Search Filters

Search Results

Found 77 results

510(k) Data Aggregation

    K Number
    K251631

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-11-24

    (180 days)

    Product Code
    Regulation Number
    868.5915
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BAG is indicated for patients requiring total or intermittent ventilatory support. Ventilation is possible with or without supplemental oxygen. The BAG is intended to be used in both pre-hospital and healthcare facility environments.

    The sizes versions also include indications for the patient population:

    Adult: The Adult model is intended for patients over 20 kg (44 lbs).
    Pediatric: The Pediatric model is intended for patients 2.5 - 20 kg (5.5 - 44 lbs).
    Newborn: The Newborn model is intended for patients 2.5 - 5 kg (5.5 -11 lbs).

    The BAG PEEP Valve is indicated for patients in need of ventilatory support with positive end-expiratory pressure (PEEP). The BAG PEEP valves can be used for all patient sizes indicated for The BAG manual resuscitator. The BAG PEEP valve is intended to be used in both pre-hospital and healthcare facility environments.

    The BAG Manometer is indicated for measuring the gas pressure provided to a patient requiring total or intermittent ventilatory support. The BAG Manometer can be used for all patient sizes indicated for The BAG manual resuscitator. The BAG Manometer is intended to be used in both pre-hospital and healthcare facility environments.

    Device Description

    The BAG is a non-sterile self-inflating manual resuscitator. It is indicated for patients requiring ventilatory support. Ventilation is possible with or without supplemental oxygen. The BAG manual resuscitator is intended for short-term use (< 24 hours per patient) and may be used multiple times on a single patient when kept free from contamination. It is available in three sizes: adult (for patients > 20 kg (44 lbs.)), pediatric (for patients 2.5-20 kg (5.5-44 lbs.)), and newborn (for patients 2.5-5 kg (5.5-11 lbs.)).

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K243861

    Validate with FDA (Live)

    Device Name
    butterflyBVM
    Date Cleared
    2025-04-25

    (130 days)

    Product Code
    Regulation Number
    868.5915
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The butterflyBVM™ when used in transport and non-clinical emergency settings (e.g., EMS, non-hospital) is a single-use resuscitator that may be manipulated to provide pulmonary resuscitation of patients including adults, adolescents, children, infants, and neonates.

    The butterflyBVM™ when used in professional healthcare facilities (e.g., hospitals) is a single-use resuscitator that may be manipulated to provide pulmonary resuscitation of patients including adults, adolescents, children, and infants.

    Device Description

    The butterflyBVM™ is a three-in-one bag-valve-mask capable of resuscitating adults, adolescents, children, infants, and neonates. The maximum tidal volumes (Vt) delivered by the butterflyBVM™ can be set via the tidal volume dial, as needed, to ranges that are generally appropriate for the size of patient receiving care. The butterflyBVM™ also has an adjustable peak inspiratory pressure (PIP) dial to help prevent barotrauma. The tidal volume and PIP are selected by aligning the dials to the Vt/PIP indicator. Patient-facing accessories such as masks, laryngeal mask airways, endotracheal tubes, end-tidal CO2 samplers, and the like can be connected to the patient connection port. Exhalation accessories such as bio filters and PEEP valves can be connected to the exhalation port. Supplemental oxygen can be added via the oxygen inlet port. The device is operated by manually squeezing the side arms together with either one or two hands. A cover is provided over the PIP dial to provide instruction to the user via images and words to first set the tidal volume for the procedure based on the assessment of the patient.

    AI/ML Overview

    The provided FDA 510(k) clearance letter and summary for the butterflyBVM device offers details about its acceptance criteria and the study conducted to demonstrate performance. Here's a breakdown of the requested information:

