510(k) Data Aggregation

    K Number
    K172903

    Validate with FDA (Live)

    Date Cleared
    2017-11-22

    (61 days)

    Product Code
    Regulation Number
    864.7320
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    HemosIL D-Dimer HS 500 is an automated latex enhanced immunoassay for the quantitative determination of D-Dimer in human citrated plasma on the ACL TOP® Family and ACL TOP Family 50 Series Systems for use, in conjunction with a clinical pretest probability (PTP) assessment model to exclude venous thromboembolism (VTE) in outpatients suspected of deep venous thrombosis (DVT) and pulmonary embolism (PE). For in vitro diagnostic use.

    Device Description

    The D-Dimer HS 500 Latex Reagent is a suspension of polystyrene latex particles of uniform size coated with the F(ab')2 fragment of a monoclonal antibody highly specific for the D-Dimer domain included in fibrin soluble derivatives. The use of the F(ab')2 fragment allows a more specific D-Dimer detection avoiding the interference of some endogenous factors like the Rheumatoid Factor. When a plasma containing D-Dimer is mixed with the Latex Reagent and the Reaction Buffer included in the HemosIL D-Dimer HS 500 kit, the coated latex particles agglutinate. The degree of agglutination is directly proportional to the concentration of D-Dimer in the sample and is determined by measuring the decrease of transmitted light caused by the aggregates (turbidimetric immunoassay).

    AI/ML Overview

    The provided document is a 510(k) summary for the HemosIL D-Dimer HS 500 device. This particular submission (K172903) is a Special 510(k), meaning it's for modifications to an already cleared device (predicate K090264) and not for a de novo clearance. Special 510(k)s typically do not require new clinical studies to demonstrate device performance against acceptance criteria, as the performance claims are carried over from the predicate device.

    The main purpose of K172903 is to add general information from peer-reviewed published literature to the Summary and Principle section of the HemosIL D-Dimer HS 500 insert sheet regarding the association of patient age with D-Dimer levels and to clarify that the device's performance has not been validated for age-adjusted cut-off values.

    Therefore, many of the requested items regarding a new clinical study and ground truth establishment would not be applicable to this specific submission. However, I can extract the acceptance criteria and performance claims that apply to the predicate device and are carried over to this modified device.

    Here's the breakdown based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    Since this is a Special 510(k) for an insert sheet modification and not a new device, the acceptance criteria and reported performance are implicitly the same as the predicate (K090264). The document explicitly states "No change to labeled performance claims, including no change to the assay cut-off."

    Performance CriterionAcceptance Criteria (from predicate K090264)Reported Device Performance (for K172903, same as predicate)
    Cut-off500 ng/mL500 ng/mL
    Linearity215 - 128000 ng/mL215 - 128000 ng/mL
    Detection Limit203 ng/mL203 ng/mL
    Indications for UseAutomated latex enhanced immunoassay for quantitative determination of D-Dimer in human citrated plasma on ACL TOP Family and ACL TOP Family 50 Series Systems for use, in conjunction with a clinical pretest probability (PTP) assessment model to exclude VTE in outpatients suspected of DVT and PE.Same as predicate.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided in this Special 510(k) summary (K172903) because no new clinical test set was used or presented for this particular submission. The performance claims are based on the predicate device (K090264). To obtain this information, one would need to refer to the 510(k) summary for K090264.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not provided in this Special 510(k) summary (K172903) as no new clinical test set (requiring expert ground truth) was used for this submission. This type of detail would be found in the original 510(k) for the predicate device (K090264) which established the clinical performance.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not provided in this Special 510(k) summary (K172903) as no new clinical test set requiring adjudication was used for this submission. This type of detail would be found in the original 510(k) for the predicate device (K090264).

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This device is an in vitro diagnostic for D-Dimer measurement, not an imaging device typically analyzed by human readers in the context of MRMC studies or AI assistance in reading. Therefore, an MRMC comparative effectiveness study, as described, is not applicable.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    The HemosIL D-Dimer HS 500 is an automated immunoassay for quantitative determination. Its performance is inherently "standalone" in the sense that it provides a quantitative result based on the immunoassay reaction, without direct human cognitive input in interpreting the raw signal for the D-Dimer concentration. The result is then used in conjunction with a clinical pretest probability (PTP) assessment model by a clinician. The validation of its quantitative accuracy would have been performed during the original clearance (K090264).

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    For an in vitro diagnostic device like a D-Dimer assay, the ground truth for performance measures like linearity, detection limit, and cut-off would typically be established by:

    • Reference methods/calibrators: For analytical performance (linearity, detection limit), certified reference materials or highly accurate reference methods would be used.
    • Clinical outcomes/diagnosis: For the diagnostic performance (sensitivity, specificity, NPV for VTE exclusion), the ground truth for DVT/PE would be established through definitive diagnostic imaging (e.g., ultrasound, CTPA, ventilation-perfusion scan) or other established clinical diagnostic criteria for the patient cohort recruited for the original predicate study (K090264).

    This specific Special 510(k) (K172903) does not provide these details as it relies on the predicate (K090264).

    8. The sample size for the training set

    This information is not provided in this Special 510(k) summary (K172903). For an immunoassay, the concept of a "training set" in the context of machine learning (as often implied by this question) is not directly applicable. However, method development and optimization would involve numerous samples, with calibration curves and reagents being "trained" or optimized during the development phase of the original predicate device (K090264), but described in terms of analytical validation rather than a machine learning 'training set'.

    9. How the ground truth for the training set was established

    As above, while not a "training set" in the common AI sense, the ground truth for optimizing assay parameters and establishing analytical performance (for the original predicate K090264) would have been established using reference standards or established, validated methods. Clinical training data, if any, for setting the original cut-off would have involved patients with confirmed (or ruled out) DVT/PE, as validated by definitive diagnostic tests. This information is not within this K172903 document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K162227

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-12-10

    (124 days)

    Product Code
    Regulation Number
    864.7320
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The STA® - Liatest® D-Di kit is an immuno-turbidimetric assay for the quantitative determination of D-dimer in venous plasma (3.2% sodium citrate) for use on STA-R®, STA Compact® and STA Satellite® analyzers by professional laboratory personnel. The STA® - Liates® D-Di is intended for use in conjunction with a clinical pretest probability (PTP) assessment model to exclude pulmonary embolism (PE) and deep venous thrombosis (DVT) in outpatients suspected of PE or DVT.

    Device Description

    STA® - Liatest® D-Di kit contains: 6 x 5-ml vials of ready-for-use Tris buffer and 6 x 6-ml vials of a suspension of microlatex particles coated with two different mouse monoclonal anti-human D-dimer antibodies (8D2 and 2.1.16) stabilized with bovine albumin.

    The test principle is based on the change in turbidity of a microparticle suspension that is measured by photometry. A suspension of latex microparticles, coated by covalent bonding with monoclonal antibodies specific for D-dimer is mixed with the test plasma for which the D-dimer level is to be assayed. An antigen-antibody reaction takes place, leading to an agglutination of the latex microparticles which causes an increase in turbidity of the reaction medium. This increase in turbidity is reflected by an increase in absorbance, the latter being measured photometrically. The increase in absorbance is a function of the D-dimer level present in the test sample.

    AI/ML Overview

    This document describes the acceptance criteria and study proving the performance of the STA® - Liatest® D-Di kit, an immuno-turbidimetric assay for the quantitative determination of D-dimer.

    1. Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the STA® - Liatest® D-Di kit were based on achieving specific Negative Predictive Value (NPV) and Sensitivity targets for the exclusion of Pulmonary Embolism (PE) and Deep Venous Thrombosis (DVT) in outpatients with low to moderate pretest probability (PTP).

