with Usability Engineering Standard | IEC 62366 |
| Compliance with Software Lifecycle Process Standard | IEC 62304 |
| Compliance with Essential performance of dental extra-oral X-ray equipment Standard | IEC 60601-2-63 |
| Compliance with Acceptance tests of Imaging performance of computed tomography X-ray equipment Standard | IEC 61223-3-4; IEC 61223-3-7 |
| Compliance with Laser Safety standard | IEC 60825-1 |
| Compliance with Risk Management Standard | ISO 14971 |
| Compliance with DICOM Standard | ISO 12052:2017 |
| Compliance with Biocompatibility Standard | ISO 10993-1, ISO 10993-5, ISO 10993-10, ISO 10993-23 |
Study Details (Based on the provided document for the current 510(k) submission)
It's crucial to reiterate that this is a Special 510(k). The document explicitly states: "The modifications of the device have no impact on the product's design, working principle, software function, software documentation level, performance characteristic and technical specification indication for use, patient population, environment of use, user needs, thus no impact on the previously conducted Bench Testing and validation. No new bench testing or validation was needed. The previously conducted and submitted verification and validation test results are still valid for the modified device..."
-
Sample size used for the test set and the data provenance:
- Test Set Sample Size: Not applicable/Not specified as new data. The submission relies on the testing of the predicate device (K232710) and general compliance with standards.
- Data Provenance: Not applicable/Not specified, as no new clinical or standalone performance studies were explicitly conducted for this submission beyond bench testing to confirm compliance with recognized standards. The submitter is Yian Medical Technology (Haining) Co., Ltd. in China.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable. This submission does not detail a study involving expert readers or ground truth establishment, as it's a Special 510(k) relying on equivalence to a previously cleared device.
-
Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- Not applicable. No such study or adjudication method is described.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No. This device is a Dental CBCT X-ray System, which is an imaging acquisition device, not an AI-based diagnostic tool for human reader interpretation improvement. The document explicitly states: "The subject of this premarket submission, did not require clinical studies to support substantial equivalence."
-
If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- No. This device is an imaging system, not a standalone algorithm. The clearance is based on the device's ability to produce images compliant with established standards, not on an algorithm's diagnostic performance. The document explicitly states: "The subject of this premarket submission, did not require clinical studies to support substantial equivalence." The bench testing cited ("Bench Testing Summary") confirms the physical and technical specifications against standards (e.g., spatial resolution, focal spot size, etc.), which are standalone performance characteristics of the device (hardware/software combined for image generation), but not a "standalone algorithm" in the typical sense of AI performance.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- Not applicable. No new clinical studies requiring external ground truth were conducted for this Special 510(k). The evaluation revolves around compliance with engineering and radiation safety standards, and performance characteristics (like spatial resolution, scan time) derived from bench testing using phantoms or calibrated equipment.
-
The sample size for the training set:
- Not applicable. This is not an AI algorithm requiring a training set in the conventional sense. The "software functions" mentioned are for image acquisition, data management, display, processing, and system settings, which are intrinsic to the operation of the imaging device itself.
-
How the ground truth for the training set was established:
- Not applicable, as there is no specific training set described for an algorithm in this context.
Ask a specific question about this device
(174 days)
The Dental Cone Beam Computed Tomography System is intended to be used by medical institutions for X-ray image diagnosis on oral and maxillofacial areas through X-ray Cone Beam Computed Tomography (CBCT).
