Found 32 results

510(k) Data Aggregation

    K Number
    K240523

    Validate with FDA (Live)

    Date Cleared
    2024-11-15

    (266 days)

    Product Code
    Regulation Number
    884.6100
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    VitaVitro Single Lumen Oocyte Retrieval Needle is intended for ultrasound guided transvaginal aspiration of oocytes from ovarian follicles for patients undergoing Assisted Reproductive procedures.

    VitaVitro Double Lumen Oocyte Retrieval Needle is intended for ultrasound guided transvaginal aspiration and flushing of oocytes from ovarian follicles for patients undergoing Assisted Reproductive procedures.

    Device Description

    The Oocyte Retrieval Needle is composed of a stainless steel needle protective sleeve, needle handle, suction catheter, silicone plug, vacuum catheter and taper joint. The device includes both dual lumen (ORND) and single-lumen needles (ORNS). Three models are available: ORNS-17G, ORNS-18G and ORND-17G. The oocyte retrieval needles are available in 17 or 18 gauge sizes and two lengths: 33 and 35 cm. The needles are provided with a suction catheter of 100 cm length and a vacuum catheter with a length of 35 cm, and the ORND-17G also includes a flushing catheter of 100 cm length. Needles have an echogenic tip that enhances the visualization of the needle tip under ultrasound.

    This product is a disposable, sterile medical device, sterilized by ethylene oxide.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the VitaVitro® Oocyte Retrieval Needle.

    This document describes a 510(k) premarket notification for a medical device, which typically compares a new device to an existing legally marketed predicate device to demonstrate substantial equivalence. It is important to note that 510(k) submissions generally do not involve clinical trials to prove efficacy against specific clinical outcomes in the same way a PMA (Premarket Approval) would. The focus is on demonstrating that the new device is as safe and effective as the predicate based on technological characteristics and performance testing.

    Therefore, many of the questions regarding "study that proves the device meets the acceptance criteria" in a clinical sense, particularly those related to human-in-the-loop performance, effect sizes, and detailed ground truth establishment from patient data, are not typically part of a 510(k) submission for this type of device. The acceptance criteria here are primarily based on non-clinical performance testing demonstrating the device's technical specifications and safety.


    1. A table of acceptance criteria and the reported device performance

    Parameter / Acceptance CriteriaReported Device Performance
    Biocompatibility
    Cytotoxicity (ISO 10993-5:2009)Non-cytotoxic
    Sensitization (ISO 10993-10:2021)Non-sensitizing
    Irritation (ISO 10993-23:2021)Non-irritating
    Mouse Embryo Assay (MEA)
    ≥80% embryos developed to expanded blastocyst at 96h (per 2021 FDA guidance)>80% embryos developed to expanded blastocyst at 96h
    Endotoxin Testing
    < 5 EU/device (LAL testing by Gel Clot assay per USP<85>)< 5 EU/device
    Sterility Testing
    Sterile, ethylene oxide, SAL of 10-6 (ISO 11135:2014)No microbial growth (meets SAL of 10-6)
    Residual Ethylene Oxide (ISO 10993-7:2008)Compliant with ISO 10993-7
    Shipping and Package Integrity Testing
    Simulated shipping per ASTM D 4169:2016 DC13Maintained integrity post-shipping
    Package integrity per ASTM F88/F88M:2021 (seal peel strength), ASTM F1929:2015 (dye penetration)Maintained package integrity (meets seal peel strength and dye penetration requirements)
    Mechanical Performance Testing
    Appearance per ISO 7864:2016Conforms to appearance standards
    Dimensional verificationDimensions verified to specifications
    Hydraulic resistance per ISO 20695:2020, Annex DMeets hydraulic resistance requirements
    Vacuum leakageNo unacceptable vacuum leakage
    Tensile strength test per ISO 20695:2020, Annex CMeets tensile strength requirements
    Flow rate per ISO 20695:2020, Annex EMeets flow rate specifications
    Puncture needle test per ISO 9626:2016, Annex B-D (Rigidity, Toughness, Corrosion resistance)Meets rigidity, toughness, and corrosion resistance standards
    Needle penetration force per ISO 7864:2016, Annex DMeets needle penetration force requirements
    Ultrasound detectability testDetectable under ultrasound
    Collection tube compatibility testCompatible with collection tubes
    Shelf-life Testing
    Maintained product specifications (endotoxin, MEA, sterility, package integrity) over proposed shelf-life (2 years)Specifications met over the 2-year shelf-life period (initial and aged samples tested for relevant parameters)

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The text does not provide specific sample sizes for each individual test (e.g., how many devices for sterility, how many mouse embryos for MEA, etc.). It only lists the types of tests performed.

    As these are non-clinical (laboratory/bench) tests for a 510(k) submission, the "data provenance" in terms of country of origin or retrospective/prospective human data is not applicable. The tests were performed on the device itself under controlled laboratory conditions, likely at the manufacturer's facility or a contracted testing lab.


    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This is not applicable as the studies are non-clinical performance and safety tests, not human-centered clinical studies that require expert-established ground truth. The "ground truth" for these tests are objective, measurable criteria defined by the relevant ISO standards and FDA guidance.


    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. Adjudication methods like "2+1" or "3+1" are used in clinical studies where human interpretation of medical images or patient data is involved to establish a "ground truth" often in cases of disagreement among experts. These performance tests rely on objective laboratory measurements and standardized methodologies.


    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC comparative effectiveness study was done, nor would it be expected for this type of device (an oocyte retrieval needle) in a 510(k) submission. This is a physical, sterile, single-use medical instrument, not an AI-powered diagnostic tool. The concept of "human readers improve with AI assistance" is irrelevant here.


