Search Filters

Search Results

Found 124 results

510(k) Data Aggregation

    K Number
    K241581

    Validate with FDA (Live)

    Date Cleared
    2025-05-16

    (347 days)

    Product Code
    Regulation Number
    876.5540
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    • The Hemodialysis Catheter is indicated for use in attaining short-term or long-term vascular access for Hemodialysis therapy and apheresis.
    • It may be inserted percutaneously and is primarily placed in the internal jugular vein of an patient.
    • Alternate insertion sites include subclavian vein as required.
    • Catheters greater than 40 cm are intended for femoral vein insertion.
    • The curved Hemodialysis Catheter is intended for internal jugular vein insertion.
    • 8-10Fr Hemodialysis Catheters can be used in pediatric patients.
    Device Description

    The Hemodialysis Catheter is indicated for use in attaining vascular access for Hemodialysis and Apheresis therapy. The catheter tubing is made of radiopaque polyurethane with a dual D cross-sectional design, providing independent arterial and venous lumens. The proximal end of the catheter features two luer connectors. The luer connectors are connected to extension legs. The extension legs are made of silicone material. Each extension leg has a catheter clamp, which is durable and can effectively ensure timely closure of the extension leg. The lumen with the red catheter clamp, which is called arterial lumen, is used for blood outflow; The lumen with the blue catheter clamp, which is called venous lumen, is used for blood return. The arterial and venous lumen priming volumes are printed directly on the ID ring of the clamp. Catheters range from approximately 13-55 cm long and are offered in straight or curved catheter configurations with cuff for long-term implantation. The Hemodialysis Catheter is packaged in a tray with legally marketed accessories intended for use during catheter placement. The Hemodialysis Catheter Kit is provided sterile, single use.

    AI/ML Overview

    The provided FDA 510(k) clearance letter and associated documentation for the Hemodialysis Catheter (K241581) primarily detail the regulatory approval process, comparison to predicate devices, and non-clinical performance testing. It does not contain information about acceptance criteria for device performance based on clinical study data, nor does it describe a study that validates the device performance against such criteria in a clinical setting.

    The document specifically states: "Clinical test: Clinical testing is not required." This indicates that the clearance was based on non-clinical performance data and substantial equivalence to legally marketed predicate devices, not on direct clinical evidence of the device's performance in patients.

    Therefore, I cannot provide the requested information regarding:

    • A table of acceptance criteria and reported device performance from a clinical study.
    • Sample sizes used for the test set and data provenance from a clinical study.
    • Number of experts used to establish ground truth and their qualifications from a clinical study.
    • Adjudication method for a clinical study.
    • Multi-reader multi-case (MRMC) comparative effectiveness study.
    • Standalone (algorithm only without human-in-the-loop performance) study.
    • Type of ground truth used in a clinical study.
    • Sample size for the training set from a clinical study.
    • How ground truth for the training set was established in a clinical study.

    The provided document focuses on the following types of performance data, which are non-clinical in nature:

    1. Acceptance Criteria and Reported Device Performance (Non-Clinical)

    While not presented in a table with specific pass/fail metrics, the document implies that the device met the requirements of the following standards and guidance:

    • FDA guidance "Implanted Blood Access Devices for Hemodialysis, issued on January 21, 2016"
    • ISO 10555-1:2013 (Sterile, single-use intravascular catheters - Part 1: General requirements)

    The following non-clinical tests were performed to demonstrate compliance:

    Test/Criteria CategoryReported Device Performance and Compliance
    Physical and Mechanical Performance
    Visual AssessmentPerformed according to requirements (implied acceptance)
    Air and Liquid Leakage TestingPerformed according to requirements (implied acceptance)
    Catheter Tensile TestingPerformed according to requirements (implied acceptance)
    Repeated ClampingPerformed according to requirements (implied acceptance)
    Priming VolumesAccurate volumes printed on ID ring of clamp (implied acceptance as per standard practice). Measured for compliance.
    Luer ConnectorConforms to ISO 80369-7:2021 (Small-bore connectors for liquids and gases in healthcare applications - Part 7: Connectors for intravascular or hypodermic applications)
    Pressure Versus Flow RatesPerformed according to requirements (implied acceptance). Data demonstrates that the difference in O.D. (size range) compared to predicate does not affect safety and effectiveness.
    Recirculation RatePerformed according to requirements (implied acceptance)
    Mechanical Hemolysis TestingPerformed according to requirements (implied acceptance)
    Chemical Tolerance TestingPerformed according to requirements (implied acceptance)
    Particulate Contamination TestingPerformed according to requirements (implied acceptance)
    Dimension TestingOverall dimensions (length, O.D.) within acceptable ranges. Data demonstrates that the difference in O.D. (broader size range, e.g., 8F~15.5F) compared to predicate (12.5Fr) does not affect safety and effectiveness. Pediatric catheter (8-10Fr) performance test results compared against K202150 (a predicate for pediatric use) and toxicological risk assessment based on children's weight proved non-inferiority.
    BiocompatibilityThe compatibility of patient-contacting materials (PU and Silicone) meets requirements. Evaluation and tests comply with ISO 10993-1: "Biological Evaluation of Medical Devices Part 1: Evaluation and testing within a risk management process."
    SterilizationEO sterile for single use. Complies with ISO 11607-1:2019 (Packaging for terminally sterilized medical devices) and ISO 11135:2014 (Sterilization of health care products -- Ethylene oxide -- Requirements for the development, validation and routine control of a sterilization process for medical devices).
    Shelf LifeAccelerated aging tests confirmed the validity of the 3-year shelf life.

    2. Sample Sized Used for the Test Set and the Data Provenance

    • Not Applicable for Clinical Data: The document explicitly states "Clinical testing is not required."
    • For Non-Clinical Data: Specific sample sizes for each non-clinical test (e.g., tensile testing, leakage testing) are not detailed in this summary document. Data provenance is implied to be from internal lab testing conducted by Haolang Medical USA Corporation, in compliance with the cited ISO standards and FDA guidance.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    • Not applicable as clinical testing with expert-established ground truth was not performed or required for this 510(k) clearance.

    4. Adjudication Method for the Test Set

    • Not applicable as clinical testing with an adjudication method was not performed or required.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    • Not applicable. This type of study is typically for AI/imaging device performance evaluation, not for a physical device like a hemodialysis catheter, and certainly not for a clearance that states "Clinical testing is not required."

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • Not applicable. This is not a software/AI device.

    7. The Type of Ground Truth Used

    • For Non-Clinical Testing: The "ground truth" for the non-clinical tests is adherence to the specified performance requirements as outlined in the referenced ISO standards (e.g., ISO 10555-1) and FDA guidance documents for hemodialysis catheters (e.g., "Implanted Blood Access Devices for Hemodialysis"). This involves objective, measurable physical, mechanical, and biological properties.

