Search Filters

Search Results

Found 69 results

510(k) Data Aggregation

    K Number
    K250550

    Validate with FDA (Live)

    Date Cleared
    2026-02-17

    (357 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    a) EG-S100XT
    This product is intended for the visualization of the upper digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, and duodenum.

    b) VS-1000
    This product is a non-sterile, reusable display device designed to display live image data from FUJIFILM single-use endoscope.

    Device Description

    A. Intended Use/Indications for Use

    a) EG-S100XT
    This product is intended for the visualization of the upper digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, and duodenum.

    b) VS-1000
    This product is a non-sterile, reusable display device designed to display live image data from FUJIFILM single-use endoscope. For information on indications for use, refer to the operation manuals of compatible FUJIFILM single-use endoscopes.

    B. Technological characteristics

    a) EG-S100XT
    This product is upper gastrointestinal endoscope, is inserted into a lumen, coelom, body cavity, or the inside of the body to provide these regions images for observation, diagnosis, and treatment. This product is inserted through the oral. The flexible tube (hereinafter "the insertion tube") follows the figure of the upper gastrointestinal tract performing a bending operation. The tip of the endoscope reaches to the duodenum. It is used for observation, diagnosis, and treatment of the esophagus, stomach, and duodenum.

    Insertion portion of this product has a mechanism (hereinafter "the bending section") which bends the tip from right to left and up and down, and the insertion tube consists of the bending section and operating portion with a knob which controls the bending section. The instrument channel which runs through the operating portion to the tip is arranged inside the insertion portion for inserting the endotherapy device. This product is a sterile, single-use device and not intended to be reprocessed.

    b) VS-1000
    This product has image processing, light source control, AE (shutter) control functions, a connector for endoscopes, and a connector for display units, and provides images for observation, diagnosis, photography, or treatment in body cavities by connecting the display unit and endoscope.

    C. Principles of Operation

    a) EG-S100XT
    The Endoscope Model EG-S100XT function on the same principles of operation as the predicate device.

    This product is lighted by LED at the tip. Its reflection forms an image on CMOS image sensor through a group of object lens placed and set in the tip of this product. An electric signal from CMOS image sensor is transmitted to the processor connected to this product. The processor converts it to a video signal and displays an image on a monitor.

    It is possible to do endoscopic fulguration, which enable a user to apply high-frequency electric current to objective regions inserting a high-frequency endotherapy device into the instrument channel of the operating portion. Since there are no exposed metal parts (conductive parts) on the endoscope surface that can be contacted in use, the endoscope can safely perform procedures using high-frequency current.

    b) VS-1000
    The product controls the light source mounted on the endoscope, receives electrical signals output from the endoscope's image sensor and outputs them as video signals to the display unit after performing image processing such as colour and image quality correction and enhancement.

    The display unit's operating console and the endoscope application software built into the display unit are used to manage patient information, adjust white balance, and provide images and patient information to network devices.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K253568

    Validate with FDA (Live)

    Date Cleared
    2026-02-06

    (81 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    These hoods are intended to be used in combination with the compatible endoscope to maintain the field of view during endoscopic procedures or examinations.

    Device Description

    These hoods are intended to be used in combination with the compatible endoscope to maintain the field of view during endoscopic procedures or examinations.

    This product is attached to the tip of our endoscope and is designed to prevent the mucosa from obstructing the surgical field of view during observation of the digestive tract or during endoscopic mucosal resection (EMR) and endoscopic submucosal dissection (ESD), thereby ensuring a clear field of view during the procedure.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K251256

    Validate with FDA (Live)

    Date Cleared
    2025-12-16

    (237 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    PENTAX Medical Video Upper GI Scopes (EG Family):
    This instrument is intended to be used with a PENTAX video processor (including light source), documentation equipment, monitor, Endotherapy Device such as a Biopsy Forceps, and other ancillary equipment for endoscopy and endoscopic surgery within the upper digestive tract including the esophagus, stomach, and duodenum.