    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is also not applicable. This device is not an algorithm or software. It is a physical medical instrument.


    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" here is defined by:

    • International Standards (e.g., ISO, ASTM): These standards specify objective, measurable criteria and methodologies acceptable for evaluating medical device performance (e.g., sterility, biocompatibility, mechanical properties).
    • FDA Guidance Documents: Specifically mentioned is the 2021 FDA guidance document for Mouse Embryo Assay, which defines the acceptance criteria for MEA performance.
    • Device Specifications: The manufacturer's own design specifications for dimensions, flow rate, vacuum maintenance, etc., which are verified through testing.

    8. The sample size for the training set

    This device does not involve a "training set" in the context of machine learning or AI. Therefore, this question is not applicable.


    9. How the ground truth for the training set was established

    As there is no training set for this device, this question is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K232163

    Validate with FDA (Live)

    Date Cleared
    2024-02-22

    (216 days)

    Product Code
    Regulation Number
    884.6100
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Lotus Single Lumen Ovum Aspiration Needle is used for ultrasound guided transvaginal aspiration of oocytes from ovarian follicles for patients undergoing Assisted Reproductive procedures.

    Lotus Double Lumen Ovum Aspiration Needle is used for ultrasound guided transvaginal aspiration and flushing of oocytes from ovarian follicles for patients undergoing Assisted Reproductive procedures.

    Device Description

    The subject devices are single use sterile devices used for ultrasound-guided transvaginal aspiration of oocytes from ovarian follicles.

    The Lotus Single Lumen Ovum Aspiration Needle consists of a stainless steel needle with a handle, which is connected to an aspiration tube with a silicone stopper at the other end. The stopper can fit onto a test tube where the oocytes are collected. The stopper is connected to a vacuum tube with a vacuum tubing connector at the other end. The vacuum tubing connector is a standard luer lock connector that enables connection to a vacuum pump to provide aspiration suction. The Lotus Single Lumen Ovum Aspiration Needle is 35 cm in length and has device variations from 21 to 16 gauge with aspiration tubing of 600 or 900 mm in length.

    The Lotus Double Lumen Ovum Aspiration Needle consists of a stainless steel needle with a Y shaped handle. The Y-shaped handle has two ports: the central port through which oocytes are aspirated via the aspiration lumen, and a secondary side port to allow flushing of follicles via the flushing lumen. The central port is connected to an aspiration tube with a silicone stopper at the other end. The stopper can fit onto a test tube where the oocytes are collected. The stopper has a vacuum connector that is a standard luer lock connector that enables connection to a vacuum pump to provide aspiration suction. The side port is connected to a flushing tube with a flushing connector at the other end, which is a standard luer lock connector that can be connected to a compatible syringe for flushing of follicles. The Lotus Double Lumen Ovum Aspiration Needle is 17 gauge with an aspiration tubing length of 900 mm and with device variations of 30 or 35 mm needle length and flushing tubing of 700 or 1000 mm in length.

    AI/ML Overview

    This document is an FDA 510(k) Premarket Notification for medical devices, specifically ovum aspiration needles. It does not describe a study involving an AI/Machine Learning device. Therefore, it is not possible to extract information about acceptance criteria or study details relevant to AI/ML device performance from this text.

    The document discusses traditional medical device testing, such as:

    • Biocompatibility testing: Cytotoxicity, Sensitization, Intracutaneous Irritation.
    • Sterilization testing: Ethylene Oxide (EO) sterilization, residual testing.
    • Simulated transportation: Adherence to ASTM D4169-22.
    • Stability and Shelf Life testing: accelerated aging, package integrity, mechanical performance (appearance, dimensions, needle stiffness, breakage resistance, corrosion resistance, tip penetration force, needle bond strength, flow rate, ultrasound detectability, tubing bond strength, leak, aspiration test, component compatibility).
    • Mouse Embryo Assay (MEA): This is a biological test to assess the non-toxicity of materials to embryos, not an AI performance metric. The acceptance criterion for 1-Cell MEA was ≥80% embryos developed to expanded blastocyst at 96 hours after 30 minute exposure.
    • Endotoxin evaluation: Acceptance criterion was ≤ 20 EU/device.

    Since the request specifically asks for details related to AI/Machine Learning device performance, and this document pertains to a traditional, non-AI medical device, I cannot fulfill the request as stated.

    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Allwin Ovum Pickup Needles are used for laparoscopic or ultrasound guided transvaginal retrieval of oocytes from ovarian follicles.

    Device Description

    Allwin Ovum Pickup Needles (ACE Single Lumen Ovum Pickup Needle, OVUMPICK Double Lumen Ovum Pickup Needle, ACE-FL Single Lumen Ovum Pickup Needle with Flushing Line, ACE-M Manual Single Lumen Ovum Pickup Needle).

    AI/ML Overview

    This FDA letter is a notification of substantial equivalence for a medical device (Allwin Ovum Pickup Needles). It does not contain information about acceptance criteria or specific studies that prove the device meets these criteria. The document explicitly states:

    "We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices..."

    This means the FDA has determined the new device is as safe and effective as a legally marketed device already on the market, based on the information provided by the manufacturer in their 510(k) submission. However, the details of the studies and acceptance criteria used by the manufacturer to demonstrate this substantial equivalence are not included in this letter. The letter is the FDA's final decision, not the manufacturer's submission itself.

    Ask a Question

    Ask a specific question about this device

    K Number
    K191291

    Validate with FDA (Live)

    Date Cleared
    2019-06-12

    (29 days)

    Product Code
    Regulation Number
    884.6100
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Wallace Dual Lumen Oocyte Recovery System is a sterile, single-use device for ultrasonic-guided transvaginal collection of oocytes from the ovarian follicles.

