510(k) Data Aggregation

    K Number
    K221851

    Validate with FDA (Live)

    Date Cleared
    2023-12-08

    (529 days)

    Product Code
    Regulation Number
    872.3930
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Lumina-PTFE Titanium is a temporarily implantable material (non-resorbable) indicated for use as a space-making barrier in the treatment of alveolar and periodontal bony defect sites.

    Device Description

    Lumina-PTFE Titanium is a synthetic, biocompatible, single-use, sterile polymeric barrier composed of a 100% polytetrafluoroethylene (PTFE) reinforced by a titanium structure. The titanium frame is embedded between two layers of PTFE. This medical device is placed between bone grafts and the periosteum in dental bone grafting procedures to stabilize and support the bone graft. The PTFE membrane helps create the space needed for bone-derived cells to repopulate and repair the defect. The PTFE membranes are designed to maintain space and conform to tissue contours. Lumina-PTFE Titanium is provided in thicknesses of 0,25mm and different sizes.

    AI/ML Overview

    The provided text is a 510(k) premarket notification for a medical device called "Lumina-PTFE Titanium." This document does not describe an AI or software device that requires acceptance criteria related to a study proving performance against a ground truth, or an MRMC study.

    Instead, the document focuses on demonstrating substantial equivalence to a previously cleared predicate device (Cytoplast™ Titanium-Reinforced PTFE Membranes). This is achieved by comparing the device's technological characteristics and presenting non-clinical performance data (e.g., biocompatibility tests, sterilization validation, material property tests like absorption, wettability, tension break, elasticity, suture pullout strength, and peel resistance).

    Therefore, I cannot provide the requested information about acceptance criteria, study design (sample size, data provenance, expert adjudication, MRMC, standalone performance), and ground truth establishment, as these are not relevant to the type of device and submission described in the input.

    The entire submission is dedicated to proving that "Lumina-PTFE Titanium" shares the same indications for use, principle of operation, design, composition, and other technical characteristics as its predicate, and that any differences do not raise new questions of safety or effectiveness. No AI or algorithm performance metrics are discussed.

    Ask a Question

    Ask a specific question about this device

    K Number
    K222549

    Validate with FDA (Live)

    Device Name
    OpenTex
    Date Cleared
    2023-05-11

    (261 days)

    Product Code
    Regulation Number
    872.3930
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    OpenTex is a temporarily implantable material (non-resorbable) for as a space-making barrier in the treatment of periodontal defects.

    Device Description

    OpenTex is a non-resorbable PTFE membrane composed of proprietary 100% polytetrafluoroethylene (PTFE) sheet with inert biological features and predictable barrier effect. OpenTex is designed to function as a physical barrier to avoid gingival cell invasion, thus providing a favorable environment for neovascularization and bone derived cells to repopulate and repair the defect. Since space-making is critical to this procedure, the membrane is sufficiently stiff to prevent spontaneous collapse but supple enough to conform easily to tissue contours. OpenTex is supplied sterile and intended for single use only. It is available in various sizes.

    AI/ML Overview

    This is a 510(k) summary for a medical device called OpenTex, a non-resorbable PTFE membrane for guided tissue regeneration in periodontal defects. The information provided is for regulatory clearance and does not involve AI or algorithms with human-in-the-loop performance. Therefore, many of the requested fields are not applicable.

    Here's an analysis of the provided text based on your request:

    1. Table of Acceptance Criteria and Reported Device Performance

    Test ItemTest MethodAcceptance CriteriaReported Device Performance (OpenTex)Predicate Device Performance (Salvin CytoSurg™)Remarks
    pHpH measurement testDifference between blank and extracts < 1.56.23 (pH)6.21 (pH)Results met criteria and were comparable.
    Dissolution / SolubilityEvaporation residue testEvaporation residue of extraction liquid $\leq$ 1.0 mg0 mg0 mgResults met criteria and were comparable.
    Tensile StrengthASTM D882$\geq$ 34 MPa (Internal)72.1524 MPa33.8444 MPaMet criteria and was higher than predicate.
    Tear ResistanceISO 6383-1$\geq$ 1.5 kgf/mm (Internal)5.9254 kgf/mm1.288 kgf/mmMet criteria and was higher than predicate.
    Suture Retention StrengthANSI/AAMI/ISO 7198Equivalent to result of predicate device. (There isn't a numerical criterion, but rather a comparative one.)1.083 MPa0.842 MPaHigher than predicate device.
    BiocompatibilityISO 10993 seriesCompliance with ISO 10993-1, -3, -5, -6, -10, -11Evaluated and presumably metNot explicitly stated; assumed compliant as predicate.Evaluated for Cytotoxicity, Irritation, Sensitization, Genotoxicity, Acute Toxicity, Subacute Toxicity, Implantation, Material mediated pyrogenicity.
    SterilizationISO 11135Sterility Assurance Level (SAL) of 10^-6SAL of 10^-6SAL of 10^-6Demonstrated compliance with standard.

