Search Filters

Search Results

Found 128 results

510(k) Data Aggregation

    K Number
    K252794

    Validate with FDA (Live)

    Date Cleared
    2026-01-08

    (128 days)

    Product Code
    Regulation Number
    870.1220
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K250590

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-07-23

    (146 days)

    Product Code
    Regulation Number
    870.1220
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MAGiC Sweep™ EP Mapping Catheter is intended for intracardiac electrophysiological recording and/or stimulation for pacing in the heart.

    Device Description

    The MAGiC Sweep™ EP Mapping Catheters are biocompatible, flexible, radiopaque fixed catheters that are 130cm in length and available in multiple electrode spacing configurations and multiple magnet spacings. They have magnets encapsulated in the device shaft to allow manipulation through the vasculature and placement by a physician controlled robotic magnetic navigation system. Each device consists of 19 electrodes plus one tip electrode and come in three (3) different electrode spacings. In addition there are twenty (20) different magnet spacings for each electrode spacing configuration. The catheters are 8F diameter tapering to 5F at the distal end of the device shaft. The variety of configurations facilitate the navigation of the device to specific areas of the heart and facilitate recording of intracardiac signals and/or stimulation in the atrial and ventricular regions of the heart during electrophysiology studies.

    The MAGiC Sweep™ EP Mapping Catheter is designed to be positioned in various endocardial and intravascular sites utilizing the Stereotaxis magnetic navigation system (MNS) along with the Stereotaxis catheter advancement system (CAS). The MAGiC Sweep™ EP Mapping Catheter is designed for recording intracardiac signals when connected to a recording system and for cardiac stimulation when connected to a stimulation system.

    The MAGiC Sweep™ EP Mapping Catheter is comprised of a Pebax catheter shaft, platinum-iridium ring electrodes, platinum-iridium electrode tip, Neodymium Boron Magnets attached to a safety cable, and Redel electrical connectors. The Neodymium Boron Magnets are encapsulated within the catheter shaft.

    AI/ML Overview

    The provided 510(k) clearance letter and summary describe the acceptance criteria and the study that proves the device meets the acceptance criteria for the Stereotaxis MAGiC Sweep™ EP Mapping Catheter.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Tests PerformedReported Device Performance/Conclusion
    Visual InspectionVisualMet specifications
    Dimensional ConformanceDimensionalMet specifications
    Electrical PerformanceContinuityMet specifications
    Leakage ImpedanceMet specifications
    Dielectric StrengthMet specifications
    Mechanical PerformanceTensile StrengthMet specifications
    TorqueMet specifications
    DeflectionMet specifications
    Shaft BucklingMet specifications
    Radio-detectabilityMet specifications
    Corrosion ResistanceMet specifications
    BiocompatibilityShort Term (<24 hours)Met specifications
    Blood ContactingMet specifications
    Packaging IntegrityPackagingMet specifications
    Functional PerformanceAnimal Studies: Catheter navigation, positioning, mappingSuccessfully navigated, positioned, and mapped heart chambers (left/right ventricles, left/right atriums)
    Safety and EffectivenessAssessment of new questions regarding safety or effectiveness in testingNo new questions raised regarding safety or effectiveness

    2. Sample Size Used for the Test Set and Data Provenance

    The document details two primary types of testing: Bench Testing and Animal Testing.

    • Bench Testing: The sample size for each specific bench test (visual, dimensional, electrical, mechanical, biocompatibility, packaging) is not explicitly stated in the provided text. It is generally understood that such testing involves a sufficient number of samples to ensure statistical validity, but the exact numbers are not given.
    • Animal Testing: The sample size for animal testing is not explicitly stated. It mentions "an animal studies (porcine model)," indicating that some number of porcine subjects were used, but the specific count is not provided.
    • Data Provenance:
      • Bench Testing: This is laboratory-based testing, so the data provenance is derived from the manufacturer's internal testing facilities.
      • Animal Testing: The animal study was conducted using a porcine model. The country of origin for the animal study data is not specified. The study was prospective as it involved actively using the device in live subjects to assess its performance.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not specify the number of experts used or their qualifications for establishing ground truth, particularly for the animal study. The animal study reports that the catheter was "advanced, positioned and successfully mapped" by presumably a researcher or veterinary expert, but no details on their number or credentials are provided. For bench testing, the "ground truth" is typically established by engineering specifications and standards, not by human experts interpreting results.

    4. Adjudication Method for the Test Set

    The document does not mention any specific adjudication method (e.g., 2+1, 3+1) for the test set. For bench and animal testing, adjudication usually means verifying that the device performance meets pre-defined engineering and biological specifications rather than a consensus among human reviewers.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not performed, nor is it applicable to this device. This device is an EP mapping catheter, a hardware medical device used for recording and stimulation, not an AI/software device that assists human readers in interpreting medical images or data. Therefore, there is no mention of human readers improving with or without AI assistance, or an effect size.

    6. Standalone Performance (Algorithm Only without Human-in-the-Loop Performance)

    Since this is a hardware device (EP mapping catheter) and not a software/AI algorithm, the concept of "standalone performance" in the context of an algorithm without human-in-the-loop performance is not applicable. Its performance is inherently linked to its physical interaction for recording and stimulation, guided by a physician using a robotic system.

