Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K013538

    Validate with FDA (Live)

    Device Name
    ETHANOL
    Date Cleared
    2001-12-14

    (51 days)

    Product Code
    Regulation Number
    862.3040
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Ethanol assay is used for the quantitative analysis of ethyl alcohol (ethanol) in human serum, plasma, or urine. The Ethanol assay should be exclusively used to detect ethanol and not other alcohols such as isopropanol or methanol. Reactivity with compounds structurally unrelated to ethanol has not been observed.

    Device Description

    Ethanol is an in vitro diagnostic assay for the quantitative analysis of ethyl alcohol (ethanol) in human serum, plasma, or urine. The assay is based on an enzymatic reaction. Active enzyme converts NAD to NADH, resulting in an absorbance change that can be measured spectrophotometrically. The increase in absorbance at 340 nm is proportional to the concentration of alcohol in the specimen.

    AI/ML Overview

    This submission describes the "Ethanol" assay, an in vitro diagnostic device for the quantitative analysis of ethyl alcohol in human serum, plasma, or urine. The study presented aims to demonstrate substantial equivalence to a predicate device, the Emit® II Plus Ethyl Alcohol assay (K993980).

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Predicate Performance)Reported Device Performance (Ethanol Assay)
    Correlation with Predicate Device (Emit® II Plus Ethyl Alcohol assay on SYVA 30R Analyzer)Acceptable correlation demonstrated.
    - Serum: Not explicitly stated as a numerical criterion, but the predicate's performance implies a high correlation is expected.- Serum: Correlation coefficient = 0.996, slope = 1.00, Y-intercept = -1.87 mg/dL
    - Urine: Not explicitly stated as a numerical criterion, but the predicate's performance implies a high correlation is expected.- Urine: Correlation coefficient = 0.997, slope = 1.01, Y-intercept = -4.85 mg/dL
    Precision (Implied from predicate or general assay expectations)Precision studies conducted.
    - Within-run Total %CV: Implied to be low for similar assays.- Low control: 5.1% Total %CV
    - Within-run Total %CV: Implied to be low for similar assays (e.g., around 1-5% for calibrators/controls).- 100 mg/dL calibrator/control: 2.8% Total %CV
    - Within-run Total %CV: Implied to be low for similar assays.- High control: 1.7% Total %CV
    Assay Range (Implied from predicate or general assay expectations)10 to 600 mg/dL
    Limit of Quantitation (Sensitivity) (Implied from predicate or general assay expectations)5 mg/dL

    Note: The acceptance criteria are not explicitly defined as pass/fail thresholds in the provided text but are inferred from the need to demonstrate "substantially equivalent performance" to the predicate device. The presented performance characteristics are typical metrics for demonstrating equivalence in IVD assays.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the specific number of human serum or urine samples used in the comparative performance studies to generate the correlation data (correlation coefficients, slopes, and y-intercepts). It only mentions "Comparative performance studies were conducted." The data provenance (e.g., country of origin, retrospective or prospective) is also not specified.

    For the precision studies, "three levels of control material" were used for the within-run study. The number of replicates for each level is not provided.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This type of information is not applicable to this submission. The "Ethanol" assay is an in vitro diagnostic (IVD) test that quantifies a biomarker (ethyl alcohol concentration). The "ground truth" for such devices is typically established by comparing its performance to a reference method or another legally marketed device (the predicate device in this case), or by using known concentrations in control materials. It does not involve interpretation by human clinicians or experts for ground truth establishment in the same way an imaging or pathology device might require.

    4. Adjudication Method for the Test Set

    The concept of an "adjudication method" (e.g., 2+1, 3+1) is not applicable here. Adjudication is used to resolve discrepancies in independent interpretations by multiple experts, typically for qualitative or semi-quantitative diagnostic tests where human judgment is involved. For a quantitative IVD assay like the Ethanol assay, the comparison is directly between the numerical results of the new device and the predicate device (or known concentrations).