    2. Sample size used for the test set and the data provenance

    The document describes non-clinical bench testing comparing the physical and chemical properties of OpenTex to its predicate device. This type of testing typically uses a small number of samples (often 3-5 per test) to evaluate material properties under controlled laboratory conditions, rather than a "test set" in the context of AI or clinical trials. The specific sample sizes for each test are not explicitly provided in the document.

    The data provenance is from laboratory testing performed by the manufacturer, Purgo Biologics Inc. The country of origin for the data is implicitly South Korea, where the company is located. The data is non-clinical performance data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This question is not applicable as the document describes bench testing of physical and chemical properties, not a clinical study involving human reads or expert ground truth establishment for diagnostic purposes. The "ground truth" here is the adherence to established international standards and internal criteria for material performance.

    4. Adjudication method for the test set

    This question is not applicable as there is no clinical test set requiring adjudication by experts.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This question is not applicable. The device is a physical bone grafting material (membrane), not an AI-assisted diagnostic or therapeutic tool. No MRMC study was performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not applicable. The device is a physical medical device, not an algorithm.

    7. The type of ground truth used

    The "ground truth" for the performance tests is based on established international standards (e.g., ASTM D882, ISO 6383-1, ANSI/AAMI/ISO 7198, ISO 10993 series, ISO 11135) and internal criteria set by the manufacturer to demonstrate substantial equivalence to the predicate device.

    8. The sample size for the training set

    This question is not applicable. There is no "training set" as this is a physical medical device for which performance is evaluated through non-clinical testing, not a machine learning model.

    9. How the ground truth for the training set was established

    This question is not applicable for the same reasons as #8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K211554

    Validate with FDA (Live)

    Date Cleared
    2023-02-24

    (646 days)

    Product Code
    Regulation Number
    872.3930
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    InnoGenic™ Non-resorbable membrane is a temporarily implantable material (non-resorbable) indicated for stabilization and support of bone grafts in alveolar bony defect sites.

    Device Description

    Not Found

    AI/ML Overview

    The provided text is a 510(k) clearance letter from the FDA for a medical device called "InnoGenic™ Non-resorbable Membrane". This document does not contain any information regarding acceptance criteria, device performance studies, sample sizes, expert qualifications, or ground truth establishment related to an AI/ML powered medical device.

    The document states that the device is a "temporarily implantable material (non-resorbable) indicated for stabilization and support of bone grafts in alveolar bony defect sites." It is a Class II device (Bone Grafting Material, 21 CFR 872.3930).

    Therefore, I cannot fulfill your request to describe the acceptance criteria and the study that proves the device meets the acceptance criteria, as the necessary information is not present in the provided text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K210797

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-08-18

    (520 days)

    Product Code
    Regulation Number
    872.3930
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Bio-MEM Ti Reinforced membranes are an implantable temporary non-resorbable device (membrane) for use as a spacer creation barrier in the treatment of local defects in the oral cavity in conjunction with tissue regeneration or augmentation. Bio-MEM Ti Reinforced membranes are intended to be submerged and clinically implanted more than 30 days and no longer than 6 months.

    Device Description

    Bio-MEM Ti Reinforced membranes are non-resorbable membranes available as Tireinforced configured of a highly shapeable Ti (Reinforced only) enveloped by an embossed polytetrafluoroethylene (PTFE) membrane. Extensive documentation demonstrates tissue compatibility of PTFE. Bio-MEM Ti Reinforced membranes are designed to avoid ingrowth of gingival soft tissue into bony defects, in order to facilitate the neovascularization and bone formation during the repair process of the defect. Bio-MEM Ti Reinforced membranes are provided STERILE and in different anatomical shapes and sizes. The titanium reinforcement is intended for space creating and shapemaintaining which minimize movements and subsequent exposure during the regenerative healing.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the Bio-MEM Ti Reinforced Membrane.

    Note: The provided text is a 510(k) summary for a medical device (Bio-MEM Ti Reinforced Membrane). 510(k) submissions typically demonstrate substantial equivalence to a predicate device rather than presenting acceptance criteria and studies in the traditional scientific sense for a novel AI/software device. Therefore, the information below will reflect that context. The questions asked are more aligned with AI device studies, and the response will clarify what information is available given the document type.

    1. Table of Acceptance Criteria and Reported Device Performance

    Strict "acceptance criteria" and "reported device performance" as one might see for a diagnostic AI algorithm are not explicitly stated in this 510(k) summary. Instead, the document focuses on demonstrating that the Bio-MEM Ti Reinforced Membrane is substantially equivalent to its predicate devices based on design, materials, technological characteristics, and performance testing against recognized standards.

    The performance testing sections describe the studies undertaken to ensure the device's safety and functionality, which implicitly means the device met the requirements of these tests.

