510(k) Data Aggregation

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NIBP cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and may be reused. It is available in neonate, infant, child and adult sizes. The cuff is not designed, sold, or intended for use except as indicated.

    Device Description

    The subject device is an accessory used in conjunction with noninvasive blood pressure measurement systems. It is available in neonatal, infant, child and adult sizes.

    The subject device includes reusable non-invasive Blood Pressure Cuff(NIBP) cuff, all the reusable blood pressure cuff has same structure, which contains Cuff with bladder and Air Hose. Air hose has single tube and double tube.

    The NIBP cuff is reusable device, Model BCS-X12 series and BCD-X12 series are made of PU synthetic leather (Cuff) and TPU (Air Hose), thereinto, the PU synthetic leather (Cuff) is the material used to contacting with the patient. Model BCS-X22 series and BCD-X22 series are made of Nylon (Cuff) and TPU (Air Hose), thereinto, and the Nylon (Cuff) is the material used to contacting with the patient's intact skin.

    The NIBP cuff comes in 28 different sizes to fit different population of different arm sizes. Of these, 14 models(BCS-X12 series and BCS-X22 series) of 28 use a single tube of air hose, while the other 14 models(BCD-X12 series and BCD-X22 series) use a double tube of air hose, which is used for different noninvasive blood pressure measurement system.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K253964

    Validate with FDA (Live)

    Date Cleared
    2026-01-09

    (29 days)

    Product Code
    Regulation Number
    870.1120
    Age Range
    0 - 0.08
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Unimed disposable blood pressure cuff (Models U1681S-C51N to U1685S-C51N) is an accessory used in conjunction with noninvasive blood pressure measurement systems. It is non-sterile and for single-patient use. The cuff is available in neonatal sizes (3-15 cm limb circumference). The cuff is not designed, sold, or intended for use except as indicated.

    Device Description

    The Unimed Disposable Neonatal NIBP Cuff (Model U1681S-C51N) is a single-patient-use accessory for non-invasive blood pressure (NIBP) measurement. The cuff is designed for neonatal limb circumferences ranging from 3 cm to 6 cm. The cuff is intended to be used with compatible blood pressure monitors employing oscillometry. It is non-sterile, and is disposed of after single-patient use to reduce cross-contamination risks.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K251745

    Validate with FDA (Live)

    Date Cleared
    2025-12-16

    (193 days)

    Product Code
    Regulation Number
    870.1120
    Age Range
    0 - 150
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Welch Allyn FlexiPort Blood Pressure Cuffs are intended to be used for patient health screening assessment to determine patient's systolic/diastolic blood pressure when used in conjunction with non-automated and automated sphygmomanometers.

    Welch Allyn FlexiPort Blood Pressure Cuffs are noninvasive blood pressure cuffs intended for use in conjunction with non-automated and automated sphygmomanometers to determine blood pressure in pediatric through adult patients.

    FlexiPort Blood Pressure Cuffs are intended to be used by a clinically trained medical professional.

    Device Description

    Welch Allyn FlexiPort Blood Pressure Cuffs product line consists of Reuse, Soft & Vinyl Blood Pressure Cuffs used for patient health screening assessment to determine the patient's systolic/diastolic blood pressure when used in conjunction with non-automated and automated sphygmomanometers. This product line was previously cleared under 510(k) premarket notification K070060 on December 18, 2007.

    Welch Allyn FlexiPort Blood Pressure Cuffs are designed with the FlexiPort, a port connector which can rotate 360 degrees to reduce tube kinking and a standardized connector for quick exchange of tubing and fittings that interface with manual or electronic sphygmomanometers. The FlexiPort feature allows the interface tubing (either one- or two-tube) to remain with the blood pressure measurement device, while the cuff can be easily removed and moved from one location to another. The cuff can then be used universally within a facility on multiple devices.