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size of Human Improvement

    A Multi-Reader Multi-Case (MRMC) comparative effectiveness study is not applicable to this device. MRMC studies are used to evaluate the impact of an AI-powered diagnostic aid on human reader performance (e.g., radiologists interpreting images). The "Ethanol" assay is an automated in vitro diagnostic test for quantifying a biomarker and does not involve human interpretation in the diagnostic process beyond performing the test and reviewing results. Therefore, there is no "human reader improvement with AI vs without AI assistance" to measure.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    The entire study presented is a standalone performance evaluation of the "Ethanol" assay. The device itself is an automated enzymatic assay. The performance characteristics (correlation, precision, range, sensitivity) are inherently "standalone" in that they describe the capability of the device itself to accurately measure ethanol concentration, without requiring active human interpretation of its diagnostic output (beyond standard laboratory procedures).

    7. The Type of Ground Truth Used

    The ground truth used for the comparative performance studies was the results obtained from the predicate device, the Emit® II Plus Ethyl Alcohol assay on the SYVA 30R Analyzer. For the precision studies, the ground truth was the known concentrations of the control materials (low control, 100 mg/dL calibrator/control, and high control).

    8. The Sample Size for the Training Set

    The document does not specify a training set sample size. This submission describes an in vitro diagnostic that uses an enzymatic reaction. This type of assay does not typically involve machine learning algorithms that require "training sets" in the conventional sense (i.e., large datasets used to train a model). Instead, it relies on established biochemical principles and calibration procedures.

    9. How the Ground Truth for the Training Set Was Established

    As noted above, this device does not utilize a training set in the machine learning context. The "ground truth" relevant to the development and calibration of such an enzymatic assay would historically involve:

    • Known concentrations of analyte: Prepared standards and calibrators with precisely measured ethanol concentrations.
    • Reference methods: Historical or established reference methods for measuring ethanol.

    These elements would be used during the assay's development and manufacturing to establish its analytical performance characteristics and ensure it accurately measures ethanol. The specific details of how these were established are not provided in this 510(k) summary, as it is focused on demonstrating equivalence to the predicate, not the de novo development process.

    Ask a Question

    Ask a specific question about this device

    K Number
    K993980

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2000-02-01

    (69 days)

    Product Code
    Regulation Number
    862.3040
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Emit® II Plus Ethyl Alcohol Assay is an enzyme assay intended for use in the quantitative analysis of ethyl alcohol in human serum or plasma.
    The Emit® II Plus Ethyl Alcohol Assay is an enzyme assay intended for use in the quantitative analysis of ethyl alcohol in human urine, plasma, and serum. The Emit® II Plus Ethyl Alcohol Assay is designed for use with most clinical chemistry analyzers.

    Device Description

    The Syva Emit® II Plus Ethyl Alcohol Assay is an enzyme assay intended for use in quantitative analysis of ethyl alcohol in human urine, serum, or plasma.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study details for the Emit® II Plus Ethyl Alcohol Assay, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Study CategoryAcceptance Criteria (Implied)Reported Device Performance
    Comparative AnalysisExcellent correlation (high correlation coefficient, slope near 1) to predicate device.Serum: Correlation of 0.986 with predicate method, slope of 1.000.
    Urine: Correlation of 0.997 with predicate method, slope of 1.000.
    PrecisionAcceptable within-run and total precision (low %CV).Within-run: %CV ranging from 1.39% to 1.55%.
    Total precision: %CV ranging from 1.61% to 1.78%.
    SpecificityAcceptable levels of cross-reactivity (low percentage).Cross-reactivity ranging from <0.1% to 14.2% for potentially cross-reacting compounds.
    SensitivityAbility to distinguish low concentrations from zero.Less than 3.0 mg/dL (lowest concentration of amphetamines that can be distinguished from 0 ng/mL with 95% confidence – Note: The text explicitly mentions "amphetamines" here, which seems like a typo given the Ethyl Alcohol assay. Assuming it means ethyl alcohol, the principle holds.)
    Endogenous InterferenceHigh recovery compared to controls (minimal interference).Serum: Bilirubin: 99% recovery, Hemoglobin: 100% recovery, Triglycerides: 97% recovery. LDH/Lactate: 0.4 mg/dL interference (not significant).
    Urine: Recoveries for interfering substances ranging from 99% to 104%.
    Spiked Sample RecoveryRecovery within an acceptable range (e.g., 90-110%).Recovery of ethanol spikes in human serum and urine ranged between 95.7% and 101.5% across a quantitative range of 10 to 600 mg/dL.
    High Sample DilutionRecovery within an acceptable range after dilution.Recovery from normal concentrations in diluted serum and urine (with water or Negative Calibrator) ranged from 97.1% to 100.9%.
    AnticoagulantsAverage recovery compared to serum control within an acceptable range.Average recovery versus serum control ranged from 95.5% to 102.3% for EDTA (K3), Sodium Citrate, Sodium Heparin, and Oxalate/Fluoride.
    LinearityConsidered linear according to CAP linearity survey acceptance criteria.Both serum and urine samples are considered linear according to the acceptance criteria of the CAP linearity survey.
    Urine PreservativesRecoveries compared to controls within an acceptable range.Recoveries of urine preservatives compared to controls ranged from 97.5% to 99.1%.
    Calibration StabilityAcceptable duration of calibration stability.25 days on the SYVA-30R Biochemical System.