    PENTAX Medical Video Colonoscopes (EC Family):
    This instrument is intended to be used with a PENTAX Video Processor (including light source), documentation equipment, Monitor, Endotherapy Device such as a Biopsy Forceps, and other ancillary equipment for endoscopy and endoscopic surgery within the lower digestive tract including the anus, rectum, sigmoid colon, colon and ileocecal valve.

    Device Description

    PENTAX Medical Video Upper GI Scopes (EG Family)
    These instruments are intended to be used with a PENTAX video processor (including light source), documentation equipment, Monitor, Endotherapy Device such as a Biopsy Forceps, and other ancillary equipment for endoscopy and endoscopic surgery within the upper digestive tract including the esophagus, stomach, and duodenum.

    PENTAX Medical Video Colonoscopes (EC Family)
    These instruments are intended to be used with a PENTAX video processor (including light source), documentation equipment, Monitor, Endotherapy Device such as a Biopsy Forceps, and other ancillary equipment for endoscopy and endoscopic surgery within the lower digestive tract including the anus, rectum, sigmoid colon, colon and ileocecal valve.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K251041

    Validate with FDA (Live)

    Date Cleared
    2025-10-31

    (211 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The GripTract-GI Endoscopic Tissue Manipulator (GripTract) Upper GI Model is an accessory intended to assist in positioning the distal end of an endoscope from the mucosal surface and assist with optical visualization, diagnosis, and endoscopic treatment.

    GripTract Upper GI Model is indicated for use in the upper gastrointestinal tract with any standard endoscope as follows:

    Endoscope Distal Tip Outer Diameter (mm)Endoscope Working Length (cm)GripTract Model#
    9.8 - 10.5103 - 110GT-GS100
    Device Description

    The GripTract-GI™ Endoscopic Tissue Manipulator (GripTract) Upper GI Model GT-GS100 is a disposable, non-sterile accessory intended to assist in positioning the distal end of an endoscope from the mucosal surface and assist with optical visualization, diagnosis, and endoscopic treatment.

    The GripTract Handpiece attaches to the endoscope control body just below the working channel. The soft End Cap with two integrated Fingers for tissue manipulation is placed on the distal end of the endoscope. Two Control Knobs on the Handpiece each operate a corresponding Finger, permitting the User to rotate and extend/retract the Fingers for tissue manipulation and visualization independent of the endoscope's movement or the presence of other tools in the endoscope's working channel.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K251708

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-10-10

    (129 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EvoEndo Model LE Gastroscope is intended for the visualization of the upper digestive tract in adults and pediatric patients, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, and duodenal bulb. The gastroscope is a sterile single-use device and can be inserted orally or transnasally. The 110 cm gastroscope is intended to be used in adult and pediatric populations. The 85 cm gastroscope is intended to be used in the pediatric population.

    The EvoEndo Controller is intended for use with an EvoEndo Endoscope for endoscopic diagnosis, treatment, and video observation.

    Device Description

    The EvoEndo Endoscopy System is comprised of three regulated components:

    • EvoEndo Model LE Single-Use Gastroscope (hereafter referred to as the EvoEndo Endoscope)
    • EvoEndo Reusable Video Cable (hereafter referred to as the Video Cable)
    • EvoEndo Controller (hereafter referred to as the Controller)

    The EvoEndo Endoscope is a sterile, single-use gastroscope intended to perform oral or transnasal diagnostic endoscopy in adult and pediatric patients. The EvoEndo Endoscope is ethylene oxide (EO) sterilized and is comprised of:

    • Handle
    • Umbilical Bundle that includes air, water, and suction lines, as well as the video connector
    • Endoscope shaft with HD Camera

    The Controller of the EvoEndo Endoscopy System translates the images or video captured by the camera at the distal end of the EvoEndo Endoscope to a monitor via an HDMI cable.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K243388

    Validate with FDA (Live)

    Date Cleared
    2025-06-27

    (239 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Disposable Distal Cap is used with the endoscope and installed to the distal end of endoscope to keep appropriate endoscopic field of view.