    MetricAcceptance Criteria (95% CI)Reported Device Performance (95% CI) - Overall Study Population
    NPV (for DVT)≥ 99.0% (as per CLSI H59-A requirements)100.0% (99.3% - 100%)
    Sensitivity (for DVT)≥ 95.0% (as per CLSI H59-A requirements)100.0% (95.8% - 100%)

    Note: The document explicitly states that the results "meet both confirmatory hypotheses relating to NPV and Sensitivity." While the predicate device's PE and DVT results are listed in the "Similarities Chart," the detailed performance data for the new device (STA® - Liatest® D-Di) in the clinical performance section is specifically for DVT exclusion. The indications for use state "to exclude pulmonary embolism (PE) and deep venous thrombosis (DVT)", but the detailed study results presented focus on DVT.

    2. Sample Size and Data Provenance

    • Test Set Sample Size: 980 samples of patients with a low or moderate PTP were included in the primary efficacy analyses. This consisted of 79 suspects of DVT and PE, and 901 suspects of DVT only. Of the 980 samples, 85 were DVT positive and 895 were DVT negative.
    • Data Provenance: The data was collected from a prospective, multi-center clinical study conducted at 16 sites across the United States, Europe, and Canada. The patients were consecutive, ambulatory outpatients presenting at emergency units or outpatient clinics suspected of having venous thromboembolism (VTE).

    3. Number of Experts and Qualifications for Ground Truth

    The document does not explicitly state the "number of experts" or their "qualifications" in the context of establishing ground truth in the traditional sense of human readers adjudicating medical images. Instead, the ground truth was established through a clinical protocol:

    • Patients with positive D-dimer results were considered for an imaging procedure (e.g., ultrasound, CT angiography).
    • Patients with negative D-dimer results were assigned to a three-month follow-up to confirm the absence of DVT/PE.
    • The "Reference" column in the results tables (Tables 1, 2) is a combination of "imaging or 3-month follow-up."

    Therefore, the establishment of ground truth was based on clinical diagnostic pathways and follow-up, rather than expert interpretation of a specific dataset for the purpose of algorithm validation in the way one might see for an imaging AI. The "experts" in this context would be the clinicians and radiologists involved in the standard of care diagnostic work-up.

    4. Adjudication Method for the Test Set

    Not applicable in the typical sense of expert adjudication of AI outputs. The ground truth was established by standard clinical practice: a positive D-dimer result led to imaging, and a negative D-dimer result led to a 3-month clinical follow-up for confirmation.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC study was done, as this is an in-vitro diagnostic (D-dimer assay), not an imaging AI diagnostic device intended to assist human readers. The study evaluates the standalone performance of the assay in a clinical pathway.

    6. Standalone Performance (Algorithm Only without Human-in-the Loop)

    Yes, this study represents a standalone (algorithm only) performance evaluation. The STA® - Liatest® D-Di kit itself is the "algorithm" or diagnostic tool, and its performance (sensitivity and NPV) is reported based on its direct results in conjunction with a clinical pretest probability assessment model. Its output (D-dimer level) is then used to guide clinical decisions (imaging vs. observation).

    7. Type of Ground Truth Used

    The ground truth was established by clinical outcomes data and definitive diagnostic imaging (e.g., ultrasound for DVT, CT angiography for PE) or clinical follow-up confirming the absence of the condition. Specifically, the "Reference" for the test set was determined by "imaging or 3-month follow-up."

    8. Sample Size for the Training Set

    The document does not specify a separate "training set" or its sample size. This is common for traditional in-vitro diagnostic assays, where product development and analytical validation are often not described in terms of "training sets" like machine learning models. The study described is primarily a clinical validation study demonstrating performance on a test set.

    9. How the Ground Truth for the Training Set Was Established

    As no specific "training set" (in the context of machine learning) is discussed for this in-vitro diagnostic device, this point is not applicable. The development and analytical validation of such assays typically involve laboratory studies (e.g., linearity, precision, interference) and internal developmental studies that establish the assay's performance characteristics, rather than a "ground truth" derived from a clinical dataset used for training, as would be the case for an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K160885

    Validate with FDA (Live)

    Date Cleared
    2016-10-27

    (210 days)

    Product Code
    Regulation Number
    864.7320
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    HemosIL D-Dimer HS is an automated latex enhanced immunoassay for the quantitative determination of D-Dimer in human citrated plasma on the ACL TOP for use in conjunction with a clinical pretest probability (PTP) assessment model to exclude venous thromboembolism (VTE) in outpatients suspected of deep venous thrombosis (DVT) and pulmonary embolism (PE).

    For in vitro diagnostic use.

    Device Description

    The D-Dimer HS Latex Reagent is a suspension of polystyrene latex particles of uniform size coated with the F(ab')2 fragment of a monoclonal antibody highly specific for the D-Dimer domain included in fibrin soluble derivatives. The use of the F(ab')2 fragment allows a more specific D-Dimer detection avoiding the interference of some endogenous factors like the Rheumatoid Factor. When a plasma containing D-Dimer is mixed with the Latex Reagent and the Reaction Buffer included in the D-Dimer HS kit, the coated latex particles agglutinate. The degree of agglutination is directly proportional to the concentration of D-Dimer in the sample and is determined by measuring the decrease of the transmitted light caused by the aggregates (turbidimetric immunoassay).

    AI/ML Overview

    This document is a 510(k) summary for the HemosIL D-Dimer HS device, which is an in vitro diagnostic (IVD) test. This specific submission (K160885) is a "Special 510(k)" to update the Limit of Detection (LoD) claim for the device.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    Key Takeaway: The "device" in this context is an in vitro diagnostic assay (HemosIL D-Dimer HS) used to quantitatively determine D-Dimer levels in human plasma. The primary focus of this specific 510(k) submission is to update a performance claim, specifically the Limit of Detection (LoD), rather than introducing a new product or a significant change in intended use.

    Given this, the questions provided in the prompt, which are commonly associated with AI/ML-driven medical imaging devices, need to be reinterpreted or noted as "not applicable" for this type of IVD 510(k) submission.


    Acceptance Criteria and Reported Device Performance

    The core acceptance criterion for this submission is the updated Limit of Detection (LoD). The device's performance is reported directly against this criterion.

    1. Table of acceptance criteria and the reported device performance:

    Acceptance CriterionOriginal Claim (Predicate)Updated Claim (Proposed/Reported Performance)
    Limit of Detection (LoD)21 ng/mL137 ng/mL

    Note: The submission states that the only change is to the LoD claim based on additional testing. All other performance claims (linearity, cut-off, etc.) remain the same as the predicate device.


    Study Details (Reinterpreted for IVD Context)

    Many of the points in the prompt are more relevant to AI/ML image analysis rather than an IVD assay. Where applicable, the information is provided. Where not applicable, it is noted.

    2. Sample size used for the test set and the data provenance:

    • Sample Size: The document states that the testing was "additional testing done to current CLSI EP17-A2 requirements" to establish the updated LoD. However, the specific number of samples (e.g., patient samples, spiked samples, replicates) used for this testing to calculate the LoD is not provided in this summary. CLSI EP17-A2 is a standard guideline for evaluation of detection capability for clinical laboratory measurement procedures. This guideline typically involves a statistically significant number of replicates (e.g., 20 or more) of blank/low-level samples across multiple runs.
    • Data Provenance: The document does not specify the country of origin of the data or whether the study was retrospective or prospective. Given it's an in vitro diagnostic, samples would typically be analyzed fresh or from appropriately stored biobanks.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not Applicable in this context. For an IVD assay like D-Dimer, the "ground truth" for the test set is established by the known concentrations of calibrators and controls, or by reference methods/materials of traceable value (e.g., a known concentration of D-Dimer in a sample matrix). It does not involve human expert interpretation in the same way an imaging study would.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not Applicable. This concept is for diagnostic interpretation, typically by human readers, not for quantitative analytical performance of an IVD assay.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not Applicable. This is an IVD assay, not an AI-assisted diagnostic imaging system. Its performance is measured analytically (e.g., precision, accuracy, LoD, linearity) against known values or reference methods, not against human reader performance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Yes, in the context of an IVD. The HemosIL D-Dimer HS device is an automated, standalone assay run on ACL TOP instruments. Its performance (e.g., LoD) is determined analytically by the instrument and reagents, independent of human interpretive input. Human operators load samples and reagents, and review results, but the analytical measurement itself is automated.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For an analytical performance evaluation like Limit of Detection, the ground truth is established based on statistical analysis of replicate measurements of samples with very low (or zero) analyte concentration. This would involve:
      • Defined Calibrators: Solutions with known, precise concentrations of D-Dimer.
      • Controls: Materials with target D-Dimer concentrations used for quality control.
      • Reference Materials/Methods: Comparison to a highly accurate reference method or certified reference materials, if available, though for LoD, it's more about statistical confidence in detecting low levels.
      • Statistical Methods: LoD is determined using statistical approaches defined by guidelines like CLSI EP17-A2, which typically involves measuring blank samples and low-level samples and calculating the concentration at which detection is statistically reliable. This does not involve "expert consensus" in the clinical diagnostic sense.