    Aspect TestedStudy TypeCriteria (Implicit)Performance (Reported)
    Strength & IntegrityBend and Delamination TestingDevice maintains structural integrity and functionality after bending and resists delamination.Confirmed the strength, integrity, and functionality of the device. (Specific quantitative results not provided, but implies pass)
    SterilizationSterilization Validation (EO Gas)Device is sterile and maintains sterility over shelf-life. Meets ISO 11135[2014] standards.Validation conducted to support EO sterilization. (Implies pass)
    Packaging Shelf-lifePackaging ValidationPackaging maintains sterility and integrity for the specified shelf-life. Meets ISO, ASTM standards.Validation conducted to support packaging shelf-life of 5 years. (Implies pass)
    BiocompatibilityCytotoxicity Testing (ISO 10993-5)No unacceptable cytotoxic effects.Cytotoxicity testing conducted. (Implies pass)
    BiocompatibilitySensitization Testing (ISO 10993-10)No unacceptable sensitization.Sensitization testing conducted. (Implies pass)
    BiocompatibilityIrritation Testing (ISO 10993-10)No unacceptable irritation.Irritation testing conducted. (Implies pass)
    BiocompatibilityAcute Systemic Toxicity Testing (ISO 10993-11)No unacceptable acute systemic toxicity.Acute systemic toxicity testing conducted. (Implies pass)
    BiocompatibilityMaterial-Mediated Pyrogenicity (USP 29 <151>)No pyrogenic response.Material-mediated pyrogenicity testing conducted. (Implies pass)
    BiocompatibilityGenotoxicity Testing (AMES, Micronucleus - ISO 10993-3)Not genotoxic.Genotoxicity (AMES and micronucleus) testing conducted. (Implies pass)
    BiocompatibilityImplantation Testing (ISO 10993-6)Acceptable local effects after implantation.Implantation testing conducted. (Implies pass)
    Ethylene Oxide ResidualsTesting against ISO 10993-7Ethylene oxide residues below acceptable limits.Testing conducted against ISO 10993-7. (Implies pass)
    Material Properties (PTFE)Testing against ASTM D4895 - 10PTFE material meets standard specifications.Testing conducted against ASTM D4895 - 10. (Implies pass)

    2. Sample Size Used for the Test Set and Data Provenance

    The document describes material and mechanical tests, and biocompatibility studies. These types of tests do not typically use "test sets" of patient data in the same way an AI algorithm would.

    • Sample Size for Test Set: Not applicable in the context of human patient data for algorithm evaluation. For the bench and in-vitro/in-vivo biocompatibility tests, specific sample sizes are not detailed in this summary document, but the fact that the tests were conducted according to ISO/ASTM/USP standards implies appropriate sample sizes were used for those specific test methods.
    • Data Provenance: Not applicable. These tests were likely performed in a laboratory setting, not with patient data.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    Not applicable. The "ground truth" for these types of device performance tests (e.g., sterility, material strength, biocompatibility) is established by adherence to validated international standards and the results of laboratory measurements, not by expert human interpretation of medical images or clinical data.

    4. Adjudication Method for the Test Set

    Not applicable. This concept belongs to the evaluation of algorithms where human assessment of data needs to be considered. For a physical device like this, performance is measured against established scientific and engineering standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. An MRMC study is a clinical study design used to compare the performance of different diagnostic methods (e.g., human readers with and without AI assistance). This 510(k) summary is for a physical medical implant (a membrane) and focuses on demonstrating its safety and substantial equivalence, not its diagnostic accuracy or the improvement of human performance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    No. This device is not an algorithm, so "standalone performance" in that context is not applicable. Its performance is measured by its material properties, biological interactions, and mechanical integrity.

    7. The Type of Ground Truth Used

    The "ground truth" for the various performance tests is either:

    • Physical/Chemical Standards: (e.g., strength measurements in Bend and Delamination testing, density, thickness).
    • Biological Standards: (e.g., absence of cytotoxicity, irritation, sensitization, genotoxicity, pyrogenicity as defined by ISO 10993 and USP standards).
    • Microbiological Standards: (e.g., sterility per ISO 11135).

    Essentially, the ground truth is adherence to internationally recognized and validated scientific and engineering standards.

    8. The Sample Size for the Training Set

    Not applicable. This device is not an AI algorithm, so there is no "training set."

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As there is no AI algorithm and no training set, this question is not relevant to the Bio-MEM Ti Reinforced Membrane.

    Ask a Question

    Ask a specific question about this device

    K Number
    K201187

    Validate with FDA (Live)

    Date Cleared
    2021-01-23

    (267 days)

    Product Code
    Regulation Number
    872.3930
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Cytoplast™ Titanium-Reinforced PTFE Membranes are a temporarily implantable material (non- resorbable) indicated for use as a space-making barrier in the treatment of alveolar and periodontal bony defect sites.