    8. The Sample Size for the Training Set

    • Not applicable. This refers to training data for AI/ML algorithms, which is not relevant to this device's clearance.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable.

    In summary, the FDA 510(k) clearance for the Hemodialysis Catheter (K241581) was granted based on a substantial equivalence determination supported by comprehensive non-clinical performance testing (physical, mechanical, biocompatibility, sterilization, and shelf-life) against recognized industry standards and FDA guidance, rather than a clinical study with patient data and expert-derived ground truth.

    Ask a Question

    Ask a specific question about this device

    K Number
    K210461

    Validate with FDA (Live)

    Device Name
    End Cap
    Date Cleared
    2021-09-14

    (210 days)

    Product Code
    Regulation Number
    876.5540
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Luer lock plug for capping male or female luer tapers for hemodialysis catheters.

    Device Description

    The Medcomp® End cap is non-vented plastic cap with a male luer. The end connects directly to the female luer of the catheter.

    AI/ML Overview

    The provided document is a 510(k) summary for a medical device called "End Cap". It details the device's description, indications for use, comparison to a predicate device, and performance data for demonstrating substantial equivalence.

    However, the questions you've asked are specifically about the acceptance criteria and study proving the device meets acceptance criteria for an AI/ML-enabled medical device. The provided document is for a physical medical device (luer lock plug) and does not involve AI or machine learning. Therefore, the questions related to AI/ML device performance (e.g., sample size for test/training sets, experts for ground truth, MRMC study, standalone performance) are not applicable to the content of this document.

    The document discusses performance testing (bench testing) to ensure the physical device meets relevant standards, but this is different from the type of performance evaluation required for an AI/ML diagnostic or predictive device.

    Here's how I can address the applicable parts of your request based on the provided document, interpreting "acceptance criteria" in the context of a physical device's regulatory submission:

    The "acceptance criteria" for this device's submission revolves around demonstrating substantial equivalence to a legally marketed predicate device by meeting recognized consensual standards for physical device performance and biocompatibility.


    1. A table of acceptance criteria and the reported device performance

    For a physical device like the End Cap, acceptance criteria are typically defined by compliance with recognized international standards (ISO standards in this case) and demonstrating safe and effective performance. The "reported device performance" is essentially that the device passed these tests, confirming its compliance.

    Acceptance Criteria (Defined by Standard)Reported Device Performance (as stated in submission)
    ISO 594-1:1986Passed (Medcomp® End Cap met requirements for Gauging, Liquid Leakage, Air Leakage, Stress Cracking)
    * Conical fittings with a 6 % (Luer) taper for syringes, needles and certain other medical equipment — Part 1: General requirements
    ISO 594-2:1998Passed (Medcomp® End Cap met requirements for Separation Force, Liquid Leakage, Air Leakage, Unscrewing Torque, Ease of Assembly, Resistance to Overriding, Stress Cracking)
    * Conical fittings with 6 % (Luer) taper for syringes, needles and certain other medical equipment — Part 2: Lock fittings
    ISO 10993-1: Biological Evaluation of Medical Devices Part 1: Evaluation and Testing Within a Risk Management ProcessPassed (Met biocompatibility requirements for an external communicating device with circulating blood contact for a limited duration (<24 hours), including tests for Cytotoxicity, Sensitization, Irritation/Intracutaneous Reactivity, and Acute Systemic Toxicity)

    Important Note: The document states that the testing was performed and the device "met" the requirements. It doesn't provide specific numerical results or raw data for these tests, as is typical for a 510(k) summary (which summarizes the full testing report).

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size: Not explicitly stated for each test, but typically these tests involve a specific number of units of the device as defined by the standards themselves or internal quality protocols.
    • Data Provenance: The testing was "Bench/Performance Data/Non-Clinical Testing" and "Biocompatibility evaluation." This implies laboratory testing performed by or for the manufacturer (Medcomp®, located in Harleysville, PA, USA). This is prospective testing, as it's specifically conducted to support the regulatory submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. For a physical device, "ground truth" is established by adherence to engineering specifications and international standards, measured by calibrated equipment and qualified lab personnel, not clinical experts interpreting images or diagnostic data.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This is for AI/ML model output, not physical device performance testing.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is not an AI/ML diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a physical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For this physical device, the "ground truth" is adherence to the specifications outlined in the ISO standards, which define acceptable functional performance (e.g., specific force required, leak rate, biocompatibility). Compliance is measured mechanically or chemically, not via clinical or expert assessment of "truth" in a diagnostic sense.

    8. The sample size for the training set

    Not applicable. This is a physical device, not an AI/ML model.

    9. How the ground truth for the training set was established

    Not applicable. This is a physical device, not an AI/ML model.


    In summary, the provided document pertains to a traditional physical medical device and therefore the acceptance criteria and study methods are fundamentally different from those for an AI/ML-enabled device. The "study" proving the device meets acceptance criteria is the successful completion of the specified bench testing and biocompatibility assessments in accordance with recognized standards, demonstrating the device's substantial equivalence to its predicate.

    Ask a Question

    Ask a specific question about this device

    K Number
    K203575

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-07-01

    (206 days)

    Product Code
    Regulation Number
    876.5540
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The GlidePath™ Retro Long-Term Hemodialysis Catheters are indicated for use in attaining short-term or long-term vascular access for hemodialysis, hemoperfusion or apheresis therapy. Access is attained via the internal jugular vein, subclavian vein, or femoral vein. Catheters longer than 40 cm are intended for femoral vein insertion.

    Device Description

    The GlidePath™ Retro Long-Term Hemodialysis Catheter features a dual-lumen polyurethane catheter shaft with optimized double-D cross-sectional design providing separate arterial and venous lumens, an external bifurcation hub with SwiftClick connector assembled by physician upon placement, and arterial and venous extension legs that connect to an external dialysis machine or blood cleansing device. The arterial (red) luer connector secures to the blood intake on the dialysis machine and the venous (blue) luer connector secures to the blood return line on the dialysis machine. Each extension leg has an atraumatic occlusion clamp which closes access to the lumen. The symmetrical catheter tip contains holes that aid in the distribution of blood flow or aid in over-the-guidewire placement. The dialysis catheter is offered in various lengths.

    AI/ML Overview

    This document describes the GlidePath™ Retro Long-Term Hemodialysis Catheter. It addresses the performance data and comparison to predicate devices, but does not describe a study involving an AI medical device or clinical performance data that would typically include acceptance criteria and reporting on sensitivity, specificity, or other performance metrics for an AI algorithm.

