Search Filters

Search Results

Found 134 results

510(k) Data Aggregation

    K Number
    K253852

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2026-02-06

    (66 days)

    Product Code
    Regulation Number
    878.4493
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K252743

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-12-22

    (116 days)

    Product Code
    Regulation Number
    878.4493
    Age Range
    18 - 74
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The STRATAFIX™ Spiral PDS™ Plus Knotless Tissue Control Devices are indicated for use in soft tissue approximation where the use of absorbable sutures is appropriate.

    Device Description

    STRATAFIX™ Spiral PDS™ Plus Knotless Tissue Control Devices are antibacterial monofilament, synthetic absorbable devices consisting of dyed (violet) polyester, poly(p-dioxanone), the empirical molecular formula of which is (C4H6O3)X. The devices contain IRGACARE® MP (triclosan), a broad spectrum antibacterial agent, at no more than 2360 μg/m. The pigment for the violet dye is D&C Violet No. 2. Polydioxanone polymer has been found to be nonallergenic, nonpyrogenic and elicits only a slight tissue reaction during absorption.

    The STRATAFIX™ Spiral PDS™ Plus Knotless Tissue Control Devices are available in Unidirectional and Bidirectional forms:

    • The STRATAFIX™ Spiral PDS™ Plus Unidirectional Knotless Tissue Control Devices consists of barbed suture material, armed with a surgical needle on one end and a fixation loop at the opposite end. The STRATAFIX™ Spiral PDS™ Plus Unidirectional Knotless Tissue Control Device is designed to anchor with a closed loop at one end and a unidirectional barbed section on the other end. Barbs are oriented in one direction to allow tissue approximation without the need to tie surgical knots.

    • The STRATAFIX™ Spiral PDS™ Plus Bidirectional Knotless Tissue Control Device consists of barbed suture material, armed with a surgical needle on each end. The device also contains an un-barbed center transition zone that facilitates the initiation of the device use. The barbs allow for tissue approximation without the need to tie surgical knots.

    While the formation of barbs in the STRATAFIX™ Spiral PDS™ Plus Devices reduce the tensile strength relative to non-barbed suture material of the same size, tying of knots in non-barbed suture materials also reduces their effective strength. For this reason, the strength of the STRATAFIX™ Spiral PDS™ Plus Devices can be compared to USP knot strength of non-barbed sutures. Additionally, USP designations for diameter are used to describe the STRATAFIX™ Spiral PDS™ Plus Devices suture material after barbing, except for minor variation in suture diameter with a maximum overage of 0.1 mm.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K253620

    Validate with FDA (Live)

    Device Name
    TroClose 1200
    Date Cleared
    2025-12-16

    (28 days)

    Product Code
    Regulation Number
    878.4493
    Age Range
    13 - 120
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TroClose1200™ bladeless trocar is intended for use in a variety of gynecologic, general and urologic endoscopic procedures to create and maintain a port of entry and to facilitate the delivery of absorbable sutures and anchors through soft tissues of the body during endoscopic/laparoscopic surgery. The trocar may be used with or without visualization for primary and secondary insertions.

    Device Description

    The TroClose1200™ is a Trocar system comprised of an Obturator and a Cannula including pre-loaded sutures. It includes a bladeless Obturator, designed to allow penetration and positioning at the required site within the abdomen, and several single use Cannulas, which serve as working channels for the procedure. The Cannulas are pre-loaded with two absorbable PGLA sutures, each attached to an absorbable PLGA anchor. The deployment of the anchors and sutures ("closure device") is achieved by utilizing two pushers within the Obturator as a deployment mechanism.

    The TroClose1200™, as one unit, consists of the following components:

    • An Obturator, which is the part of a trocar that allows the insertion of the trocar (after routine scalpel-made incision). An Obturator is for single patient use. For this purpose, the Company's Obturator is bladeless, similar to the predicate Obturator. In addition, the Company's Obturator has a set of 2 pushers that deploy the anchors with sutures attached from the Cannula for later closure.
    • A Cannula, which is the part of a trocar that establishes the working channel through which the surgeon introduces surgical tools while keeping the CO2 in the abdominal lumen. The Cannula is for single use and is similar to the predicate trocar Cannula. In addition, the Company's Cannula has 2 absorbable anchors attached to absorbable sutures, at 180º to each other, as the closure device, which is similar to the predicate closure device. The suture (thread) is made of absorbable PLGA and the anchors are made of absorbable PLGA and are designed to close the access port by suturing the abdominal wall's fascia.