    Acceptance Criteria and Device Performance

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriterionReported Device Performance (butterflyBVM)Relevant Standard / Context
    Peak Inspiratory Pressure (PIP) Control (Target ≤20cmH2O)1 failure with 20.3cmH2O max PIP deliveredComparative study against Ambu Spur II
    Peak Inspiratory Pressure (PIP) Control (Target ≤30cmH2O)1 failure with 30.4cmH2O max PIP deliveredComparative study against Ambu Spur II
    Default Max PIP Delivery0 failures with 40cmH2O max PIP deliveredDefault setting of the device
    Delivered Volume One Hand (Minimum)30 mlN/A (listed as comparable to predicate)
    Delivered Volume One Hand (Maximum)565 mlN/A (listed as comparable to predicate)
    Delivered Volume Two Hands (Minimum)70 mlN/A (listed as comparable to predicate)
    Delivered Volume Two Hands (Maximum)715 mlN/A (listed as comparable to predicate)
    Delivered Oxygen Concentration (≥85% FiO2 at 15 LPM)Demonstrated nominal oxygen concentrations through actual testing per EN ISO 10651-4 and met consensus standard.EN ISO 10651-4
    Patient valve function after contamination with vomitusPassedISO 10651-4
    Mechanical strength / Drop testPassedISO 10651-4
    Immersion in waterPassedISO 10651-4
    Minimum number of cyclesPassedISO 10651-4
    Expiratory resistancePassedISO 10651-4
    Inspiratory resistancePassedISO 10651-4
    Patient valve malfunctionPassedISO 10651-4
    Resuscitator dead-space and rebreathingPassedISO 10651-4
    Biocompatibility (Cytotoxicity, Irritation, Sensitization)PassedISO 10993-5, 10993-10, 10993-11, 10993-23
    Biocompatibility (Breathing Gas Pathways - Particulate Matter, Volatile Organic Compounds, Leachables)PassedISO 18562-1, 18562-2, 18562-3, 18562-4
    Shelf Life6 monthsN/A (Compared to predicate, difference does not raise safety/effectiveness questions)
    Operating Temperature-18°C to +50°CN/A (Same as predicate)
    Storage Conditions-40°C to +60°CN/A (Same as predicate)

    Details of the Study Proving Acceptance Criteria

    The primary study mentioned directly addressing quantitative performance criteria is a "Comparative Testing" study titled "In a study of the butterflyBVM versus the Ambu Spur II [Merrell et al. 2023]".

    2. Sample Size for the Test Set and Data Provenance

    • Sample Size: The document does not explicitly state a sample size for the comparative testing portion of the PIP Dial study. It mentions failures (e.g., "1 failure", "23 failures", "32 failures"), which implies multiple tests were conducted, but the total number of tests or devices tested is not provided.
    • Data Provenance: The study is referenced as [Merrell et al. 2023]. Given this is a 510(k) submission, it is assumed to be an internal, pre-market engineering or bench-top study conducted by Compact Medical, Inc., or a contracted firm. The country of origin and whether it was retrospective or prospective is not explicitly stated, but such a study would typically be prospective bench testing.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    • Not applicable. The study described is a bench testing/engineering study, not a clinical or human perception study requiring expert ground truth setting. The ground truth (target pressures, delivered volumes, oxygen concentrations) is established by measurement devices and consensus standards (ISO 10651-4).

    4. Adjudication Method for the Test Set

    • Not applicable. This was a bench testing study comparing device measurements to target values and predicate device performance. No human adjudication was involved.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    • No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The study described is a bench testing comparison of device performance parameters. It does not involve human readers (clinicians) interacting with cases or assessing improvements with AI assistance.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • This is not an AI/Software as a Medical Device (SaMD). The butterflyBVM is a manual emergency ventilator. Therefore, the concept of standalone algorithm performance (without human-in-the-loop) is not applicable. The device's performance is inherently tied to a human operator, but the presented data focuses on the device's mechanical and functional capabilities.

    7. The Type of Ground Truth Used

    • The ground truth for the performance parameters (PIP, delivered volume, oxygen concentration) was established through measurement against recognized consensus standards (primarily ISO 10651-4) using calibrated laboratory equipment. For comparative testing, the performance of the predicate device (Ambu Spur II) also served as a comparative benchmark to demonstrate substantial equivalence.

    8. The Sample Size for the Training Set

    • Not applicable. This is a mechanical medical device, not an AI/machine learning product. Therefore, there is no "training set" in the context of AI model development.