    Therefore, I cannot provide the information requested in your prompt regarding AI device acceptance criteria and study details because the provided text is for a traditional medical device (a catheter) and does not contain information about an AI-powered device or its performance study.

    The document focuses on demonstrating substantial equivalence of the catheter to a predicate device through non-clinical testing of various physical and functional characteristics.

    Ask a Question

    Ask a specific question about this device

    K Number
    K211410

    Validate with FDA (Live)

    Date Cleared
    2021-06-04

    (29 days)

    Product Code
    Regulation Number
    876.5540
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The GlidePath™ 13F long-term hemodialysis catheters are indicated for use in attaining short-term or long-term vascular access for hemodialysis, hemoperfusion or apheresis therapy. Access is attained via the internal jugular vein, subclavian vein, or femoral vein. Catheters longer than 34 cm are intended for femoral vein insertion.

    Device Description

    The GlidePath™ 13F Long-Term Hemodialysis Catheter is a vascular access device, intended for use in attaining short-term or long-term vascular access for hemodialysis, hemoperfusion or apheresis therapy and features a dual-lumen, optimized double-D cross-sectional catheter shaft with a fixed symmetrical tip design. The catheter has separate arterial and venous lumens, a molded bifurcation and, extending from the bifurcation, arterial (red) luer and venous (blue) luer and extension legs which connect to an external dialysis machine or blood cleansing device. Each extension leg has an atraumatic occlusion clamp which closes access to the lumen. The symmetrical catheter tip contains holes that aid in the distribution of blood flow and aid in overthe-guidewire placement. The GlidePath™ 13F Long-Term Hemodialysis Catheters are packaged in a tray with legally marketed accessories intended for use during catheter placement. The device is intended for single patient use only.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device, the GlidePath™ 13F Long-Term Hemodialysis Catheter. This submission aims to demonstrate substantial equivalence to a legally marketed predicate device, the GlidePath™ Long-Term Hemodialysis Catheter (K190527).

    The document details various performance tests conducted to support this claim, but it does not describe a study involving an AI/Machine Learning algorithm or human reader improvement with AI assistance. Therefore, many of the requested items related to AI device performance and clinical studies are not applicable or cannot be answered from the provided text.

    The acceptance criteria mentioned are general ("met all predetermined acceptance criteria for design verification activities as specified by applicable standards, guidance, test protocols and/or customer inputs") rather than specific quantitative metrics for an AI model's performance.

    Here's an analysis based on the provided text, addressing the applicable points:

    Acceptance Criteria and Device Performance

    The document states that the device "met all predetermined acceptance criteria for design verification activities as specified by applicable standards, guidance, test protocols and/or customer inputs." However, a specific table of quantitative acceptance criteria and reported numerical performance values is not provided in the text. The performance data section lists the types of tests conducted:

    Test Characteristic / Performance CriteriaReported Device Performance
    Catheter TensileMet acceptance criteria (Implied: results demonstrate substantial equivalence)
    Catheter Tip StiffnessMet acceptance criteria (Implied: results demonstrate substantial equivalence)
    Catheter Tip Damage after FlexureMet acceptance criteria (Implied: results demonstrate substantial equivalence)
    Flow RateMet acceptance criteria (Implied: results demonstrate substantial equivalence)
    RecirculationMet acceptance criteria (Implied: results demonstrate substantial equivalence)
    Mechanical HemolysisMet acceptance criteria (Implied: results demonstrate substantial equivalence)
    Catheter Assembly LeakMet acceptance criteria (Implied: results demonstrate substantial equivalence)
    Catheter BurstMet acceptance criteria (Implied: results demonstrate substantial equivalence)
    Surface InspectionMet acceptance criteria (Implied: results demonstrate substantial equivalence)
    Stylet to Luer TensileMet acceptance criteria (Implied: results demonstrate substantial equivalence)
    Stylet Assembly LeakMet acceptance criteria (Implied: results demonstrate substantial equivalence)
    Catheter to Tunneler Rod TensileMet acceptance criteria (Implied: results demonstrate substantial equivalence)
    Tunneler Rod MalleabilityMet acceptance criteria (Implied: results demonstrate substantial equivalence)

    Note: The text explicitly states that "The results from these tests performed in accordance with standards and FDA guidance, demonstrate that the technical characteristics and performance criteria of the GlidePath™ 13F Long-Term Hemodialysis Catheter is substantially equivalent to the predicate..."

    Study Details (Applicable to Non-AI Device Testing)

    This document describes a submission for a traditional medical device (a catheter), not specifically an AI/Machine Learning device. Therefore, many of the questions related to AI study design (like training/test sets, ground truth establishment for AI, expert consensus for AI interpretation, MRMC studies) are not directly applicable to this premarket notification. The "performance data" refers to physical and functional performance testing of the catheter itself, not diagnostic or clinical accuracy.

    1. Sample sizes used for the test set and the data provenance: The document does not specify the number of units tested for each characteristic (e.g., how many catheters were subjected to tensile testing). It also does not discuss "data provenance" in the context of clinical data for an AI model, as this is a physical device.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth in this context would be objective physical measurements or established engineering standards, not expert clinical interpretation.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable, as this is not an AI-powered device under review.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable, as this is not an AI-powered device under review.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): For the performance tests listed, the "ground truth" would be the established engineering and medical device standards (e.g., ISO standards, FDA guidance on non-clinical testing) against which the catheter's physical/functional properties were measured.
    7. The sample size for the training set: Not applicable, as this is not an AI device that requires a training set.
    8. How the ground truth for the training set was established: Not applicable, as this is not an AI device.

    In summary, the provided document is a 510(k) premarket notification for a physical medical device (a hemodialysis catheter) and details its engineering and physical performance testing to demonstrate substantial equivalence to a predicate device, rather than the performance of an AI/ML algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K203767

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-04-02

    (100 days)

    Product Code
    Regulation Number
    876.5540
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Pristine™ Long-Term Hemodialysis Catheters are indicated for use in attaining short-term vascular access for hemodialysis, apheresis and infusion. Access is attained via the internal jugular vein, or femoral vein. Catheters longer than 40 cm implant length are indicated for femoral vein insertion. Catheters may be inserted percutaneously.

    Device Description

    The Pristine™ Long-Term Hemodialysis Catheter is a chronic hemodialysis catheter consisting of a dual lumen radiopaque shaft with a pre-formed split tip, which enables long-term vascular access for hemodialysis, apheresis, and infusion. The proximal end of the catheter features two color-coded luer adapters. The luer adapters are connected to clear extension tubes. Each extension tube contains a clamp and is connected to the catheter bifurcation and suture wings (hub). The distal end of the catheter hub is connected to the dual lumen catheter shaft. The shaft contains a cuff and extends to a symmetrical split tip. The design of the catheter's distal tip includes a split, symmetric Y-Tip™ with notches and without side-holes or slots. The symmetric Y-Tip™ design allows a spatial separation between the distal ends of the two lumens. The Pristine™ Long-Term Hemodialysis Catheter is provided as a sterile, single use device, and is sterilized using a validated ethylene oxide process. The dialysis catheters are offered in various implant lengths, varying from 19 cm to 55 cm as depicted below.