The system consists of X-ray tube assemblies, collimators, flat panel detector, control devices (the touch screen, the control panel, the console panel and the exposure switch), auxiliary positioning devices(the lifting seat and the chin rest) and U-shape gantry (the X-ray tube assembly support, the base and the Detector Support), and workstation. The system is an open design that allows patients to sit upright during a procedure. An electric powered seat is built into the scanner for proper patient positioning. The system has 2 configurations: the Configuration 1 has two X-ray tube assemblies(X-ray tube assembly A and X-ray tube assembly B) and two collimators(A and B), the Configuration 2 only has X-ray tube assembly A and collimator A. The proposed device is a CBCT dental system utilizes X-ray to obtain cross-sectional images of patient's oral and maxillofacial areas. The image software can reconstruct data and clearly display the 3D anatomical structure of the oral and maxillofacial areas, and achieve the multi-slice at any level, accurate measurement, contour plots and so on in the workstation software, greatly facilitating the clinical application of stomatology. The Lifting Seat moves up and down via buttons on the system Control Panel allowing positioning and adjustment of patients at different heights; the Touch Screen facilitates the adjustment of projection parameters and the control of actions of the device; the Gating and the shielding door are the safety interlock of the X-ray beam exposure to prevent the dose leakage caused by abnormal operation; the Exposure Switch controls the exposure output of the whole machine in real time during the scanning process. There are four FOV sizes; 20cm x 18cm high resolution is for diagnostic imaging of maxillary region, mandibular region, TMJ region and other craniofacial region, and it is only available for configuration 1; 17cm x 10.5cm high resolution , 8cm x8cm Super Resolution and 4cm x4cm Super Resolution are for diagnostic imaging of maxillary region and mandibular region. Under High Resolution mode, patients will receive lower radiation dose compared to Super Resolution. Dose radiation for each mode is different for pediatric and adult patients.
The provided text does not contain detailed information about the acceptance criteria or a specific study proving the device meets those criteria for AI performance. The document is a 510(k) clearance letter for a Dental Cone Beam Computed Tomography System, which is a medical imaging device, not an AI-powered diagnostic tool.
The "Clinical Performance Data" section mentions that "Sample clinical images obtained from the subject device were provided to demonstrate its imaging performance. An ABR-certified radiologist with over 5 years of experience post-certification independently reviewed the sample scans and evaluated essential image quality. The overall image quality was acceptable for all cases and image types in various scanning modes for both adult and pediatric applications."
However, this review is about the imaging performance of the CBCT system itself, not about the performance of an AI algorithm within the system. There are no mentions of:
- Specific AI acceptance criteria (e.g., sensitivity, specificity, AUC)
- A test set for AI
- Ground truth establishment methods for AI
- MRMC studies or standalone AI performance
- Training set details for AI
Therefore, I cannot fulfill your request for information related to AI acceptance criteria and study data based on the provided text. The document focuses on the substantial equivalence of the CBCT system as an imaging device to a predicate device, covering aspects like electrical safety, EMC, biocompatibility, and general image quality assessment by a human radiologist.
To provide the information you requested, the input document would need to be a 510(k) for an AI/ML-powered device, or indicate that the "Dental Cone Beam Computed Tomography System" includes an AI component whose performance was evaluated for clearance.
Ask a specific question about this device
(83 days)
RCT600 is CBCT and panoramic x-ray imaging system with cephalometric. Which is intended to radiographic examination of the dento-maxillofacial, sinus and TMJ structure for adult and pediatric patients. Cephalometric image is also includes wrist to obtain carpus images for growth and maturity assessment for orthodontic treatment. The device is to be operated and used by dentists or other legally qualified heath care professionals.
RCT600 provides 3D computed tomography for scanning hard tissues such as bone and teeth. By rotating the C-arm, which houses a high-voltage generator, an X-ray tube and a detector on each end, CBCT images of dental maxillofacial structures are obtained by recombining data scanned from the same level at different angles. Functionalities include panoramic image scanning for obtaining images of whole teeth, and a cephalometric option for obtaining cephalometric images.
The RCT600 device, for which K243903 is the 510(k) number, is a Computed Tomography X-Ray System used for radiographic examination of dento-maxillofacial, sinus, and TMJ structures. It also includes cephalometric imaging, specifically wrist images for growth and maturity assessment for orthodontic treatment.
The acceptance criteria are implicitly defined by the safety and effectiveness information provided in the 510(k) summary, specifically by demonstrating substantial equivalence to the predicate device, RCT700 (K213226). The testing performed aims to prove that despite some differences (deletion of one-shot Ceph option, different scan Ceph detector, changes in magnification, and scan times), the RCT600 maintains similar safety and effectiveness characteristics.