    9. How the Ground Truth for the Training Set was Established

    • Not applicable, as explained in point 8.
    Ask a Question

    Ask a specific question about this device

    K Number
    K221841

    Validate with FDA (Live)

    Device Name
    EOlife®
    Manufacturer
    Date Cleared
    2023-03-18

    (267 days)

    Product Code
    Regulation Number
    868.5915
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    EOlife is intended for use with emergency manual resuscitation devices to measure ventilatory flows and display visual guide on the insufflated volume, and ventilation frequency to ensure adequate ventilation of adult cardiopulmonary arrest patient during cardiopulmonary resuscitation (CPR) performed by healthcare professionals.

    Device Description

    EOlife is a medical device dedicated to healthcare professionals to help them providing manual ventilation during cardiopulmonary resuscitation (CPR). EOlife is intended to be connected to any standard manual resuscitator for adults including a bag and a mask or endotracheal tube (ET tube) and supraglottic airway (SGA) device and to be used during manual ventilation of an adult cardiopulmonary arrest patient. EOlife is a portable device composed of an electronic control unit including an embedded software, a removable and rechargeable battery pack and of a single use flow sensor: FlowSense. EOlife does not present direct contact with the patient. Only the ventilation air flow is in contact with the internal part of FlowSense (indirect contact with the patient). During manual ventilation, EOlife measures the ventilation parameters (insufflated volumes, tidal volumes, ventilation frequencies, ) and gives real-time feedback to the user about the quality of the ventilation provided to the patient compared to the 2020 AHA (American Heart Association) Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care, Part 3: Adult Basic and Advanced Life Support. The aim is to help the user ventilate according to the recommended frequencies (10 cycles per minute) and volumes (6-8 ml/kg of ideal body weight).

    AI/ML Overview

    The provided text describes the EOlife device, a medical device intended to assist healthcare professionals in manual ventilation during cardiopulmonary resuscitation (CPR) by measuring ventilatory flows and displaying visual guides on insufflated volume and ventilation frequency. The document is a 510(k) premarket notification by Archeon Medical to the FDA, asserting substantial equivalence to a predicate device.

    Here's an analysis of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document includes a "Comparison of Technological Characteristics with Predicate Device" (Table 1) and a "Summary of Performance Testing" (Table 2). The most relevant section for acceptance criteria related to product performance is the "Measuring range and accuracy of ventilation values" and "Measuring range and accuracy of ventilation frequencies" within Table 1.

    Acceptance Criteria (from Subject Device)Reported Device Performance and Discussion (from Subject Device, Predicate Device, and Discussion columns in Table 1)
    Insufflated Volume (Vi): Operating range 50 to 2000 mlVi: Operating range 50 to 2000 ml RMSE: 15.7 ml
    Tidal Volume (Vt): Operating range 0 to 2000 mlVt: Operating range 0 to 2000 ml RMSE: 40 ml
    Ventilation Frequency: 1 to 60 bpm, ± 1 bpmVentilation Frequency: 1 to 60 bpm, ± 1 bpm
    Real time emergency feedback: Based on 2020 AHA GuidelinesComparison to Predicate: Both devices (EOlife and L770-CPR Resuscitation Timer) provide real-time feedback for manual ventilation quality during CPR. The predicate device uses 2005 AHA Guidelines, while EOlife uses 2020 AHA Guidelines. The discussion states this is "Same" because "Both devices were based on the current relevant AHA guideline in force." This implies that adhering to the most current guidelines is an acceptable performance characteristic.
    Mechanical design and compatibility (with environment and resuscitator)Bench Testing Summary: Verification of the mechanical design and compatibility of the device with its environment and with the manual resuscitator was performed. Comparison between features (mechanical design, assembly with patient interface and manual resuscitator, ventilation setting) and predicate device was done to demonstrate substantial equivalence.
    Software functionality, safety, and effectivenessSoftware Testing Summary: IEC 62304 testing met requirement of the standard. Software testing included design, development, verification, validation, and traceability. Verification and validation demonstrated that the software functions correctly as designed and operates the device according to device design specifications and requirements.
    Usability/Human FactorsUsability Assessment Summary: Usability assessment was completed including critical tasks identification through use-based risk analysis, rounds of formative and summative testing according to IEC 62366-1 and FDA guidance. The study demonstrated that users were able to successfully and safely use the device.