    Device Description

    The Wallace Dual Lumen Oocyte Recovery System consists of a 33 cm dual lumen stainless steel needle with a plastic needle hub, and aspiration, flushing, and vacuum tubing. The needle hub acts as a handle and is designed for the user to hold between the thumb and index finger. It has two ports: the central port through which oocytes are aspirated via the central lumen, and a secondary side port to allow flushing of follicles via the secondary lumen that is attached to the flushing tubing. The needle aspiration tubing connects to a silicone bung that allows connection to a sample tube. The silicone bung is also connected to the vacuum tubing that allows connection to a vacuum source. The Wallace Dual Lumen Oocyte Recovery System is available in 16 or 17-gauge color-coded sizes (16G-Blue and 17G-Red). The needles are available in three aspiration tubing lengths: 500, 750, and 950 mm. The system includes flushing tubing of 700 mm length and vacuum tubing of 500 mm length. The device is provided sterile and is for single-use only.

    AI/ML Overview

    The provided text is a 510(k) summary for a medical device (Wallace Dual Lumen Oocyte Recovery System) and does not contain information about studies related to AI or algorithm performance. The content focuses on demonstrating substantial equivalence to a predicate device through non-clinical performance testing of physical attributes, such as needle size, biocompatibility, stability, shelf life, and mechanical performance.

    Therefore, I cannot provide the requested information regarding acceptance criteria and studies that prove a device meets criteria in the context of AI/algorithm performance, as these details are not present in the given document.

    The document indicates that the device is a sterile, single-use system for ultrasonic-guided transvaginal collection of oocytes and does not mention any AI or software components.

    Ask a Question

    Ask a specific question about this device

    K Number
    K182959

    Validate with FDA (Live)

    Date Cleared
    2019-05-10

    (197 days)

    Product Code
    Regulation Number
    884.6100
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Wallace Dual Lumen Oocyte Recovery System in a sterile single-use device for ultrasonic-guided transvaginal collection of oocytes from the ovarian follicles.

    Device Description

    The Wallace Dual Lumen Oocyte Recovery System consists of a dual lumen stainless steel needle with a plastic needle hub, and aspiration, flushing, and vacuum tubing. The needle hub acts as a handle and is designed for the user to hold between the thumb and index finger. It has two ports: the central port through which oocytes are aspirated via the central lumen, and a secondary side port to allow flushing of follicles via the secondary lumen that is attached to the flushing tubing. The needle aspiration tubing connects to a silicone bung that allows connection to a sample tube. The silicone bung is also connected to the vacuum tubing that allows connection to a vacuum source.

    The Wallace Dual Lumen Oocyte Recovery System consists of a 17 gauge needle that is 33 cm long, and three aspiration tubing lengths: 500, 750, and 950 mm. The system includes flushing tubing of 700 mm length and vacuum tubing of 500 mm length. The device is provided sterile and is for single-use only.

    AI/ML Overview

    Based on the provided text, the device in question is the Wallace Dual Lumen Oocyte Recovery System, an ovulation retrieval needle. The document does not describe a study involving an AI/algorithmic device or human readers. Instead, it details the non-clinical performance testing conducted to demonstrate substantial equivalence to a predicate device.

    Therefore, many of the requested sections (e.g., sample size for test/training sets, data provenance, number of experts for ground truth, MRMC study, standalone performance, training set ground truth establishment) are not applicable as they pertain to the evaluation of an AI or algorithm, which is not the subject of this 510(k) summary.

    Here's an analysis of the available information:

    Acceptance Criteria and Device Performance for the Wallace Dual Lumen Oocyte Recovery System

    The device under review is an assisted reproduction needle, not an AI or algorithm. The acceptance criteria and performance data provided relate to the physical and biological characteristics of the device to demonstrate its safety and effectiveness, primarily through comparison to a legally marketed predicate device.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document describes various non-clinical performance tests and their acceptance criteria (implicitly met as the conclusion states the device passed all testing).

    Test TypeAcceptance CriteriaReported Device Performance
    Sterilization TestingPer ISO 11135:2014; Sterility Assurance Level (SAL) 10^-6Passed all testing in accordance with internal requirements and applicable standards. Implies SAL 10^-6 was met.
    Ethylene Oxide (EtO) ResidualsPer ISO 10993-7: 2008 / (R)2012Passed all testing in accordance with internal requirements and applicable standards.
    Biocompatibility: CytotoxicityPer ISO 10993-5: 2009Passed.
    Biocompatibility: IrritationPer ISO 10993-10: 2010Passed.
    Biocompatibility: SensitizationPer ISO 10993-10: 2010Passed.
    Stability and Shelf Life (Accelerated Aging)Two years shelf life (tested at baseline and after 2 years accelerated aging per ASTM F1980-16)Passed. Device demonstrated stable performance over a two-year shelf life.
    Package Integrity (Seal Tensile Strength)Per ASTM F88/F88M-15Passed.
    Package Integrity (Seal Peel)Per ASTM F1886/F1866M-16Passed.
    Mouse Embryo Assay (MEA)≥80% embryos expanded to blastocyst at 96 hours (for one-cell mouse embryos incubated in extracts of the device)Passed.
    Endotoxin≤20 EU/device (using Gel-Clot Limulus Amoebocyte Lysate (LAL) method per ANSI/AAMI ST72-2011 and USP <85>)Passed.
    Mechanical Performance TestingSpecific criteria not detailed, but tests included: Joint strength tensile testing; Needle point penetration force testing; Needle stiffness; Tubing flow rate; Tubing leak testingPassed all mechanical performance tests.
    Simulated Shipping & DistributionPer ISTA 3A: 2008; subsequent visual inspection, bubble leak per ASTM F2096-11, and mechanical performance testing (same as above)Passed. Device and packaging maintained integrity and performance after simulated shipping and distribution conditioning, including visual inspection, bubble leak, and mechanical performance testing.

