Found 124 results
510(k) Data Aggregation
(261 days)
The Eyer 2 is accompanied by accessories: frontal module for the ocular surface (1pc), dock station (charging station) (1pc), eye cap (1pc), lens protector (1 pc), storage case (1 pc), cleaning cloth (1 pc), allen wrench (1 pc), quick start guide (1 pc), welcome card (1 pc), shipping box (1 pc), power supply (1 pc), slit-lamp adapter (1 pc), silica gel bags (2 pcs).
Eyer 2 is designed for use in a medical environment by healthcare professionals. Captured images and videos are used for documentation and consultation. The images and videos are securely stored in an internal smartphone application database.
For the retinal function, the Eyer 2 is designed for non-mydriatic fundus imaging. In non-mydriatic imaging, no mydriasis is needed because infrared light is used for targeting the fundus and white light is flashed when an image is taken. The pupil does not respond to the infrared light so examination is convenient for the patient. With small pupils, it is recommended to use mydriatic drops. Eyer 2 has fixation targets for the patient to fixate on during imaging. The middle fixation target provides a macula-center image. It is possible to fix the optical disc in the center by selecting the appropriate point.
For the ocular surface and surrounding areas function, Eyer 2 has an ocular surface module with white, blue, and infrared light sources for imaging the eye surface and surrounding areas; in this configuration, the device does not make contact with the patient.
The transfer of images to a PC is carried out via DICOM, DICOMWEB, FTPS, or local folder connections, with the client responsible for the connection and subsequent storage.
The Eyer 2 energy comes from the smartphone that has a rechargeable Li-Ion battery and is charged when the device is docked on the charge station, which is connected to the mains by a power supply cable.
N/A
Ask a specific question about this device
(158 days)
FOP NM-10 is an ophthalmic non-mydriatic digital camera using iPhone SE3 which captures images of the fundus of the human eye and surface as well as surrounding areas of the human eye.
The FOP NM-10 is designed for imaging of the fundus and surface as well as surrounding area of the human eye, in a compact, portable format, with an inherent simplicity built in its operation and use. The application software (Remidio FOP) is loaded on top of the iOS operating system to enable the user to view, save, archive and retrieve the captured images of fundus and surface area of the human eye. It is a compact, battery-operated device that enables high-quality fundus photography without requiring pupil dilation. The device can be used in Hospitals and Ophthalmic clinics. The FOP NM-10 incorporates a 40° field of view (FOV) optical system optimized for high-resolution retinal imaging. The device employs infrared and white LED illumination to achieve adequate retinal illumination without causing discomfort to the patient. The device is built on a modular smartphone-based design, ensuring portability and ease of operation. The Remidio FOP app enables wireless control of illumination intensity, focus adjustments, capturing the images in Auto and manual modes, saving, editing and archiving images.
Physical Specifications
- a) Field of View: 40° (for a minimum pupil size of 3mm)
- b) Material: ABS, Silicone Rubber
- c) Diopter Correction: -16D to +16D
- d) Working Distance: 33mm
Output and Performance Characteristics
- a) Illumination (Light Source): Infrared LED and Cool White LED
- b) Camera Resolution: 64 line pairs per millimeter (lp/mm) (at the center)
- c) ISO Range: ISO 200 and 400
- d) Focus Adjustment: Manual and Tap to Focus
- e) Internal Fixation: 8 fixation points
- f) Operating temperature: 10° to 40°C
- g) Relative Humidity: 10% to 85%
- h) Atmospheric Pressure: 750 to 1060 hpa
N/A
Ask a specific question about this device
(175 days)
Optina-4C™ is a non-contact, high resolution digital imaging device which is suitable for photographing, displaying, storing, and transferring images of the retina. The Optina-4C is indicated for the viewing of the posterior segment of the eye at multiple wavelengths (colors) under mydriatic conditions.
Images are intended to aid clinicians in the detection, diagnosis, and management of ocular health and disease.
The Optina-4C™ (MHRC-C1N) is a non-contact mydriatic retinal camera (also called a fundus camera) designed to provide eye care professionals with a series of images of the retina obtained at specific wavelengths (colors). The Optina-4C processes images using algorithms that take advantage of the spectral information to produce 2D en face images of the retina as an aid to clinicians in the diagnosis of age-related macular degeneration (AMD), retinal pigment epithelium (RPE) disruptions and lesions, retinopathies, and other retinal diseases.
