510(k) Data Aggregation
(26 days)
InnoClip™ Clip Applier is indicated for patients undergoing laparoscopic surgical procedures involving occlusion of blood vessel, ducts and other tubular structures, and for radiographic marking.
The InnoClip™ Clip Applier is a disposable clip cartridge contains 10, 16 or 20 implantable titanium clips, and should be used with "Taiwan Surgical" Reusable Clip Applier Handle. The product line includes two shaft diameters, 10 mm and 5 mm. Titanium clip sizes are 5.5 × 9.0 mm and 5.5 × 11 mm for the 10 mm applier, and 4.6 × 9.1 mm for the 5 mm applier. The applier is designed to be introduced through applicable trocar sleeves, or larger sleeves with a fitted converter.
N/A
Ask a specific question about this device
(17 days)
InnoClip™ Disposable Clip Applier is indicated for patients undergoing laparoscopic surgical procedures involving occlusion of blood vessel, ducts and other tubular structures, and for radiographic marking.
The InnoClip™ Disposable Clip Applier consists of a handle piece and 10, 16 or 20 implantable titanium clips. The product line includes two shaft diameters, 10 mm and 5 mm. Titanium clip sizes are 5.5 × 9.0 mm and 5.5 × 11 mm for the 10 mm applier, and 4.6 × 9.1 mm for the 5 mm applier. The applier is designed to be introduced through applicable trocar sleeves, or larger sleeve with a fitted converter.
N/A
Ask a specific question about this device
(261 days)
The Disposable Polymer Ligation Clips are intended for use in procedures involving ligation of vessels or tissue structures. Surgeons should apply the appropriate size clip for the size of the vessel or tissue structure to be ligated such that the clip completely encompasses the vessel or tissue structure.
The Disposable polymer ligation clips includes Polymer Clips and Multiple polymer clips. The devices are sterilized by EO, single-use, non-absorbable, non-active implantable devices designed for use in general surgical procedures that require vessel or tissue ligation. The Disposable polymer ligation clips are compatible with the reusable, non-sterile Clip appliers which are made by Kangji. The clips are placed around the tissue and closed with the grip of a clip applier. These clips are made of medical grade acetyl homopolymer and supplied in different sizes and packaged in different clip quantity.
Polymer Clips are housed in a colored cartridge which is consist of base and platen. The clips shall be placed and fired in clip applier one by one. Polymer Clips are available in four sizes: M, ML, L, XL.
Multiple polymer clips (model: KJ-JZJDML, KJ-JZJDL, KJ-JZJDXL) are available in 3 sizes (ML, L, XL), housed in a colored cartridge and has 3 clips which can be placed in clip applier in one time and fired constantly totally 3 times.
Multiple polymer clips (except model: KJ-JZJDML, KJ-JZJDL, KJ-JZJDXL) are available in ML and L size, housed in a striped cartridge. The cartridge which contains 5-15 clips will be placed in clip applier. The clips will be delivered individually advance after each firing.
N/A
Ask a specific question about this device
(90 days)
The Hem-o-lok™ PurplePlus™ Large Polymer Ligating Clips are intended for use in procedures involving the ligation of vessels or tissue structures. Surgeons should apply the appropriate size clip for the size of the vessel or tissue structure to be ligated such that the clip completely encompasses the vessel or tissue structure.
Hem-o-lok™ PurplePlus™ Large Polymer Ligating Clips are single-use, non-active implantable devices designed for use in general surgical procedures that require vessel or tissue ligation. The clips are manufactured from a non-absorbable acetal polymer and are provided prepackaged in color-coded cartridges, which are provided as single-use, sterile devices.
The provided FDA 510(k) clearance letter pertains to a new version of a physical medical device (surgical clips), not an AI/Software as a Medical Device (SaMD).
Therefore, the document does not contain any of the information requested in points 2-9, as these points are specific to the evaluation and validation of AI/SaMD products, which involve concepts like:
- Acceptance criteria for AI performance (e.g., sensitivity, specificity, AUC): This is relevant for diagnostic or predictive AI systems.
- Sample sizes for test sets, data provenance, ground truth establishment, expert adjudication, MRMC studies, standalone performance, and training set details: These are critical components of validating AI model performance to ensure its accuracy, robustness, and generalizability.