    GORDIAN's TroClose1200™ is manufactured from routinely used medical device biocompatible materials. Like its predicates, this device is provided sterile (by EtO) and intended for single patient (Obturator) and single use (Cannula) only.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K252820

    Validate with FDA (Live)

    Date Cleared
    2025-10-03

    (29 days)

    Product Code
    Regulation Number
    878.4493
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Atramat® PGLA90 Poly(glycolide-co-L-lactide) Surgical Sutures are indicated for use in general soft tissue approximation and/or ligation, including use in ophthalmic procedures, but not for use in cardiovascular and neurological tissues.

    Device Description

    The subject devices, ATRAMAT® PGLA90 (Calypso/Odysseus), are modifications to the previously cleared Atramat® PGLA90 Poly(glycolide-co-L-lactide) Surgical Suture (K072859). The predicate device is a synthetic, absorbable, sterile surgical suture composed of a copolymer of glycolide and L-lactide, which is indicated for soft tissue approximation.

    ATRAMAT® PGLA90 (Calypso/Odysseus) devices integrate the predicate suture into a metal part with a pre-tied knot mechanism. This modification changes the operating method from traditional hand-tying to a pre-tied knot closure.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K233265

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-10-11

    (378 days)

    Product Code
    Regulation Number
    878.4493
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    OPTIME® R sutures are intended for use in general soft tissue approximation including ophthalmic procedures, but not in cardiovascular or neurological procedures.

    Device Description

    OPTIME®R sutures are intended to be used in suturing in open or laparoscopic surgery. These sutures are available with a disposable needle attachment. OPTIME®R is a sterile, multiple-stranded (multifilament), synthetic bioabsorbable thread made from a polymerization of polyglycolic acid (PGA) used in suturing when short-term support and rapid strength loss are needed. To perform suturing on soft tissue wounds involves a series of precise steps. The surgeon grasps the needle, pushes it through tissue, pulls it through and repeats for continuous suturing. Further, knots are tied, excess thread is cut, and used needles are single use and safely disposed in accordance with hospital procedure.

    AI/ML Overview

    The provided text is an FDA 510(k) Pre-Market Notification for a medical device called OPTIME® R, which is an absorbable surgical suture. The document primarily focuses on establishing substantial equivalence to a predicate device (Safil Quick Aesculap, K031286) based on technological characteristics and performance testing.

    Here's an analysis of the provided information regarding acceptance criteria and the study that proves the device meets them:

    1. A table of acceptance criteria and the reported device performance:

    The document states that the OPTIME® R suture was subjected to the requirements of the United States Pharmacopeia (U.S.P) monograph for Synthetic Absorbable Sutures. The following tests and their corresponding USP standards are mentioned:

    Acceptance Criteria / TestReported Device Performance
    Diameter (following USP 43 <861>)Meets USP performance requirements for absorbable sutures
    Tensile strength (following USP 43 <881>)Meets USP performance requirements for absorbable sutures
    Needle attachment (following USP 43 <871>)Meets USP performance requirements for absorbable sutures
    Length (following USP 43 - Absorbable surgical suture)Meets USP performance requirements for absorbable sutures
    Resorption propertiesEssentially complete at 42 days (matches predicate's absorption time)
    Biocompatibility (following ISO 10993)Biocompatibility tests performed and results indicate compliance
    Sterilization methodsGamma sterilization (matches predicate)
    Shelf-life5 years
    Ideal storageStored in original packaging, <25°C, dry, protected from light
    Packaging36 units/carton, individual units in tearable aluminum pouch, then secondary pouch

    Specific details about the exact numerical acceptance criteria (e.g., minimum tensile strength in Newtons, acceptable diameter range in mm) are not explicitly provided in this document beyond stating compliance with USP standards. The document broadly indicates that the device "meets USP performance requirements."