Utilizing Hyperspectral Imaging (HSI) technology, the Optina-4C captures a series of 92 high-resolution images at precise 5 nm wavelength increments across the visible and near-infrared (NIR) spectrum on a ~31° field-of-view. Image acquisition takes under two seconds, with only 0.92 seconds of light exposure time. During this time, the camera scans through all wavelengths from 905 to 450 nm, spending 10 milliseconds per 5 nm step.
The Optina-4C processes the images to provide spectrally calibrated, normalized, and registered images in a structured Hyperspectral Cube format for scroll-through viewing. The spectral resolution supports visualization and differentiation of drusenoid deposit types, as well as high contrast viewing of the retinal chromophores hemoglobin, melanin, and carotenoids (combination of lutein and zeaxanthin). Images can be viewed, stored or transferred in their native .h5 format, or as DICOM or PNG formats. Optina-4C images offer sharp, vivid, high-contrast coronal views of the posterior segment of the eye for clinical assessment and evaluation by eye care professionals.
The HSI NormReg Cube is a processed data set of 91 hyperspectral images that have undergone intensity normalization and spatial registration. Normalization corrects for the non-uniform illumination profile of the imaging system and the spectral sensitivity of the camera, while registration compensates for intra-scan eye movement to ensure consistent alignment across frames. The result is a hyperspectral image cube with enhanced spatial and spectral fidelity, referred to as the HIS-NormReg Cube.
N/A
Ask a specific question about this device
(266 days)
3nethra neo HD FA:
3nethra neo HD FA is used as a wide-field retinal imaging digital camera for photo documentation of ocular diseases that manifest in infants. The device acquires only images and does not provide any pathological analysis or diagnosis for treatment.
3nethra neo HD:
3nethra neo HD is used as a wide-field retinal imaging digital camera for photo documentation of ocular diseases that manifest in infants. The device acquires only images and does not provide any pathological analysis or diagnosis for treatment.
The 3nethra neo-HD FA is a handheld wide-angle fundus imaging system for neonatal screening. It is designed to acquire, display, store and transmit images of the posterior and anterior surfaces of human eye. The images assist clinicians in the evaluation and documentation of visual health in retinopathy of prematurity (ROP) and other problems. It operates in contact with the cornea of the eye under test. 3nethra neo HD FA uses continuous white light for operation and hence need pupil dilation. It also equipped with blue light source and green filters for fundus fluorescein angiography (FFA).
The 3nethra neo-HD is a handheld wide-angle fundus imaging system for neonatal screening. It is designed to acquire, display, store and transmit images of the posterior and anterior surfaces of human eye. The images assist clinicians in the evaluation and documentation of visual health in retinopathy of prematurity (ROP) and other problems. It operates in contact with the cornea of the eye under test. 3nethra neo HD uses continuous white light for operation and hence need pupil dilation. The 3nethra neo HD is a lower end variant of the 3nethra neo HD FA without the Fluorescein Angiography feature.
The provided 510(k) clearance letter and summary describe the acceptance criteria and a study to prove the device's performance. Here's a breakdown of the information requested:
The device in question, 3nethra neo HD FA and 3nethra neo HD, are ophthalmic cameras intended for wide-field retinal imaging and photo documentation of ocular diseases in infants. They acquire images only and do not provide pathological analysis or diagnosis.
Acceptance Criteria and Reported Device Performance
The core acceptance criterion for this device, based on the provided document, is substantial equivalence to existing predicate devices, particularly in terms of image quality for the FA functionality. The study focuses on demonstrating that the fluorescent angiography (FA) images produced by the subject device are comparable in quality to those produced by a predicate device.
| Acceptance Criterion | Reported Device Performance |
|---|---|
| Safety and Effectiveness | Demonstrated through compliance with electrical safety (IEC 60601-1, IEC 60601-1-2) and light safety standards (ISO 15004-2, ANSI Z80.36) and comparable image quality to a cleared reference device for its intended use. |
| Image Quality (for FA images, specifically for 3nethra neo HD FA) | All study images produced by the 3nethra neo HD FA were found to be of comparable quality to those produced by the RetCam 3 reference device (K182263). |
Study Details
The study described is a clinical observational study focused on image quality comparison.