Here’s a breakdown of the information requested based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are based on demonstrating performance similar to the predicate device (RCT700, K213226) and meeting recognized international standards for medical electrical equipment and X-ray imaging devices.
| Acceptance Criteria Category | Specific Acceptance Criteria (Inferred from Substantial Equivalence and Standards Compliance) | Reported Device Performance (Summary from Submission) |
|---|---|---|
| Safety | Compliance with IEC 60601-1, IEC 60601-1-3, IEC 60601-1-6, IEC 60601-2-63 standards. | Complied with listed IEC standards for electrical, mechanical, and environmental safety. |
| EMC | Compliance with IEC 60601-1-2 standard. | Complied with IEC 60601-1-2 for EMC testing. |
| Software Safety | Compliance with FDA Guidance for "Content of Premarket Submissions for Device Software Functions" and "Cybersecurity in Medical Devices". Software classified as "Basic Documentation Level" of concern. | Validated according to FDA guidance. Risk analysis of software determined the differences do not affect safety and effectiveness. |
| Image Quality (Non-Clinical) | Performance metrics (MTF, DQE, NPS) demonstrating no significant differences from predicate according to IEC 61223-3-4 and IEC 61223-3-7. | All test results satisfactory, indicating safety and effectiveness. Key image quality performance metrics (MTF, DQE, and NPS) demonstrated no significant differences between the two devices. |
| Diagnostic Quality (Clinical) | Generation of images of diagnostic quality for all dental modalities. | Sample clinical images for each dental modality were submitted and reviewed by a licensed practitioner, who deemed them to be of acceptable quality for the intended use. |
| Intended Use | Device performs as intended for radiographic examination of dento-maxillofacial, sinus, TMJ, and wrist for growth assessment. | Clinical testing confirmed the features of RCT600 (CBCT, panoramic, cephalometric acquisitions) worked as intended. |
| Technological Characteristics | Similar technological characteristics to the predicate (imaging modes, X-ray source, materials). | Fundamental technological characteristics are similar (PANO, CEPH (Optional), CBCT). Changes in magnification and scan times were noted but deemed not to affect safety/effectiveness. |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Test Set: The document mentions "Sample clinical images" were submitted for each imaging dental modality and that features were "clinically tested." However, it does not specify the number of cases or patients used in the clinical imaging test set.
- Data Provenance: The document does not explicitly state the country of origin of the data. It mentions "Clinical images were gathered from all detectors of the RCT600 system." The applicant is Ray Co., Ltd., based in SOUTH KOREA. Given the context, it is reasonable to infer the data either originated from or was managed by the manufacturer. The study is prospective, as it involved gathering clinical images with the new device.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts
- Number of Experts: "A licensed practitioner" (singular) reviewed the sample clinical images.
- Qualifications of Experts: The expert was an "independent, licensed practitioner/clinician DDS/ABOMR."
- DDS stands for Doctor of Dental Surgery, indicating a dental professional.
- ABOMR stands for American Board of Oral and Maxillofacial Radiology, indicating specialization in dental and maxillofacial radiology. No years of experience are specified.
4. Adjudication Method for the Test Set
The document states, "A licensed practitioner reviewed the sample clinical images and deemed them to be of acceptable quality for the device intended use." This indicates a single-reader review without an explicit adjudication process defined for multiple readers.
5. If a Multi-reader Multi-case (MRMC) Comparative Effectiveness Study was done
- No, an MRMC comparative effectiveness study was not done. The document describes a "clinical testing" where sample images were reviewed by a single licensed practitioner to confirm diagnostic quality. There is no mention of human readers improving with or without AI assistance, as this is an imaging device, not an AI diagnostic algorithm.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
- This question is not directly applicable to the RCT600, as it is an X-ray imaging system, not an AI diagnostic algorithm. The "performance" being evaluated implicitly includes the algorithm within the imaging device that processes and reconstructs the images. The focus is on the imaging system's ability to produce diagnostically acceptable images for human interpretation. The clinical evaluation verifies the diagnostic quality of the images produced by the device, which is an assessment of its standalone imaging performance.