    2. Sample size used for the test set and the data provenance:

    • The document does not specify the sample sizes used for the Diameter, Tensile Strength, Needle Attachment, Length, or Resorption testing.
    • The data provenance is from Peters Surgical, a French company (Immeuble Aurélium, 1 course de l'Ile Seguin, Boulogne-Billancourt, Hauts-De-Seine 92100 France). The studies appear to be prospective testing conducted by the manufacturer to demonstrate compliance. There is no mention of country of origin of the data beyond the manufacturer's location.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

    • This information is not applicable in this context. The "ground truth" for a medical device like a surgical suture is established by objective engineering and biological performance standards (USP, ISO) rather than expert consensus on interpretation. There are no images or subjective assessments involved that would require expert adjudication or establishing ground truth in the way it's typically described for AI/diagnostic devices.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • This is not applicable. As mentioned above, the tests are objective measurements against predefined standards. There is no subjective assessment requiring adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • This is not applicable. OPTIME® R is an absorbable surgical suture, a physical medical device, not an AI or diagnostic software. Therefore, MRMC studies are not relevant.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • This is not applicable. OPTIME® R is a physical surgical suture, not an algorithm or AI product.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The "ground truth" for this device's performance is based on established national and international standards for absorbable sutures. Specifically:
      • United States Pharmacopeia (USP) monographs for physical and mechanical properties (Diameter, Tensile Strength, Needle Attachment, Length).
      • ISO 10993 series for biocompatibility (Cytotoxicity, Sensitization, Irritation, Material mediated pyrogenicity, Acute Systemic Toxicity, Sub-acute/sub-chronic Toxicity, Genotoxicity, Chemical characterization and toxicological risk assessment, Implantation, Degradation).
      • Resorption time is a characteristic feature verified against the predicate and established through testing.

    8. The sample size for the training set:

    • This is not applicable. There is no "training set" as this is a physical device, not a machine learning model.

    9. How the ground truth for the training set was established:

    • This is not applicable for the same reason as point 8.
    Ask a Question

    Ask a specific question about this device

    K Number
    K232372

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-07-12

    (339 days)

    Product Code
    Regulation Number
    878.4493
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ADVANTIME® synthetic absorbable surgical sutures are intended for use in general soft tissue approximation and/or ligation, but not for use cardiovascular surgery, microsurgery, ophthalmic surgery, and neurological tissue.

    Device Description

    ADVANTIME® sutures are intended to be used in suturing and/or tissue ligation in open or laparoscopic surgery. These sutures are available with a disposable needle attachment.

    ADVANTIME® is a single use, sterile, single-stranded (monofilament), synthetic bioabsorbable thread made of a poliglecaprone (a copolymer of glycolide and E- caprolactone of low tissue reactivity). Sutures are commonly used in digestive, urological, aesthetic, gynecological and intradermal surgery. To perform suturing on soft tissue wounds involves a series of precise steps. The surgeon grasps the needle, pushes it through tissue, pulls it through and repeats for continuous suturing. The same applies to double-armed needles. Further, knots are tied, excess thread is cut, and used needles are single use and safely disposed in accordance with hospital procedure.

    AI/ML Overview

    The provided document is a 510(k) summary for a surgical suture device (ADVANTIME Absorbable Suture) and does not describe an AI/ML powered device. Therefore, the questions regarding acceptance criteria, study details, and performance measures (such as sample size for test/training sets, expert ground truth, adjudication methods, MRMC studies, or standalone performance of an algorithm) related to AI/ML devices are not applicable.

    The document discusses the substantial equivalence of the ADVANTIME Absorbable Suture to a predicate device based on physical and biological performance testing for a medical device that is not an AI/ML device.

    Here's the information that is relevant from the provided text regarding this medical device:

    1. Acceptance Criteria and Reported Device Performance (Non-AI/ML Device)

    The device was subjected to the requirements of the United States Pharmacopeia (U.S.P.) monograph for Synthetic Absorbable Sutures.

