Search Filters

Search Results

Found 45 results

510(k) Data Aggregation

    K Number
    K240865

    Validate with FDA (Live)

    Date Cleared
    2024-10-23

    (209 days)

    Product Code
    Regulation Number
    862.1680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The IDS-iSYS Free Testosterone assay is an in vitro diagnostic device intended for the quantitative determination of free testosterone in human serum or the IDS system. Measurement of free testosterone is used in the diagnosis and treatment of disorders involving the male sex hormones (androgens), including primary and secondary hypogonadism, impotence in male and in females; hirsutism (excessive hair) and virilization) due to tumors, polycystic ovaries and androgenital syndromes.

    Device Description

    The IDS-iSYS Free Testosterone assay consists of a reagent cartridge. The reagent cartridge contains multiple reagents:

    • MP3: Magnetic particles coated with Streptavidin in a phosphate Pluronic buffer with sodium azide (NaN3) as preservative (< 0.1%). 1 bottle, 2.5 mL.
    • CONJ: Testosterone labelled with an acridinium ester derivative, in phosphate buffer containing bovine protein with ProClin 300 as preservative (< 0.0015%). 1 bottle, 3.5 mL
    • Ab-BIOT: Monoclonal anti-Testosterone labelled with Biotin in MES buffer with ProClin 300 as preservative (<0.0015%). 1 bottle, 7.5 mL
    • Cal A: Human serum matrix containing human free testosterone with sodium azide and ProClin 300 as preservatives (<0.1% and 0.0024% respectively). 1.0 mL per bottle
    • Cal B: Human serum matrix containing human free testosterone with sodium azide and ProClin 300 as preservatives (<0.1% and 0.0024% respectively). 1.0 mL per bottle
    AI/ML Overview

    The provided document describes the analytical performance of the IDS-iSYS Free Testosterone assay but does not detail a device performance study with acceptance criteria in the typical format of a clinical trial for an AI/ML medical device. Instead, it focuses on the analytical characteristics of the in vitro diagnostic device, comparing it to a predicate device and demonstrating its performance through various laboratory tests.

    Here's an attempt to structure the information based on your request, with the understanding that not all requested points are directly applicable to this type of IVD submission:

    1. Table of Acceptance Criteria and Reported Device Performance

    For an in vitro diagnostic device like the IDS-iSYS Free Testosterone, "acceptance criteria" and "reported device performance" are typically defined by analytical performance characteristics, such as sensitivity, precision, linearity, and interference. The document presents these values but does not explicitly state pre-defined acceptance criteria for each measurement that would be found in a clinical study protocol. However, we can infer performance targets based on the data presented and common medical device standards (e.g., CLSI guidelines).

    Acceptance Criteria (Inferred/Generic for IVD)Reported Device Performance (IDS-iSYS Free Testosterone)
    Analytical Sensitivity
    Limit of Blank (LoB)0.08 pg/mL
    Limit of Detection (LoD)0.17 pg/mL
    Limit of Quantitation (LoQ)0.40 pg/mL (with CV < 20%)
    Precision (Repeatability)
    CV% for Low concentration (e.g., < 1 pg/mL)7.0% (S1: 0.7 pg/mL)
    CV% for Mid concentration1.4% (S6: 10.8 pg/mL)
    CV% for High concentration2.2% (S10: 56.0 pg/mL)
    Precision (Within-Laboratory)
    CV% for Low concentration (e.g., < 1 pg/mL)9.9% (S1: 0.7 pg/mL)
    CV% for Mid concentration3.7% (S6: 10.8 pg/mL)
    CV% for High concentration3.4% (S10: 56.0 pg/mL)
    Precision (Reproducibility)
    CV% (System to System, Low Conc.)11.1% (S2: 1.2 pg/mL)
    CV% (System to System, High Conc.)3.9% (S10: 57.9 pg/mL)
    CV% (Lot to Lot, Low Conc.)7.8% (S2: 1.2 pg/mL)
    CV% (Lot to Lot, High Conc.)4.7% (S10: 58.9 pg/mL)
    Linearity Range
    Linear measurement range0.12 pg/mL to 68.12 pg/mL
    Interference/Cross-Reactivity
    Cross-reactivity % for various compoundsMostly < 0.01% (e.g., 11-Deoxycortisol, Estradiol)
    Bias for interfering agents (e.g., Bilirubin, Hemoglobin)≤ ±10% (observed at specified thresholds)
    Method Comparison (vs. commercially available ELISA)
    Correlation Coefficient (r)0.98
    Slope1.02 (95% CI: 0.97 to 1.06)
    Intercept-0.02 pg/mL (95% CI: -0.26 to 0.07)
    Matrix Comparison (vs. Serum)
    Correlation Coefficient (r) for SST, K2 EDTA, Li Heparin, Na HeparinAll 0.99 or 1.00
    Slope for SST, K2 EDTA, Li Heparin, Na Heparin0.96 - 0.97

    2. Sample Size Used for the Test Set and Data Provenance

    The "test set" in this context refers to the samples used for analytical validation studies.

    • Analytical Limits (LoB, LoD, LoQ): 60 replicates of 4 blank samples and 6 low-concentration samples for LoB/LoD; 105 replicates of 7 low-concentration samples for LoQ.
    • Repeatability: 10 serum samples, 80 replicates per sample.
    • Reproducibility (System/Operator): 9 serum samples, 75 replicates per sample.
    • Reproducibility (Lot-to-Lot): 9 serum samples, 75 replicates per sample.
    • Linearity: The number of unique samples is not specified, but the study evaluated the measurement procedure's linearity from 0.12 pg/mL to 68.12 pg/mL.
    • Cross-Reactivity: 2 samples (1.0 and 15 pg/mL free testosterone) spiked with various cross-reactants.
    • Interference: 2 samples (1.0 and 45.0 pg/mL free testosterone) spiked with various interfering agents.
    • Method Comparison: 241 samples. The document does not specify the country of origin or if these were retrospective or prospective samples, but they are patient samples covering a wide range of concentrations.
    • Matrix Comparison: 40 samples for each matrix (SST, K2 EDTA, Li Heparin, Na Heparin) compared against serum.
    • Expected Values:
      • Females: 130 (21-39 yrs), 57 (40-59 yrs), 67 (>=60 yrs)
      • Males: 129 (21-39 yrs), 138 (40-59 yrs), 42 (>=60 yrs)
        The provenance of these subjects is not stated, but they are described as "apparently healthy adults and children."

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    This information is not applicable to this document. The "ground truth" for an IVD device like this is typically established by reference methods or gravimetric preparation of calibrators/controls, not by human expert opinion as would be the case for image-based AI/ML diagnostics. The values are quantitative measurements of a biochemical marker.

    4. Adjudication Method for the Test Set

    This information is not applicable. Adjudication methods (e.g., 2+1, 3+1) are common in studies where human readers interpret data (like medical images) and their agreement, or lack thereof, needs to be resolved. For an IVD, the "ground truth" is a measured concentration, and the accuracy is assessed against reference standards or established methods.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    This information is not applicable. An MRMC study assesses how human readers' performance changes with and without AI assistance. This device is an in vitro diagnostic assay, directly measuring a biomarker without human interpretation in the workflow described.

    6. Standalone (Algorithm Only) Performance

    The performance data presented throughout the document (LoB, LoD, LoQ, precision, linearity, cross-reactivity, interference, method comparison, matrix comparison) is the standalone performance of the IDS-iSYS Free Testosterone assay. This device is a fully automated assay system, and its performance is evaluated independent of human interpretive steps.

    7. Type of Ground Truth Used

    The ground truth for the analytical studies is generally based on:

    • Known concentrations: For LoB, LoD, LoQ, linearity, cross-reactivity, and interference, samples are prepared with known or target concentrations of free testosterone and potential interfering substances.
    • Comparative methods: For method comparison, a "commercially available quantitative free testosterone ELISA" serves as the comparative method against which the IDS-iSYS Free Testosterone is measured.
    • Reference Intervals: For expected values, "95% reference interval for apparently healthy adults and children" was calculated using a non-parametric method, likely referring to the distribution of measurements in that population.

    8. Sample Size for the Training Set

    This information is not provided and is typically not applicable in the same way it would be for an AI/ML device. For an IVD assay, "training" involves the development and optimization of the assay reagents, protocols, and calibration, rather than training a machine learning model on a distinct dataset. The "training set" for an IVD refers to the samples used to develop and refine the assay's performance characteristics and establish its calibration curve, which is distinct from the analytical validation samples.

    9. How the Ground Truth for the Training Set Was Established

    This information is not explicitly provided in the document. For an IVD, the "ground truth" for developing a training set (i.e., for calibration) typically involves:

    • Gravimetric preparation: Precisely weighing and dissolving a known amount of the analyte (free testosterone) in a suitable matrix to create primary calibrators with accurate, traceable concentrations.
    • Reference methods: Using highly accurate and validated reference methods (e.g., LC-MS/MS, though not specified here) to assign values to calibrators or control materials.
    • Standardization: Following established industry and regulatory standards (e.g., CLSI guidelines) for calibrator preparation and value assignment to ensure accuracy and traceability.
    Ask a Question

    Ask a specific question about this device

    K Number
    K233480

    Validate with FDA (Live)

    Device Name
    Access SHBG
    Date Cleared
    2024-02-08

    (105 days)

    Product Code
    Regulation Number
    862.1680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Access SHBG assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of Sex Hormone Binding Globulin levels in human serum and plasma using the Access Immunoassay Systems. The Access SHBG assay is indicated for use in the assessment of androgen disorders.

    Device Description

    The Access SHBG assay is a sequential two-site immunoenzymatic ("sandwich") assay. The Access SHBG assay consists of the reagent pack, calibrators and QCs. Other items needed to run the assay include substrate and wash buffer. The Access SHBG assay reagent pack, Access SHBG assay calibrators, Access SHBG QCs, along with the UniCel Dxl Wash Buffer II are designed for use with the Dxl 9000 Access Immunoassay Analyzer in a clinical laboratory setting.

    AI/ML Overview

    The FDA 510(k) clearance letter and accompanying summary describe the Acceptance Criteria and the study proving the Access SHBG assay meets these criteria.

    Acceptance Criteria and Device Performance

    The provided document focuses on the analytical performance of the Access SHBG assay on the Dxl 9000 Access Immunoassay Analyzer, comparing it to a previously cleared predicate device (Access SHBG on Access 2 instrument). The acceptance criteria are largely derived from CLSI (Clinical and Laboratory Standards Institute) guidelines for various analytical performance characteristics.