    Device Description

    Disposable Distal Cap is used with the endoscope and installed to the distal end of endoscope to keep appropriate endoscopic field of view. Disposable Distal Cap is a single structural product, a total of 3 types, and each type is divided into different specifications according to different inner diameter and the diameter of attaching endoscopic distal end. The proposed devices are EO sterilized to achieve the Sterility Assurance Level (SAL) of 10⁻⁶ and placed in a sterility maintenance package to ensure a shelf life of one year.

    AI/ML Overview

    The provided FDA 510(k) clearance letter and summary for the "Disposable Distal Cap" does not describe an AI medical device. Therefore, the requested information regarding acceptance criteria, study details, human reader performance, AI-specific ground truth, and training set information is not applicable and cannot be extracted from the given document.

    The document primarily focuses on demonstrating substantial equivalence to a predicate device (K984358) based on intended use, technological characteristics, and various performance and biocompatibility tests typical for a physical medical accessory. These tests ensure the device's physical and biological safety and performance, not the performance of an AI algorithm.

    Below is a breakdown of what can be extracted or inferred from the provided text, but it will not fulfill the requirements for an AI device as those details are absent.


    Acceptance Criteria and Device Performance (Non-AI Device)

    Since this is not an AI device, there are no AI-specific performance metrics like sensitivity, specificity, AUC, etc. The acceptance criteria revolve around physical and biological attributes. The document states that the proposed device, "Disposable Distal Cap," is substantially equivalent to the predicate device, "Disposable Distal Attachment (K984358)". This substantial equivalence forms the primary "acceptance criterion" for clearance.

    Table of Acceptance Criteria and Reported Device Performance (Based on Substantial Equivalence and Performance Data provided for a Non-AI Device):

    Item / Acceptance CriteriaReported Device Performance (Disposable Distal Cap)Remark
    Intended UseUsed with the endoscope and installed to the distal end of endoscope to keep appropriate endoscopic field of view.SE (Substantially Equivalent) to predicate's intended use ("keep the suitable depth of endoscope's view field").
    ConfigurationSingle structure, a component is a productSE to predicate.
    SterilitySupplied SterileSE to predicate.
    Sterilization MethodEO (Ethylene Oxide)SE to predicate. Sterilization validated per ISO 11135:2014+A1:2018 to achieve SAL of 10⁻⁶.
    Maximum Diameter (Specifications)11.35, 11.8, 12.4, 13.4, 14, 15, 15.7SE to predicate. Also demonstrated through Dimension Test.
    CompatibilityCompatible with different diameters of endoscopySimilar to predicate ("Compatible with different diameters of Olympus endoscopy"). Demonstrated through Endoscope Compatibility Test.
    Single UseYesSE to predicate.
    PackagingOne product housed in protective cover and sealed pouchSE to predicate. Packaging integrity tested per ASTM F1980-21 and ISO 11607-1/2:2019.
    Shelf LifeOne yearSimilar to predicate (three years). Shelf-life testing demonstrated three-year stability via accelerated aging per ASTM F1980-21 and ISO 11607-1/2:2019. The "Remark" states "Similar" but the performance data mentions demonstrating "three-year stability", indicating it meets the predicate's shelf life.
    BiocompatibilityComply with ISO 10993-1SE to predicate. Tests conducted: Cytotoxicity, Sensitization, Irritation, Acute Systemic Toxicity, Material Mediated Pyrogenicity.
    LabelingConforms to 21 CFR part 801SE to predicate.
    Functional Performance (Specific tests conducted for the device)Dimension Test: Conducted. Endoscope Compatibility Test: Conducted. Endoscopic Field of View Test: Conducted. Connection Force Test: Conducted.These tests confirm the physical and functional aspects of the cap as intended. Side-by-side comparison testing was reported against the predicate device.