    2. Sample Size Used for the Test Set and the Data Provenance

    • Sample Size: Not explicitly stated for each test (e.g., number of needles tested for mechanical performance, number of packages for integrity). However, the methods refer to standard test procedures (e.g., ISO, ASTM), which typically define sample sizes.
    • Data Provenance: The document describes non-clinical laboratory testing of the device itself, not human clinical data or imaging data. Therefore, concepts like country of origin or retrospective/prospective are not applicable.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    • Not Applicable. This device is a physical medical instrument, not an AI/algorithmic software. "Ground truth" in this context refers to the defined scientific and engineering standards and methods for assessing the device's physical, chemical, and biological properties, which are established by standard organizations (e.g., ISO, ASTM, USP) and verified by qualified laboratory personnel.

    4. Adjudication Method for the Test Set

    • Not Applicable. As this is non-clinical performance testing of a physical device, there is no "adjudication" of expert opinions in the sense of image interpretation. Test results are objective measurements against defined standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    • No. This is not an AI/algorithmic device or diagnostic tool. An MRMC study would be irrelevant.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    • Not Applicable. This is a physical medical device. "Standalone performance" in this context refers to the device's ability to meet its specifications independently, which is what the non-clinical performance testing aimed to demonstrate.

    7. The Type of Ground Truth Used

    • Defined Standards and Measurable Performance Metrics: The "ground truth" for this device's performance is based on established international and national standards (e.g., ISO, ASTM, ANSI/AAMI, USP) for sterility, biocompatibility, mechanical properties, and packaging integrity. Tests evaluate quantitative and qualitative measures directly against these benchmarks (e.g., specific thresholds for endotoxin, percentage of embryo development, strength limits, leak absence).

    8. The Sample Size for the Training Set

    • Not Applicable. This refers to an AI/algorithmic device, not a physical medical instrument. There is no concept of a "training set" for the type of evaluation described.

    9. How the Ground Truth for the Training Set was Established

    • Not Applicable. Again, this is for AI/algorithmic development. The "ground truth" for evaluating this physical device is established through adherence to recognized industrial and medical device testing standards and protocols.
    Ask a Question

    Ask a specific question about this device

    K Number
    K171625

    Validate with FDA (Live)

    Date Cleared
    2018-01-12

    (224 days)

    Product Code
    Regulation Number
    884.6100
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Single Lumen Ovum Aspiration Needles are used for laparoscopic or ultrasound guided transvaginal aspiration and flushing of oocytes from ovarian follicles for patients undergoing Assisted Reproductive procedures.

    Device Description

    Single Lumen Ovum Aspiration Needles consist of a stainless-steel needle with a manipulating handle which may be connected to a fluorinated ethylene-propylene (FEP) or polytetrafluorethylene (PTFE) tube, which in turn is connected to a Silicone stopper (bung). The stopper has a Luer lock fitting directly to the stopper or connected to a length of tube to allow connection to a vacuum pump which provides aspiration suction. The stopper is placed in a 14mL test tube which acts as a collection reservoir. Needles are available in 16 to 21 gauge sizes and lengths from 25 to 35 cm. Needles have an echogenic tip that enhances the visualization of the needle tip under ultrasound. The Single Lumen Ovum Aspiration Needles are sterile and single-use devices.

    AI/ML Overview

    Here's an analysis of the provided FDA 510(k) summary, specifically addressing the acceptance criteria and the study that proves the device meets them:

    The document is a 510(k) premarket notification for the "Single Lumen Ovum Aspiration Needles." This is a medical device, and the focus of the regulatory claim is substantial equivalence to a predicate device, not necessarily a demonstration of clinical superiority or a deep dive into AI performance. Therefore, many of the typical AI/ML acceptance criteria and study details you'd expect are not present.

    The "acceptance criteria" here largely refer to the performance standards and modifications made to demonstrate that the new device is as safe and effective as the predicate device.


    1. Table of acceptance criteria and the reported device performance

    Acceptance Criteria (Test Performed)Reported Device Performance
    Biocompatibility (ISO 10993-1:2009)Passed for Cytotoxicity, Sensitization, and Irritation.
    - Cytotoxicity (ISO 10993-5:2009)Passed
    - Sensitization (ISO 10993-10:2010)Passed
    - Irritation (ISO 10993-10:2010)Passed
    Mouse Embryo Assay≥80% development to blastocyst at 72 hours (compared to control group).
    Endotoxin testing (USP <85>)< 20 EU/Device
    Mechanical performance testingPassed for all specific aspects listed below.
    - Negative pressure leak testPassed
    - Tensile strength of the tubing to cannulaPassed
    - Tensile strength of the tubing to bungPassed
    - Tensile strength of joint between cannula and handlePassed
    - Tensile strength of 4 Fr tubingPassed
    - Tensile strength of the joint between guide needle cannula and hubPassed
    - Needle stiffnessPassed
    Stability testingPassed for all specific aspects listed below after three years of aging.
    - Negative pressure leak test after three years of agingPassed
    - Tensile testing of joint between handle and needle cannula after three years of agingPassed
    - Torque testing of joint between guide needle cannula and hub after three years of agingPassed

    2. Sample size used for the test set and the data provenance

    The document does not specify exact sample sizes for each test set beyond mentioning "device extracts" for the Mouse Embryo Assay. The studies are pre-clinical/design verification and validation studies rather than clinical trials with human participant data. Therefore, concepts like "country of origin of the data" or "retrospective/prospective" as relevant to patient data provenance do not directly apply here. These studies are conducted by the manufacturer (William A. Cook Australia Pty Ltd).