    The following are the types of FlexiPort Blood Pressure Cuffs available with the FlexiPort feature.
    a. Welch Allyn FlexiPort Reusable Blood Pressure Cuffs (also referred as Reuse Blood Pressure Cuffs).
    b. Welch Allyn FlexiPort Soft Blood Pressure Cuffs (also referred as Soft Blood Pressure Cuffs).
    c. Welch Allyn FlexiPort Vinyl Blood Pressure Cuffs (also referred as Vinyl Blood Pressure Cuffs).

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K250765

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-09-03

    (174 days)

    Product Code
    Regulation Number
    870.1120
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The cuff is compatible with non-invasive blood pressure monitors to measure blood pressure.

    It is intended for use by healthcare professionals (prescription use) and by laypersons in home settings (over-the-counter use).

    The cuff is designed for adult patients. It is non-sterile and reusable.

    Device Description

    The primary components of the AViTA blood pressure cuff include an inflatable bladder and an outer fabric layer with a hook-and-loop fastening system. The cuff is designed to wrap securely around the patient's upper arm, allowing for accurate measurement of blood pressure when used with AViTA blood pressure monitors.

    The AViTA blood pressure cuffs are available in five different sizes to accommodate a range of arm circumferences:

    ModelLimb Range (Unit: cm)
    6NRUI-K22-42 cm
    6NRLI-K33-42 cm
    6NRXI-K42-52 cm
    6NRSI-K15-22 cm
    6NRMI-K22-33 cm

    During operation, the cuff inflates and deflates to occlude and release arterial blood flow, utilizing the oscillometric method to detect systolic and diastolic pressure. The cuff is constructed with durable materials to withstand repeated inflation cycles and ensure long-term performance.

    The AViTA blood pressure cuff is designed for compatibility with AViTA blood pressure monitors, ensuring optimal performance and measurement accuracy.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K251045

    Validate with FDA (Live)

    Date Cleared
    2025-07-15

    (103 days)

    Product Code
    Regulation Number
    870.1120
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Unimed disposable blood pressure cuff (Models U1682S-C51N to U1685S-C51N) is an accessory used in conjunction with noninvasive blood pressure measurement systems. It is non-sterile and for single-patient use. The cuff is available in neonatal sizes (4–15 cm limb circumference). The cuff is not designed, sold, or intended for use except as indicated.

    Device Description

    The Unimed Disposable Neonatal NIBP Cuffs (Models U1682S-C51N, U1683S-C51N, U1684S-C51N, U1685S-C51N) are single-patient-use accessories for non-invasive blood pressure (NIBP) measurement. Each cuff is designed for neonatal limb circumferences ranging from 4 cm to 15 cm (Neonate 2 through Neonate 5) as follows.

    • U1682S-C51N: 4 cm - 8 cm.
    • U1683S-C51N: 6 cm - 11 cm.
    • U1684S-C51N: 7 cm - 13 cm.
    • U1685S-C51N: 8 cm - 15 cm.
      The cuffs are intended to be used with compatible blood pressure monitors employing oscillometry. They are non-sterile, and are disposed of after single-patient use to reduce cross-contamination risks.
    AI/ML Overview

    This 510(k) clearance letter is for a Disposable Neonatal NIBP Cuff. It does not describe an AI/ML powered medical device, nor does it detail a study involving AI/ML.

    Therefore, many of the requested criteria related to AI/ML device validation (e.g., sample size for test/training sets, data provenance, number of experts for ground truth, MRMC study, standalone performance, etc.) are not applicable to this document.

    The document primarily focuses on establishing substantial equivalence to a predicate device through non-clinical testing and material biocompatibility.