-
Sample size used for the test set and the data provenance:
- The document states "of all study images," implying that all images collected within the clinical observational study were used for the comparison. However, the exact numerical sample size (number of images or patients) for the test set is not explicitly stated in the provided text.
- Data Provenance: The document does not explicitly state the country of origin of the data or whether the study was retrospective or prospective. It is described as a "clinical observational study," which typically implies prospective data collection, but this is not confirmed.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- The document does not explicitly state the number of experts used or their specific qualifications (e.g., number of years of experience, specific certifications). It simply refers to the primary objective being to demonstrate comparability to a cleared reference device. The process of expert evaluation for "comparable quality" is not detailed.
-
Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- The document does not describe any specific adjudication method for evaluating image quality. It states that "all study images were found to be of comparable quality," but the process by which this finding was made (e.g., consensus, majority vote) is not provided.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and if so, what was the effect size of how much human readers improve with AI vs. without AI assistance:
- No, a multi-reader multi-case (MRMC) comparative effectiveness study was not performed or described in the provided text.
- The device is described as an "Ophthalmic Camera" that "acquires only images and does not provide any pathological analysis or diagnosis for treatment." Therefore, there is no AI assistance component for human readers to be compared against. The study focused on assessing the image quality of the camera itself, not the impact of AI on human interpretation.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- This question is not applicable as the device is an imaging camera that acquires images for human interpretation, not an algorithm providing a standalone diagnostic or analytical output. The study assessed the quality of the images produced by the device.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- The ground truth for the "comparable image quality" assessment appears to be based on a clinical observational study comparing the subject device's images to those from a predicate device (RetCam 3). The implicit ground truth is the accepted, established quality of images from the predicate device.
- The document does not mention pathology or outcomes data as a direct ground truth for image quality assessment in this context. The study aimed to show that the images themselves were of sufficient quality for their intended purpose (photo documentation) by comparing them to a cleared device.
-
The sample size for the training set:
- The document does not mention a training set. This is because the device is an imaging camera, not a machine learning or AI algorithm that requires a training set.
-
How the ground truth for the training set was established:
- This question is not applicable as there is no mention of a training set for this device.
Ask a specific question about this device
(55 days)
The Resolve Fundus Camera is an automatic eye-fundus camera intended for taking digital images of a human retina with or without the use of a mydriatic agent. It is intended for use as an aid to clinicians in the evaluation and diagnosis of ocular health.
The Resolve Fundus Camera is a fundus camera designed to perform fundus observation, automatic pupil tracking and focusing, automatic image capture, and image preservation. The Resolve Fundus Camera is used for non-mydriatic observation and capturing of retinal images. The fundus camera employes three internal imaging systems to operate: illumination system, imaging system and observation system. Auto-alignment and auto-focus algorithms are used to automatically find and capture desired images of the fundus.
The fundus camera features an internal, movable three-dimensional platform that allows for switching and precise positioning of the left and right eyes. The imaging process is done one eye at a time (left eye and then the right eye). The internal, moveable three-dimensional platform contains two types of cameras: a set of infrared cameras (one on each side) as well as one main camera.
The observation system is used to detect and track the patient's pupil. Next, the illumination system determines the precise location of the ocular fundus after the pupil is located by the observation system. The system precisely finds the relationship between the position of the ocular fundus at different diopters, and the lens of the main camera.
Once the location of the fundus is found, a charge-coupled device (CCD) automatically captures a still image of the fundus through the main camera. The autofocus system utilizes a beam splitter to split a beam of light into two fine beams, which then image the ocular fundus onto the CCD. When the focus is at its sharpest position, the split beams align horizontally. At this point, the LED white light illumination system, installed in the imaging module on the three-dimensional movable platform, emits uniform white light of appropriate intensity to illuminate the ocular fundus. The imaging system captures the fundus information onto the CCD, and the received signals are displayed in real-time on a liquid crystal display (LCD) through the control system. Doctors can visually assess the patient's ocular fundus on the LCD screen.
The provided FDA 510(k) clearance letter for the Optain Health Resolve Fundus Camera focuses primarily on establishing substantial equivalence to a predicate device (Next Sight Srl Nexy). For this type of device (an ophthalmic camera), the clearance process in this document does not necessitate a detailed clinical study for performance evaluation that would typically involve acceptance criteria for diagnostic accuracy (e.g., sensitivity, specificity) of an AI algorithm. Instead, the performance data presented is focused on demonstrating that the device meets safety and basic functional standards, similar to the predicate device.