    7. Type of Ground Truth Used

    • Bench Testing: The ground truth for bench testing was established by adherence to existing standards and test methods. This means the performance metrics (e.g., electrical continuity, tensile strength, biocompatibility profiles) were compared against pre-defined engineering specifications and regulatory limits.
    • Animal Testing: The ground truth for animal testing was established by observational functional performance during live animal studies. This involved observing whether the catheter could be successfully advanced, positioned, and used to map the defined heart chambers. This can be considered a form of "outcomes data" in a controlled experimental setting.

    8. Sample Size for the Training Set

    The document does not mention a training set. This is expected because the MAGiC Sweep™ EP Mapping Catheter is a physical medical device (hardware), not an artificial intelligence (AI) or machine learning (ML) algorithm that requires a training set. The development and testing of this device involve engineering design, material selection, and rigorous physical and biological testing, not data-driven model training.

    9. How the Ground Truth for the Training Set Was Established

    As there is no training set for this hardware device, the concept of establishing ground truth for a training set is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K250310

    Validate with FDA (Live)

    Date Cleared
    2025-06-27

    (144 days)

    Product Code
    Regulation Number
    870.1220
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VIKING Fixed Curve Diagnostic Catheter is indicated for use to diagnose cardiac arrhythmias.

    Device Description

    The Viking Fixed Curve Diagnostic Catheters (also referred to as Viking Catheters) are intended for use in electrophysiology procedures to record, stimulate and pace in the diagnosis of cardiac arrhythmias.

    The Viking Catheters receive electrical signals from the cardiac tissue, which are transmitted via the electrode wiring through the connector and cable to an amplifier, where the signals are displayed on an EP recording system for the physician to view and interpret in diagnosing cardiac arrhythmias. The physician inserts the diagnostic catheter into the vasculature percutaneously. To reach the right side of the heart, either the internal jugular vein, the subclavian veins, or the femoral veins can be accessed. To reach the left ventricle, the retrograde approach via the femoral artery can be used. In some cases, the radial artery may be used to access the targeted cardiac chamber. Using fluoroscopic guidance, the physician then navigates the diagnostic catheter to the target site using the catheter steering mechanism, as confirmed by the intracardiac electrograms and/or pacing threshold, confirming electrode to endocardium interface.

    Viking Catheters are radiopaque, flexible, insulated catheters. The catheters have a polymer shaft that houses the electrical wiring that connects the electrodes to a connector at the proximal end of the catheter; each electrode wire is soldered into this connector. When the catheter is used, the connector is attached to a sterile cable that connects to the hospital recording and pacing equipment.

    The Viking Catheters are offered pre-packaged sterile and are a single use product not to be reused, reprocessed, or resterilized. The method of sterilization is ethylene oxide (EO).

    AI/ML Overview

    The provided FDA 510(k) clearance letter and summary for the VIKING™ Fixed Curve Diagnostic Catheter do not contain any information regarding acceptance criteria and studies demonstrating device performance in the context of an AI/algorithm-based diagnostic device.

    The document describes a medical device, specifically a diagnostic catheter, and the FDA's determination of substantial equivalence to a predicate device. The performance data discussed relates to:

    • Bioburden Testing: To confirm bioburden levels were not adversely affected.
    • Design Verification Testing: To demonstrate adherence to product specifications at baseline and after accelerated aging (25 months/761 days).
    • Packaging Verification Testing: To ensure packaging meets design input and maintains sterility.

    These tests are standard for physical medical devices to ensure safety, sterility, functionality, and shelf-life, not for evaluating the diagnostic accuracy or effectiveness of an AI algorithm.

    Therefore, I cannot provide the requested information for acceptance criteria and studies related to an AI diagnostic device because the provided text is for a physical diagnostic catheter and does not involve AI or algorithmic performance evaluation.

    To directly answer your numbered points based only on the provided text, without making assumptions:

    1. A table of acceptance criteria and the reported device performance: Not applicable for AI performance from this document. The document presents performance data for physical device characteristics (bioburden, design verification, packaging verification) to support substantial equivalence, but no specific acceptance criteria or reported values are detailed in the summary itself.
    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective): Not applicable for AI performance from this document. The document refers to "Design Verification Testing" and "Packaging Design Verification testing" but does not specify sample sizes or data provenance in the context of diagnostic accuracy.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience): Not applicable for AI performance from this document.
    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable for AI performance from this document.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable, as this is a physical diagnostic catheter, not an AI-assisted diagnostic tool.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable, as this is a physical diagnostic catheter, not an AI algorithm.
    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc): Not applicable for AI performance from this document. The "ground truth" for the catheter's physical performance would be its adherence to engineering specifications and international standards.
    8. The sample size for the training set: Not applicable, as this is a physical diagnostic catheter, not an AI algorithm.
    9. How the ground truth for the training set was established: Not applicable, as this is a physical diagnostic catheter, not an AI algorithm.

    In summary, the provided FDA document pertains to the clearance of a physical medical device (a catheter) and does not involve the evaluation of an AI algorithm or its diagnostic performance.

    Ask a Question

    Ask a specific question about this device

    K Number
    K240366

    Validate with FDA (Live)

    Date Cleared
    2024-11-01

    (269 days)

    Product Code
    Regulation Number
    870.1220
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    EP.XT, Dynamic Tip, and Dynamic XT Unidirectional Steerable Diagnostic Catheters are indicated for use to diagnose cardiac arrhythmia.

