The PreemptiveAI Clinical SDK is intended for noninvasive spot measurement of pulse rate. The measurement is based on video recording of the subject's fingertip using a smartphone camera. It is designed for use when the subject is still. It is indicated for individuals 18 years of age or older not requiring critical care or continuous vital signs monitoring. It is not intended for use in patients with known or suspected heart arrhythmias.
The PreemptiveAI Clinical SDK is a Software as a Medical Device (SaMD) that utilizes a smartphone to noninvasively measure pulse rate (PR). Designed for integration into third-party mobile applications, the SDK operates on Android and iOS devices, enabling spot-check pulse rate measurements.
The PreemptiveAI Clinical SDK captures a 30-second fingertip video with the smartphone's back camera. The software extracts a photoplethysmography (PPG) waveform, partitions that 30-second signal into overlapping 10-second windows, and applies an autocorrelation-based signal quality check. A proprietary deep learning algorithm then calculates pulse rate for every window that passes the quality screen. The value shown to the user is the average of all passing windows, and if every window fails the screen, no rate is reported and a "Low Signal Quality" prompt is displayed.
The SDK outputs a single pulse-rate value; it is not intended for automated analysis or arrhythmia detection. The PreemptiveAI Clinical SDK is not intended for use with alarm systems.
Here's a breakdown of the acceptance criteria and study details based on the provided FDA 510(k) clearance letter for the PreemptiveAI Clinical SDK:
Acceptance Criteria and Device Performance
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Pulse Rate Range: 50-125 bpm | Bench testing confirmed accuracy across the claimed 50–125 bpm operational range. |
| Pulse Rate Performance (Error Level): +/- 3 bpm ARMS (Average Root Mean Square) | The clinical study demonstrated PreemptiveAI Clinical SDK measures pulse rate within ± 3 BPM ARMS. |
Study Details
1. Sample Size for Test Set and Data Provenance:
- Sample Size: 111 participants.
- Data Provenance: The document states participants were "recruited from a general population," implying the data is from a human clinical study. The country of origin is not explicitly mentioned but is likely the USA given the FDA clearance. The study appears to be prospective as it involved recruiting participants and obtaining new measurements.
2. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:
- This information is not provided in the document. The document states "The pulse rate obtained by the PreemptiveAI Clinical SDK was compared to the heart rate obtained by the reference device." It doesn't specify how the ground truth from the reference device was established or if experts were involved in its interpretation for this study.
3. Adjudication Method for the Test Set:
- This information is not provided in the document. The reference to comparing the SDK's pulse rate to a "reference device" suggests a direct comparison rather than a human adjudication process for the AI's output against independent expert assessments.
4. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done:
- No, an MRMC comparative effectiveness study was not explicitly mentioned or implied. The study focused on the standalone performance of the AI device against a "reference device," not on how human readers improve with AI assistance.
- Effect Size of Human Improvement with AI: Not applicable, as an MRMC study was not conducted.
5. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done:
- Yes, a standalone performance study was conducted. The document states, "The pulse rate obtained by the PreemptiveAI Clinical SDK was compared to the heart rate obtained by the reference device," indicating that the SDK's output was directly evaluated without human intervention in the pulse rate determination process.
6. The Type of Ground Truth Used:
- The ground truth was established using a "reference device." While the specific type of reference device is not detailed (e.g., medical-grade ECG, oximeter, etc.), it served as the comparator for the SDK's pulse rate measurements. It is a form of outcomes data in the sense that it provides a measured, objective physiological parameter.
7. The Sample Size for the Training Set:
- This information is not provided in the document. The document describes the clinical study and bench testing, but not the training data specifics for the proprietary deep learning algorithm.
8. How the Ground Truth for the Training Set Was Established:
- This information is not provided in the document.
Summary of Missing Information:
The provided document, being an FDA clearance letter, focuses primarily on the successful outcome and key aspects of the regulatory submission. It does not delve into the detailed methodology of the AI model's development, including:
- Specifics about the reference device used for ground truth in the clinical study.
- The number and qualifications of experts involved in ground truth establishment (if any, beyond the reference device).
- Details on the training set (size, ground truth methods).
- An MRMC study was not stated, so no human-in-the-loop performance change was evaluated.
Ask a specific question about this device
(314 days)
HEARTBEAM AIMIGo SYSTEM:
The HeartBeam AIMIGo™ System is a portable non-invasive recorder intended to record, store, and transfer a patient's 3-Lead (in three-directions) electrocardiogram (ECG) acquired from 5 electrodes. The device is intended to be used by adult patients in either a clinical setting or at home. The device does not conduct cardiac analysis and can be used with an ECG Viewer software system for manual interpretation of non-life-threatening arrhythmias by a physician or healthcare professional.
HEARTBEAM 12-L ECG SYNTHESIS SOFTWARE:
The HeartBeam 12-L ECG Synthesis Software synthesizes a 12-L ECG from the HeartBeam AIMIGo 3-Leads (in three-directions) recording device, producing a visual 12-L ECG representation that is similar, but not identical, to the same leads of a standard diagnostic 12-L ECG.
The synthesized 12-L ECG output is solely intended for manual assessment of normal sinus rhythm and the following non-life-threatening arrhythmias: sinus arrhythmia, sinus tachycardia, sinus bradycardia, atrial premature complexes, atrial fibrillation, and ventricular premature complex. The synthesized 12-L ECG output is not intended for the assessment of any other arrhythmia or conditions (including but not limited to: other atrial arrhythmias, ventricular arrhythmias, hypertrophy, conduction disorders, myocardial infarction or ischemia, pacemaker functions, localization of arrhythmia foci, ECG wave abnormalities, and/or any other disorder). The software does not conduct cardiac analysis and is not intended to replace a standard 12-L ECG. The 12-L ECG Synthesis Software is intended for adult use only.
