Search Filters

Search Results

Found 37 results

510(k) Data Aggregation

    K Number
    K242383

    Validate with FDA (Live)

    Device Name
    Apex Locator
    Date Cleared
    2025-03-07

    (207 days)

    Product Code
    Regulation Number
    N/A
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Apex Locator is used to detect the apex of root canal. This device must only be used in hospital environments, clinics or dental offices by qualified dental personnel.

    Device Description

    Apex Locator is an oral equipment used for root canal measurements. It features an LCD color display with functional buttons, enabling users to conveniently view parameters such as battery status, test wire connection status, and apex position etc. Users can easily adjust sound level, brightness, and reference settings using intuitive buttons, and perform functional checks of both the device and cable. The device is intended to be sterilized prior to use.

    AI/ML Overview

    The provided text is a 510(k) summary for a medical device called "Apex Locator". It details a comparison to a predicate device and states that non-clinical tests were performed. However, it explicitly states that no clinical test data was used to support the decision of substantial equivalence. This means that a study proving the device meets acceptance criteria, as typically understood in the context of clinical trials or performance evaluations with human subjects/data, was not conducted or reported in this submission. The "acceptance criteria" and "device performance" mentioned here refer to specifications confirmed through non-clinical testing and comparison to the predicate.

    Given this, I cannot provide information on:

    • The study that "proves the device meets the acceptance criteria" in a clinical sense.
    • Sample size used for the test set and data provenance (as no clinical test data was used).
    • Number of experts used to establish ground truth or their qualifications.
    • Adjudication method.
    • Multi-reader multi-case (MRMC) comparative effectiveness study.
    • Standalone (algorithm only) performance.
    • Type of ground truth (as no clinical data).
    • Sample size for the training set (as no clinical data).
    • How ground truth for the training set was established (as no clinical data).

    I can, however, provide the acceptance criteria based on the non-clinical tests and the device performance as reported in the comparison table and discussion of differences.


    Non-Clinical Acceptance Criteria and Reported Device Performance

    The provided document describes acceptance criteria and reported "performance" based on non-clinical tests and comparison to a predicate device. The primary method for demonstrating performance and safety was through compliance with recognized standards and verification that design differences do not affect safety and effectiveness.

    Key takeaway: This submission focuses on engineering, safety, and functional conformance rather than clinical performance data.

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria (Standard / Metric)Reported Device Performance (Proposed Device)
    Electrical Safety (Compliance)Compliance with IEC 60601-1
    EMC (Compliance)Compliance with IEC 60601-1-2
    Usability Engineering (Compliance)Compliance with IEC 62366-1
    Software (Compliance)Compliance with IEC 62304
    Biocompatibility (Compliance)Compliance with ISO 10993-1 for patient-contacting components (Lip hook: Stainless steel; File clip: stainless steel, plastic PBT). Also meets ISO 10993-11.
    Power Supply (Type)Lithium battery (DC3.7V)
    Charger Power Supply (Input)AC100V-240V, 50Hz ~ 60Hz
    Charger OutputDC 5V 1A, Class II (adaptor)
    ComponentsMeasuring wire, File clip, Lip hook, Touch Probe, Tester
    Measurement Accuracy±0.5mm
    SterilizationLip clip and file clip are user sterilized by steam sterilization.
    Battery Safety (Compliance)Li-ion battery complies with IEC 62133-2
    Function TestMeets requirements of pre-defined acceptance criteria and intended uses. (Specifics not detailed beyond general statement).

    Explanation of "Study that proves the device meets the acceptance criteria":

    The "study" referenced in the document is a series of non-clinical tests and a detailed comparison to a legally marketed predicate device. The conclusion of substantial equivalence is based on demonstrating that the proposed device:

    1. Complies with relevant recognized safety and performance standards: These include IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), IEC 62366-1 (usability), IEC 62304 (software), ISO 10993-1 and ISO 10993-11 (biocompatibility), and IEC 62133-2 (battery safety).
    2. Performs the same intended function as the predicate device.
    3. Has fundamental technological characteristics that are substantially equivalent to the predicate device.
    4. Any differences in technological characteristics (e.g., dimensions, weight, battery capacity, means of input) do not raise new questions of safety or effectiveness. For each noted difference, the submitter provided a justification that it "would not affect its safety and effectiveness" or "could not effects the performance and safety."
    5. Achieves the same specified performance metric as the predicate device where applicable (e.g., Measurement Accuracy: ±0.5mm).

    The document explicitly states: "No clinical test data was used to support the decision of substantial equivalence." Therefore, there is no clinical study that "proves the device meets the acceptance criteria" in terms of patient outcomes or comparative clinical performance. The acceptance is based on engineering principles, compliance with standards, and a robust non-clinical comparison to a predicate device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K242765

    Validate with FDA (Live)

    Date Cleared
    2025-01-17

    (126 days)

    Product Code
    Regulation Number
    N/A
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Electronic Apex Locator is used to support the determination of the working length during the endodontic treatment. The use of this product is intended exclusively for duly qualified dental practitioners.