7. The Type of Ground Truth Used
The ground truth for the clinical image review was based on expert consensus (from the singular licensed practitioner) regarding the "acceptable quality for the device intended use" of the sample clinical images.
8. The Sample Size for the Training Set
The document does not specify a sample size for a training set. The device being described is an X-ray imaging system, not a machine learning algorithm that typically requires a separate training set. The "software" component mentioned underwent validation, but no details of training data for the software's inherent functions (e.g., image reconstruction algorithms) are provided, as these are typically validated through engineering and bench testing rather than a separate clinical "training set" in the context of device clearance.
9. How the Ground Truth for the Training Set was Established
As no explicit training set for an AI algorithm is mentioned, the concept of ground truth establishment for a training set is not applicable in the context of this device's submission description. The software validation relies on risk analysis and compliance with FDA guidance documents rather than a specific clinical training dataset and ground truth for learning.
Ask a specific question about this device
(144 days)
Green X 12 SE (Model : PHT-40CHS) is intended to produce panoramic, cephalometric, or 3D digital x-ray images. It provides diagnostic details of the dento-maxillofacial, sinus, and TMJ for adult and pediatric patients. The system also utilizes carpal images for orthodontic treatment. The device is to be operated by healthcare professionals.
Green X 12 SE (Model : PHT-40CHS) is an advanced 4-in-1 digital X-ray imaging system that incorporates PANO, CEPH(optional), CBCT and MODEL Scan imaging capabilities into a single system. Green X 12 SE (Model : PHT-40CHS), a digital radiographic imaging system, acquires and processes multi-FOV diagnostic images for dentists. Designed explicitly for dental radiography. Green X 12 SE (Model : PHT-40CHS) is a complete digital X-ray system equipped with imaging viewers, an X-ray generator and a dedicated SSXI detector. The digital CBCT system is based on a CMOS digital X-ray detector. The CMOS CT detector is used to capture 3D radiographic images of the head, neck, oral surgery, implant and orthodontic treatment. Green X 12 SE (Model : PHT-40CHS) can also acquire 2D diagnostic image data in conventional PANO and CEPH modes.
The provided document does not contain information regarding a study that proves the device meets specific acceptance criteria in the context of AI performance, clinical trials, or comparative effectiveness studies with human readers. The document describes a Computed Tomography X-Ray System named Green X 12 SE (PHT-40CHS) and its substantial equivalence to a predicate device (Green X 12 (PHT-75CHS)).
The performance data section primarily focuses on:
- Substantial Equivalence: Demonstrating that the new device is equivalent to the predicate device despite minor changes in detector technology and some feature deletions.
- Technical Performance Testing: Verification against international standards (IEC 61223-3-5) for general image quality indicators like Contrast, Noise, CNR, and MTF, and dosimetric performance (DAP).
- Safety and EMC: Compliance with relevant IEC standards for electrical, mechanical, environmental safety, and electromagnetic compatibility.
- Software Verification and Validation: Adherence to FDA guidance for device software functions.
Therefore, many of the requested categories for AI-related studies and acceptance criteria are not applicable based on the provided text.
Here's a breakdown of the available information:
1. Table of Acceptance Criteria and Reported Device Performance:
The document does not explicitly state "acceptance criteria" in a table format for AI performance. Instead, it refers to equivalence to a predicate device and compliance with international performance standards for Computed Tomography X-ray systems.