    Device Description

    Cytoplast™ Titanium-Reinforced PTFE Membranes are placed between bone grafts and the periosteum in dental bone grafting procedures to stabilize and support the bone graft. The PTFE membrane helps create the space needed for bone-derived cells to repopulate and repair the defect.

    Cytoplast™ Titanium-Reinforced PTFE Membranes are composed of proprietary 100% polytetrafluoroethylene sheets reinforced with a titanium frame. The titanium frame is embedded between two layers of PTFE. PTFE is biologically nearly inert and a tissue-compatible material.

    The PTFE membranes are designed to maintain space and conform to tissue contours.

    Cytoplast™ Titanium-Reinforced PTFE Membranes are provided in two different thicknesses (approximately 150 µm & 250 µm) and are pre-shaped in a variety of shapes and sizes.

    AI/ML Overview

    The provided text describes a 510(k) summary for a medical device and does not contain information about an AI/ML powered device, a comparative effectiveness study, or details related to establishing ground truth for a test set in the context of an AI/ML device.

    Therefore, I cannot provide the requested information in the format of a table for acceptance criteria and device performance as it pertains to AI/ML or a study that proves a device meets AI/ML specific acceptance criteria. The document describes a traditional medical device (Cytoplast™ Titanium-Reinforced PTFE Membranes) and its substantial equivalence to predicate devices based on non-clinical testing.

    Here's a breakdown of what can be extracted in relation to the questions asked, adapted to the context of the provided document:

    Acceptance Criteria and Device Performance (Non-AI/ML context)

    Acceptance Criteria (Bench Tests)Reported Device Performance (Cytoplast™ Titanium-Reinforced PTFE Membranes)
    Biocompatibility (per ISO 10993)Confirmed (Cytotoxicity, Irritation, Toxicity, Implantation)
    Sterility Assurance Level (per ISO 11135-1:2007)Achieved 10⁻⁶ SAL
    Shelf-life (per ISO 11607-1)Met (4 years)
    Tensile Strength (per ASTM D638-14)Tensile strength ≥ reference device (Bio-Gide® Resorbable Bilayer Membrane)
    Lamination Strength (Internal Test)Lamination is acceptable

    Further Details (where applicable to the provided document):

    1. Sample size used for the test set and the data provenance: Not applicable in the context of AI/ML. The "test set" here refers to the samples of the Cytoplast™ Titanium-Reinforced PTFE Membranes subjected to various non-clinical bench tests (e.g., for tensile strength, lamination). Specific sample sizes for each test are not explicitly stated in the summary, but the tests were performed on the device itself.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth as typically understood for AI/ML models (e.g., expert-labeled images) is not relevant here. The "truth" for these non-clinical tests is determined by standardized protocols (e.g., ASTM D638-14 for tensile strength) and material characterization.
    3. Adjudication method for the test set: Not applicable. This document describes physical and material property testing, not subjective assessment requiring adjudication.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI/ML device, and no MRMC study was conducted.
    5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable. This is not an AI/ML device.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable in the AI/ML sense. The "ground truth" for the non-clinical tests is established by industry standards, physical measurements, and chemical/biological assays. For instance, biocompatibility is evaluated against established ISO 10993 cytotoxicity, irritation, and toxicity standards. Tensile strength is objectively measured.
    7. The sample size for the training set: Not applicable. There is no AI/ML model involved, and therefore no training set.
    8. How the ground truth for the training set was established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K171774

    Validate with FDA (Live)

    Date Cleared
    2017-10-19

    (127 days)

    Product Code
    Regulation Number
    872.3930
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    RPM™ Reinforced PTFE Mesh is a temporarily implantable material (non-resorbable) indication and support of bone grafts in alveolar bony defect sites.

    Device Description

    RPM™ Reinforced PTFE Mesh is placed between bone grafts and the periosteum in dental bone grafting procedures to stabilize and support the bone graft. The PTFE mesh helps create the space needed for bone-derived cells to repopulate and repair the defect. RPM™ Reinforced PTFE Mesh is composed of proprietary 100% polytetrafluoroethylene sheets reinforced with a titanium frame. The titanium frame is embedded between two layers of PTFE. PTFE is a biologically inert and tissue-compatible material. RPM™ Reinforced PTFE Mesh is manufactured with circular 0.66 mm diameter macropores to allow direct contact between the bone graft and the periosteum. Direct contact between the periosteum and bone graft allows naturally occurring revascularization and infiltration of cells. The PTFE mesh is designed to maintain space and conform to tissue contours. RPM™ Reinforced PTFE Mesh is provided pre-shaped in a variety of shapes and sizes.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the RPM™ Reinforced PTFE Mesh, a medical device for bone grafting. The document focuses on demonstrating substantial equivalence to predicate and reference devices through non-clinical testing.