    AI/ML Overview

    Acceptance Criteria and Device Performance Study for Pristine™ Long-Term Hemodialysis Catheter (K203767)

    The information provided describes the substantial equivalence determination for the Pristine™ Long-Term Hemodialysis Catheter (K203767) by comparing it to a predicate device, the Pristine™ Hemodialysis Catheter (K182443). The regulatory submission focuses on demonstrating that the new device meets the same performance criteria as the previously cleared predicate device, with additional testing for specific design changes.

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document states that the subject device met "all of the design verification and validation requirements previously reported for the predicate device. Pristine™ Hemodialysis Catheter (K182443, cleared May 31, 2019)."

    For the specific design changes in the subject device, the verification testing was performed to the "exact same acceptance criteria as K182443."

    The following table summarizes the acceptance criteria categories mentioned and the general reported performance. Specific numerical acceptance values are not provided in this document but are implied to be identical to those established for the predicate device.

    Acceptance Criteria CategoryReported Device Performance
    Dimensional and Workmanship Analysis (Luers)Met established acceptance criteria (same as K182443).
    Luer OcclusionMet established acceptance criteria (same as K182443).
    Luer Assembly TensileMet established acceptance criteria (same as K182443).
    Leak DecayMet established acceptance criteria (same as K182443).
    Tunneler (Shaft to Tunneler Tensile)Met Bard's acceptance criteria (additional test for design change).
    Tunneler (Damage after Removal)Met Bard's acceptance criteria (additional test for design change).
    Sterile Barrier Visual InspectionMet Bard's acceptance criteria (additional test for design change).
    Sterile Barrier Integrity (Bubble Leak, Dye Leak)Met Bard's acceptance criteria (additional test for design change).
    Minimum Seal WidthMet Bard's acceptance criteria (additional test for design change).
    External Literature Pouch DamageMet Bard's acceptance criteria (additional test for design change).
    Label LegibilityMet Bard's acceptance criteria (additional test for design change).
    Tray CrackingMet Bard's acceptance criteria (additional test for design change).
    Component Movement and DamageMet Bard's acceptance criteria (additional test for design change).
    Peel Force Test (Poly to Poly, Poly to Tyvek)Met Bard's acceptance criteria (additional test for design change).
    All other original verification and validation testing for the predicate deviceRemained the same and was unchanged, implying established acceptance criteria were met.

    2. Sample Size Used for the Test Set and Data Provenance:

    The document states: "All samples used for final testing of each product attribute were representative of finished products." However, specific numerical sample sizes for each test are not provided.

    The data provenance is implied to be from internal laboratory testing conducted by the manufacturer, C. R. Bard, Inc. The document does not specify country of origin for the data nor explicitly state if it was retrospective or prospective, but given the context of design verification and validation, it would be considered prospective testing of newly manufactured devices.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    This information is not applicable and not provided in the document. The study described is a device performance verification and validation study against established engineering and performance criteria, not a study involving human interpretation of medical images or expert consensus for ground truth.

    4. Adjudication Method for the Test Set:

    This information is not applicable and not provided in the document. As this is a physical device performance study, there's no "adjudication" in the sense of reconciling multiple expert opinions. Test results are compared against pre-defined engineering acceptance criteria.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

    No, an MRMC comparative effectiveness study was not done. The document describes a design verification and validation study for a medical device against pre-defined engineering acceptance criteria, not a study evaluating human reader performance with or without AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    No. This document pertains to a physical medical device (catheter), not an algorithm or AI system. Therefore, standalone algorithm performance is not relevant.

    7. The Type of Ground Truth Used:

    The "ground truth" in this context is the pre-established engineering and performance specifications and acceptance criteria for the device's physical and functional attributes. These criteria are derived from applicable standards, guidance documents, test protocols, and customer inputs, which would have been established internally by the manufacturer and validated during the predicate device's clearance.

    8. The Sample Size for the Training Set:

    This information is not applicable and not provided. This is a device performance study, not a machine learning model development or validation study; therefore, there is no "training set."

    9. How the Ground Truth for the Training Set Was Established:

    This information is not applicable and not provided. As there is no training set, the establishment of ground truth for it is irrelevant to this document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K202150

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-11-18

    (107 days)

    Product Code
    Regulation Number
    876.5540
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The GlidePath™ 7.5F Long-Term Hemodialysis Catheter is indicated for use in attaining short-term or long-term vascular access in pediatric, adolescent, and adult patients for hemodialysis, hemoperfusion or apheresis as determined by the prescribing physician. Access is attained via the internal jugular vein, subclavian vein, or femoral vein.

    Device Description

    The GlidePath™ 7.5F Long-Term Hemodialysis Catheter features a dual-lumen shaft with optimized double-D cross-sectional designs providing separate arterial and venous lumens, a molded bifurcation and, extending from the bifurcation, arterial and venous extension legs that connect to an external dialysis machine or blood cleansing device. The arterial (red) luer connector secures to the blood intake on the dialysis machine and the venous (blue) luer connector secures to the blood return line on the dialysis machine. Each extension leg has an atraumatic occlusion clamp which closes access to the lumen. Long-term dialysis catheters are packaged in a tray with legally marketed accessories intended for use during catheter placement. The symmetrical catheter tip contains holes that aid in the distribution of blood flow or aid in over-the-guidewire placement. The symmetrical catheter tip contains holes that aid in the distribution of blood flow or aid in over-the-guidewire placement. The dialysis catheter is offered in various lengths. This line extension of GlidePath™ has been designed with smaller patients in mind and for physicians that may prefer a smaller diameter catheter, so product offerings include smaller lumen diameter and shorter lengths, 8 cm to 19 cm, tip to cuff measurement. Additionally, GlidePath™ 7.5F Long-Term Hemodialysis Catheters have been designed to reach adequate flow rate requirements for pediatric patients or smaller adult patients for whom a physician desires a smaller catheter size.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the GlidePath™ 7.5F Long-Term Hemodialysis Catheter. The document focuses on demonstrating substantial equivalence to a predicate device rather than detailing specific acceptance criteria and a human-in-the-loop study for a device that meets acceptance criteria for an AI/ML based device.

    Therefore, based solely on the provided text, I cannot provide the information requested in your prompt regarding acceptance criteria and a study proving an AI/ML device meets those criteria. The document describes a medical device (a catheter) and its non-clinical performance testing.