The HeartBeam AIMIGo™ System captures a 3-lead (in three-directions) electrocardiogram (ECG) recording of patients who have been prescribed a HeartBeam AIMIGo™ System for recording an ECG remotely. The HeartBeam AIMIGo™ System records and transmits the 3-lead ECG signal which can be displayed as rhythm strips on a compatible ECG viewer. The HeartBeam AIMIGo™ System does not provide any analysis of the ECG data nor provide any recommendation toward a clinical diagnosis and is not intended to be used with automated ECG analysis systems. It reports a series of 3-lead ECG rhythm strips for manual interpretation. The hardware platform is designed to be functional and effective with any compatible software designed for the purpose of displaying ECG waveforms for clinical review. The HeartBeam AIMIGo™ System is provided by prescription only.
The system allows the patient (or their caregiver or healthcare provider) the capability to perform the following:
- Record a 3-lead (in three-directions) ECG with the recording device and accompanying mobile Patient Application
- Transmit the recorded 3-lead ECG signal to the cloud server, which can then be accessed by a clinician via the Clinician Portal for review and manual interpretation.
This submission includes a modification to the HeartBeam System involving the addition of a new software feature, the HeartBeam 12-L Synthesis Software. Like the HeartBeam AIMIGo system, the patient-activated recording of a 3-L ECG is obtained and the HeartBeam 12-L Synthesis Software can synthesize a 12-L ECG signal from the 3-lead ECG signal recorded from the HeartBeam AIMIGo Device of the cleared system. The synthesized 12-L ECG output signal is intended to be an adjunct set of ECG information that complements but does not replace a standard 12-L ECG, to aid in the manual assessment of non-life-threatening arrhythmias.
The synthesized 12-L ECG signal has not been studied or evaluated for safe and effective use for other cardiac conditions outside of manual non-life-threatening arrhythmias and is contraindicated for the assessment of any urgent, serious, or life-threatening conditions or arrhythmias.
The new software works with the HeartBeam system, which retains all functionalities and performance characteristics of the previously cleared system. No changes to the hardware, intended use, or fundamental technological characteristics were made to incorporate the new software. Additionally, the synthesis software works with the standard 12-L ECG system, the EDAN SE-1515 DX-12. The EDAN SE-1515 DX-12 systems provide a patient's compatible standard 12-L ECG signal, which is used as one of the inputs for the synthesis software to create the personalized transformation matrix for each patient.
N/A
Ask a specific question about this device
(394 days)
The MEMO Patch M is intended to record, transfer and store single-channel electrocardiogram via USB transmission to PC. The monitor is intended for use by healthcare professionals, patients with known or suspected heart conditions and health-conscious individuals. The ECG data is intended to supplement other patient data and is not intended for automated analysis. The device has not been tested and it is not intended for pediatric use.
The MEMO Patch M is a continuous electrocardiogram (ECG) recording device to record, store, and transfer single channel ECGs and is designed for a fully retrospective review of data that was recorded over the specified wear period. The MEMO Patch M has Bluetooth module to verify ECG signals with mobile device and to authenticate the activation of MEMO Patch M. The device is in the form of an ECG patch that records ECG signals. As well the patch connects with a cradle and transmits ECG data via USB communication to PC. Healthcare professional actives the ECG patch, connects the patch with the cradle to the PC, and accesses raw ECG data on the ECG dataloader viewer. The MEMO Patch M includes a battery powered electronic unit that is used with off-the-shelf (OTS) disposable medical grade gel electrodes for long term monitoring. The adhesive electrodes should be replaced by the user every 24 hours or when it no longer adheres to skin. The device is only intended for manual interpretation, and does not provide any automated ECG analysis. The intended clinical use of the ECG is to display the waveforms of the P, QRS, and T waves of the measured electrocardiogram. The intended clinical use of the ECG waveform to be limited to the discrimination of normal sinus rhythm from easily identifiable, non-lethal arrhythmias. The device is not intended for additional analysis or the detection of specific cardiac conditions, such as ischemia, myocardial infarction, and ldft ventricular hypertrophy, etc. The Ambulatory ECG Monitor is intended to capture and continuous electrocardiogram information for long-term monitoring. The device records continuously for the entire wear period, up to 8 days. The MEMO Patch M is a prescription use device and the recorded ECG data is intended to be used with patient-triggered events. When a patient presses the event marking button during a symptom, this event is recorded and displayed on the PDF report of raw ECG data. However, there is no ability to detect specific cardiac events on the PDF report.
N/A
Ask a specific question about this device
(330 days)
The Hexoskin Medical System (HMS) is intended for use at home, or at a healthcare facility, on the order of a licensed medical professional, to record, display and store the following physiological data: a) 3 leads of electrocardiogram; b) heart rate; c) respiratory rate; and d) activity. The device is intended for use when the clinician decides to evaluate the physiologic signals of adult patients as an aid to diagnosis and treatment. The HMS ECG recordings are indicated for the manual assessment of cardiac rhythm disturbances. The device does not produce alarms and is not intended for active patient monitoring (real-time). The device is not intended for use as life supporting equipment on high-risk patients such as critical care patients. The device is not intended for use in the presence of a pacemaker.
The HMS consists of four main components - three physical components and one software user interface (UI) component. The physical components are the HMS Recorder, the HMS Shirt, and the Cable. The Recorder is a rechargeable data acquisition module that connects to the Shirt. The Shirt is available in a wide range of configurations, catering to diverse patient body types and builds. The Cable is a USB to micro-USB cable that connects to the Recorder for data transfer and charging. The UI component is a PC-based, stand-alone software that displays the recordings stored on the Recorder.
This device is not an apnea monitor. Users should not rely on its respiration monitoring for detecting cessation of breathing. The Hexoskin Medical System does not provide any automatic analysis or diagnosis.
N/A
Ask a specific question about this device
(264 days)
The VSH101 is designed to record, transmit, and store single channel electrocardiogram (ECG) data via Bluetooth communication to compatible Bluetooth enabled devices. The device is intended for use by healthcare professionals, individuals with known or suspected cardiac conditions, and health-conscious users. The ECG data serves as a supplementary source of patient information and is not intended for automated analysis.
VitalSigns 1-Lead Holter is an ambulatory and Bluetooth-based wireless communication ECG measurement solution designed to allow users to record, store, transmit, and display single-channel ECG data.