    Here's an analysis of the acceptance criteria and study information provided:

    1. Table of Acceptance Criteria and Reported Device Performance:

    TestAcceptance CriteriaReported Device Performance
    Tensile StrengthTensile strength ≥ reference device (Bio-Gide®)Tensile strength ≥ reference device (Bio-Gide®)
    Lamination StrengthLamination strength ≥ reference device (Cytoplast™ Regentex™ Titanium 250)Lamination strength ≥ reference device (Cytoplast™ Regentex™ Titanium 250)
    Suture Retention ForceSuture retention force ≥ reference device (Cytoplast™ Regentex™ Titanium 250)Suture retention force ≥ reference device (Cytoplast™ Regentex™ Titanium 250)
    CytotoxicityNon-cytotoxic (passed ISO 10993-5)Non-cytotoxic
    IrritationNon-irritant (passed ISO 10993-10)Non-irritant

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not explicitly state the numerical sample sizes used for each in vitro bench test. It mentions that "in vitro product characterization testing was performed" and "A series of bench tests were conducted."
    The data provenance is from non-clinical testing (in vitro characterization studies and biocompatibility studies). No information about country of origin for the data is provided, but it's generated as part of a US FDA submission. The studies are prospective in the sense that they were conducted specifically for this submission to evaluate the device.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts:

    Not applicable. The "ground truth" in this context is established by the design and execution of the scientific tests based on recognized standards (ASTM, ISO). There are no human "experts" establishing ground truth in the way one would for image interpretation in AI. The testing itself is the ground truth.

    4. Adjudication Method for the Test Set:

    Not applicable. This is not a study involving human interpretation or subjective assessment that would require adjudication. The test results are objective measurements against established standards and comparisons to reference devices.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This submission is for a physical medical device (reinforced PTFE mesh), not an AI/imaging diagnostic device. Therefore, no MRMC study was conducted or is relevant.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    Not applicable. This is not an algorithm or AI device.

    7. The Type of Ground Truth Used:

    The ground truth used for these non-clinical tests is based on:

    • Performance against established industry standards: ASTM D638-14 for tensile strength, and ISO 10993-5 and ISO 10993-10 for biocompatibility.
    • Comparative performance against legally marketed predicate/reference devices: The specified predicate device (Osteo-Mesh™ TM-300) and reference devices (Cytoplast™ Regentex™ Titanium 250 and Bio-Gide® Resorbable Bilayer Membrane) serve as benchmarks for tensile strength, lamination strength, and suture retention force.

    8. The Sample Size for the Training Set:

    Not applicable. This is not an AI device or a study involving a "training set" in that sense.

    9. How the Ground Truth for the Training Set was Established:

    Not applicable. Refer to point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K160493

    Validate with FDA (Live)

    Date Cleared
    2016-09-08

    (199 days)

    Product Code
    Regulation Number
    872.3930
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Salvin CytoSurg™ Non-Resorbable PTFE Membrane is a temporarily implantable material (nonresorbable) for use as a space-making barrier in the treatment of periodontal defects.

    Device Description

    The Salvin CytoSurg™ Non-Resorbable PTFE Membrane is composed of microporous PTFE material. It has a nominal thickness of 250 microns and is supplied in two sizes (25mm x 30mm and 12mm x 24mm). The membranes are single use and are supplied in sterile sealed pouches.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification for a medical device called the Salvin CytoSurg™ Non-Resorbable PTFE Membrane. This document focuses on demonstrating substantial equivalence to a predicate device, rather than providing a study proving performance against specific acceptance criteria in the context of an AI/ML device.

    Therefore, the requested information regarding acceptance criteria, study details, sample sizes, expert involvement, adjudication methods, MRMC studies, standalone performance, ground truth, and training set details for an AI/ML device cannot be extracted from this document.

    However, I can extract the performance data and the conclusion related to the device's substantial equivalence to its predicate, as presented in the document.

    Acceptance Criteria and Reported Device Performance (as related to substantial equivalence):

    Acceptance Criteria CategorySpecific Tests PerformedReported Device Performance
    Physical PropertiesTensile StrengthBench testing demonstrated substantial equivalence to the predicate device
    Tear ResistanceBench testing demonstrated substantial equivalence to the predicate device
    Geometric ComparisonsDemonstrated substantial equivalence to the predicate device
    BiocompatibilityISO 10993-5 / Tests CytotoxicityMet their acceptance criteria
    ISO 10993-6 / Tests Local Effects After ImplantationMet their acceptance criteria
    ISO 10993-10 / Tests for irritation and skin sensitizationMet their acceptance criteria
    ISO 10993-11 / Tests for Systemic ToxicityMet their acceptance criteria
    Other ValidationsPackaging ValidationMet their acceptance criteria
    Sterilization ValidationMet their acceptance criteria

    Since this is not an AI/ML device, the following information is not applicable and cannot be extracted from the provided text:

    1. Sample size used for the test set and the data provenance
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
    3. Adjudication method for the test set
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and the effect size
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
    7. The sample size for the training set
    8. How the ground truth for the training set was established
    Ask a Question

    Ask a specific question about this device

    K Number
    K151344

    Validate with FDA (Live)

    Date Cleared
    2015-11-19

    (184 days)

    Product Code
    Regulation Number
    872.3930
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Cytoflex® Tef-Guard® Ti-Enforced membrane is a temporarily, non-resorbable, implantable material for use as a space-making barrier in the treatment of periodontal defects and augmentation of alveolar ridge in accordance with guided tissue regeneration principle.

