Search Results
Found 3 results
510(k) Data Aggregation
(239 days)
Ask a specific question about this device
(196 days)
Uriscreen is a non-quantitative rapid screen for the detection of Urinary Tract Infection. Uriscreen detects both bacteria and/or white blood cells, common indicators of Urinary Tract Infection.
Uriscreen is a non-quantitative rapid screen, which detects catalase. Urinary catalase may be associated with White Blood Cells, bacteria and other cells, some of which may be indicators of urinary tract infection. Only Adults should use the test.
Uriscreen - Uriscreen is a rapid screen for the detection of Urinary Tract Infection. Uriscreen detects Catalase activity. Most organisms that commonly cause Urinary Tract Infection are Catalase positive. Additionally, Catalase activity is an indicator of white blood cells in urine. White blood cells are a common indicator of Urinary Tract Infection.
Uriscreen tubes contain a reagent powder, which enables Catalase detection. Urine is added to the tube to the line indicated, a few drops of hydrogen peroxide solution are added. Production of foam indicates Catalase activity. Catalase activity is indicative of bacteria and/or white blood cells in the urine. Most bacteria that commonly cause urinary tract infection are Catalase positive. The presence of bacteria and/or white blood cells in the urine are clearly indicative of Urinary Tract Infection.
The manufacturer did not define acceptance criteria. Here's a summary of the performance data from the studies:
1. Acceptance Criteria and Reported Device Performance:
| Study/Comparison | Performance Metric | Reported Performance |
|---|---|---|
| Phase I: Uriscreen vs. Multistix Nitrite/LE | Sensitivity | 100.00% |
| (Lay consumer) | Specificity | 87.34% |
| Accuracy | 89.13% | |
| Phase I: Uriscreen vs. Uri-Test Nitrite in Urine | Sensitivity | 80.00% |
| (Lay consumer) | Specificity | 88.31% |
| Accuracy | 86.96% | |
| Phase I: Uriscreen vs. Reference Culture Method | Sensitivity | 100.00% |
| (Lay consumer) | Specificity | 85.19% |
| Accuracy | 86.96% | |
| Phase II: Consumer vs. Medical Technologist | Sensitivity | 100.00% |
| Specificity | 100.00% | |
| Accuracy | 100.00% | |
| Phase III: Uriscreen vs. Culture (>100,000 CFU/ml) | Sensitivity | 95.35% |
| (Clinical Labs) | Specificity | 72.92% |
| Accuracy | 79.86% | |
| Phase III: Uriscreen vs. Culture (10,000-100,000 CFU/ml) | Sensitivity | 82% |
| (Clinical Labs) | Specificity | 87% |
| Accuracy | 84% |
2. Sample Size and Data Provenance (Test Set):
- Phase I (Lay Consumer comparisons):
- Multistix Nitrite/LE vs. Uriscreen: Total of 92 samples (13 Uriscreen+/Multistix+, 10 Uriscreen+/Multistix-, 0 Uriscreen-/Multistix+, 69 Uriscreen-/Multistix-).
- Uri-Test Nitrite in Urine vs. Uriscreen: Total of 92 samples (12 Uriscreen+/Nitrite+, 9 Uriscreen+/Nitrite-, 3 Uriscreen-/Nitrite+, 68 Uriscreen-/Nitrite-).
- Culture vs. Uriscreen: Total of 92 samples (11 Uriscreen+/Culture+, 12 Uriscreen+/Culture-, 0 Uriscreen-/Culture+, 69 Uriscreen-/Culture-).
- Provenance: Not explicitly stated, but implies collected as part of a clinical trial by the sponsor, likely prospective, with consumers performing the test on their own urine.
- Phase II (Consumer vs. Medical Technologist):
- Sample Size: 65 blinded samples (water only and water + yeast at 10 CFU/ml).
- Provenance: Samples tested by participants at a social club and MDC Associates. Prospective, with "blinded samples."
- Phase III (Clinical Laboratories):
- Uriscreen vs. Culture (>100,000 CFU/ml): 139 samples (41 Uriscreen+/Culture+, 26 Uriscreen+/Culture-, 2 Uriscreen-/Culture+, 70 Uriscreen-/Culture-).
- Uriscreen vs. Culture (≥10,000 to >100,000 CFU/ml): 139 samples (58 Uriscreen+/Culture-Pos, 9 Uriscreen+/Culture-Neg, 13 Uriscreen-/Culture-Pos, 59 Uriscreen-/Culture-Neg).
- Provenance: Conducted in three CLIA Registered Clinical Laboratories, including patients presenting with symptoms of Urinary Tract Infection. Likely prospective.
3. Number and Qualifications of Experts for Ground Truth (Test Set):
- Phase I: Not explicitly stated. For "reference culture method," it implies standard laboratory procedures, but the number and qualifications of experts interpreting these are not detailed.
- Phase II: "A trained Medical Technologist" confirmed all consumer results. Number of technologists not specified.
- Phase III: "Culture" results were used as ground truth. This implies microbiology laboratory professionals performed and interpreted these cultures. The number and specific qualifications of these experts are not explicitly stated, but standard clinical laboratory practices would apply.
4. Adjudication Method (Test Set):
- The document does not describe an adjudication method for conflicting results between experts.
- In Phase II, a single "trained Medical Technologist" confirmed consumer results, implying a direct comparison rather than an adjudication process for conflicting expert opinions.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
- No MRMC comparative effectiveness study was done to assess how much human readers improve with AI vs. without AI assistance. The Uriscreen is a rapid diagnostic test, not an AI-assisted diagnostic tool for human readers. Phase II did compare consumer interpretation to a medical technologist, but this was to validate consumer use, not human reader improvement with AI.