| Criterion Type (Inferred from Text) | Acceptance Criteria (Inferred) | Reported Device Performance |
|---|---|---|
| Image Quality | Performed equivalently to the predicate device according to IEC 61223-3-5. | Contrast, Noise, CNR, and MTF were measured and demonstrated that the subject device performed equivalently to the predicate device in general image quality. |
| Dosimetric Performance | DAP measurements should be the same as the predicate device under identical exposure conditions. | DAP measurements in PANO, CEPH, and CBCT modes were the same as the predicate device under the same X-ray exposure conditions (exposure time, tube voltage, tube current). |
| Software Functionality | Compliance with FDA guidance for device software functions. Software criticality assessed as "basic documentation." | Software verification and validation were conducted and documented as recommended by FDA guidance. The Green X 12 SE provides EzDent-i (K241114) for 2D viewing and Ez3D-i (K231757) for 3D viewing, both previously cleared. |
| Safety and EMC | Compliance with IEC 60601-1:2005+AMD1:2012+AMD2:2020, IEC 60601-1-3:2008+AMD1:2013+AMD2:2021, IEC 60601-2-63:2012+AMD1:2017+AMD2:2021, and IEC 60601-1-2:2014+AMD1:2020. | Electrical, mechanical, environmental safety, and EMC testing were performed per specified IEC standards. All test results were satisfactory. The device also conforms to NEMA PS 3.1-3.18 (DICOM Set). |
| Manufacturing Standards | Conformance with 21 CFR Part 1020.30, 1020.31, 1020.33, and 21 CFR 820.30. | The manufacturing facility is in conformance with relevant EPRC standards. Adequate design and development controls (according to 21 CFR 820.30) were in place. |
2. Sample size used for the test set and the data provenance: Not applicable. The document discusses bench testing and comparison to a predicate device, not a test set of patient cases for AI evaluation.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. This information is relevant for AI (CAD) performance studies, which are not described here.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: No MRMC study was described. The device is an X-ray imaging system, not an AI-assisted diagnostic tool in the sense of a CADe/CADx system.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. The device is a physical imaging system.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable in the context of AI performance. For the technical performance aspects, the "ground truth" would be established by the specifications and measurements according to IEC standards.
8. The sample size for the training set: Not applicable. This device is an imaging system, not an AI algorithm requiring a training set in the conventional sense.
9. How the ground truth for the training set was established: Not applicable.
Ask a specific question about this device
(137 days)
Green X 21 (Model: PHT-90CHO) is intended to produce panoramic, cephalometric, or 3D digital Xray images. It provides diagnostic details of the dento-maxillofacial, sinus, TMJ, and ENT for adult and pediatric patients. The system also utilizes carpal images for orthodontic treatment. The device is to be operated by healthcare professionals.
Green X 21 (Model : PHT-90CHO) is an advanced 6-in-1 digital X-ray system specifically designed for 2D and 3D dental radiography. This system features six imaging modalities: PANO, CEPH (optional), DENTAL CT, ENT CT, MODEL, and FACE SCAN, all integrated into a single unit.
Green X 21's imaging system is based on digital TFT detectors, accompanied by imaging viewers and an X-ray generator.
The materials, safety characteristics, X-ray source, indications for use and image reconstruction/MAR(Metal Artifact Reduction) algorithm of the subject device are same as the predicate device (PHT-75CHS (K210329)). The subject device differs from the predicate device in the following ways: It is equipped with new X-ray detectors for CT/PANO and CEPH modification results in a different maximum FOV provided in a single scan for CT mode compared to the predicate device. For CEPH modality, the subject device utilizes a one- shot imaging capture method.
Additionally, the subject device includes new modalities such as ENT CT and FACE SCAN with a face scanner, along with new software functions, including Auto Pano, Smart Focus, and Scout.
The provided document describes a 510(k) premarket notification for the "Green X 21 (PHT-90CHO)" digital X-ray system. The core of the submission is to demonstrate substantial equivalence to a predicate device, the "Green X 18 (PHT-75CHS)". The document focuses heavily on comparing technical specifications and performance data to support this claim, particularly for the new components and features of the Green X 21.
However, it's crucial to note that this document does not describe a study that proves the device meets specific acceptance criteria in the context of an AI/human-in-the-loop performance study. Instead, it focuses on engineering performance criteria related to image quality and safety for an X-ray imaging device, and comparing these to a predicate device. The "acceptance criteria" discussed are primarily technical specifications and performance benchmarks for the X-ray system components (detectors, imaging quality metrics), rather than clinical performance metrics in disease detection with AI assistance.