    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable to this type of regulatory submission. For a medical device like an ovum aspiration needle, "ground truth" is established through standardized laboratory testing and engineering specifications to confirm physical and biological properties, not expert clinical interpretation of data.


    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable. Adjudication methods like 2+1 or 3+1 are used for establishing ground truth in clinical imaging studies or other diagnostic assessments involving multiple human readers, which is not the nature of the studies described here.


    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    There was no MRMC study conducted. This device is a physical medical instrument (needle) and does not involve AI or human readers in the context of diagnostic interpretation.


    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This information is not applicable. The device is not an algorithm or an AI product.


    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the performance data presented is derived from standardized laboratory testing, engineering specifications, and established biological assays. For example:

    • Biocompatibility: Conformity to ISO 10993 standards.
    • Mouse Embryo Assay: Direct observation of embryo development under controlled conditions.
    • Endotoxin testing: Quantitative measurement against USP <85> limits.
    • Mechanical testing: Measurements against predefined engineering specifications for strength, leak resistance, and stiffness.

    8. The sample size for the training set

    This information is not applicable. The device is not an AI/ML system and therefore does not have a "training set" in the computational sense. The design verification and validation tests are performed on representative samples of the manufactured device.


    9. How the ground truth for the training set was established

    This information is not applicable as there is no training set for an AI/ML algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K171611

    Validate with FDA (Live)

    Date Cleared
    2017-12-15

    (197 days)

    Product Code
    Regulation Number
    884.6100
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Double Lumen Ovum Aspiration Need for laparoscopic or ultrasound guided transvaginal aspiration and flushing of oocytes from ovarian follicles for patients undergoing Assisted Reproductive procedures.

    Device Description

    Double Lumen Ovum Pick-up Needles consist of a double lumen stainless steel needle with a manipulating handle. The manipulating handle has two ports: the central port through which oocytes are aspirated via the central lumen, and a secondary side port to allow flushing of follicles via a secondary lumen. The needle is intended to be used to aspirate oocytes from follicles of human ovaries via a laparoscopic or ultrasound guided transvaginal path.

    The Double Lumen Ovum Aspiration Needles are sterile and single-use devices.

    The Double Lumen Ovum Aspiration Needles are available in 16 or 17 gauge sizes and three lengths: 30, 33 and 35 cm. The needles are provided with an aspiration tube of 75 cm length and a vacuum tube with a length of 70 or 100 cm.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Test/Performance MetricAcceptance Criteria (Threshold/Standard)Reported Device Performance (Result)
    BiocompatibilityCytotoxicityConformance to ISO 10993-5:2009Conformed
    SensitizationConformance to ISO 10993-10:2010Conformed
    IrritationConformance to ISO 10993-10:2010Conformed
    Embryo CompatibilityMouse Embryo Assay (MEA)Not explicitly stated, but implied to show comparable development to control groupPercent of embryos developed to expanded blastocyst stage within 72 hours, assessed in comparison with the control group (implied positive result)
    Sterility/PurityEndotoxin TestingConformance to USP <85>Conformed
    Mechanical PerformanceNegative Pressure Leak TestNot explicitly stated, but implied to meet functional requirementsPerformed and implied to meet requirements
    Tensile Strength (Tubing to Cannula)Not explicitly stated, but implied to meet functional requirementsPerformed and implied to meet requirements
    Tensile Strength (Tubing to Bung)Not explicitly stated, but implied to meet functional requirementsPerformed and implied to meet requirements
    Durability/Shelf LifeNegative Pressure Leak Test (after aging)Not explicitly stated, but implied to maintain functional requirements after 3 yearsPerformed after three years of real time aging and implied to maintain requirements

    2. Sample Size Used for the Test Set and Data Provenance

    The document describes design verification and validation activities for a medical device (Double Lumen Ovum Aspiration Needles), not a study involving human data, algorithms, or test sets in the context of medical imaging or AI.

    Therefore, the concepts of "sample size used for the test set," "data provenance," "country of origin of the data," and "retrospective or prospective" are not applicable to this type of device submission. The tests were performed on the device itself or its materials.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of those Experts

    This information is not applicable as the document describes physical/material testing of a medical device, not a diagnostic or AI-driven system requiring expert ground truth establishment.

    4. Adjudication Method for the Test Set

    This information is not applicable for the reasons stated above.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable. This submission is for a physical medical device, not an AI-based diagnostic tool.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    This information is not applicable. This submission is for a physical medical device, not an algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" for this device is based on established industry standards (e.g., ISO, USP) and internal functional requirements for biocompatibility, mechanical performance, and sterility. For the Mouse Embryo Assay (MEA), the "ground truth" would be the development of a control group of embryos.

    8. The Sample Size for the Training Set

    This information is not applicable. There is no "training set" as this is not an algorithm or AI model.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable. There is no "training set" or associated ground truth establishment for an AI model.

    Ask a Question

    Ask a specific question about this device

    K Number
    K172050

    Validate with FDA (Live)

    Date Cleared
    2017-11-21

    (138 days)

    Product Code
    Regulation Number
    884.6100
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Intended for flushing and/or aspiration of oocytes from ovarian follicles.