    Here is a table summarizing the acceptance criteria and the reported device performance for the analytical studies:

    Performance CharacteristicAcceptance Criteria (from CLSI guidelines or internal design)Reported Device Performance (Access SHBG on Dxl 9000)
    Method Comparison (compared to predicate)R$^2$ $\ge$ 0.95 with a slope equal to 1.00 $\pm$ 0.09R = 1.00, R$^2$ = 0.99, Slope = 1.01 (95% CI: 1.00-1.03), Intercept = -0.019 (95% CI: -0.46 - 0.29)
    Precision (Within-Laboratory Imprecision)$\le$ 0.14 nmol/L SD at concentrations $\le$ 2 nmol/L $\le$ 7.0% CV at concentrations > 2 nmol/LSample 1 (0.82 nmol/L): SD = 0.04, CV = 4.6% Sample 2 (18 nmol/L): SD = 0.5, CV = 2.7% Sample 3 (47 nmol/L): SD = 1.3, CV = 2.7% Sample 4 (90 nmol/L): SD = 2.6, CV = 2.9% Sample 5 (198 nmol/L): SD = 5.1, CV = 2.6%
    LinearityThe assay must demonstrate linearity throughout its analytical measuring interval.Linear throughout the analytical measuring interval of 0.33 nmol/L - 200 nmol/L.
    Limit of Blank (LoB)Not explicitly stated as a number, but derived from the study.0.005 nmol/L (maximum observed)
    Limit of Detection (LoD)Not explicitly stated as a number, but derived from the study.0.01 nmol/L
    Limit of Quantitation (LoQ)Not explicitly stated as a number, but derived from the study.0.06 nmol/L

    Study Details

    The provided document describes analytical verification studies for the Access SHBG assay, not a clinical study involving human readers or AI. Therefore, some of the requested information (e.g., MRMC studies, human reader improvement with AI, ground truth for training AI models, number of experts for AI ground truth) is not applicable or cannot be extracted from this document as it pertains to an immunoassay device, not an AI/ML-based diagnostic system.

    Here's the relevant information that can be extracted:

    2. Sample Size and Data Provenance:

    • Method Comparison: A total of 151 samples were evaluated. The data provenance is implied to be clinical samples (patient samples are mentioned in the CLSI guideline for method comparison), though specific country of origin or whether they were retrospective/prospective is not stated. Given the context of a 510(k) submission for an in-vitro diagnostic, it is highly likely these were de-identified retrospective clinical samples.
    • Precision: Five serum samples with varying SHBG concentrations were used. Each sample was assayed in duplicate with two runs per day, over 20 days, on three Dxl 9000 Access Immunoassay Analyzer systems, three reagent lots, and three calibrator lots. This totals 80 measurements per sample (5 samples * 2 duplicates * 2 runs/day * 20 days / 1 instrument * 1 reagent lot * 1 calibrator lot used for reported results).
    • Linearity: A native low sample and a spiked high sample were used, along with seven mixtures in between.
    • LoB, LoD, and LoQ: Four distinct blank samples were used for LoB. Six to seven samples were used for LoD. 12-13 serum samples were used for LoQ.

    3. Number of Experts and Qualifications for Test Set Ground Truth:

    • This is an immunoassay device, not an image-based AI system. The ground truth for analytical studies like linearity, precision, and limits of detection is established by the assay itself (measurements of known concentrations, serial dilutions, etc.) and validated against established analytical method guidelines (CLSI). No human experts are involved in establishing ground truth for these analytical performance characteristics in the way they would be for image interpretation tasks.

    4. Adjudication Method for Test Set:

    • Not applicable. Adjudication methods (e.g., 2+1, 3+1) are typically used in clinical studies where human readers provide interpretations (e.g., radiology reads) that need to be reconciled to form a ground truth. For analytical performance studies of an immunoassay, the "truth" is based on the chemical and instrument measurements and statistical analysis against predefined acceptance criteria.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • No, an MRMC comparative effectiveness study was not done. This type of study is relevant for AI systems assisting human readers in diagnostic tasks, such as radiology image interpretation. This document describes the analytical performance of an immunoassay for quantitative determination of SHBG levels, which does not involve human readers interpreting AI output.

    6. Standalone Performance:

    • Yes, the entire document focuses on the "standalone" analytical performance of the Access SHBG assay on the Dxl 9000 Access Immunoassay Analyzer. The performance described (method comparison, precision, linearity, limits) is the direct output of the instrument and reagents, without human interpretation "in the loop" beyond standard laboratory procedures for operating the instrument.

    7. Type of Ground Truth Used:

    • For analytical performance studies, the "ground truth" is established through:
      • Reference measurements/Predicate device: For method comparison, the results from the previously cleared predicate device (Access SHBG on Access 2 instrument) served as the comparative "truth."
      • Known concentrations/mixtures: For linearity, samples with known or precisely diluted concentrations were used.
      • Replicate measurements: For precision, repeated measurements establish the variability.
      • Blank and low-level samples: For LoB, LoD, and LoQ, these are the "truth" against which the assay's detection capabilities are measured.
      • CLSI Guidelines: The studies were designed and evaluated according to CLSI guidelines, which represent a form of accepted scientific and statistical "truth" for validating laboratory assays.

    8. Sample Size for Training Set:

    • This document describes the validation of an immunoassay device, not an AI/ML model. Therefore, there isn't a "training set" in the context of machine learning. The device's underlying chemistry and physics are "trained" during its development and manufacturing process, but not through a data-driven training set in the way an AI algorithm is.

    9. How Ground Truth for Training Set was Established:

    • Not applicable, as there is no AI/ML training set in the context of this immunoassay device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K223405

    Validate with FDA (Live)

    Date Cleared
    2023-01-13

    (65 days)

    Product Code
    Regulation Number
    862.1680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Access Testosterone assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of total testosterone levels in human serum and plasma using the Access Immunoassay Systems.

    Measurement of testosterone are used in the diagnosis and treatment of disorders involving the male sex hormones (androgens), including primary and secondary hypogonadism, delayed or precocious puberty, impotence in males and, in females hirsutism (excessive hair) and virilization) due to tumors, polycystic ovaries, and adrenogenital syndromes.

    Device Description

    The Access Testosterone assay is a competitive binding immunoenzymatic assay. The Access Testosterone assay consists of the reagent pack and calibrators. Other items needed to run the assay include substrate and wash buffer. The Access Testosterone assay reagent pack, Access Testosterone assay calibrators, along with the UniCel Dxl wash buffer II are designed for use with the Dxl 9000 Access Immunoassay Analyzer in a clinical laboratory setting.

    AI/ML Overview

    The provided text describes the analytical performance verification of an in vitro diagnostic device (Access Testosterone Assay) and its substantial equivalence to a previously cleared device, rather than an AI-powered medical device performing image analysis or similar tasks that typically involve human experts for ground truth and MRMC studies.

    Therefore, many of the requested elements for an AI device (like human expert adjudication, MRMC studies, effect size of human improvement with AI, or detailed ground truth establishment for a training set) are not applicable to this document.

    However, I can extract the relevant information regarding the acceptance criteria and study that proves the device meets them for this in vitro diagnostic device.

    Here's the information based on the provided text:

    Acceptance Criteria and Device Performance for Access Testosterone Assay

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance CriteriaReported Device Performance
    Method ComparisonR² ≥ 0.90 and Slope of 1.00 ± 0.14R² = 0.98, Slope = 0.95 (95% CI: 0.93, 0.98), Intercept = 0.028 (95% CI: -0.015, 0.071)
    Linearity (Analytical Measuring Interval)Linear throughout the analytical measuring interval of 0.4 - 16.0 ng/mLLinear throughout the analytical measuring interval of 0.4 - 16.0 ng/mL
    Imprecision (Within-Laboratory)≤ 0.14 ng/mL SD at concentrations ≤ 1.4 ng/mLSample 1 (0.71 ng/mL): SD = 0.04 (meets criteria)
    ≤ 10.0% CV at concentrations > 1.4 ng/mLSample 2 (2.0 ng/mL): CV = 4.4% (meets criteria)
    Sample 3 (4.8 ng/mL): CV = 3.9% (meets criteria)
    Sample 4 (7.1 ng/mL): CV = 4.6% (meets criteria)
    Sample 5 (8.6 ng/mL): CV = 5.9% (meets criteria)
    Sample 6 (14 ng/mL): CV = 7.8% (meets criteria)
    Limit of Blank (LoB)Not explicitly stated as a numerical acceptance criterion, but determined and accepted.LoB = 0.2 ng/mL
    Limit of Detection (LoD)Not explicitly stated as a numerical acceptance criterion, but determined and accepted.LoD = 0.4 ng/mL
    Limit of Quantitation (LoQ)Not explicitly stated as a numerical acceptance criterion, but determined and accepted.20% Within-Laboratory CV LoQ = 0.4 ng/mL

    2. Sample Size Used for the Test Set and Data Provenance

    • Method Comparison: 108 serum samples.
    • Linearity: Low sample (native serum), high sample (native serum spiked with Testosterone antigen), and mixtures of low and high samples. (Specific N not further detailed for mixtures, but implies multiple points across measuring interval).
    • Imprecision: Six (6) serum samples (2 native, 3 pooled native, 1 spiked native). Each tested in replicates of two (2) per run, with two (2) runs per day for a minimum of 20 days on each instrument and reagent lot combination. (N=88 for each sample shown in table corresponds to total replicates across the study design).
    • LoB: Four blank samples.
    • LoD: Seven serum samples containing low levels of Testosterone.
    • LoQ: Twelve serum samples containing low levels of Testosterone.
    • Data Provenance: Not explicitly stated (e.g., country of origin, retrospective/prospective). However, given it's an IVD, samples would be collected for analytical performance testing.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts

    • Not applicable (N/A). This device is an in vitro diagnostic assay for quantitative determination of a hormone level. Ground truth (or reference values) for analytical performance are established through established analytical methods, reference materials (e.g., USP Reference Material for Standardization and Traceability), and quantitative measurement procedures, not by human expert opinion or interpretation of images.

    4. Adjudication Method for the Test Set

    • N/A. Adjudication by human experts is not relevant for establishing the analytical performance of an in vitro diagnostic assay.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and its effect size of human readers improve with AI vs. without AI assistance

    • N/A. This is not an AI-powered device involving human readers or interpretation. It's an in vitro diagnostic assay.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • Yes, in essence. The studies described (Method Comparison, Linearity, Imprecision, LoB/LoD/LoQ) demonstrate the analytical performance of the device itself (the Access Testosterone Assay on the Dxl 9000 Access Immunoassay Analyzer) in a standalone, automated manner. There is no human interpretation component where performance needs to be evaluated.

    7. The Type of Ground Truth Used

    • Analytical Ground Truth:
      • Reference Method/Predicate Device: For method comparison, the "previously cleared system" (Access Testosterone assay on the Access 2 instrument) serves as the reference for comparison, indicating that its results are considered the comparative "ground truth."
      • Known Concentrations/Standards/Reference Materials: For linearity, imprecision, and limit studies, samples with known or characterized concentrations (e.g., native serum, spiked samples, blank samples) and established reference materials (USP Reference Material) are used.
      • Statistical Models/Calculations: LoB, LoD, and LoQ are determined statistically based on measurements of blank and low-level samples, following CLSI guidelines.