Therefore, many of the requested points regarding AI algorithm performance (like specific acceptance criteria for diagnostic accuracy, sample sizes for test sets, expert adjudication methods, MRMC studies, standalone performance, and ground truth establishment for training/test sets) are not explicitly described or required for this particular regulatory submission type, given the device's classification and stated indications for use. The device's indication is for "taking digital images" and "aid to clinicians in the evaluation and diagnosis of ocular health," implying that the device primarily performs an imaging function, and any "diagnosis" is still ultimately made by the human clinician using the image.
Below is an attempt to answer the questions based only on the provided text. Where information is not available, it will be stated as such.
Device: Optain Health Resolve Fundus Camera
Indications for Use: The Resolve Fundus Camera is an automatic eye-fundus camera intended for taking digital images of a human retina with or without the use of a mydriatic agent. It is intended for use as an aid to clinicians in the evaluation and diagnosis of ocular health.
Acceptance Criteria and Reported Device Performance
The document describes the device's performance in terms of compliance with recognized consensus standards rather than diagnostic performance metrics (e.g., sensitivity, specificity, accuracy) for an AI algorithm. The acceptance criteria are implicit in meeting the requirements of these standards.
1. Table of Acceptance Criteria and Reported Device Performance:
| Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|
| Electrical Safety (IEC 60601-1) | Complies with IEC 60601-1 |
| Electromagnetic Compatibility (IEC 60601-1-2) | Complies with IEC 60601-1-2 |
| Device Usability (IEC 60601-1-6) | Complies with IEC 60601-1-6 |
| Ocular Light Hazard Protection (ANSI Z80.36) | Complies with ANSI Z80.36 |
| Biocompatibility (ISO 10993-1, -5, -10, -23) | Complies with ISO 10993-1, -5, -10, -23 (for surface device, intact skin, limited duration) |
| Fundus Camera Performance (ISO 10940) | Complies with ISO 10940 |
| General Ophthalmic Instrument Requirements (ISO 15004-1) | Complies with ISO 15004-1 |
| DICOM Compliance (NEMA PS 3.1-3.20) | Complies with NEMA PS 3.1-3.20 (DICOM Set) |
| Battery Safety (IEC 62133-2) | Complies with IEC 62133-2 |
| Software Verification & Validation | "Software verification and validation activities were performed to ensure the device performed as intended and software documentation appropriate for the Basic documentation set." (No specific metrics provided in this summary) |
2. Sample size used for the test set and the data provenance:
- The document states "Non-clinical testing was performed," and "Performance testing" but does not specify a "test set" in the context of a dataset for evaluating an AI algorithm's diagnostic performance.
- The tests are primarily related to general device safety, function, and image quality standards compliance, not the diagnostic accuracy of an AI.
- Data Provenance: Not specified, but likely laboratory or engineering test data, not patient data from a specific country or retrospective/prospective study.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable/Not specified in the provided document, as no specific diagnostic ground truth for an AI algorithm's performance is described. The device's function is to aid clinicians, not to output a diagnosis via AI.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Not applicable/Not specified, for the same reasons as above.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No MRMC study is described in this 510(k) summary. The device's clearance is based on substantial equivalence to a predicate ophthalmic camera, not on a claim of AI-assisted diagnostic improvement.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- No standalone AI performance study is described. The device is an "automatic eye-fundus camera" which "aids clinicians." There is no mention of an AI algorithm producing a diagnostic output independently of a human.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- Not applicable/Not specified, as the performance evaluation is focused on device functionality and safety standards, not diagnostic accuracy against a clinical ground truth.
8. The sample size for the training set:
- Not applicable/Not specified. The document does not describe a training set for an AI algorithm. If there are "auto-alignment and auto-focus algorithms," these might involve machine learning, but the document does not detail their training data or performance evaluation beyond general "software verification and validation."
9. How the ground truth for the training set was established:
- Not applicable/Not specified, for the same reasons as above.
Ask a specific question about this device
(20 days)
Sentinel Camera is a non-mydriatic medical digital camera that is intended to capture digital images of the human eye.