VitalSigns 1-Lead Holter consists of the following three components:
- VS Electrode Patch "VSP101"
- ECG Recorder Host "VSH101"
- iOS APP "VSHealth"
When the ECG Recorder Host "VSH101" is fully charged and connected wirelessly via Bluetooth to the iOS app "VSHealth", it can instantly obtain the user's ECG data. VSHealth assists in transmitting, displaying, recording, and storing ECG data.
After the "VSH101" is fully charged, the device, through the VSHealth app, allows users to continuously use and record ECG data for 24 hours in their daily routines, whether at home or in a work environment.
The provided FDA 510(k) clearance letter and summary for the VitalSigns 1-Lead Holter (VSH101) primarily focus on demonstrating substantial equivalence to a predicate device through comparison of features and adherence to various safety and performance standards. However, it does not include explicit acceptance criteria tables or detailed study results for specific device performance metrics that would be typically found in a clinical or performance validation report.
The document states that "software validation, performance test and usability test have been completed to demonstrate that the differences between these parameters would not impact the safety and effectiveness of the subject device. The subject device has also undergone all safety and performance tests, and the results complied with the test requirements." It also mentions "Performance testing - IEC 60601-2-47 test" and "Disposable ECG electrode test."
Given the information provided in the 510(k) summary, I will infer the acceptance criteria based on the mentioned compliance standards and the general nature of an ECG Holter device, and then describe what is stated about the testing.
Inferred Acceptance Criteria and Reported Device Performance
The 510(k) summary for the VitalSigns 1-Lead Holter (VSH101) does not provide a specific table of acceptance criteria with corresponding performance metrics. Instead, it relies on demonstrating compliance with relevant international standards and equivalence to a predicate device. Based on the mentioned standards (e.g., IEC 60601-2-47 for ambulatory ECG recorders, ANSI/AAMI EC12 for disposable ECG electrodes), the general acceptance criteria for such a device would relate to the accuracy, signal quality, and reliability of ECG signal acquisition.
Here's a table based on the inferred acceptance criteria from the context of ECG device standards and the information stated in the document:
| Acceptance Criteria Category | Specific Metric (Inferred) | Acceptance Threshold (Inferred from Standards & Equivalence) | Reported Device Performance (as stated in document) |
|---|---|---|---|
| ECG Signal Quality | Artifact/Noise Levels, Baseline Wander, Frequency Response | Compliance with IEC 60601-2-47 (e.g., specific limits for noise, linearity, gain accuracy) | "The patch provides stable conductivity, low impedance" (for electrodes). "Performance is equivalent to IEC 60601-2-47 for all devices." "The results complied with the test requirements." |
| ECG Electrode Performance | Impedance, Biocompatibility, Adhesion | Compliance with ANSI/AAMI EC12 (e.g., biocompatibility (cytotoxicity, irritation, sensitization), acceptable impedance range, adhesion properties over time) | "The VS Electrode Patch has been tested in accordance with ANSI/AAMI EC12, confirming compliance with established safety and performance standards." "The patch provides stable conductivity, low impedance, and biocompatibility." |
| Data Acquisition & Storage | Continuous Recording Duration | At least 24 hours (explicitly stated function) | "Allows users to continuously use and record ECG data for 24 hours." "Data storage: 24 hours." |
| Device Functionality & Reliability | Wireless Communication Reliability, Battery Life, Software Functionality | Reliable Bluetooth communication, adequate battery life for intended use, proper software operation (no critical errors) | "Connected wirelessly via Bluetooth." "Allows users to continuously use and record ECG data for 24 hours." "Software validation... has been completed." |
| Safety | Electrical Safety, Electromagnetic Compatibility (EMC), Biocompatibility | Compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, ISO 10993-1 | "Meet all requirements for all design, biocompatibility, electrical, EMC safety and cybersecurity protection." "VSH101 has been tested and complies with the requirements of Clause 8.5.5.2 – Energy reduction test." "Pass all testing." |
Study Proving Device Meets Acceptance Criteria
The provided 510(k) summary describes that comprehensive testing was conducted, primarily focusing on compliance with recognized consensus standards and demonstrating substantial equivalence to a predicate device, rather than a single large-scale clinical/performance study with detailed outcome metrics.
Here's an analysis based on the information provided:
-
A table of acceptance criteria and the reported device performance:
- As detailed above, a direct table is not provided in the 510(k) summary. The acceptance criteria are inferred from the listed compliance standards (e.g., IEC 60601-2-47, ANSI/AAMI EC12). The reported performance is generally stated as "complied with the test requirements," "met its pre-defined criteria," or "confirmed compliance."
-
Sample sizes used for the test set and the data provenance:
- Sample Size: Not explicitly stated for performance testing. For compliance with standards like IEC 60601-2-47 or ANSI/AAMI EC12, testing typically involves a sufficient number of device units or electrodes to statistically demonstrate compliance with the standard's requirements (e.g., a batch of electrodes, multiple device samples). The document does not specify if patient data was used for performance testing beyond what is implied by the "24 hours" recording capability.
- Data Provenance (Country of Origin, Retrospective/Prospective): Not specified. Standard compliance testing is typically done in a lab setting rather than involving patient data in a "retrospective" or "prospective" clinical study design for 510(k) submissions unless a specific clinical performance claim needs to be proven. The focus here is on engineering verification and validation against standards.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- This information is not provided in the 510(k) summary. The nature of the testing described (compliance with standards like IEC 60601-2-47 for signal quality, ANSI/AAMI EC12 for electrodes, and general electrical safety) suggests controlled laboratory testing and engineering validation, which typically does not involve human expert adjudication of ECG readings as a "ground truth" for device performance in this context. The device's indication for use explicitly states, "The ECG data serves as a supplementary source of patient information and is not intended for automated analysis," implying that human interpretation remains key. Therefore, ground truth establishment by experts for automated analysis is not applicable.
