Search Filters

Search Results

Found 31 results

510(k) Data Aggregation

    K Number
    K222831

    Validate with FDA (Live)

    Date Cleared
    2023-09-13

    (359 days)

    Product Code
    Regulation Number
    864.7290
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    CRYOcheck Factor VIII Deficient Plasma with VWF is for clinical laboratory use as a deficient substrate in the quantitative determination of Factor VIII activity in 3.2% citrated human plasma based on the activated partial thromboplastin time (APTT) assay. It is intended to be used in identifying factor VIII deficiency and as an aid in the management of hemophilia A in individuals aged 2 years and older. For in vitro diagnostic use

    Device Description

    CRYOcheck Factor VIII Deficient Plasma with VWF is normal human citrated plasma which has been immunodepleted of factor VIII and to which an exogenous source of human von Willebrand Factor (vWF) has been added and buffered with HEPES. Factor VIII has been assayed at less than 1% of normal activity levels and vWF antigen and activity are >50%. It will be provided to users frozen in small-volume aliquots (25 vials of 1.0 mL, and 25 vials of 1.5 mL). Vials will be packaged into boxes; these will be frozen during the manufacturing process and will be shipped and stored frozen until use to preserve the integrity of the components

    AI/ML Overview

    The provided FDA 510(k) summary for the "CRYOcheck Factor VIII Deficient Plasma with VWF" describes various performance studies and their results. Based on the document, here's a structured breakdown of the acceptance criteria and the study details:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly present a table of pre-defined acceptance criteria for each performance characteristic in a pass/fail format. Instead, it presents the results of various studies and implies that these results met internal or regulatory expectations for substantial equivalence to the predicate device. However, we can infer performance targets based on the documented results and common regulatory expectations for in vitro diagnostic devices.

    Here's a table summarizing the reported device performance, with inferred acceptance "criteria" based on the conclusions drawn in the report:

    Performance CharacteristicInferred Acceptance Criteria (Based on reported success)Reported Device Performance
    Precision (Multi-Reagent Lot)<10% CV for controls and high FVIII samples; <0.5 SD for low/very low FVIII samples (implied as 'acceptable')Aggregated Data (Lots 1, 2, and 3): - CRYOcheck Reference Control Normal: 5.7% CV - CRYOcheck Abnormal 1 Reference Control: 6.2% CV - CRYOcheck Abnormal 2 Reference Control: 5.2% CV - High FVIII Plasma Sample: 6.0% CV - Low FVIII Plasma Sample: 0.2 SD (11.1% CV) - Very Low FVIII Plasma Sample: 0.0 SD (24.1% CV) (Note: 24.1% CV for "Very Low" indicates higher variability, but the SD is 0.0, suggesting it met the SD criterion for this range)
    Reproducibility (Multi-Reagent Lot Site-to-Site)<10% CV for controls and high FVIII samples; <0.5 SD for low/very low FVIII samples (implied as 'acceptable')Pooled 3-Site Data: - Reference Control Normal: 6.04% CV - Abnormal 1 Reference Control: 4.75% CV - Abnormal 2 Reference Control: 7.68% CV - High FVIII Plasma Sample: 5.07% CV - Low FVIII Plasma Sample: 0.18 SD (9.39% CV) - Very Low FVIII Plasma Sample: 0.04 SD (38.56% CV) (Note: 38.56% CV for "Very Low" indicates higher variability, but the SD is 0.04, suggesting it met the SD criterion for this range)
    Linearity/Assay Reportable RangeDemonstration of linearity across the claimed reportable range (implied as 'acceptable' based on graphical analysis or statistical fit).Linearity Range: 0 to 230% FVIII activity. The study concluded this range is supported.
    Reference IntervalEstablishment of a clinically meaningful and statistically derived reference interval from a healthy population.Reference Interval: 62 to 163 % FVIII activity (calculated as the non-parametric 95% confidence interval, 2.5th to 97.5th percentiles).
    Shelf-Life StabilityMaintenance of FVIII activity within acceptable limits over the claimed shelf-life.A shelf-life stability claim of 24 months when stored at -40 °C to -80 °C, based on data up to 25 months (real-time study ongoing).
    In-Use StabilityMaintenance of FVIII activity within acceptable limits for the claimed on-board and refrigerated stability periods.A stability claim of 24 hours on board the instrument and at 2-8 °C.
    InterferencesNo significant interference from common substances at specified concentrations.No interference from Hemoglobin (≤1000 mg/dL), Intralipid (≤2000 mg/dL), Bilirubin (conjugated, ≤4.0 mg/dL), Bilirubin (unconjugated, 40 mg/dL), Lupus Anticoagulant (≤1.8 dRVVT ratio), Warfarin (≤ INR ratio 2.72). Certain anticoagulants (Rivaroxaban, fondaparinux, dabigatran, emicizumated heparin, and low molecular weight heparin) were shown to interfere.
    Recovery of FVIII ReplacementsAccurate evaluation (100 ± 25%) of the potency of specified FVIII replacement therapies.Accurately evaluated (100 ± 25%) Advate (92.84%), Afstyla* (97.38% with 2x correction factor), Elocate (94.90%), Jivi (104.49%), Novoeight (113.33%), and Wilate (94.92%) at concentrations 0.05 to 1.0 IU/mL.
    Method ComparisonDemonstrating substantial equivalence to the predicate device via Passing-Bablok regression (e.g., slope close to 1, intercept close to 0, high correlation coefficient).Overall (N=366): Slope = 1.16 (CI: 1.15 to 1.18), Intercept = -0.08 (CI: -0.25 to 0.13), Pearson Correlation Coefficient (r) = 0.96. Predicted biases at medical decision levels were reported (e.g., 0.37% at 1% FVIII, 13.70% at 100% FVIII). The study concluded it performed equivalently.
    Sample IntegritySupport for claimed fresh sample stability and frozen storage stability, including freeze-thaw cycles.Supported a fresh sample stability claim of 2 hours at room temperature and a frozen storage claim of 1.5 months at ≤-70 ℃, including up to two freeze-thaw cycles.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Multi-Reagent Lot Precision: Not explicitly stated as a "test set" sample size for patient plasma, but control samples were tested 80 replicates per sample per lot across 3 lots.
    • Multi-Reagent Lot Site to Site Reproducibility: 270 replicates per sample (control or plasma) for the pooled data, across 3 sites and 3 lots. Patient samples included high, low, and very low FVIII plasma.
    • Linearity/Assay Reportable Range: Fifteen sample dilutions, each tested in quadruplicate.
    • Reference Interval: 136 normal, ostensibly healthy individuals.
    • Shelf-Life Stability: Five replicates of one normal and two abnormal reference controls and two patient plasma samples per timepoint, across 3 lots.
    • In-Use Stability: Five replicates of one normal and two abnormal reference controls and two patient plasma samples per timepoint, across 3 lots.
    • Interferences: 10 replicates of spiked samples and 10 replicates of control for each possible interferent.
    • Recovery of FVIII Replacements: Congenital FVIII deficient plasma spiked with six FVIII replacement therapies at seven concentrations (total 42 conditions).
    • Method Comparison: 366 human plasma samples (from normal ostensibly healthy individuals, patients with congenital or acquired hemophilia A, patients with hemophilia B, patients with von Willebrand disease, hemophilia A patients on FVIII replacement therapies, and patients with other factor deficiencies). These samples were distributed across four sites.
    • Sample Integrity: 94 plasma samples.