    8. The sample size for the training set:

    • Not Applicable in the AI/ML sense. This is not an AI/ML algorithm that requires a "training set." The assay is a chemical and immunoturbidimetric reaction. The development of such an assay involves extensive R&D and optimization, but not typically a "training set" in the machine learning paradigm.

    9. How the ground truth for the training set was established:

    • Not Applicable. See point 8. The "ground truth" for developing the assay itself involves ensuring the reagents specifically bind to D-Dimer and that the turbidimetric measurement accurately correlates with its concentration. This is achieved through chemical and biological validation, not through annotated training data.
    Ask a Question

    Ask a specific question about this device

    K Number
    K151534

    Validate with FDA (Live)

    Date Cleared
    2015-07-06

    (28 days)

    Product Code
    Regulation Number
    864.7320
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    HemosIL D-Dimer HS is an automated latex enhanced immunoassay for the quantitative determination of D-Dimer in human citrated plasma on the ACL TOP for use in conjunction with a clinical pretest probability (PTP) assessment model to exclude venous thromboembolism (VTE) in outpatients suspected of deep venous thrombosis (DVT) and pulmonary embolism (PE)

    For in vitro diagnostic use.

    Device Description

    The D-Dimer HS Latex Reagent is a suspension of polystyrene latex particles of uniform size coated with the F(ab')2 fragment of a monoclonal antibody highly specific for the D-Dimer domain included in fibrin soluble derivatives. The use of the F(ab')2 fragment allows a more specific D-Dimer detection avoiding the interference of some endogenous factors like the Rheumatoid Factor. When a plasma containing D-D-Dimer is mixed with the Latex Reagent and the Reaction Buffer included in the D-Dimer HS kit, the coated latex particles agglutinate. The degree of agglutination is directly proportional to the concentration of D-Dimer in the sample and is determined by measuring the decrease of the transmitted

    AI/ML Overview

    This document, K151534, is a 510(k) premarket notification for the HemosIL D-Dimer HS device. The submission is a "Special 510(k)" to add information about the association of patient age with D-Dimer levels to the device's insert sheet. This means there are no changes to the device's performance, indications for use, or operating principles. Therefore, there is no new study conducted to demonstrate that the device meets new acceptance criteria. Instead, the focus is on the substantial equivalence to the predicate device (K070927) based on the absence of changes to performance claims.

    Given that this is a Special 510(k) submission for minor labeling changes and explicitly states "No change to labeled performance claims", there isn't a new study presented in this document proving the device meets new acceptance criteria. The existing performance acceptance criteria and their proof are derived from the original clearance of the predicate device (K070927).

    However, I can extract the reported performance of the device (which remains unchanged from the predicate) and details relevant to a typical device submission, even though a new study isn't detailed here for new acceptance criteria.


    1. Table of Acceptance Criteria and Reported Device Performance

    Since this is a Special 510(k) for labeling changes, the acceptance criteria and reported performance are implicitly those established during the original clearance of the predicate device (K070927). The document explicitly states "No change to labeled performance claims."

    Acceptance CriteriaReported Device Performance
    Indications for UseAutomated latex enhanced immunoassay for quantitative determination of D-Dimer in human citrated plasma on ACL TOP, for use with clinical pretest probability (PTP) assessment model to exclude VTE in outpatients suspected of DVT and PE.
    AnalyteD-Dimer
    MethodologyLatex-enhanced immunoturbidimetric assay
    AnalyzersACL TOP Family
    Sample TypeCitrated Plasma
    Cut-off230 ng/mL
    Linearity150 – 69000 ng/mL
    Detection Limit21 ng/mL
    Performance ClaimsNo change to labeled performance claims from predicate (K070927)

    2. Sample size used for the test set and the data provenance
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
    4. Adjudication method for the test set
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
    8. The sample size for the training set
    9. How the ground truth for the training set was established

    The provided document (K151534) is a Special 510(k) submission that does not include a new study to establish performance for the device. It states, "No change to labeled performance claims, including no change to the assay cut-off." This submission is solely to add general information to the device's insert sheet regarding the association of patient age with D-Dimer levels, citing "peer-reviewed published literature."

    Therefore, the information requested in points 2-9 about test sets, experts, adjudication, MRMC studies, standalone performance, ground truth, and training sets is not available in this document. Such details would have been part of the original 510(k) submission (K070927) for the predicate device, which established its performance characteristics. This current submission does not involve new clinical or performance studies for the device itself.

    Ask a Question

    Ask a specific question about this device

    K Number
    K141144

    Validate with FDA (Live)

    Device Name
    STA LIATEST D-DI
    Manufacturer
    Date Cleared
    2014-09-03

    (124 days)

    Product Code
    Regulation Number
    864.7320
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The STA® - Liatest® D-Di kit is an immuno-turbidimetric assay for the quantitative determination of D-dimer in venous plasma (in 3.2% sodium citrate) for use on STA-R®, STA Compact® and STA Satellite® analyzers by professional laboratory personnel. The STA® - Liates® D-Di is intended for use in conjunction with a clinical pretest probability (PTP) assessment model to exclude pulmonary embolism (PE) and as an aid in the diagnosis of deep venous thrombosis (DVT) in outpatients suspected of PE or DVT.

    Device Description

    STA® - Liatest® D-Di kit contains: 6 x 5-ml vials of ready-for-use Tris buffer and 6 x 6-ml vials of a suspension of microlatex particles coated with two different mouse monoclonal anti-human D-dimer antibodies (8D2 and 2.1.16) stabilized with bovine albumin.

    The test principle is based on the change in turbidity of a microparticle suspension that is measured by photometry. A suspension of latex microparticles, coated by covalent bonding with monoclonal antibodies specific for D-dimer is mixed with the test plasma for which the D-dimer level is to be assayed. An antigen-antibody reaction takes place, leading to an agglutination of the latex microparticles which causes an increase in turbidity of the reaction medium. This increase in turbidity is reflected by an increase in absorbance, the latter being measured photometrically. The increase in absorbance is a function of the D-dimer level present in the test sample.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study details for the STA® - Liatest® D-Di device, based on the provided document:

    Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the STA® - Liatest® D-Di device are implicitly tied to its ability to
    safely rule out Pulmonary Embolism (PE) in patients with low or moderate pretest probability (PTP).
    The key performance metric for exclusion of PE is the Negative Predictive Value (NPV).

    Performance MetricAcceptance Criteria (Implied by Study)Reported Device Performance (Overall Study Population)
    Sensitivity for PEHigh sensitivity to minimize false negatives97.0 % (95% CI: 91.6 % - 99.4 %)
    NPV for PE ExclusionHigh NPV to safely rule out PE99.7 % (95% CI: 99.2 % - 100.0 %)
    Specificity for PE(No explicit threshold, but reported for completeness)75.5 % (95% CI: 72.8 % - 78.1 %)
    PPV for PE(No explicit threshold, but reported for completeness)25.5 % (95% CI: 23.5 % - 27.7 %)
    Clinical Cut-offD-dimer level < 0.50 µg/ml (FEU) for PE exclusion0.50 µg/ml (FEU)

    Study Details

    1. Sample Size used for the test set and the data provenance:

      • Total Test Set Sample Size: 1130 samples of patients with a low or moderate PTP.
        • Prospective Study Population: 1060 samples.
        • US Banked Samples (retrospective): 70 samples.
      • Data Provenance: The study involved multiple centers across the United States, Europe, and Canada for the prospective data. The banked samples were from the US. The overall study is a mix of prospective and retrospective data.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
      The document does not specify the number or qualifications of experts used to establish the ground truth. The ground truth was established through a combination of imaging procedures for positive PE cases and a 3-month follow-up for negative PE cases.