The Sentinel Camera is accompanied by the following accessories: a charging base, eye cup, and power cable. The Sentinel Camera is designed for use in a medical environment. Captured images are used for documentation and consultation. Sentinel Camera has internal memory where captured images are stored. The Sentinel Camera is designed for non-mydriatic retinal imaging. In non-mydriatic imaging no mydriasis is needed because infrared light is used for targeting the retina and white light is flashed when an image is taken. The pupil does not respond to the infrared light, so examination is convenient for the patient. With small pupils, it is recommended to use mydriatic drops. The Sentinel Camera has three internal fixation targets for the patient to fixate on during imaging. The middle fixation target provides a macula-centered image. The left and right fixation targets provide disc-centered images. The transfer of images to the AI Optics Server is carried out via Wi-Fi communication. The Sentinel Camera has a rechargeable Li-ion battery that is charged when the camera is placed on charging base, which is connected to the mains by the power cable.
The provided text describes a 510(k) summary for the "Sentinel Camera," an ophthalmic camera. However, it does not contain information about acceptance criteria or a study that proves the device meets those criteria, particularly in the context of an AI-powered diagnostic or assistive tool. The document focuses on comparing the Sentinel Camera to a predicate device (Optomed Aurora Camera) based on their technical characteristics and safety/performance standards.
The 510(k) summary explicitly states: "Performance data for the Sentinel Camera, including extensive testing against recognized standards, supports the substantial equivalence of the Sentinel Camera to the predicate device. Both devices underwent similar testing to external standards." It concludes that the device "meets the necessary safety and performance criteria, with no new questions of safety or effectiveness."
Crucially, there is no mention of any AI component within the Sentinel Camera's functionality in this document, nor any studies that would involve AI performance metrics (like sensitivity, specificity, or human-AI collaboration studies). The device is described as "a non-mydriatic medical digital camera that is intended to capture digital images of the human eye."
Therefore, based on the provided text, I cannot answer the questions regarding acceptance criteria and studies proving the device meets them in the context of AI performance. The document does not provide such information.
If this device were to incorporate AI functionality, the information requested (acceptance criteria, study design, sample sizes, ground truth establishment, expert qualifications, etc.) would be crucial for its regulatory submission. However, this particular document does not detail any such AI-related performance studies.
Ask a specific question about this device
(109 days)
Verily Numetric Retinal Camera is a non-contact, high resolution digital imaging device which is suitable for photographing, displaying and storing images of the retina segment of the eye to be evaluated under non-mydriatic conditions.
Verily Numetric Retinal Camera is indicated for in-vivo viewing of the posterior segment of the eye. The images are intended for use as an aid to clinicians in the evaluation, diagnosis, and documentation of ocular health.
The Verily Numetric Retinal Camera (VNRC), the subject device, is a tabletop fundus camera that provides non-mydriatic, color, posterior segment images of the eye as an aid to clinicians in the evaluation and diagnosis of eye disease. It is a non-contact, high-resolution digital imaging device that is suitable for photographing, displaying, and storing images of the retina to be evaluated under non-mydriatic conditions.
The Verily Numetric Retinal Camera takes images of the fundus in the following manner:
Focusing: The patient interactively focuses an image shown on a microdisplay to the best qualitative focus by turning a knob on an external peripheral device which modulates the camera optics.
Alignment: The patient is shown a fixation target on the microdisplay and through an interactive session with position feedback, the patient aligns their pupil to the camera. The pupil location is determined using eye-tracking.
Image Capture: White light from LEDs illuminates the fundus. The light reflection from the fundus is captured by a color CMOS sensor.
The Verily Numetric Retinal Camera has the following acceptance criteria and reported device performance:
1. Table of Acceptance Criteria and Reported Device Performance
| Performance Item | Acceptance Criteria | Reported Device Performance |
|---|---|---|
| Resolving power (center) | ≥ 60 line pairs/mm at the center of the field | Pass |
| Resolving power (mid field) | ≥ 40 line pairs/mm at the mid field (r/2) | Pass |
| Resolving power (periphery) | ≥ 25 line pairs/mm at the periphery of the field (r) | Pass |
| Field of view | 45 degrees | Pass |
| Pixel pitch tolerance | < 7% | Pass |
| Alignment illumination | The alignment illumination is done via Near Infrared (NIR) LEDs during eye tracking. The intensity of the NIR LEDs is controlled by SW. | Pass |
| Flash illumination | The flash illumination is done via white LEDs during the image capture. The intensities and durations of the white LEDs are controlled by SW. | Pass |
| Range of focus | -20D to +20D (no compensating lens) | Pass |
| Minimum pupil size | 3.0mm | Pass |
| Alignment | Interactive alignment using 3D tracking. | Pass |
| Image quality | The gradability of images captured by VNRC is comparable to the predicated device on the same population. | Pass |
2. Sample Size and Data Provenance for Test Set
The document mentions that non-clinical performance testing for the Verily Numetric Retinal Camera includes analysis using raw data collected via a retrospective clinical study. It also states that the "gradability of images captured by VNRC is comparable to the predicated device on the same population."