    Device Description

    The Electronic Apex Locator is a microprocessor-controlled device used for root canal length determination. The Electronic Apex Locator assesses the electrical resistance of the tooth and surrounding tissues while the file is inserted and moving along the root canal. By detecting the change of the impedance value of the tooth apex to the AC current response of two different frequencies, two signals are output that are proportional to the change of the AC current response of two different frequencies corresponding to the tooth apex, and the two response signals are compared, and the digital quantity of their ratio is output. The digital quantity represents the position of the electrode at the tooth apex.

    Advanced user interface implemented in Electronic Apex Locator is based on high resolution TFT (Thin Film Transistor) color graphic display. It provides presentation of endodontic file movement inside the canal. In addition, the interface displays the mute or volume level and power status.

    The product is functionally operated by three buttons: 1) On/off button of the main unit: Short press once to hear a beep is to start up the machine, short press once again to shut down, and long press no response. 2) Volume button: Under the startup state, press once to increase one volume. After reaching the maximum volume, press once to enter the mute mode. Press four times as a cycle, and there will be no responds when long press. 3) DEMO button: Under the startup state, long press to enter the DEMO process, and the screen will demonstrate the process of measuring the length.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study information for the Electronic Apex Locator (Alpha Apex I), based on the provided document:


    1. Table of Acceptance Criteria and Reported Device Performance

    Feature/MetricAcceptance CriteriaReported Device Performance
    Measurement Accuracy±0.5mm±0.5mm

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the numerical sample size used for the accuracy testing test set (e.g., number of teeth, number of measurements). It only mentions that "Performance testing to verify the apex position measurement accuracy according to internal technical requirements has been conducted."

    The data provenance is not specified. It is unclear if the data was collected retrospectively or prospectively, or its country of origin.


    3. Number of Experts Used to Establish Ground Truth and Their Qualifications

    The document does not provide information on the number of experts used or their qualifications to establish ground truth for the performance testing.


    4. Adjudication Method for the Test Set

    The document does not describe any specific adjudication method (e.g., 2+1, 3+1, none) used for the test set.


    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No multi-reader multi-case (MRMC) comparative effectiveness study was mentioned or performed. The device is an electronic apex locator, which is typically used directly by a single dental practitioner, not for multi-reader interpretation of images or data.


    6. Standalone Performance Study

    Yes, a standalone performance study was conducted. The "Accuracy Testing" section states, "Performance testing to verify the apex position measurement accuracy according to internal technical requirements has been conducted and demonstrated that the subject device meets the requirement of ±0.5mm." This indicates the algorithm's performance was evaluated independently.


    7. Type of Ground Truth Used

    The document does not explicitly state the specific type of ground truth used for the accuracy measurements (e.g., direct visual measurement through transparent tooth, X-ray confirmation, histological analysis). It only refers to "internal technical requirements" for verifying accuracy.


    8. Sample Size for the Training Set

    The document does not provide information on the sample size for any training set. This is likely because the Electronic Apex Locator is a hardware device based on electrical impedance principles, and while it contains a microprocessor and software, the core "measurement" algorithm is typically determined through engineering design and calibration rather than machine learning training on a large dataset in the way a diagnostic AI would be. The software validation (Alpha Apex I.1.0.0) focuses on general software development life cycle processes, not necessarily on a training and testing paradigm for an AI model.


    9. How Ground Truth for the Training Set Was Established

    As no training set is explicitly mentioned or implied for a machine learning model, there is no information on how its ground truth would have been established. The device relies on physical principles and calibrated measurements rather than a trained AI model in the conventional sense.

    Ask a Question

    Ask a specific question about this device

    K Number
    K231990

    Validate with FDA (Live)

    Device Name
    Apex Locator
    Manufacturer
    Date Cleared
    2024-02-13

    (223 days)

    Product Code
    Regulation Number
    N/A
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    APEX LOCATOR is a device intended for locating the apex of the root canal during endodontic treatments.

    Device Description

    APEX LOCATOR is a device used in dentistry to estimate the position of the root apex of teeth during root canal preparation. It is used connected with the Cefla dental units. Its principle of operation relies on measuring the electrical impedance of dental tissue, which varies according to tissue density and composition. The APEX LOCATOR is an accessory device for the Cefla dental units used to estimate the position of the root apex of teeth during root canal preparation. The kit is composed of the following elements: file clip (2 pieces), lip hook (4 piece), external wiring (1 piece).

    AI/ML Overview

    The provided text is a 510(k) summary for an "Apex Locator" device. It outlines the device, its intended use, and its substantial equivalence to a predicate device. However, it does not include information about AI/ML models, clinical studies for AI device performance, or acceptance criteria and proven performance related to such models.

    The document explicitly states: "Clinical performance testing was not conducted." This directly indicates that there is no data to address the points in your request that pertain to acceptance criteria and study proving performance for an AI/ML device.

    Therefore, I cannot provide the requested information because the document describes a traditional medical device (an Apex Locator based on electrical impedance measurement) and not an AI/ML-driven device. The review focuses on substantial equivalence based on technological characteristics and non-clinical performance (electrical safety, EMC, usability, biocompatibility, reprocessing validation), not on AI/ML model performance.