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not explicitly state the sample sizes used for most of the individual tests. It mentions "human serum or plasma" and "human urine, serum, or plasma" as the sample types.

    • Provenance: The document does not specify the country of origin of the data. It is generally implied to be from the United States given the FDA submission.
    • Retrospective/Prospective: The document does not specify whether the data was retrospective or prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    This information is not provided in the document. The studies performed are laboratory-based analytical performance studies, not clinical studies requiring expert interpretation of results for ground truth establishment in the traditional sense. The ground truth for these types of assays is established through reference methods or known concentrations, not expert consensus.

    4. Adjudication Method for the Test Set:

    This information is not applicable as the studies are analytical performance assessments of an in vitro diagnostic device, not studies involving human interpretation or adjudication of cases.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

    No, an MRMC comparative effectiveness study was not done. This type of study is relevant for imaging or diagnostic tools where human readers interpret results, often with and without AI assistance. The Emit® II Plus Ethyl Alcohol Assay is an enzyme assay, which is an automated chemical analysis, not subject to human reader interpretation in the same way.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    Yes, the studies described are essentially standalone performance studies for the Emit® II Plus Ethyl Alcohol Assay. The device operates as an algorithm/assay without human intervention in the interpretation of the assay result itself. Its performance is measured directly against established analytical metrics and a predicate device.

    7. The Type of Ground Truth Used:

    The ground truth for these analytical studies is primarily established through:

    • Reference Methods/Predicate Device: For comparative analysis, the predicate device (Emit® Ethyl Alcohol Assay K903153) served as the reference standard.
    • Known Concentrations/Spiked Samples: For precision, sensitivity, spiked sample recovery, endogenous interference, high sample dilution, anticoagulant, and urine preservative studies, the ground truth would be the known concentrations or expected values of ethyl alcohol or interferents.
    • CAP Linearity Survey Acceptance Criteria: For linearity, the ground truth/standard is defined by these external criteria.

    8. The Sample Size for the Training Set:

    This information is not applicable/not provided. The Emit® II Plus Ethyl Alcohol Assay is an enzyme assay, which is a chemical reaction-based method, not a machine learning or AI algorithm that requires a "training set" in the computational sense. Its parameters are chemically and biochemically defined, not iteratively learned from data.

    9. How the Ground Truth for the Training Set Was Established:

    This information is not applicable. As stated above, this device does not utilize a training set in the way an AI/ML algorithm would. The "ground truth" equivalent for developing such an assay would involve established biochemical principles, reagent formulations, optimization experiments, and quality control standards.

    Ask a Question

    Ask a specific question about this device

    K Number
    K963197

    Validate with FDA (Live)

    Device Name
    ON.SITE ALCOHOL
    Manufacturer
    Date Cleared
    1996-09-19

    (35 days)

    Product Code
    Regulation Number
    862.3040
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ON•SITE Alcohol is recommended for professional use only and is not intended for overthe-counter sale to the general public. Professional use includes applications in healthcare, corrections and drug treatment.