    Data Provenance: The document does not specify the country of origin for any of the patient samples. All studies appear to be prospective in design, as they describe the testing of samples under controlled conditions to evaluate device performance characteristics.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This device is an in vitro diagnostic (IVD) reagent for determining Factor VIII activity in human plasma, which is a quantitative measurement, not an interpretative one like image analysis. Therefore, the concept of "experts establishing ground truth" in the way it applies to diagnostic algorithms (e.g., radiologists for imaging) is not directly relevant here.

    The "ground truth" for the performance studies is established by:

    • The known concentrations/characteristics of reference controls.
    • The actual FVIII activity levels as measured by a legally marketed predicate device (HemosIL Factor VIII Deficient Plasma) in the method comparison study.
    • The carefully prepared spiked samples for linearity and recovery studies.
    • The clinical diagnosis categories of patient samples (e.g., congenital hemophilia A).

    There is no mention of human experts (e.g., pathologists, hematologists) adjudicating "ground truth" for individual cases in these studies beyond ensuring appropriate sample selection and handling. The performance is assessed against quantitative benchmarks and comparison with a known predicate.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    As mentioned above, the device performs a quantitative measurement, not an interpretative diagnosis requiring adjudication. Therefore, no adjudication method (like 2+1 or 3+1 for expert consensus on image interpretation) was used or is applicable for this type of IVD device testing. The "ground truth" is analytical and quantitative.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No. An MRMC comparative effectiveness study is not applicable as this is an in vitro diagnostic reagent, not an AI-powered diagnostic tool that assists human readers. The device itself performs the measurement, and its performance is evaluated directly through analytical studies, not through its impact on human reader performance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This refers to the performance of the device itself (the reagent in combination with instrumentation) without human interpretation affecting the measurement result. All the analytical performance studies (precision, reproducibility, linearity, stability, interferences, recovery, method comparison, sample integrity) represent the standalone performance of the CRYOcheck Factor VIII Deficient Plasma with VWF in conjunction with the specified automated APTT assay systems. The "human-in-the-loop" here refers to the user performing the test, but the performance characteristics relate to the analytical accuracy and reliability of the automated test method itself.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    The ground truth used in these studies is primarily analytical and quantitative:

    • Known concentrations/values of commercially available reference controls.
    • Known FVIII activity levels measured by a predicate device for method comparison.
    • Expected FVIII activity levels in carefully fabricated samples (e.g., highly concentrated plasma diluted with congenital hemophilia A plasma for linearity, spiked samples for recovery).
    • Clinical classification of patient samples (e.g., "patients with congenital or acquired hemophilia A") as the basis for sample selection in the method comparison and reference interval studies, though specific individual diagnoses are not the "ground truth" for each quantitative measurement, rather the context.

    8. The sample size for the training set

    This document describes a premarket notification (510(k)) for an in vitro diagnostic reagent. IVD reagents are not typically "trained" in the machine learning sense. Therefore, there is no training set in the context of an AI/ML algorithm. The studies detailed are analytical verification and validation studies to demonstrate performance and establish substantial equivalence to a predicate device.

    9. How the ground truth for the training set was established

    Since there is no training set for this type of device, this question is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K180486

    Validate with FDA (Live)

    Date Cleared
    2018-03-22

    (27 days)

    Product Code
    Regulation Number
    864.7290
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    HemosIL Factor XII Deficient Plasma immunodepleted of factor XII and intended for the in vitro diagnostic quantitative determination of factor XII activity in citrated plasma, based on the activated partial thromboplastin time (APTT) assay, on IL Coagulation Systems.

    Device Description

    Factor XII activity in a patient's plasma is determined by performing a modified activated partial thromboplastin time test (APTT). Patient plasma is diluted and added to plasma deficient in factor XII. Correction of the clotting time of the deficient plasma is proportional to the concentration (% activity) of that factor in the patient plasma, interpolated from a calibration curve.

    AI/ML Overview

    The provided text describes a 510(k) submission for a medical device called "HemosIL Factor XII Deficient Plasma." This submission is a "Special 510(k)" because the changes are limited to an updated on-board instrument stability claim, which means only specific performance characteristics related to this change were tested.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    Key Takeaway from the Document: This is not a study proving the overall efficacy or safety of a new diagnostic algorithm or AI system. Instead, it's a regulatory submission for a minor modification to an existing reagent used in a diagnostic test. The study described focuses solely on validating a change in the reagent's "on-board instrument stability" claim, meaning how long the reagent is stable once placed on the testing instrument. Therefore, many of the typical acceptance criteria and study aspects found in AI/algorithm submissions (like MRMC studies, expert consensus on images, etc.) are not applicable here.


    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implicitly defined by standard laboratory validation practices for a changed stability claim, specifically referencing CLSI EP25-A. The performance is the outcome of this validation.

    Acceptance Criterion (Implied)Reported Device Performance
    Validation of "on-board instrument stability" claimNew claim of 6 Hours for on-board stability on ACL TOP Family and ACL TOP Family 50 Series coagulation systems. (Previously 24 Hours)
    Compliance with CLSI EP25-A requirements for validation of stabilityTesting was conducted "under design control and in accordance with CLSI EP25-A."
    Continued equivalence to predicate device in other performance characteristicsStated: "HemosIL Factor XII Deficient Plasma ... share the same Intended Use/ Indications for Use, same test principle, same formulation and the same performance characteristics, except for the updated on-board instrument claim."

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document does not explicitly state the sample size (number of patient samples, runs, or measurements) used for the stability testing. It only mentions "additional testing done to current CLSI EP25-A requirements" and that "verification testing to establish the modified on-board instrument stability claim... was conducted under design control."
    • Data Provenance: The document does not specify the country of origin of the data or whether the testing was retrospective or prospective. It implies the testing was conducted by Instrumentation Laboratory Co. as part of their design control process. Given it's a stability study for a reagent, it would inherently be a prospective study conducted in a laboratory setting.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • Not Applicable. This is not a study involving expert readers interpreting medical images or data where "ground truth" is established by human consensus. The "ground truth" for a reagent's stability is determined by laboratory measurements and adherence to pre-defined acceptance limits for analytical performance (e.g., maintaining accuracy, precision within a certain range over time).

    4. Adjudication Method for the Test Set

    • Not Applicable. As this is a laboratory stability study for a reagent, there is no "adjudication" of human interpretations involved. The "adjudication" would be based on objective analytical measurements compared against pre-defined statistical criteria (e.g., drift, bias within acceptable limits according to CLSI EP25-A).