    Ask a Question

    Ask a specific question about this device

    K Number
    K232717

    Validate with FDA (Live)

    Date Cleared
    2023-12-22

    (108 days)

    Product Code
    Regulation Number
    N/A
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    support the dentist in the determination of the working the endodontic treatment. The use of this product is intended exclusively for duly qualified dental practitioners.

    Device Description

    Apex locators are equipment used for working length measurement during root canal treatment. The devices are powered by a built-in lithium battery, can be charged by a USB adaptor. The devices employ a LCD screen to display the status of file in the root canal and the relative distance between file tip and reference point.

    The devices are reusable medical devices initially supplied as non-sterile to the user and requiring the user to clean and disinfect the surface of device. And their accessory, Dental root canal file clip, Dental lip hook and probe are reusable, initially supplied as non-sterile to the user, and requiring the user to sterile for initial use, as well as to reprocess after each use.

    The devices include a main unit, a measurement wire A, a measurement wire B, a dental root canal file clip, a dental lip hook, a probe and an adaptor.

    Main unit: measure and display working length.

    Measurement wire A: connect file clip and lip hook with main unit.

    Measurement wire B: connect lip hook and main unit with EM-Motor root canal preparation device.(The EM-Motor root canal preparation device is not included with the subject device and is sold separately.)

    File clip: Connect to file.

    Lip hook: hook patient's lip.

    Touch Probe: promote to measure molar.

    Power Adapter: use for charging.

    Tester: use for checking whether the device is normal.

    AI/ML Overview

    The provided document is a 510(k) summary for the Apex Locator (Model: AL-Pex, AL-Pex+). It outlines why the device is considered substantially equivalent to a predicate device, but it does not contain the specific details of a study with acceptance criteria and reported device performance in the format requested.

    The document states:

    • "Clinical testing is not required." This means there isn't a clinical study with real patient data to generate the performance metrics typically seen in such tables (sensitivity, specificity, accuracy, etc.) against a human or definitive ground truth.
    • The performance data section primarily lists compliance with standards (electrical safety, EMC, biocompatibility) and "Accuracy verification: Internal method." This suggests the accuracy was verified through internal, non-clinical means.

    Therefore, I cannot extract the information required for:

    1. A table of acceptance criteria and the reported device performance: No such table is in the document for the device's main function (determining working length). The "Accuracy: ±0.5mm" mentioned in the comparison table is likely the design specification and presumed achieved, but no study results are provided to explicitly "prove" it meets this with a test set.
    2. Sample size used for the test set and data provenance: No information on a test set (e.g., number of roots/teeth tested) or its provenance (country, retrospective/prospective) is provided, as clinical testing was not required.
    3. Number of experts used to establish ground truth and qualifications: Not applicable as no clinical study leveraging human expert ground truth is described.
    4. Adjudication method: Not applicable.
    5. MRMC comparative effectiveness study: Not applicable, as no human reader study is described.
    6. Standalone performance: No detailed standalone performance metrics (like accuracy with a confidence interval) are presented from a formal study. The "±0.5mm" is a general specification.
    7. Type of ground truth used: Not explicitly stated for accuracy verification beyond "Internal method." For devices like apex locators, the ground truth is typically a known or precisely measured length, often in experimental setups or ex-vivo teeth, rather than clinical outcomes or pathology.
    8. Sample size for the training set: Not applicable, as this is a traditional medical device, not an AI/ML-driven software device that would have a "training set" in the context of machine learning.
    9. How ground truth for training set was established: Not applicable.

    In summary, the provided FDA document focuses on confirming substantial equivalence through device design, intended use, and compliance with general medical device safety and performance standards (electrical safety, EMC, biocompatibility, reprocessing validation), rather than presenting specific quantitative performance data from a clinical or extensive non-clinical study that would fit the requested criteria for acceptance and proof of performance. The "Accuracy: ±0.5mm" found in the comparison table is a functional specification, not a reported result from a study described in the document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K212178

    Validate with FDA (Live)

    Date Cleared
    2022-07-20

    (373 days)

    Product Code
    Regulation Number
    N/A
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Root Apex Locator is used to detect the apex of root canal. This device must only be used in hospital environments, clinics or dental offices by qualified dental personnel.

    Device Description

    The Root Apex Locator C-ROOT I is a oral equipment used for root canal measurement. The device includes a TFT colour display with touch panel displays parameters such as battery status, connection status of test wire and apex position, etc. Users can also set and modify the sound level, brightness level, DR'S CHOICE via a touch panel, and provide a functional check of the device and cable.

    AI/ML Overview

    The Root Apex Locator is a dental device used to detect the apex of the root canal. The acceptance criteria and the study proving the device meets these criteria are detailed below, primarily based on the accuracy testing.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Apex position measurement accuracy within ±0.5mmThe Root Apex Locator C-ROOT I meets the requirement of ±0.5mm accuracy for apex position measurement.
    Compliance with general safety and performance standards (e.g., electrical safety, biocompatibility, EMC)The device complies with all listed standards, including ISO 10993-5, 10993-10, 10993-11, IEC 60601-1, IEC 60601-1-2, IEC 80601-2-60, IEC 62133-2, and IEC 60601-1-6.
    Software verification and validationSoftware documentation for moderate level of concern per FDA guidance.
    Sterilization validationCleaning/Disinfection Validation and Sterilization of components as per FDA Guidance.