    Device Description

    The unidirectional steerable diagnostic catheters acquire electrical signals from the cardiac tissue, which are transmitted via the electrode wiring through the connector and cable to an amplifier, where the signals are displayed on an EP recording system for the physician to view and interpret in diagnosing cardiac arrhythmias. The physician inserts the diagnostic catheter(s) percutaneously via either the anterior approach using the internal jugular vein or the subclavian vein or via the femoral vein or for a retrograde approach, via the femoral artery. In some cases, the radial artery may be used to access the targeted cardiac chamber. Using fluoroscopic guidance, the physician then navigates the diagnostic catheter to the target site using the catheter steering mechanism as confirmed by the intracardiac electrograms and/or pacing threshold, confirming electrode to endocardium interface.

    The steerable diagnostic electrode catheters are radiopaque, flexible, insulated catheters with a polymer shaft, each having a mechanical operating mechanism. In each catheter the polymer shaft connects to the catheter handle whereby the curve or loop is actuated by a single pull cable. The EP+XT Catheter has a white rotating handle mechanism which provides curve actuation of the distal tip. The Dynamic Tip and Dynamic XT Catheters have a plunger mechanism, which, when moved forward or back, results in curvature of the distal tip.

    The EP•XT, Dynamic Tip, and Dynamic XT Catheters are offered pre-packaged sterile and are a single use product not to be reused, reprocessed, or resterilized. The method of sterilization is ethylene oxide (EO).

    AI/ML Overview

    This document describes the FDA's decision to clear Boston Scientific's EP.XT, Dynamic Tip, and Dynamic XT Unidirectional Steerable Diagnostic Catheters for marketing based on a determination of substantial equivalence to predicate devices. This clearance is not based on a study proving the device meets specific performance criteria for diagnostic accuracy in the same way an AI/ML diagnostic would be; rather, it relates to the device's physical and functional characteristics and its ability to acquire electrical signals.

    Therefore, the requested information (acceptance criteria and study details related to diagnostic performance, human reader improvement with AI, ground truth, expert adjudication, etc.) is not applicable to this 510(k) submission. This is a clearance for a physical medical device (catheter) used to acquire data, not a software-based diagnostic algorithm.

    Here's a breakdown based on the provided document, addressing what is relevant and noting what is not:

    1. Acceptance Criteria and Reported Device Performance

    The "acceptance criteria" here are based on the device's physical and functional performance, sterilization, and packaging, demonstrating it is substantially equivalent to a predicate device. There are no diagnostic performance metrics (e.g., sensitivity, specificity, AUC) reported because this is not a diagnostic algorithm.

    Acceptance Criteria CategoryReported Device Performance (Summary)
    BiocompatibilityImplied by "substantially equivalent ... to predicate device." Specific testing not detailed in summary.
    SterilizationEO (Ethylene Oxide) sterilization validated; packaging maintains sterility.
    Mechanical PerformanceDemonstrated through "Design Verification Testing" at baseline and 37 months accelerated aging. Ensured adherence to "applicable product specifications." Examples of specific tests would likely include steerability, durability, electrical signal acquisition integrity, etc., but these are not detailed in the summary.
    Packaging Integrity"Packaging Design Verification testing... meet[s] predetermined design input specification and maintain sterility in conformance with ISO 11607-1."

    2. Sample Size Used for the Test Set and Data Provenance

    N/A. This 510(k) summary does not describe a clinical study for diagnostic performance with a test set of patient data. The "test set" for this device would refer to units of the manufactured catheters themselves undergoing engineering and bench testing. The document states "Design verification testing was performed on the EP.XT, Dynamic Tip, and Dynamic XT Catheters," implying a sample of the manufactured devices were tested, but specific numbers are not provided in this summary.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    N/A. This is not a diagnostic algorithm requiring expert-established ground truth on patient data. The ground truth for this device's performance relates to engineering specifications and international standards (e.g., ISO 11607-1 for packaging, ISO 11737-1 for bioburden).

    4. Adjudication Method for the Test Set

    N/A. Not applicable to a physical device's engineering and performance testing.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. An MRMC study is relevant for evaluating the impact of AI on human reader performance for diagnostic tasks. This device is a diagnostic tool that acquires electrical signals, not an AI-powered diagnostic system.

    6. If a Standalone Performance Study Was Done

    No, not in the context of an algorithm's diagnostic performance. The "performance data" described (Bioburden, Design Verification, Packaging Verification) are standalone tests of the device's physical and functional attributes against engineering specifications and regulatory standards.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's clearance is based on:

    • Engineering Specifications: The design inputs and performance requirements set by Boston Scientific for the catheters.
    • International Standards: E.g., ISO 11737-1 for bioburden, ISO 11607-1 for packaging, and likely other ISO standards for biocompatibility and electrical safety (though not explicitly listed in the summary).
    • Predicate Device Performance: The primary "ground truth" for 510(k) clearance is that the new device performs "substantially equivalent" to legally marketed predicate devices, meaning it has similar technological characteristics and the same intended use.

    8. The Sample Size for the Training Set

    N/A. This is not an AI/ML device that requires a training set of data.

    9. How the Ground Truth for the Training Set Was Established

    N/A. Not applicable as there is no training set for an AI/ML algorithm.

    In summary: The provided document is a 510(k) clearance letter for a medical device (catheter) based on substantial equivalence. It does not contain information about acceptance criteria or studies related to the diagnostic accuracy of an AI/ML algorithm, nor does it describe human-in-the-loop performance studies. The "performance data" refers to the physical and functional characteristics of the catheter itself, not its diagnostic accuracy in interpreting patient data.