    Here's the information that can be extracted from the provided text, primarily related to the non-clinical testing performed to meet acceptance criteria for a physical medical device:


    Acceptance Criteria and Device Performance for Disposable Neonatal NIBP Cuff

    The acceptance criteria for this device are primarily based on established international standards for non-invasive blood pressure cuffs and biocompatibility. The performance is demonstrated through non-clinical testing, confirming adherence to these standards.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Standard Compliance)Reported Device Performance (Test Results)
    ISO 81060-1:2007 (Non-invasive sphygmomanometers – Requirements and test methods for non-automated measurement type)The device met all design specifications based on testing conducted in accordance with this standard.
    ISO 80601-2-30:2018 (Medical electrical equipment — Part 2-30: Particular requirements for basic safety and essential performance of automated non-invasive sphygmomanometers)The device met all design specifications based on testing conducted in accordance with this standard.
    Biocompatibility - ISO 10993-5 (Tests for in vitro cytotoxicity)The materials used in the proposed cuffs demonstrated biocompatibility, indicating suitability for intended use.
    Biocompatibility - ISO 10993-10 (Tests for skin sensitization)The materials used in the proposed cuffs demonstrated biocompatibility, indicating suitability for intended use.
    Biocompatibility - ISO 10993-23 (Tests for irritation)The materials used in the proposed cuffs demonstrated biocompatibility, indicating suitability for intended use.
    Pressure LimitsThe device operates within the accepted range of 0-300 mmHg, which is the same as the predicate device.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not specified in the provided document. The non-clinical testing likely involved a representative sample of each cuff model (U1682S-C51N, U1683S-C51N, U1684S-C51N, U1685S-C51N).
    • Data Provenance: The document states "Non-clinical tests were conducted..." but does not specify the country of origin where the testing was performed. Given the manufacturer's location (Shenzhen, China), it's plausible the testing was conducted there or at a certified lab supporting the manufacturer. The testing is described as non-clinical, implying it was laboratory-based product verification and validation, not retrospective or prospective human clinical data collection.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This is not applicable. This document describes the clearance of a physical medical device (NIBP cuff) via substantial equivalence, not an AI/ML algorithm. Therefore, there is no mention of "ground truth" derived from expert consensus on medical images or data. The "ground truth" for this device would be the defined specifications and performance requirements established by the relevant ISO standards.

    4. Adjudication Method for the Test Set

    This is not applicable. Adjudication methods like 2+1 or 3+1 are typically used in clinical trial settings, especially for AI/ML performance evaluation where multiple human readers might disagree on findings. For non-clinical device testing like this, adherence to predefined test protocols and specifications is the primary method of evaluation.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a MRMC comparative effectiveness study was not done. This type of study is relevant for evaluating the impact of AI assistance on human reader performance, which is not applicable to a non-AI medical device like a blood pressure cuff.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    No, a standalone performance evaluation of an algorithm was not done. This is not an AI/ML device.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's performance is established by compliance with recognized consensus standards (ISO 81060-1, ISO 80601-2-30, and ISO 10993 series for biocompatibility). The device's performance metrics are evaluated against the specific requirements and test methods outlined in these standards.

    8. The Sample Size for the Training Set

    This is not applicable. There is no training set mentioned as this is not an AI/ML device.

    9. How the Ground Truth for the Training Set Was Established

    This is not applicable. There is no training set mentioned as this is not an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K242623

    Validate with FDA (Live)

    Date Cleared
    2024-10-31

    (58 days)

    Product Code
    Regulation Number
    870.1120
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The reusable blood pressure cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and may be reused. It is available in neonate, infant, child and adult sizes. The cuff is not designed, sold, or intended for use except as indicated.

    The disposable blood pressure cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and for single-patient use. It is available in neonate size. The cuff is not designed, sold, or intended for use except as indicated.

    Device Description

    The device comprises tubing attached to an inelastic sleeve with an integrated inflatable bladder that is wrapped around the patient's limb and secured by hook and loop closure. The device tubing is connected to a non-invasive blood pressure measurement system. It is available in neonatal, infant, child and adult sizes. The subject device includes disposable blood pressure cuff and reusable blood pressure cuff and contains Cuff with/without bladder and single/dual tubes Air Hose.