    8. The Sample Size for the Training Set

    • N/A. This is an in vitro diagnostic device, not an AI/machine learning model that undergoes a "training" phase with a data set in the conventional sense. The "training" in manufacturing involves calibrating the instrument and reagents according to established protocols, not learning from a dataset.

    9. How the Ground Truth for the Training Set was Established

    • N/A. As above, there is no "training set" in the context of an AI model for this device. Calibration and quality control procedures ensure the analytical performance by using calibrators and controls with known values, which are traceable to reference materials (e.g., USP Reference Material).
    Ask a Question

    Ask a specific question about this device

    K Number
    K211685

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-05-06

    (338 days)

    Product Code
    Regulation Number
    862.1680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Immunoassay for the in vitro quantitative determination of testosterone in human serum and plasma.
    The electrochemiluminescence immunoassay "ECLIA" is intended for use on the cobas e 601 immunoassay analyzer.Measurements of testosterone are used in the diagnosis and treatment of disorders involving the male sex hormones (androgens), including primary and secondary hypogonadism, delayed or precocious puberty, impotence in males and, in females hirsutism (excessive hair) and virilization) due to tumors, polycystic ovaries, and adrenogenital syndromes.

    Device Description

    The Elecsys Testosterone II immunoassay makes use of a competitive test principle using streptavidin-coated microparticles and electrochemiluminescence detection. Results are determined using a calibration curve that is generated specifically on each instrument by a 2point calibration and master curve provided with the reagent bar code. The Elecsys Testosterone II reagent kit consists of a Reagent Pack (R1, R2, and M (Streptavidin-coated microparticles)).

    AI/ML Overview

    The Elecsys Testosterone II device is an immunoassay for the in vitro quantitative determination of testosterone in human serum and plasma. The device is intended for use on the cobas e 601 immunoassay analyzer.

    1. Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the Elecsys Testosterone II are based on various analytical performance characteristics. The table below summarizes some of the key performance criteria and the reported performance of the candidate device (Elecsys Testosterone II, K211685) in comparison to its predicate device (Elecsys Testosterone II, K093421).

    FeatureAcceptance Criteria (Predicate)Reported Performance (Candidate)
    PrecisionRefer to CLSI EP05-A21-day precision: Evaluated according to CLSI guideline EP05-A3. Protocol: 2 replicates of 2 controls and 5 samples, 2 runs/day over 21 days with 1 reagent lot. Repeatability and intermediate precision calculated. 5-day precision: Evaluated on one cobas e 601 analyzer according to CLSI guideline EP05-A3. Protocol: 5 aliquots of each control (PreciControl Universal Level 1 & 2) and human serum samples per run, 1 run/day for 5 days with 3 lots. Repeatability and intermediate precision calculated.
    LoB1.2 ng/dL1.50 ng/dL or 0.052 nmol/L
    LoDSame as predicate2.50 ng/dL or 0.087 nmol/L
    LoQSame as predicate12.0 ng/dL or 0.416 nmol/L
    Measuring Range2.50-1500 ng/dL or 0.087-52.0 nmol/L2.50-1500 ng/dL or 0.087-52.0 nmol/L (defined by Limit of Detection and maximum of master curve).
    Cross-ReactivityVarious specific valuesDHEA-S: 0.003%, Androstenedione: 3.15%, Danazol: 0.504%, Estradiol: 0.211%, Ethisterone: 3.57%, 19-Norethisterone: 5.51%, Norgestrel: 0.539%, △5-Androstene-3β17β-diol: 0.289%, Testosterone propionate: 0.718%, 5α-Androstane-3β, 17β-diol: 2.15%, 5α-Dihydrotestosterone: 1.30%, 11β-OH-Testosterone: 20.6%, 11keto-Testosterone: 4.87%, Prednisone: n.d., Prednisolone: n.d., Progesterone: 0.009%, Cortisol: n.d., Cortisone: n.d., Dexamethasone: n.d., Estrone: n.d., DHEA: 0.014%.
    Biotin Interference< 30 ng/mLNo biotin interference in serum concentrations up to 1200 ng/mL.
    Special Drug InterferenceNoneTestosterone Undecanoate or Nandrolone strongly interfered with Testosterone and produced elevated recovery values.
    Method ComparisonPassing/Bablok: y = 0.961x - 0.015, T = 0.975; Linear regression: y = 0.958x - 0.006, r = 0.999Passing/Bablok: y = 0.948x + 0.023, T = 0.976; Linear regression: y = 0.950x - 0.871, r = 0.998. The data from analytical studies demonstrate the device is as safe, as effective, and performs as well as or better than the predicate.

    2. Sample Size Used for the Test Set and Data Provenance

    • Precision (5-Day): 5 aliquots of each control (PreciControl Universal Level 1 and Level 2) and human serum samples per run, 1 run per day for 5 days with 3 lots. (Implies multiple individual human serum samples but the exact number isn't specified, and the controls are commercial products).
    • Precision (21-Day): 2 replicates of 2 controls and 5 samples, 2 runs per day over 21 days using 1 reagent lot. (Implies 5 individual human serum samples, plus two controls).
    • Linearity: One human serum sample with high analyte content was serially diluted (25 steps).
    • Endogenous Interferences: Not explicitly stated, but implies analysis of testosterone concentrations with and without spiked interfering substances.
    • Biotin Interference: Not explicitly stated, but implies multiple testosterone samples spiked with biotin.
    • Common Drug Interferences: Not explicitly stated, but implies multiple samples tested with common drugs.
    • Special Drug Interferences: Samples (with testosterone concentrations near 0.5 ng/mL and near 5.0 ng/mL) were divided into aliquots and spiked with Testosterone Undecanoate. The number of samples is not explicitly stated.
    • Analytical Specificity/Cross-Reactivity: One testosterone level (0.5 ng/mL) in human serum matrix was spiked with potential cross-reactants. The number of samples is not explicitly stated.
    • Sample Matrix Comparison: At least 40 serum/plasma pairs (Native or spiked samples).
    • Method Comparison to Predicate: 168 samples (Native single samples and spiked single samples).

    Data Provenance: The studies used human serum and plasma samples, as well as control materials. The specific country of origin is not explicitly stated, but Roche Diagnostics has facilities in Indianapolis, IN (USA), Mannheim, Germany, and Penzberg, Germany, suggesting a potential international origin or a combination of origins for their sample collection. The studies described are analytical performance evaluations, which are typically prospective in design for device validation.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This device is an in vitro diagnostic (IVD) quantitative assay for testosterone. The "ground truth" for such assays typically refers to the reference method (e.g., ID-GC/MS) or the established value of control materials. It does not involve expert readers reviewing images or clinical data to establish a diagnostic truth. Therefore, the concept of "experts" in the context of interpretation of results is not directly applicable here. The establishment of ground truth for accuracy and calibration is based on:

    • Traceability/Standardization: ID-GC/MS ("Isotope Dilution - Gas Chromatography/Mass Spectrometry"). This is a highly accurate and precise analytical method that serves as a reference.
    • Calibrators: Testosterone II CalSet II (Calibrators 1 and 2) are used, which are themselves traceable to the ID-GC/MS method.

    4. Adjudication Method for the Test Set

    Not applicable. As described above, this is an IVD assay, not an imaging-based diagnostic where adjudication of expert readings would be necessary. The performance is assessed by comparing results to reference methods or known concentrations, or through statistical analyses of precision and linearity.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No MRMC comparative effectiveness study was mentioned. This type of study is typically relevant for interpretative diagnostic devices (e.g., radiology AI tools) where human readers interpret cases with and without AI assistance. The Elecsys Testosterone II is a laboratory assay that provides a quantitative numerical result, and therefore, does not involve human readers interpreting "cases" in the same way.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    The performance studies described (Precision, LoB, LoD, LoQ, Linearity, Interferences, Cross-Reactivity, Sample Matrix Comparison, Method Comparison) are all standalone performance evaluations of the assay system (reagents + instrument) without human-in-the-loop interpretation. The device's primary function is to quantitatively measure testosterone, and its performance is assessed directly against analytical standards and reference methods.

    7. The Type of Ground Truth Used

    • Reference Method: ID-GC/MS (Isotope Dilution - Gas Chromatography/Mass Spectrometry) serves as the primary reference method for traceability and standardization.
    • Known Concentrations: Control materials (e.g., PreciControl Universal) and spiked samples with known concentrations of analytes and interferents are used.
    • Predetermined Clinical Values: Samples for method comparison include those that span the measuring range, implying a range of clinically relevant testosterone levels.

    8. The Sample Size for the Training Set

    The document describes analytical performance studies for validation (test set), but does not explicitly mention a "training set" in the context of algorithm development. Immunoassays like the Elecsys Testosterone II are typically developed and optimized using a range of samples during the research and development phase to establish reagent formulations, reaction kinetics, and calibration curves. However, the exact sample sizes for this initial developmental "training" are not usually detailed in 510(k) summaries, which focus on the validation data.

    9. How the Ground Truth for the Training Set Was Established

    Similar to point 8, the document does not detail the "training set" for algorithm development. For IVD assays, the "training" involves establishing the master curve and optimizing reagents to accurately measure testosterone across its dynamic range. This optimization would rely on:

    • Characterized reference materials: Samples with known testosterone concentrations, often established by highly accurate reference methods like ID-GC/MS.
    • Clinical samples: A broad range of human serum and plasma samples from various populations (males, females, different age groups, healthy, and those with relevant disorders) would be used to ensure the assay performs reliably across the intended use population.
    • Spiked samples: Samples with known additions of testosterone or interfering substances to evaluate various aspects of assay performance.
    Ask a Question

    Ask a specific question about this device

    K Number
    K201908

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-10-15

    (98 days)

    Product Code
    Regulation Number
    862.1680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The LIAISON® Testosterone xt is a direct, competitive, chemiluminescence immunoassay (CLIA) intended for the quantitative determination of testosterone in human serum and EDTA plasma on the LIAISON® XL Analyzer. The assay is intended for in vitro diagnostic use.

    Measurement of testosterone is used in the diagnosis and treatment of disorders involving the male sex hormones (androgens), including primary and secondary hypogonadism, delayed or precocious puberty, impotence in males and, in females hirsutism (excessive hair) and virilization (masculinization) due to tumors, polycystic ovaries, and adrenogenital syndromes.

    The test has to be performed on the LIAISON® XL Analyzer.

    Device Description

    The LIAISON® Testosterone xt assay's method for quantitative determination of testosterone is a direct, competitive, chemiluminescence immunoassay (CLIA). Specific antibody to testosterone is bound to magnetic particles (solid phase) and testosterone is linked to an isoluminol derivative. During the incubation, testosterone is dissociated from its binding protein and competes with labeled testosterone for binding sites on the antibody. After the incubation, the unbound material is removed with a wash cycle. Subsequently, the starter reagents are added and a flash chemiluminescent reaction is initiated. The light signal is measured by a photomultiplier as relative light units (RLU) and is inversely proportional to the concentration of testosterone present in calibrators, controls, or samples.