    Information NOT Applicable/Present for an AI Device:

    1. Sample size used for the test set and the data provenance: Not an AI device, so no "test set" in the context of AI algorithm evaluation. The testing involved physical device samples.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not an AI device. No ground truth established by experts for algorithmic performance.
    3. Adjudication method: Not an AI device. No adjudication method for algorithmic output.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not an AI device. This type of study is irrelevant.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not an AI device.
    6. The type of ground truth used: Not an AI device. "Ground truth" for this device refers to meeting established engineering and biological safety standards (e.g., passing a cytotoxicity test, maintaining a certain connection force).
    7. The sample size for the training set: Not an AI device. No training set.
    8. How the ground truth for the training set was established: Not an AI device. No training set.

    In summary, the provided document details the regulatory clearance for a physical medical accessory, not an AI medical device. Therefore, the questions specifically related to AI device performance and study design cannot be answered from this text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K250900

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-06-16

    (83 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EvoEndo Model LE Gastroscope is intended for the visualization of the upper digestive tract in adults and pediatric patients, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, and duodenal bulb. The gastroscope is a sterile single‐use device and can be inserted orally or transnasally.

    The EvoEndo Controller is intended for use with an EvoEndo Endoscope for endoscopic diagnosis, treatment, and video observation.

    Device Description

    The EvoEndo Endoscopy System is comprised of two regulated components:

    • EvoEndo Model LE Single‐Use Gastroscope (hereafter referred to as the EvoEndo Endoscope)
    • EvoEndo Controller (hereafter referred to as the Controller)

    The EvoEndo Endoscope is a sterile, single‐use gastroscope intended to perform oral or transnasal diagnostic endoscopy in adult and pediatric patients. The EvoEndo Endoscope is ethylene oxide (EO) sterilized and is comprised of:

    • Handle
    • Umbilical Bundle that includes air, water, and suction lines, as well as the video connector
    • Endoscope shaft with HD Camera

    The Controller of the EvoEndo Endoscopy System translates the images or video captured by the camera at the distal end of the EvoEndo Endoscope to a monitor via an HDMI cable.

    AI/ML Overview

    The provided FDA 510(k) clearance letter and summary for the EvoEndo Single-Use Endoscopy System (K250900) do not describe a study involving acceptance criteria for device performance in the context of an AI/human reader study. Instead, this submission focuses on extending the indicated age range for an existing medical device (an endoscope system) and increasing its shelf life.

    The "performance testing summary" sections (1.7, 1.8, 1.9, 1.10) refer to:

    • Non-clinical functional and performance testing (e.g., visual, dimensional, tip deflection, air/water/suction flow rates, optical performance, electrical safety, EMC). These tests have specific acceptance criteria typical for medical device engineering, but they are not related to AI performance, clinical accuracy, or human reader performance.
    • Biocompatibility testing
    • Human factors/usability engineering testing
    • Literature review to support the age extension. This is a scientific review, not a performance study with acceptance criteria of the type requested in the prompt.

    Therefore, I cannot fulfill most of the requested information regarding acceptance criteria and studies typical for AI-enabled medical devices or comparative effectiveness studies with human readers. The document does not contain information about:

    • A table of acceptance criteria and reported device performance (in the context of clinical accuracy or AI performance).
    • Sample sizes used for a test set (clinical images) or the data provenance.
    • Number of experts or their qualifications to establish ground truth for a test set.
    • Adjudication methods for a test set.
    • Multi-Reader Multi-Case (MRMC) comparative effectiveness study or related effect sizes.
    • Standalone (algorithm only) performance.
    • Type of ground truth used for clinical validation.
    • Sample size for a training set or how its ground truth was established.

    The prompt's requirements are designed for AI/imaging analysis devices, which this clearance document does not pertain to. This document is for a physical endoscopy system.