-
Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- None is explicitly described. Based on the content, there was no clinical study described that would require a ground truth panel or adjudication method for ECG event interpretation. The testing focuses on technical performance compliance.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No MRMC study was done or described. The device is a "1-Lead Holter" that records ECG data; it does not perform "automated analysis" or include AI assistance. Its purpose is to provide raw ECG data for supplementation of patient information, not for diagnostic interpretation by an algorithm. Therefore, an MRMC study assessing AI assistance is not relevant to this device's claims or function.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Based on the indication for use, "The ECG data serves as a supplementary source of patient information and is not intended for automated analysis," there is no algorithm for diagnostic interpretation. Therefore, no standalone algorithm performance study was done or described.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- As noted above, for the type of device and the presented summary, complex ground truth derived from expert consensus, pathology, or outcomes data is not applicable for the performance testing described. The "ground truth" for compliance testing is largely defined by the technical specifications and requirements within the IEC/AAMI standards themselves (e.g., known input signals for signal quality, chemical assays for biocompatibility).
-
The sample size for the training set:
- Not applicable / Not provided. The device records raw ECG data and does not perform automated analysis using a trained algorithm. Therefore, no "training set" for AI/ML model development is mentioned or required for this type of device based on its intended use.
-
How the ground truth for the training set was established:
- Not applicable. As there is no AI/ML model for automated analysis that requires a training set, the establishment of ground truth for such a set is not discussed.
Ask a specific question about this device
(35 days)
The QT ECG System is intended to acquire, record and process an electrocardiographic signal so that it can be transmitted digitally via Bluetooth technology to a cell-phone or mobile device, then to a remote location. The QT ECG System is indicated for use on infants, children and adults. It is designed to be operated by adults in the home, or by healthcare workers in non-acute care clinical facilities (such as nursing homes, skilled nursing facilities), and in professional healthcare facilities (such as clinics, hospitals, and ambulances) to record and transmit a 12- lead ECG and rhythm strip in near real-time to enable review at a physician's office, hospital or other medical receiving centers.
The QT ECG system is a non-defibrillator-proof, hand-held, cordless 12-lead electrocardiograph (ECG) system with Bluetooth connectivity. The QT ECG system consists of 5 maior components:
- The QT ECG Recorder-Compact device that records 12-lead, resting electrocardiograms, then transmits the recorded data to a mobile device (smartphone, tablet, etc.) paired via Bluetooth. A Bluetooth-enabled mobile device (not included) is needed to operate the QT ECG Recorder, and to send the recorded rhythm strip to a cardiologist or licensed physician for review.
- The QT ECG Electrode Strip-Disposable, patented electrodes that are prepositioned on a self-adhesive strip.
- The QT ECG App - Software that lets the user uses their mobile device to operate the QT ECG recorder, then send the recorded data via cloud to a certified medical professional for review.
- Analysis The analysis module provides ECG measurement from the collected data. It does not make any interpretation of the intervals provided based on factors such as heart rate, QRS duration, etc.
- Web Service - The web service provides an interface for communication. The recorded ECG data is saved temporarily on the mobile device until it is transferred via the Internet to the cloud server. The QT ECG System does not have monitoring capabilities and does not have diagnostic alarm function. The QT ECG System is indicated for use on infants, children and adults. It is designed to be operated by adults in the home, or by healthcare workers in non-acute care clinical facilities (such as nursing homes, skilled nursing facilities), and in professional healthcare facilities (such as clinics, hospitals, and ambulances) to record and transmit a 12- lead ECG and rhythm strip in near real-time to enable review at a physician's office, hospital or other medical receiving centers.
The provided text describes the QT ECG (QTERD100) device and its submission for FDA clearance. The submission, K233521, is a Special 510(k) comparing the device to a previously cleared version (K220795). The changes primarily relate to an expanded indication for use to include Emergency Medical Services (EMS) environments and updated environmental operating conditions and supported mobile devices.
Here's an analysis of the acceptance criteria and study data based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The document presents a comparison between the subject device (K233521) and the original device (K220795), focusing on the differences. The "Acceptance Criteria" column explicitly states the criteria for the tests conducted to validate these differences.
| Device Change | Acceptance Criteria | Summary of Results |
|---|---|---|
| Indications for Use (addition of EMS) | Pass the test according to IEC 60601-1-12. | Pass. The risk is controlled. The document: Risk Management Plan, Risk Control Report, Risk Management Summary, Report of IEC 60601-1-12 |
| Storage and transport temperature | (Implicit: device functions within new temperature range) | (No explicit "Pass" for this row, but overall "Pass" for the change) |
| Operating Temperature | (Implicit: device functions within new temperature range) | (No explicit "Pass" for this row, but overall "Pass" for the change) |
| Storage and Transport Atmospheric Pressure | (Implicit: device functions within new atmospheric pressure range) | (No explicit "Pass" for this row, but overall "Pass" for the change) |
| Operating Atmospheric Pressure | (Implicit: device functions within new atmospheric pressure range) | (No explicit "Pass" for this row, but overall "Pass" for the change) |
| Ingress Protection Rating (IP33 vs IP22) | (Implicit: device meets new IP33 rating) | (No explicit "Pass" for this row, but the "Summary of Results" column isn't provided for this specific change. However, the overall conclusion states all safety and performance tests were passed.) |
| Package | N/A (change in packaging format, no performance impact expected) | N/A |
| Quick guide | N/A (simplification for risk reduction, no performance impact expected) | N/A |
| Mobile device (updated supported models/OS) | Pass criteria according to test reports: Unit Test, Integration Test, System Test | Pass. The risk is controlled. The document: Risk_Management_Plan, Risk_Control_Report, Risk_Management_Summary, Software Test reports: Unit Test, Integration Test, System Test |
2. Sample Size Used for the Test Set and Data Provenance
The provided document does not specify sample sizes for any of the tests mentioned (e.g., IEC 60601-1-12, software tests).
The data provenance is also not explicitly stated in terms of country of origin or whether it was retrospective or prospective. The document mentions tests being conducted "according to IEC 60601-1-12," which implies standard compliance testing, but doesn't detail the origin of the test data.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
The document does not provide information regarding the use of experts to establish a ground truth for the test set, nor does it mention their number or qualifications. The tests conducted appear to be primarily engineering and safety compliance tests rather than clinical studies requiring expert ground truth for interpretation of ECG data.