    The LIAISON® Testosterone xt is an in vitro diagnostic device consisting of reagents provided in individual compartments within a plastic container called the Reagent Integral. The components provided in the unitized Reagent Integral include: PMP (paramagnetic particles), conjugate and assay buffer. All reagents in the integral are supplied ready to use. The assay configuration for the LIAISON® Testosterone xt allows for the performance of 100 tests.

    The two-point calibrators are provided in the same kit box, but separate from the Reagent Integral. The two-point calibrators are supplied ready to use.

    The LIAISON® Testosterone xt assay is performed on the LIAISON® XL Analyzer (Model 10050; originally FDA cleared under K103529), a fully automated system with continuous loading combining the chemiluminescence technology with magnetic microparticles as the solid phase.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the LIAISON® Testosterone xt device, extracted from the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Performance CharacteristicAcceptance Criteria (Predicate Device)Reported Device Performance (LIAISON® Testosterone xt)
    Assay Measuring Range0.16 - 15.0 ng/mLExtended: 0.024 – 15.0 ng/mL
    Limit of Blank (LoB)≤ 0.031 ng/mLImproved: ≤ 0.005 ng/mL
    Limit of Detection (LoD)0.098 ng/mLImproved: 0.010 ng/mL
    Limit of Quantitation (LoQ)0.160 ng/mLImproved: 0.024 ng/mL
    Method Comparisony = 0.9390x – 0.1002 (vs. commercially available immunoassay) 95% CI for slope: 0.92 to 0.96 95% CI for intercept: -1.88 to 1.80 ng/dLy = 0.99x – 1.77 ng/dL (vs. CDC HoSt Testosterone RMP ID-LC-MS/MS values) 95% CI for slope: 0.97 to 1.02 95% CI for intercept: -3.22 to -0.35 ng/dL
    Precision (Total/Across Lots %CV)7.9% – 14.0%Improved: 3.5% – 7.9 %
    Linearity (Serum)Observed Analyte = 0.9942x – 16.062; R² = 0.9959Observed Analyte = 0.995x + 0.0346; R² = 0.9928
    Linearity (SST Serum)Observed Analyte = 1.0188x - 14.531; R² = 0.9965Observed Analyte = 1.0225x - 57.853; R² = 0.9914
    Linearity (EDTA plasma)Observed Analyte = 1.0057x - 13.029; R² = 0.9913Observed Analyte = 1.0337x - 31.889; R² = 0.9955
    Recovery (Mean %)97%99%
    Recovery (Range %)91% - 105%93% - 105%
    Open Use Stability: Reagent Integral4 weeksExtended: 8 weeks
    Open Use Stability: Calibrators4 weeksExtended: 8 weeks
    Calibration Curve Stability7 daysExtended: 28 days

    2. Sample Size for the Test Set and Data Provenance

    The provided document does not explicitly state the sample sizes used for each specific test (e.g., method comparison, precision, linearity, recovery). It only summarizes the results in "Table 6-2: Comparison to Predicate Device" and refers to "verification and validation activities" without detailing the exact number of samples for each.

    The data provenance is not specified in terms of country of origin. The studies appear to be retrospective analyses of device performance characteristics, as they involve testing the LIAISON® Testosterone xt against existing methods or conditions to demonstrate equivalence and improved performance over the predicate device.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This information is not explicitly provided in the document. For the "Method Comparison," it refers to "commercially available immunoassay" and "CDC HoSt Testosterone RMP ID-LC-MS/MS values." The latter, CDC HoSt Testosterone RMP ID-LC-MS/MS, likely represents a highly accurate and standardized reference method, which serves as a widely accepted ground truth in laboratory medicine, rather than relying on individual expert consensus for each measurement. No human experts are mentioned for ground truth establishment.

    4. Adjudication Method for the Test Set

    No explicit adjudication method is mentioned. The studies focus on direct quantitative analytical performance comparisons rather than subjective human interpretation.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable. The LIAISON® Testosterone xt is an in vitro diagnostic device, specifically a chemiluminescence immunoassay (CLIA), for the quantitative determination of testosterone. It is an automated laboratory test, not an AI-assisted diagnostic imaging or interpretation tool that involves human readers. Therefore, an MRMC comparative effectiveness study involving human readers and AI is not relevant to this device.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes, this is effectively a standalone device performance study. The LIAISON® Testosterone xt is an automated system where the measurement of testosterone is performed by the analyzer (LIAISON® XL Analyzer) without human intervention in the analytical process itself. The performance metrics (LoB, LoD, LoQ, precision, linearity, recovery) are all measures of the algorithm's and instrument's direct analytical capability.

    7. The Type of Ground Truth Used

    The ground truth used for method comparison was:

    • For the predicate device: A "commercially available immunoassay."
    • For the LIAISON® Testosterone xt: "CDC HoSt Testosterone RMP ID-LC-MS/MS values." This is a highly accurate, reference method-based measurement, representing the gold standard for testosterone quantification. Other ground truths were established by controlled experiments for limits (e.g., dilution series for LoB, LoD, LoQ), spiked samples for recovery, and replicated measurements for precision.

    8. The Sample Size for the Training Set

    The document does not explicitly mention a "training set" in the context of machine learning or AI. This device is an immunoassay, not an AI model that requires a distinct training phase with a labeled dataset in the typical sense. Its development would involve analytical validation using various samples to optimize reagent formulation, assay parameters, and calibration, but not as a machine learning training set.

    9. How the Ground Truth for the Training Set was Established

    As explained above, there is no "training set" in the machine learning sense for this immunoassay. The development and validation of an immunoassay involve:

    • Reference materials: Use of certified reference materials or reference methods (like ID-LC-MS/MS) to establish accurate values for calibrators and controls.
    • Spiked samples: Samples with known concentrations of testosterone added.
    • Clinical samples: Testing a range of patient samples representing the intended use population, comparing results to established methods.
    • Statistical analysis: Extensive statistical methods are used during development and validation to ensure accuracy, precision, linearity, and other performance characteristics.
    Ask a Question

    Ask a specific question about this device

    K Number
    K191350

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-12-20

    (214 days)

    Product Code
    Regulation Number
    862.1680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Radioimmunoassay for the quantitative measurement of free testosterone in human serum. This assay is intended for in vitro diagnostic use.

    Free testosterone test is used in the diagnosis and treatment of disorders involving the male sex hormones (androgens), including primary and secondary hypogonadism, delayed or precocious puberty, impotence in males and, in female's hirsutism (excessive hair) and virilization (masculinization) due to tumors, polycystic ovaries, and adrenogenital syndromes.

    Device Description

    The radioimmunoassay of free testosterone is a competitive assay. The procedure follows the basic principle of radioimmunoassay where there is competition between a radioactive and a non-radioactive antigen for a mixed number of antibody binding sites. The amount of [125]]-labeled testosterone analog bound to the antibody is inversely proportional to the concentration of unlabeled free testosterone present. The separation of the free and bound antigen is achieved by decanting or aspirating the antibody-coated tubes. A standard curve is constructed and unknown free testosterone values are obtained from the curve by interpolation.

    Kit Content:
    Free Testosterone Antibody-Coated Tubes: 2 x 50 tubes (ready-to-use)
    125|-labeled Testosterone Analog Tracer (YELLOW): one 22 mL vial (ready-to-use)
    Calibrators: one 1.0 mL vial labeled 0, and seven 0.5 mL vials labeled 1-7 (ready-to-use)
    Controls: two 0.5 mL vials labeled 1, 2 (ready-to-use)

    AI/ML Overview

    The provided text describes the Immunotech ACTIV\u00AE Free Testosterone RIA device, a radioimmunoassay for the quantitative measurement of free testosterone in human serum. This is an in vitro diagnostic device and the information provided is a 510(k) summary, which is typically for demonstrating substantial equivalence to a legally marketed predicate device.

    Here's an analysis of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present "acceptance criteria" for overall device performance in the form of pass/fail thresholds for clinical utility. Instead, it provides a summary of the analytical performance characteristics and reference ranges, which are the results of validation studies conducted to support the device's claims and demonstrate its performance.

    Performance CharacteristicAcceptance Criteria (Implicit from CLSI guidelines and typical IVD requirements)Reported Device Performance
    Sensitivity- Limit of Blank (LoB) < 0.1 pg/mL (typical for low-end sensitivity) - Limit of Detection (LoD) < 0.2 pg/mL (typical) - Limit of Quantitation (LoQ) < 0.5 pg/mL (typical)LoB: 0.05 pg/mL LoD: 0.13 pg/mL LoQ: 0.37 pg/mL
    SpecificityHigh specificity for free testosterone, low cross-reactivity with related molecules.Antibody highly specific for free testosterone. Low cross-reactivity (< 0.1% for most, up to 0.116% for 19-Nortestosterone in depleted serum) for various related compounds.
    Precision (Repeatability)Coefficient of Variation (CV) <= 15% (typical for IVDs, especially at lower concentrations)<= 13.2% for serum samples
    Precision (Within-Laboratory)Coefficient of Variation (CV) <= 20% (typical for IVDs)<= 19.3% for serum samples
    Measurement RangeRange should cover clinically relevant concentrations.0.37 pg/mL (LoQ) to ~100 pg/mL (highest calibrator)
    InterferenceMinimal interference from common endogenous substances and medications at specified concentrations.Tested against various substances (e.g., hemoglobin, bilirubin, biotin, ibuprofen, cholesterol, certain hormones) at specified concentrations, implying no significant interference.
    LinearityLinear response across the entire measuring range.Linear throughout the measuring range of the kit (0.30 \u2013 114.11 pg/mL).
    Cross-reactivityLow cross-reactivity to structurally similar compounds.Very low cross-reactivity (often ND or < 0.1%) for a panel of 20+ compounds in pooled normal serum and pooled depleted serum. Specific percentages are provided.
    Method ComparisonGood agreement/correlation with the predicate device (e.g., slope close to 1, intercept close to 0, high R-value).n=278; Range: 0.26 \u2013 101.21 pg/mL; Slope: 1.0078; Intercept: 0.0759; R: 0.9956 (indicating very good agreement with the predicate)
    Expected Values/Reference RangesEstablished stratification for different populations (age, sex, physiological state).Comprehensive tables for Men (by age group), Women (by cycle phase, menopausal status, and contraceptives), Boys (by age group), and Girls (by age group).

    2. Sample Size Used for the Test Set and Data Provenance

    The document describes specific sample sizes for different analyses:

    • Method Comparison: N = 278 samples were used to compare the proposed device with the predicate. The dataProvenance is implicitly retrospective as it compares results from two versions of a device, likely using banked samples or samples run consecutively.
    • Expected Values/Reference Ranges:
      • Men: N = 120 (30 per age group: 20-29, 30-39, 40-49, 50+).
      • Women: N = 120 (32 follicular, 32 luteal, 21 preovulatory peak, 13 postmenopausal, 19 contraceptives \u2013 totaling 117 individual samples given in the subcategories, implying some overlap or missing a few for the total of 120 "Random Women").
      • Boys: N = 114 (18 infant, 48 child, 30 adolescent in one grouping; 36 6m-9y, 20 10-11y, 20 12-13y in another grouping - these age ranges seem to be different categorizations of the same underlying data or separate cohorts).
      • Girls: N = 119 (34 infant, 55 child, 30 adolescent in one grouping; 69 6m-9y, 20 10-12y, 20 13-16y in another grouping - similar to boys, likely different categorizations or cohorts).