    Acceptance Criteria (USP 43)Reported Device Performance
    Diameter (<861>)Meets USP requirements with minor variations (within FDA guidance allowance)
    Tensile strength (<881>)Meets USP requirements
    Needle attachment (<871>)Meets USP requirements
    Length (Absorbable surgical suture)Meets USP requirements
    Extractable color (Absorbable surgical suture)Meets USP requirements
    Resorption profilesEvaluated for each material; data provided in-vitro
    StabilityTesting completed
    Biocompatibility (ISO 10993-1)All tested aspects met (Cytotoxicity, Sensitization, Irritation, Material mediated pyrogenicity, Acute Systemic Toxicity, Sub-acute/sub-chronic/chronic toxicity, Genotoxicity, Carcinogenicity, Chemical characterization and toxicological risk assessment, Implantation, Degradation, Sterilization residue testing)

    2. Sample size used for the test set and the data provenance: Not applicable as this is not an AI/ML device study. Performance was evaluated based on physical and biological tests according to USP standards and ISO 10993-1.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable as this is not an AI/ML device. Ground truth for physical properties is defined by USP standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable as this is not an AI/ML device. Performance is objectively measured against USP standards.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable as this is not an AI/ML device.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable as this is not an AI/ML device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): For the physical characteristics, the ground truth is defined by the United States Pharmacopeia (USP) standards and test methodologies. For biological evaluation, the ground truth is established by standardized testing protocols detailed in ISO 10993-1.

    8. The sample size for the training set: Not applicable as this is not an AI/ML device.

    9. How the ground truth for the training set was established: Not applicable as this is not an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K221527

    Validate with FDA (Live)

    Device Name
    StitchKit
    Date Cleared
    2023-02-14

    (264 days)

    Product Code
    Regulation Number
    878.4493
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    StitchKit® Suture Delivery Canister facilitates minimally invasive robotic surgery by transporting suture to the operative field and removing used needles after suturing. The suture contained within the device is intended for soft tissue approximation where use of the specific absorbable or non-absorbable sutures contained within it is appropriate.

    Device Description

    StitchKit® is a sterile, single-use plastic canister that is provided pre-loaded with suturing materials with attached needles. The device facilitates endoscopic robotic surgery by introducing multiple strands of suture to the surgical site at one time and allowing for the safe retrieval of the needles. The canister is sized to be passed through a ≥12 mm trocar or passed through an incision sized for an 8 mm trocar under direct endoscopic visualization. As suturing is completed with each strand, the used needle is placed into a compartment within the canister for safekeeping until the entire canister is removed either through the ≥ 12 mm trocar or through the 8mm trocar incision in tandem with the 8mm trocar, also under direct endoscopic visualization using the attached retrieval string. It is supplied sterile.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device called StitchKit®. This document primarily focuses on demonstrating substantial equivalence to a predicate device, specifically for a minor modification related to the device's insertion and removal method (allowing use through an 8mm trocar incision in addition to a 12mm trocar).

    Crucially, this document does NOT describe the acceptance criteria or a study proving the device meets performance criteria for a complex AI/ML medical device. The device in question is a suture delivery canister, a physical surgical tool, not an AI or imaging device. Therefore, many of the requested points related to AI/ML device validation (like expert adjudication, MRMC studies, ground truth for training/test sets, effect size of human reader improvement with AI assistance, and standalone AI performance) are not applicable to this submission.

    The "clinical data" mentioned (retrospective case review n=422) is cited in the context of supporting the modification to the instructions for use, not for establishing the core performance of an AI algorithm. The performance data section refers to "functional performance test consisting of simulated surgical testing in an animal model" to demonstrate compatibility with the new insertion/removal technique, which is a physical device test, not an AI/ML study.

    Therefore, I cannot fulfill most of your request for an AI/ML device, as the provided document pertains to a physical surgical device and its minor instruction for use modification.

    However, I can extract the safety and efficacy information relevant to this device (the StitchKit® suture delivery canister) as presented in the document, acknowledging that it doesn't align with the detailed AI/ML validation questions.


    Based on the provided document for the StitchKit® Suture Delivery Canister (a non-AI device):

    1. A table of acceptance criteria and the reported device performance

    The document does not specify formal, quantitative acceptance criteria for device performance in the typical sense of accuracy, sensitivity, or specificity, as it's a physical tool. Instead, the "performance" demonstrated relates to its functional compatibility with a new surgical technique.