    AI/ML Overview

    The provided document is a 510(k) summary for a blood pressure cuff and does not contain detailed information about acceptance criteria, reported device performance for specific metrics, or any study proving such performance. The document explicitly states:

    "No clinical study is included in this submission"

    Therefore, I cannot provide the requested information regarding acceptance criteria and a study proving the device meets them because such information is not present in the provided text. The document focuses on demonstrating substantial equivalence to a predicate device primarily through non-clinical data and compliance with voluntary standards for design specifications.

    However, I can extract the information that is present:

    8. Non-clinical data

    • Performance testing: Non-clinical tests were conducted to verify that the proposed devices met all design specifications.
    • Compliance with standards: The test results demonstrated that the proposed device complies with ISO 81060-1, Non-Invasive Sphygmomanometers - Part 1: Requirements and Test Methods for Non-Automated Measurement Type, First Edition 2007.

    Based on this, here's what can be stated about the available information:

    1. A table of acceptance criteria and the reported device performance: Not provided. The document mentions compliance with ISO 81060-1, which would contain performance requirements for non-automated sphygmomanometers (blood pressure cuffs). However, the specific acceptance criteria used for the device and the reported device performance against those criteria are not detailed.
    2. Sample sized used for the test set and the data provenance: Not provided. The document only mentions "non-clinical tests" and "performance testing" but does not give details about sample sizes or data provenance.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable, as no clinical study or expert-based ground truth establishment is mentioned. The testing focused on technical specifications and standards compliance.
    4. Adjudication method for the test set: Not applicable.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done: No. The document explicitly states "No clinical study is included in this submission."
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This device is a blood pressure cuff, which is a physical accessory, not an algorithm.
    7. The type of ground truth used: Not applicable for performance evaluation as no clinical data or comparative performance study is provided. For demonstrating substantial equivalence, the ground truth indirectly relies on the established performance and safety of the predicate device and the standards.
    8. The sample size for the training set: Not applicable. This is a non-AI/ML device.
    9. How the ground truth for the training set was established: Not applicable. This is a non-AI/ML device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K240165

    Validate with FDA (Live)

    Date Cleared
    2024-03-20

    (58 days)

    Product Code
    Regulation Number
    870.1120
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is intended to be used by medical professionals or in the home for the measurement of systolic and diastolic pressure on new borns, infants, children, young adults. The device is intended to be manually attached to a patient and manually inflated along with a manual method for detecting Korotkoff sounds.

    Device Description

    The ANEROID SPHYGMOMANOMETER and ANEROID SPHYGMOMANOMETER WITH STETHOSCOPE are manual blood pressure monitor used to measure blood presure.It consists of a cuff,a inflation bulb,a pressure gauge, an air valve and a stethoscope(optional).

    Cuff: The cuff is made of a durable fabric and is wrapped around the upper arm of the patient. It inflates and deflates during the measurement process. The material of outside layer is nylon cloth.

    Inflation Bulb: The inflation bulb is a rubber bulb that connects to the cuff and is used to manually inflate the cuff with air. The material is Neoprene or Silicon rubber or Nature latex or PVC.

    Pressure Gauge: The pressure gauge is a circular dial with a needle that indicates the pressure within the cuff. It is calibrated in millimeters of mercury (mmHg) and shows the systolic and diastolic blood pressure readings. The material is Aluminum or ABS.

    Air Valve: The valve is a control mechanism that allows for the release of air from the cuff once the measurement is complete.

    Stethoscope: Although not a part of the device itself, a stethoscope is typically used in conjunction with an Aneroid Sphygmomanometer to listen for the blood flow sounds, known as Korotkoff sounds, in the brachial artery during the measurement. Three model are single head Stethoscope(Model: TY-S01),Dual Head Stethoscope(Model: TY-S05) and Spraque Rappaport Stethoscope(Model: TY-S04) individually.These stethoscopes are listed in FDA database.