    2. Sample Size Used for the Test Set and Data Provenance:

    The document explicitly states: "The subject EOlife did not require clinical studies to support substantial equivalence." Instead, it relies on non-clinical (bench) testing, usability testing, and comparison to a predicate device.

    • Test Set Description: The "test set" in this context refers primarily to the testing conducted on the physical device and its software, rather than a dataset of patient cases.
    • Sample Size:
      • For the performance metrics (Vi, Vt, Freq), the specific number of tests or samples used to derive the RMSE values is not provided.
      • For biocompatibility, electrical safety, EMC, battery, power supply, mechanical, and software testing, the sample size is also not explicitly stated. These types of tests typically involve a limited number of devices to demonstrate compliance with standards.
      • For usability testing, "rounds of formative and summative testing" were performed, but the number of participants or test scenarios is not specified.
    • Data Provenance: The testing was related to the device itself (hardware, software). No patient data is mentioned as being used for testing. The manufacturer, Archeon Medical, is based in Besançon, France. Therefore, the testing would presumably have been conducted there or by affiliated labs.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    This information is not provided as the submission focuses on non-clinical testing and comparison to a predicate device, not on diagnostic performance against human expert-established ground truth on a patient dataset. The "ground truth" for the device's measurements (volume, frequency) would be established by calibrated measurement equipment used in the bench testing.

    4. Adjudication Method for the Test Set:

    This information is not applicable as there is no human expert-based ground truth or adjudication process described for clinical effectiveness. The assessment is against technical specifications and established standards through bench testing.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance:

    A multi-reader multi-case (MRMC) comparative effectiveness study was not done. The submission explicitly states: "The subject EOlife did not require clinical studies to support substantial equivalence." The device is positioned as an "adjunct tool" providing real-time feedback, not as an AI diagnostic tool that human readers would interpret. Therefore, there is no discussion of human reader improvement with or without AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    The performance testing (bench testing) described in Table 2, specifically the "Accuracy assessment of the values measured and displayed by EOlife," can be considered a form of standalone performance assessment of the device's measurement capabilities. The software within the device operates autonomously to measure and display parameters ("algorithm only"). However, this is for measurement accuracy, not for a diagnostic or interpretive task that would typically be associated with "standalone AI performance" in imaging or other clinical decision support contexts. The device's primary function is to provide real-time guidance to a human user, making it inherently "human-in-the-loop" in its intended use.

    7. The Type of Ground Truth Used:

    • For the measurement accuracy of volume and frequency values during bench testing, the ground truth would be established by precision laboratory instruments and simulated ventilation profiles (e.g., using a test lung and a ventilator simulator with known, calibrated outputs).
    • For biocompatibility, electrical safety, EMC, mechanical, and software validation, the ground truth is compliance with international consensus standards (e.g., ISO, IEC, UN).
    • For usability, the ground truth is the successful and safe completion of critical tasks by users as assessed in human factors testing.

    8. The Sample Size for the Training Set:

    This information is not applicable/not provided. The EOlife device is described as having "an embedded software that contains algorithms," but the document does not indicate that these algorithms (for measuring ventilatory flows and providing guidance) were developed using machine learning or required a "training set" of data in the typical sense. It appears to be based on programmed logic and physics models rather than statistical learning from a large dataset.

    9. How the Ground Truth for the Training Set Was Established:

    This information is not applicable/not provided for the same reasons as point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K212905

    Validate with FDA (Live)

    Device Name
    Sotair Device
    Date Cleared
    2022-08-18

    (339 days)

    Product Code
    Regulation Number
    868.5915
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Sotair™ device is intended for use with a manual resuscitator with air-tight connections, in non-breathing patients that require flow controlled ventilation using ambient air or a supplemental oxygen source. The Sotair device is a single use, disposable device added to a manual resuscitator and can be used for in-hospital, emergency, and transport care. The adult Sotair device comprises a flow-limiting valve that limits the inspiratory flow enabling providers to ventilate approximately 55 liters-per-minute (LPM). The flow-limiting valve is intended to minimize gastric inflation during manual ventilation. The Sotair device can be disabled by removing the manual resuscitator to its conventional operation. The Sotair device is intended for adult use only.