    3. Adjudication method for the test set:
      The document does not explicitly describe a specific adjudication method for the test set results. Diagnosis of PE was based on imaging or clinical follow-up. For patients with suspected PE, those with positive D-dimer results were considered for an imaging procedure. Patients with negative D-dimer results were considered not to have PE and were assigned to a three-month follow-up. This implies a diagnostic pathway rather than a direct adjudication process by a panel of experts on each case.

    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
      No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This device is an in-vitro diagnostic assay (D-dimer test) and not an AI-assisted diagnostic imaging device that would typically involve human readers. Therefore, the concept of human readers improving with or without AI assistance is not applicable here.

    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
      Yes, a standalone performance study was done for the STA® - Liatest® D-Di device. The reported sensitivity, specificity, NPV, and PPV are for the device's performance alone in determining D-dimer levels and, in conjunction with PTP, for PE exclusion. While the device is "intended for use in conjunction with a clinical pretest probability (PTP) assessment model," the performance metrics presented specifically reflect the diagnostic accuracy of the D-dimer assay within that pathway.

    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
      The ground truth was established through a combination of imaging procedures (for positive PE cases) and outcomes data (3-month follow-up for negative PE cases).

    7. The sample size for the training set:
      The document does not specify a separate training set sample size. The study focuses on demonstrating the performance of the device in a clinical setting based on a defined test set. For an in-vitro diagnostic like this, the "training" (e.g., assay optimization, establishing parameters) is typically internal to the manufacturer's development process and not part of the clinical performance study reported for regulatory submission in the same way an AI model's training set would be.

    8. How the ground truth for the training set was established:
      As no separate "training set" is described in the context of this regulatory submission, how its ground truth was established is not provided. The ground truth for the test set was established as described in point 6.

    Ask a Question

    Ask a specific question about this device

    K Number
    K112818

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-07-31

    (307 days)

    Product Code
    Regulation Number
    864.7320
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    VIDAS® D-Dimer Exclusion II™ is an automated quantitative test for use on the instruments of the VIDAS family for the immunoenzymatic determination of fibrin degradation products (FbDP) in human plasma (sodium citrate, CTAD) using the ELFA technique (Enzyme Linked Fluorescent Assay). VIDAS D-Dimer Exclusion II is indicated for use in conjunction with a clinical pretest probability assessment model to exclude deep vein thrombosis (DVT) and pulmonary embolism (PE) disease in outpatients suspected of DVT or PE.

    Device Description

    The VIDAS® D-Dimer Exclusion II assay is an automated quantitative test for use on the instruments of the VIDAS family for the determination of fibrin degradation products (FbDP) in human plasma (sodium citrate) using the ELFA (Enzyme-Linked Fluorescent Assay) technique. VIDAS D-Dimer Exclusion II is indicated for use in conjunction with a clinical pretest probability assessment model to exclude deep vein thrombosis (DVT) and pulmonary embolism (PE) disease in outpatients suspected of DVT or PE.

    The assay principle combines a two-step enzyme immunoassay sandwich method with a final fluorescent detection (ELFA). The Solid Phase Receptacle (SPR), a pipette tip-like device, serves as the solid phase as well as the pipetting device for the assay. The assay reagents are ready-to-use and pre-dispensed in the sealed reagent strips (STRs). The individual kit components are described in detail on the following pages.

    All of the assay steps are performed automatically by the instrument. The reaction medium is cycled in and out of the SPR several times.

    First, the sample is taken by the SPR, diluted and then cycled in and out of the SPR several times. The antigen binds to the anti-FbDP immunodobulins coated on the SPR. Unbound components are eliminated during a washing step. In the second step, the conjugate that contains an alkaline phosphatase labeled anti-FbDP monoclonal antibody is cvcled in and out of the SPR to form a sandwich. Unbound components are eliminated during the washing steps.

    A detection step is then performed. The substrate (4-Methyl-umbellifery) phosphate) is cycled in and out of the SPR. The conjugate enzyme catalyzes the hydrolysis of this substrate into a fluorescent product (4-Methyl-umbelliferone), the fluorescence of which is measured at 450 nm. The intensity of fluorescence is proportional to the concentration of antigen present in the sample. At the end of the assay, results are automatically calculated by the instrument in relation to the calibration curve stored in memory. The results are then printed.

    AI/ML Overview

    Acceptance Criteria and Study for VIDAS® D-Dimer Exclusion II Assay

    This analysis focuses on the information provided in the 510(k) Summary for the VIDAS® D-Dimer Exclusion II Assay (K112818).

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document primarily focuses on demonstrating substantial equivalence to a predicate device rather than explicitly stating acceptance criteria for new device performance. However, by comparing the VIDAS® D-Dimer Exclusion II (DEX2) with its predicate (VIDAS® D-Dimer Exclusion), we can infer the performance metrics that were deemed acceptable for the new device. The "reported device performance" are the values presented for the VIDAS® D-Dimer Exclusion II Assay.

    MetricAcceptance Criteria (Inferred from Predicate)Reported Device Performance (VIDAS® D-Dimer Exclusion II)
    Intended UseExclusion of DVT & PE in outpatients with pretest probability modelExclusion of DVT & PE in outpatients with pretest probability model
    Assay TechniqueELFAELFA
    AutomatedYesYes
    Assay DurationSimilar to or improved from 35 min, 35 sec19 minutes, 57 seconds (Improved)
    SPR CoatingAnti-FbDP monoclonal mouse antibodiesAnti-FbDP monoclonal mouse antibodies (No preservative)
    Calibration FrequencyImproved from 14 days28 days (Improved)
    Sample TypeCitrated plasmaCitrated or CTAD plasma (Expanded)
    Assay Range45-10,000 ng/mL45-10,000 ng/mL
    Clinical Cut-Off500 ng/mL500 ng/mL
    Linearity45-5,000 ng/mL45-10,000 ng/mL (Improved)
    Detection Limit≤ 45 ng/mL≤ 45 ng/mL
    Hook Effect400,000 ng/mL400,000 ng/mL
    Specificity (Fibrinogen)No cross-reactivity<10 g/L (No cross-reactivity)
    Specificity (FbDP X)No cross-reactivity<10 µg/mL (No cross-reactivity)
    Specificity (FbDP Y)No cross-reactivity<10 µg/mL (No cross-reactivity)
    Specificity (FbDP D)Cross-reactivity (10-100 µg/mL), but not clinically significant in target populationCross-reactivity (10-100 µg/mL), but not clinically significant in target population
    Total Precision (CV)< 7.1% at various concentrationsCV 6.6 % at 277.97 ng/mL CV 5.9 % at 544.14 ng/mL CV 6.0% at 7,788.88 ng/mL
    Interference (Hemoglobin)No interference up to 300 µmol/LNo interference up to 300 µmol/L
    Interference (Lipemia)No interference up to 20 g/LNo interference up to 30 g/L (Improved)
    Interference (Bilirubin)No interference up to 537 µmol/LNo interference up to 537 µmol/L
    Interference (Rheumatoid Factor)No interference up to 396 IU/mLNo interference up to 400 IU/mL (Slightly improved)
    Drug InterferenceN/A (Predicate did not perform separate study)47 analytes tested, no interference observed
    Normal Values96% less than 500 ng/mL90% less than 500 ng/mL

    Note on Acceptance Criteria: The document is a 510(k) summary for substantial equivalence. The "acceptance criteria" are not explicitly stated as quantitative thresholds but are implicitly met by demonstrating that the new device's performance is either equivalent to or improved compared to the predicate device across various analytical and clinical characteristics. For instance, if the predicate had a linearity range of X, and the new device has a linearity range of Y (where Y >= X), then Y would be considered acceptable. The "Normal Values" for DEX2 (90% less than 500 ng/mL) is slightly lower than the predicate (96%), but in the context of a qualitative D-Dimer Exclusion assay where values below the cut-off are considered negative, this might still be deemed acceptable if the overall clinical sensitivity and specificity for DVT/PE exclusion remain robust.