Since the Hem-o-lok™ PurplePlus™ Large Polymer Ligating Clips are physical devices, their acceptance criteria and proof of efficacy are based on benchtop non-clinical performance testing and biocompatibility assessments, which are detailed in the "Performance Data" section (Section J) of the 510(k) summary.
Here's how to interpret the provided document in the context of your request:
1. A table of acceptance criteria and the reported device performance
The document lists the following non-clinical performance tests and biocompatibility assessments that were conducted. While explicit "acceptance criteria" are not numerically stated in this summary for each test (as they typically would be in a full test report), the statement "in order to ensure the device performed equivalently to the predicate" implies that the new device's performance in these tests met established safety and efficacy standards, likely mirroring or demonstrating similar performance to the predicate device.
| Acceptance Criteria (Implied) / Performance Test Category | Reported Device Performance (Summary) |
|---|---|
| Non-clinical performance testing: | Conducted following product sterilization, environmental conditioning, simulated distribution, and accelerated aging. The device ensured "performed equivalently to the predicate." |
| - Clip Latching | Met performance standards. |
| - Clip Resistance to Leakage | Met performance standards. |
| - Clip Removal | Met performance standards. |
| Biocompatibility Endpoints (per ISO 10993-1): | All assessed endpoints met standards. |
| - Cytotoxicity (ISO 10993-5) | Assessed and met. |
| - Sensitization (ISO 10993-10) | Assessed and met. |
| - Intracutaneous Reactivity (ISO 10993-10) | Assessed and met. |
| - Systemic toxicity (acute) (ISO 10993-11) | Assessed and met. |
| - Subchronic toxicity (subacute) (ISO 10993-11) | Assessed and met. |
| - Implantation (ISO 10993-6) | Assessed and met. |
| - Material Mediated Pyrogenicity (ISO 10993-11) | Assessed and met. |
| - Chronic Toxicity (ISO 10993-11) | Assessed and met. |
| - Genotoxicity (ISO 10993-3) | Assessed and met. |
| - Carcinogenicity (ISO 10993-3) | Assessed and met. |
| - Chemical Characterization (ISO 10993-18) | Assessed and met. |
For the remaining points (2-9), the information is not applicable to a physical device like surgical clips.
- 2. Sample sized used for the test set and the data provenance: Not applicable to physical clips. Testing involves benchtop models or animal studies for implantation, not human data sets.
- 3. Number of experts used to establish the ground truth... and qualifications: Not applicable. Ground truth for physical clips is established through engineering and biological testing standards, not expert annotation of data.
- 4. Adjudication method: Not applicable.
- 5. If a multi reader multi case (MRMC) comparative effectiveness study was done: Not applicable. MRMC studies are for AI/SaMD to compare human performance with and without AI assistance.
- 6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This concept applies to AI algorithms.
- 7. The type of ground truth used: For physical devices, ground truth for efficacy and safety is established through validated test methods (e.g., mechanical strength testing, leak resistance, biocompatibility standards, often with reference to predicate devices).
- 8. The sample size for the training set: Not applicable. This relates to AI model development.
- 9. How the ground truth for the training set was established: Not applicable.
In summary, the provided document details the testing performed for a physical medical device (surgical clips) to demonstrate its substantial equivalence to a predicate device, focusing on non-clinical performance and biocompatibility rather than AI/software performance metrics.
Ask a specific question about this device
(142 days)
ZIPCLIP is indicated for use in surgical procedures to occlude blood vessels.
The GEM ZIPCLIP (ZIPCLIP) is a sterile, single-use automatic microclip applier. Each device contains 15 titanium microclips. When applied on a vessel, microclips remain with the patient as permanent implants. ZIPCLIP is for prescription use only and the use environment is the operating room. The ZIPCLIP device is a pre-loaded, disposable, single-patient use mechanical assembly comprised of plastic and metal components. The applier device has (2) scissor-like handles that, when driven medially, close the distal jaws forming a closed clip the device in the open configuration. When released, the handles return to their resting state and the closed clip disengages from the applier. Simultaneously, a new clip is automatically loaded into the distal jaws for consecutive firing. When ZIPCLIP is empty a lockout clip (anodized gold in color) deploys between the jaws, preventing them from closing. The ZIPCLIP is a mechanical assembly comprised of plastic and metal components. The clips are composed exclusively of titanium (Grade 1) and are supplied sterile in a preloaded channel that is incorporated into the number of clips per applier is fifteen (15). The clips are stacked and contained internal to the device. Clips cannot be reloaded once the stack is deployed and the applier is disposed of once emptied. Note, the ZIPCLIP preloaded microclips that can be used with the ZIPCLP device.