    Device Description

    Cytoflex® Tef-Guard® Ti-Enforced membranes are a multi-layer, non-resorbable membrane intended to be surgically placed under the muco-periosteum to aid in tissue regeneration in accordance with the guided tissue regeneration principle. It is a passive, non-load bearing material with a titanium frame enclosed within two exterior layers of ePTFE material. The titanium reinforcement is intended for space and shape maintenance to contain bone grafting material and to minimize graft migration during wound healing. The membranes are designed to reduce the migration and establishment of gingival tissue derived cells into bony defects thus providing a more favorable environment for neovascularization and bone derived cells to repopulate and repair the defect. The membranes are intended to be submerged and implanted for more than 30 days and up to 6 months. The membrane is supplied sterile and available in a variety of shapes and sizes for single use only.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the Cytoflex® Tef-Guard® Ti-Enforced Membrane:

    Important Note: The provided document is a 510(k) Summary, which focuses on demonstrating substantial equivalence to a predicate device rather than comprehensive clinical efficacy testing. Therefore, many of the typical elements of a study proving a device meets acceptance criteria (like human reader studies, ground truth establishment for a test set, etc.) are not applicable in this context. The acceptance criteria here largely revolve around device performance characteristics and biocompatibility, demonstrated through non-clinical bench testing.


    Acceptance Criteria and Study to Prove Device Meets Acceptance Criteria

    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't explicitly state quantitative acceptance criteria for each test in a table format. Instead, it states that the device "was found to meet the requirements of the pre-defined acceptance criteria" for the listed ISO standards and that "The test results demonstrate that the subject device is substantially equivalent to the primary predicate device in functional performance."

    However, based on the performance testing section and the comparison table with predicate devices, we can infer the types of criteria and the reported performance:

    Acceptance Criteria CategorySpecific Criteria (Inferred)Reported Device Performance
    Material & Design Equivalence- Thickness: Not to exceed a certain deviation from predicate (e.g., within 0.07mm difference).- Density: Not to exceed a certain deviation from predicate (e.g., within 0.5 g/dm² difference).- Material Composition: ePTFE with Titanium frame.- Design: Multi-layer, non-resorbable, space-making barrier, trimmable, flexible.- Thickness: 0.25mm (Predicate 1: 0.23mm; Predicate 2: 0.3mm). Considered "Substantially Equivalent."- Density: 3.5 g/dm² (Predicate 1: 4.0 g/dm²; Predicate 2: 3.5 g/dm²). Considered "Substantially Equivalent."- Material Composition: ePTFE membranes reinforced with a Titanium frame. Considered "Substantially Equivalent."- Design: A variety of configurations, trimmable and shaped to fit bony defect. Considered "Substantially Equivalent."
    Functional Performance- Tensile Loading: Performance similar to predicate.- Bending: Performance similar to predicate.- Flexibility: Performance similar to predicate.- Delaminating Potential: Performance similar to predicate."The test results demonstrate that the subject device is substantially equivalent to the primary predicate device in functional performance." (No specific quantitative values provided for the subject or predicate, just a statement of equivalence).
    Biocompatibility- Cytotoxicity: No evidence of cytotoxicity (ISO 10993-5).- Local Effects after Implantation: No adverse local effects (ISO 10993-6).- Irritation & Skin Sensitization: No irritation or sensitization (ISO 10993-10).- Systemic Toxicity: No systemic toxicity (ISO 10993-11)."Found to meet the requirements of the pre-defined acceptance criteria" for ISO 10993-5, -6, -10, -11. "Biocompatible: Yes."
    Sterility- Sterility Assurance Level (SAL): SAL < 10-6.- Sterilization Process Control: Meets requirements of ISO 11135.- Ethylene Oxide Residuals: Meets limits of ISO 10993-7."Sterile by EtO gas, SAL < 10-6." "Found to meet the requirements of the pre-defined acceptance criteria" for ISO 11135 and ISO 10993-7. "Sterility: Sterile."

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify sample sizes for the performance (tensile loading, bending, flexibility, delamination) or biocompatibility tests. These tests are typically conducted in a laboratory setting on a representative number of device samples.

    • Provenance: This is a non-clinical 510(k) submission. The data provenance would be from laboratory bench testing and materials characterization, not from patient data or clinical studies.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This is not applicable as there was no test set requiring expert interpretation or ground truth establishment for clinical performance. The "ground truth" for the non-clinical tests would be the established scientific and engineering principles and the acceptance limits defined by the relevant ISO standards and predicate device performance.