    2. Sample Size Used for the Test Set and Data Provenance

    The document mentions "Internal test method" for accuracy testing. It does not explicitly state the sample size used for the test set or the data provenance (e.g., country of origin, retrospective or prospective nature of the data).

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    The document does not specify the number of experts used or their qualifications for establishing ground truth for the test set. It only mentions "Internal test method" for accuracy.

    4. Adjudication Method for the Test Set

    The document does not describe any specific adjudication method (e.g., 2+1, 3+1) for the accuracy test set.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not done. The study focuses on the device's standalone accuracy and compliance with various non-clinical standards, and substantiates equivalence to a predicate device. There is no mention of human readers improving with AI assistance, as this is not an AI-powered device.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes, a standalone performance test was done for the device's accuracy. The "Accuracy Testing" section explicitly states that the "Internal test method demonstrated that the apex position measurement accuracy of the proposed Root Apex Locator C-ROOT I meets the requirement of ±0.5mm." This indicates a measurement of the device's intrinsic performance.

    7. The Type of Ground Truth Used

    The type of "ground truth" for the accuracy testing is implied to be a precise measurement of the actual apex position, against which the device's measurement is compared. The document does not specify if this was established by expert consensus, pathology, or outcomes data. However, for a root apex locator, the ground truth would typically be a verified measurement of the true working length (e.g., direct measurement on extracted teeth or high-resolution imaging).

    8. The Sample Size for the Training Set

    The device described is a medical instrument (Root Apex Locator), not an AI/ML algorithm that typically requires a 'training set'. Therefore, the concept of a training set and its sample size is not applicable in this context. The performance evaluation is based on non-clinical testing for accuracy, safety, and compliance with standards.

    9. How the Ground Truth for the Training Set was Established

    As this is not an AI/ML algorithm, there is no training set and thus no ground truth establishment for a training set. The device's functionality is based on bioelectrical principles for sensing the root apex, not on learned patterns from a training dataset.

    Ask a Question

    Ask a specific question about this device

    K Number
    K210789

    Validate with FDA (Live)

    Device Name
    EQ-PEX
    Date Cleared
    2021-10-20

    (218 days)

    Product Code
    Regulation Number
    N/A
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    EQ-PEX is an Electronic Apex Locator designed for use in measuring the working length of root canals.

    Device Description

    EQ-PEX is Electronic Apex Locator and accessories to be used to treat of patients. It consists of the main body and lip hook, probe cord and 2 kinds of file holder. The device is used to track the position of the file in the root canal based on the impedance of two different frequencies. The device is intended for measuring the length of the root canal treatments. The components of the device include the main body, probe cord, lip book, file holder, AC adapter, and power cord.

    The screen displays measurement information of the root canal length at the current location of the file and displays a number and graph. Stainless steel hand files are intended to be used with the subject device. Lip hook is intended to be placed on the opposite lip of the tooth to be worked on. The file is slowly inserted into the root canal. When the file approaches the apex reference value and alarm will beep and the text apex will appear on the measurement screen.

    AI/ML Overview

    The provided text describes the 510(k) summary for the EQ-PEX Electronic Apex Locator. Based on the document, here's a description of the acceptance criteria and the study that proves the device meets them:

    1. A table of acceptance criteria and the reported device performance:

    The primary performance criterion for the EQ-PEX, as stated and compared to the predicate device, is Accuracy of Apex.

    Acceptance CriteriaReported Device Performance (EQ-PEX)Predicate Device (I-ROOT 100) Performance
    Accuracy of Apex<±0.5mm<±0.5mm

    This table indicates that the EQ-PEX meets the same accuracy specification as its predicate device. This suggests that the acceptance criterion for accuracy was to demonstrate performance equivalent to or better than the predicate.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    The document mentions "Comparative performance testing of the accuracy of the apex locator and usability validation." However, it does not specify the sample size for this performance testing.

    Regarding data provenance:

    • Country of Origin: Not explicitly stated for the testing data. The manufacturer is Meta Systems Co., Ltd., located in Seongnam-si, Gyeonggi-do, Korea. It's plausible the testing was conducted in Korea or an affiliated lab.
    • Retrospective or Prospective: The document does not specify whether the data used for performance testing was retrospective or prospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    The document does not provide information on the number of experts or their qualifications used to establish ground truth for the performance test set.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    The document does not provide information on the adjudication method used for the test set.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    An Electronic Apex Locator like EQ-PEX is a measuring device designed to assist a clinician in a procedure, not an AI diagnostic tool that assists human readers in interpreting medical images. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance is not applicable to this device and was not mentioned as being performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    The device itself is a standalone measuring instrument. The performance testing of "Accuracy of Apex" at <±0.5mm is an assessment of the device's standalone performance in measuring root canal length. While a user operates it, the accuracy being measured is the accuracy of the device's measurement output. So, yes, a standalone performance test was done to assess its accuracy.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    For an Electronic Apex Locator, the ground truth for working length measurement is typically established by direct, precise measurements, often through radiographic methods (e.g., radiography with a file in place, or CBCT), or potentially by the physical measurement of extracted teeth. The document does not explicitly state the specific method used to establish the ground truth for the "Accuracy of Apex" performance testing, but it would almost certainly be an objective, highly precise measurement technique.