Here is an analysis of the provided text regarding the acceptance criteria and study for the device, organized according to your request.
Please note: The provided document is a 510(k) summary for a medical device (GEM ZIPCLIP, an implantable clip) and primarily focuses on demonstrating substantial equivalence to a predicate device. It describes performance testing but does not detail acceptance criteria in the format of specific thresholds for metrics like sensitivity, specificity, accuracy, or human reader improvement, which are typical for AI/ML-based medical devices or diagnostic tools. The document describes a traditional medical device (a surgical clip) and its mechanical performance rather than the performance of a software algorithm. Therefore, many of your requested points related to AI/ML device studies (e.g., ground truth establishment for training sets, MRMC studies, standalone algorithm performance, AI assistance effect size) are not applicable to this document's content.
Acceptance Criteria and Device Performance for GEM ZIPCLIP
As per the provided 510(k) Summary for the GEM ZIPCLIP, the device is a sterile, single-use automatic microclip applier containing titanium microclips for occluding blood vessels. The performance testing described is focused on the mechanical and functional aspects of the device, rather than the diagnostic capabilities typical of AI/ML software.
1. Table of Acceptance Criteria and Reported Device Performance
The document does not provide a formal table of specific, quantitative acceptance criteria (e.g., success rates, tensile strength thresholds) with corresponding numerical reported performance values. Instead, it states the purpose of the testing and implies successful meeting of implicit performance criteria necessary for safe and effective occlusion.
| Acceptance Criteria Category (Implied) | Description from Document | Reported Device Performance (Implied) |
|---|---|---|
| Functional Performance | Reliable deployment of clips for blood vessel ligation. | "Bench performance testing to verify that ZIPCLIP reliably deploys the clips and ligates vessels was performed." |
| Occlusion Effectiveness | Safe and effective occlusion of blood vessels. | "ZIPCLIP deployment of microclips on simulated vessels and microclip ligation of simulated vessels was performed using benchtop models to verify the ZIPCLIPs ability to safely and effectively occlude blood vessels." |
| MR-Conditional Labeling | Device is appropriate for MR-Conditional labeling. | "Additional testing was performed to verify the ZIPCLIP microclip is appropriate for MR-Conditional labeling..." |
| Safety and Effectiveness | Safe and effective for the intended user population. | "...and validate that ZIPCLIP is safe and effective for the intended user population." |
| Substantial Equivalence | No new questions of safety or effectiveness compared to the predicate device. | "The design/ technological differences were found to not affectiveness and are supported by design verification and validation activities performed. The non-clinical testing performed supports the subject device for its intended use." |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Test Set: The document does not specify the exact number of test units (e.g., appliers, clips) or the number of vessels or simulations used in the bench testing. It broadly states that "bench performance testing" was performed.
- Data Provenance: The testing was "bench testing" conducted by Baxter (Synovis Micro Companies Alliance is part of Baxter, as indicated by the contact email). This implies a controlled laboratory environment.
- Country of Origin: Not explicitly stated, but typically assumed to be the country of the manufacturer or its testing facilities (likely USA, given the FDA submission).
- Retrospective or Prospective: Not applicable as this relates to clinical data. The bench testing would be considered prospective in its execution (i.e., tests were planned and executed to gather performance data).
3. Number of Experts Used to Establish Ground Truth and Qualifications
This concept is not applicable to this type of device and study. The "ground truth" for a surgical clip is its physical performance (e.g., does it close properly, does it occlude a vessel, does it stay closed). This is assessed via engineering and mechanical testing, not by expert interpretation of images or clinical outcomes in the diagnostic sense.