    Device Description

    The Sotair™ device is a universal safety accessory device that is compatible with all manual resuscitators and airway interfaces that have standard ISO 5356-1:2015 15/22 mm connections. The Sotair device's flow-limiting function has been tested with the following bag resuscitators: [List of compatible resuscitators provided in the document]. The Sotair device is not compatible with manual resuscitators that have a built-in flow limiting device. The Sotair device uses a flapper valve with a blade design to limit the flow rate.

    AI/ML Overview

    The provided text describes the Sotair™ Device, a flow-limiting valve for manual resuscitators, and its equivalence to a predicate device. However, the text does not include a table of acceptance criteria and reported device performance in the specific format requested. It also does not contain information on sample sizes, data provenance, ground truth establishment methods, or the involvement of experts in establishing ground truth for a test set. There's no mention of a multi-reader multi-case (MRMC) comparative effectiveness study or a standalone algorithm performance study, as this is a medical device, not an AI/software product in the traditional sense described by those elements of the prompt.

    Based on the available information, here's a breakdown of what can be extracted:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document describes various performance tests conducted. While it states that the testing "met the requirements" or "demonstrated that the Sotair device met the product specifications and also performed equivalent to the predicate device," it does not provide specific acceptance values or the quantitative results from these tests in a table format.

    Test CategoryAcceptance Criteria (Implied)Reported Device Performance
    BiocompatibilityMet requirements of ISO 10993 and ISO 18562"Overall, the testing met the requirements of ISO 10993 and ISO 18562."
    Mechanical TestingMet product specifications and performed equivalent to predicate"The performance data demonstrated that the Sotair device met the product specifications and also performed equivalent to the predicate device in the Valve Functionality Testing..."
    Human Factors / UsabilityMet criteria for use with intended users, uses, and environments"The study concluded that the Sotair device met the criteria for use with the intended users, uses, and use environments. The overall result... confirm that the Sotair device is designed, functions and meets the expectations and need for use."

    2. Sample size used for the test set and the data provenance:

    • Test Set Sample Size: Not explicitly stated for any of the tests.
    • Data Provenance: The studies were conducted by SafeBVM Corporation (submitter) and its contracted consultants/labs. The country of origin for the data is not specified beyond the submitter's address in Chesterfield, MO, USA. The studies are described as "performance testing" and "simulated-use Human Factors Engineering / Usability Engineering (HFE/UE) Summative / Validation Study," implying prospective testing rather than retrospective data analysis.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not provided. The "ground truth" for the device's performance appears to be established through adherence to engineering standards (ISO) and functional testing against design specifications and a predicate device, rather than expert consensus on a diagnostic output.

    4. Adjudication method for the test set:

    Not applicable or not described in the document, as it pertains to a physical medical device's performance against predefined criteria and standards.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. The Sotair™ Device is a mechanical medical device (flow-limiting valve), not an AI-powered diagnostic or assistive tool for human readers. Therefore, an MRMC study related to AI assistance is not relevant to this submission.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Not applicable. The device is a mechanical component of a manual resuscitator, not an algorithm. Its performance is inherent to its physical design and function.

    7. The type of ground truth used:

    The ground truth used for evaluating the Sotair™ device's performance is based on:

    • Biocompatibility Standards: Adherence to ISO 10993 and ISO 18562.
    • Mechanical and Functional Specifications: Meeting predetermined product specifications for various mechanical properties and functionality (e.g., flow rate, leakage, resistance, valve function).
    • Equivalence to Predicate Device: Direct comparison testing to demonstrate equivalent performance to the legally marketed predicate device (Smart Bag MO) in Valve Functionality Testing.
    • Human Factors/Usability Criteria: Meeting usability criteria for intended users, uses, and use environments as per FDA guidance.

    8. The sample size for the training set:

    Not applicable. The device is a mechanical product, not an AI model requiring a training set. The descriptions relate to performance testing and usability studies.