However, the specific sample size used for the test set is not explicitly stated in the provided text. The country of origin of the data is also not specified.
3. Number and Qualifications of Experts for Ground Truth (Test Set)
The document does not provide information on the number of experts used to establish ground truth for the test set or their specific qualifications (e.g., "radiologist with 10 years of experience"). It only mentions that the images are "intended for use as an aid to clinicians in the evaluation, diagnosis, and documentation of ocular health" and that "the gradability of images captured by VNRC is comparable to the predicated device on the same population." This implies clinical evaluation but lacks details on the expert panel.
4. Adjudication Method for the Test Set
The document does not specify an adjudication method (e.g., 2+1, 3+1, none) for the test set.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
A multi-reader multi-case (MRMC) comparative effectiveness study was not explicitly described in the provided text in the typical sense of measuring human reader improvement with AI assistance. The document compares the performance of the device itself (Verily Numetric Retinal Camera) to a predicate device (NFC-700) regarding image quality and technical specifications. It notes that "the gradability of images captured by VNRC is comparable to the predicated device on the same population," implying a comparison of image utility for human readers, but no effect size on human reader improvement with AI assistance is provided as the device is a camera, not an AI diagnostic tool.
6. Standalone (Algorithm Only) Performance Study
The Verily Numetric Retinal Camera is a hardware device (ophthalmic camera) that captures images, not a standalone AI algorithm. Therefore, a standalone (algorithm only) performance study as typically understood for AI-based diagnostic software was not conducted or described. The focus is on the camera's ability to produce high-quality, gradable images.
7. Type of Ground Truth Used (for Test Set)
The document primarily refers to evaluating "image quality" and "gradability." While not explicitly stated as "expert consensus ground truth," the comparison of "gradability of images captured by VNRC... comparable to the predicated device on the same population" strongly implies that human expert evaluation of image quality was used as a form of ground truth for assessing the usability of the images for clinical purposes. There is no mention of pathology or outcomes data being used as ground truth for effectiveness.
8. Sample Size for the Training Set
The document does not provide a sample size for a training set. This is expected as the device is a camera, not an AI algorithm that typically requires a training set.
9. How Ground Truth for Training Set Was Established
As the device is a camera and not an AI algorithm, the concept of a "training set" and establishing "ground truth for the training set" does not directly apply in the context of this submission. The camera's performance is evaluated against technical specifications and a comparison to a predicate device.
Ask a specific question about this device
(28 days)
Ask a specific question about this device
(221 days)
Optina-4C™ is intended to capture images of the retina at multiple wavelengths (colors) under mydriatic conditions.
Optina-4C™ (model: MHRC-C1N) is a mydriatic fundus camera (also called retinal camera) that presents eye care practitioners (optometrists and ophthalmologists) with a series of the retina obtained sequentially at specific wavelengths (colors) in the spectral range 905 nm to 450 nm in steps of 5 nm.
Pictures of the retina are obtained on a field-of-view of 31.5° without contact with the eye. The patient is positioned in front of the device with the chin on the chinrest and forehead on the forehead rest and a positioning system is used to align the camera relative to the patient's eye. An external fixation target is available to guide the patient's eye. A focus wheel allows for the accommodation for eye refractive error in the range of -15 to +15 diopters. The images are displayed on a monitor and can be saved on the computer for future consultation.
The illumination light of Optina-4C™ is provided by a Tunable Laser Source (TLS). The TLS selects a narrow band of light from a broadband white illumination source. Only a single monochromatic band can output the TLS at a time. The alignment of the retinal camera relative to the patient's eye phase is performed with the illumination light set at a wavelength of 700 nm. The image acquisition phase consists of the sequential acquisition of a series of 92 monochromatic images at wavelengths from 905 nm in steps of 5 nm. Each frame is captured with an exposure time of 10 ms, resulting in a total acquisition time of 920 ms.