4. Adjudication Method for the Test Set
Not Applicable. Adjudication methods (like 2+1, 3+1) are typically used in clinical studies where multiple human readers or experts are involved in interpreting complex data (e.g., medical images) to establish a consensus "ground truth." For the mechanical testing of a surgical clip, direct measurements and observable outcomes determine performance.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No, a MRMC study was not done. This type of study is specifically designed for evaluating diagnostic devices, especially those involving AI, to measure the impact of AI assistance on human reader performance. As the GEM ZIPCLIP is a mechanical surgical device and not a diagnostic tool, an MRMC study is not applicable. Therefore, there is no effect size reported for human readers improving with AI assistance.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
Not Applicable. The GEM ZIPCLIP is a physical medical device (surgical clip applier), not a software algorithm. Therefore, standalone algorithm performance is not relevant.
7. The Type of Ground Truth Used
The "ground truth" for this device's performance testing is:
- Physical Verification/Measurement: Successful deployment, proper closure of clips, and effective occlusion on "simulated vessels" using "benchtop models."
- Material Properties Testing: Verification of titanium material (Grade 1) and MR-Conditional properties.
- Functional Observation: Observing the mechanical actions of the applier (e.g., automatic loading, lockout mechanism).
Essentially, the "ground truth" is established by direct engineering and biomechanical testing demonstrating the functional capabilities and material characteristics of the device.
8. The Sample Size for the Training Set
Not Applicable. This device is not an AI/ML algorithm, so there is no "training set."
9. How the Ground Truth for the Training Set Was Established
Not Applicable. As there is no training set for an AI/ML algorithm, this question is not relevant to the GEM ZIPCLIP device.
Ask a specific question about this device
(787 days)
The SOLID CLIP™ Single Use Clip Applier is intended for patients undergoing laparoscopic surgical procedures and can achieve dissection and ligation of blood vessels and other tubular structures for radiographic marking.
The SOLID CLIP™ Single Use Clip Applier is a sterile, single patient use device designed for an appropriately sized trocar sleeve or larger sized trocar sleeve with the use of a converter. It consists of a shaft with an outer diameter of 10 or 5 mm and length of 33 cm, a jaw for forming ligating clips, a pistol handle, an actuation trigger, a 360 degrees rotational knob, and a clip counter.
The shaft with outer diameter 10 mm was preloaded 10, 15, 20 ML (Medium-Large) or 10, 15 L (Large) titanium clips, outer diameter 5 mm shaft only preloaded 10, 15 ML (Medium-Large) titanium clips. Squeezing the trigger places a titanium clip in the jaws and closes the jaws to form the clip on the occlusion of the vessel or tubular structure.
The design of SOLID CLIP™ Single Use Clip Applier provides secured placement of the clip to the targeted vessel and enables the surgeon to apply the clips during a laparoscopic procedure without the need for withdrawing and reinserting the device each time a clip is put in place and closed.
The Medscope Biotech Co., Ltd. SOLID CLIP™ Single Use Clip Applier (K221495) demonstrated substantial equivalence to its predicate device through a series of non-clinical tests.
1. Table of Acceptance Criteria and Reported Device Performance:
The document provides a comparison of features and performance between the subject device (SOLID CLIP™ Single Use Clip Applier) and the predicate device (Endo Clip™ II Clip Applier, K143644). While specific acceptance criteria for each test result are not explicitly listed, the "Comparison" column indicates whether the subject device's performance is "Identical" or "Similar" and if it meets the criteria.