6. Standalone (Algorithm Only) Performance:
- The Uriscreen itself is a standalone test that produces a result (foam for positive, no foam for negative). The performance data presented (e.g., Sensitivity, Specificity, Accuracy) are representative of the device's standalone performance compared to various benchmarks. There is no separate "algorithm only" component as it's a reagent-based test.
7. Type of Ground Truth Used:
- Phase I:
- Predicate devices (Bayer Multistix Leukocyte esterase/Nitrite and Uri-Test Nitrite in Urine).
- Reference culture method.
- Phase II: Blinded samples (water/yeast) with results confirmed by a "trained Medical Technologist."
- Phase III: Reference culture method, with a positive culture defined as >100,000 CFU/ml or ≥10,000 to >100,000 CFU/ml based on the specific analysis. This is a definitive laboratory standard for bacterial infection.
8. Sample Size for the Training Set:
- This device is a rapid diagnostic test using visual interpretation of a chemical reaction. It is not an algorithm or AI model that typically requires a "training set" in the computational sense. Therefore, there is no training set mentioned or applicable for this type of device.
9. How Ground Truth for the Training Set Was Established:
- As stated above, there is no "training set" for this type of device. The development of the Uriscreen likely involved R&D and optimization of the chemical formulation and reaction parameters based on scientific principles of catalase detection, rather than machine learning training on a dataset with established ground truth.
Ask a specific question about this device
(222 days)
The Bactis 160US is an automated, computer assisted, optically based in vitro diagnostic instrument intended for the rapid detection and enumeration of bacteria in urine specimens. Urine samples are stained, filtered through a membrane, and excited with light of a defined wavelength range so as to produce a distinct fluorescent signal. Using fluorescent microscopy and a CCD camera, the image of the sample is captured. Using digitized image processing, the microscopic images are analyzed and the concentration of bacteria present in the specimen is determined by applying a proprietary software algorithm.
The Bactis 160US is indicated for rapid urine screening for the purpose of determining the occurrence of bacteriuria. Evidence of bacteriuria is an important factor for clinicians to consider in making a diagnosis of urinary tract infection.
The Bactis 160US is based on direct screening of urine using optical image analysis. Thirty (30) ul of well-mixed urine is mixed with fluorescent stains that bind to bacteria and other particles present in the specimen. The stained samples are filtered through a membrane. The dispersed and stained organisms are excited with a filtered lamp to produce a distinct fluorescent signal. Using fluorescence microscopy and a CCD camera the image of the sample is captured. These microscopic images are then analyzed. Using digitized image processing techniques, the bacteria are localized and enumerated, and their concentration is determined. Bacterial concentration for each specimen can be then reported. Alternatively, the concentration can be compared to a predetermined threshold (according to each laboratory's policy) and each urine sample can be categorized according to the infection category of "positive" or "negative". This method provide results within 25 minutes.
The Bactis system is comprised of the following main components:
- Control and Data Management System (CMS) 1.
- Specimen Processor and Analyzer (SP&A) Unit 2.
The urine screening process is controlled and monitored by the operator via the CMS, The CMS is comprised of two PC computers. An optional printer can be used to provide diagnostic reports. In the SP&A Unit, the specimens are processed and analyzed.
Here's a breakdown of the acceptance criteria and study information for the Bactis 160US, based on the provided text:
Acceptance Criteria and Device Performance
| Acceptance Criteria | Reported Device Performance (Bactis 160US) |
|---|---|
| Sample size | 4561 |
| Threshold CFU/ml | $\ge 10^5$ |
| Sensitivity | 87.5% - 90.2% |
| Specificity | 75% - 86.7% |
| Negative Predictive Value | 96.2% - 97.5% |
Study Details
2. Sample size used for the test set and the data provenance:
- Test Set Sample Size: 4561 random urine specimens.
- Data Provenance: The study was conducted prospectively at two different clinical sites. The specific country of origin is not explicitly stated, but the company is based in Herzliya, Israel.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- The document implies that the "standard reference methods for detecting bacteriuria" and "standard reference culture procedures" were used as ground truth. It does not specify the number of experts, their qualifications, or if a dedicated panel of experts was involved in establishing the ground truth for the test set.
4. Adjudication method for the test set:
- The document states that the studies were conducted under "double blinded conditions." This implies that those performing the Bactis 160US tests and those interpreting the standard reference methods were unaware of each other's results. However, it does not explicitly mention an adjudication method (like 2+1, 3+1, etc.) in case of discrepancies.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done:
- No, a multi-reader multi-case (MRMC) comparative effectiveness study comparing human readers with and without AI assistance was not performed or described. The study compared the device's performance against standard reference cultures.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
- Yes, the performance data presented is for the Bactis 160US as an automated, computer-assisted system. It reports the algorithm's performance (sensitivity, specificity, NPV) in detecting bacteria in urine specimens against culture results. While an operator loads samples and monitors the process, the analysis itself is automated and driven by the algorithm.
7. The type of ground truth used:
- The ground truth used was standard reference culture methods for detecting bacteriuria. The document also mentions "standard reference procedures such as expanded culture" and assessing "growth inhibitory substances in urine."
8. The sample size for the training set:
- The document does not specify the sample size used for the training set. It only describes the clinical studies conducted for performance validation.
9. How the ground truth for the training set was established:
- This information is not provided in the document. The document focuses on the performance study and its ground truth, not on the development or training of the proprietary software algorithm.
Ask a specific question about this device
Page 1 of 1