    2. Sample Size Used for the Test Set and Data Provenance

    The document states "A summary of the non-clinical results is presented in the table below," referring to the detailed comparison table. This table includes analytical performance data.

    • Sample Size for Test Set: The specific sample sizes for each analytical performance test (e.g., linearity, precision, interference, specificity) are not provided in this summary. The table indicates values for these tests (e.g., CV at specific ng/mL levels), implying that tests were conducted, but the number of samples or replicates used for each test is not detailed.
    • Data Provenance: Not explicitly stated. Given that bioMérieux, Inc. is the submitter (located in Hazelwood, MO, USA), the studies were likely conducted in the USA or through their global facilities. The summary does not specify whether the data is retrospective or prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This is a an in vitro diagnostic device measuring D-Dimer levels. For such devices, "ground truth" for analytical performance studies is typically established by:

    • Reference Methods: Comparing the new assay's results against established reference methods or highly accurate laboratory techniques.
    • Reference Materials: Using certified reference materials or plasma samples with known D-Dimer concentrations.
    • Clinical Diagnosis: For clinical performance, the ground truth for DVT/PE would be established by definitive diagnostic imaging (e.g., ultrasonography, CTPA) and/or medical record review, not by "experts" in the sense of adjudicating images.

    The summary does not mention any "experts" being used to establish ground truth for the analytical test set in the way one would for diagnostic imaging. Instead, the performance is compared to the predicate device and established analytical methodologies.


    4. Adjudication Method for the Test Set

    Adjudication methods (like 2+1, 3+1) are primarily used in studies involving subjective interpretation, often in imaging or clinical assessments where multiple readers provide opinions, and inconsistencies need resolution.

    Since the VIDAS® D-Dimer Exclusion II Assay is an automated quantitative immunoassay, the results are numerical and objective. Therefore, an adjudication method in the sense of resolving conflicting interpretations is not applicable to the analytical performance tests described in this summary. The "results are automatically calculated by the instrument," which minimizes the need for human adjudication of raw data.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. An MRMC comparative effectiveness study is not mentioned. Such studies are typically conducted for diagnostic imaging devices where human readers interpret medical images, and the AI's impact on their performance is assessed. The VIDAS® D-Dimer Exclusion II Assay is an in vitro diagnostic device, and its performance is assessed analytically and clinically, not through human-in-the-loop image interpretation.


    6. If a Standalone (i.e., Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Yes, in essence. The VIDAS® D-Dimer Exclusion II is an automated assay. Its "performance" as described (assay range, linearity, precision, specificity, etc.) is the standalone performance of the instrument and reagents. The results are automatically calculated by the instrument.

    However, it's crucial to note that the device's intended use is "in conjunction with a clinical pretest probability assessment model." This means that while the assay itself is standalone, its clinical utility for DVT/PE exclusion requires human interpretation of the assay result combined with a clinical pretest probability assessment. The output of the device (D-Dimer concentration) is an algorithmic result, but its application in patient management involves a human clinician.


    7. The Type of Ground Truth Used

    For the analytical performance characteristics summarized, the ground truth would be established by:

    • Reference Standards/Materials: For linearity, detection limit, hook effect, and precision, samples with known, precisely measured concentrations of D-Dimer or related substances are used.
    • Interference Substances: Known concentrations of interfering substances (e.g., hemoglobin, bilirubin, rheumatoid factor, drugs) are added to samples to determine their impact.
    • Reference Assays: For method comparison (implicitly to the predicate, and typically to other established D-Dimer assays), results are compared against those from a recognized reference method.

    For the stated Intended Use (exclusion of DVT/PE in outpatients), the ultimate ground truth in clinical studies (which are not detailed in this analytical performance summary but are foundational for such devices) would be definitive clinical diagnosis of DVT or PE, typically established by objective imaging studies (e.g., compression ultrasonography for DVT, CT Pulmonary Angiography for PE) and clinical follow-up. This summary focuses on the analytical performance, not the full clinical validation study details.


    8. The Sample Size for the Training Set

    This document is a 510(k) summary for an immunoassay, not a machine learning or AI-based device that would typically have a separate "training set" for an algorithm. The "training set" concept is not applicable here in the conventional AI/ML sense.

    For an immunoassay, the "training" equivalent would be the extensive R&D and validation work done to optimize reagent concentrations, incubation times, antibody selection, and assay parameters. This process would involve many experiments with various sample types and concentrations, but it's not described as a discrete "training set" like in AI.


    9. How the Ground Truth for the Training Set Was Established

    As noted above, the concept of a "training set" in the AI/ML context with an established ground truth is not applicable to this immunoassay. The development and optimization of the assay would rely on:

    • Biochemical Principles: Understanding antigen-antibody reactions, enzyme kinetics, and fluorescent detection.
    • Calibrators and Controls: Precisely manufactured substances with known concentrations, used to establish the calibration curve and monitor assay performance.
    • Validation Studies: Iterative experiments using a wide range of patient samples (with D-dimer levels determined by reference methods or clinical diagnosis) to optimize and verify performance characteristics like sensitivity, specificity, and linearity.

    These activities during the assay development process serve a similar function to a training set in that they guide the refinement of the assay, but the specifics are not detailed in this regulatory summary.

    Ask a Question

    Ask a specific question about this device

    K Number
    K110303

    Validate with FDA (Live)

    Date Cleared
    2011-05-16

    (104 days)

    Product Code
    Regulation Number
    864.7320
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Stratus® CS Acute Care™ D-dimer assay (DDMR) is an in vitro diagnostic test for the quantitative measurement of cross-linked fibrin degradation products (D-dimer) in human citrated or heparinized plasma on the Stratus® CS analyzer. The Stratus® CS Acute Care™ DDMR assay is intended for use in conjunction with a non-high clinical pretest probability (PTP) assessment model to exclude pulmonary embolism (PE) disease and as an aid in the diagnosis of venous thromboembolism (VTE) [deep vein thrombosis (DVT) or pulmonary embolism (PE)]. This assay is for use by trained health care professionals in the clinical laboratory and point of care (POC) settings.

    Device Description

    The Stratus® CS Acute Care™ D-Dimer method is a two-site sandwich assay based upon solid phase Radial Partition Immunoassay (RPIA) technology. In this procedure, dendrimer linked monoclonal antibody is added to the center portion of a square piece of glass fiber paper in the DDMR TestPak. This antibody recognizes a distinct antigenic site on the D-dimer molecule. Sample is then added onto the paper where it reacts with the immobilized antibody. After a short incubation, a conjugate consisting of enzyme-labeled monoclonal antibody directed against a second distinct antigenic site on the D-dimer molecule is pipetted onto the reaction zone of the paper. During this second incubation period, enzyme-labeled antibody reacts with the bound D-dimer, forming an antibodyantigen-labeled antibody sandwich. The unbound labeled antibody is later eluted from the field of view of the Stratus® CS STAT Fluorometric Analyzer (Stratus® CS analyzer) by applying a substrate wash solution to the center of the reaction zone. By including substrate for the enzyme within the wash solution, initiation of enzyme activity occurs simultaneously with the wash. The enzymatic rate of the bound fraction increases directly with the concentration of D-dimer in the sample. The reaction rate can then be measured by an optical system that monitors the reaction rate via front surface fluorescence. All data analysis functions are performed by the microprocessor within the analyzer.

    AI/ML Overview

    Acceptance Criteria and Device Performance for Stratus® CS Acute Care™ D-Dimer Assay

    This document summarizes the acceptance criteria and performance of the Stratus® CS Acute Care™ D-Dimer assay as described in the provided 510(k) summary (K110303).