    9. How the ground truth for the training set was established:

    Not applicable. There is no training set for this type of medical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K210288

    Validate with FDA (Live)

    Date Cleared
    2021-08-26

    (205 days)

    Product Code
    Regulation Number
    868.5915
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Single patient use manual resuscitator for use in hospital, transport, emergency, and post-hospital care to temporary ventilate a patient for the given body mass ranges of: Infant: less than or equal to 10 kg Child: less than or equal to 23 kg Adult: greater than 23 kg

    Device Description

    Disposable manual resuscitator is disposable, medical device, which temporarily augment ventilation in patients during ventilatory insufficiency or ventilator failure. Disposable manual resuscitator uses a duck-bill valve in the non-rebreathing valve assembly, attaches the non-rebreathing valve directly onto the resuscitation bag and includes an oxygen enrichment (reservoir) system. Disposable manual resuscitator may be used in hospital, transport, emergency, and post hospital care to temporary ventilate a patient. Disposable manual resuscitator is for single patient used. It comprises of mask, oxygen tube, reservoir bag and resuscitator bag. It is used to temporary ventilate a patient for the given body mass ranges of: Infant - Less than or equal to 10 kg Child - less than or equal to 23 kg Adult - Greater than 23 kg.

    AI/ML Overview

    The provided text is a 510(k) premarket notification for a medical device (Disposable Manual Resuscitator). This document outlines the manufacturer's claim of "substantial equivalence" of their device to a legally marketed predicate device.

    Crucially, this document is NOT a study report for an AI/ML powered medical device, nor does it contain information about establishing ground truth, multi-reader multi-case studies, or training sets in the context of AI/ML.

    The "Performance Data" and "Summary of Device Testing" sections refer to the physical and functional performance of the manual resuscitator as per established medical device standards (e.g., ISO 10651-4, ISO 5356-1, ISO 10993). These are tests for mechanical and material properties, not an AI algorithm.

    Therefore, I cannot extract the information required for bullet points 1 through 9 from the provided text, as those points are specific to the validation of AI/ML systems.

    Here's a breakdown of why the requested information cannot be found in this document:

    • Acceptance Criteria for AI/ML and Reported Device Performance (Table 1): The document's "Performance Data" section in Table 1 refers to physical characteristics of a manual resuscitator (e.g., "Ventilation Bag Volume," "Max Delivered Volume," "Expiratory resistance," "Supplemental Oxygen%"). These are not acceptance criteria or performance metrics for an AI/ML algorithm.
    • Sample Size for Test Set and Data Provenance: This pertains to data used to evaluate AI. The document describes testing of physical products, not data sets.
    • Number of Experts and Qualifications for Ground Truth: No AI is involved, so no ground truth established by experts is mentioned.
    • Adjudication Method: Not applicable for a non-AI physical device.
    • Multi Reader Multi Case (MRMC) Comparative Effectiveness Study: This is for evaluating changes in human performance with AI assistance. Not relevant here.
    • Standalone (Algorithm Only) Performance: Not applicable as it's not an algorithm.
    • Type of Ground Truth Used: Not applicable.
    • Sample Size for Training Set & How Ground Truth for Training Set was Established: These concepts are central to AI/ML model development. The document describes product manufacturing and testing processes, not AI training.

    In summary, the provided document details the regulatory submission for a physical medical device (a manual resuscitator) and demonstrates its substantial equivalence to a predicate device based on physical and functional testing. It does not involve any artificial intelligence or machine learning components, and thus the requested details regarding AI/ML acceptance criteria and validation studies are not present.

    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Flexicare's Adult Single Use Resuscitator Bag is intended for manual pulmonary resuscitation and emergency respiratory support of adult patients with a body weight of more than 66lbs (30kg). For use with ambient air and supplemental oxygen if required. For use by CPR-trained personnel only within hospital environments.

    Flexicare's Pediatric Single Use Resuscitator Bag is intended for manual pulmonary resuscitation and emergency respiratory support of infants and children with a body weight of 22lbs to 66lbs (10-30kg). For use with ambient air and supplemental oxygen if required. For use by CPR-trained personnel only within hospital environments.

    Flexicare's Infant Single Use Resuscitator Bag is intended for manual pulmonary resuscitation and emergency respiratory support of neonates and infants with a body weight of up to 22lbs (10kg). For use with ambient air and supplemental oxygen if required. For use by CPR-trained personnel only within hospital environments.