    8. The sample size for the training set:

    The document does not specify a training set sample size. Electronic Apex Locators, particularly those based on impedance measurements, typically rely on established physical principles and calibrated hardware rather than deep learning models that require large training datasets in the conventional sense. The "training" for such devices is primarily in their design, calibration, and manufacturing process.

    9. How the ground truth for the training set was established:

    As mentioned in point 8, the concept of a "training set" with established ground truth, as applied to AI/machine learning models, does not directly apply to this type of medical device which relies on electrical impedance measurement principles. The "ground truth" for its operation is inherent in the laws of physics governing electrical circuits and biological tissues, and its accuracy is validated through performance testing against known objective measurements.

    Ask a Question

    Ask a specific question about this device

    K Number
    K203836

    Validate with FDA (Live)

    Date Cleared
    2021-03-30

    (90 days)

    Product Code
    Regulation Number
    N/A
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BOMEDENT Apex locator and WISMY Apex locator support the determination of the working length during the endodontic treatment. The use of this product is intended exclusively for duly qualified dental practitioners.

    Device Description

    The proposed BOMEDENT Apex locator and WISMY Apex locator are microprocessor-controlled devices can support the dentist during endodontic treatments for the proximity of the file to the reference point for endodontic working length determination. To do so, the proposed BOMEDENT Apex locator and WISMY Apex locator assess the electrical resistance of the tooth and surrounding tissues while the file is inserted and moving along the root canal. The proposed BOMEDENT Apex locator and WISMY Apex locator products configuration consist of the following components: (1) Power Adapter (2) USB cable (3) Measuring cable A (4) File clip (5) Lip hook (6) Measuring cable C (Optional)

    AI/ML Overview

    Here's a summary of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    Device: BOMEDENT Apex locator, WISMY Apex locator


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific CriteriaReported Device Performance
    Performance TestingAccuracyEquivalent to primary predicate device (Propex IQ® Apex Locator, K191806)
    Safety & PerformanceElectrical SafetyComplies with IEC 60601-1:2005+A1:2012, IEC 80601-2-60:2012
    EMCElectromagnetic CompatibilityComplies with IEC 60601-1-2:2014
    Battery SafetyRechargeable Li-ion Battery SafetyComplies with IEC 62133:2012
    BiocompatibilityIrritation & Skin SensitizationComplies with ISO 10993-10:2010
    BiocompatibilityIn Vitro CytotoxicityComplies with ISO 10993-5:2009
    Software ValidationSoftware SafetyComplies with FDA Guidance "General Principles of Software Validation; Final Guidance for Industry and FDA Staff, Document issued on: January 11, 2002" and "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" (Moderate level of concern)
    Reprocessing ValidationCleaning, Low Level Disinfection, and SterilizationComplies with FDA Guidance and standards (AAMI TIR 30, AAMI TIR 12, ISO 17665-1, and ISO17665-2) for lip clip and file clip

    Study Proving Acceptance Criteria:

    The document states that multiple tests were conducted to demonstrate the device meets the acceptance criteria. These include:

    • Internal test method for Accuracy Testing: This test shown that the performance of the proposed BOMEDENT Apex locator and WISMY Apex locator are equivalent to that of primary predicate device Propex IQ® Apex Locator (K191806).
    • Compliance with International Standards:
      • IEC 60601-1:2005+A1:2012 (Medical electrical equipment – Part 1: General requirements for basic safety and essential performance)
      • IEC 80601-2-60:2012 (Medical electrical equipment - Part 2-60: Particular requirements for the basic safety and essential performance of dental equipment)
      • IEC 60601-1-2:2014 (Electromagnetic Compatibility)
      • IEC 62133:2012 (Secondary cells and batteries safety requirements)
      • ISO 10993-10:2010 (Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization)
      • ISO 10993-5:2009 (Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity)
    • Compliance with FDA Software Guidance: "General Principles of Software Validation; Final Guidance for Industry and FDA Staff, Document issued on: January 11, 2002" and "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices".
    • Cleaning, Low Level Disinfection, and Sterilization validation: Per FDA Guidance Document, AAMI TIR 30, AAMI TIR 12, ISO 17665-1, and ISO17665-2.

    Detailed Information about the Study:

    • The document mentions "Internal test method" for accuracy testing, but provides no specific details such as test parameters, methodology, or results beyond the claim of equivalence.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size: Not specified in the provided text for the accuracy testing.
    • Data Provenance: Not specified. It only states "Internal test method." No information on country of origin or whether it was retrospective/prospective.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:

    • Not applicable/Not specified. The accuracy testing seems to be a technical performance comparison against a predicate device, not involving expert-established ground truth in a clinical setting.