    AI/ML Overview

    The provided submission describes an Aneroid Sphygmomanometer. Here's an analysis of the acceptance criteria and the study proving the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance (Reference Standard)
    Accuracy± 3 mmHg (ISO 81060-1:2007)
    BiocompatibilityComplies with ISO 10993-5:2009, ISO 10993-10:2021, ISO 10993-23:2021
    Safety & PerformanceComplies with ISO 81060-1:2007

    The document explicitly states that the "Accuracy" of the device is ± 3 mmHg, which is also the performance reported. For "Biocompatibility" and "Safety & Performance", compliance with the specified ISO standards is the stated performance.

    Note: The submission describes an aneroid sphygmomanometer, which is a mechanical device. The concept of an "AI device" with human-in-the-loop performance or standalone performance, sample sizes for test/training sets, adjudication methods, and expert qualifications for ground truth establishment, are not applicable to this type of device. The provided text is a 510(k) summary for a traditional medical device, not an AI/ML-powered device.

    Therefore, the following points are answered with "Not Applicable" (N/A) due to the nature of the device.

    2. Sample size used for the test set and the data provenance
    N/A - This device is a mechanical blood pressure monitor, not an AI/ML-powered device that uses a "test set" in the context of data. The performance testing refers to compliance with a standard (ISO 81060-1), which involves testing the device itself, not an algorithm's performance on a dataset.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
    N/A - See explanation above.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
    N/A - See explanation above.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
    N/A - This is a mechanical device, not an AI-assisted one.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
    N/A - This is a mechanical device, not an AI algorithm.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)
    N/A - The ground truth for this device's accuracy would be established by direct comparison to a reference standard, as defined by ISO 81060-1, which likely involves a calibrated pressure standard.

    8. The sample size for the training set
    N/A - This is a mechanical device, not an AI/ML-powered device that requires a "training set."

    9. How the ground truth for the training set was established
    N/A - See explanation above.

    Ask a Question

    Ask a specific question about this device

    K Number
    K233276

    Validate with FDA (Live)

    Date Cleared
    2023-12-21

    (83 days)

    Product Code
    Regulation Number
    870.1120
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Reusable Blood Pressure Cuff is an accessory used in conjunction with non-invasive blood pressure measurement systems. The cuff is non-sterile and reusable. It is available in adult size.

    Device Description

    The proposed device is an accessory used in conjunction with non-invasive blood pressure measurement systems, and wrap around the patient's limb. The cuff is non-sterile and reusable. The model difference of the proposed device includes bladder type, tube, color, joint model and size.

    AI/ML Overview

    The provided text is a 510(k) summary for a Reusable Blood Pressure Cuff. It focuses on demonstrating substantial equivalence to predicate devices rather than proving a device's performance against specific acceptance criteria for an AI/ML medical device.

    Therefore, the document does not contain the information required to describe the acceptance criteria and the study that proves an AI/ML device meets those criteria.

    Specifically, the document does not provide:

    • A table of acceptance criteria and reported device performance: The document lists standards the device complies with (IEC 80601-2-30, ISO 81060-1, ISO 10993-5, ISO 10993-10), but these are general safety and performance standards for a mechanical device, not AI/ML acceptance criteria like sensitivity, specificity, or AUC. No specific performance metrics are reported against quantitative acceptance criteria.
    • Sample size and data provenance for a test set: No mention of a test set, data provenance, or sample size for evaluating AI/ML algorithm performance.
    • Number and qualification of experts for ground truth: Not applicable, as this is not an AI/ML device requiring expert-established ground truth.
    • Adjudication method: Not applicable.
    • MRMC comparative effectiveness study: Not applicable.
    • Standalone (algorithm only) performance: Not applicable.
    • Type of ground truth used: Not applicable.
    • Sample size for the training set: Not applicable.
    • How ground truth for the training set was established: Not applicable.