| Feature/Performance | Acceptance Criteria (Implied by Predicate) | SOLID CLIP™ Performance | Comparison/Comments |
|---|---|---|---|
| Clip Formation | Pass (Predicate also Pass) | Pass | Identical |
| Clipped Gap | Similar to 0.18 mm (Predicate) | 0.17 mm | Similar |
| Clipped Step Difference | Similar to 0.02 mm (Predicate) | 0.06 mm | Similar (All meet the criteria) |
| Clipped Pull-out Force | Similar to 258 g (Predicate) | 314 g | Identical |
| Clipped Slip Force | Similar to 933 g (Predicate) | 884 g | Similar (All meet the criteria) |
| Airtight Capability | Pass (Predicate also Pass) | Pass | Identical |
| Trigger Initial Force | Similar to 1.7 kg (Predicate) | 2.3 kg | Similar (All meet the criteria) |
2. Sample size used for the test set and the data provenance:
The document does not explicitly state the sample sizes for the specific functional performance tests (e.g., number of clips tested for formation, gap, pull-out, slip force, airtight capability, or trigger force). The tests are described as "Non-Clinical Test" and "Performance Study (Functional Testing)", indicating in-vitro testing. The data provenance is not specified in terms of country of origin, but it is clear these are non-clinical (laboratory/bench) tests, not human or animal studies for these specific performance metrics.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
Not applicable. This device is not an AI/software device that requires expert review for ground truth establishment. The performance tests are based on objective physical measurements and comparisons to a predicate device's established performance.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
Not applicable. Adjudication methods are typically relevant for studies involving human interpretation or subjective assessments, which is not the case for these non-clinical, objective functional tests.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
Not applicable. This is a medical device (clip applier), not an AI-powered diagnostic or assistive tool for human readers.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
Not applicable. This is a medical device, not a standalone algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
For the non-clinical performance tests, the "ground truth" or reference points are largely based on:
- Predicate Device Performance: The performance characteristics of the legally marketed predicate device (Endo Clip™ II Clip Applier, K143644) serve as a benchmark for comparison.
- Established Engineering/Performance Standards: Implied adherence to good engineering practice and the ability to achieve the intended function safely and effectively.
- Reference Standards: For other tests like biocompatibility, sterilization, and MR compatibility, specific ISO and ASTM standards (e.g., ISO 10993, ISO 11135, ASTM F2052) serve as the "ground truth" for compliance.
8. The sample size for the training set:
Not applicable. This is a medical device, not a machine learning model requiring a training set.
9. How the ground truth for the training set was established:
Not applicable. As this is not an AI/ML device, there is no training set or associated ground truth establishment process.
Ask a specific question about this device
(113 days)
Hem-o-lok® Ligating Clips are intended for use in procedures involving ligation of vessels or tissue structures. Surgeons should apply the appropriate size clip for the size of the vessel or tissue structure to be ligated such that the clip completely encompasses the vessel or tissue structure.
Hem-o-lok® Ligating Clips are single-use, non-active implantable devices designed for use in general surgical procedures that require vessel or tissue ligation. The clips are available in four sizes (Medium, Medium-Large, Large, Extra Large), allowing the end user to ligate a wide range of vessels and tissue structures. Hem-o-lok® Ligating Clips are manufactured from a non-absorbable acetyl polymer and are provided prepackaged in color-coded cartridges, which are provided as single-use, sterile devices.
Accessories to the Hem-o-lok® Ligating Clips include manual clip appliers and removers for use in both general open and endoscopic procedures. Both the appliers and removers are multiple use, non-sterile devices that require cleaning and sterilization prior to each use.
The provided text is a 510(k) summary for the Hem-o-lok® Ligating Clips. This document primarily focuses on establishing substantial equivalence to a predicate device based on technological characteristics and performance testing. It does not describe a study involving an AI/Machine Learning device or a diagnostic device that would require the kind of clinical performance study to prove the device meets acceptance criteria as typically understood for AI-based medical devices.
The product codes FZP, and regulation 21 CFR 878.4300, specifically describe an "Implantable Clip," which is a physical surgical device, not a software or AI-based diagnostic tool.
Therefore, the requested information regarding acceptance criteria, study details (sample size for test/training sets, experts, adjudication, MRMC studies, standalone performance, ground truth, etc.) for an AI/ML device's performance cannot be extracted from this document, as it pertains to a different type of medical device entirely.
The "Performance Data" section (J.) describes non-clinical benchtop verification testing for physical characteristics like clip latching, resistance to leakage, and clip removal, as well as biocompatibility testing according to ISO 10993 standards. This is for a physical implantable device, not an AI or diagnostic software.
In summary, this document is not relevant to the posed question about a study proving an AI/ML device meets acceptance criteria.
Ask a specific question about this device
(167 days)
angiOCCLUDE ligating clips are intended for use in procedures involving ligation of vessels or tissue structures. Surgeons should apply the appropriate size clip for the size of vessel or tissue structure to be ligated based upon their experience, judgment, and needs.
AngiOCCLUDE ligating clips are permanent implant, non-absorbable, sterile, surgical clips made from an implantable grade of Titanium and are available in six different sizes (micro, small-wide, medium, medium-large and large).