    ON•SITE Alcohol yields only a "qualitative" result. A positive result indicates that alcohol (ethanol) is present at a concentration of approximately 0.02% when performed according to the instructions.

    Confirmation of ON•SITE Alcohol results using gas chromatography is recommended if quantitative results are desired.

    Device Description

    ON•SITE Alcohol is a self contained, disposable enzymatic test device that provides rapid detection of alcohol in saliva or urine. It includes all reagents and supplies necessary to perform the test.

    AI/ML Overview

    Here's an analysis of the provided text, focusing on the acceptance criteria and the study details:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are not explicitly stated in a quantitative form, but rather implied by the design intention and the performance characteristics observed in the 'modified' device compared to the 'current' device. The key goal was to increase the reaction time and slightly raise the limit of detection.

    Criteria (Implied)Current Device PerformanceModified Device PerformanceTarget/Acceptance (Implied)
    Accuracy & Precision (Detection at 0.008% w/v EtOH)13/20 positives in 2 mins, Mean Reaction Time: 111 sec0/20 positives in 2 mins, Mean Reaction Time: 147 secReduced detection at 0.008% (i.e., fewer positives at 2 mins, increased reaction time)
    Accuracy & Precision (Detection at 0.02% w/v EtOH)20/20 positives in 2 mins, Mean Reaction Time: 81 sec12/20 positives in 2 mins, Mean Reaction Time: 115 secMaintain detection at 0.02% (or slightly reduced as a side effect of shifted LOD), increased reaction time
    Accuracy & Precision (Detection at 0.032% w/v EtOH)20/20 positives in 2 mins, Mean Reaction Time: 61 sec20/20 positives in 2 mins, Mean Reaction Time: 100 secMaintain detection at 0.032%, increased reaction time
    Sensitivity (Reaction Times across various EtOH concentrations)Range of mean reaction times from 1 min 6 sec to 1 min 47 secRange of mean reaction times from 1 min 29 sec to 2 min 41 secConsistently increased reaction times across all tested concentrations

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size:
      • Accuracy and Precision: 20 replicates for each of the three ethanol controls (0.008%, 0.02%, 0.032% w/v). This means 20x3 = 60 tests for the current device and 60 tests for the modified device, totaling 120 tests for this part.
      • Sensitivity: 10 replicates for each of the eight ethanol controls (0.005% to 0.0225% w/v). This means 10x8 = 80 tests for the current device and 80 tests for the modified device, totaling 160 tests for this part.
    • Data Provenance: Not specified within the document (e.g., country of origin, type of samples used - saliva/urine). Given the context of a 510(k) summary, it's implied to be laboratory testing of controlled ethanol solutions, not human samples. It is a prospective laboratory study comparing a modified device to its predicate.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    Not applicable. The ground truth was established by laboratory-prepared ethanol solutions of known concentrations. The "results were read" likely implies visual interpretation by a technician, but their number or qualifications are not specified as it's not an expert-driven ground truth.

    4. Adjudication Method for the Test Set

    Not applicable. The outcomes were objective measurements (time to positive result) and counts of positive/negative results at a specific time point, based on predefined ethanol concentrations. No adjudication process for differing interpretations of subjective data would be needed.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC study was not done. This study is a technical performance comparison between two versions of a diagnostic device in a laboratory setting, not a study involving human readers interpreting cases.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, this was a standalone performance study. The device's reaction to known concentrations of alcohol was evaluated directly, without human interpretation of results beyond simple visual observation of a color change, and timing that change. There is no "algorithm" in the modern sense relevant to AI.

    7. The Type of Ground Truth Used

    The ground truth used was objectively known concentrations of ethanol in laboratory-prepared control solutions.

    8. The Sample Size for the Training Set

    Not applicable. This device is an enzymatic test, not an AI/machine learning model, so there is no training set in the computational sense. The "training" for the device's development would have been implicit in chemical formulation and calibration, not a data-driven training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set in the context of an AI/ML model. If interpreted very broadly as the development process for the chemical formulation, the ground truth would still be based on known chemical reactions and desired performance characteristics to achieve the intended "shift in reaction time and detection levels."

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1