Therefore, many of the requested items (e.g., sample size for test set, number of experts for ground truth, adjudication method, MRMC study, standalone performance) are not applicable or not detailed in this submission because the device in question is an X-ray imaging system, not an AI-based diagnostic tool. The "new software functions" mentioned (Auto Pano, Smart Focus, Scout) are described as image reconstruction or manipulation tools, not AI algorithms for clinical diagnosis.
Here's a breakdown based on the information available, addressing the points where possible and noting when information is absent or not relevant to the provided text:
Acceptance Criteria and Reported Device Performance
Given that this is an X-ray imaging device and not an AI diagnostic tool, the acceptance criteria are generally related to image quality, safety, and functional performance, benchmarked against standards and the predicate device.
Table of Acceptance Criteria and Reported Device Performance (as inferred from the document):
| Acceptance Criteria Category | Specific Criteria (Inferred) | Reported Device Performance (Green X 21) |
|---|---|---|
| I. Imaging Performance (New X-ray Detectors) | ||
| 1. CT/PANO Detector (Jupi1012X) | - Modulation Transfer Function (MTF) & Detective Quantum Efficiency (DQE) & Noise to Power Spectrum (NPS): Performance comparable or superior to predicate (Xmaru1524CF Master Plus OP). | - MTF (CT/PANO): Jupi1012X showed more stable or superior performance for DQE, MTF, and NPS, particularly better stability in high-frequency regions. Jupi1012X could distinguish up to 3.5 line pairs (MTF 10% criterion), compared to 2.5 line pairs for predicate. |
| - Pixel Size: Similar to predicate device. | - Pixel Size (CT/PANO): "Very similar" to predicate. Image test patterns demonstrated test objects across the same spatial frequency range without aliasing. | |
| 2. CEPH Detector (Venu1012VD) | - MTF, DQE, NPS: Performance comparable or superior to predicate (Xmaru2602CF), despite predicate having lower NPS (noise). | - MTF (CEPH): Venu1012VD exhibited superior performance in DQE, MTF, and NPS (except predicate's better NPS due to lower noise). Higher MTF values indicate sharper images. |
| - Pixel Size: Similar to predicate device. | - Pixel Size (CEPH): Similar to predicate's 100 µm (non-binning). Image test patterns demonstrated test objects across the same spatial frequency range without aliasing. | |
| 3. Overall Diagnostic Image Quality | - Equivalent to or better than predicate device. | - "Equivalently or better than the predicate device in overall image quality." |
| II. Compliance with Standards and Guidelines | ||
| 1. IEC 61223-3-5 (CT) | - Quantitative testing for noise, contrast, CNR, MTF 10%. | - All parameters met the standards specified. |
| 2. IEC 61223-3-4 (Dental X-ray) | - Quantitative assessment for line pair resolution and low contrast performance (PANO/CEPH). | - All parameters met the standards specified. |
| 3. Software/Firmware (Basic Documentation Level) | - Adherence to FDA Guidance "Content of Premarket Submissions for Device Software Functions." | - Software verification and validation conducted and documented. |
| 4. Safety & EMC | - Compliance with IEC 60601-1, IEC 60601-1-3, IEC 60601-2-63 standards for electrical, mechanical, environmental safety and performance; IEC 60601-1-2 for EMC. | - Testing performed and results were satisfactory. |
| III. Functional Equivalence / Performance of New Modalities/Functions | ||
| 1. ENT CT Modality | - Meet image quality standards of Dental CT; limit radiation exposure to ENT region. | - "Adheres to the same image quality standards as the Dental CT modality." Specifically designed to limit radiation exposure to ENT region. |
| 2. FACE SCAN Modality | - Intended for aesthetic consultations, not clinical diagnostic; meets internal criteria. | - "Not designed for clinical diagnostic purposes." "Meets the internally established criteria and has been designed to perform as intended." |
| 3. Auto Pano, Smart Focus, Scout Modes | - Should function as intended, similar to previously cleared devices (Green X 12, K231796). | - Evaluated according to IEC 61223-3-4 and IEC 61223-3-5; "both standard requirements were met." |
| IV. Dosimetric Performance (DAP) | ||
| 1. PANO Modality DAP | - Similar DAP to predicate device under same exposure conditions. | - "DAP measurement results are similar" when tested under same exposure conditions (High Resolution Mode). |
| 2. CEPH Modality DAP | - Performance balance between increased DAP (due to one-shot type) and reduced exposure time/motion artifacts. | - DAP "more than twice that of the predicate device" (due to one-shot vs. scan-type), but "utilizes a one-shot type, operating with approximately one-fourth the exposure time... This reduces motion artifacts." |
| 3. CT Modality DAP | - Overall DAP performance balanced with FOV and image quality. | - "Slight increase in DAP for the subject device" for most comparable/equivalent FOVs. However, "maximum FOV provided by the subject device demonstrated a reduced radiation dose compared to the predicate device." |
Study Details (as far as applicable and available)
-
Sample size used for the test set and the data provenance:
- Test Set Description: The "test set" in this context is referring to data collected through bench testing using phantoms for image quality assessment, and potentially clinical images for subjective comparison, rather than a clinical trial cohort.