    1. Table of Acceptance Criteria and Reported Device Performance

    The study evaluated the performance of the Stratus® CS Acute Care™ D-Dimer assay against a clinical cutoff of 450 ng/mL [µg/L] (FEU) for the exclusion of Pulmonary Embolism (PE).

    Acceptance Criteria (Implied by Predicate Device Equivalence and Clinical Relevance for PE Exclusion):
    While explicit numerical acceptance criteria for sensitivity, specificity, and NPV are not directly stated as pass/fail thresholds, the study aims to demonstrate substantial equivalence to the Innovance D-Dimer assay and clinical utility for PE exclusion, implying that the performance metrics should be comparable to or ideally exceed those required for safe and effective PE exclusion. For D-dimer assays used to exclude PE, high sensitivity and negative predictive value (NPV) are critical. The reported values are generally considered acceptable for this application.

    Reported Device Performance:

    Citrated Plasma: In all patients

    InstrumentPE Patients (n)Cutoff ng/mL [µg/L] FEUSensitivity (CL) %Specificity (CL) %NPV (CL) %
    Stratus® CS analyzer65545098.9 (94.1 – 100.0)42.5 (38.3 – 46.7)99.6 (97.7 – 100.0)

    Citrated Plasma: In Patients with low and moderate pre-test probability

    InstrumentPE Patients (n)Cutoff ng/mL [µg/L] FEUSensitivity (CL) %Specificity (CL) %NPV (CL) %
    Stratus® CS analyzer62545098.7 (92.9 – 100.0)43.0 (38.8 – 47.3)99.6 (97.7 – 100.0)

    Heparinized Plasma: In all patients

    InstrumentPE Patients (n)Cutoff ng/mL [µg/L] FEUSensitivity (CL) %Specificity (CL) %NPV (CL) %
    Stratus® CS analyzer42745098.3 (91.1 – 100.0)29.7 (25.1 – 34.7)99.1 (95.0* – 100.0)
    *95.04

    Heparinized Plasma: In Patients with low and moderate pre-test probability

    InstrumentPE Patients (n)Cutoff ng/mL [µg/L] FEUSensitivity (CL) %Specificity (CL) %NPV (CL) %
    Stratus® CS analyzer40145097.9 (88.7 – 100.0)29.9 (25.2 – 35.0)99.1 (94.9 – 100.0)

    2. Sample Size Used for the Test Set and Data Provenance

    • Citrated Plasma: 730 consecutive patients initially, 655 patients remaining after exclusions for final analysis.
    • Heparinized Plasma: 468 consecutive patients initially, 427 patients remaining after exclusions for final analysis.
    • Data Provenance: The study was a multi-center study. The document does not specify the countries of origin but states it's a "multi-center study." The data is prospective, collected from consecutive patients presenting to the emergency department with suspected PE.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    The document does not explicitly state the number of experts used to establish the ground truth or their qualifications in detail. However, the ground truth for PE diagnosis appears to be established through a combination of imaging methods (e.g., spiral CT and/or VQ scan) for patients with positive D-dimer and/or high PTP, and a three-month follow-up for patients with negative D-dimer/low/moderate PTP or negative imaging results. This implies that the diagnosis of PE was made by medical professionals (e.g., radiologists, clinicians) utilizing established diagnostic protocols but does not specify individual expert roles for establishing a "ground truth" for the test set.

    4. Adjudication Method for the Test Set

    The document does not describe an explicit adjudication method (e.g., 2+1, 3+1) for establishing the ground truth diagnoses. The diagnostic process involved imaging (spiral CT, VQ scan) and clinical follow-up for three months. It is implied that standard clinical practices and interpretations were followed for these diagnostic procedures, rather than a specific consensus or adjudication process among multiple independent experts to establish the ground truth for each case.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study involving human readers with and without AI assistance was not performed. This study focuses on the performance of the device (assay) itself in a standalone manner, rather than as an aid to human interpretation.

    6. If a Standalone Study Was Done

    Yes, a standalone study of the algorithm's performance (the D-dimer assay) was done. The reported sensitivity, specificity, and negative predictive value are directly from the assay's performance against the established clinical ground truth for PE. There is no human-in-the-loop component for these reported performance metrics.

    7. The Type of Ground Truth Used

    The ground truth used was a combination of:

    • Imaging Data: Spiral CT and/or VQ scans were used for patients with positive D-dimer results and/or high pre-test probability (PTP).
    • Outcomes Data (Clinical Follow-up): Patients with negative D-dimer results and low or moderate PTP (or negative imaging results) were followed for three months to evaluate potential development of PE. This follow-up serves as an outcome-based ground truth to confirm the absence of PE.

    8. The Sample Size for the Training Set

    The document does not specify a separate training set. This 510(k) submission describes a clinical validation study, implying that the D-dimer assay and its cutoff of 450 ng/mL [µg/L] (FEU) were already established, likely based on previous research or predicate device characteristics. The study described here acts as a test set for validating the performance of the assay rather than for training a new algorithm.

    9. How the Ground Truth for the Training Set Was Established

    As no explicit training set is described, the method for establishing ground truth for a training set is not applicable to the information provided. The study focuses on evaluating the assay's performance on a validation dataset.

    Ask a Question

    Ask a specific question about this device

    K Number
    K093626

    Validate with FDA (Live)

    Date Cleared
    2010-11-29

    (370 days)

    Product Code
    Regulation Number
    864.7320
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For the quantitative determination of cross-linked fibrin degradation products (D-dimers) in human plasma on Siemens Healthcare Diagnostics and Sysmex® Coagulation Systems. The INNOVANCE® D-Dimer assay is intended for use in conjunction with a non-high clinical pretest probability (PTP) assessment model to exclude deep vein thrombosis (DVT) and pulmonary embolism (PE).

    Device Description

    Polystyrene particles covalently coated with a monoclonal antibody (8D3) are aggregated when mixed with samples containing D-dimer. The D-dimer crosslinkage region has a stereosymmetrical structure, i.e. the epitope for the monoclonal antibody occurs twice. Consequently, one antibody suffices in order to trigger an aggregation reaction, which is then detected turbidimetrically via the increase in turbidity.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the INNOVANCE® D-Dimer assay:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are not explicitly stated as distinct numerical targets in the provided text. However, the study aims to demonstrate that the device performs with sufficient sensitivity and Negative Predictive Value (NPV) to reliably exclude DVT, especially in patients with an unlikely pre-test probability. Based on the presented data, the implied acceptance criteria would be a sensitivity and NPV close to or at 100% for the intended use.

    MetricAcceptance Criteria (Implied)Reported Device Performance (All Patients)Reported Device Performance (Unlikely PTP Patients)
    Sensitivity≥ 96% (based on lower CL)100.0% (96.1 – 100.0% CL)100.0% (83.9 – 100.0% CL)
    SpecificityNot directly, but >30% (for practical use)34.5% (29.4 – 39.9% CL)37.0% (31.0 – 43.4% CL)
    NPV≥ 96% (based on lower CL)100.0% (96.8 – 100.0% CL)100.0% (96.0 – 100.0% CL)

    Study Proving Device Meets Acceptance Criteria:

    The study referenced is a multi-center clinical evaluation evaluating the INNOVANCE® D-Dimer assay on the BCS® / BCS® XP System to exclude DVT.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set:
      • Total Patients: 455 consecutive patients initially enrolled.
      • Patients for Final Analysis: 426 patients (29 were excluded)
      • Patients with unlikely pre-test probability: 267 patients.
    • Data Provenance: The study was a "multi-center study," suggesting data was collected from multiple clinical sites. The text doesn't explicitly state the country of origin, but the manufacturer is German, and the contact information is for the US. It is a prospective study as it involved "consecutive patients presenting to the emergency department" and follow-up for three months.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The text does not specify the number of experts used or their qualifications to establish the ground truth for the test set.