    Here's what I can extract and address from your prompt based on the provided text, and what I cannot address:

    Information NOT Found in the Provided Text (relevant to AI/ML device studies):

    • A table of acceptance criteria and the reported device performance for an AI/ML device. The document lists performance tests for a physical catheter, not acceptance criteria for an AI/ML algorithm.
    • Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective). This is specific to data used for AI/ML model validation, which is not discussed.
    • Number of experts used to establish the ground truth for the test set and the qualifications of those experts. Not applicable to a physical catheter.
    • Adjudication method (e.g. 2+1, 3+1, none) for the test set. Not applicable.
    • If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance. No mention of human readers or AI assistance.
    • If a standalone (i.e. algorithm only without human-in-the-loop performance) was done. No mention of an algorithm.
    • The type of ground truth used (expert consensus, pathology, outcomes data, etc). Not applicable to a physical catheter.
    • The sample size for the training set. Not applicable.
    • How the ground truth for the training set was established. Not applicable.

    What CAN be stated from the provided text (related to the physical catheter):

    The document states:
    "The subject device, the GlidePath™ 7.5F Long-Term Hemodialysis Catheter, meets all predetermined acceptance criteria of design verification and validation as specified by applicable standards, guidance, test protocols and/or customer inputs."

    This indicates that prior to the 510(k) submission, the manufacturer established and met certain acceptance criteria for the physical device's performance. However, the document does not detail these specific criteria in a table or provide the raw results against them. It only lists the types of in-vitro tests performed to demonstrate substantial equivalence.

    List of In-Vitro Performance Tests Performed for the GlidePath™ 7.5F Long-Term Hemodialysis Catheter (as a proxy for components of the "study"):

    The following in-vitro tests were performed based on FDA Guidance Documents and internal Risk Assessment procedures:

    • Catheter Tip (Damage After Flexure)
    • Tip Tensile
    • Tunneler (Shaft to Tunneler Tensile)
    • Catheter Tip Stiffness
    • Surface Inspection
    • Catheter Insertion Over Split Sheath Introducer
    • Assembly Leak Resistance
    • Flow Rates
    • Catheter Collapse
    • Catheter Tensile (Shaft to Bifurcation)
    • Catheter Tensile (Extension Leg to Bifurcation)
    • Catheter Tensile (Extension Leg to Connector)
    • Cuff Securement
    • Burst
    • Recirculation
    • Mechanical Hemolysis
    • Catheter Shaft Stiffness
    • Catheter Radiopacity
    • Thumb Clamps
    • Extension Legs (Knitting, Flow)

    Conclusion regarding your query:

    The provided FDA 510(k) clearance letter and summary concern a physical medical device (a catheter), not an AI/ML-based device. Therefore, the specific details requested about an AI/ML device's acceptance criteria, test set characteristics (sample size, provenance, expert ground truth, adjudication), MRMC studies, standalone performance, and training set information are not present in this document. The document confirms that the physical catheter met its predetermined acceptance criteria for design verification and validation through in-vitro testing.

    Ask a Question

    Ask a specific question about this device

    K Number
    K202176

    Validate with FDA (Live)

    Date Cleared
    2020-09-03

    (30 days)

    Product Code
    Regulation Number
    876.5540
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Symetrex® Long Term Hemodialysis Catheter & Symetrex® Long Term Hemodialysis Catheter with Sideholes is a symmetric tip dual lumen catheter designed for chronic hemodialysis and apheresis. It may be inserted percutaneously or by cut down. Catheters with greater than 37cm implant length are indicated for femoral placement.

    Long term, greater than 30 days, vascular access for Hemodialysis and Apheresis treatments.

    Device Description

    The Symetrex® Long Term Hemodialysis Catheter & Symetrex® Long Term Hemodialysis Catheter with Sideholes is a chronic, 15.5 French, dual lumen, radiopaque catheter made of polyurethane. It has a polyester retention cuff and two female luer adapters. The retention cuff promotes tissue ingrowth to anchor the catheter in the subcutaneous tunnel. The luer adapters are identical in color to indicate the reversibility of this catheter. This catheter features symmetrical side channels with a distal tip configuration designed to separate the intake flow from the output flow in both directions.

    AI/ML Overview

    The provided document is a 510(k) summary for a medical device, the Symetrex® Long Term Hemodialysis Catheter & Symetrex® Long Term Hemodialysis Catheter with Sideholes. It states that the device is substantially equivalent to a previously cleared predicate device.

    However, this document does not contain the kind of detailed information requested regarding acceptance criteria and a study proving device performance, particularly related to AI/software performance metrics like sensitivity, specificity, or human reader improvement. The "performance testing" referenced in the document (Table 6.2) refers to non-clinical bench testing and adherence to standards for physical properties, materials, packaging, and biocompatibility, not a clinical study or a study involving AI with human experts.

    Therefore, I cannot provide an answer to your request based on the provided text. The document focuses on demonstrating substantial equivalence through non-clinical testing of the physical catheter and its materials.

    Ask a Question

    Ask a specific question about this device

    K Number
    K200627

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-06-22

    (104 days)

    Product Code
    Regulation Number
    876.5540
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The GlidePath™ 10F long-term hemodialysis catheter is indicated for use in attaining short-term vascular access in pediatric, adolescent and adult patients for hemodialysis, hemoperfusion or apheresis as determined by the prescribing physician. Access is attained via the internal jugular vein, or femoral vein, Catheters longer than 22 cm are intended for femoral vein insertion, depending on patient anatomy and size.

    Device Description

    The Glidepath™ 10F Long-Term Hemodialysis Catheter features a dual-lumen shaft with optimized double-D cross-sectional designs providing separate arterial and venous lumens, a molded bifurcation and, extending from the bifurcation, arterial and venous extension legs that connect to an external dialysis machine or blood cleansing device. The arterial (red) luer connector connects to the blood intake on the dialysis machine and the venous (blue) luer connector connects to the blood return line on the dialysis machine. Each extension leg has an atraumatic occlusion clamp which closes access to the lumen. Long-term dialysis catheters are packaged in a tray with legally marketed accessories intended for use during catheter placement. The symmetrical catheter tip contains holes that aid in the distribution of blood flow or aid in overthe-guidewire placement. The dialysis catheters are offered in various lengths. This line extension of GlidePath™ has been designed with smaller patients in mind and for physicians that may prefer a smaller diameter catheter, so product offerings include the smaller lumen diameter, 10 F, and shorter lengths, 10 cm to 23 cm. Additionally, Glidepath™ 10F Long-Term Hemodialysis Catheter has been designed to reach adequate flow rate requirements for pediatric patients or smaller adult patients for whom a physician desires a smaller catheter size.