The provided text describes a 510(k) submission for the "angiOCCLUDE Ligating Clips" aiming for substantial equivalence to the Teleflex Medical Horizon Ligating Clips. This document primarily focuses on regulatory approval based on comparison to a predicate device and does not contain information about acceptance criteria or a study proving that the device meets those criteria in the way a clinical study would for an AI-powered diagnostic device.
However, I can extract the "performance data" that was summarized to show substantial equivalence, which can be thought of as criteria for demonstrating equivalence to the predicate device.
Here's the breakdown based on your request, with an important note that this is not a clinical study of device performance against specific clinical acceptance metrics, but rather a comparison study for substantial equivalence according to FDA 510(k) regulations:
1. Table of Acceptance Criteria and Reported Device Performance
For medical devices seeking 510(k) clearance, the "acceptance criteria" are generally tied to demonstrating substantial equivalence to a legally marketed predicate device. In this instance, the criteria are based on direct comparison to the Teleflex Medical Horizon Ligating Clips.
| Acceptance Criteria (for Substantial Equivalence to Predicate) | Reported Device Performance (angiOCCLUDE Ligating Clips) |
|---|---|
| Mechanical/Functional Equivalence: | |
| - Closure over vessels | Both will close over vessels. |
| - Clip closure testing (security in applier, release from applier's jaws) | Proven that the clips are secure in the applier and release from applier's jaws equivalent to Horizon. |
| - Mechanism of action (cartridge/clip design, symmetrical leg pivoting) | Same between angiOCCLUDE and Horizon; cartridge and clip design allows for an applier to easily load a clip, and clips are then removed for use. When the clip is closing, the symmetrical legs pivot in from a centered apex as clips are closed. |
| - Dimensional comparison (open and closed state within applier) | Resulted in having the same function as Horizon when clips are applied and closed. |
| - Dimensional/geometry comparisons (using precise overlays) for clip size | Proven that angiOCCLUDE and Teleflex Horizon clips are the same size across the five provided sizes (MC, SML, M, M/L, and LG). |
| Material Equivalence: | |
| - Clip material | Same clip material: implantable grade of titanium. |
| - Material comparisons (formed and sterilized material safety) | Proven that the formed and sterilized material used for the angiOCCLUDE clip is as safe and the same as the material used in Teleflex Horizon. |
| Clinical Application Equivalence: | |
| - Indications and Contraindications | Same clinical application for indications and contraindications. |
| Manufacturing/Convenience Equivalence: | |
| - Cartridge color coding | Cartridges are color coded to match applier ring colors. |
| - Sterilization processing | Uses traditional EO sterilization processing. |
| - Packaging | Uses peel open blister packs with either 6 or 24 clips each. |
| - Clip forming method | Uses 4-slides to form the clips. |
2. Sample Size Used for the Test Set and Data Provenance
The document does not specify a "test set" in the context of clinical or AI performance evaluation. Instead, the "study" is a comparison conducted by the manufacturer against the predicate device. The sample sizes for the quantitative and qualitative performance analyses are not explicitly stated (e.g., number of clips tested for closure, number of units measured for dimensions). The "data provenance" is internal testing and comparison performed by A2 Medical Systems, LLC.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
This information is not applicable and not provided in the document. The study described is a technical and material comparison, not one requiring expert human interpretation or establishment of ground truth in a diagnostic sense.
4. Adjudication Method
Not applicable as there is no human interpretation or diagnostic ground truth to adjudicate.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No MRMC study was done. This device is a surgical clip, not a diagnostic imaging device that would typically involve human readers or AI assistance in interpretation.
6. Standalone Performance Study (Algorithm Only Without Human-in-the-Loop Performance)
This device is a physical surgical clip, not an algorithm. Therefore, a standalone performance study without human-in-the-loop performance, in the context of AI/algorithms, is not applicable. The performance studies conducted were mechanical and material comparisons, as outlined in the table above.
7. Type of Ground Truth Used
The "ground truth" here is the established design, material properties, and performance characteristics of the legally marketed predicate device (Teleflex Medical Horizon Ligating Clips), against which the new device is compared. This is based on demonstrable physical properties and existing regulatory acceptance of the predicate.
8. Sample Size for the Training Set
Not applicable. This is not an AI/machine learning device.