The 92 images can be visualized one by one in Optina-4C™ acquisition software. The retinal images captured by Optina-4C™ are monochromatic images having a spectral bandwidth of "10 nm centered on the wavelength indicated in the upper left corner of the visualization pane, with a spectral accuracy of 7.5 nm.
The provided text is a 510(k) summary for the Optina-4C (MHRC-C1N) device, which is a mydriatic fundus camera. The document primarily focuses on demonstrating substantial equivalence to a predicate device (MHRC-C1) and outlines various performance tests conducted to support this claim.
However, the document does not contain specific acceptance criteria for the device's performance regarding its intended use of capturing retinal images, nor does it present a detailed study that proves the device meets such criteria in terms of clinical accuracy or utility.
Instead, the document focuses on:
- Technological equivalence: Stating that the main elements of the hardware, illumination, and monochromatic images are the same as the predicate.
- Verification of safety and regulatory compliance: Covering software evaluation, electrical safety, electromagnetic compatibility, biocompatibility, and compliance with recognized consensus standards for ophthalmic cameras (ISO 15004-1:2020, ISO 10940:2009, ANSI Z80.36:2021).
- Spectral accuracy and reliability of retinal images: This was evaluated in an "eye model using a reference material with tabulated spectral bands," but no specific acceptance criteria or performance metrics (e.g., sensitivity, specificity, accuracy for a particular clinical task) are provided from this evaluation.
Therefore, I cannot populate the requested table and information as the necessary details regarding acceptance criteria, reported device performance related to a specific clinical task, test set characteristics, expert involvement, adjudication, or a multi-reader multi-case study are not present in the provided text. The document describes tests to ensure the device functions safely and in a manner similar to its predicate, but not how well it performs a particular diagnostic task against established clinical benchmarks or ground truth.
Missing Information:
- No specific acceptance criteria for diagnostic performance (e.g., image quality metrics, sensitivity/specificity for detecting a condition).
- No reported clinical performance metrics directly related to the device's ability to 'capture images' that translates to a clinical outcome.
- No details on a clinical study where the device's images were evaluated for a specific purpose against a ground truth.
The document seems to describe the device as an image capture device, and the "performance testing" focuses on its technical specifications and safety rather than its diagnostic accuracy for a particular condition.
Ask a specific question about this device
(259 days)
Zilia Ocular is a non-contact imaging device for capturing, displaying and storing images of the retina under mydriatic conditions.
Zilia Ocular is intended for use as an aid to clinicians in the evaluation and documentation of ocular health.
The Zilia Ocular Fundus Camera (Zilia Ocular FC) is a non-contact, stand-alone device designed to image the eye fundus.
The device includes a base, a head support, an onboard computer with a 15.6'' touch-screen display and an optical head. A joystick, situated at the base of the device, controls the movements of the optical head to enable adequate positioning. The touch-screen display also includes a user interface allowing the user to control the device and acquire and display patient data and images.
The Zilia Ocular FC uses near-infrared (NIR) light-emitting diodes (LEDs) as illumination during alignment to the patient's pupil, and a white LED during focusing and imaging of the eye fundus. The region of the eye fundus being imaged can be modified by changing or moving the fixation target. The images are saved and stored on the onboard computer.
The dimensions of the Zilia Ocular FC are 620 mm (L) by 360 mm (W) by 500 mm (H) and it weighs about 32 kg.
The provided document is a 510(k) Summary for the Zilia Ocular Fundus Camera (Zilia Ocular FC). It describes the device, its intended use, and performance testing conducted to demonstrate substantial equivalence to predicate and reference devices.
However, the document does not contain specific acceptance criteria, reported device performance data against those criteria, or the details of a clinical study (such as sample size, data provenance, expert ground truth establishment, adjudication methods, MRMC studies, or standalone algorithm performance).
The document explicitly states: "Clinical Performance Data: This section is not applicable because clinical data was not provided for this 510(k) submission."
Therefore, I cannot provide the information requested in points 1, 2, 3, 4, 5, 6, 7, 8, and 9 based on the provided text. The submission relied on non-clinical performance testing (electrical safety, EMC, recognized standards, risk management, biocompatibility, processing of healthcare products, software evaluation, and usability) to demonstrate substantial equivalence, rather than clinical performance data.