    4. Adjudication Method for the Test Set

    The text does not explicitly state an adjudication method for the test set. The diagnostic certainty (ground truth) appears to have been established through a combination of imaging methods (compression ultrasound and/or venography) for positive D-dimer results, and a three-month follow-up for negative D-dimer and negative imaging results. This implicitly suggests that the attending clinicians/radiologists' assessments, informed by these methods, served as the ground truth.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

    No, an MRMC comparative effectiveness study involving human readers with and without AI assistance was not done. This study focuses on the standalone performance of the D-dimer assay.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Yes, a standalone performance study was done. The results presented (Sensitivity, Specificity, NPV) are for the INNOVANCE® D-Dimer assay performed at a specific cutoff, independent of human interpretation of the D-dimer value itself, other than applying the PTP model. The assay's performance is then compared to the established ground truth of DVT.

    7. The Type of Ground Truth Used

    The ground truth used was a combination of:

    • Imaging methods: Compression ultrasound and/or venography for patients with positive D-dimer results.
    • Outcomes data: Three-month follow-up for patients with negative D-dimer results and those with negative imaging results, to evaluate for the potential development of DVT.

    8. The Sample Size for the Training Set

    The document does not specify a separate training set or its sample size. The study described appears to be a validation study of the device against clinical outcomes, rather than a development and training study for an algorithm. D-dimer assays traditionally involve established biochemical principles and cutoffs, rather than machine learning algorithms that require explicit training sets.

    9. How the Ground Truth for the Training Set Was Established

    Since no explicit training set or machine learning algorithm development is described, the method for establishing ground truth for a training set is not applicable in this context. The 0.50 mg/L (FEU) cutoff is a pre-established clinical cutoff for D-dimer assays.

    Ask a Question

    Ask a specific question about this device

    K Number
    K090264

    Validate with FDA (Live)

    Date Cleared
    2010-02-05

    (367 days)

    Product Code
    Regulation Number
    864.7320
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    HemosIL D-Dimer HS 500 is an automated latex enhanced immunoassay for the quantitative determination of D-Dimer in human citrated plasma on the ACL TOP® Family Systems for use, in conjunction with a clinical pretest probability (PTP) assessment model to exclude venous thromboembolism (VTE) in outpatients suspected of deep venous thrombosis (DVT) and pulmonary embolism (PE).

    HemosIL D-Dimer HS 500 Controls are intended for the quality control of the D-Dimer HS 500 assay performed on the ACL TOP® Family Systems.

    For in vitro diagnostic use.

    Device Description

    HemosIL D-Dimer HS 500: The D-Dimer HS 500 Latex Reagent is a suspension of polystyrene latex particles of uniform size coated with the F(ab')2 fragment of a monoclonal antibody highly specific for the D-Dimer domain included in fibrin soluble derivatives. The use of the F(ab'), fragment allows a more specific D-Dimer detection avoiding the interference of some endogenous factors like the Rheumatoid Factor. When a plasma containing D-Dimer is mixed with the Latex Reagent and the Reaction Buffer included in the HemosIL D-Dimer HS 500 kit, the coated latex particles agglutinate. The degree of agglutination is directly proportional to the concentration of D-Dimer in the sample and is determined by measuring the decrease of transmitted light caused by the aggregates (turbidimetric immunoassay).

    HemosIL D-Dimer HS 500 Controls: The Low and High D-Dimer HS 500 Controls are prepared by means of a dedicated process and contain different concentrations of partially purified D-Dimer obtained by digestion of Factor XIIIa cross-linked human fibrin with human plasmin.

    AI/ML Overview

    Here's an analysis of the provided text, outlining the acceptance criteria and the study data that supports the device's performance, as requested:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the HemosIL D-Dimer HS 500 assay are primarily focused on its sensitivity and negative predictive value (NPV) for excluding Venous Thromboembolism (VTE) in outpatients suspected of Deep Venous Thrombosis (DVT) and Pulmonary Embolism (PE), when used in conjunction with a clinical pretest probability (PTP) assessment model. The reported performance from the outcome study and multi-center management study directly addresses these criteria.

    Acceptance Criteria (Implied for VTE Exclusion)Reported Device Performance (95% CI) - Overall Population
    For Exclusion of DVT:
    High Sensitivity100.0% (96.0%-100.0%)
    High Negative Predictive Value (NPV)100.0% (97.2%-100.0%)
    For Exclusion of PE:
    High Sensitivity100.0% (93.2%-100.0%)
    High Negative Predictive Value (NPV)100.0% (97.4%-100.0%)
    Precision:ACL TOP Family (Mean/CV%)
    Acceptable precision (low CV%) for controlsLow Control: 877 U/mL (2.9% within-run, 8.9% total)
    High Control: 2469 U/mL (2.5% within-run, 7.3% total)
    Acceptable precision for plasma poolPlasma Pool: 423 U/mL (7.2% within-run, 9.5% total)
    Method Comparison:n=100
    Good correlation with predicate device (VIDAS)Slope: 1.00, r: 0.981

    Study Proving Device Meets Acceptance Criteria:

    The device's ability to meet the acceptance criteria is demonstrated through two primary clinical studies and supporting analytical studies:

    • "Outcome Study" (Section 2, paragraph 3)
    • "Multi-center Management Study" (Section 3, paragraph 1)
    • "Precision" study (Section 2, paragraph 1)
    • "Method Comparison" study (Section 2, paragraph 2)

    These studies collectively show that the HemosIL D-Dimer HS 500, with a clinical cut-off of 500 ng/mL, consistently achieved 100% sensitivity and 100% negative predictive value for both DVT and PE in the studied outpatient populations when used in conjunction with a PTP assessment. This indicates that no VTE events were missed by the test, effectively "excluding" the condition when the D-Dimer result was negative. The precision and method comparison studies demonstrate the analytical reliability and equivalence to existing commercial methods.

    2. Sample Size Used for the Test Set and Data Provenance

    Due to the nature of this in-vitro diagnostic device (IVD), there isn't a traditional "test set" in the sense of a medical imaging AI algorithm. Instead, there are clinical samples used in performance evaluations.

    • Outcome Study (Clinical Performance):

      • Sample Size: 295 frozen samples.
      • Data Provenance: Patients admitted consecutively to an emergency unit with suspected PE or DVT. The country of origin is not explicitly stated but implies a clinical setting. The samples are retrospective as they were "frozen samples" from past admissions.
    • Multi-center Management Study (Clinical Performance):

      • Sample Size: 747 samples (401 suspected DVT, 346 suspected PE).
      • Data Provenance: Patients admitted consecutively to the emergency unit with suspected DVT or PE at four hospitals. The country of origin is not explicitly stated. This appears to be a prospective study because it describes patient management pathways and follow-up, indicating data collected as patients were presented.
    • Precision and Method Comparison Studies: These appear to be in-house laboratory studies using controls, plasma pools, and patient samples (n=100 for method comparison). Further specifics on provenance are not detailed.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The concept of "experts" establishing ground truth in this context is different from image-based AI where radiologists annotate images. Here, the ground truth for VTE (DVT and PE) status was established through "standard objective tests" and clinical follow-up, not by a panel of human reviewers interpreting the D-Dimer test itself.

    • Outcome Study: VTE positive cases (75 samples) were confirmed by "standard objective tests." VTE negative cases (220 samples) were also confirmed. The specific tests (e.g., ultrasound, CT angiography) are not detailed, nor are the number or qualifications of the clinicians/technicians who performed or interpreted these objective tests.

    • Multi-center Management Study: VTE status for DVT and PE was confirmed through a combination of imaging techniques (for positive D-Dimer or high PTP patients) and importantly, 3-month clinical follow-up for patients with negative D-Dimer and low/moderate PTP. Again, the number and specific qualifications of the clinicians involved in these diagnostic methods or follow-ups are not specified.