    However, I can extract information related to the device and its predicate, as well as the justification for the Indication for Use extension:

    Device: EvoEndo Single-Use Endoscopy System
    Submission: K250900 (for indication extension and shelf-life extension)
    Predicate Device: EvoEndo Single-Use Endoscopy System (K213606)


    Summary of Device and Changes:

    • Subject Device: EvoEndo Single-Use Endoscopy System (K250900)
      • Components: EvoEndo Model LE Single-Use Gastroscope & EvoEndo Controller
      • Intended Use: Visualization of the upper digestive tract (esophagus, stomach, duodenal bulb) for observation, diagnosis, and endoscopic treatment in adult and pediatric patients (extended from "patients over the age of five years" in predicate). Sterile, single-use, oral or transnasal insertion.
      • Key Differences from Predicate (K213606):
        1. Indicated Age: Removal of "in patients over the age of five years."
        2. Shelf Life: Extended from 6 months to 14 months.
    • Predicate Device: EvoEndo Single-Use Endoscopy System (K213606)
      • Intended Use: Same as above, but "in patients over the age of five years."
      • Shelf Life: 6 months.

    Acceptance Criteria and "Study" for Indication Extension:

    Since this is not an AI-enabled device or a diagnostic device relying on accuracy metrics, the "acceptance criteria" here relate to demonstrating equivalence and safety for the extended indications and shelf life.

    Acceptance Criteria (Implicit for this type of submission)Reported "Device Performance" (or Justification)
    Safety and Effectiveness for Extended Age RangeClinical Literature Search: "Published literature was evaluated for the safety of endoscopy procedures in patients under 5 years as well as review anatomical considerations of using the EvoEndo Endoscopy system in the smaller patient population. A risk/benefit evaluation was also completed. The evaluation supported the removal of the specific age of 5 years and above."
    Extended Shelf Life ValidationAdditional Testing: "Additional testing done using the protocol described in K213606 was performed to extend the shelf life of the subject device to 14 months." (Implies meeting pre-defined stability/performance criteria over 14 months).
    Maintenance of Predicate Device PerformanceSubstantial Equivalence Claim: "No changes to the design of the device as compared to K213606." Previous performance testing (biocompatibility, non-clinical functional, optical, photobiological, human factors, electrical safety, EMC, software) conducted on the predicate device (K213606) is deemed applicable to the subject device and supports substantial equivalence.

    Regarding the specific questions in the prompt, based solely on the provided document:

    1. A table of acceptance criteria and the reported device performance: As explained above, the "acceptance criteria" for this specific submission relate to the justification for the extended Indication for Use and shelf life, not typical AI/diagnostic performance. The table above attempts to capture this contextually.
    2. Sample size used for the test set and the data provenance: Not applicable. No "test set" of clinical images or data was used for a performance validation in this submission type. The justification for age extension was a literature review.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for clinical data (like diagnostic accuracy) was not established or presented in this 510(k) for an endoscope.
    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI-assisted device.
    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable. This is not an algorithm.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): For the current submission, the "ground truth" for the expanded age indication was supported by a clinical literature review and anatomical considerations, along with a risk/benefit evaluation. This is not "ground truth" in the sense of a definitive diagnostic label for performance evaluation.
    8. The sample size for the training set: Not applicable. There is no AI training set mentioned.
    9. How the ground truth for the training set was established: Not applicable.

    In conclusion, the provided FDA document is for the clearance of an endoscope and its extended indications, not an AI-enabled diagnostic device. Therefore, the detailed requirements for AI/diagnostic performance studies cannot be addressed from this document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K243512

    Validate with FDA (Live)

    Date Cleared
    2025-06-10

    (210 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    FUJIFILM Endoscope Models EG-840T and EG-840TP is intended for the visualization of the upper digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, and duodenum.

    FUJIFILM Endoscope Model EG-840N is intended for the visualization of the upper digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, and duodenum. This product can be inserted orally or nasally.

    Never use this product for any other purposes.

    Device Description

    The insertion portion of the device has a mechanism (hereinafter "the bending portion") which bends the tip from right to left and up and down, and a flexible tube (hereinafter "the flexible portion") consists of the bending portion and operating portion with a knob which controls the bending portion. The forceps channel which runs through the operating portion to the tip is arranged inside the insertion portion for inserting the surgical instrument. The insertion portion of the endoscopes comes into contact with the mucosal membrane.