9. How the Ground Truth for the Training Set Was Established
Not applicable. This is not an AI/machine learning device. The "ground truth" for the comparison was the predicate device's characteristics.
Ask a specific question about this device
(27 days)
The LIGACLIP Endoscopic Rotating Multiple Clip Applier is intended for use on tubular structures or vessels wherever a metal ligating clip is indicated. The tissue being ligated should be consistent with the size of the clip.
The LIGACLIP 12mm L and 10mm M/L Endoscopic Rotating Clip Appliers are sterile, single-patient use instruments designed to provide a means of ligation through surgical trocars. The instruments deliver titanium clips that individually advance after each firing. The shafts of these devices are made of a low glare material that minimizes reflective distortion. The are designed to rotate 360 degrees in either direction. The configuration of the Subject devices, Ligaclip® 12mm Land 10mm M/L Endoscopic Rotating Multiple Clip Appliers consist of a pistion knob, and a shaft. The shaft is made of a low glare material that mininizes reflective distal end of the shaft are the jaws, which form ligating dips. The force to squeeze the trigger increases when no clips remain in the device. The shaft contains a yellow clip counter indicator bar, which appears yellow when only 3 clips or fewer remain in the device.
The provided document is a 510(k) Premarket Notification from the U.S. FDA for the LIGACLIP Endoscopic Rotating Multiple Clip Applier. It aims to demonstrate substantial equivalence to previously marketed predicate devices.
The document does not describe a study involving artificial intelligence or human readers for diagnostic purposes. Instead, it refers to a medical device that physically applies clips during surgery. Therefore, many of the requested items related to AI device performance, such as MRMC studies, ground truth establishment for AI models, and training/test set sample sizes for AI, are not applicable to this submission.
However, I can extract information related to the acceptance criteria and the study performed for this specific device, to the extent that it is described.
Acceptance Criteria and Study for LIGACLIP Endoscopic Rotating Multiple Clip Applier
The device under review is an endoscopic clip applier, not an AI-powered diagnostic tool. The "performance" in this context refers to the device's mechanical functionality and usability, not diagnostic accuracy.
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria (Success Criteria) | Reported Device Performance |
|---|---|
| Usability of Instructions for Use (IFU) | The usability testing conducted to evaluate the usability and acceptance of the IFU met the success criteria. The study demonstrated that the evaluated IFU steps can be performed as intended by representative users without a pattern of use error, close calls, or use difficulty. |
| Device Functionality (Minor Component Design Change) | The verification testing conducted to evaluate the change to the device component met the success criteria. This testing demonstrated acceptable device functionality performance of the subject device and ensured it meets existing finished good specifications of the predicate device. |
2. Sample size used for the test set and the data provenance:
- Usability Study: The document mentions "representative users" but does not specify the exact sample size for the usability study.
- Verification Testing: No specific sample size is mentioned for the verification testing.
- Data Provenance: Not specified, but generally, such studies are conducted by the manufacturer. The submission is from Ethicon Endo Surgery, LLC in Guaynabo, Puerto Rico. The studies are non-clinical (usability and verification of a mechanical device). These would be prospective tests conducted for the purpose of this submission.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Usability Study: The "ground truth" for the usability study is whether "representative users" can perform the IFU steps as intended without errors. This isn't about expert medical diagnosis. The number and qualifications of these "representative users" are not specified.
- Verification Testing: The "ground truth" for verification testing is meeting existing finished good specifications. This involves engineering and quality control, not medical experts.
4. Adjudication method for the test set:
- Not applicable in the context of this device. Adjudication methods like 2+1 or 3+1 are used for expert consensus on medical image interpretations, which is not relevant here.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No. An MRMC study is not applicable as this device is a surgical instrument and not an AI-powered diagnostic tool.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not applicable. This is a mechanical surgical device, not an algorithm.
7. The type of ground truth used:
- Usability Study: The ground truth was the ability of representative users to successfully follow the Instructions for Use without significant errors or difficulties.
- Verification Testing: The ground truth was the ability of the device to meet existing finished good specifications, based on a minor design change to an internal component. This would involve objective measurements against predefined engineering specifications.
8. The sample size for the training set:
- Not applicable. This device does not use a "training set" in the context of machine learning or AI models. It's a physical medical device.