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    • No. An MRMC study is relevant for evaluating human performance (e.g., diagnostic accuracy of radiologists) with and without AI assistance. This submission is for a reagent, not an AI system or diagnostic algorithm that interacts with human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • No. This is not an algorithm or AI system. It's a laboratory reagent. The "performance" being evaluated is the chemical stability of the reagent on an automated instrument.

    7. The Type of Ground Truth Used

    • Analytical Performance Data / Reference Method. The "ground truth" for reagent stability in this context would be defined by the reagent's ability to produce accurate and precise results when tested immediately after reconstitution (time zero) and then over the claimed stability period (e.g., 6 hours). This would be compared against a reliable reference method or pre-established acceptable analytical performance ranges. The CLSI EP25-A guideline would specify how this analytical "ground truth" is established and evaluated.

    8. The Sample Size for the Training Set

    • Not Applicable. There is no "training set" as this is not a machine learning model or AI algorithm. This is a chemical reagent.

    9. How the Ground Truth for the Training Set was Established

    • Not Applicable. As there is no training set, this question is not relevant.
    Ask a Question

    Ask a specific question about this device

    K Number
    K102908

    Validate with FDA (Live)

    Device Name
    NOFACT IX
    Date Cleared
    2011-12-19

    (444 days)

    Product Code
    Regulation Number
    864.7290
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    NoFact IX Deficient Plasma is a human plasma immunodepleted of Factor IX and intended for the quantitative determination of Factor IX activity in citrated plasma from patients suspected of FIX deficiency. FIX activity is based on the activated partial thromboplastin time. For in vitro diagnostic use.

    Device Description

    NoFact IX Deficient Plasma is a human plasma immunodepleted of Factor IX and intended for the quantitative determination of Factor IX activity in citrated plasma from patients suspected of FIX deficiency. FIX activity is based on the activated partial thromboplast time. For in vitro diagnostic use.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study detailed in the provided 510(k) summary for the NoFact IX Deficient Plasma device:

    Important Note: The provided document is a 510(k) summary for an in vitro diagnostic device, specifically a deficient plasma product used in laboratory tests. It is not an AI or imaging device, and therefore many sections of your request (like "Number of experts used to establish the ground truth," "Adjudication method," "MRMC comparative effectiveness study," and "Standalone performance") are not applicable to this type of medical device submission. I will address only the relevant information from the document.


    Acceptance Criteria and Device Performance for NoFact IX Deficient Plasma

    The device, NoFact IX Deficient Plasma, is an in vitro diagnostic intended for the quantitative determination of Factor IX activity. Its performance is demonstrated through a comparison study with a legally marketed predicate device (STA Deficient IX) and precision testing.

    1. Table of Acceptance Criteria and Reported Device Performance

    Since this is a 510(k) for an in vitro diagnostic, the "acceptance criteria" are not explicitly stated with numerical thresholds in the provided summary in the same way they might be for an AI model. Instead, the acceptance is based on demonstrating "substantial equivalence" to a predicate device through a strong correlation in quantitative measurements and acceptable precision. The reported performance metrics are derived from this equivalence study and precision assessment.

    Performance MetricAcceptance Criteria (Implied for Substantial Equivalence)Reported Device Performance (NoFact IX vs. Predicate)
    Comparative Study (Linear Regression)All Labs (N=233)
    - Coefficient of Determination (r²)Strong correlation with predicate device (typically r² > 0.90 is desired for quantitative comparison)0.915
    - Correlation Coefficient (r)Strong linear relationship0.956
    - SlopeClose to 1 (indicating proportional agreement)0.858
    - InterceptClose to 0 (indicating minimal constant bias)5.729
    Precision (CV%) - Normal ControlAcceptable within-run and lot-to-lot variability for an IVD measuring coagulation factors (target typically <10% for CV)Within-run: 5.1%, Lot-to-Lot: 1.1%, Within-Device: 7.2%
    Precision (CV%) - Abnormal ControlAcceptable within-run and lot-to-lot variabilityWithin-run: 4.6%, Lot-to-Lot: 2.9%, Within-Device: 7.7%
    Precision (CV%) - Low FIX PatientAcceptable within-run and lot-to-lot variabilityWithin-run: 5.7%, Lot-to-Lot: 2.2%, Within-Device: 7.8%

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Comparative Study (Test Set): 233 plasma samples.
    • Data Provenance: The plasma samples were analyzed across three different laboratories ("sites"). The document does not specify the country of origin, nor whether the samples were collected retrospectively or prospectively. It just indicates they were "plasma samples."

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This is not applicable as this is an in vitro diagnostic device where the "ground truth" is established by the highly controlled, quantitative measurement from an established predicate device and internal controls, not expert interpretation of images or clinical data.

    4. Adjudication Method for the Test Set

    This is not applicable for an in vitro diagnostic device that quantifies a substance in plasma. There is no human interpretation or adjudication involved in generating the raw measurement data.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

    This is not applicable. An MRMC study is relevant for imaging or diagnostic tools where human readers interpret results, and the AI is meant to assist that interpretation. This device is a reagent used in an automated laboratory test.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This is not applicable in the context of AI algorithms. This device is a reagent; its "standalone" performance is assessed by its ability to accurately measure Factor IX activity in a laboratory setting, which is directly addressed by the comparative study and precision data.

    7. The Type of Ground Truth Used

    The "ground truth" in this context is established by:

    • The Stago PTT-A FIX assay using the predicate STA IX deficient plasma. The comparison seeks to show that the NoFact IX Deficient Plasma yields equivalent quantitative results to this established method.
    • Known control plasma samples (Normal Control Plasma, Abnormal Control Plasma) and pooled patient plasma with a known Low FIX concentration for precision testing.

    8. The Sample Size for the Training Set

    This is not applicable. This is not an AI or machine learning device that utilizes a "training set." The device is a manufactured biological reagent.

    9. How the Ground Truth for the Training Set Was Established

    This is not applicable for the reasons stated above.

    Ask a Question

    Ask a specific question about this device

    K Number
    K102851

    Validate with FDA (Live)

    Device Name
    NOFACT VIII
    Date Cleared
    2011-12-19

    (446 days)

    Product Code
    Regulation Number
    864.7290
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    NoFact VIII Deficient Plasma is a human plasma immunodepleted of Factor VIII and intended for the quantitative determination of Factor VIII activity in citrated plasma from patients suspected of FVIII deficiency. FVIII activity is based on the activated partial thromboplastin time. For in vitro diagnostic use.

    Device Description

    NoFact VIII Deficient Plasma is a human plasma immunodepleted of Factor VIII and intended for the quantitative determination of Factor VIII activity in citrated plasma from patients suspected of FVIII deficiency. FVIII activity is based on the activated partial thromboplastin time. For in vitro diagnostic use.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study details for the NoFact VIII Deficient Plasma device, based on the provided 510(k) summary:

    1. Table of Acceptance Criteria and Reported Device Performance

    The 510(k) summary for NoFact VIII Deficient Plasma focuses on demonstrating substantial equivalence to a predicate device (Stago VIII Deficient Plasma) through performance comparison and precision evaluation. While explicit "acceptance criteria" for each metric are not stated numerically, the narrative implies that the performance must be comparable to the predicate device and demonstrate acceptable precision.