    The tip of the insertion portion is called the "Distal end" which contains the Imaging section, Distal cap, Objective lens, Air/water nozzle, Water jet nozzle (Except EG-840N), Instrument channel outlet, and Light guide. The bending portion is controlled by knobs on the control portion/operation section to angulate the distal end to certain angles.

    The Flexible portion refers to the long insertion area between the Bending portion and the Control portion (a part of Non-insertion portion). This portion contains light guides (glass fiber bundles), air/water channels, a forceps/suction channel, a CrOS image sensor, and cabling. The glass fiber bundles allow light to travel through the endoscope to illuminate the body cavity, thereby providing enough light to the CMOS image sensor to capture an image and display the image on a monitor. The forceps channel is used to introduce biopsy forceps and other endoscopic accessories, as well as providing suction.

    The control portion/operating section provides a grip to grasp the endoscopes and contains mechanical parts to operate the endoscopes. This section includes a Forceps inlet, which allows endoscope accessories to be introduced. The Scope connector connects the endoscopes to the light source.

    AI/ML Overview

    The provided FDA 510(k) clearance letter pertains to endoscopes, which are hardware devices, not AI/ML software. Therefore, the information requested regarding acceptance criteria, study details, and data provenance for an AI/ML device is not explicitly present in the provided text.

    The document focuses on demonstrating substantial equivalence to a predicate device through various non-clinical tests (electrical safety, biocompatibility, endoscope-specific testing, software validation, and reprocessing validation) rather than a clinical effectiveness study involving human readers or AI performance metrics.

    However, I can extract information about the device's performance specifications that were tested to prove its general functionality and safety, which can be interpreted as fulfilling certain "acceptance criteria" for a physical medical device.

    Here's a breakdown of the available information based on your request, with the caveat that it does not directly address AI/ML performance:


    1. A table of acceptance criteria and the reported device performance

    The document lists various performance specifications that the device met. It does not provide specific numerical acceptance criteria alongside reported performance values in a table format. Instead, it states that "The subject device met performance specifications in the following additional testing." This implies that the device did meet predefined internal thresholds for these parameters.

    Acceptance Criterion (Performance Specification Tested)Reported Device Performance
    Field of viewMet performance specifications
    Diameter of forceps channelMet performance specifications
    Uneven illuminationMet performance specifications
    Bending capabilityMet performance specifications
    Viewing directionMet performance specifications
    Color reproducibilityMet performance specifications
    Rate of suctionMet performance specifications
    ResolutionMet performance specifications
    Air volumeMet performance specifications
    Working lengthMet performance specifications
    LG outputMet performance specifications
    Water volumeMet performance specifications
    Electrical safetyMet specified standards (ANSI/AAMI ES 60601-1:2012, IEC 60601-1-2:2020, IEC60601-1-6:2020 and IEC 60601-2-18:2009)
    BiocompatibilityMet specified standards (ISO 10993-1:2018, ISO 10993-5:2009, and ISO 10993-10:2010), in accordance with FDA guidance.
    Endoscope specific testingMet specified standards (ISO 8600-1:2015, ISO 8600-3:2019, and ISO 8600-4:2014)
    Software specific testing (Validation)Met specified standards (IEC 62304:2015), in accordance with FDA guidance.
    Cleaning, disinfection, and sterilization instructions validationMet FDA guidance, demonstrated substantial equivalence in performance to predicate device after reprocessing.

    2. Sample size used for the test set and the data provenance

    • Sample Size: Not specified for any of the tests. The document mentions "testing" or "validation activities" without detailing the number of devices or trials involved.
    • Data Provenance: Not specified. As this is a 510(k) for a physical medical device (endoscope), the testing would typically be conducted in a laboratory setting by the manufacturer, rather than involving patient data in the context of imaging performance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. The tests performed are engineering and safety validations. There is no mention of human expert involvement for establishing "ground truth" in the context of image interpretation or diagnosis, as this is not an AI/ML diagnostic tool.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This is a concept typically used in clinical studies involving interpretation by multiple readers, which is not described here.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI-assisted device. The document describes a traditional endoscope.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. The device itself is an endoscope, not an algorithm. Software validation was conducted (IEC 62304:2015), but this relates to the software controlling the endoscope's functions, not a standalone diagnostic AI.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the listed performance specifications (field of view, resolution, etc.), the "ground truth" would be established by objective physical measurements using calibrated equipment and engineering standards. For biocompatibility, it's based on biological response to materials, and for reprocessing, it's based on sterility and decontamination efficacy. There is no "ground truth" in the diagnostic sense as there would be for an AI/ML algorithm.