4. Adjudication Method for the Test Set
Since there's no mention of experts establishing a ground truth or interpreting data, there's no information about an adjudication method.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
The document does not mention a multi-reader multi-case (MRMC) comparative effectiveness study. The device is described as acquiring, recording, and transmitting ECG signals for review by a physician, but there's no study presented that evaluates the performance of human readers with or without AI assistance, or an effect size. The "Analysis module" for the device provides ECG measurements but "does not make any interpretation."
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
The document describes the device's "Analysis module" which "provides ECG measurement from the collected data. It does not make any interpretation of the intervals provided based on factors such as heart rate, QRS duration, etc." This indicates that the device itself performs some algorithmic processing (measurement) but does not provide diagnostic interpretation. Therefore, a standalone diagnostic performance study (algorithm only) as typically understood for AI-based diagnostic devices, is not described here. The presented studies focus on safety, electrical performance, and software validation.
7. Type of Ground Truth Used
Given the nature of the tests (compliance with standards like IEC 60601-1-12, software validation), the "ground truth" would be defined by the technical specifications of the standards themselves (e.g., passing specific electrical tests, proper software function as per design documents) rather than expert consensus, pathology, or outcomes data related to ECG interpretation. The document does not describe studies requiring such clinical ground truth.
8. Sample Size for the Training Set
The document does not mention a training set sample size. Since the device's "Analysis module" provides measurements but no interpretation, it's unlikely to involve machine learning models that require extensive training data in the traditional sense for diagnostic capabilities. The software validation tests mentioned ("Unit Test, Integration Test, System Test") are for functional correctness rather than training predictive models.
9. How the Ground Truth for the Training Set Was Established
As no training set is described, there is no information on how ground truth for a training set was established.
Ask a specific question about this device
(30 days)
The V-Patch Cardiac Monitor is intended to record, transfer and store single-channel electrocardiogram (ECG) data via Bluetooth transmission to Bluetooth enabled devices. The monitor is intended for use by healthcare professionals, patients with known or suspected heart conditions and health-conscious individuals. The ECG data is intended to supplement other patient data and is not intended for automated analysis. The device has not been tested and it is not intended for pediatric use.
The V-Patch Cardiac Monitor ("V-Patch CM") is a continuous electrocardiogram (ECG) recording device to record, store, and transfer single channel ECGs and is designed for remote ECG data collection and remote monitoring. The device is in the form of an ECG patch that records ECG signals and transmits directly to a Bluetooth™ enabled device. The V-Patch CM has Bluetooth module for transferring ECG data from V-Patch CM to a Bluetooth enabled computing or mobile devices to be accessible to healthcare professionals. The V-Patch CM includes a battery powered electronic unit that is used with off-the-shelf (OTS) disposable medical grade gel electrodes for long term monitoring. The adhesive electrodes should be replaced by the user every 24 hours or when it no longer adheres to skin. The V-Patch CM is a prescription use device and the recorded ECG data is intended to be used with other patient data. The device does not provide any automated ECG analysis.
The provided document does not contain information about specific acceptance criteria, device performance, sample sizes used, data provenance, expert involvement, adjudication methods, MRMC studies, or standalone performance for the V-Patch™ Cardiac Monitor.
The document is a 510(k) summary for the V-Patch™ Cardiac Monitor (K222842) and focuses on establishing substantial equivalence to a predicate device (eCordum™ Cardiac Monitor, K193296). It outlines the device description, indications for use, and a comparison of technological characteristics, stating that the two devices are "identical in form and function..., using the same components, materials, and software... and the same manufacturing process."
Instead of presenting performance data against specific acceptance criteria, it lists nonclinical testing activities that were successfully passed, indicating compliance with relevant standards. These tests confirm the device's electrical safety, electromagnetic compatibility, biocompatibility, and functionality for ambulatory ECG recording, but they do not provide quantitative performance metrics against pre-defined acceptance criteria.
The document explicitly states: "No clinical testing is required to demonstrate substantial equivalence to the predicate eCordum Cardiac Monitor (K193296)." This implies that no specific clinical study was conducted to prove primary performance criteria for the V-Patch™ Cardiac Monitor itself.
Therefore, I cannot provide the requested information based on the given text.
Ask a specific question about this device
(192 days)
The QT ECG System is intended to acquire, record and process an electrocardiographic signal so that it can be transmitted digitally via Bluetooth technology to a cell-phone or mobile device, then to a remote location. The QT ECG System is indicated for use on infants, children and adults. It is designed to be operated by adults in the home, or by healthcare workers in non-acute care clinical facilities (such as nursing facilities), to record and transmit a 12-lead ECG and rhythm strip in near real-time to enable review at a physician's office, hospital or other medical receiving centers.
The OT ECG system is a non-defibrillator-proof, hand-held, cordless 12-lead electrocardiograph (ECG) system with Bluetooth connectivity. The OT ECG system consists of 5 major components:
- ·The QT ECG Recorder—Compact device that records 12-lead, resting electrocardiograms, then transmits the recorded data to a mobile device (smartphone, tablet, etc.) paired via Bluetooth. A Bluetooth-enabled mobile device (not included) is needed to operate the QT ECG Recorder, and to send the recoded rhythm strip to a cardiologist or licensed physician for review.
- . The QT ECG Electrode Strip-Disposable, patented electrodes that are prepositioned on a self-adhesive strip.
- The QT ECG App - Software that lets the user uses their mobile device to operate the QT ECG recorder, then send the recorded data via cloud to a certified medical professional for review.
- . Analysis - The analysis module provides ECG measurement from the collected data. It does not make any interpretation of the intervals provided based on factors such as heart rate, ORS duration, etc.
- . Web Service — The web service provides an interface for communication.
The recorded ECG data is saved temporarily on the mobile device until it is transferred via the Internet to the cloud server. The QT ECG System does not have monitoring capabilities and does not have diagnostic alarm function. The QT ECG System is intended for use on infants, children and adults patients to acquire ECG signals to be transmitted wirelessly via Bluetooth to a mobile device, and then over the Internet to a web service. The QT ECG System is designed to be used by a patient to record and transmit ECG data to a physician's office, hospital or other medical receiving center for review.