    MetricAcceptance Criteria (Implied)Reported Device Performance (NoFact VIII Deficient Plasma)
    Comparative PerformanceHigh correlation and agreement with predicate device.Linear Regression: y = 0.8453x + 4.2111 (across all labs)
    (NoFact vs. Predicate)Slope close to 1, Intercept close to 0, R² close to 1.R² = 0.9683 (across all labs)
    Site 1: Slope 0.861, Intercept 2.8, R² 0.991
    Site 2: Slope 0.914, Intercept 2.5, R² 0.986
    Site 3: Slope 0.831, Intercept 5.9, R² 0.953
    Precision (CV%)Acceptable precision for quantitative measurement methods.Normal Control Plasma (91.6% FVIII): Within-run CV 4.2%, Lot-to-Lot CV 0.63%, Within-Device CV 6.8%
    (Often defined by CLSI guidelines or industry standards for similar assays)*Abnormal Control Plasma (33.3% FVIII): Within-run CV 4.9%, Lot-to-Lot CV 3.8%, Within-Device CV 8.0%
    Low FVIII Pooled Patient Plasma (11.8% FVIII): Within-run CV 5.7%, Lot-to-Lot CV 0.0%, Within-Device CV 8.5%

    Note: The document states "NoFact VIII Deficient Plasma provided acceptable precision" without numerical thresholds, implying adherence to general CLSI guidelines for precision for such assays.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: 233 frozen plasma samples.
    • Data Provenance: The samples were analyzed across three different laboratories. The document does not specify the country of origin but implies a multi-site study within a regulatory context that often points to domestic (US) or internationally recognized sites. The data is retrospective as it refers to "frozen plasma samples."

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This type of in vitro diagnostic device (IVD) for factor deficiency measurement typically does not use human experts to establish "ground truth" for individual samples in the way a diagnostic imaging AI might. Instead, the "ground truth" for the test set is established by the measurement of Factor VIII activity using the predicate device's reagent (Stago FVIII deficient plasma) coupled with the Stago PTT-A FVIII assay. The comparison is between the new device's performance and the established performance of the legally marketed predicate. Therefore, no experts were specifically used to establish ground truth in the context of interpretation or diagnosis for each sample.

    4. Adjudication Method for the Test Set

    As explained above, this study design compares the new device’s performance to an existing, validated method (the predicate device). There is no "adjudication" in the sense of resolving discrepancies between multiple human readers or between an AI and a human. The "ground truth" is the result obtained with the predicate device, and the new device's results are compared directly to those values.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This device is an in vitro diagnostic reagent, not an AI-powered diagnostic system that assists human readers in interpreting complex data like medical images. Therefore, the concept of "human readers improving with AI assistance" does not apply here.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes, in spirit, a standalone performance was done. The "NoFact VIII Deficient Plasma" is a reagent. Its performance is evaluated directly in an assay (Stago PTT-A FVIII assay) on an automated analyzer (STA Compact). The results (Factor VIII activity) are generated solely by the interaction of the sample, the reagent, and the analyzer, and then compared to the results obtained with the predicate reagent on the same analyzer using the same assay. There is no human "loop" involved in generating the primary measurement data, only in setting up the assay and validating the results.

    7. The Type of Ground Truth Used

    The ground truth used for performance comparison was the quantitative Factor VIII activity values obtained from the predicate device (Stago VIII Deficient Plasma) using the Stago PTT-A FVIII assay. This represents a "reference standard" established by an already legally marketed and validated assay.

    8. The Sample Size for the Training Set

    The document does not provide information on a "training set." This is typical for an in vitro diagnostic reagent. Reagents are generally developed and then their performance is validated against predicate devices or established methods. There isn't an "AI model" that requires a distinct training phase with labeled data in the same way an imaging algorithm would. The development (analogous to training) would involve optimizing the reagent's formulation and manufacturing processes, but this isn't detailed as a "training set" in the context of performance studies.

    9. How the Ground Truth for the Training Set Was Established

    Since no specific "training set" is mentioned for an AI model, the concept of establishing ground truth for it is not applicable here. The focus is on validating the performance of the final reagent product against a recognized standard (the predicate device).

    Ask a Question

    Ask a specific question about this device

    K Number
    K110237

    Validate with FDA (Live)

    Date Cleared
    2011-07-08

    (163 days)

    Product Code
    Regulation Number
    864.7290
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    HemosIL Factor VIII deficient plasma is human plasma, depleted of Factor VIII, which is intended for the in vitro diagnostic quantitative determination of Factor VIII activity in citrated plasma, based on the activated partial thromboplastin time (APTT) assay, on the ACL TOP® Family analyzers. HemosIL Factor VIII deficient plasma is indicated for use on patients who are suspected of congenital or acquired deficiency based on the activated partial thromboplastin time (APTT) assay results.

    Device Description

    The assay determines the functional activity of Factor VIII by measuring the degree of prolongation of activated partial thromboplastin time in the presence of a contact activator, thromboplastin, phospholipids and calcium ions. Factor VIII activity is correlated with the prolongation of the clotting time of the Factor VIII deficient plasma to which diluted patient sample has been added.

    AI/ML Overview

    Here's a breakdown of the requested information based on the provided text, focusing on the acceptance criteria and the study that proves the device meets them:

    Acceptance Criteria and Device Performance

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document primarily details performance characteristics rather than explicit "acceptance criteria" for regulatory approval in every instance. However, it implicitly defines satisfactory performance through comparisons to the predicate device and established linearity/precision targets. Due to the nature of the submission (Reagent formulation change), the primary acceptance criterion is demonstrating comparable performance to the predicate device.