    Acceptance Criteria (Implied)Reported Device Performance
    Compatibility with 8 mm trocar site incision insertion/removalDemonstrated through "functional performance test consisting of simulated surgical testing in an animal model." No specific numerical performance metric provided beyond "compatible."
    Maintenance of Substantial Equivalence to PredicateConcluded based on: unchanged Indication for Use and Technological Characteristics, functional test data, clinical data (retrospective case review), and comparison to predicate device including risk analysis.
    Functionality in transporting suture and removing used needles(Presumed to be maintained from the predicate device and inherent to the device's design, as no change to core functionality is described.)

    2. Sample size used for the test set and the data provenance

    • Test Set (for the modification validity):
      • Sample Size: n=422 (Cited as a "retrospective case review clinical study").
      • Data Provenance: Conducted by "an academic medical center." The country of origin is not specified but is implicitly in the US due to FDA submission. The study was retrospective.
    • Functional Performance Test: Conducted in an "animal model." Sample size for this specific test is not provided.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not Applicable. This is a physical device, not an AI/ML diagnostic or image analysis device where "ground truth" would be established by experts interpreting medical data. The "ground truth" for the retrospective case review would likely be the surgical outcomes or successful use reported in the patient records.

    4. Adjudication method for the test set

    • Not Applicable. As above, no expert adjudication process is described or relevant for this type of device and study.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. No MRMC study was done, as this is not an AI-assisted diagnostic device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable. No standalone algorithm performance was evaluated, as this is not an algorithm-based device.

    7. The type of ground truth used

    • For the retrospective case review (n=422), the "ground truth" would be derived from clinical records and surgical outcomes related to the use of the device in actual patient cases, rather than expert interpretation of images or pathology. The document doesn't explicitly state how "success" or "performance" was quantified from these cases, but it supports the safety/effectiveness of the device, particularly with the proposed instruction modification.
    • For the animal model functional test, the "ground truth" was direct observation of the device's physical handling and compatibility with the 8mm trocar site.

    8. The sample size for the training set

    • Not Applicable. There is no "training set" as this is not an machine learning model.

    9. How the ground truth for the training set was established

    • Not Applicable. No training set was used.
    Ask a Question

    Ask a specific question about this device

    K Number
    K221744

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-11-15

    (152 days)

    Product Code
    Regulation Number
    878.4493
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    STRATAFIX™ Spiral MONOCRYL™ Plus Unidirectional Knotless Tissue Control Device is indicated for use in soft tissue approximation where the use of absorbable sutures is appropriate.

    Device Description

    The STRATAFIX™ Spiral MONOCRYL™ Plus Unidirectional Knotless Tissue Control Device is an antibacterial monofilament, synthetic absorbable single-use surgical suture device composed of a copolymer of glycolide and (epsilon) s-caprolactone. The device contains IRGACARE®: MP (triclosan), a broad-spectrum antibacterial agent, at no more than 2360 ug/m. The colorant employed is D&C Violet No.2, in <0.050% w/w concentration. Poliglecaprone 25 copolymer has been found to be nonpyrogenic and elicits only a slight tissue reaction during absorption. The subject device is intended for professional use only.

    Same as the currently marketed undyed version, the subject device consists of a unidirectional barbed suture material, armed with a surgical needle on one end and an anchoring fixation loop at the opposite end. The barbs are oriented in one direction to allow tissue approximation without the need to tie surgical knots.

    While the formation of barbs in the subject STRATAFIX™ Spiral MONOCRYL™ Plus Unidirectional Device reduces the tensile strength relative to non-barbed suture material of the same size, tying of knots in non-barbed suture materials also reduces their effective strength. For this reason, the strength of the STRATAFIX™ Spiral MONOCRYL™ Plus Unidirectional Device can be compared to USP knot strength of non-barbed sutures. The actual diameter of the non-barbed section fiber is one size greater than the designated size with a maximum overage of 0.1 mm.

    The subject device is sterilized by Ethylene Oxide, and is available in lengths of 6, 9, and 12 inches, with a USP diameter size 2-0 and 3-0 (metric sizes 3.0 and 2.0), and single armed with various needle sizes.