    Device Description

    The Follicle Aspiration Set. Reduced Single Lumen proposed in this submission is intended for ultrasound-guided transvaginal aspiration and flushing of oocytes from ovarian follicles. This product is comprised of the following components:

    • A stainless steel needle .
    • A silicone cork to be pushed into the opening of a sampling tube .
    • A length of TPE-O tubing for aspiration and flushing ●
    • An adapter for flushing .
      The subject device is provided sterile and is for single use only. It comes in four versions with the same design but different dimensions, as described below.
    AI/ML Overview

    The provided text is a 510(k) summary for the Vitrolife Sweden AB Follicle Aspiration Set, Reduced Single Lumen. This document demonstrates the device's substantial equivalence to a predicate device, rather than providing a detailed study report for meeting specific performance criteria typically associated with AI/ML-enabled devices.

    Therefore, many of the requested categories (such as sample size for test set, number of experts for ground truth, adjudication method, MRMC study, standalone performance, training set sample size, and ground truth establishment for training set) are not applicable to this type of device and submission. This is a medical device clearance based on engineering performance and material compatibility, not on diagnostic accuracy derived from AI analysis of data.

    However, I can extract the acceptance criteria and reported performance for the engineering and biological tests conducted.

    Acceptance Criteria and Device Performance for Vitrolife Sweden AB Follicle Aspiration Set, Reduced Single Lumen

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Sterilization ValidationMet ISO 11137-2:2013 standards. (Specific performance metric not detailed, but implied compliance).
    Endotoxin Testing≤ 1.2 EU/device (determined by LAL method per USP <85>).
    Breakage and Elasticity TestingMet requirements in ISO 9626:2016 for stainless steel needle. (Specific performance metric not detailed, but implied compliance).
    Accelerated Shelf-Life Testing (3-year shelf-life):
    - Mouse Embryo Assay (MEA): ≥ 80% embryos expanded to blastocysts within 96 hours.One-cell mouse embryos exposed to device extracts. Percent of embryos developed to expanded blastocyst stage within 96 hours were assessed in comparison with the control group. (Specific percentage not detailed, but implied compliance indicating ≥ 80% was achieved or demonstrated equivalence to control).
    - Visual InspectionPerformed per ASTM F1886/F1886M-09. (Specific results not detailed, but implied compliance).
    - Dye Penetration TestingPerformed per ASTM F1929-12. (Specific results not detailed, but implied compliance).
    - Seal Strength TestingPerformed per ASTM F88/F88M-09. (Specific results not detailed, but implied compliance).
    - Vacuum Testing: Operate as intended without damage under -500 mmHg vacuum pressure.Demonstrated operation as intended without signs of damage.
    - Flow Testing: Flow rates within anticipated ranges under intended use conditions.Demonstrated flow rates within anticipated ranges.
    - Tensile Strength Testing: Tensile strength at joints within anticipated ranges.Demonstrated tensile strength at joints within anticipated ranges.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify exact sample sizes for each test set (e.g., number of devices for endotoxin testing, number of embryos for MEA). The tests are typically laboratory-based engineering and biological assessments compliant with recognized standards. Data provenance is not described in terms of geographical origin or retrospective/prospective study design as would be relevant for clinical data. These are intrinsic performance tests of the device itself.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Not applicable. The "ground truth" for these types of tests is generally defined by the specific test method, instrument calibration, and adherence to international standards (e.g., ISO, ASTM, USP). There are no human "experts" establishing ground truth in the context of diagnostic interpretation for these performance evaluations. Qualified laboratory personnel perform and interpret the results according to the established protocols.

    4. Adjudication Method for the Test Set

    Not applicable. These are quantitative and qualitative laboratory tests with defined pass/fail criteria from international standards. There is no ambiguous "case" interpretation requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not done. This device is an aspiration set, primarily a mechanical and sterile instrument, not an imaging or diagnostic device that involves human interpretation of results.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is not an algorithm or AI-enabled system. Its performance is entirely mechanical, material-based, and biological compatibility.

    7. The Type of Ground Truth Used

    The ground truth is based on established international and national standards and methodologies for medical device performance, sterility, material safety, and biocompatibility. Examples include:

    • Sterilization: ISO 11137-2:2013
    • Endotoxin: USP <85>
    • Needle Properties: ISO 9626:2016
    • Embryo Toxicity: Mouse Embryo Assay (MEA), a recognized biological test.
    • Packaging Integrity: ASTM F1886/F1886M-09, ASTM F1929-12, ASTM F88/F88M-09

    8. The Sample Size for the Training Set

    Not applicable. This device does not use machine learning, so there is no "training set."

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As there is no training set, this question is irrelevant to the device submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K162881

    Validate with FDA (Live)

    Date Cleared
    2017-07-11

    (270 days)

    Product Code
    Regulation Number
    884.6100
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Kitazato OPU Needles consisting of:

    • · Kitazato OPU Needle with Connection Tube (model number Type2)
    • · Kitazato OPU Reduced Needle (model number Type3)
    • · Kitazato OPU Two-Stage Reduced Needle (model number Type4)
      are intended to obtain oocytes from ovarian follicles.
    Device Description