    4. Adjudication Method for the Test Set

    This is not applicable as there was no clinical test set requiring expert adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    This is not applicable. No MRMC clinical study was performed or required for this 510(k) submission. The device is a physical implant, not an imaging or diagnostic AI device that would involve human readers.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    This is not applicable. The device is a physical medical implant, not a software algorithm.

    7. Type of Ground Truth Used

    For the non-clinical testing:

    • Bench Test Results: The "ground truth" for functional performance tests (tensile, bending, flexibility, delamination) was derived from direct comparison to the performance of the chosen predicate device, assessed through quantitative measurements. The acceptance criterion was "substantially equivalent" performance to the predicate.
    • ISO Standard Requirements: For biocompatibility and sterilization, the "ground truth" was established by the requirements and limits defined within the specified ISO standards (e.g., no cytotoxicity detected according to ISO 10993-5).

    8. Sample Size for the Training Set

    This is not applicable. There was no "training set" in the context of an AI/ML algorithm. The device is a manufactured physical product.

    9. How the Ground Truth for the Training Set Was Established

    This is not applicable. There was no "training set" for this device.


    Summary of Approach:

    The K151344 submission for the Cytoflex® Tef-Guard® Ti-Enforced Membrane relied on demonstrating substantial equivalence to existing predicate devices. This was achieved primarily through:

    • Comparison of Device Characteristics: Showing similar materials, design, indications for use, thickness, density, and sterility to the predicate devices.
    • Non-Clinical Performance Testing: Bench testing for functional performance (tensile loading, bending, flexibility, delamination) was conducted, and the results confirmed that the subject device performed "substantially equivalent" to one of the predicate devices.
    • Biocompatibility and Sterilization Testing: Meeting the requirements of relevant ISO standards for biocompatibility (cytotoxicity, local effects, irritation, systemic toxicity) and sterilization (EtO sterilization, residuals).

    Clinical data was specifically stated as "not required to establish the substantial equivalence." This means the FDA accepted the non-clinical data as sufficient for demonstrating safety and effectiveness comparable to the legally marketed predicate devices.

    Ask a Question

    Ask a specific question about this device

    K Number
    K143327

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2015-04-13

    (144 days)

    Product Code
    Regulation Number
    872.3930
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    An implantable temporary non-resorbable device (membrane) for use as a spacer creation barrier in the treatment of local defects in the oral cavity in conjunction with tissue regeneration or augmentation. Neoss membranes are intended to be submerged and clinically implanted more than 30 days with an expected duration of implantation up of 6 months.

    Device Description

    The Neoss Ti Reinforced Membrane a multi-layer non-resorbable dental membrane intended to be surgically placed beneath the muco-periosteum to aid in regenerative healing in 1) bone or 2) bone/periodontal ligament defects of the oral cavity. The Neoss Ti Reinforced Membrane is composed of two layers of PTFE membrane material enclosing a Titanium(Ti) mesh. The Neoss Ti Reinforced Membrane is a passive, non-load bearing material. The small pores of the PTFE membrane material, allows passage of fluid and vapor; as well as a controlled cellular ingrowth for stability and a barrier for soft tissue penetration. The Ti mesh is unalloyed surgical grade Titanium ISO 5832-2 Grade 2. The titanium mesh reinforcement is intended for space creating and shape-maintaining which minimizes movements and subsequent exposure while implanted. Neoss Ti Reinforced Membrane is provided pre-shaped in a variety of shapes and sizes detailed below. The device is supplied sterile and single use.

    AI/ML Overview

    This document is a 510(k) Pre-Market Notification from the FDA regarding the "Neoss Ti Reinforced Membrane." It is a regulatory approval document and not a study report that details acceptance criteria and device performance in a clinical context.

    Therefore, the requested information about acceptance criteria and a study proving device performance cannot be extracted from this document.

    The document states:

    • "Clinical Testing: Clinical data was not required to establish the substantial equivalence of the Neoss Ti Reinforced Membrane." This explicitly indicates that a clinical study, with its associated methodologies for acceptance criteria and performance measurement, was not performed or submitted as part of this specific FDA cleared.

    The "Performance Testing" section mentions tests conducted according to ISO and ASTM standards (e.g., burst strength, bend testing, biocompatibility). However, these are engineering and material property tests to ensure the device meets specified quality and safety characteristics, not clinical performance acceptance criteria. The document lists the standards applied, but does not provide the specific acceptance criteria values or the reported performance values of the Neoss Ti Reinforced Membrane against those criteria. It only states that these tests were done "in order to confirm the strength, integrity and functionality of the device."

    The document focuses on establishing "substantial equivalence" to predicate devices based on indications for use, technological characteristics, and material composition, rather than presenting a standalone study with defined acceptance criteria and performance results in a table.