    Performance MetricAcceptance Criteria (Implicit/Explicit)Reported Device Performance
    Precision (CV%)Within acceptable limits for diagnostic assays (not explicitly stated as a numerical target in this summary, but typical for CLSI EP05-A2). The predicate device's performance would serve as the benchmark.Within Run CV%: 3.7 - 4.1% (ACL TOP 500 CTS) Total CV%: 3.9 - 6.4% (ACL TOP 500 CTS) (These values are for various control levels with SynthASil reagent and a representative instrument/lot).
    Linearity (R², Slope, Intercept)R² close to 1.0, slope close to 1.0, and intercept close to 0.0, indicating proportional response across the analytical range.R²: 0.995 - 0.998 Slope: 0.964 - 1.045 Intercept: -1.621 - 2.261 (Demonstrated linearity for activities up to 150% for both reagent lots and ACL TOP Family instruments).
    Analytical RangeDefined operational range for Factor VIII activity.0.1 - 150% (with SynthASil) 0.5 - 150% (with APTT-SP)
    InterferenceNo significant interference from common interferents at specified concentrations (e.g., hemoglobin, triglycerides, bilirubin, FVIII inhibitors, Lupus anticoagulant antibodies).No effect from hemoglobin up to 530 mg/dL, triglycerides up to 2000 mg/dL, bilirubin up to 150 mg/dL, Factor VIII inhibitors up to 0.1BU, and Lupus anticoagulant antibodies.
    Method Comparison (In-House)Slope: Close to 1.0 Correlation Coefficient (R): Close to 1.0 (indicating strong agreement with the predicate device). "Satisfy the product specifications for method comparison and demonstrate comparable performance."With SynthASil: Slope: 1.064 (ACL TOP), 1.112 (ACL TOP 500 CTS); R: 0.9885 (ACL TOP), 0.9923 (ACL TOP 500 CTS) With APTT-SP: Slope: 0.893 (ACL TOP), 0.913 (ACL TOP 500 CTS); R: 0.9884 (ACL TOP), 0.9926 (ACL TOP 500 CTS) Conclusion: Satisfies product specifications and demonstrates comparable performance.
    Method Comparison (Field Site)Slope: Close to 1.0 Correlation Coefficient (R): >0.98 "Statistically similar performance to the predicate device."Slope: 0.871 - 1.138 Correlation Coefficient (R): >0.9829 Conclusion: Performance is statistically similar.
    Stability/Shelf LifeDemonstrated shelf life under specified storage conditions. The claimed shelf life would need to be supported by stability data.12 months when stored at 2-8°C (based on accelerated stability, real-time ongoing).

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Precision Study:

      • Sample Size: 80 replicates per instrument/lot per control level (e.g., 4 control levels x 3 lots x 3 instruments x 80 = 2880 individual measurements if all were run, but data is shown for one representative instrument/reagent/lot).
      • Data Provenance: Not explicitly stated, but clinical laboratory samples (normal and abnormal types) were used. Implied to be prospective within the context of the study design (20 days, 2 runs/day, 2 replicates/run). Country of origin is not specified, but the applicant is US-based.
    • Linearity Study:

      • Sample Size: 4 replicates for each Factor VIII sample level.
      • Data Provenance: Not explicitly stated, but generated in-house using prepared Factor VIII samples ranging from <1% to >150%. Implied prospective within the study.
    • Interference Study:

      • Sample Size: Not specified, but involved testing samples spiked with various interferents.
      • Data Provenance: Not specified, likely in-house laboratory testing, prospective.
    • In-House Method Comparison Study:

      • Sample Size: Not explicitly stated for the regression analysis. The table shows 'N' for the ACL TOP 500 CTS as 90 (likely 90 samples tested).
      • Data Provenance: "In-House Study," likely prospective laboratory testing.
    • Field Site Study:

      • Sample Size:
        • OUS site: n = 109
        • US #1 site: n = 125
        • US #2 site: n = 102
      • Data Provenance: Conducted at "2 US sites and one OUS site." The document states "Both normal and abnormal samples were tested." Implied prospective testing at these sites.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This device is an in vitro diagnostic reagent kit for quantitative determination of Factor VIII activity. The "ground truth" for Factor VIII activity in the test samples is typically established by assigned values from reference materials or by established laboratory methods using calibrated instruments, not by expert human interpretation like in imaging or pathology. The document mentions "assigned values" for the HemosIL Calibration plasma (for calibration curves) and "assigned values" for samples in the linearity study, but does not detail how these assignments were made in terms of human experts. It's a quantitative chemical assay, where "ground truth" is analytical, not based on expert consensus for a "reading."


    4. Adjudication Method for the Test Set

    Not applicable. This is a quantitative diagnostic assay, not a diagnostic imaging or pathology device where human interpretation and adjudication would be typically required for a "test set" in the sense of establishing truth for diagnostic accuracy. The performance is assessed against known or assigned values for Factor VIII activity.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a diagnostic reagent, not an AI-powered diagnostic aid for human readers. It functions as a standalone laboratory test.


    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes. This device (HemosIL Factor VIII deficient plasma) is a reagent used in an automated assay on ACL TOP Family analyzers. Its performance characteristics (precision, linearity, interference, method comparison) are evaluated as a standalone component of the in vitro diagnostic system, independent of direct human interpretive "reading" of the results in the same way an imaging device would be read. The "performance" described is the analytical performance of the reagent used with the specified instruments.


    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    For the analytical studies (precision, linearity, method comparison), the "ground truth" is based on:

    • Reference materials/assigned values: HemosIL Calibration plasma values are assigned for Factor VIII Activity and used to calibrate the standard curve. In the linearity study, samples had "assigned values."
    • Established methods: The predicate device itself serves as the "truth" for comparative purposes in the method comparison studies.
    • Known concentrations: For interference studies, samples are spiked with known concentrations of interferents.

    This is primarily an analytical ground truth derived from laboratory standards, calibrated materials, and established predicate device performance.


    8. The Sample Size for the Training Set

    Not applicable. This is an IVD reagent, not a machine learning or AI-based device that requires a "training set" in the context of algorithm development. The reagent itself does not learn or get "trained."


    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no "training set" for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K050661

    Validate with FDA (Live)

    Date Cleared
    2005-05-02

    (48 days)

    Product Code
    Regulation Number
    864.7290
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    HemosIL Factor II Deficient Plasma is human plasma immunodepleted of factor II and intended for the in vitro diagnostic quantitative determination of factor II activity in citrated plasma, based on the prothrombin time (PT) assay, on IL Coagulation and ELECTRA Systems.

    Device Description

    HemosIL Factor II Deficient Plasma is human plasma immunodepleted of factor II and intended for the in vitro diagnostic quantitative determination of factor II activity in citrated plasma, based on the prothrombin time (PT) assay, on IL Coagulation and ELECTRA Systems.

    Abnormalities of the extrinsic pathway factors are determined by performing a modified prothrombin time (PT) test. Patient plasma is diluted and added to a plasma deficient in factor II. Correction of the clotting time of the deficient plasma is proportional to the concentration (% activity) of the factor II in the patient plasma, interpolated from a calibration curve.

    AI/ML Overview

    The provided 510(k) summary for the HemosIL Factor II Deficient Plasma does not explicitly state acceptance criteria in the form of pre-defined thresholds that the device had to meet. Instead, the study demonstrates that the new device is "substantially equivalent" to predicate devices by showing comparable performance.

    However, we can infer performance metrics that were deemed acceptable for substantial equivalence based on the presented data. The study primarily relies on method comparison and precision data.

    Here's a breakdown of the information requested, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    As noted, explicit "acceptance criteria" are not given. The study aims to demonstrate substantial equivalence to predicate devices. We can infer that correlation coefficients (r) close to 1.0, slopes close to 1.0, and small intercepts compared to predicate devices, along with acceptable precision (%CV), were considered evidence of substantial equivalence.