- Sample Size: Not specified in terms of number of patient images. The testing was conducted on the device itself using phantom studies (e.g., test patterns for MTF, line pairs for resolution, phantoms for low contrast).
- Data Provenance: The bench tests were conducted in a laboratory ("in a laboratory using the same test protocol as the predicate device"). The "Clinical consideration" section for image quality evaluation implies some clinical image review, but the origin (e.g., country, retrospective/prospective) of these potential clinical images is not detailed.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not Applicable in the traditional sense of an AI study. Ground truth for an X-ray imaging device's performance is typically established through quantitative measurements using phantoms and technical specifications, not expert consensus on medical findings.
- The document mentions "Image Quality Evaluation Report and Clinical consideration" and concludes "the subject device performed equivalently or better than the predicate device in overall image quality." This implies some form of expert review for subjective image quality, but the number or qualifications of these "experts" are not detailed.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Not Applicable. As noted above, this is not an AI diagnostic study with a human-in-the-loop component requiring adjudication of disease findings. The evaluations are primarily technical and quantitative measurements comparing physical properties and image output.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No. An MRMC study was not described. This is an X-ray imaging system, not an AI diagnostic software that assists human readers. The new software functions (Auto Pano, Smart Focus, Scout) described are image manipulation/reconstruction features, not AI for diagnostic assistance.
-
If a standalone (i.e. algorithm only, without human-in-the-loop performance) was done:
- Not applicable in the context of an AI algorithm. The device is an X-ray system; its "standalone" performance refers to its ability to produce images according to technical specifications, which was assessed through bench testing (MTF, DQE, NPS, etc.). There is no mention of a diagnostic AI algorithm that operates standalone.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- Quantitative Phantom Measurements and Technical Specifications: For image quality (MTF, DQE, NPS, line pair resolution, contrast, noise, CNR), the ground truth is established by physical measurements using standardized phantoms and reference values as per IEC standards.
- Predicate Device Performance: A key "ground truth" for substantial equivalence is the established performance of the predicate device. The new device's performance is compared against this benchmark.
- Internal Criteria: For functionalities like FACE SCAN, "internally established criteria" were used as a benchmark.
-
The sample size for the training set:
- Not Applicable. This document describes an X-ray imaging device, not an AI model that requires a training set. The software functions mentioned (Auto Pano, Smart Focus, Scout) are described as computational algorithms for image reconstruction or enhancement, not machine learning models that learn from a training dataset.
-
How the ground truth for the training set was established:
- Not Applicable. No training set for an AI model is described.
Ask a specific question about this device
(114 days)
FinScan F350 is CBCT, panoramic and cephalometric x-ray imaging system. The device is intended to radiographic examination of the dento-maxillofacial and TMJ structure for diagnostic support for adult and pediatric patients. Cephalometric image also includes wrist to obtain carpus images for growth and maturity assessment for orthodontic treatment.
The device is to be operated and used by dentists or other legally qualified health care professionals.