    8. The sample size for the training set

    Not applicable. This is not an AI/ML device that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K242635

    Validate with FDA (Live)

    Date Cleared
    2025-05-30

    (269 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Endoscopic Distal Attachment is intended to be used in combination with compatible endoscopes to maintain the field of view during endoscopic procedures such as mucosal resection.

    Device Description

    The Endoscopic Distal Attachment is an additional device, made of TPU designed to attach to the distal end of the endoscope. The device is composed of three main portions: an attaching portion, a distal portion, and a drain portion. The attaching portion is a wider diameter opening which is used to connect the attachment to a compatible endoscope; the distal portion is the ending portion of the attachment which tapers into narrower diameter opening. The hole, which is only available for some models, forms drain portion which prevent the fluids lodging on the surface of the endoscope.

    The Endoscopic Distal Attachment can be used in combination with compatible endoscopes to maintain an appropriate endoscopic field of view.

    The Endoscopic Distal Attachment has six models, depending on different surface design (flat or incline), with or without hole and with scale or without scale. The models include models of flat face without hole, flat face with hole, inclined face without hole, inclined face with hole, flat face without hole with scale and flat face.

    AI/ML Overview

    This document is a 510(k) clearance letter for a medical device called "Endoscopic Distal Attachment (AF-D series)". Based on the provided text, the device is not an AI/Software as a Medical Device (SaMD). It is a physical accessory made of TPU that attaches to endoscopes.

    Therefore, many of the requested bullet points, such as "MRMC comparative effectiveness study," "standalone (i.e. algorithm only) performance," "sample size for training set," and "how ground truth for training set was established," are not applicable to this type of device and are not discussed in the clearance letter.

    The clearance relies heavily on bench testing and biocompatibility data to demonstrate substantial equivalence to a predicate device, rather than clinical study data or AI performance metrics.

    Here's an analysis of the provided text in relation to your questions:

    1. A table of acceptance criteria and the reported device performance

    The document does not present a formal table of acceptance criteria with quantified reported device performance in the precise way you've requested for an AI/SaMD. Instead, it lists the types of performance tests conducted and states that the device "met all design specifications" and "demonstrated safety and essential performance."

    The acceptance criteria are implicitly defined by the standards and the comparison to the predicate device. The performance is summarized as meeting these criteria.

    Acceptance Criteria CategorySpecific Tests/Standards MentionedReported Device Performance (Summary)
    BiocompatibilityISO 10993-1, ISO 10993-5 (Cytotoxicity), ISO 10993-10 (Skin Sensitization, Irritation), ISO 10993-11 (Acute Systemic Toxicity, Pyrogen Test)Evaluation conducted in accordance with FDA guidance and ISO standards. Implicitly, the device passed these tests to demonstrate biocompatibility.
    Sterilization ValidationISO 11135 (EO sterilization), ISO 11737-2 (Tests of sterility), ISO 10993-7 (EO sterilization residuals)EO sterilization validated according to applicable standards. Implicitly, the device maintained sterility and acceptable residuals.
    Shelf Life & Sterile Barrier SystemASTM F1980 (Accelerated Aging), ISO 11607-1 & -2 (Packaging), ASTM F1929 (Seal Leaks by Dye Penetration), ASTM D3078 (Flexible Packaging Leaks by Bubble Emission), DIN 58593-6 (Microbial Barrier Testing), ASTM F88/F88M (Seal Strength), ASTM D4169 (Shipping Containers)Validated according to applicable standards. Implicitly, the device maintained sterility and integrity of the sterile barrier system for the declared shelf life (3 years).
    Performance Data – BenchAppearance, Sizes, Endoscope Compatibility, Operational performance (ISO 8600-1, ISO 8600-4), Safety against tissue damage, Connection reliability, Resistance against endoscope damage in combination assembly with compatible endoscope"Performance tests were implemented on both the subject device and the predicate device to demonstrate substantial equivalence." "All performances including mechanical property and chemical property are verified to confirm the performance of the subject device is substantially equivalent to the predicate device." "Favorable test results demonstrated the subject device's safety on such issue concerns."