    4. Adjudication Method for the Test Set:

    • Not applicable/Not specified. This information is not relevant for the type of technical performance testing described.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • No, a MRMC comparative effectiveness study was not done. The document states, "Clinical testing is not required."
    • Effect Size: Not applicable.

    6. Standalone Performance (Algorithm Only without Human-in-the-Loop Performance):

    • Yes, a standalone performance test was done in the form of "Accuracy Testing" using an "Internal test method" which evaluated the device's ability to determine working length. The device itself (an apex locator) is a standalone measurement tool, indicating its performance without human intervention (beyond operating it).

    7. Type of Ground Truth Used:

    • Not explicitly stated as a "ground truth" in clinical terms. For the accuracy testing, the performance of the predicate device (Propex IQ® Apex Locator, K191806) serves as the benchmark against which the subject device's performance is compared for demonstrating equivalence. This implies a technical or simulated ground truth established by comparing measurement outputs, rather than a clinical outcome or pathology.

    8. Sample Size for the Training Set:

    • Not applicable/Not specified. The device is a measurement tool, not an AI/machine learning algorithm requiring a "training set" in the traditional sense.

    9. How the Ground Truth for the Training Set Was Established:

    • Not applicable. As there is no mention of a training set for an AI/ML algorithm, this information is not relevant.
    Ask a Question

    Ask a specific question about this device

    K Number
    K191806

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-10-03

    (90 days)

    Product Code
    Regulation Number
    N/A
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Propex IQ® Apex Locator supports the dentist in the determination of the working length during the endodontic treatment. The use of this product is intended exclusively for duly qualified dental practitioners.

    Device Description

    The proposed Propex IQ® Apex Locator is a microprocessor-controlled device that supports the dentist during endodontic treatments for the determination of the file to the reference point for endodontic working length determination. To do so, the proposed Propex IQ® Apex Locator assesses the electrical resistance of the tooth and surrounding tissues while the file is inserted and moving along the root canal. The proposed Propex IQ® Apex Locator product configuration consists of the following components:

    • The proposed Propex IO® Apex Locator
    • The proposed Propex IQ® Apex Locator Measurement cable
    • The Proposed X-Smart IO® Measurement Cable
    • The proposed Propex IQ® Apex Locator External Tester
    • The Propex IQ® Disposable Barrier Sleeve
    • The proposed Propex IQ® Lip clip
    • The proposed Propex IQ® File clip
    • The proposed Propex IQ® Clamp
    • Propex IQ® Charger and AC Adapter Plugs
    AI/ML Overview

    The document describes the Dentsply Sirona Propex IQ Apex Locator and its non-clinical performance data to demonstrate substantial equivalence to its predicate devices.

    Here's an analysis of the acceptance criteria and study data:

    1. Table of Acceptance Criteria and Reported Device Performance

    Test CategoryAcceptance Criteria (Implied)Reported Device Performance
    BiocompatibilityIdentical materials and fabrication methods to predicate device.The patient contacting part (lip clip) is composed of identical materials and fabricated using the same methods as the primary predicate device (K131907).
    Electrical SafetyConformance to IEC 60601-1:2005 + CORR. 1:2006 + CORR. 2:2007 + A1:2012.The proposed Propex IQ® Apex Locator conforms to the standard requirements.
    Electromagnetic Compatibility (EMC)Conformance to IEC 60601-1-2:2014.The proposed Propex IQ® Apex Locator conforms to the standard requirements.
    Software Verification & ValidationConformance to IEC 62304:2015 and FDA guidance "General Principles of Software Validation".Software Verification and Validation tests were conducted in accordance with the specified standards and guidance.
    Accuracy Testing (vs. RAYPEX® 6)Performance equivalent to primary predicate device RAYPEX® 6 (K131907).The performance of the proposed Propex IQ® Apex Locator is equivalent to that of the primary predicate device RAYPEX® 6 (K131907).
    Accuracy Testing (vs. Root ZX II)Performance equivalent to reference device Root ZX II Apex Locator (K071190).The performance of the proposed Propex IQ® Apex Locator is equivalent to that of the reference device: Root ZX II Apex Locator (K071190).
    Disinfectant ResistanceResistance to chemical disinfectants supporting substantial equivalence.Applicable device components were assessed for resistance to chemical disinfectants and the results support substantial equivalence.
    Information Display LatencyLatency supporting substantial equivalence.Device display latency in all modes was assessed and the results support substantial equivalence.
    Hardware Requirements Verification (Battery, Audio, Temp, Ingress, Mating Components)Performance meeting device hardware requirements.Verification testing of hardware requirements (battery performance, audio alarm, temperature, liquid ingress, mating components connection cycling) was included, and results support substantial equivalence.
    Wireless CoexistenceMaintained functional wireless performance in the presence of unintended signals.The proposed Propex IQ® Apex Locator maintained functional wireless performance in the presence of unintended signals.
    Usability StudyUsers able to use the device correctly and safely without use error.Usability Studies conducted showed that users were able to use the device correctly and safely without use error.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample sizes for the "Accuracy Testing" or "Disinfectant Resistance Testing." It refers to "Internal test method" for accuracy, suggesting it was conducted in a controlled lab or in-house environment rather than a clinical setting. Given the nature of the device (apex locator), the "test set" for accuracy would likely involve extracted teeth or simulated root canals.