    The document states:

    • "6. Non-Clinical Test Conclusion" and "6.1 Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards: IEC 80601-2-30: 2018..., ISO 81060-1:2007..., ISO 10993-5: 2009..., ISO 10993-10: 2010..."
    • "6.2 Human Factors / Usability Testing Usability testing is not required per the risk profile of this product."
    • "7. Clinical Test Conclusion No clinical study is included in this submission."

    This unambiguously indicates that the submission is for a physical medical device (a blood pressure cuff), assessed through non-clinical bench testing for compliance with manufacturing and biocompatibility standards, and not an AI/ML driven diagnostic or assistive device that would require the specified performance evaluations.

    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The disposable blood pressure cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and for single-patient use. It is available in neonatal, pediatric and adult sizes. The cuff is not designed, sold, or intended for use except as indicated.

    The reusable blood pressure cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and may be reused. It is available in neonatal, pediatric and adult sizes. The cuff is not designed, sold, or intended for use except as indicated.

    Device Description

    The device comprises tubing attached to an inelastic sleeve with an integrated inflatable bladder that is wrapped around the patient's limb and secured by hook and loop closure. The device tubing is connected to a non-invasive blood pressure measurement system.

    The subject device is categorized into two types of models according to its usage.

    Reusable Blood Pressure Cuff models:

    1201B1,12L2B1,1202B1,12L3B1,1204B1,1205B1,1206B1,1207B1-1#,1207B1-2#,1207B1-

    3#,1207B1-4#,1207B15#,1201W2,12L2W2,1202W2,12L3W2,1203W2,

    1204W2,1205W2,1206W2 and 1207W2

    Disposable Blood Pressure Cuff models:

    1201D1,12L2D1,1202D1,12L3D1,1204D1,1205D1,1204D1,1207D1-1#,1207D1-2#,1207D1-

    3#,1207D1-4#,1207D1-5#

    Connector models: PLINMA-03-127-003, PLINMA-03-T, PLINMA-03-Z3

    AI/ML Overview

    The provided text is a 510(k) summary for a Blood Pressure Cuff. It focuses on demonstrating substantial equivalence to predicate devices rather than proving the device meets specific acceptance criteria through a clinical study with detailed performance metrics. The information about a study proving device performance is limited.

    Here's an analysis based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document primarily refers to meeting ISO standards and does not present a table of specific acceptance criteria (e.g., specific accuracy ranges for blood pressure measurements) versus reported device performance in a clinical setting. It states that the device was "tested according to ISO 81060-1 Non-invasive sphygmomanometers - Part 1: Requirements and test methods for non-automated measurement type". However, it does not provide the results of this testing in a quantifiable form or against specific, pre-defined acceptance criteria.

    2. Sample Size Used for the Test Set and Data Provenance:

    The document states, "The subject and predicate devices are exactly the same. Since the blood pressure cuffs are identical, no further clinical testing is necessary." This implies that no separate test set with human subjects was used for this 510(k) submission to evaluate the performance of the device in terms of blood pressure measurement accuracy. The data provenance is assumed to be from the predicate device's testing, but details are not provided here.

    3. Number of Experts Used to Establish Ground Truth and Qualifications:

    Since no separate clinical performance testing was conducted for this submission, no experts were used to establish ground truth from newly collected data for this specific device.

    4. Adjudication Method for the Test Set:

    Not applicable, as no dedicated clinical performance test set was used for this submission.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    Not applicable, as this is a blood pressure cuff and not an AI-powered diagnostic imaging device. The document does not describe any human reader involvement or AI assistance.

    6. Standalone Performance Study:

    A standalone performance study focused on technical compliance with ISO 81060-1 was mentioned ("The subject device has been tested according to ISO 81060-1 Non-invasive sphygmomanometers - Part 1: Requirements and test methods for non-automated measurement type"). However, the results of this testing, which would quantify its performance against the ISO standard's requirements, are not provided in this summary. The summary only attests that the device "met the requirements."