    Ask a Question

    Ask a specific question about this device

    K Number
    K233900

    Validate with FDA (Live)

    Date Cleared
    2024-09-06

    (270 days)

    Product Code
    Regulation Number
    870.1220
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Nordica PV Cryo Mapping Catheter is indicated for multiple electrophysiological mapping of cardiac structures (i.e., recording and stimulation only). The Nordica PV Cryo Mapping Catheter is designed to obtain electrograms in the atrial regions of the heart.

    Device Description

    The Nordica PV Cryo Mapping Catheter is a 3F sterile, single use, multi-electrode, diagnostic catheter designed to map cardiac signals during ablation procedures. The catheter is provided with a fixed loop diameter of 15mm or 20mm with either 5 or 4 evenly spaced, radiopaque pairs of electrodes. The proximal end of the catheter contains an electrical connection for the EP electrical cable that integrates with electrophysiology lab recording systems. Once deployed in the vasculature through the central lumen of a catheter, a circular or spiral shape is established such that the electrodes contact the endocardial surface. This allows for recording, pacing, and interrogation of electrical conduction between the left atrium and the pulmonary veins.

    AI/ML Overview

    The provided text describes the submission for a 510(k) premarket notification for the "Nordica PV Cryo Mapping Catheter" and focuses on demonstrating substantial equivalence to a predicate device.

    Crucially, this document focuses on the safety and performance of a medical device (a catheter) through various engineering, sterilization, biocompatibility, and animal studies. It explicitly states: "This section is not applicable as the subject and predicate devices do not contain any software components or utilize software for their intended use." and "Human Clinical Performance Testing: Clinical testing was not required to demonstrate substantial equivalence to the predicate device."

    Therefore, the request for acceptance criteria and a study proving a device meets the acceptance criteria in the context of AI/software performance, human readers, ground truth, and training/test sets is not applicable to this document.

    The document details the following types of testing for the physical medical device:

    1. A table of acceptance criteria and the reported device performance:

    The document provides a "Table 1: Comparison of Technological Characteristics with the Predicate Device" which implicitly serves as a comparison of the subject device's features against the predicate's known characteristics, thereby establishing the "acceptance criteria" through equivalence. The "reported device performance" is then demonstrated through various specific tests.

    Feature / Acceptance Criteria CategoryPredicate Device (Implied Acceptance)Subject Device (Reported Performance)Rationale for Substantial Equivalence
    Product CodeDRFDRFSame Product Code as Predicate Device.
    Regulation21 CFR 870.122021 CFR 870.1220Same Regulation Number as Predicate.
    Indications for Use"multiple electrode electrophysiological mapping of the cardiac structures of the heart, i.e. recording or stimulation only. The Achieve Advance mapping catheter is designed to obtain electrograms in the atrial regions of the heart.""multiple electrophysiological mapping of cardiac structures (i.e., recording and stimulation only). The Nordica PV Cryo Mapping Catheter is designed to obtain electrograms in the atrial regions of the heart."Same Indications for Use as Predicate Device.
    Intended Use"to map intracardiac structures of the heart.""to map intracardiac structures of the heart."Same Intended Use as Predicate Device.
    SterilizationEthylene Oxide SterilizedEthylene Oxide Sterilized (Validated to ISO 11135:2014+A1:2019, SAL 10^-6 or better)Same Sterilization Method as Predicate Device (with updated standard compliance).
    Single UseYesYesSame Single Use as Predicate Device.
    BiocompatibilityComplies with ISO 10993-1 (as specified in K153139)Complies with ISO 10993-1 (2018) and FDA Guidance. Tested endpoints: ISO MEM Elution, ASTM Hemolysis, SC5b-9 Complement Activation Assay, ASTM Partial Thromboplastin Time (PTT), In-Vivo Thrombogenicity Evaluation [GLP Animal Study], ISO Guinea Pig Maximization Sensitization Test, ISO Intracutaneous Study in Rabbits, ISO Material Mediated Pyrogen Study (GLP), ISO Acute Systemic Toxicity Study in Mice. All met acceptance criteria.Same Biocompatibility compliance as Predicate Device - subject device complies to current revision of the standard.
    Performance (General Standard)Per ISO 10555-1Per ISO 10555-1:2013/AMD 1:2017 (for general catheter performance)Same Performance standards as Predicate Device - subject device complies to current revision of the standard.
    Performance Testing (Specific)Joint Strength, Corrosion Resistance, Radiopacity, Radial Loop Compliance, Distal Stiffness, Axial Load, Torque Response, Electrical Continuity, Turns to Failure, Kink Resistance, Stiffness, Life Testing.Dimensional, Atraumatic Tip & Surface, Joint Strength / Peak Tensile Force, Corrosion Resistance, Radiopacity (verified in GLP animal study), Radial Loop Compliance, Distal Stiffness, Axial Load, Torque Response, Electrical Continuity, Turns to Failure, Device Functionality Kink Resistance Stiffness Life Testing. All met design input acceptance criteria.Same as predicate device - testing for subject device performed in accordance with latest versions of standards. Requirements established for these tests were validated for their intended use through V&V testing.
    Outer Diameter3F (1mm)3F (1mm)Same Outer Diameter size as predicate device.
    Effective Length146 cm149 cmEquivalent effective length; demonstrated compatibility, safe and effective use through V&V studies.
    Distal End ShapeCircular LoopCircular LoopSame Distal End Shape as predicate device.
    Loop Diameter15mm & 20mm (K162892), 25 mm (K153139)15mm & 20mmSame Loop Diameters as predicate device.
    Number of Electrodes8 (15mm loop), 10 (25mm loop)8 (15mm loop), 10 (20mm loop)Same number of electrodes as predicate device. (Note: Predicate has 25mm loop option not present in subject device, but the 15mm and 20mm subject options match predicate's corresponding electrode counts).
    Electrode SpacingUnpaired, Spacing Between Electrodes: 4.1mm (15mm loop), 5.8mm (20mm loop), 5.8mm (25mm loop)Paired Electrodes, Spacing Between Pairs: 7.3 mm (15mm loop/8 electrodes), 7.3mm (20mm loop/10 electrodes); Spacing Between Electrodes: 2 mm (15mm and 20mm loops)Similar electrode spacing to predicate device. Differences due to paired electrodes; demonstrated safety and performance through V&V testing.
    Delivery MethodDelivered Through a Delivery Catheter: YesDelivered Through a Delivery Catheter: YesSame delivery method as predicate device.
    MaterialsCatheter Body Tubing: Nitinol / Pebax (Distal), Stainless Steel / Polyimide (Proximal). Loop Material: Nitinol insulated with PET (Pebax Covered). Electrode Material: Not Specified. Tip: Not Specified.Catheter Body Tubing: Nitinol / Pebax (Distal), Stainless Steel (Proximal). Loop Material: Nitinol covered with Pebax. Electrode Material: 90% Platinum / 10% Iridium. Tip: Loctite 4311 Medical Grade Adhesive.Subject and predicate devices constructed of equivalent materials. Differences qualified through biological safety and V&V testing.
    Sterilization Residuals(Implicitly meets standards based on predicate's clearance)Met acceptance criteria below limits specified in EN ISO 10993-7:2008 + A1:2022.Compliance with updated standard.
    Shelf-Life(Implicitly demonstrated for predicate)Accelerated aging per ASTM F1980-21. Package Integrity (label, bubble leak ASTM F2096-11(2019), seal visual, seal strength ASTM F88/F88M-21) and Product V&V performed after aging.Demonstrated through testing.
    Electrical Safety / EMC(Implied compliance for predicate)Complies with IEC 60601-1-2 Ed. 4.1 (2020-09) for Class A Emissions; IEC 60601-1:2005 (and amendments); IEC 60601-1-6:2010 (and amendments) for usability.Compliance with updated standards.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Test set for Biocompatibility: "All tested subject devices met the acceptance criteria for each biological endpoint test". The exact number of devices tested for each biological endpoint (e.g., ISO MEM Elution, Hemolysis, etc.) is not specified, but it implies a sufficient sample size was used for each test to provide robust results.
    • Test set for Benchtop Studies: "The results of benchtop testing demonstrated the design outputs of the subject device met the design input acceptance criteria required to demonstrate substantial equivalence with the predicate device." Similarly, exact sample sizes for each benchtop test (e.g., buckle force, catheter deflection, etc.) are not provided.
    • Pre-Clinical GLP Animal Study: A "chronic GLP animal study" was performed on a "canine animal model". The exact sample size (number of animals) is not specified.
    • Data Provenance: The document does not specify the country of origin for the data or whether the studies were retrospective or prospective, beyond stating it was a "chronic GLP animal study" which implies a prospective, controlled study in a laboratory setting. All testing appears to be internal (Synaptic Medical Corporation) or contracted to ISO/ASTM compliant labs.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This is not applicable as the studies are engineering, biological, and animal model evaluations, not human-read AI evaluations.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. The device has no software components or AI.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the various tests, the "ground truth" is defined by the established scientific and engineering standards and methods:

    • Biocompatibility: Defined by the acceptance criteria of ISO 10993 series standards (e.g., no cytotoxicity, no hemolysis, no sensitization, no pyrogenicity, etc.).
    • Sterilization: Defined by ISO 11135:2014+A1:2019 (achieving SAL of 10^-6 or better) and EN ISO 10993-7:2008 + A1:2022 (residual limits).
    • Shelf-Life: Defined by ASTM F1980-21 for accelerated aging and specific packaging integrity tests (ASTM F2096-11(2019), ASTM F88/F88M-21).
    • Benchtop Performance: Defined by ISO 10555-1:2013/AMD 1:2017 and internal "design input acceptance criteria" for various physical properties (e.g., buckle force, tensile strength, electrical continuity).
    • Pre-Clinical GLP Animal Study: Evaluation of "chronic safety, performance, and usability" in a canine model, presumably against pre-defined physiological endpoints and observational criteria.

    8. The sample size for the training set

    Not applicable, as this device does not involve AI/machine learning.

    9. How the ground truth for the training set was established

    Not applicable, as this device does not involve AI/machine learning.

    Ask a Question

    Ask a specific question about this device

    K Number
    K233397

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-12-02

    (60 days)

    Product Code
    Regulation Number
    870.1220
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Cosine-10 Diagnostic Catheter is indicated for electrophysiological mapping (stimulation and recording) of cardiac structures, including in the coronary sinus.

    Device Description

    The Cosine-10 Diagnostic Catheter is a sterile, single-use, steerable decapolar mapping catheter with a bidirectional deflecting tip intended for electrophysiological mapping of cardiac structures (i.e., stimulation and recording). The catheter contains 10 electrodes spaced in 5 pairs, including a distal dome electrode. Two electrode spacing configurations are available: 2-8-2 (AFR-00017) or 2-5-2 (AFR-00018). All catheter electrodes may be used for recording or stimulation. The catheter can be used with a compatible introducer sheath (7.5Fr minimum).