    The data provenance is non-clinical performance data, primarily from internal testing and conformance to international standards. There is no mention of country of origin for the data; it is assumed to be performed by the manufacturer, Dentsply Sirona. The tests are retrospective in the sense that they are conducted on a developed device to prove its performance against established standards and predicates.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    The document does not specify the number or qualifications of experts used to establish ground truth for the non-clinical tests. For "Accuracy Testing," the "ground truth" would likely be the actual measured length of the root canal established by precise laboratory methods or a gold-standard manual measurement. This might involve trained technicians or dental professionals, but their specific roles and numbers are not detailed.

    4. Adjudication Method for the Test Set

    No adjudication method is mentioned as this was a non-clinical study focused on direct measurement and conformance to standards, not subjective interpretation.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC comparative effectiveness study was mentioned. The device is not an AI-assisted diagnostic tool; it is an apex locator, a measurement device for determining root canal length. There is no "human reader" component in the context of interpretation or AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The accuracy testing on the Propex IQ Apex Locator and its comparison with predicate devices served as a standalone performance assessment of the device's ability to measure root canal length. The device itself operates to provide a measurement, and its performance was evaluated independently.

    7. The Type of Ground Truth Used

    For the accuracy testing, the ground truth would be the actual physical measurements of root canal length, likely established by high-precision manual or digital methods that are considered the gold standard for such measurements in dental research (e.g., direct visualization after sectioning, or radiographic measurements validated post-extraction). The document states "Internal test method," implying a controlled experimental setup to determine true root canal length.

    8. The Sample Size for the Training Set

    The document does not mention a "training set" as this device is not described as being based on machine learning or AI that requires training data. It is a measurement device based on electrical resistance principles.

    9. How the Ground Truth for the Training Set was Established

    Not applicable, as there is no mention of a training set for an AI/ML algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K181087

    Validate with FDA (Live)

    Date Cleared
    2018-12-18

    (237 days)

    Product Code
    Regulation Number
    N/A
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Apex Locator, DPEX III is a microprocessor controlled device intended for locating the apex of root canal.

    Device Description

    Apex Locator, DPEX III is an apex locator intended for precise localization of root canal apex. Operation of Apex Locator, DPEX III is fully automatic, no manual calibrations or adjustments are required. The measuring signal is analyzed, and automatic adjustments are made if needed. The device may operate within different conditions in the root canal. Apex Locator, DPEX III may only be used with stainless steel or nickel titanium endodontic files. The scale indication on the apex locator screen does not represent a distinct length or distance in mm or other linear units. It merely indicates the file progression towards the apex.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the Apex Locator, DPEX III. It aims to demonstrate substantial equivalence to a predicate device, the RAYPEX 6.

    Here's an analysis of the acceptance criteria and study information based solely on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document states that the acceptance criteria for performance were met by demonstrating equivalence to the predicate device through performance comparison testing.

    Acceptance Criteria CategorySpecific Criteria (Implicitly based on stated equivalence)Reported Device Performance (Apex Locator, DPEX III)
    Measurement PerformanceClinically acceptable measurement of apex location, equivalent to predicate device (RAYPEX 6)."The result obtained with the Apex Locator, DPEX III is clinically acceptable since they are equivalent to the results of the predicate device."
    Electrical Safety (EMC)Compliance with IEC 60601-1:2005+CORR.1 (2006)+CORR.2(2007)+AM1(2012) and IEC 60601-1-2: 2014."The safety and EMC testing confirm that the subject device complies with the same FDA recognized safety and EMC standards as the primary predicate device. All of the predetermined acceptance criteria were met."
    BiocompatibilityCompliance with ISO10993-5 (Cytotoxicity), ISO10993-10 (Skin Irritation), ISO10993-10 (Oral Mucosa Irritation), ISO10993-10 (Skin Sensitization)."The test result shows that the subject device complies with the same FDA recognized biocompatibility standards as predict device. All of the pre-determined acceptance criteria were met."
    Sterilization ValidationCompliance with ISO 17665-1:2006, ISO 11737-2:2009, and ANSI AAMI ST79:2017 for accessories."The steam sterilization validation study supports the labeling of sterilization method for the accessories subject to be sterilized at the clinical site. All of the pre-determined acceptance criteria were met."

    Note: The document does not provide quantitative specific acceptance criteria (e.g., "accuracy must be within X mm") or numerical performance data for the device. Instead, it relies on demonstrating equivalence to a legally marketed predicate device.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document mentions "using extracted tooth" for the performance comparison test, but it does not specify the sample size (i.e., number of extracted teeth) used for this test set.
    • Data Provenance: The data provenance is retrospective, as it involves testing performed by the manufacturer to demonstrate equivalence. The country of origin of the data is not explicitly stated, but the manufacturer is "Guilin Woodpecker Medical Instrument Co., Ltd." in P.R. China.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    This information is not provided in the document. The performance comparison test relies on a comparison against the predicate device, and the method for establishing "ground truth" for the apex location is not described beyond "using extracted tooth." There is no mention of human experts defining ground truth for the device's accuracy.