    Performance MetricImplied "Acceptance Range" (for substantial equivalence)Reported Device Performance (Range Across Systems)
    Method Comparison (vs. Predicate/Reference Device)
    Correlation Coefficient (r)Close to 1.0 (e.g., typically > 0.95 or 0.98 for strong correlation in equivalent devices)0.9855 to 0.9954
    SlopeClose to 1.0 (e.g., typically 0.95-1.05)1.0357 to 1.0603
    InterceptClose to 0 (e.g., often within a small clinically acceptable range)-4.6831 to 1.0196
    Within Run Precision (%CV)Typically a low percentage, e.g., < 5% or < 10% depending on analyte and clinical context.2.3% to 6.2%
    Between Run Precision (%CV)Typically a low percentage, e.g., < 5% or < 10% depending on analyte and clinical context.2.0% to 6.4%

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size (Method Comparison): Approximately 60 citrated plasma samples for the in-house method comparison study. A separate clinical study used n=61 samples.
    • Data Provenance:
      • Country of Origin: Not explicitly stated, but the submission is from Massachusetts, USA. The studies are described as "in-house" and "clinical," suggesting they were conducted by the manufacturer or collaborators, likely within the US, but this is not definitively stated.
      • Retrospective or Prospective: Not explicitly stated. The description "evaluating approximately 60 citrated plasma samples" and "clinical study" doesn't specify the collection method. However, for method comparison, samples are often either prospectively collected or selected from an existing bank that represents the intended use population.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts

    • Ground Truth Establishment: For in-vitro diagnostic (IVD) devices like this, "ground truth" for method comparison and precision is established by the reference method (the predicate device) and by validated control materials, rather than by human expert consensus or pathology review.
    • Number and Qualifications of Experts: Not applicable in the traditional sense for this type of IVD performance study. The "truth" is based on the performance of a legally marketed predicate device and well-characterized control materials.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not applicable. This is a quantitative IVD method comparison study where results are numerical and compared against a reference method, not a subjective interpretation requiring adjudication among experts.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

    • MRMC Study: No, an MRMC comparative effectiveness study was not conducted. This type of study is typically relevant for diagnostic imaging where human readers interpret results. The HemosIL Factor II Deficient Plasma is an in-vitro diagnostic reagent designed for automated analyzers.

    6. If a Standalone Study Was Done

    • Standalone Study: Yes, the performance data presented is for the algorithm/device (HemosIL Factor II Deficient Plasma) operating on various IL Coagulation and ELECTRA systems. The "method comparison" directly compares the new device's performance against legally marketed predicate devices, demonstrating its standalone functionality in achieving comparable results. The precision studies also show its standalone performance.

    7. The Type of Ground Truth Used

    • Type of Ground Truth: The "ground truth" for assessment of the new device is based on the performance of the legally marketed predicate devices (Hemoliance Factor II Deficient Plasma and HemosIL Factor II Deficient Plasma predicate versions) and using known normal and abnormal control materials.

    8. The Sample Size for the Training Set

    • Training Set Sample Size: The document does not describe a "training set" in the context of machine learning or AI models. This device is a diagnostic reagent, and its development typically involves formulation optimization and analytical verification rather than algorithmic training on a dataset. Therefore, this concept is not directly applicable.

    9. How the Ground Truth for the Training Set Was Established

    • Ground Truth for Training Set: Not applicable, as there is no "training set" in the context of an AI/ML algorithm for this IVD device. The development of such a reagent involves analytical chemistry, reagent formulation, and stability studies, rather than training on a data set with an established ground truth.
    Ask a Question

    Ask a specific question about this device

    K Number
    K043459

    Validate with FDA (Live)

    Date Cleared
    2005-02-09

    (56 days)

    Product Code
    Regulation Number
    864.7290
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    HemosIL Factor XII Deficient Plasma is human plasma immunodepleted of factor XII and intended for the in vitro diagnostic quantitative determination of factor XII activity in citrated plasma, based on the activated partial thromboplastin time (APTT) assay, on IL Coagulation and ELECTRA Systems.

    Device Description

    HemosIL Factor XII Deficient Plasma is human plasma immunodepleted of factor XII and intended for the in vitro diagnostic quantitative determination of factor XII activity in citrated plasma, based on the activated partial thromboplastin time (APTT) assay, on IL Coagulation and ELECTRA Systems. Abnormalities of the intrinsic pathway factors are determined by performing a modified activated partial thromboplastin time (APTT) test. Patient plasma is diluted and added to a plasma deficient in factor XII. Correction of the clotting time of the deficient plasma is proportional to the concentration (% activity) of the factor XII in the patient plasma, interpolated from a calibration curve.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the HemosIL Factor XII Deficient Plasma, based on the provided text:

    Acceptance Criteria and Reported Device Performance

    For this specific device, the acceptance criteria are not explicitly stated as numerical cutoffs. Instead, the study aims to demonstrate substantial equivalence to predicate devices. This means that the new device's performance, as measured by method comparison and precision, should be comparable to the legally marketed predicates.

    The summary highlights:

    • Method Comparison: Consistency in results (slope and correlation coefficients) when compared to predicate devices across different IL systems. A slope close to 1 and a correlation coefficient (r) close to 1 indicate strong agreement.
    • Precision: Acceptably low variability (Within run %CV and Between Run %CV) for both normal and abnormal controls.
    Performance MetricSpecific Criterion (Implied for Substantial Equivalence)Reported Device Performance
    Method Comparison (New vs. Predicate HemosIL Factor XII Deficient Plasma on ACL Family)
    ACL 3000 (n=71)Slope ≈ 1, r ≈ 1Slope: 0.9284, r: 0.9882
    ACL 10000 (n=71)Slope ≈ 1, r ≈ 1Slope: 0.9116, r: 0.9704
    ACL Advance (n=70)Slope ≈ 1, r ≈ 1Slope: 1.0065, r: 0.9687
    ACL TOP (n=72)Slope ≈ 1, r ≈ 1Slope: 1.0739, r: 0.9551
    Method Comparison (New vs. Predicate Hemoliance Factor XII Deficient Plasma on ELECTRA)
    E1600C (n=73)Slope ≈ 1, r ≈ 1Slope: 0.9918, r: 0.9863
    Within Run Precision (%CV)Low %CV for both normal and abnormal controls
    ACL 9000 Normal Control1.9 %CV
    ACL 9000 Special Test Control Level 22.6 %CV
    ACL Futura Normal Control3.0 %CV
    ACL Futura Special Test Control Level 22.3 %CV
    ACL TOP Normal Control4.7 %CV
    ACL TOP Special Test Control Level 26.1 %CV
    ELECTRA 1600C Normal Control3.4 %CV
    ELECTRA 1600C Special Test Control Level 23.5 %CV
    Between Run Precision (%CV)Low %CV for both normal and abnormal controls
    ACL 9000 Normal Control3.4 %CV
    ACL 9000 Special Test Control Level 23.3 %CV
    ACL Futura Normal Control2.9 %CV
    ACL Futura Special Test Control Level 23.8 %CV
    ACL TOP Normal Control3.1 %CV
    ACL TOP Special Test Control Level 23.2 %CV
    ELECTRA 1600C Normal Control7.8 %CV
    ELECTRA 1600C Special Test Control Level 24.3 %CV