    The data provenance is not explicitly stated (e.g., country of origin). Given the manufacturer is based in the Czech Republic, the samples could be from Europe, but this is not confirmed. The nature of these studies (establishing reference ranges) suggests they are prospective or retrospective collections specific for this type of validation.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This section is not applicable to this type of device and study. The device, Immunotech ACTIV\u00AE Free Testosterone RIA, is an in vitro diagnostic (IVD) test that measures a quantitative biomarker concentration (free testosterone in serum). The "ground truth" for such a device is established through:

    • Reference measurement procedures (RMPs) or reference methods, which are highly accurate and precise analytical techniques.
    • Certified reference materials (CRMs) with known concentrations.
    • Comparisons to a legally marketed predicate device (as is the case here).

    Experts, clinical consensus, or pathology are typical for image-based diagnostic AI, not for quantitative chemical assays.

    4. Adjudication Method for the Test Set

    This section is not applicable for this type of in vitro diagnostic device. Adjudication methods (like 2+1, 3+1) are typically used in clinical studies or studies involving human interpretation (e.g., radiology for AI interpretation) to establish a consensus ground truth. For an IVD, the measurement itself using a validated method is the "truth," or the comparison is made against a validated predicate device.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. MRMC studies are specific to evaluating devices (often AI-powered) that assist human readers (e.g., radiologists, pathologists) in making diagnoses. This device is a quantitative immunoassay, not a reader-assisted diagnostic tool.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, the performance characteristics provided (sensitivity, specificity, precision, measurement range, interference, linearity, cross-reactivity, method comparison) represent the standalone performance of the assay itself, as an in vitro diagnostic test. There is no human intervention in the result generation beyond operating the instrument and following the assay procedure.

    7. The Type of Ground Truth Used

    The ground truth for evaluating the analytical performance of the device is established using:

    • Internal standards for calibrator values.
    • Reference materials/samples with known concentrations for sensitivity, linearity, and interference studies (though not explicitly detailed, this is standard practice for CLSI guidelines).
    • Comparison to a legally marketed predicate device (ACTIV\u00AE Free Testosterone RIA, K952281) for method comparison, which serves as a high-quality comparator given its prior FDA clearance.
    • Clinically characterized samples (e.g., from different age groups, sexes, and physiological states) to establish expected values/reference ranges.

    The document implicitly uses these standard analytical validation approaches rather than, for instance, pathology or outcomes data (which would be for clinical utility, not analytical performance).

    8. The Sample Size for the Training Set

    This document only details validation and verification activities, not the development or "training" of the assay. For a traditional immunoassay like this RIA, there isn't a "training set" in the context of machine learning or AI. The assay's "design" is based on biochemical principles (antibody-antigen binding) rather than statistical learning from a dataset. The formulation of reagents, selection of antibody, and optimization of the protocol would involve internal R&D, but not a formally defined "training set" with ground truth in the AI sense.

    9. How the Ground Truth for the Training Set Was Established

    As stated above, this is not applicable for a traditional immunoassay. The concept of a "training set" and "ground truth" establishment in the context of AI does not directly translate to the development process of this type of diagnostic kit.

    Ask a Question

    Ask a specific question about this device

    K Number
    K191533

    Validate with FDA (Live)

    Date Cleared
    2019-08-28

    (79 days)

    Product Code
    Regulation Number
    862.1680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ADVIA Centaur® Testosterone II (TSTII)

    The ADVIA Centaur® Testosterone II (TSTI) assav is for in the quantitative determination of total testosterone (bound and unbound) in human serum and plasma using the ADVIA Centaur XP system.

    Measurements of testosterone are used in the diagnosis and treatment of disorders involving the male sex hormones (androgens), including primary and secondary hypogonadism, delayed or precocious puberty, impotence in males and, in females, hirsutism (excessive hair) and virilization) due to tumors, polycystic ovaries, and adrenogenital syndromes.

    ADVIA Centaur® SHBG

    The ADVIA Centaur® SHBG assay is an in vitro diagnostic immunoassay for the quantitative determination of sex hormone-binding globulin (SHBG) in human serum and plasma using the ADVIA Centaur XP system.

    The ADVIA Centaur SHBG assay is intended for use as an aid in the diagnosis of androgen disorders.

    Device Description

    The re-standardized ADVIA Centaur® Testosterone II (TSTII) and the ADVIA Centaur® SHBG are in vitro diagnostic immunoassays for use on the ADVIA Centaur® XP system. The ADVIA Centaur® Testosterone II (TSTII) assay quantitatively determines total testosterone in human serum and plasma. The ADVIA Centaur® SHBG assay quantitatively determines sex hormone-binding globulin (SHBG) in human serum and plasma. The submission includes details on the ingredients of the reagents for both assays.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the ADVIA Centaur® Testosterone II (TSTII) and ADVIA Centaur® SHBG based on the provided text, formatted as requested:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Performance CharacteristicAcceptance Criteria (ADVIA Centaur® Testosterone II (TSTII))Reported Device Performance (ADVIA Centaur® Testosterone II (TSTII))Acceptance Criteria (ADVIA Centaur® SHBG)Reported Device Performance (ADVIA Centaur® SHBG)
    Detection Limits
    Limit of Blank (LoB)(Implicitly good performance relative to predicate)2.50 ng/dL(Not explicitly provided in this document)(Previously established, see K151986 & K091867)
    Limit of Detection (LoD)(Implicitly good performance relative to predicate)5.00 ng/dL (95% probability)(Not explicitly provided in this document)(Previously established, see K151986 & K091867)
    Limit of Quantitation (LoQ)(Total CV of 20%)7.00 ng/dL (at 20% CV)(Not explicitly provided in this document)(Previously established, see K151986 & K091867)
    Precision(Implicitly good performance relative to predicate)See table below for detailed CVs across various samples and concentrations. Generally, Within-Run CVs are lower than Within-Lab CVs.(Not explicitly provided in this document)(Previously established, see K151986 & K091867)
    LinearityBias from linear fit estimate < 10%Bias from the linear fit estimate was <10% for all 9 samples.(Not explicitly provided in this document)(Previously established, see K151986 & K091867)
    Method ComparisonAcceptable correlation with predicate device and reference methodVs ID-LC-MS/MS: y = 0.97(x) - 0.22 ng/dL (r = 0.98) Vs Predicate ADVIA Centaur TSTII: y = 1.04(x) - 4.14 ng/dL (r = 1.00) Vs Dimension Vista LOCI Total Testosterone (Adult & Pediatric): y = 1.01(x) - 3.32 ng/dL (r = 0.99) Vs Dimension Vista LOCI Total Testosterone (Pediatric only): y = 1.01(x) - 4.74 ng/dL (r = 1.00)(Implicitly good agreement with predicate with new reference intervals not affecting analytical performance)(Previously established, see K151986 & K091867; this submission focused solely on new reference intervals)
    Reference IntervalsEstablished in accordance with CLSI guideline EP28-A3cNew reference intervals established and provided in detail for adult and pediatric populations, stratified by age and Tanner stage.Established in accordance with CLSI guideline EP28-A3cNew reference intervals established for adult male and female populations.
    Biotin InterferenceNo significant effect on the assay at 30 ng/mL% Bias ranged from 2% to 6% for 30 ng/mL biotin at various analyte concentrations. Higher biotin levels (e.g., 250, 500 ng/mL) showed significant bias.Less than or equal to 10% change in results at 300 ng/mL% Bias ranged from -3% to -9% at 300 ng/mL biotin for various analyte concentrations. Higher biotin levels (e.g., 600, 1200 ng/mL) showed greater bias.

    ADVIA Centaur® Testosterone II (TSTII) Precision Results:

    SamplenMean (ng/dL)SD (Repeatability)CV (Repeatability)SD (Within-Lab)CV (Within-Lab)
    Control 180257.389.063.514.385.6
    Control 280636.5745.147.153.128.3
    Control 3801021.9362.286.181.027.9
    MDP18020.921.36.21.868.9
    MDP28073.573.544.85.247.1
    MDP380312.8113.44.323.997.7
    MDP480776.6442.595.558.267.5
    MDP5801123.8359.225.3109.579.7
    Female Patient Serum Pool8015.861.8411.62.3915.1

    2. Sample sizes used for the test set and the data provenance:

    • ADVIA Centaur® Testosterone II (TSTII):

      • Precision: 80 replicates per sample, totaling 9 samples (3 controls, 5 medical decision pools, 1 female patient serum pool). Data provenance not explicitly stated, but typically internal lab data.
      • Linearity: 9 serially diluted samples. Data provenance not explicitly stated.
      • Method Comparison:
        • Vs ID-LC-MS/MS: 108 adult male and female serum samples (7.27-1394.00 ng/dL). Data provenance not explicitly stated, but implied to be patient samples.
        • Vs Predicate ADVIA Centaur TSTII: 108 individual male and female adult serum samples (10.93-1439.77 ng/dL). Data provenance not explicitly stated.
        • Vs Dimension Vista LOCI Total Testosterone: 124 serum samples (79 adult, 45 pediatric). Data provenance not explicitly stated.
          • Pediatric: 45 samples (31 females, ages 7-18; 14 males, ages 22 months-18 years).
      • Reference Intervals: Samples from apparently healthy, normal adult females and males, and pediatric populations. Collected prospectively.
        • Adult Male (Under 50): 250 samples
        • Adult Male (50 and Over): 135 samples
        • Adult Female (Under 50): 224 samples
        • Adult Female (50 and Over): 151 samples
        • Pediatric Male (by Tanner Stage): 101-129 samples per stage
        • Pediatric Male (by Age): 27-149 samples per age group
        • Pediatric Female (by Tanner Stage): 49-138 samples per stage
        • Pediatric Female (by Age): 145-174 samples per age group
      • Biotin Interference: Not explicitly stated, but typically in-house spiked samples.
    • ADVIA Centaur® SHBG:

      • Reference Intervals: Samples from apparently healthy, normal adult females and males. (Presumably collected prospectively, similar to TSTII).
        • Adult Male (Under 50): 250 samples
        • Adult Male (50 and Over): 135 samples
        • Adult Female (Under 50): 224 samples
        • Adult Female (50 and Over): 151 samples
      • Biotin Interference: Not explicitly stated, but typically in-house spiked samples.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • For both devices: The ground truth for the ADVIA Centaur® Testosterone II (TSTII) method comparison was established using ID-LC-MS/MS (Isotope Dilution Liquid Chromatography-Mass Spectrometry/Mass Spectrometry), with samples assigned concentrations by the CDC HoSt RMP ID-LC/MS/MS. This is an analytical reference method, not expert human readers.
    • For the ADVIA Centaur® SHBG, the ground truth for traceability/standardization is stated as WHO 2nd International Standard (08/226), an internationally recognized analytical standard.