    AI/ML Overview

    This document describes a 510(k) premarket notification for a medical device, specifically a surgical suture. It does not contain information about an AI/ML medical device, which would typically include acceptance criteria, study details (sample size, data provenance, ground truth establishment, expert qualifications, etc.), or details of human reader studies (MRMC).

    Therefore, I cannot provide the requested information regarding acceptance criteria and the study that proves the device meets them, as the provided text pertains to a traditional medical device (surgical suture) and not an AI/ML device. The document focuses on demonstrating substantial equivalence to a predicate device through non-clinical performance testing rather than the detailed clinical validation studies common for AI/ML products.

    Ask a Question

    Ask a specific question about this device

    K Number
    K212810

    Validate with FDA (Live)

    Date Cleared
    2022-08-05

    (336 days)

    Product Code
    Regulation Number
    878.4493
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    MONOFIX PGCL, knotless wound closure device is indicated for use in soft tissue approximation where the use of absorbable suture is appropriate. MONOFIX PGCL, knotless wound closure device should not be used to approximate tissue under tension and the use of barbed suture has not been shown to be safe in the closure of hernias, laparotomy incisions, bowel anastomosis or suture lines, vascular suture lines, and vaginal cuff closure in controlled clinical or preclinical studies for minimally invasive or open surgery.

    Device Description

    MONOFIX PGCL, knotless wound closure device, comprised of Poly(glycolide-co-carprolactone)(PGA-PCL) is a synthetic absorbable monofilament undyed-suture. It is available sterile, developed by Samyang Holdings Corp., Ltd. The MONOFIX PGCL, knotless wound closure device consists of a Barb-type suture and a surgical needle, with a suture at one end and a stopper shaped end for tissue anchoring at the stopper is attached to the end of the suture and has a triangular shape with one side length of 1.8 to 3.3 mm and a thickness of 0.7 to 1.0 mm. The device is designed with small uni-directional barbs in size 0.18-0.55 mm along the length and stopper that eliminate the need to tie knots during approximation. The device is available in lengths of 15 to 60cm and diameter sizes 1 through 3-0 with needles in 1/2 circle, 5/8 circle, 1/2 curved and straight shapes attached to each end.

    AI/ML Overview

    This appears to be an FDA 510(k) clearance letter and a summary of the predicate device comparison for a surgical suture, not a diagnostic AI/ML device. Therefore, the requested information regarding acceptance criteria, study details (sample size, data provenance, expert consensus, MRMC studies, standalone performance), and ground truth establishment, which are typical for AI/ML device submissions, are not applicable in this context.

    The document describes the non-clinical tests performed to demonstrate the device's compliance with established standards for surgical sutures. These include:

    • Bench tests: To verify design specifications, focusing on:
      • USP <861> Suture Diameter
      • USP <871> Suture-Needle attachment
      • USP <881> Tensile Strength Surgical Suture
      • USP Monographs: Absorbable Surgical Suture
      • EP Monographs: Sutures, Sterile Synthetic Absorbable Monofilament
    • Sterilization validation: ISO 11135:2014, ISO 10993-7
    • Sterile barrier system testing: ASTM F 88
    • Accelerated aging stability testing: ASTM F 1980
    • Biocompatibility testing: A comprehensive suite of ISO 10993 standards (e.g., ISO 10993-5, -10, -11, -4, -12, -3, -6) and ASTM F756-17.
    • In vivo tensile strength: Performed to evaluate tensile strength retention rate.

    The "Acceptance Criteria" in this document would refer to the specific thresholds and performance requirements defined by these standards (e.g., minimum tensile strength, maximum/minimum diameter, acceptable levels of residuals, successful sterilization). The "Reported Device Performance" would be the results from the tests confirming the device met these criteria.

    Therefore, it is impossible to provide the table and study details as requested because this document does not pertain to an AI/ML device. The study outlined is a non-clinical, benchtop and biocompatibility assessment, not a clinical performance study with human readers or AI algorithms.