    The Kitazato OPU Needles are single-use sterile devices that are used for ultrasound-guided transvaginal collection of oocytes from ovarian follicles.
    All of the Kitazato OPU Needle have a similar general design and device materials. All of the Kitazato OPU Needles are composed of a 304 stainless steel needle, acrylonitrile butadiene styrene (ABS) needle hub, PVC connection tube, and a silicone stopper that is used to connect to a collection tube and vacuum pump.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Kitazato OPU Needles, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    CriteriaAcceptance CriterionReported Device Performance
    Endotoxin< 20 EU/device (per ANSI/AAMI ST72-2011)< 20 EU/device
    Mouse Embryo Assay (MEA)≥ 80% expanded blastocysts at 96 hours (1-Cell MEA)≥ 80% expanded blastocysts at 96 hours (1-Cell MEA)
    Sterilization ValidationPer ISO 11135:2014 and ISO 10993-7:2008Passed validation
    Package Integrity (Dye penetration)Per ASTM F929-15Passed testing
    Package Integrity (Seal strength)Per ASTM F88/F88M-15Passed testing
    Transport TestingPer ASTM D4169-09Passed testing
    Shelf-life (MEA)≥ 80% expanded blastocysts at 96 hours (1-Cell MEA)Met at 3 years
    Shelf-life (Endotoxin)< 20 EU/deviceMet at 3 years
    Shelf-life (Tensile testing - needle to hub)Internal requirements metMet at 3 years
    Shelf-life (Tensile testing - tubing to hub)Internal requirements metMet at 3 years
    Shelf-life (Bending elasticity of needle)Return to normal position after bending 8° from straightMet at 3 years
    Shelf-life (Folding strength of needle)No fracture when folded at 5 mm curvature radius to 90°Met at 3 years
    Shelf-life (Device appearance)No burrs, scratches, damageMet at 3 years
    Shelf-life (Dimensional assessments)Internal requirements metMet at 3 years
    Aspiration pressure testingDevices operated as intended (no deformation, blockages, damage)Operated as intended
    Biocompatibility (Cytotoxicity)Acceptable results per ISO 10993-5:2009No signs of cytotoxicity
    Biocompatibility (Intracutaneous reactivity)Acceptable results per ISO 10993-10:2010No signs of irritation reactions
    Biocompatibility (Sensitization)Acceptable results per ISO 10993-10:2010No signs of sensitization reactions

    2. Sample Size Used for the Test Set and Data Provenance:

    The document describes non-clinical performance tests. It does not specify a "test set" in the context of clinical data or human subjects. The testing involved various batches of the devices for each test. For example:

    • Mouse Embryo Assay (MEA): "1-Cell mouse embryos were incubated in extracts of the subject device..." - The specific number of embryos or repetitions is not provided.
    • Shelf-life studies: Performed on "real-time aged devices." The number of devices tested for each shelf-life parameter (MEA, endotoxin, tensile, bending, folding, appearance, dimensional) is not specified.

    The data provenance is not explicitly stated in terms of country of origin or whether it's retrospective/prospective, as these are non-clinical lab tests performed on the manufactured devices.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    Not applicable. This report details non-clinical laboratory testing (e.g., chemical, mechanical, biological assays), not clinical studies requiring expert interpretation of medical images or data. Therefore, there is no "ground truth" established by experts in the typical sense for clinical AI performance studies.

    4. Adjudication Method for the Test Set:

    Not applicable, as this is non-clinical laboratory testing.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No. The document describes non-clinical performance testing of the device itself (e.g., material properties, sterility, embryotoxicity) and does not involve human readers or AI assistance.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

    Not applicable. This is a medical device (needles), not an AI algorithm. The performance is assessed based on its physical and biological characteristics, not computational output.

    7. Type of Ground Truth Used:

    The "ground truth" for these non-clinical tests is based on:

    • Quantitative measurements: e.g., endotoxin levels (< 20 EU/device), percentage of expanded blastocysts (≥ 80%), specific dimensions.
    • Qualitative observations: e.g., absence of deformation/blockages, no fracture, no burrs/scratches, no signs of cytotoxicity/sensitization/irritation.
    • Adherence to established standards: e.g., ANSI/AAMI ST72-2011, ISO 11135:2014, ISO 10993-7:2008, ASTM F929-15, ASTM F88/F88M-15, ASTM D4169-09, ISO 10993-1-2009, ISO 10993-5:2009, ISO 10993-10:2010.

    8. Sample Size for the Training Set:

    Not applicable. This document pertains to the regulatory submission for a physical medical device, not an AI model. There is no concept of a "training set" in this context.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable. As there is no training set, the establishment of ground truth for it is irrelevant here.

    Ask a Question

    Ask a specific question about this device

    K Number
    K161970

    Validate with FDA (Live)

    Date Cleared
    2017-06-14

    (331 days)

    Product Code
    Regulation Number
    884.6100
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Intended for flushing and/or aspiration of oocytes from ovarian follicles.

    Device Description

    Follicle Aspiration Sets included in this submission are intended for ultrasound-guided transvaginal aspiration and flushing of oocytes from ovarian follicles. Four types of Follicle Aspiration Sets are included in this submission, and are described in more detail below.

    The Follicle Aspiration Sets are radiation sterilized and are for single use only. The materials used in these devices include: stainless steel, TPE-O tubing, silicone, medical grade adhesive, polypropylene, and methyl methacrylate acrylonitrile butadiene styrene.

    Follicle Aspiration Set, Single Lumen: This device consists of a single-lumen stainless steel needle, aspiration tubing, and a silicone rubber cork for attachment to a collection tube. One version (17132) also includes vacuum tubing attached to the silicone rubber cork for attachment to the vacuum pump used for aspiration. This device also includes an adapter for flushing of ovarian follicles and removal of blockages. The needles for the majority of the versions of this device type include echogenic markings at the tip of the needle to aid in visualization during aspiration procedures.

    Follicle Aspiration Set, Double Lumen: This device consists of a double lumen stainless steel needle, tubing for flushing of ovarian follicles and removal of blockages, tubing for aspiration, and a silicone rubber cork for attachment to a collection tube. One version (17129) also includes vacuum tubing attached to the silicone rubber cork for attachment to the vacuum pump used for aspiration. The needles for these versions of the device include echogenic markings at the tip of the needle to aid in visualization during aspiration procedures.