The provided text is a 510(k) Summary for the QT Medical, Inc. QT ECG device (K220795). This document primarily focuses on establishing substantial equivalence to a predicate device (K180157) and outlines various non-clinical testing performed to show safety and performance. However, it does not contain the detailed clinical study information typically required to describe acceptance criteria and the study that proves the device meets those criteria for an AI/ML enabled medical device.
The document states:
- "All the test results demonstrate OT ECG meets the requirements of its pre-defined acceptance criteria and indication for use, and is substantially equivalent to the predicate device." (Page 5)
- "Although there are some different specifications between two devices, the software validation, performance test and usability test have been completed to demonstrate that the differences between these parameters would not impact the safety and effectiveness of the subject device. The subject device has also undergone all safety and performance tests, and the results complied with the test requests." (Page 14)
These statements indicate that acceptance criteria exist and were met through testing. However, the document lists these tests as "Non-clinical Testing" including biocompatibility, software validation, cybersecurity, electromagnetic compatibility, electrical safety, performance (of electrode and recorder, durability, wireless interference), and human factors (usability). These are engineering and regulatory compliance tests, not a clinical study proving diagnostic or predictive performance against specific clinical acceptance criteria for an AI/ML algorithm's output.
The device description on page 5 explicitly mentions:
- "Analysis - The analysis module provides ECG measurement from the collected data. It does not make any interpretation of the intervals provided based on factors such as heart rate, QRS duration, etc."
- "The QT ECG System does not have monitoring capabilities and does not have diagnostic alarm function." (Page 5)
This indicates that the "Analysis module" performs basic ECG measurements (e.g., intervals, durations) but does not provide diagnostic interpretations or AI-driven insights that would require validation through a clinical study with expert ground truth. The device's function is data acquisition and transmission, not automated diagnosis or interpretation. Therefore, a clinical study of the device's diagnostic performance, as might be done for an AI/ML device that provides interpretations, is not described because it appears not to be within the scope of this device's intended use or capabilities.
Given the information provided in the document, I cannot create the requested table or answer most of the questions because the device does not include an AI/ML component for diagnostic interpretation and the detailed clinical study information is not present. The "performance test" mentioned in the non-clinical testing section likely refers to technical specifications for ECG signal acquisition and transmission, not clinical outcome or interpretative accuracy.
Therefore, the following table and answers are based on the absence of such information in the provided document, reflecting that this device's clearance did not hinge on an AI/ML diagnostic performance study.
Acceptance Criteria and Device Performance (Based on available information)
The provided submission does not detail acceptance criteria related to AI/ML diagnostic performance as the device's "Analysis module" performs raw ECG measurement without interpretation or diagnostic capabilities. The acceptance criteria mentioned are general for device functionality, safety, and substantial equivalence to a predicate device.
| Acceptance Criteria Category | Specific Criterion (if available from this document) | Reported Device Performance (if available from this document) |
|---|---|---|
| AI/ML Diagnostic/Interpretative Performance | Not applicable; device performs measurement, not interpretation. | Not applicable; device performs measurement, not interpretation. |
| Biocompatibility | - (Met predefined criteria) | "test results demonstrate OT ECG meets the requirements of its pre-defined acceptance criteria." |
| Software Validation & Cybersecurity | - (Met predefined criteria) | "test results demonstrate OT ECG meets the requirements of its pre-defined acceptance criteria." |
| Electromagnetic Compatibility & Electrical Safety | - (Met predefined criteria) | "test results demonstrate OT ECG meets the requirements of its pre-defined acceptance criteria." |
| Performance (general device) | - (Met predefined criteria for electrode, recorder, durability, measurement, wireless interference) | "test results demonstrate OT ECG meets the requirements of its pre-defined acceptance criteria." |
| Human Factors (Usability) | - (Met predefined criteria) | "test results demonstrate OT ECG meets the requirements of its pre-defined acceptance criteria." |
| Substantial Equivalence | Demonstrated equivalence to K180157 | "QT ECG is substantially equivalent to the predicate device." |
Answers to Specific Questions (Based on available information):
-
A table of acceptance criteria and the reported device performance:
(See table above. Note: This device is not described as having an AI/ML diagnostic component, so the clinical performance criteria typically associated with such devices are not present.) -
Sample sizes used for the test set and the data provenance:
The document mentions general "validation activities" and "performance tests" but does not specify sample sizes for test sets related to clinical performance or diagnostic accuracy. The data provenance (e.g., country of origin, retrospective/prospective) for any specific performance testing is also not provided. The testing described appears to be primarily engineering and regulatory compliance testing rather than clinical study data. -
Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):
Not applicable/Not provided. Since the device states it "does not make any interpretation" and "does not have diagnostic alarm function," there's no indication that expert-established ground truth for diagnostic or interpretative performance was required or used in this submission. -
Adjudication method (e.g., 2+1, 3+1, none) for the test set:
Not applicable/Not provided. No clinical diagnostic test set or adjudication method is described in this document. -
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
No MRMC study was done or described. The device does not appear to provide AI assistance for human interpretation, as its analysis module only provides measurements, not interpretations. -
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
No standalone diagnostic performance study was done or described as the device's analysis module explicitly states it "does not make any interpretation." -
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
Not applicable/Not provided regarding diagnostic ground truth, for reasons stated above. The ground truth for the non-clinical performance tests would be defined by engineering standards and specifications. -
The sample size for the training set:
Not applicable/Not provided. As the device's "Analysis module" performs measurements rather than AI-driven interpretations, there's no mention of a training set for a machine learning algorithm. -
How the ground truth for the training set was established:
Not applicable/Not provided. (See answer to question 8).
Ask a specific question about this device
(280 days)
The Study Watch with Irregular Pulse Monitor is indicated for use by adult patients (22 years and older) who have been diagnosed with, or are susceptible to developing, atrial fibrillation enabling them to monitor and record their heart rhythms. Study Watch is also intended to record, store, transfer, and display single-channel electrocardiogram (ECG) rhythms.
The Study Watch is intended for prescription use only.