    No human experts were used to establish the ground truth for the test set in the traditional sense of image or clinical interpretation. The methods employed are laboratory analytical techniques. When discussing reference intervals for pediatric subjects, it mentions that "Samples were also clinically characterized according to Tanner Stage," which would implicitly involve clinical experts (pediatricians or endocrinologists) for classification, but the number and qualifications are not specified in this document.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • None. The ground truth was established by objective analytical reference methods (ID-LC-MS/MS) and international standards (WHO).

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This document describes laboratory diagnostic devices (immunoassays) for quantitative determination of biomarkers, not AI-assisted image analysis or clinical decision support tools that involve human readers/interpreters in a comparative effectiveness study.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • Yes, this describes standalone performance. The ADVIA Centaur® Testosterone II (TSTII) and ADVIA Centaur® SHBG are automated in-vitro diagnostic assays. The performance metrics (precision, linearity, method comparison, detection limits, etc.) represent the performance of the algorithm/instrument itself without human intervention in the measurement process, beyond sample loading and general operation according to the Instructions for Use.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • Analytical Reference Methods and International Standards:
      • For the ADVIA Centaur® Testosterone II (TSTII), the primary ground truth for method comparison was ID-LC-MS/MS (specifically, CDC HoSt RMP ID-LC/MS/MS).
      • For the ADVIA Centaur® SHBG, the primary ground truth for traceability/standardization is the WHO 2nd International Standard (08/226).
      • For reference intervals, apparently healthy individuals were identified, and for pediatric subjects, Tanner Stage classification was used, which is a clinical assessment.

    8. The sample size for the training set:

    • Not explicitly stated in this document. The document discusses the performance of a re-standardized assay (ADVIA Centaur® Testosterone II (TSTII)) and new reference intervals for another (ADVIA Centaur® SHBG). It does not detail the development or "training" process of the original assays or any underlying algorithms, which would typically involve a separate, often larger, dataset. The studies described are primarily for validation of the modified device and new reference intervals.

    9. How the ground truth for the training set was established:

    • Not explicitly stated in this document. As mentioned above, this document focuses on validation studies of a modified/re-standardized device and new reference intervals for an existing device, rather than the initial development and "training" of the assays. The original development would have involved similar analytical ground truth methods, but the specifics are not provided here.
    Ask a Question

    Ask a specific question about this device

    K Number
    K190121

    Validate with FDA (Live)

    Device Name
    IDS SHBG
    Date Cleared
    2019-06-17

    (143 days)

    Product Code
    Regulation Number
    862.1680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The IDS SHBG assay is an in vitro diagnostic device intended for the quantitative determination of SHBG in human serum or plasma on the IDS System. Results are to be used as an aid in the diagnosis of androgen disorders

    Device Description

    The IDS SHBG assay is an in vitro diagnostic device intended for the quantitative determination of sex hormone binding globulin (SHBG) in human serum and plasma on the IDS-iSYS Multi-Discipline Automated System. Results are to be used in conjunction with other clinical and laboratory data to assist the clinician in the diagnosis of androgen disorders.

    The assay is based on chemiluminescence technology. 5 uL of patient sample or calibrators are incubated with biotinylated monoclonal anti-SHBG antibody, an acridinium labelled monoclonal anti-SHBG conjugate and streptavidin labelled magnetic particles. The magnetic particles are captured using a magnet and a wash step performed to remove any unbound analyte. Trigger reagents are added; the resulting light emitted by the acridinium label is directly proportional to the concentration of analyte in the original sample.

    The IDS SHBG assay is an in vitro diagnostic device consisting of ready to use reagents provided in individual compartments within the reagent cartridge.

    The reagent cartridge contains:

    • Magnetic particles magnetic particles coated with streptavidin in a phosphate buffer containing preservatives
    • -Biotin antibody - monoclonal anti-SHBG labelled with biotin in a buffer containing proteins and preservatives
    • Conjugate monoclonal anti-SHBG labelled with an acridinium ester derivative in a buffer containing proteins and preservatives The calibrators consist of:
    • Calibrators A and B are included in the assay kit. The calibrators consist of a human serum matrix with defined concentrations of SHBG and preservatives. Together with a lot specific master calibration curve, the calibrators will be used to perform adjustment of the master calibration curve.
    AI/ML Overview

    The provided document is a 510(k) summary for the IDS SHBG assay, an in vitro diagnostic device for the quantitative determination of Sex Hormone Binding Globulin (SHBG). The document details the device's performance characteristics and compares it to a predicate device to demonstrate substantial equivalence.

    Here's an analysis of the acceptance criteria and the study proving the device meets them, based on the provided text:

    Key Takeaway: This document describes the validation of an in vitro diagnostic (IVD) assay, not an AI/ML-based diagnostic device that typically involves human readers or image analysis. Therefore, many of the requested categories (e.g., number of experts, adjudication method, MRMC studies, human reader improvement with AI, standalone AI performance) are not applicable to this type of device and study. The ground truth in this context is established by reference methods or validated reference materials, typical for IVD assays.


    1. Table of Acceptance Criteria and Reported Device Performance

    The document describes several analytical performance characteristics that serve as "acceptance criteria" for the IDS SHBG assay. These are primarily related to the accuracy, precision, limits of detection, and specificity of the assay.

    Performance MetricAcceptance Criteria (Implied)Reported Device Performance
    Precision (Reproducibility)Within-run and total precision (CV%) to be within acceptable limits for a quantitative assay.Within Run CV: 1.7% to 3.7% across various SHBG concentration levels (e.g., IQC 1: 1.7% at 5.57 nmol/L; CV 3: 3.7% at 201.67 nmol/L). Total CV: 3.2% to 5.0% across various SHBG concentration levels (e.g., IQC 3: 3.2% at 96.82 nmol/L; CTL3: 5.0% at 93.43 nmol/L). (Compared favorably to predicate which reported 2.5-3.8% within run and 3.1-6.5% total precision.)
    Linearity/Reportable RangeAssay to be linear over its claimed measuring range; accuracy demonstrated for automated dilution.Measuring Range: Linear from 1.60 to 180.00 nmol/L for serum and K2 EDTA plasma (R2 values of 0.999 and 0.998 respectively). Reportable Range: 0.30 to 720.00 nmol/L with automated 1:4 dilution for samples >180 nmol/L. Recovery (Automated Dilution): 87% to 100% when compared to predicate for samples in the 180-720 nmol/L range (Mean recovery: 93%).
    Detection Limits (LoB, LoD, LoQ)Limits should be sufficiently low to meet diagnostic needs.LoB: 0.01 nmol/L LoD: 0.15 nmol/L LoQ: 0.30 nmol/L
    Traceability of CalibratorCalibrator values must fall within specified acceptable ranges and internal quality controls within their respective ranges with defined precision.Calibrator A: 0.10 to 0.30 nmol/L, precision CV ≤ 11%. Verified through internal QC procedures. Calibrator B: 110.00 to 130.00 nmol/L, precision CV ≤ 8%. Verified through internal QC procedures. Internal QC controls: CV ≤ 11% for IQC1, ≤ 8% for IQC2 and IQC3. Results from Table 11 show these were met (e.g., IQC1 Total CV 4.1%, IQC2 Total CV 3.4%, IQC3 Total CV 3.2%).
    Analytical Specificity (Interference)No significant interference from common biological substances and exogenous compounds up to tested concentrations.No significant interference found for: - Triglycerides: up to 3000 mg/dL - Haemoglobin: up to 500 mg/dL - Bilirubin (conjugated/unconjugated): up to 40 mg/dL - Total Protein: up to 12 g/dL - Biotin: up to 6000 ng/mL (and 1500 ng/mL) - Rheumatoid Factor: up to 7000 IU/mL - HAMA: up to 3000 ng/mL - Cholesterol: up to 456 mg/dL - Various common drugs (e.g., Acetaminophen, Ibuprofen, Ascorbic acid, Creatinine, Dopamine, Tetracycline, Tolbutamide, Tolazamide, Uric Acid).
    Analytical Specificity (Cross-Reactivity)No significant cross-reactivity with structurally similar compounds or other biological substances up to tested concentrations.Low/No significant cross-reactivity observed (<1% typically) for: - AFP (-0.3% to 0.5%) - Thyroxin binding globulin (0.0% to -0.1%) - Transferrin (0.0%) - Cortisol (0.0%) - 11-deoxycortisol (0.0% to -0.1%) - 5a-dihydroxytestosterone (0.0%) - Testosterone (0.0%) - Fibrinogen (0.0%) - Corticosteroid binding globulin (0.0%) - Thyrotropin (TSH) (0.0%) - Plasminogen (0.0%) - Human IgA (0.0%) - Human IgG (0.0%) Notable exceptions: Estradiol (-7.3% to -3.2%) and Thyroglobulin (5.1% to 90.2%), where the higher value indicates specific interference for Thyroglobulin.
    Method ComparisonGood correlation and agreement with a predicate device.Correlation Coefficient (r): 0.989 (n=136 samples, range 2.54 - 172.12 nmol/L). Passing-Bablok Regression: Slope 0.9112, Intercept 0.1556 nmol/L.
    Matrix ComparisonPerformance across different sample matrices (serum, serum gel, K2 EDTA plasma) should be comparable.Good correlation: - Gel tube vs. Serum: r=0.999, Mean bias 0.7% - K2 EDTA vs. Serum: r=0.998, Mean bias -1.3% (n=69 samples, range 0.51 to 238.45 nmol/L)
    Expected Values/Reference RangeEstablished typical ranges for different populations.Males 21-49y: 11.47 – 58.07 nmol/L (n=165) Males >50y: 14.85 – 65.21 nmol/L (n=180) Premenopausal Females: 20.30 – 140.18 nmol/L (n=206) Postmenopausal Females: 11.30 – 127.31 nmol/L (n=120) (Based on 671 apparently healthy adults from the United States).