    Follicle Aspiration Set, Single Lumen Luer with Tubing: This device consists of a single lumen stainless steel needle, and aspiration tubing with a luer connection at the end. The needle is provided with echogenic markings at the tip of the needle to aid in visualization during aspiration procedures.

    Follicle Aspiration Set, Single Lumen, Luer: This device consists of a single lumen stainless steel needle with a luer connection at the base of the needle. The needle is provided with echogenic markings at the tip of the needle to aid in visualization during aspiration procedures.

    AI/ML Overview

    The provided text describes the Follicle Aspiration Set, outlining its design, intended use, and comparison to a predicate device. It also includes information about performance testing to establish substantial equivalence.

    Here's a breakdown of the requested information:

    Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Mouse Embryo Assay (MEA): ≥ 80% expanded blastocysts within 96 hours (1-Cell MEA)Met: > 80% expanded blastocysts within 96 hours (1-Cell MEA)
    Endotoxins: < 1.2 EU/deviceMet: ≤ 1.2 EU/device
    Vacuum Testing: Operate as intended under -500 mmHg vacuum without signs of damageMet: Devices were shown to operate as intended without signs of damage.
    Flow Testing: Flow rates within anticipated ranges under intended use conditionsMet: Testing was conducted to assess that flow rates were within anticipated ranges under intended use conditions.
    Sterilization Validation per ISO 11137-2:2013Met: Confirmed by Sterilization Validation per ISO 11137-2:2013.
    Biocompatibility: Non-cytotoxicMet: Device extracts assessed were non-cytotoxic.
    Biocompatibility: Non-sensitizingMet: Device extracts assessed were non-sensitizing.
    Biocompatibility: Non-irritatingMet: Device extracts assessed were non-irritating.
    Mechanical Testing (Breakage, Bend, Stiffness per ISO 9626:2016)Met: Testing conducted per ISO 9626:2016. (Specific results not detailed but implied to be sufficient for a substantial equivalence determination).
    Tensile Strength TestingMet: Testing conducted. (Specific results not detailed but implied to be sufficient for a substantial equivalence determination).
    Shelf-life (3 years): 1-cell MEA ≥80% expanded blastocysts within 96 hoursMet: Conducted to ensure acceptance criteria are met over 3 years shelf-life. (Implied to meet the criterion).
    Shelf-life (3 years): Standard Test Method for Seal Strength of Flexible Barrier Materials per ASTM F88 / F88M - 09Met: Conducted to ensure acceptance criteria are met over 3 years shelf-life. (Implied to meet the criterion).
    Shelf-life (3 years): Dye Penetration per ASTM F1929 12Met: Conducted to ensure acceptance criteria are met over 3 years shelf-life. (Implied to meet the criterion).
    Shelf-life (3 years): Visual Inspection per ASTM F1886 / F1886M 09Met: Conducted to ensure acceptance criteria are met over 3 years shelf-life. (Implied to meet the criterion).
    Shelf-life (3 years): Tensile Strength TestingMet: Conducted to ensure acceptance criteria are met over 3 years shelf-life. (Implied to meet the criterion).

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not specify human test subjects or clinical data for the "test set" in the context of typical AI/software studies. The testing described is primarily non-clinical, involving laboratory and mechanical evaluations of the physical device. Therefore, information regarding human sample size, country of origin, or retrospective/prospective nature is not applicable here.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable. The device is a physical medical instrument (Follicle Aspiration Set), and the testing performed relates to its physical, chemical, and biological properties (e.g., sterilization, endotoxins, MEA, mechanical strength, biocompatibility). It does not involve the interpretation of data by human experts in the way an AI diagnostic tool would.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable. The testing methods described are objective laboratory or standardized mechanical tests, not subjective assessments requiring expert adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    A multi-reader multi-case (MRMC) comparative effectiveness study was not conducted. This type of study is relevant for AI-powered diagnostic or assistive technologies, where human performance is measured with and without AI. The Follicle Aspiration Set is a physical medical device, not an AI system.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    A standalone performance evaluation (as typically understood for an algorithm) was not conducted because the device is a physical instrument, not an algorithm. The "standalone" performance here refers to the device's intrinsic mechanical, chemical, and biological properties, which were evaluated through the performance testing listed.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the performance testing is based on established scientific and engineering standards and methods:

    • Mouse Embryo Assay (MEA): The development of mouse embryos to the expanded blastocyst stage is a biological endpoint widely accepted in ART for assessing the embryotoxicity of materials.
    • Endotoxin testing (USP<85>, LAL): Standardized biochemical tests for the presence and quantity of endotoxins.
    • Sterilization Validation (ISO 11137-2:2013): International standard for radiation sterilization.
    • Mechanical Testing (ISO 9626:2016, Tensile Strength): International and industry standards for needle properties and material strength.
    • Biocompatibility Testing (ISO 10993-5, 10, 1): International standards for evaluating the biological effects of medical devices.
    • Packaging Testing (ASTM F88/F88M, F1929, F1886/F1886M): American Society for Testing and Materials standards for package integrity.

    These are objective, quantitative, and qualitative measures against defined standards or biological responses, rather than subjective ground truths established by human experts or outcomes data in a clinical trial.

    8. The sample size for the training set

    This information is not applicable. The device is a manufactured medical instrument, not a machine learning model, so there is no "training set."

    9. How the ground truth for the training set was established

    This information is not applicable, as there is no "training set" for this physical device.

    Ask a Question

    Ask a specific question about this device