    AI/ML Overview

    The provided text describes the regulatory clearance of a medical device, the GlidePath 10F Long-Term Hemodialysis Catheter, and its substantial equivalence to predicate devices. However, the document does not contain the specific details required to answer all parts of your request regarding acceptance criteria and the study that proves the device meets those criteria, especially in the context of an AI/ML device.

    This document is a 510(k) summary for a physical medical catheter, not an AI/ML device. Therefore, questions relating to AI/ML device performance (e.g., test sets, ground truth establishment by experts, adjudication methods, MRMC studies, standalone algorithm performance, training set details) are not applicable to the information provided.

    The document discusses performance data for the physical catheter, but these tests are related to mechanical and physical properties of the device, not diagnostic accuracy or AI performance.

    Here's a breakdown of what can be extracted from the provided text, and what cannot:

    1. A table of acceptance criteria and the reported device performance:

    The document lists various in vitro tests performed to demonstrate substantial equivalence, and states that the "subject device... meets all predetermined acceptance criteria of design verification and validation as specified by applicable standards, guidance, test protocols and/or customer inputs." However, it does not provide a table detailing the specific acceptance criteria (e.g., numerical thresholds for flow rates, tensile strength) or the reported numerical performance data for the GlidePath 10F Long-Term Hemodialysis Catheter against those criteria. It only lists the types of tests performed.

    Acceptance Criteria (Category)Reported Device Performance
    Catheter Tip (Damage After Flexure)Met predetermined acceptance criteria
    Tip TensileMet predetermined acceptance criteria
    Tunneler (Shaft to Tunneler Tensile)Met predetermined acceptance criteria
    Catheter Tip StiffnessMet predetermined acceptance criteria
    Surface InspectionMet predetermined acceptance criteria
    Catheter Insertion Over Split Sheath IntroducerMet predetermined acceptance criteria
    Assembly Leak ResistanceMet predetermined acceptance criteria
    Flow RatesMet predetermined acceptance criteria (comparable to pediatric reference device)
    Catheter CollapseMet predetermined acceptance criteria
    Catheter Tensile (Shaft to Bifurcation)Met predetermined acceptance criteria
    Catheter Tensile (Extension Leg to Bifurcation)Met predetermined acceptance criteria
    Catheter Tensile (Extension Leg to Connector)Met predetermined acceptance criteria
    Cuff SecurementMet predetermined acceptance criteria
    BurstMet predetermined acceptance criteria
    RecirculationMet predetermined acceptance criteria (comparable to pediatric reference device)
    Mechanical HemolysisMet predetermined acceptance criteria (comparable to pediatric reference device)
    Catheter Shaft StiffnessMet predetermined acceptance criteria
    Catheter RadiopacityMet predetermined acceptance criteria
    Thumb ClampsMet predetermined acceptance criteria
    Extension Legs (Knitting, Flow)Met predetermined acceptance criteria

    Note: The phrase "Met predetermined acceptance criteria" is inferred from the statement "The subject device... meets all predetermined acceptance criteria of design verification and validation." Specific numerical values for the criteria or the device's performance are not provided.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

    The document does not specify sample sizes for the "test set" in the context of diagnostic performance, nor does it describe data provenance for patient data. The "performance data" refers to in vitro mechanical/physical tests on the device itself, not a clinical study on patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

    Not applicable, as this is not an AI/ML device or a diagnostic device relying on expert interpretation for ground truth.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    Not applicable.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable, as this is not an AI/ML device.

    6. If a standalone (i.e. algorithm only, without human-in-the-loop performance) was done:

    Not applicable, as this is not an AI/ML device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    For the physical device performance tests, the "ground truth" or reference is established by the specifications in applicable standards, guidance documents, and internal test protocols. For example, a "burst" test would have a specified pressure it must withstand, and the device either passes or fails that engineering specification.

    8. The sample size for the training set:

    Not applicable, as this is not an AI/ML device.

    9. How the ground truth for the training set was established:

    Not applicable, as this is not an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K190527

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-03-05

    (367 days)

    Product Code
    Regulation Number
    876.5540
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The GlidePath™ Long-Term Hemodialysis Catheters are indicated for use in attaining short-term vascular access for hemodialysis, hemoperfusion or apheresis therapy. Access is attained via the internal jugular vein, subclavian vein, or femoral vein. Catheters longer than 40cm are intended for femoral vein insertion.

    The HemoStar™ and HemoStar™ XK Long-Term Hemodialysis Catheters are indicated for use in attaining short-term or long-term vascular access for hemodialysis, hemoperfusion or apheresis therapy. Access is attained via the internal jugular vein, subclavian vein or femoral vein. Catheters greater than 40 cm are intended for femoral vein insertion.

    Device Description

    The GlidePath™, HemoStar™, and HemoStar™ XK Long-Term Hemodialysis Catheters feature a dual-lumen shaft with double-D cross-sectional designs providing separate arterial and venous lumens, a molded bifurcation and, extending from the bifurcation, arterial and venous extension legs that connect to an external dialysis machine or blood cleansing device. The arterial (red) luer connector connects to the blood intake on the dialysis machine and the venous (blue) luer connector connects to the blood return line on the dialysis machine. Each extension leg has an atraumatic occlusion clamp which closes access to the lumen. Long-term dialysis catheters are packaged in a tray with legally marketed accessories intended for use during catheter placement. Both the staggered and symmetrical catheter tips contain holes that aid in the distribution of blood flow or aid in over-the-quidewire placement. The dialysis catheters are offered in various lengths in straight or Alphacurve™ configurations. Alphacurve™ configurations feature a pre-formed, ~225 ° curve. Long-term dialysis catheters are tunneled to provide greater distance between the skin and entry site into the vein and provide a physical barrier to the migration of skin organisms.

    The HemoStar™ and HemoStar™ XK Long-Term Dialysis Catheters feature a fixed, staggered distal tip. This tip design is identical to that of the predicate devices, HemoGlide™ Star Series and HemoGlide™ Star Series XK Long-Term Dialysis Catheters, cleared August 12, 2005 under K051748.

    The GlidePath™ Long-Term Dialysis Catheters, launched in February 2013, are a modification of the HemoStar™ Long-Term Dialysis Catheters. GlidePath™ catheters feature an optimized catheter shaft extrusion profile and a modified tip which is fixed and symmetrical, allowing for reduced luminal pressure during use. The GlidePath™ catheters also include a preloaded stylet for ease of placement. This accessory was previously cleared under K090101. The changes to the catheter were documented via internal note to file prior to launch in 2013.