The Study Watch with Irregular Pulse Monitor is a miniaturized physiological data monitoring device that is intended to record, store, transfer and display single-channel electrocardiogram (ECG) rhythms. The Study Watch with Irregular Pulse Monitor is intended to notify the user in the event of an irregular pulse, such as atrial fibrillation (AF), and recommend acquisition of an ECG. The device utilizes a band and proprietary watch that incorporates two dedicated sensing electrodes to obtain a single-channel ECG measurement. In practice, a healthcare professional (HCP) may prescribe the Study Watch with Irregular Pulse Monitor to a patient and recommend the capture and transmission of one or more ECG rhythms on a daily basis for analysis by the HCP. The patient uses prompts provided on the graphical user interface (GUI) of the watch to collect a real-time 45-second ECG measurement. The patient is able to capture an ECG rhythm on-demand by navigation to the ECG menu, upon receipt of an irregular pulse notification. While an animated ECG waveform is briefly displayed on the GUI of the watch, the patient will not have direct access to the collected waveform. Using the provided charging dock (Cradle) and Verily's Study Hub or Watch App, the ECG measurements are securely transferred to Verily's cloud server and are viewed by the HCP using a web portal. Additionally, data may be securely transferred to interoperable devices for further analysis. The web portal is solely intended for use by the HCP to view time-stamped ECG waveforms collected by the patient and does not include any analysis features. Collectively, the Study Watch consists of the wearable watch and band, the Study Hub or Watch app (used to transfer data from the watch to the cloud), a Cradle, and the web portal.
Below is a detailed description of the acceptance criteria and the study proving the device meets these criteria, based on the provided document.
Device Name: Study Watch with Irregular Pulse Monitor
1. Table of Acceptance Criteria and Reported Device Performance
| Metric | Acceptance Criteria (Threshold) | Reported Device Performance (Home Use) |
|---|---|---|
| Per-interval Sensitivity | Not explicitly stated (only "exceeds pre-specified thresholds") | 96.1% (95% CI: 92.7 - 98.0) |
| Per-interval Specificity | Not explicitly stated (only "exceeds pre-specified thresholds") | 98.1% (95% CI: 97.2 - 99.1) |
Note: The document states the algorithm "exceeds the pre-specified per-interval performance thresholds for sensitivity and specificity," but the exact numerical thresholds are not provided within the given text.
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Test Set: The document mentions "thousands of ECG data collected from iRhythm Technologies' Zio devices" for algorithm training but does not explicitly state the sample size for the validation (test) set. However, it describes the "Study Watch AF Detection At Home (NCT04546763) data" as consisting of "free-living, multi-day PPG recordings obtained from the Study Watch, along with ECG-based rhythm labels obtained from reference iRhythm Technologies Zio XT Patches worn simultaneously."
- Data Provenance: The study was conducted in the USA, with regional demographics provided: Midwest (8%), Mountain (32%), West (32%), Northeast (14%), South (13%). The data was collected prospectively in a "free-living (home use) setting."
3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications
- Number of Experts: Not specified.
- Qualifications of Experts: The algorithm was trained using ECG data that had undergone "Certified Cardiographic Technician (CCT) review." It can be inferred that similar qualified personnel were involved in establishing the ground truth for the validation set, as the reference standard (Zio XT Patches) typically requires expert interpretation.
4. Adjudication Method for the Test Set
- The document does not explicitly state an adjudication method (e.g., 2+1, 3+1). It only mentions that ECG-based rhythm labels were "obtained from reference iRhythm Technologies Zio XT Patches worn simultaneously" and that the training data underwent "Certified Cardiographic Technician (CCT) review."
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size
- No, an MRMC comparative effectiveness study was not done in the context of human readers improving with AI vs. without AI assistance. This study focused on the standalone performance of the irregular pulse monitoring algorithm.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
- Yes, a standalone performance study was done. The described "Clinical Performance Data" section focuses solely on the algorithm's performance (sensitivity and specificity) in detecting irregular pulses based on PPG data, using a reference ECG as ground truth, without human-in-the-loop evaluation.
7. The Type of Ground Truth Used
- Clinical Ground Truth: The ground truth for validation was established using ECG-based rhythm labels obtained from reference iRhythm Technologies Zio XT Patches worn simultaneously with the Study Watch. This is a robust clinical ground truth for cardiac rhythm analysis.
8. The Sample Size for the Training Set
- The algorithm was trained using "thousands of ECG data collected from iRhythm Technologies' Zio devices." A specific number is not provided beyond "thousands."
9. How the Ground Truth for the Training Set Was Established
- The ground truth for the training set was established through "ECG data collected from iRhythm Technologies' Zio devices, which have undergone Certified Cardiographic Technician (CCT) review." This indicates human expert review by qualified professionals.
Ask a specific question about this device
(107 days)
KardiaMobile 6L is intended to record, store and two-channel electrocardiogram (ECG) rhythms. In single channel mode, KardiaMobile 6L can record Lead-I. In two channel mode, KardiaMobile 6L can record Lead-I and Lead-II simultaneously and derive Lead-III and unipolar limb leads aVR, aVF and aVL. KardiaMobile 6L also displays ECG rhythms and output of ECG analysis from AliveCor's KardiaAI platform including detecting the presence of normal sinus rhythm, atrial fibrillation, bradycardia, and others. KardiaMobile 6L is intended for use by healthcare professionals, patients with known or suspected heart conditions and health conscious individuals. The device has not been tested and is not intended for pediatric use.