    AI/ML Overview

    The provided text is a 510(k) Premarket Notification from the FDA for a medical device called the Cosine-10TM Diagnostic Catheter. This document primarily focuses on establishing substantial equivalence to a predicate device and does not contain the detailed acceptance criteria or the specific study results related to AI/algorithm performance that your request asks for.

    Therefore, I cannot provide the requested information, which includes:

    1. A table of acceptance criteria and the reported device performance: This document does not specify quantitative acceptance criteria or detailed performance metrics from a study related to an algorithm's diagnostic accuracy (e.g., sensitivity, specificity, AUC). It only states that "all acceptance criteria were met" for various engineering and biological tests (sterilization, biocompatibility, design verification, etc.).
    2. Sample size used for the test set and the data provenance: Not mentioned.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not mentioned.
    4. Adjudication method for the test set: Not mentioned.
    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and the effect size: Not mentioned, as this device is a physical catheter, not an AI/software.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable, as this is a physical diagnostic catheter.
    7. The type of ground truth used: Not applicable in the context of an AI study. The document mentions "electrophysiological mapping" and "stimulation and recording" as the device's function.
    8. The sample size for the training set: Not applicable, as this is not an AI/ML device.
    9. How the ground truth for the training set was established: Not applicable.

    The document describes the device, its intended use, and its similarities and differences to a predicate device (DECANAV Mapping Catheter K231312). It states that "Safety and performance testing applicable to the subject device was completed to ensure it performs as intended per product specifications and requirements/user needs." The list of completed testing includes:

    • Sterilization validation and adoption
    • Biocompatibility
    • Design verification
    • Packaging validation
    • Design validation
    • Pre-clinical animal testing

    These are standard engineering and biological tests for a physical medical device, not studies of an AI's diagnostic performance. The document concludes that "No questions of safety or effectiveness are raised as a result of the testing. The subject device is as safe and effective as the predicate device and the subject device is considered substantially equivalent to the predicate device."

    Ask a Question

    Ask a specific question about this device

    K Number
    K233207

    Validate with FDA (Live)

    Date Cleared
    2023-10-27

    (29 days)

    Product Code
    Regulation Number
    870.1220
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Polaris X catheter is indicated for use to diagnose cardiac arrhythmia.

    Device Description

    The Polaris X Unidirectional Steerable Diagnostic Catheter is a sterile, single use, unidirectional, steerable, diagnostic catheter intended for temporary use in electrophysiological studies for intracardiac stimulation (pacing) and/or recording of electrical potentials. The catheter consists of a distal electrode segment, a shaft and a handle. The distal electrode segment consists of 10 electrodes (decapolar); 1 tip electrode that is 2 mm in length and 9 ring electrodes in a variety of electrode spacings, as detailed in Table 1. The distal electrode segment is capable of forming a 270° standard curve configuration. The shaft of the Polaris X catheter has a 6 French (F) outer diameter and a useable length of 105 cm. The proximal end of the shaft is connected to a handle, which contains a patented thumbslide steering control that actuates the curve. This catheter is compatible with most commercially available recording and mapping systems and connects to these systems via a cable, which is available separately.

    AI/ML Overview

    The provided text is a 510(k) Summary for a medical device called the Polaris X™ Unidirectional Steerable Diagnostic Catheter. It describes the device, its intended use, and compares it to a predicate device to establish substantial equivalence.

    However, this document does not contain information about the device's acceptance criteria, performance against those criteria, or details of a study (such as a clinical trial or performance investigation) that proves the device meets specific acceptance criteria.

    The section titled "8. Performance Data" explicitly states "Not applicable for Special 510(k)." This indicates that for this type of submission (a Special 510(k)), the FDA did not require new performance data to demonstrate safety and effectiveness, as it relies on the substantial equivalence to a previously cleared predicate device.

    Therefore, I cannot provide the requested information based on the given text. The text does not describe an AI/Software as a Medical Device (SaMD), but rather a physical diagnostic catheter. As such, concepts like "ground truth," "multi-reader multi-case (MRMC) comparative effectiveness study," "standalone performance," "training set," or "adjudication methods" are not applicable in this context.

    Ask a Question

    Ask a specific question about this device

    K Number
    K232651

    Validate with FDA (Live)

    Date Cleared
    2023-09-29

    (29 days)

    Product Code
    Regulation Number
    870.1220
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Blazer Dx-20 Catheter is indicated for use to diagnose cardiac arrhythmia.

    Device Description

    The Blazer Dx-20 Bidirectional Steerable Diagnostic Catheter is a sterile, single use, steerable, diagnostic catheter intended for temporary use in electrophysiological studies for intracardiac stimulation (pacing) and/or recording of electrical potentials. The catheter is provided in two different bidirectional curve configurations, Medium (25 mm diameter) and Super Large (56 mm diameter), to assist the physician in reaching the desired anatomical location.

    The distal electrode segment of the catheter contains an electrode tip and 19 electrode rings which are designed to carry electrical signals for endocardial stimulation (pacing) and/or recording. The electrode rings are placed on the distal segment of the catheter in various electrode spacing configurations, as described in Table 1. A 7F torqueable shaft connects the distal electrode segment to an ergonomically designed cylindrical handle.

    The catheter handle contains both a steering knob and tension control knob. The degree of tip deflection of the catheter is controlled by the steering knob. The adjustable tension control knob can be tightened to hold the tip in the desired position

    This catheter is compatible with most commercially available recording and mapping systems and connects to these systems via a cable, which is available separately.