    AI/ML Overview

    The provided text describes the 510(k) summary for the Bard Access Systems' GlidePath, HemoStar, and HemoStar XK Long-Term Hemodialysis Catheters. It details performance testing conducted to demonstrate substantial equivalence to predicate devices, but does not provide specific acceptance criteria values or detailed reported device performance in a table format. It also does not include information on sample sizes for test sets, data provenance, expert involvement for ground truth, adjudication methods, MRMC studies, standalone algorithm performance, or training set details as these are not relevant for a medical device clearance focused on physical and fluid dynamics performance instead of AI/software performance.

    Therefore, I cannot populate all the requested sections. However, I can summarize the information related to device performance and the study that "proves" the device meets acceptance criteria, based on the provided text.


    Acceptance Criteria and Device Performance for GlidePath, HemoStar, and HemoStar XK Long-Term Hemodialysis Catheters

    The provided 510(k) summary indicates that the devices underwent a series of bench tests to demonstrate substantial equivalence to predicate devices. The "study" mentioned here refers to these non-clinical bench tests. The report states that the devices met all predetermined acceptance criteria, implying that the performance observed during these tests was within acceptable limits established for each test. However, the specific quantitative acceptance criteria values and the reported device performance measurements are not detailed in this summary document.

    1. Table of Acceptance Criteria and Reported Device Performance

    Test NameAcceptance Criteria (Specific Values NOT Provided in Document)Reported Device Performance (Specific Measurements NOT Provided in Document)
    HemoStar™ and HemoStar XK™ Alphacurve™:
    Guidewire RemovalDefined criteria for ease and integrity of guidewire removal (e.g., no damage, smooth passage).Met predetermined acceptance criteria (implies smooth removal without issues).
    Priming VolumeDefined volumetric range for priming.Met predetermined acceptance criteria (implies volume within range).
    Shaft Bifurcation TensileDefined tensile strength thresholds for the bifurcation point (e.g., withstand specified force without failure).Met predetermined acceptance criteria (implies sufficient tensile strength).
    Flow RateDefined flow rate thresholds (e.g., maintain specified flow rate at given pressure).Met predetermined acceptance criteria (implies flow rates within acceptable ranges).
    Shaft TensileDefined tensile strength thresholds for the catheter shaft.Met predetermined acceptance criteria (implies sufficient shaft tensile strength).
    Burst TestingDefined pressure thresholds for burst resistance (e.g., withstand specified pressure without bursting).Met predetermined acceptance criteria (implies burst resistance meets thresholds).
    Kink TestingDefined criteria for resistance to kinking and recovery from kinking (e.g., no permanent occlusion after kinking, flow maintained).Met predetermined acceptance criteria (implies good kink resistance and recovery).
    GlidePath™ Modifications (including Alphacurve™):
    Shaft Stiffness With and Without StyletDefined stiffness ranges for ease of insertion and patient comfort.Met predetermined acceptance criteria (implies appropriate shaft stiffness).
    Catheter-to-Tunneler DetachmentDefined force thresholds for securement and detachment characteristics.Met predetermined acceptance criteria (implies secure attachment and controlled detachment).
    Tunneler Tip BendDefined criteria for flexibility and integrity of the tunneler tip (e.g., bend without breaking, return to original shape).Met predetermined acceptance criteria (implies appropriate tunneler tip properties).
    Stylet RemovalDefined criteria for ease and integrity of stylet removal.Met predetermined acceptance criteria (implies smooth stylet removal without issues).
    Guidewire removal and Visual InspectionDefined criteria for ease of guidewire removal and absence of visual defects after removal.Met predetermined acceptance criteria (implies smooth guidewire removal and no visual defects).
    Kink TestingDefined criteria for resistance to kinking and recovery from kinking.Met predetermined acceptance criteria (implies good kink resistance and recovery).
    Catheter Tip TensileDefined tensile strength thresholds for the catheter tip.Met predetermined acceptance criteria (implies sufficient tip tensile strength).
    Occlusion Simulation against vessel wall Initial PlacementDefined criteria for resistance to occlusion during initial placement (e.g., maintain flow, no collapse against vessel wall).Met predetermined acceptance criteria (implies minimal occlusion risk during initial placement).
    Occlusion Simulation against vessel wall Flow MethodDefined criteria for resistance to occlusion during use against a vessel wall (e.g., maintain flow, no collapse during pumping).Met predetermined acceptance criteria (implies maintained flow and resistance to collapse).
    Tip StiffnessDefined stiffness ranges for the catheter tip (e.g., for atraumatic insertion and positioning).Met predetermined acceptance criteria (implies appropriate tip stiffness).
    Flow and CollapseDefined criteria for maintained adequate flow rates and resistance to catheter collapse under negative pressure.Met predetermined acceptance criteria (implies good flow characteristics and collapse resistance).
    Recirculation Glass Heart ModelDefined acceptable levels of recirculation (e.g., less than a specified percentage).Met predetermined acceptance criteria (implies low levels of recirculation).
    RadiopacityDefined criteria for visibility under fluoroscopy (e.g., clearly visible radiopaque markers, good overall visibility).Met predetermined acceptance criteria (implies adequate radiopacity).
    Assembly TensileDefined tensile strength thresholds for assembled components.Met predetermined acceptance criteria (implies sufficient assembly tensile strength).
    Assembly Leak ResistanceDefined pressure thresholds for leak resistance of assembled components.Met predetermined acceptance criteria (implies leak-free assembly at specified pressures).
    BurstDefined pressure thresholds for burst resistance.Met predetermined acceptance criteria (implies burst resistance meets thresholds).
    Cuff TensileDefined tensile strength thresholds for the ingrowth cuff.Met predetermined acceptance criteria (implies sufficient cuff tensile strength).
    Priming VolumeDefined volumetric range for priming.Met predetermined acceptance criteria (implies volume within range).
    Introducer Sheath PassageDefined criteria for smooth and damage-free passage through an introducer sheath.Met predetermined acceptance criteria (implies smooth passage without damage).

    Summary of Device Performance:
    The document concludes that the subject devices "met all predetermined acceptance criteria of design verification and validation as specified by applicable standards, guidance, test protocols and/or customer inputs." This statement indicates that all tests listed above were successfully passed according to their respective acceptance criteria.

    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not specified in the provided summary.
    • Data Provenance: The tests were "bench tests" performed by Bard Access Systems, Inc. The data is non-clinical. No country of origin is specified, but the company is located in Utah, USA. The data is non-clinical and would be considered prospective in the context of device development and testing, meaning the tests were designed and executed to evaluate the new/modified devices.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    • This is not applicable as the study involves bench testing of physical device characteristics, not interpretation of medical images or patient outcomes. The "ground truth" for these tests is established by engineering specifications, recognized standards, and established test protocols.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • This is not applicable for bench testing. Test results are typically objective measurements against predefined specifications.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • This is not applicable. The devices are physical hemodialysis catheters, not AI software or diagnostic imaging aids.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • This is not applicable. The devices are physical hemodialysis catheters.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • For bench testing, the "ground truth" is defined by engineering specifications, recognized industry standards (e.g., ISO, ASTM relevant to medical devices), and internal test protocols that establish the functional and safety requirements for the catheter's physical properties, fluid dynamics, and material integrity.