    2. Sample Size Used for the Test Set and Data Provenance

    • Precision/Reproducibility: 14 serum-based samples at different SHBG concentration levels covering the assay range. Tested across 21 days for one kit lot. (Table 11 shows individual sample IDs, CTLs, CALBs, IQCs)
    • Linearity/Assay Reportable Range:
      • Linearity: A high human serum sample and a low human serum sample, plus 14 evenly spaced dilutions created by mixing high and low samples.
      • Automated Post-dilution Accuracy: 9 native samples with known SHBG concentrations obtained from the predicate device.
    • Traceability of Calibrator:
      • Value assignment of Internal Reference Standards (IRs): At least 24 runs using three iSYS instruments, 2 kit lots, 3 replicates for each run. Serial dilutions of WHO 2nd international standard 08/266 used.
      • Value assignment of kit calibrators: At least 20 runs (14 runs using previous IR lot, 6 runs using international standard) using one iSYS instrument, 2 replicates for each run.
      • Value assignment verification: Internal quality controls at 3 SHBG levels tested in five replicates in one run and on each of three different iSYS instruments.
      • Correlation study (2-point calibration vs. IS 08/266 curve): 189 samples (139 serum, 50 K2 EDTA plasmas). Tested in two replicates using three lots (MB1, MB2, MB3) on one iSYS instrument.
    • Detection Limit (LoB, LoD, LoQ):
      • LoB: LoB sample run in 12 replicates for each of 5 runs over 3 days, by one operator, on one different instrument for each of 3 manufacture batches (MB1, MB2B, MB3) = total 60 replicates per lot.
      • LoD: 7 LoD samples measured in duplicate. For each of 3 kit lots, 5 assays over 3 days by one operator on a different instrument = total 70 replicates per lot.
      • LoQ: Panel of 9 samples measured in singlicate two times per day. For each of 3 kit lots, 10 assays over 5 days by one operator on a different instrument = total 90 replicates per lot.
    • Analytical Specificity (Interference): Two serum samples (low and high SHBG conc.) spiked with potential interferents. Control samples were also run. Number of replicates for each specific test not detailed but implied to be sufficient for statistical comparison.
    • Analytical Specificity (Cross-Reactivity): Low and high SHBG samples spiked with various cross-reactants. Number of replicates for each specific test not detailed.
    • Method Comparison: 136 samples, selected to represent a wide range of SHBG concentrations (2.54 - 172.12 nmol/L).
    • Matrix Comparison: 69 samples (68 native, 1 diluted) covering a range of 0.51 to 238.45 nmol/L.
    • Expected Values/Reference Range: 671 serum samples from apparently healthy adults (21-77 years old). Data Provenance: From the United States. Retrospective/Prospective: Not explicitly stated, but typically such studies for establishing reference ranges are retrospective collections of banked samples or a prospective study designed to collect samples from a healthy population.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Not Applicable (N/A).

    This is an in vitro diagnostic (IVD) assay quantifying a biomarker (SHBG). "Ground truth" for an IVD assay is established through:

    • Reference Methods: Such as the WHO 2nd international standard 08/266 for SHBG, against which the calibrators are standardized.
    • Known Concentrations: Use of accurately prepared standards, spiked samples, or samples extensively characterized by reference methods.
    • Clinical Data: For expected values, SHBG concentrations are measured in samples from defined healthy populations.

    Therefore, the concept of "experts establishing ground truth" in the way it applies to image interpretation or AI-assisted diagnostics (e.g., radiologists labeling images) is not relevant here. The ground truth is analytical and based on metrological traceability to international standards.


    4. Adjudication Method for the Test Set

    N/A.

    Adjudication methods (like 2+1, 3+1) are relevant for subjective interpretations (e.g., radiology reads) where discrepancies between readers need to be resolved to establish a definitive ground truth. For a quantitative IVD assay like IDS SHBG, the "reading" is a numerical output from the instrument based on chemical reactions. Accuracy is determined by comparison to reference materials or established methods, not by human adjudication of interpretations.


    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    N/A.

    This is an IVD assay, not an AI/ML diagnostic for human interpretation. No MRMC study or human reader improvement with AI assistance is applicable.


    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    N/A.

    This is not an AI/ML algorithm. The "device" is a fully automated immunoassay system (IDS-iSYS Multi-Discipline Automated System) that performs the biochemical analysis and reports a quantitative result without human "interpretation" of a signal beyond reading the numerical output. Its performance is evaluated as a standalone analytical system.


    7. The Type of Ground Truth Used

    The ground truth for the analytical performance studies (precision, linearity, detection limits, specificity, method comparison) is based on:

    • International Reference Standards: Specifically, the WHO 2nd international standard for SHBG (IS 08/266) for traceability and calibration. This is the primary reference.
    • Known Concentrations/Spiked Samples: Samples with defined, known concentrations of SHBG or interferents, prepared in a laboratory setting.
    • Comparison to a Legally Marketed Predicate Device: The Siemens ADVIA Centaur SHBG assay (K151986) was used as a comparative method to assess agreement and accuracy (e.g., for method comparison and automated dilution accuracy).
    • Healthy Population Data: For expected values/reference ranges, a large cohort of apparently healthy individuals were tested to establish population-specific normal ranges.

    8. The Sample Size for the Training Set

    N/A (for AI/ML 'training set' in the traditional sense).

    For an IVD assay, the equivalent of a "training set" would be the samples and calibrators used during the assay's development and optimization phases to set parameters, establish reagent formulations, and fine-tune the system. This information is typically proprietary development data and is not explicitly detailed as a 'training set' in 510(k) summaries, which focus on the final validation/test data.

    The closest analogous "training" or "calibration" process mentioned is the traceability and value assignment of calibrators:

    • Value assignment of secondary standards (IRs) involves at least 24 runs (using 3 instruments, 2 kit lots, 3 replicates) comparing to the WHO international standard.
    • Value assignment of IDS SHBG kit calibrators A and B involves at least 20 runs (1 instrument, 2 replicates) using the secondary standards and IS-08/266.
      These processes are used to establish the calibration curve for the assay.

    9. How the Ground Truth for the Training Set Was Established

    N/A (for AI/ML 'training set').

    As explained in point 8, the concept of a "training set" for an IVD assay is different from that for AI/ML. The "ground truth" for establishing the calibration and parameters of the assay is based on:

    • Metrological traceability to the WHO 2nd international standard for SHBG (IS 08/266). This involved serial dilutions of the international standard in SHBG-depleted human serum to create reference points.
    • Use of internal reference calibrators (IRs) that were themselves value-assigned against the WHO standard.

    The goal is that the assay's measurements accurately reflect the true concentration of SHBG as defined by an international reference.

    Ask a Question

    Ask a specific question about this device

    K Number
    K182521

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-01-29

    (138 days)

    Product Code
    Regulation Number
    862.1680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    FastPack® IP SHBG is a chemiluminescent immunoassay intended for the quantitative determination of Sex Hormone Binding Globulin in human serum and plasma on the FastPack® System. The FastPack® IP SHBG assay is intended for use as an aid in the diagnosis of androgen disorders.

    Device Description

    The FastPack® IP Sex Hormone Binding Globulin Immunoassay employs a sandwich immunoassay principle. Endogenous SHBG in a patient sample, calibrator, or control is dispensed into a FastPack® reagent pack. In the reagent pack, the sample binds with a monoclonal anti-SHBG antibody covalently linked to alkaline phosphatase (ALP) and a different monoclonal anti-SHBG antibody linked to biotin will bind to streptavidin coated paramagnetic particles (PMP). After incubation, washing steps (using a Tris buffer containing detergents) occur to separate bound from unbound anti-SHBG monoclonal antibody-ALP, a chemiluminogenic substrate mixture is added to the system. This mixture contains indoxyl-3-phosphate, a substrate for ALP, and lucigenin (N,N dimethyl-9,9'-biacridinium dinitrate). ALP dephosphorylates indoxyl-3-phosphate to indol-3-ol, which subsequently undergoes oxidation. As a result, lucigenin is reduced to form a dioxetane structure that is cleaved to yield N-methylacridone. This compound produces a sustained luminescent glow following excitation. The raw relative luminescence units (RLUs) generated are measured by a photomultiplier tube in the FastPack® Analyzer and are directly proportional to the concentration of SHBG in the sample. The entire reaction sequence takes place at 37 ± 0.5 ℃ and is protected from external light.

    AI/ML Overview

    Here's an analysis of the provided text regarding the FastPack® IP Sex Hormone Binding Globulin Immunoassay's acceptance criteria and studies:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state formal "acceptance criteria" for all performance characteristics in a separate section. However, based on the comparative effectiveness study against a predicate device and the presented performance data, we can infer some implied acceptance ranges or targets for the new device.

    Performance CharacteristicAcceptance Criteria (Implied)Reported Device Performance
    Precision (%CV)Similar to or better than predicate (≤ 5.5% for all listed). Stated target: Within-run: ≤ 10%, Between-run: ≤ 8%, Between-day: ≤ 8%Reagent lot 1, analyzer 1, calibrator lot 1: * Within-Run: 2.74% - 6.95% * Between-Run: 0.00% - 7.79% * Between-Day: 0.00% - 4.96% * Total: 3.21% - 7.06%Reagent lot 2, analyzer 2, calibrator lot 2: * Within-Run: 4.16% - 7.74% * Between-Run: 0.00% - 6.56% * Between-Day: 1.25% - 7.44% * Total: 5.04% - 11.53%Reagent lot 3, analyzer 3, calibrator lot 3: * Within-Run: 3.4% - 9.60% * Between-Run: 0.0% - 7.21% * Between-Day: 0.0% - 6.05% * Total: 4.0% - 12.05%
    LinearityAssay linear within a specified range (Predicate: 0.33 - 200 nmol/L)Linear from LOQ (0.80 nmol/L) to 174 nmol/L
    Interfering SubstancesNo interference at specified levels (similar to predicate)No interference with listed compounds (e.g., conjugated bilirubin 40 mg/dL, hemoglobin 1.0 g/dL, lipid 1000 mg/dL, d-Biotin 0.2 mg/dL) and cross-reactants (e.g., Transferrin 0.5 g/dL, Testosterone 2.5 mg/dL, etc.) at specified concentrations. Rheumatoid factor up to 1000 IU/mL and human anti-mouse IgG up to 4 µg/mL also showed no cross-reactivity. Six known heterophile samples did not generate detectable interference.
    Method ComparisonStrong correlation to predicate (Predicate R2 = 0.94)R = 0.985, Slope (95% CI): 0.993 (0.967-1.019), y-intercept (95% CI): -0.614 (-2.21 to 0.982), R2 = 0.971
    Sample Type EquivalenceEquivalence between serum and plasmaStrong correlation between serum and lithium-heparin plasma via Passing-Bablok regression: Slope (95% CI): 0.960 (0.920-1.00), y-intercept (95% CI): 1.859 (-0.89 to 4.61), R = 0.990, R2 = 0.979. Absolute bias 1.117 nmol/L, % Bias 1.928%.
    LOBNot explicitly stated for acceptance, but a calculated value0.08 nmol/L SHBG
    LODNot explicitly stated for acceptance, but a calculated value0.20 nmol/L SHBG
    LOQLowest sample with < 20% CV0.80 nmol/L SHBG

    2. Sample Size and Data Provenance for Test Set (Performance Studies)

    • Precision: 7 serum patient samples, tested in duplicate over 20 non-consecutive days, resulting in 240 replicate determinations per sample.
    • LOB/LOD: 180 replicate determinations of a blank sample (LOB) and 180 replicate determinations of four low-level samples (LOD).
    • Linearity: 11 concentration levels generated by intermixing high and low patient samples, each tested in quadruplicate.
    • Interferences: Two serum samples (low and high SHBG concentrations) spiked with known interfering concentrations.
    • Serum and Plasma Equivalence: Blood collections from 54 volunteers.
    • Expected Values/Reference Intervals: N=613 male (n=304) and female (n=309) apparently healthy individuals.
    • Method Comparison: 158 human serum samples.