    AI/ML Overview

    The provided document describes a 510(k) premarket notification for the BLAZER™ Dx-20 Bidirectional Steerable Diagnostic Catheter. This type of submission, particularly a "Special 510(k)," is often used when modifications to a previously cleared device do not affect its fundamental safety or effectiveness.

    Based on the information given, the document does not contain the detailed performance study information requested in the prompt. The text explicitly states:

    • "8. Performance Data: Not applicable for this Special 510(k)."

    This indicates that clinical performance data and acceptance criteria for a new study were not required for this particular regulatory submission because the device modifications were minor ("The proposed labeling modifications of the Blazer Dx-20 catheter do not impact the device's substantial equivalence to the previously cleared version of this device."). The submission primarily relies on the substantial equivalence to a predicate device (K211375).

    Therefore, it is not possible to fill out a table of acceptance criteria and reported device performance or describe a study that proves the device meets the acceptance criteria using only the provided text. The document does not describe:

    • A table of acceptance criteria and reported device performance.
    • Sample sizes for test sets, data provenance.
    • Number of experts, their qualifications, or adjudication methods for ground truth.
    • Whether MRMC studies were done, or their effect sizes.
    • Standalone algorithm performance.
    • Type of ground truth used.
    • Sample size for training sets or how ground truth was established for training sets (as this is a medical device, not an AI/ML algorithm requiring training sets).

    The document is a regulatory notice for a physical medical device, not an AI/ML algorithm, which explains why many of the questions (especially those related to AI model training and ground truth establishment) are not applicable. The primary evidence presented is the substantial equivalence to a predicate device, not new performance study data.

    Ask a Question

    Ask a specific question about this device

    K Number
    K230503

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-04-25

    (60 days)

    Product Code
    Regulation Number
    870.1220
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Blazer Dx-20 Catheter Cable:

    The BSC Cables are intended to be used with the BSC Cardiac Diagnostic Catheters during an electrophysiology procedure for intracardiac stimulation (pacing) and/or recording of electrical potentials.

    Polaris X Catheter Cable:

    The BSC Cables are intended to be used with the BSC Cardiac Diagnostic Catheters during an electrophysiology procedure for intracardiac stimulation (pacing) and/or recording of electrical potentials.

    Device Description

    Blazer™ Dx-20 Cable:

    The subject Blazer Dx-20 Catheter cable (sold separately) is used with the Blazer Dx-20 Bidirectional Duodecapolar Catheter during an electrophysiology procedure. The Blazer Dx-20 Catheter Cable provides an electrical connection between the Blazer Dx-20 Diagnostic Catheter and the Electrophysiology (EP) Recording System. The cable has one connector that directly connects to the diagnostic catheter and multiple connection points that connect to an EP Recording System.

    The main trunk consists of a jacketed 20-conductor cable that is split into individual leads by means of an over-molded splitter yoke. The proximal end of the cable has 20 universal 2 mm shrouded male pins that are typical to electrophysiology recording and stimulation systems. The Blazer Dx-20 cable is provided sterile using an EO sterilization process. The cable is reusable and can be reprocessed (i.e. re-sterilized) up to 10 times using either EO or STERRAD™ sterilization methods. The Blazer Dx-20 cable is to be used in a fully-equipped electrophysiology lab by physicians thoroughly trained in invasive cardiac mapping and ablation procedures.

    Polaris X™ Cable:

    The subject Polaris X Catheter Cable (sold separately) is used with the Polaris X Steerable Decapolar Mapping Catheters during an electrophysiology procedure. The Polaris X cable provides an electrical connection between the Polaris X Catheter and the Electrophysiology (EP) Recording System. The cable has one connector that directly connects to the diagnostic catheter and multiple connection points that connect to an EP Recording System.

    The main trunk consists of a jacketed 10-conductor cable that is split into individual leads by means of an over-molded splitter yoke. The proximal end of the cable has 10 universal 2 mm shrouded male pins that are typical to electrophysiology recording and stimulation systems. The Polaris X cable is provided sterile using an EO sterilization process. The cable is reusable and can be reprocessed (i.e. re-sterilized) up to 10 times using EO sterilization method. The Polaris X cable is to be used in a fully-equipped electrophysiology lab by physicians thoroughly trained in invasive cardiac mapping and ablation procedures.

    AI/ML Overview

    The provided document, a 510(k) summary for the Blazer™ Dx-20 Catheter Cable and Polaris X™ Catheter Cable, does not contain information about acceptance criteria or a study proving the device meets those criteria in the context of AI/ML device performance. The document describes the device, its intended use, comparison with predicate devices, and performance testing related to bioburden, reusability (manual cleaning and resterilization), and general design verification.

    Therefore, I cannot provide the detailed information requested in the prompt based on the provided text, as it pertains to medical device regulatory clearance for cables used in electrophysiology procedures, not an AI/ML powered device.

    Specifically, the document does not include:

    1. A table of acceptance criteria and reported device performance for an AI/ML algorithm. The performance data section refers to bioburden testing, reusable device testing (cleaning and resterilization), and design verification testing for the physical cables.
    2. Sample sizes used for a test set or data provenance related to AI/ML.
    3. Number of experts or their qualifications for establishing ground truth for an AI/ML test set.
    4. Adjudication methods for an AI/ML test set.
    5. Information about a multi-reader multi-case (MRMC) comparative effectiveness study or human reader improvement with AI assistance.
    6. Standalone (algorithm only) performance data.
    7. Type of ground truth used for AI/ML (expert consensus, pathology, outcomes data, etc.).
    8. Sample size for an AI/ML training set.
    9. How ground truth for an AI/ML training set was established.