    2. Sample sizes used for the test set and the data provenance

    • Sample Size for Test Set: Not explicitly stated in terms of number of devices or number of tests conducted for each category. The document indicates "performance tests were implemented on both the subject device and the predicate device," implying sufficient samples were used to conduct the various bench tests. This is a physical device, so "test set" would refer to the number of physical devices and components tested.
    • Data Provenance: The testing was conducted by Alton (Shanghai) Medical Instruments Co. Ltd. (China). The data origin is from their internal testing. The document implies these are prospective tests conducted specifically for this submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • N/A. This is a physical non-AI medical device. The "ground truth" for its performance is established through adherence to engineering specifications, material properties, and standardized testing protocols (e.g., ISO, ASTM standards), not through expert consensus on images or clinical outcomes.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • N/A. Adjudication methods like 2+1 or 3+1 are typically used for establishing ground truth in clinical imaging studies or pathology where human annotators are involved. For this physical device, performance is evaluated against objective, measurable criteria defined by engineering standards and specifications.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No. This is a physical device, not an AI/SaMD. MRMC studies are not applicable.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • N/A. This is a physical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • For this physical device, the "ground truth" (or basis for verification) comes from engineering specifications, material science principles, and international consensus standards (e.g., ISO for biocompatibility and sterilization, ASTM for packaging and mechanical properties). The device's performance is compared against the established limits or requirements of these standards and against the performance of the predicate device.

    8. The sample size for the training set

    • N/A. This is a physical device, not an AI/SaMD. There is no "training set."

    9. How the ground truth for the training set was established

    • N/A. This is a physical device, not an AI/SaMD. There is no "training set" or corresponding ground truth establishment process in the AI sense.
    Ask a Question

    Ask a specific question about this device

    K Number
    K250296

    Validate with FDA (Live)

    Date Cleared
    2025-05-01

    (90 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Disposable Distal Attachment D-201 Series have been designed to be attached to the distal end of the endoscope to keep the suitable depth of endoscope's view field within the gastrointestinal tract.

    Device Description

    The Disposable Distal Attachment D-201 Series is comprised of seven (7), sterile, single-use, distal attachments of varying diameters. Each distal attachment is a short transparent plastic tube which is attached to the distal end of an endoscope to facilitate observation of tissues during endoscopic procedures. A subset of the models includes a drainage side hole that allows egress of water and blood during use. The Disposable Distal Attachment D-201 Series is to be used with compatible endoscopes.

    AI/ML Overview

    The provided FDA 510(k) clearance letter and summary for the "Disposable Distal Attachment D-201 Series" does not describe a study involving an AI/Medical Imaging device that requires ground truth, human readers, or a test set as described in the prompt.

    Instead, this document describes a traditional medical device (a physical accessory for an endoscope) and the performance testing conducted to demonstrate its substantial equivalence to a predicate device. The performance data focuses on biocompatibility, sterilization, packaging, shelf-life, mechanical testing, and human factors. There is no mention of an algorithm, image analysis, or clinical performance in terms of diagnostic accuracy or reader improvement.

    Therefore, I cannot fulfill the request to describe the acceptance criteria and the study that proves the device meets the acceptance criteria using the provided information, as the requested elements (ground truth, expert consensus, MRMC studies, training/test sets for an algorithm, etc.) are not applicable to this type of device and its clearance process.

    The document states: "Clinical data is not required to demonstrate substantial equivalence." This further confirms that no studies involving analysis of images or AI performance were conducted or necessary for this device's clearance.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 7