    7. Type of Ground Truth Used:

    For the "performance data" mentioned, the ground truth would be the reference measurements and methods defined within the ISO 81060-1 standard for non-invasive sphygmomanometers. For the biocompatibility testing, the ground truth would be the established scientific methods and criteria for evaluating cytotoxicity, sensitization, and irritation.

    8. Sample Size for the Training Set:

    Not applicable, as this is a medical device (blood pressure cuff) and not an AI/ML algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable.

    Summary of Acceptance Criteria and Study (based on available information):

    Acceptance Criteria (Implied)Reported Device Performance (Summary)
    Compliance with ISO 81060-1 for non-automated measurement type"The subject device has been tested according to ISO 81060-1... and met the requirements."
    Biocompatibility (Cytotoxicity, Sensitization, Irritation)"The test results show that the proposed device has no issue of cytotoxicity, sensitization, or skin irritation."
    Functional equivalence to predicate devicesStated that "The subject and predicate devices are exactly the same."
    Limb Circumference Range (as per ISO 81060-1 and labeling)"the size of limb circumference of subject device met the requirements of ISO 81060-1, and the limb circumference was marked in the labeling."
    Tube Configuration (meeting ISO 81060-1)"The subject device met the requirements of ISO 81060-1."

    Study Details:

    • Test Set Sample Size: Not applicable for clinical performance testing, as it was deemed unnecessary due to the device being "exactly the same" as predicates. For technical testing to ISO 81060-1, the sample size is implicitly defined by the standard's requirements for testing a physical device.
    • Data Provenance: Not specified for the ISO 81060-1 technical testing, but typically conducted in a laboratory setting.
    • Experts for Ground Truth: Not applicable for clinical performance as no new human data was collected. Ground truth for technical testing is established by standard operating procedures and reference equipment.
    • Adjudication Method: Not applicable for clinical performance.
    • MRMC Study: No.
    • Standalone Study: Yes, a technical standalone study was performed to ISO 81060-1 and biocompatibility standards, with results indicating compliance, although specific performance metrics are not listed in this summary.
    • Type of Ground Truth: ISO standard requirements and established biocompatibility testing protocols.
    • Training Set Sample Size: Not applicable (not an AI/ML device).
    • Training Set Ground Truth Establishment: Not applicable.

    The document primarily relies on demonstrating substantial equivalence to predicate devices and adherence to relevant ISO standards, rather than presenting a detailed clinical performance study specific to this submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K231961

    Validate with FDA (Live)

    Device Name
    NIBPCuff
    Date Cleared
    2023-08-30

    (58 days)

    Product Code
    Regulation Number
    870.1120
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The reusable blood pressure cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and may be reused. It is available in neonate, infant, child and adult sizes. The cuff is not designed, sold, or intended for use except as indicated.

    The disposable blood pressure cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and for single-patient use. It is available in neonate, infant, child and adult sizes. The cuff is not designed, sold, or intended for use except as indicated.

    Device Description

    The proposed device is an accessory used in conjunction with noninvasive blood pressure measurement systems. It is available in neonatal, infant, child and adult sizes.

    The proposed device includes disposable blood pressure cuff and reusable blood pressure cuff. Both disposable and reusable blood pressure cuff have the same structure, which contains Cuff with bladder and single-tube Air Hose.

    AI/ML Overview

    This document is a 510(k) Premarket Notification from the FDA regarding a non-invasive blood pressure (NIBP) cuff. It focuses on demonstrating substantial equivalence to a predicate device rather than providing a detailed study report of the device's accuracy or efficacy as one might find for a more complex AI-driven medical device.

    Therefore, the requested information, particularly concerning "acceptance criteria" and "study that proves the device meets the acceptance criteria" in the context of an AI/algorithm-driven device (like an effect size of human readers improving with AI, or standalone algorithm performance), is largely not applicable to this document. This document primarily describes the physical characteristics, intended use, and manufacturing standards of a blood pressure cuff.

    However, I can extract the relevant information where available, focusing on performance data related to blood pressure cuffs.