Dental Cone-beam Computed Tomography (model: FinScan F350) uses cone-beam computed tomography (CBCT) technology through X-ray cone-beam, panoramic radiography, cephalometric radiography to produce images of the dento-maxillofacial and TMJ structure to provides diagnostic details for dental clinics and dental hospitals.
This product mainly consists of column, rotating unit, tube head assembly, CBCT/panoramic detector, cephalometric detector, the workstation and image processing software.
The provided text describes the 510(k) premarket notification for the Dental Cone-beam Computed Tomography (model: FinScan F350) and its substantial equivalence to a predicate device. However, the document does not contain explicit acceptance criteria or a comprehensive study demonstrating that the device meets specific performance criteria in a quantitative manner as typically expected for device performance studies.
The document focuses on regulatory compliance, outlining how the FinScan F350 is substantially equivalent to the predicate device RCT700 (K213226) based on technological characteristics and adherence to recognized standards. It mentions "Performance bench testing" and "Clinical images acquired using FinScan F350 were evaluated by two US board certified experienced dentists to be of acceptable clinical effectiveness," but it does not provide specific metrics or the study details requested.
Here's an analysis based on the available information, noting what is not present:
1. Table of Acceptance Criteria and Reported Device Performance:
The document mentions performance bench testing that measured spatial resolution, CNR (Contrast-to-Noise Ratio), and homogeneity for CT image quality. It states that "The results demonstrate that the subject device is as effective, and perform as well as the predicate device." However, specific numerical acceptance criteria or the reported performance values for these metrics are NOT provided in the text.
2. Sample Size Used for the Test Set and Data Provenance:
The document briefly mentions "Clinical images acquired using FinScan F350 were evaluated..." but does not specify the sample size (number of images or cases) used for this evaluation, nor does it provide information on the data provenance (e.g., country of origin, retrospective or prospective nature of the image acquisition).
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:
"Clinical images acquired using FinScan F350 were evaluated by two US board certified experienced dentists to be of acceptable clinical effectiveness for the proposed indications for use."
- Number of experts: 2
- Qualifications: US board certified experienced dentists. (Specific years of experience are not mentioned).
4. Adjudication Method for the Test Set:
The document states "evaluated by two US board certified experienced dentists," but it does not specify an adjudication method (e.g., whether consensus was required, if a third expert was used in case of disagreement, or if individual evaluations were aggregated).
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:
No, an MRMC comparative effectiveness study is NOT mentioned in the provided text. The document describes a substantial equivalence submission, which typically focuses on demonstrating that a new device is as safe and effective as a legally marketed predicate device, rather than quantifying performance improvements with AI assistance. The device is described as an imaging system, not an AI-based diagnostic tool that assists human readers.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done:
The FinScan F350 is a medical imaging device (CBCT, panoramic, cephalometric x-ray system) that produces images for diagnostic support. It includes "image processing software." The performance evaluation mentioned (bench testing and clinical image evaluation by dentists) is about the general image quality and clinical effectiveness of the imaging system itself. The concept of "standalone algorithm performance" (without human-in-the-loop) as typically applied to AI/CADe systems is not directly applicable or discussed for this device based on the provided text, as it's an image acquisition system, not primarily a standalone diagnostic algorithm.
7. The Type of Ground Truth Used:
For the "clinical images evaluation report," the ground truth seems to be implicitly expert judgment/consensus of the two US board-certified experienced dentists, who deemed the image quality "acceptable clinical effectiveness." Pathology or outcomes data are not mentioned as ground truth sources for this evaluation.
8. The Sample Size for the Training Set:
The document does not mention any training set size. This is expected as the document describes a hardware imaging device with associated standard image processing software, not a machine learning or AI algorithm that requires a "training set" in the context of deep learning. The software mentioned (FinScan and FinScanFW) performs control, image acquisition, reconstruction, storage, browsing, and processing, which are typically deterministic software functions, not machine learning models.
9. How the Ground Truth for the Training Set was Established:
As no training set is mentioned as applicable to this type of device (traditional imaging system with processing software, not an AI/ML algorithm), this information is not applicable and not provided.
Ask a specific question about this device