9. How the ground truth for the training set was established:
- Not applicable. As there is no training set for an AI model, there is no ground truth establishment in that sense. The device's design and functionality are based on engineering principles and prior predicate devices.
Ask a specific question about this device
(65 days)
The Weck Auto Endo5® Hem-o-lok® ML automatic endoscopic ligating clip appliers are indicated for use as delivery devices for Hem-o-lok® ML non-absorbable polymer ligating clips. These appliers are designed for use with 5/5.5mm cannulas.
Hem-o-lok® ligating clips are intended for use in procedures involving ligation of vessels or tissue structures. Surgeons should apply the appropriate size clip for the size of the vessel to or tissue structure to be ligated so that the clip completely encompasses the vessel or tissue structure.
The Auto Endo5® 5mm Automatic Endoscopic 35cm Applier is an automatic, endoscopic applier that is pre-loaded with fifteen (15) Hem-o-lok® medium-large, non-absorbable polymer ligating clips. The applier is a sterile, disposable device that is intended to be used during laparoscopic procedures when ligation of vessels or tissue structures is necessary. The Auto Endo5® 5mm Automatic Endoscopic 35cm Applier employs a trigger grip handle which is housed in a body assembly. The applier is 49cm long with a working length of 35cm. The device is designed for use with a 5/5.5mm cannula and includes a knob to allow 360° rotation of the applier shaft for clip positioning using the index finger of the gripping hand.
This document describes the Weck Auto Endo5 5mm Automatic Endoscopic 35cm Applier (AE05ML), a device intended for delivering Hem-o-lok ML non-absorbable polymer ligating clips during laparoscopic procedures. The provided information is from an FDA 510(k) summary (K230480) and does not contain details about a study evaluating AI performance. Therefore, I cannot address aspects related to AI/human reader studies, ground truth establishment, or specific acceptance criteria for an AI device.
However, I can extract information related to the performance data and acceptance criteria for the medical device itself, as described in the 510(k) summary.
1. A table of acceptance criteria and the reported device performance
The document does not provide a table with explicit acceptance criteria and corresponding performance metrics in a quantitative format as might be seen for an AI device. Instead, it states that:
| Acceptance Criterion (Implied) | Reported Device Performance |
|---|---|
| Performance after sterilization, environmental conditioning, and simulated distribution | Non-clinical performance testing was conducted following these conditions. The purpose was to "support device modifications and ensure the device performed equivalently to the predicate." (Implies successful performance upholding equivalence). |
| Usability and design validation in a simulated use environment | "Usability and design validation... was conducted to document that the user was able to operate the system as intended, and the product conformed to user needs." (Implies successful demonstration of intended operation and user satisfaction). |
| Substantial Equivalence to Predicate | The conclusion states: "Based upon the performance and comparative test results, the proposed Auto Endo5® 5mm Automatic Endoscopic 35cm Applier is substantially equivalent in performance to the predicate device cleared to market via 510(k) K152081." (Implies all necessary performance aspects were met to establish substantial equivalence). The modifications "do not introduce any new issues of safety and effectiveness." |
2. Sample size used for the test set and the data provenance
The document does not specify a "test set" sample size in the context of clinical data for human or AI performance. The performance testing mentioned ("Non-clinical performance testing") likely refers to mechanical, functional, and usability evaluations of the device itself. No explicit sample sizes for these tests are provided, nor is the provenance of any data specified (e.g., country of origin, retrospective/prospective).
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This information is not applicable as the document describes a medical device, not an AI system that requires expert-established ground truth for a test set.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This information is not applicable as the document does not describe a study involving expert adjudication for a test set.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This information is not applicable. The document describes a medical device (surgical clip applier), not an AI system, and therefore no MRMC study or AI assistance evaluation was conducted or is relevant.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
This information is not applicable as the device is not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The concept of "ground truth" as typically applied to AI performance evaluation is not relevant here. The device's performance is assessed against engineering specifications, functional requirements, and equivalence to a predicate device. This would involve objective measurements, material analysis, mechanical testing, and usability assessments, rather than an expert-determined "ground truth."
8. The sample size for the training set
This information is not applicable as the device is not an AI system.
9. How the ground truth for the training set was established
This information is not applicable as the device is not an AI system.
Ask a specific question about this device