    The document primarily focuses on demonstrating the substantial equivalence of the new catheter cables to existing predicate devices based on physical and functional characteristics, sterility, and reprocessing capabilities.

    Ask a Question

    Ask a specific question about this device

    K Number
    K210313

    Validate with FDA (Live)

    Date Cleared
    2022-12-30

    (695 days)

    Product Code
    Regulation Number
    870.1220
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    EasyFinder™ Fixed Curve Diagnostic Catheter is used for electrogram recording and cardiac stimulation during diagnostic electrophysiology studies.

    Device Description

    The EasyFinder™ Fixed Curve Diagnostic Catheter is a sterile, single-use diagnostic catheter with multiple electrodes and a fixed distal curve. It has been designed to facilitate electrophysiological mapping of cardiac structures in 28 models and 6 kinds of curves.

    The catheter has a high-torque shaft with a fixed curve section containing an array of platinum electrodes. The high-torque shaft allows the plane of the curved tip to be rotated to facilitate accurate positioning of the catheter tip at the desired site.

    The catheter is connected to separately cleared diagnostic electrophysiology equipment via an accessory, the FORLNK™ Cable for Mapping Catheter.

    AI/ML Overview

    The provided document, a 510(k) summary for the EasyFinder™ Fixed Curve Diagnostic Catheter, indicates that no clinical study was included in the submission. Therefore, information regarding acceptance criteria derived from a clinical study, sample sizes for test and training sets, expert qualifications, adjudication methods, MRMC studies, or standalone algorithm performance, as requested, is not available from this document.

    The substantial equivalence determination for this device is based on non-clinical performance testing and comparison to a predicate device.

    Here's the information that can be extracted:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't explicitly define "acceptance criteria" in a quantitative sense as might be seen for a new AI/software device. Instead, it states that the device complies with specific industry standards and regulatory guidance, which serve as the "acceptance criteria" for demonstrating safety and effectiveness. The "reported device performance" is its compliance with these standards.

    Acceptance Criterion (Standard/Guidance)Reported Device Performance
    Biocompatibility Verification: ISO 10993-1:2018 and FDA's modified ISO guidelinesThe biological safety of the catheter was verified as per the requirements.
    Mechanical Verification: ISO 10555-1:2013Mechanical testing was performed to verify compliance of the catheter with the standard.
    Electrical Verification: IEC 60601-1:2005, IEC 60601-1-2:2014, and IEC 60601-2-27:2011Electrical testing was performed to verify compliance of the catheter and accessory cable with applicable standards.
    Bench Validation: Radiopacity of catheter; design and function of accessory cableValidation testing was performed for radiopacity and to evaluate the design and function of the cable.
    Cleaning and Sterilization Validation: ISO 11135:2014, AAMI TIR 28:2009, and FDA guidance documentValidation testing was performed to demonstrate that the catheter and accessory cable could be cleaned and sterilized in accordance with the standards. -------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
    Shelf Life Validation: Demonstrate shelf lifeValidation testing was performed to demonstrate a shelf life of three years.
    Indications for Use Equivalence: Same as predicate deviceEasyFinder™ Fixed Curve Diagnostic Catheter is used for electrogram recording and cardiac stimulation during diagnostic electrophysiology studies (SE to predicate).
    Configuration Equivalence: Cable connector, Tip electrode, Ring electrode, High-torque shaft (Same as predicate device)The device has a cable connector, tip electrode, ring electrode, and high-torque shaft (SE to predicate).
    Dimensional Comparison: Catheter effective length, Length of color code, Electrode length, Electrode diameter, Electrode spacingCatheter effective length (Similar to predicate), Length of color code (SE to predicate), Electrode length (Similar to predicate), Electrode diameter (SE to predicate), Electrode spacing (SE to predicate).
    Outside diameter of catheter shaft: Equivalent to predicate deviceOutside diameter of catheter shaft (SE to predicate).
    Curve shape: Similar to predicate deviceCurve shape (Similar to predicate).
    Functional Performance: Comply with ISO 10555-1Comply with ISO 10555-1 (SE to predicate).
    Biological Characteristics: Comply with ISO 10993-1Comply with ISO 10993-1 (Similar to predicate).
    Intended Healthcare Environment: Professional healthcare environmentProfessional healthcare environment (SE to predicate).
    Sterilization Method: EO sterilizedEO sterilized (SE to predicate).
    Single Use: YesYes (SE to predicate).
    Shelf Life: 3 years3 years (SE to predicate).

    2. Sample size used for the test set and the data provenance:

    • Not Applicable. No clinical study or separate test set in the context of an AI/software device performance evaluation was conducted or reported. Performance was based on non-clinical (bench) testing against established standards and comparison to a predicate device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not Applicable. No clinical study or expert-adjudicated test set in the context of an AI/software device performance evaluation was conducted or reported.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not Applicable. No expert adjudication was conducted or reported.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not Applicable. This is a hardware medical device (diagnostic catheter), not an AI-assisted diagnostic tool for human readers. No MRMC study was conducted or reported.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Not Applicable. This is a hardware medical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Not Applicable. For the non-clinical tests, the "ground truth" was established by the specifications defined in the referenced international standards (e.g., ISO, IEC). For equivalence to the predicate, the characteristics of the predicate device served as the reference.

    8. The sample size for the training set:

    • Not Applicable. No training set, in the context of an AI/software device, was used or reported.

    9. How the ground truth for the training set was established:

    • Not Applicable. No training set was used.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 13