    Here's an analysis based on the provided text:

    Key Takeaway: This 510(k) submission primarily relies on physical and material testing and comparison to a predicate device to demonstrate substantial equivalence, not a clinical study proving its performance against specific physiological accuracy criteria in humans, nor an AI-driven system's performance.


    1. A table of acceptance criteria and the reported device performance

    The document doesn't present a table of acceptance criteria and corresponding reported device performance in the same way one would for an AI model's diagnostic accuracy. Instead, it refers to compliance with established standards for blood pressure cuffs.

    • Acceptance Criteria (as implied by standards compliance):

      • Max. Leakage: < 4mm Hg/min (for both reusable and disposable cuffs, identical to predicate)
      • Biocompatibility: Compliance with ISO 10993-5 (Cytotoxicity) and ISO 10993-10 (Irritation and Skin Sensitization) for patient-contacting materials.
      • Non-Invasive Sphygmomanometers Performance: Compliance with ISO 81060-1 (Requirements and Test Methods for Non-Automated Measurement Type).
    • Reported Device Performance:

      • Max. Leakage: Met the < 4mm Hg/min criteria (implied by "The test results demonstrated that the proposed device complies with the following standards").
      • Biocompatibility: Cytotoxicity, Sensitization, and Irritation tests were conducted and the device "Comply with ISO 10993-5; Comply with ISO 10993-10."
      • ISO 81060-1 Compliance: The device met "all design specifications as was Substantially Equivalent (SE) to the predicate device" and complied with ISO 81060-1. This standard outlines accuracy requirements for NIBP devices (e.g., mean difference and standard deviation of blood pressure measurements compared to a reference). While the specific numerical results (mean difference, standard deviation) are not provided in this summary, compliance with this standard indicates these performance metrics were met.

    2. Sample size used for the test set and the data provenance

    • Sample Size: The document does not specify human clinical trial sample sizes. The testing mentioned is primarily non-clinical (materials, leakage, mechanical). For ISO 81060-1, compliant NIBP devices typically undergo clinical validation with a certain number of subjects (e.g., ANSI/AAMI/ISO 81060-2 specifies requirements, often involving over 85 subjects for accuracy validation). However, this 510(k) summary likely relies on the predicate device's clinical validation and the subject device's adherence to relevant parts of the standard (ISO 81060-1, which focuses on non-automated devices and test methods, rather than the clinical validation part 2). The document notes "Non-clinical tests were conducted to verify that the proposed devices met all design specifications as was Substantially Equivalent".
    • Data Provenance: Not specified. Based on the manufacturer and consultant location (Shenzhen, China), it's likely the non-clinical tests were conducted in China. The "retrospective or prospective" nature is not applicable here as no clinical trial data is presented.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not Applicable. This is not an AI/algorithm-driven device requiring expert adjudication for ground truth (e.g., radiologists reviewing images). The acceptance criteria relate to physical properties and compliance with medical device standards for blood pressure cuffs.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable. See point 3.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. This is a hardware medical device (blood pressure cuff), not an AI-driven diagnostic or assistive tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable. This is a hardware medical device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • The "ground truth" for this device's performance is established by compliance with engineering and biocompatibility standards (e.g., ISO 81060-1 for performance, ISO 10993 for biocompatibility). For accuracy, the "ground truth" is typically a reference blood pressure measurement device.

    8. The sample size for the training set

    • Not Applicable. This device is not an AI/machine learning model and therefore does not have a "training set."

    9. How the ground truth for the training set was established

    • Not Applicable. See point 8.

    In summary, this 510(k) notification focuses on the manufacturing, physical characteristics, and material safety (biocompatibility) of blood pressure cuffs, demonstrating their substantial equivalence to a legally marketed predicate device by complying with relevant international standards. The criteria and studies are for a traditional medical device, not an AI-powered system that would involve clinical performance metrics and ground truth establishment involving human interpretation or large datasets.