    8. The sample size for the training set:

    • This is not applicable. There is no machine learning or AI component to this device.

    9. How the ground truth for the training set was established:

    • This is not applicable. There is no machine learning or AI component to this device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K182443

    Validate with FDA (Live)

    Date Cleared
    2019-05-31

    (266 days)

    Product Code
    Regulation Number
    876.5540
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Pristine™ Hemodialysis Catheter is indicated for acute and chronic hemodialysis, apheresis and infusion. It may be inserted percutaneously or by cut down. Catheters with greater than 40cm implant length are indicated for femoral placement.

    Device Description

    Pristine™ Hemodialysis Catheter is a chronic hemodialysis catheter consisting of a dual lumen radiopaque shaft with a pre-formed split tip, which enables a long-term vascular access for hemodialysis and apheresis. The proximal end of the catheter features two color-coded luer adapters. The luer adapters are connected to clear extension tubes. Each extension tube contains a clamp and is connected to the catheter junction and suture wings (hub). The distal end of the catheter hub is connected to the dual lumen catheter shaft. The shaft contains a cuff and extends to a symmetrical split tip. The design of the catheter's distal tip includes a split, symmetric tip with notches and without side-holes nor slots. The symmetric tip design allows a spatial separation between the distal ends of the two lumens. The Pristine™ Hemodialysis Catheter is packed along with a Tunneler and two sealing caps in a vented blister tray and lid, sealed within a Tyvek pouch. The packed catheter and accessories set is provided as a sterile, single-use device, and is sterilized using a validated ethylene oxide process.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the Pristine™ Hemodialysis Catheter. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving that the device meets specific acceptance criteria in a clinical study against a defined ground truth.

    Therefore, many of the requested points, such as detailed acceptance criteria with reported device performance numbers, sample sizes for test and training sets, expert qualifications, adjudication methods, multi-reader multi-case studies, and standalone algorithm performance, are not applicable or not provided in this type of regulatory document.

    However, based on the information given, here's what can be extracted:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document states that "All tests passed and met the predefined acceptance criteria," but it does not provide the specific numerical acceptance criteria or the quantitative reported performance for most of the tests. It only lists the tests performed.

    Test TypeAcceptance Criteria (Not explicitly stated numerically, but implied as "met predefined acceptance criteria")Reported Device Performance (Not explicitly stated numerically, but implied as "passed")
    Biocompatibility TestingCompliance with ISO 10993-1, ISO 10993-4, ISO 10993-5, ISO 10993-6, ISO 10993-10, ISO 10993-11, ISO 10993-12, ISO 10993-17, ANSI/AAMI T72:2011/(R)2016 and ASTM F 756 standards.All tests completed with passing results.
    Cytotoxicity StudyCompliance with ISO 10993-5: 2009Passed
    Sensitization TestCompliance with ISO 10993-10: 2010Passed
    Irritation StudyCompliance with ISO 10993-10:2010Passed
    Acute Systemic Injection StudyCompliance with ISO 10993-11: 2010Passed
    Pyrogen StudyCompliance with ISO 10993-11:2010 and ANSI/AAMI T72:2011/(R)2016Passed
    HemocompatibilityCompliance with ASTM F 756, ISO 10993-12:2012Passed
    Complement ActivationCompliance with ANSI/AAMI/ISO 10993-12Passed
    ThrombogenicityCompliance with ISO 10993-4Passed
    Implantation (13 weeks)Compliance with ISO 10993-6Passed
    Extractable/leachable AnalysisCompliance with ISO 10993-17Passed
    Subacute/subchronic toxicity, Genotoxicity, Chronic toxicity and CarcinogenicityEvaluation by extractable/leachable analysis according to ISO 10993-17Passed
    Tunneler and Caps BiocompatibilitySame applicable ISO standards as above for individual tests with similar natureAll tests completed with passing results.
    Sterilization, Packaging, Shelf LifeCompliance with ISO 11135 and support of labeled shelf life.All tests successfully completed.
    Bench Testing (Per FDA guidance Implanted Blood Access Devices for Hemodialysis, 2016)Predefined acceptance criteria for: Dimensional Attributes, Distal Tip Visual Inspection, Luer Dimensions, Shaft Radiopacity, Flow vs. Pressure, Nominal Flow vs. Pressure, Air leakage during aspiration, Leak, Chemical resistance (Conditioning), Clamp fatigue (Conditioning), Tensile, Kink resistance, Priming volume, Tip recirculation, Mechanical Hemolysis, Hydration, Accessories compatibility analysis.All tests passed and met the predefined acceptance criteria.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not specified. The document refers to "testing" and "performance data" without detailing the number of devices or components tested for each bench or biocompatibility test.
    • Data Provenance: Not explicitly stated as retrospective or prospective clinical data. The tests performed are laboratory/bench tests and biocompatibility studies, not clinical studies on human subjects. The data would be generated in a lab setting.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • Not applicable in the context of this document. The "ground truth" for the tests mentioned (biocompatibility, bench testing) is defined by established engineering and biological standards (e.g., ISO, ASTM). These are objective measurements against specified criteria, not subjective expert interpretations that require consensus.

    4. Adjudication Method for the Test Set

    • Not applicable. As described above, the tests are objective measurements against defined standards, not subjective assessments requiring an adjudication process.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

    • No, a Multi Reader Multi Case (MRMC) comparative effectiveness study was not done. This type of study is typically performed for diagnostic devices, especially those involving human interpretation of images, to assess the impact of AI on reader performance. The Pristine Hemodialysis Catheter is a physical medical device, not a diagnostic AI system.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    • No, a standalone performance assessment (in the context of an algorithm) was not done. This device does not involve an algorithm for performance, but rather physical and biological performance characteristics.

    7. The Type of Ground Truth Used

    • The "ground truth" for the performance data in this submission is based on adherence to international and national standards (e.g., ISO, ASTM, ANSI/AAMI) for medical device performance and biocompatibility. For bench testing, this means meeting predefined engineering specifications and functional requirements. For biocompatibility, it means demonstrating that the materials are safe for biological contact as per established protocols.

    8. The Sample Size for the Training Set

    • Not applicable. This device is a physical medical device, not an AI/ML algorithm that requires a training set.

    9. How the Ground Truth for the Training Set was Established

    • Not applicable. No training set is used for this type of device submission.