    4. Adjudication Method for the Test Set

    No adjudication method is described. Given the type of device (apex locator) and the testing described (comparison using extracted teeth), a human adjudication process by multiple experts is not indicated as part of the performance testing.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. The document explicitly states: "No clinical study is included in this submission." This indicates that there was no human-in-the-loop study comparing human readers with and without AI assistance.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    The performance comparison test "using extracted tooth" functions as a standalone performance test for the device's ability to locate the apex. It evaluates the device's accuracy against the predicate without direct human interaction influencing the measurement outcome during the test itself. However, it's not an "algorithm" in the sense of a software AI model; it's a physical device measurement.

    7. The Type of Ground Truth Used

    The "ground truth" for the performance comparison test appears to be established by the measurement of the predicate device (RAYPEX 6). The aim was to show the new device's measurements were equivalent. For an apex locator, the true anatomical apex location in an extracted tooth would be the actual ground truth, but the document frames the comparison against the predicate as the basis for performance acceptance. It's essentially "predicate device performance" as the de facto standard for this submission.

    8. The Sample Size for the Training Set

    Not applicable/Not provided. This device is an electrical impedance-based apex locator, not an AI/ML device that requires a training set in the conventional sense. Its "knowledge" is embedded in its electrical design and algorithms, which are determined during engineering and design, not through a data-driven training process.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable/Not provided. As stated above, this is not an AI/ML device that uses a "training set" with established ground truth.

    Ask a Question

    Ask a specific question about this device

    K Number
    K171867

    Validate with FDA (Live)

    Device Name
    Dr’s Finder NEO
    Date Cleared
    2018-03-30

    (281 days)

    Product Code
    Regulation Number
    N/A
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Dr's Finder NEO is intended for detecting the apex of root canal.

    Device Description

    The device allows the relative position of a dental file and the apex to be determined electrically. Using a standard dental file inserted into the root canal as an electrode, the device emits very small electrical currents having frequencies of 400 Hz, and 10400 Hz. The current between the file and mouth is measured at each of these frequencies, and the readout of the relative proximity to the apex will appear on the meter.

    AI/ML Overview

    The provided text describes the Dr's Finder NEO, an apex locator device, and its substantial equivalence to predicate devices. Here's a breakdown of the requested information based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Derived from Primary Predicate Accuracy)Reported Device Performance (Dr's Finder NEO)
    Accuracy: ±0.2 mmAccuracy: ±0.2 mm

    2. Sample Size Used for the Test Set and Data Provenance

    The document mentions "comparison testing on Dr's Finder NEO and the predicate device (RCM-7, K090925) was performed to compare the measurement performance of locating apex root using extracted tooth and a stainless steel file."

    • Test Set Sample Size: Not explicitly stated. The phrase "extracted tooth" implies multiple teeth were used, but a specific number is not provided.
    • Data Provenance: The study used "extracted tooth," suggesting an in vitro setting. The country of origin for the data is not specified, but the manufacturer is Good Doctors Co., Ltd. from the Republic of Korea. The study appears to be prospective in nature as it was conducted for this 510(k) submission.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    • Number of Experts: Not specified.
    • Qualifications of Experts: Not specified. The ground truth method described below does not inherently require human expert interpretation as the measurement is electrical.

    4. Adjudication Method for the Test Set

    Not applicable. The "ground truth" for an apex locator's accuracy is typically established by direct measurement relative to the actual apical foramen in an ex vivo setting, rather than through expert adjudication of images or clinical findings.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and the Effect Size of Human Improvement

    • MRMC Study: No, an MRMC comparative effectiveness study was not done. The study conducted was a technical performance comparison between two devices, not an evaluation of human reader performance with or without AI assistance.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) was Done

    Yes, a standalone performance test was done. The document states, "The comparison testing on Dr's Finder NEO and the predicate device (RCM-7, K090925) was performed to compare the measurement performance of locating apex root using extracted tooth and a stainless steel file." This describes the device operating on its own to locate the apex.

    7. The Type of Ground Truth Used

    The ground truth explicitly stated for accuracy is ±0.2mm. For an apex locator, this ground truth is typically established by:

    • Direct Measurement/Physical Measurement: In an ex vivo setting (e.g., extracted teeth), the actual length to the anatomical apex is physically measured (e.g., with a microscope or calipers after sectioning) and compared to the device's reading. The document's mention of "extracted tooth" supports this type of ground truth.

    8. The Sample Size for the Training Set

    The document does not mention a training set. Apex locators like the Dr's Finder NEO are typically based on electrical impedance principles and calibrated during manufacturing, rather than "trained" in the machine learning sense with a data set.

    9. How the Ground Truth for the Training Set Was Established

    As no training set is discussed or implied by the technology, this question is not applicable based on the provided information.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 4