    Therefore, I cannot fulfill your request for:

    1. A table of acceptance criteria and reported device performance.
    2. Sample sizes used for the test set or data provenance (as no clinical test set was described).
    3. Number of experts or their qualifications (as no ground truth establishment for a clinical test set was described).
    4. Adjudication method (as no clinical test set was described).
    5. MRMC comparative effectiveness study (as no human reader study was described).
    6. Standalone performance (as it refers to clinical performance, which was not assessed in a study described here).
    7. Type of ground truth used (as no clinical ground truth was established).
    8. Sample size for the training set (as this is not a machine learning device).
    9. How ground truth for the training set was established (as this is not a machine learning device).
    Ask a Question

    Ask a specific question about this device

    K Number
    K132325

    Validate with FDA (Live)

    Date Cleared
    2013-10-29

    (95 days)

    Product Code
    Regulation Number
    872.3930
    Panel
    Dental
    Age Range
    2 - 120
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    GBR Configurations: To aid in bone regeneration and augmentation in oral surgery for: Extraction socket site preservation; Immediate implant placement at time of extraction or delayed placement when additional bone regeneration is desire; Ridge augmentation; Sinus elevation; Stable barrier for the containment of bone grafting materials. GTR Configurations: To aid in the regeneration and integration of periodontal tissue components. For use as an adjunct in periodontal surgical treatment to supplement the reparative process following scaling and root planning for: Class II furcations; intrabony defects; recession type defects.

    Device Description

    The GUIDOR® Bioresorbable Matrix Barrier is a two-layered (sandwich) structure with spacers between the layers that provide stability to the matrix and maintain an adequate space to aid in tissue regeneration and prevention of epithelial downgrowth. The GUIDOR® Matrix is bioresorbable to allow for singlestep surgery, eliminating the need for a second surgery to remove the matrix.

    AI/ML Overview
    1. Acceptance Criteria and Reported Device Performance:
    Feature/CharacteristicAcceptance Criteria (Implicit)Reported Device Performance
    Substantial EquivalenceThe device should have the same intended use, similar indications, technological characteristics, and principles of operation as legally marketed predicate devices. It must be as safe and effective as the predicate devices.The GUIDOR® Bioresorbable Matrix Barrier is stated to have the "same intended use and similar indications, technological characteristics, and principles of operation as its predicate devices." It is concluded to be "as safe and effective as the predicate devices," thereby demonstrating substantial equivalence.
    Technological CharacteristicsShould be similar to predicate devices, specifically regarding its two-layered (sandwich) structure, inclusion of spacers, and being bioresorbable. The material composition (synthetic) should also be comparable to relevant predicates.The device is described as a "two-layered (sandwich) structure with spacers" and is "bioresorbable." It shares "the same technical characteristics and operating principles as the Guidor™ predicate (K912724)" and is "composed of synthetic materials similar to the Guidor™ predicate (K912724), Inion GTR Biodegradable Membrane System (K033074), and Resolut® Adapt LT Regenerative Membrane (K051267) predicates." It is also designed to be sterile.
    Material PropertiesBiocompatible and sterile for its intended use."Results of biocompatibility and sterility testing conclude that the GUIDOR® Bioresorbable Matrix Barrier is biologically safe for its intended use."
    Clinical EfficacyShould aid in bone regeneration and augmentation for specific GBR configurations and aid in regeneration and integration of periodontal tissue components for specific GTR configurations, similar to established performance of predicate devices."Results from retrospective clinical data and human studies reported in the literature for both synthetic resorbable membranes and resorbable collagen membranes support that the GUIDOR® Bioresorbable Matrix Barrier is substantially equivalent to those of the previously accepted and clinically successfully used bioresorbable membranes for similar indications."

    Study Information:

    1. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

      • The document does not explicitly state a specific "test set" sample size or its country of origin.
      • The clinical evidence provided is based on "retrospective clinical data and human studies reported in the literature." This suggests the data is retrospective, but the origin country is not specified.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not applicable. The study relies on "retrospective clinical data and human studies reported in the literature" rather than a specific expert-adjudicated test set for this 510(k) submission. No information is provided about the number or qualifications of experts involved in the original studies forming the literature review.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not applicable. No explicit adjudication method is mentioned for a test set. The submission relies on a literature review.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not applicable. This device is a bioresorbable matrix barrier, not an AI or imaging diagnostic device. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance is not relevant or described.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Not applicable. This is a physical medical device, not an algorithm, so standalone algorithm performance is not relevant.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The "ground truth" is implicitly derived from the established clinical outcomes and efficacy reported in the existing "retrospective clinical data and human studies reported in the literature" for both synthetic resorbable membranes and resorbable collagen membranes. This effectively means outcomes data from previously conducted clinical studies.
    7. The sample size for the training set:

      • Not applicable. This submission does not describe an AI/machine learning model with a distinct "training set." The evidence for substantial equivalence is based on existing published clinical literature.
    8. How the ground truth for the training set was established:

      • Not applicable, as there is no described training set for an algorithm in this submission.