    Study Information

    1. Sample size used for the test set and the data provenance:

      • Sample Size (Method Comparison): Approximately 70-73 citrated plasma samples per comparison (e.g., 71 for ACL 3000, 70 for ACL Advance, 73 for ELECTRA 1600C).
      • Sample Size (Precision): Not explicitly stated how many unique samples were used to create the controls, but precision was assessed over multiple runs (n=80) for each instrument and control type.
      • Data Provenance: "In-house method comparison study." This suggests the data was generated by the manufacturer, likely in a controlled laboratory setting. The country of origin is not specified, but the manufacturer is based in Lexington, MA, USA. The study appears to be prospective for the new device as it's being compared to established predicates.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not applicable. This device is an in vitro diagnostic (IVD) for quantitative determination of factor XII activity, not an image-based diagnostic or clinical decision support system that relies on expert human interpretation for ground truth. The "ground truth" (or reference standard) in this context is the measurement obtained from the predicate devices.
    3. Adjudication method for the test set:

      • Not applicable. As above, this is an IVD device measuring a quantitative value, not subject to expert adjudication in the way an imaging study or qualitative diagnostic might be. The comparison is objective, based on measured values.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not applicable. This is an IVD device, not an AI-assisted diagnostic tool that involves human readers interpreting results.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Yes, this is a standalone device. Its performance is evaluated directly on patient samples (or controls) by comparing its measurements to those of predicate devices. The "algorithm" here is the biochemical assay and the instrument's processing of the reaction, without human interpretation as a performance variable.
    6. The type of ground truth used:

      • For the method comparison, the "ground truth" or reference standard for comparison was the measurements obtained from the predicate devices (HemosIL Factor XII Deficient Plasma on ACL Family and Hemoliance Factor XII Deficient Plasma on ELECTRA). These are established, legally marketed devices.
      • For precision, the ground truth is against the statistical mean of the measurements themselves, for "Normal Control" and "Special Test Control Level 2" materials.
    7. The sample size for the training set:

      • Not applicable in the typical sense of machine learning. This device functions as a biochemical assay, not an AI/ML model that requires a "training set" to learn from data. Its performance is based on the inherent chemical and biological principles of the assay and its manufacturing quality, not a learned algorithm.
    8. How the ground truth for the training set was established:

      • Not applicable for the reasons stated above.
    Ask a Question

    Ask a specific question about this device

    K Number
    K040362

    Validate with FDA (Live)

    Date Cleared
    2004-07-12

    (150 days)

    Product Code
    Regulation Number
    864.7290
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DiaPharma Factor X Kit is an in vitro diagnostic test kit for the quantitative determination of Factor X activity in human citrated plasma.

    Factor X activity is useful for monitoring patients on oral anticoagulant therapy (warfarin) where baseline PT values may be prolonged and INR results are not reliable, such as in OAC patients with lupus inhibitors. The DiaPharma FX kit provides health care providers with a tool for with Tapac Innovelor The DiaPharma Factor X Kit is also useful for screening for factor X deficiencies.

    Device Description

    Not Found

    AI/ML Overview

    The provided document does not contain information about acceptance criteria or a study that proves a device meets such criteria. This document is a 510(k) premarket notification letter from the FDA to DiaPharma Group, Inc. for their DiaPharma Factor X Kit. It primarily confirms the substantial equivalence of the device to a legally marketed predicate device and outlines regulatory compliance requirements.

    Therefore, I cannot fulfill your request for the specific information regarding acceptance criteria and a study demonstrating compliance. This type of detailed performance data is typically found in a clinical study report or a more extensive technical file, not in the FDA's initial 510(k) clearance letter.

    Ask a Question

    Ask a specific question about this device

    K Number
    K034007

    Validate with FDA (Live)

    Date Cleared
    2004-02-13

    (51 days)

    Product Code
    Regulation Number
    864.7290
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    HemosIL Factor VIII Deficient Plasma is human plasma immunodepleted of factor VIII and intended for the in vitro diagnostic quantitative determination of factor VIII activity in citrated plasma, based on the activated partial thromboplastin time (APTT) assay, on IL Coagulation and ELECTRA Systems.

    Device Description

    HemosIL Factor VIII Deficient Plasma is human plasma immunodepleted of factor VIII and intended for the in vitro diagnostic quantitative determination of factor VIII activity in citrated plasma, based on the activated partial thromboplastin time (APTT) assay, on IL Coagulation and ELECTRA Systems.

    Abnormalities of the intrinsic pathway factors are determined by performing a modified activated partial thromboplastin time (APTT) test. Patient plasma is diluted and added to a plasma deficient in factor VIII. Correction of the clotting time of the deficient plasma is proportional to the concentration (% activity) of the factor VIII in the patient plasma, interpolated from a calibration curve.

    AI/ML Overview

    The provided text describes the 510(k) summary for the HemosIL Factor VIII Deficient Plasma device, which is an in vitro diagnostic intended for quantitative determination of factor VIII activity. The study presented focuses on demonstrating substantial equivalence to predicate devices, rather than establishing specific clinical acceptance criteria in the way an AI/ML device would.

    However, I can extract the performance data provided and frame it in a table, even if explicit "acceptance criteria" are not stated as pass/fail thresholds in the document. The study's aim was to show comparable performance to predicate devices.

    1. A table of acceptance criteria and the reported device performance

    For a diagnostic device like this, "acceptance criteria" are typically met by demonstrating performance (like correlation and precision) that is comparable to or better than a legally marketed predicate device. The document states that the device is "substantially equivalent... in performance, intended use and safety and effectiveness."

    Performance MetricAcceptance Criteria (Implicit for Substantial Equivalence to Predicate)Reported Device Performance
    Method Comparison vs. Hemoliance Factor VIII Deficient Plasma (ELECTRA System)Similar slope and correlation coefficient to predicateE1400C: Slope = 0.9518, r = 0.9873
    Method Comparison vs. IL Test Factor VIII Deficient Plasma (ACL Family)Similar slope and correlation coefficient to predicateACL 300: Slope = 0.9391, r = 0.9942 ACL Advance: Slope = 1.0073, r = 0.9906
    Within-Run Precision (CV%)Acceptable precision for clinical use (relative to predicate or industry standard)ACL 9000: Normal Control (3.1%), Low Abnormal (2.2%) ACL Futura: Normal Control (2.7%), Low Abnormal (3.4%) ELECTRA 1600C: Normal Control (4.8%), Low Abnormal (4.1%)
    Between-Run Precision (CV%)Acceptable precision for clinical use (relative to predicate or industry standard)ACL 9000: Normal Control (3.1%), Low Abnormal (3.3%) ACL Futura: Normal Control (3.6%), Low Abnormal (4.0%) ELECTRA 1600C: Normal Control (6.0%), Low Abnormal (4.6%)

    2. Sample size used for the test set and the data provenance

    • Test Set Sample Size:
      • Method Comparison: Approximately 60 citrated plasma samples (30 normal and 30 abnormal) were used for each predicate comparison.
      • Within-Run and Between-Run Precision: Not explicitly stated as a number of samples, but assessed over "multiple runs (n=80) on different instruments using a specific lot of APTT reagent and both normal and abnormal samples." This refers to the number of runs, not individual patient samples. Typically, controls are run multiple times.
    • Data Provenance: Not specified in the document (e.g., country of origin, retrospective or prospective).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This type of in vitro diagnostic device study does not typically involve human expert adjudication for ground truth in the same way an AI image analysis device would. The "ground truth" for method comparison is the result obtained from the predicate device (another established diagnostic test). For precision, the "ground truth" is the expected value of the control sample. Therefore, this section is not applicable.