    4. Adjudication Method for the Test Set

    Not applicable in the conventional sense of human reviewers adjudicating results. The "adjudication" of definitive VTE status relied on a combination of objective diagnostic tests and clinical follow-up, which are the accepted standards for determining the presence or absence of DVT/PE. The clinical pretest probability (PTP) assessment using the Wells model served as an initial clinical judgment tool, guiding further diagnostic pathways, but the ultimate ground truth was based on objective methods and patient outcomes.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done as described for AI assistance. This device is an in-vitro diagnostic (IVD) assay, a lab test intended to provide quantitative results directly, not to assist human readers in interpreting complex images or data in the same way an AI algorithm might augment a radiologist. The study evaluates the assay's performance in a clinical pathway, not the improvement of human readers with AI assistance.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, the performance data presented is for the standalone performance of the HemosIL D-Dimer HS 500 assay, specifically the quantitative determination of D-Dimer. The "human-in-the-loop" aspect here refers to the physician integrating the D-Dimer result with the clinical pretest probability (PTP) assessment model. The device itself (the assay) provides a direct quantitative reading (e.g., 500 ng/mL) without human interpretation to arrive at that numerical value. The clinical utility is then determined by how this standalone result integrates into a diagnostic algorithm. The sensitivity and NPV reported are based on the assay's output.

    7. The Type of Ground Truth Used

    The ground truth used for the clinical studies (Outcome Study and Multi-center Management Study) was primarily:

    • Objective Diagnostic Tests: For confirmed VTE cases (DVT and PE), standard objective tests (e.g., imaging techniques like ultrasound for DVT, CT angiography for PE) were used. These are considered highly definitive.
    • Outcomes Data (Clinical Follow-up): Crucially, for patients with negative D-Dimer and low/moderate PTP, the absence of VTE was confirmed by 3-month clinical follow-up where the development of DVT or PE was monitored. This represents outcomes data as ground truth, which is a very strong form of evidence for exclusion criteria.

    8. The Sample Size for the Training Set

    The provided summary does not explicitly mention a "training set" in the context of machine learning. This is an in-vitro diagnostic device, which is typically developed through analytical validation (reagent formulation, calibration, interference studies) and clinical validation.

    • The "Method Comparison" study (n=100) could be considered part of the validation data used to refine or confirm equivalence, but it's not a training set for an algorithm in the AI sense.
    • The "Precision" study uses controls and a plasma pool, which are analytical samples for assessing assay performance, not a training set.

    If earlier development involved optimizing the assay's parameters (e.g., antibody concentration, reaction conditions), that would have used various experimental samples, but a defined "training set" like in AI is not applicable here.

    9. How the Ground Truth for the Training Set Was Established

    As there isn't a "training set" in the AI sense for this IVD, this question is not directly applicable. For analytical development and validation, the "ground truth" would be established through:

    • Defined D-Dimer concentrations: For calibrators and controls used in precision studies.
    • Comparison to established reference methods: For method comparison, the predicate device (VIDAS D-Dimer Exclusion Assay) served as the reference for comparison. The ground truth for the samples tested would be their D-Dimer concentration as determined by the predicate device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K091916

    Validate with FDA (Live)

    Date Cleared
    2009-10-29

    (121 days)

    Product Code
    Regulation Number
    864.7320
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For the quantitative determination of cross-linked fibrin degradation products (D-dimers) in human plasma on Siemens Healthcare Diagnostics and Sysmex® Coagulation Systems. The INNOVANCE® D-Dimer assay is intended for use in conjunction with a non-high clinical pretest probability (PTP) assessment model to exclude pulmonary embolism (PE) disease and as an aid in the diagnosis of venous thromboembolism (VTE) [deep vein thrombosis (DVT) or pulmonary embolism (PE)].

    Device Description

    Polystyrene particles covalently coated with a monoclonal antibody (8D3) are aggreated when mixed with samples containing D-dimer. The D-dimer crosslinkage region has a stereosymmetrical structure, i.e. the epitope for the monoclonal antibody occurs twice. Consequently, one antibody suffices in order to trigger an aggregation reaction, which is then detected turbidimetrically via the increase in turbidity.

    AI/ML Overview

    The INNOVANCE® D-Dimer assay was evaluated in a multicenter clinical study to validate its ability to assist in the exclusion of pulmonary embolism (PE).

    Here's a breakdown of the acceptance criteria and study details:

    1. Acceptance Criteria and Reported Device Performance:

    The document doesn't explicitly state quantitative acceptance criteria in a dedicated section. However, the performance is reported as key metrics (Sensitivity, Specificity, and Negative Predictive Value) with 95% Confidence Limits (CL), which are typically compared against pre-defined thresholds for acceptance in clinical trials. Given the context of a 510(k) submission for exclusion of PE, a high Negative Predictive Value (NPV) is crucial.

    MetricAcceptance Criteria (Implied)Reported Device Performance (All Patients)Reported Device Performance (Low/Moderate PTP Patients)
    SensitivityHigh (especially for PE exclusion)98.9% (93.9 - 100.0%)98.6% (92.5 - 100.0%)
    SpecificityAdequate (to avoid unnecessary imaging)39.6% (35.5 - 43.8%)40.4% (36.3 - 44.7%)
    NPVVery High (critical for rule-out)99.6% (97.5 - 100.0%)99.6% (97.5 - 100.0%)

    Note: The "Acceptance Criteria (Implied)" column reflects the expected performance for a D-dimer assay used to exclude PE. A very high NPV is generally the primary focus for such an exclusion test.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size:
      • Initial patients presenting: 701
      • Patients available for final analysis (test set): 647
    • Data Provenance: The study was a multicenter study, suggesting data was collected from multiple clinical sites. It was a prospective study, as patients were "consecutive patients presenting to the emergency department with suspected PE." The country of origin is not explicitly stated, but the manufacturer is based in Germany and the contact information is in the US, indicating an international context or a US-based multicenter study.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    This information is not explicitly stated in the provided text. The document mentions that "Patients with a positive D-dimer result and/or a high PTP were evaluated by imaging methods, e.g. spiral CT and/or VQ scan." It also states, "Patients with a negative D-dimer result and a low or moderate PTP (these patients underwent imaging at the physician's discretion), and patients with negative imaging results, were followed for three months to evaluate potential development of PE." This implies that the ultimate diagnosis of PE (the ground truth) was established through a combination of imaging, clinical follow-up for events, and physician assessment, but the number or specific qualifications of experts involved in reviewing these results are not detailed.

    4. Adjudication Method for the Test Set:

    The document does not explicitly describe an adjudication method for conflicting interpretations of imaging or follow-up results. The ground truth seems to be derived from a combination of recognized medical diagnostic procedures (imaging, follow-up) as performed by attending physicians, rather than a separate, blinded expert adjudication panel of the test results themselves.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This was not an MRMC comparative effectiveness study in the context of human readers vs. AI. The device is an in-vitro diagnostic (IVD) assay (D-dimer blood test), not an AI imaging or diagnostic algorithm designed to assist human readers. Therefore, this section is not applicable.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Yes, this was a standalone performance study of the INNOVANCE® D-Dimer assay. The study evaluated the direct output of the assay (D-dimer levels) in conjunction with clinical pretest probability (PTP) assessment, without explicit human-in-the-loop performance comparison for the assay itself. The physician's role was to use the assay's results as part of a diagnostic algorithm (along with PTP).

    7. The Type of Ground Truth Used:

    The ground truth for the presence or absence of PE was established through a combination of:

    • Imaging methods: e.g., spiral CT and/or VQ scan for patients with positive D-dimer and/or high PTP, and for some patients with negative D-dimer and low/moderate PTP at physician's discretion.
    • Outcomes data/Clinical Follow-up: Three-month follow-up for patients with negative D-dimer and low or moderate PTP (and those with negative imaging) to evaluate potential development of PE.

    This is a robust method leveraging established clinical diagnostic pathways for PE.

    8. The Sample Size for the Training Set:

    The document does not provide information about a separate training set. This study describes the clinical validation of the INNOVANCE® D-Dimer assay, implying that the assay itself (and its cutoff values) were already developed and "trained" (if applicable to the assay's development process) prior to this multicenter clinical validation. IVD assays typically undergo extensive R&D and analytical validation before clinical studies like this for regulatory submission.

    9. How the Ground Truth for the Training Set Was Established:

    As no specific training set is detailed for this 510(k) submission, the method for establishing its ground truth is not provided.