    Data Provenance (Implied): The studies refer to "patient samples," "serum samples," and "human serum/plasma." While direct country of origin is not specified, it's generally understood that such studies for FDA submission are conducted under controlled laboratory conditions, likely within the US or by institutions adhering to similar regulatory standards. The data is retrospective in the sense that these are collected samples used for performance evaluation, not part of real-time clinical use for new diagnoses in a prospective clinical trial.

    3. Number of Experts and Qualifications for Ground Truth

    This device is an in vitro diagnostic (IVD) immunoassay that measures a biomarker (SHBG concentration). The concept of "ground truth" here is the true concentration of SHBG in a sample, or the accuracy of the measurement against a reference method.

    • No human experts (like radiologists reading images) are used to establish ground truth for individual measurements of SHBG concentration.
    • For Method Comparison: The "ground truth" or reference standard for comparison is the predicate device (Beckman Coulter Access Sex Hormone Binding Globulin assay K083867). This predicate device itself had previously established performance.
    • For Reference Intervals (Expected Values): "Apparently healthy individuals with no known pre-existing endocrine disorders" were used. The reference intervals (2.5th-97.5th percentiles) are derived statistically from a population, rather than established by individual expert diagnosis for each sample.

    4. Adjudication Method for the Test Set

    Not applicable. As described above, the "ground truth" for this IVD is either the measurement by a predicate device or statistically derived normal ranges, not expert interpretation requiring adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    Not applicable. This is an automated immunoassay for quantitative determination of a biomarker, not an imaging device or AI algorithm that assists human readers in interpreting cases. There is no "human-in-the-loop" component in the direct measurement process to be evaluated for improvement with AI assistance.

    6. Standalone (Algorithm Only) Performance

    Yes, the studies presented are all standalone performance evaluations of the FastPack® IP Sex Hormone Binding Globulin Immunoassay system (device + assay + analyzer). The device operates independently to provide quantitative results.

    7. Type of Ground Truth Used

    • For Method Comparison: The measurements from the predicate device (Beckman Coulter Access Sex Hormone Binding Globulin assay) served as the comparative "truth" or reference.
    • For Limit of Quantitation (LOQ): Defined by statistical criteria (lowest concentration with < 20% CV).
    • For Linearity: Based on spiked samples with known concentrations and statistical fit to a linear model.
    • For Interfering Substances/Cross-reactivity: Based on known spiked concentrations of potential interferents and lack of significant deviation from unspiked controls.
    • For Expected Values/Reference Intervals: Statistically derived from a large population of apparently healthy individuals.
    • Traceability: The assay is traceable to WHO 082/266 reference material, which serves as a primary reference for SHBG concentration.

    8. Sample Size for the Training Set

    Not applicable in the typical AI/ML sense. This is a conventional in vitro diagnostic device, not an AI/ML system that requires a "training set" to learn from data. The device's performance characteristics are established through analytical validation studies (precision, linearity, interference, method comparison, etc.) as described. The calibration curve for the assay is established using known calibrator materials, but this is part of the daily operational setup, not a "training set" in the context of AI.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable for the same reasons as #8. The assay relies on known concentrations of calibrators (traceable to WHO reference material) to establish its quantitative scale, not on a "ground truth" established for an AI training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K181017

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-07-20

    (101 days)

    Product Code
    Regulation Number
    862.1680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Free Testosterone AccuBind® ELISA Test System is an Enzyme Immunoassay (EIA) for the quantitative measurement of free testosterone in human serum. Measurement of free testosterone is used in the diagnosis and treatment of disorders involving the male sex hormones (androgens), including primary and secondary hypogonadism, impotence in males and in females; hirsutism (excessive hair) and virilization (masculinization) due to tumors, polycystic ovaries and androgenital syndromes.

    Device Description

    The kit consists of seven (7) vials of serum reference calibrators for Free Testosterone with two (2) controls (one low and one high); one (1) vial of Testosterone (Analog)-horseradish peroxidase (HRP) conjugate in a protein stabilizing matrix; one 96-well testosterone antibody-coated microplate; one (1) vial of concentrated wash solution; two (2) vials for tetramethy(benzidine (TMB) substrate solution preparation; and one (1) vial of stop reaction solution.

    AI/ML Overview

    The provided document describes the analytical performance of the Monobind Inc. Free Testosterone AccuBind® ELISA Test System, an in vitro diagnostic device, rather than an AI/ML-driven device. Therefore, many of the requested criteria often associated with AI/ML device studies (e.g., number of experts for ground truth, MRMC studies, human-in-the-loop performance, training set details) are not applicable to this type of medical device.

    However, I can extract the relevant information regarding the device's acceptance criteria and the study that proves it meets them as presented in the document.

    Here's the breakdown:

    1. Table of Acceptance Criteria and Reported Device Performance

    For this type of in-vitro diagnostic device, "acceptance criteria" are typically defined by regulatory standards and good laboratory practices (e.g., CLSI guidelines). The document presents analytical performance data against these established industry benchmarks rather than explicit numerical acceptance criteria beyond what is internally defined for successful assay development and validation (e.g., precision specifications).

    Performance MetricAcceptance Criteria (Implicit from Industry Standards/Good Practice)Reported Device Performance (Representative Lot / Combined Lot)
    PrecisionLow %CV for within-run and total precision across concentrations.Within-Run CV: 3.6-7.9% (Representative Lot), 4.23-5.95% (Combined Lot)
    Total CV: 7.8-12.4% (Representative Lot), 7.20-9.43% (Combined Lot)
    Linearity (Accuracy)High correlation coefficient (R) and slope close to 1.y = 1.0149x - 0.6028, R = 0.9888 (Excellent linearity)
    RecoveryPercent recovery close to 100% for spiked samples.98.7% - 105.2% (Across 5 patient samples)
    Reagent/Kit StabilityDemonstrates stability for stated shelf life and open-vial stability.2 years (unopened kit), 61 days (open kit, calibrators, controls, enzyme reagent, TMB, plate)
    Sample StabilityDemonstrates stability for specified storage conditions.61 days (serum at 2-8°C), 31 days (frozen serum at -20°C)
    Detection LimitsDefined LOB, LOD, LOQ based on CLSI EP17-A.LoB: 0.0295 pg/ml; LoD: 0.0519 pg/ml; LoQ: 0.0519 pg/ml
    Cross-ReactivityGenerally low cross-reactivity (<10% difference) with specified interfering substances.Generally 0.000% for most compounds; highest reported up to 0.647% (for 11-KetoTestosterone). No significant cross-reactivity (>10% difference) observed.
    Method ComparisonStrong correlation with predicate device (high R, slope near 1, low intercept).y = 1.017x - 0.244, Correlation Coefficient = 0.997 (Excellent agreement)
    InterferencesNo significant interference at specified concentrations for various substances.Most substances showed no significant interference at the tested concentrations.

    2. Sample Size Used for the Test Set and Data Provenance

    • Precision Study:
      • Sample Size: For the representative lot, 3 control samples were tested, with 80 measurements per sample (duplicate, two times per day for 20 days). 3 serum pools were also tested in the same manner (N=32 readings for the representative lot table, N=80 for combined lot table). For the combined lot precision, 3 controls and 3 patient samples (serum pools) were used, with a total of 80 measurements per sample.
      • Data Provenance: Not explicitly stated, but typical for in vitro diagnostic device validation, these would be controlled laboratory studies using clinical samples (serum pools and control materials). No mention of country of origin or retrospective/prospective clinical data for the performance evaluation (this differs from the reference range determination).
    • Linearity Study: 11 concentrations of sample preparations.
    • Recovery Study: 5 serum samples (containing different levels of endogenous testosterone).
    • Reference Range Determination: 261 male and female serum samples.
      • Data Provenance: Not specified for the 261 samples, but typically these samples are collected under ethical guidelines from a relevant population.
    • Cross-Reactivity Study: Specific compounds were tested with male serum spiked samples and blank serum spiked samples. No specific number of replicates per compound is given, but "aliquots from pool of human serum" are mentioned for testosterone cypionate and undecanoate.
    • Method Comparison Study: 137 samples.
      • Data Provenance: Not explicitly stated, but these would be clinical samples with varying testosterone levels.
    • Interferences Study: Charcoal-stripped human serum spiked with known concentrations of interferent.

    3. Number of Experts Used to Establish the Ground Truth and Qualifications of Experts

    • Not Applicable. For an in-vitro diagnostic assay for quantitative measurement of an analyte like Free Testosterone, the "ground truth" is established by the analytical method itself (the assay's ability to accurately measure the target analyte) and validated against reference methods or calibrated materials. There are no human "experts" establishing image-based ground truth as would be the case for AI/ML diagnostic tools. The predicate device serves as a comparative "ground truth" for method comparison.

    4. Adjudication Method for the Test Set

    • Not Applicable. This is an in-vitro diagnostic assay, not an AI/ML system requiring human adjudication of results. The results are quantitative measurements.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No. This is an in-vitro diagnostic assay, not an imaging AI/ML device that assists human readers. Therefore, an MRMC study is not relevant or performed.

    6. Standalone (Algorithm Only) Performance

    • Yes (inherently). The device itself is the "standalone" measurement system. Its performance (precision, linearity, recovery, sensitivity, specificity, interference) is evaluated on its own. There isn't a separate "human-in-the-loop" component as would be found in an AI-assisted diagnostic workflow.

    7. The Type of Ground Truth Used

    The ground truth for this device's performance validation is established through:
    * Known Concentrations: For precision (control materials with known values), linearity (prepared concentrations), recovery (spiked samples with known additions), detection limits (analyzing blanks and low-concentration samples).
    * Reference Methods/Predicate Device: For method comparison, the predicate device (EiAsy Free Testosterone EIA) serves as the comparator.
    * Established Analytical Principles: The fundamental biochemical reactions and measurement principles of the ELISA platform.
    * CLSI Guidelines: Adherence to established Clinical and Laboratory Standards Institute (CLSI) guidelines (e.g., EP06-A, C28-A3, EP17-A, EP07-A2) for validation studies provides the framework for defining acceptable performance.

    8. The Sample Size for the Training Set

    • Not Applicable. This is an in-vitro diagnostic assay, not an AI/ML device that requires a distinct "training set" in the context of machine learning model development. The development and optimization of the assay's reagents and protocol are analogous to "training" in a general sense, but no specific numerical sample size is defined as a training set for an algorithm.

    9. How the Ground Truth for the Training Set Was Established

    • Not Applicable. As above, there is no AI/ML "training set." The "ground truth" for developing such an assay comes from fundamental biochemical understanding, chemical synthesis of reagents, and iterative optimization of assay conditions to achieve desired analytical performance characteristics. This involves standard laboratory development practices common to IVD manufacturing.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 5