    4. Adjudication method for the test set

    Not applicable. As noted above, this study compares the device's measurements to predicate device measurements or established control values, not to adjudicated expert opinions.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is an in vitro diagnostic reagent, not an AI-assisted diagnostic tool that would involve human readers or image interpretation.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes, this is a standalone device. The performance data presented (method comparison, precision) reflects the device's intrinsic analytical performance when run on the specified coagulation systems. It's a reagent that works with an instrument to produce a result, without direct human-in-the-loop interpretation of that result in the way an AI would assist.

    7. The type of ground truth used

    • For Method Comparison: The "ground truth" was the results obtained from the predicate devices (Hemoliance Factor VIII Deficient Plasma and IL Test Factor VIII Deficient Plasma) on their respective analytical systems.
    • For Precision: The "ground truth" was the assigned values of normal and low abnormal control plasmas.

    8. The sample size for the training set

    Not applicable. This is a conventional in vitro diagnostic device, not an AI/ML device that requires a "training set" in that sense. The device's formulation and manufacturing processes are developed to achieve the intended performance characteristics.

    9. How the ground truth for the training set was established

    Not applicable, as there is no "training set" in the context of an AI/ML algorithm for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K031829

    Validate with FDA (Live)

    Date Cleared
    2003-07-31

    (48 days)

    Product Code
    Regulation Number
    864.7290
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    HemosIL Factor IX Deficient Plasma is human plasma immunodepleted of Factor IX and intended for the in vitro diagnostic quantitative determination of Factor IX activity in citrated plasma, based on the activated partial thromboplastin time (APTT) assay, on IL Coagulation and ELECTRA Systems.

    Device Description

    HemosIL Factor IX Deficient Plasma is human plasma immunodepleted of Factor IX and intended for the in vitro diagnostic quantitative determination of Factor IX activity in citrated plasma, based on the activated partial thromboplastin time (APTT) assay, on IL Coagulation and ELECTRA Systems. Abnormalities of the intrinsic pathway factors are determined by performing a modified activated partial thromboplastin time (APTT) test. Patient plasma is diluted and added to a plasma deficient in factor IX. Correction of the of the deficient plasma is proportional to the concentration (% activity) of the factor IX in the patient plasma, interpolated from a calibration curve.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the HemosIL Factor IX Deficient Plasma device, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't explicitly state quantitative acceptance criteria in terms of specific slope and correlation coefficient ranges. Instead, it implies that the device is acceptable if it demonstrates "substantial equivalence" to predicate devices through method comparison and satisfactory precision.

    Test TypeAcceptance Criteria (Implied)Reported Device Performance
    Method ComparisonDemonstrate "substantial equivalence" to predicate devices via slope and correlation coefficient. Specifically, slopes ideally close to 1.0 and correlation coefficients ideally close to 1.0.Vs. Hemoliance Factor IX Deficient Plasma (on ELECTRA IT Cast ci - E1600C):n=65, Slope = 1.0049, r = 0.9584Vs. IL Test Factor IX Deficient Plasma (on ACL Family - ACL 3000):n=76, Slope = 1.1076, r = 0.9793
    Precision (Within-Run)Acceptable reproducibility of results. (No specific quantitative criteria provided).ACL 6000: Normal 5.8% CV, Low Abnormal 8.5% CVACL 9000: Normal 3.0% CV, Low Abnormal 2.3% CVACL Futura: Normal 4.4% CV, Low Abnormal 6.0% CVELECTRA 1400C: Normal 8.4% CV, Low Abnormal 8.0% CVELECTRA 1800C: Normal 8.8% CV, Low Abnormal 8.8% CV
    Precision (Between-Run)Acceptable reproducibility of results across different runs. (No specific quantitative criteria provided).ACL 6000: Normal 2.6% CV, Low Abnormal 4.1% CVACL 9000: Normal 3.8% CV, Low Abnormal 5.9% CVACL Futura: Normal 2.9% CV, Low Abnormal 2.7% CVELECTRA 1400C: Normal 8.3% CV, Low Abnormal 7.6% CVELECTRA 1800C: Normal 5.7% CV, Low Abnormal 6.4% CV

    Study Description:

    The study to prove the device meets acceptance criteria involved two main parts: Method Comparison and Precision.

    Method Comparison Study:

    • Sample size used for the test set:
      • 65 citrated plasma samples for comparison against Hemoliance Factor IX Deficient Plasma on the E1600C.
      • 76 citrated plasma samples for comparison against IL Test Factor IX Deficient Plasma on the ACL 3000.
    • Data provenance: Not explicitly stated, but "field site studies" suggest prospective collection. Country of origin not specified.
    • Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. This study compares the new device's measurements against established predicate devices, which serve as the reference.
    • Adjudication method for the test set: Not applicable. Direct comparison of numerical results.
    • If a multi-reader multi-case (MRMC) comparative effectiveness study was done: No, this is an in-vitro diagnostic device measuring an analyte, not an imaging or diagnostic device requiring human interpretation of cases.
    • If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Yes, the device's performance is measured in a standalone fashion, comparing its output to that of predicate devices. There is no "human-in-the-loop" component in the measurement itself.
    • The type of ground truth used: The results obtained from the predicate devices (Hemoliance Factor IX Deficient Plasma and IL Test Factor IX Deficient Plasma) served as the comparative 'ground truth' or reference for demonstrating substantial equivalence.
    • The sample size for the training set: Not applicable. This is not a machine learning model requiring a training set.
    • How the ground truth for the training set was established: Not applicable.

    Precision Study (Within-Run and Between-Run):

    • Sample size used for the test set: "multiple runs (n=80)" for each instrument/control combination (40 measurements for Normal Control, 40 measurements for Low Abnormal Control on each of the 5 instruments listed, for a total of 80 measurements per instrument/control pair over the multiple runs).
    • Data provenance: Not explicitly stated, but implies laboratory testing.
    • Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Precision is about the device's own consistency.
    • Adjudication method for the test set: Not applicable.
    • If a multi-reader multi-case (MRMC) comparative effectiveness study was done: No.
    • If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Yes, this is a standalone performance metric of the device's reproducibility.
    • The type of ground truth used: Not applicable. This study assesses the consistency of the device's measurements on control samples, rather than comparing to an independent ground truth. The "control" samples themselves have established target ranges, but the study measures the device's variability around the mean.
    • The sample size for the training set: Not applicable.
    • How the ground truth for the training set was established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 4