KardiaMobile 6L is a trans-telephonic (transmission by smartphone) electrocardiogram (ECG) event recorder that records, stores, transfers, and analyzes single-channel or two channel ECG rhythm recordings. KardiaMobile 6L provides output of one or six ECG leads, including Lead I, Lead II, Lead III, aVL, aVR and aVF. The device utilizes the computing power of iOS-based or Android-based devices (referred to as "Mobile Computing Platforms" (MCP) within this 510(k)) to record and analyze ECG signals. KardiaMobile 6L consists of KardiaMobile 6L Hardware (portable small wireless hardware with electrodes) and the Kardia Core app. which is installed on an MCP (i.e., iOS or Android devices). KardiaMobile 6L Hardware uses Bluetooth to transmit the ECG signal from the electrodes to the Kardia Core app on the MCP, which then displays the recorded ECG on the MCP's screen. The device is intended to be used by patients with known or suspected heart conditions and health conscious individuals as well as by healthcare professionals (HCPs) who want to remotely monitor their patient's heart health. The device is available for Over-the-Counter (OTC) as well prescription use. The Kardia Core app provides the complete ECG recording and analysis workflow, from acquisition of the signal from the KardiaMobile 6L hardware, to the display of the ECG and analysis results, to printing of the ECG rhythm strip. The app utilizes KardiaAI (K181823, K201985) to provide ECG analysis, which includes the determinations of Normal Sinus Rhythm, Atrial Fibrillation, Bradycardia, Tachycardia, or Unclassified to OTC users and additionally, Sinus Rhythm determinations (Sinus Rhythm with Wide ORS, Sinus Rhythm with Supraventricular Ectopy, and Sinus Rhythm with Premature Ventricular Contractions) for prescription-use only users.
The document provided is a 510(k) summary for the AliveCor KardiaMobile 6L device. It describes the device, its intended use, and its substantial equivalence to a previously cleared predicate device.
Based on the provided information, the current submission (K220350) is for a minor modification (software reorganization and API addition) to an existing device (K210753, KardiaMobile 6L). Therefore, the study described here is primarily focused on demonstrating that these software changes do not adversely affect the device's safety or effectiveness as previously established.
Here's a breakdown of the requested information based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance:
The document explicitly states that the subject device (K220350) is being compared to its predicate (K210753). The performance criteria are implicitly met by demonstrating that the changes do not raise new questions of safety or effectiveness, as stated in the "Nonclinical Testing Summary" and "Conclusions" sections.
| Acceptance Criteria (Implied by Predicate Equivalence) | Reported Device Performance (Subject Device K220350) |
|---|---|
| Maintain all functionalities of the predicate device (KardiaMobile 6L K210753) | All functionalities (ECG acquisition, display, analysis, PDF generation) maintained. |
| ECG Analysis Determinations (KardiaAI K181823) for OTC and Rx Only users remain equivalent for: Atrial Fibrillation, Normal Sinus Rhythm, Tachycardia, Bradycardia, Unclassified, Unreadable. | Performances for these determinations are stated as "No difference" compared to the predicate. |
| ECG Analysis Determinations (KardiaAI K201985) for Rx Only users remain equivalent for: Atrial Fibrillation, Sinus Rhythm, Sinus Rhythm with Wide QRS, Sinus Rhythm with Supraventricular Ectopy, Sinus Rhythm with Premature Ventricular Contractions, Normal Sinus Rhythm, Tachycardia, Bradycardia, Unclassified, Unreadable. | Performances for these determinations are stated as "No difference" compared to the predicate. |
| No degradation in data acquisition specifications: Frequency Response (0.5Hz – 40Hz), Number of ECG electrodes (Three dry electrodes), Number of ECG channels (Single-channel and two-channel), Resolution (16 bits), Sample Rate (300 Samples/Second). | All data acquisition specifications are stated as "No difference" compared to the predicate. |
| No degradation in power supply specifications: Battery (1 Lithium Manganese Dioxide Coin Cells), Battery Life (100 hours operational typical). | All power supply specifications are stated as "No difference" compared to the predicate. |
| Successful integration of an Application Program Interface (API) for other mobile applications. | API for integrating with other mobile applications is a new feature. Verification activities focused on software changes were successfully passed. |
| All verification test methods previously used for K183319 and K210753 are successfully passed. | The subject device successfully passed the software verification test methods previously used in K183319 and K210753. |
2. Sample size used for the test set and the data provenance:
The document states: "No clinical testing was required for the minor changes made to the software." This implies that the current submission (K220350) did not involve a new clinical test set with human subjects to prove performance, but rather relied on nonclinical testing, specifically software verification activities.
Therefore, information on the sample size for a test set (e.g., number of ECGs, number of patients) or data provenance (country of origin, retrospective/prospective) directly related to a new clinical performance study for K220350 is not provided as it was explicitly deemed not required. The performance is being established by comparing to the predicate and ensuring the software changes did not introduce new risks. The previous predicate devices (K210753, K183319) would have had their own clinical data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
Not applicable, as no new clinical testing was performed for this specific submission (K220350). The performance claims are based on the substantial equivalence to the predicate, K210753, which would have established its own ground truth using experts.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
Not applicable, as no new clinical testing was performed for this specific submission (K220350).
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
Not applicable. The document states "No clinical testing was required for the minor changes made to the software." This submission is not an MRMC study and does not evaluate AI assistance for human readers. It focuses on software reorganization and API addition.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
The device description indicates that the Kardia Core app utilizes KardiaAI (K181823, K201985) to provide ECG analysis, detecting conditions like Normal Sinus Rhythm, Atrial Fibrillation, Bradycardia, Tachycardia, or Unclassified. This implies a standalone algorithmic analysis component. However, the current submission (K220350) primarily focuses on the software reorganization and API addition and explicitly states no new clinical testing was required. The performance of these AI determinations would have been established in the previous predicate clearances (K181823 and K201985), not in this particular submission.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
Not directly addressed for this specific submission (K220350), as no new clinical data was collected. For the underlying KardiaAI platform (K181823, K201985), the ground truth for ECG analysis typically involves expert consensus of cardiologists reviewing the ECGs.
8. The sample size for the training set:
The document does not provide details of the training set size, as this submission is not about training a new AI model but rather a software modification to an already cleared device that utilizes existing AI algorithms (KardiaAI K181823, K201985). The training data for those AI algorithms would have been described in their respective 510(k) submissions.
9. How the ground truth for the training set was established:
Not directly addressed for this specific submission (K220350), as it pertains to the training of the underlying KardiaAI algorithms (K181823, K201985) rather than the software reorganization. For such algorithms, ground truth for training is typically established by expert adjudication/consensus of ECG interpretations by qualified medical professionals (e.g., cardiologists).