    Ask a Question

    Ask a specific question about this device

    K Number
    K211792

    Validate with FDA (Live)

    Device Name
    StitchKit
    Date Cleared
    2021-07-16

    (36 days)

    Product Code
    Regulation Number
    878.4493
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    StitchKit® Suture Delivery Canister facilitates minimally invasive robotic surgery by transporting suture to the operative field and removing used needles after suturing. The suture contained within the device is intended for soft tissue approximation where use of the specific absorbable or non-absorbable sutures contained within it is appropriate.

    Device Description

    StitchKit® is a suture delivery canister which facilitates endoscopic robotic surgery by introducing multiple strands of different suturing materials to the surgical site at one time and allowing for the safe retrieval of the needles. It is sized to be passed through a ≥12 mm trocar. As suturing is completed with each strand, the used needle is placed into a disposal compartment within the canister for safekeeping until the entire canister is removed through the trocar using the attached retrieval string. It is supplied sterile. Each StitchKit® contains strands of existing legally-marketed Suturing Materials, cleared for use within the device.

    AI/ML Overview

    The provided document is a 510(k) summary for the StitchKit® Suture Delivery Canister, which is a Class II medical device. This document details the device's indications for use, technological characteristics, and substantiates its substantial equivalence to predicate devices, rather than presenting a study to prove acceptance criteria in the typical sense of a human-in-the-loop AI effectiveness study.

    The device itself is a delivery system for existing sutures and not an AI-powered diagnostic or assistive device. Therefore, many of the requested criteria related to AI performance, AI effect size on human readers, expert ground truth, and training data are not applicable to this submission.

    Here's an analysis of the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the StitchKit® are based on demonstrating substantial equivalence to legally marketed predicate devices. The performance data provided focuses on the physical and chemical properties of the device and its included sutures.

    Acceptance Criteria CategorySpecific CriteriaReported Device Performance / Compliance
    Technological Characteristics- Canister is legally marketed and unchanged. - Sutures are legally marketed and have not been modified, only trimmed to length and packaged within the StitchKit® canister.- Canister is confirmed to be legally marketed and unchanged. - Sutures are confirmed to be legally marketed, not modified other than trimming and packaging for the StitchKit®.
    Performance Testing- Comparative Physico-Chemical analysis - LAL Pyrogen testing - Material Mediated Pyrogen testing - Stability Evaluations (including USP suture tensile strength, diameter, and needle attachment testing) - Comparative in-vitro simulated biodegradation testing for absorbable sutures- Testing performed to verify substantial equivalence to predicate devices. - Specific results or thresholds are not provided in this summary, but the conclusion states that the device "has been shown to be substantially equivalent to its predicate devices" based on this data.
    Indications for Use- Facilitates minimally invasive robotic surgery by transporting suture and removing used needles. - Suture intended for soft tissue approximation where use of specific absorbable/non-absorbable sutures is appropriate.- The device's design directly supports these indications. No performance metrics are given here for success rate of needle removal or ease of transport, rather it's implicit in its design that it fulfills these functions.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify a "test set" in the context of an AI study. Instead, it refers to various forms of performance testing on the device and its components. The document doesn't provide specific sample sizes for these tests (e.g., number of sutures tested for tensile strength or number of canisters tested for pyrogenicity). The provenance of the data (e.g., country of origin, retrospective/prospective) is also not specified, as this type of information is typically not included in a 510(k) summary for a physical device.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    Not applicable. As this is not an AI or diagnostic device, there is no "ground truth" established by experts in the context of image interpretation or clinical diagnosis. The performance criteria are based on objective physical/chemical tests.

    4. Adjudication Method (e.g., 2+1, 3+1, none) for the Test Set

    Not applicable. As there are no expert interpretations or subjective assessments requiring adjudication, no such method was used.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    Not applicable. The StitchKit® is a physical surgical delivery device and does not involve AI or human readers for diagnostic assistance. Therefore, an MRMC study and AI effect size are irrelevant.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    Not applicable. This device does not involve an algorithm.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for this device's performance validation is based on established engineering and material science standards (e.g., USP standards for suture properties, LAL pyrogen testing standards, physicochemical analysis). It's not clinical "ground truth" derived from patient outcomes or expert consensus on a diagnosis.

    8. The Sample Size for the Training Set

    Not applicable. There is no AI component requiring a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. There is no AI component requiring a training set and associated ground truth.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 14