    Device Description

    Flexicare's Single Use Resuscitator Bags are Single Use devices For manual ventilation of a patient by trained operators in emergency/critical situations where short term ventilation is demanded by the patient's medical condition (e.g. inadequate or no breathing). Manual squeezing of the resuscitator bag forces air or air /O2 mixture into the patient's lungs via a face mask or pre-positioned airway tube with a 15mm male connection. A reservoir bag is present at the distal end of the device which fills with supplementary O2 (if using). This can be squeezed into the resuscitator for administration if required.

    Flexicare's Single Use Resuscitator Bags feature a one-way duck bill valve at the patient end to deliver fresh gas to the patient whilst preventing ingress of exhaled air and potential rebreathing. A one-way valve is also present at the distal end of the device, ensuring that upon squeezing the Resuscitator bag air only travels toward the patient, and upon release of a squeezed bag fresh ambient air with/without supplementary O2 is drawn in. Flexicare's Single Use Resuscitator Bags incorporate a pressure relief valve. This valve limits the pressure within the resuscitator and patient lungs to 60cmHzO (Adult) or 40cmHzO (Adult, Pediatric, Infant). One variant of Flexicare's Adult Single Use Resuscitator Bags does not feature a pressure relief valve.

    Flexicare's Single Use Resuscitator Bags are comprised of disposable components including a compression bag, valves, tubing and connectors. The Single Use Resuscitator Bags are intended for Adult. Pediatric and Infant patients, are supplied non sterile and are for use by CPR-trained personnel only within a hospital and/ pre-hospital environments.

    AI/ML Overview

    The provided text describes the acceptance criteria and a study demonstrating the Flexicare Single Use Resuscitator Bags meet these criteria by showing substantial equivalence to predicate devices.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally based on meeting recognized standards and performance characteristics comparable to the predicate devices. The "Outcome" column in the table below indicates that the Flexicare device "Passed" each test, implying that it met the respective criteria.

    Test / Characteristic comparedAcceptance Criteria (or Standard Met)Reported Device Performance (Outcome)
    Visual inspectionComparison testSubstantially equivalent
    Valve Function after contamination with vomitusBS EN ISO 10651-4:2009Pass
    Mechanical shock - Drop testing- (Implied to meet performance of predicate / safety standards)Pass
    Immersion in Water- (Implied to meet performance of predicate / safety standards)Pass
    Supplementary Oxygen & Delivered Oxygen Concentration- (Implied to meet performance of predicate / safety standards)Pass
    Expiratory Resistance- (Implied to meet performance of predicate / safety standards. Specific values provided in comparison tables indicate a required resistance range)Pass
    Inspiratory Resistance- (Implied to meet performance of predicate / safety standards. Specific values provided in comparison tables indicate a required resistance range)Pass
    Patient Valve Malfunction- (Implied to meet performance of predicate / safety standards)Pass
    Tidal Volume - Minimum Delivered Volume- (Implied to meet performance of predicate / safety standards. Specific values provided in comparison tables indicate a required volume range)Pass
    Pressure Limitation - Pressure Relief Blow off- (Implied to meet performance of predicate / safety standards. Specific values provided in comparison tables indicate a required pressure)Pass
    Conical Connector complianceISO 5356-1:2004Pass
    Leak testing- (Implied to meet performance of predicate / safety standards)Pass
    Drop testing- (Implied to meet performance of predicate / safety standards)Pass
    Cytotoxicity, Irritation, Sensitization, Systemic Toxicity, Extractables & LeachablesISO 10993-10:2010, ISO 10993-5:2009, ISO 10993-11:2009, ISO 10993-17:2009Pass
    Tubing resistance to gas flowBS EN 13544-2:2002+A1:2009Pass
    Tensile strength – connector to tubing- (Implied to meet performance of predicate / safety standards)Pass
    Tensile strength – connector to O2 spout- (Implied to meet performance of predicate / safety standards)Pass
    Resistance to kinking- (Implied to meet performance of predicate / safety standards)Pass
    Accelerated AgeingASTM F1980Pass
    Particulate EmissionsBS ISO 18562-2 2017Pass

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample size (number of devices) used for each specific test in the "test set". However, it mentions "Verification tests were performed to establish the safety and efficacy of Flexicare's Adult Single Use Resuscitator Bag." and similarly for Pediatric and Infant versions. The tests would likely involve a statistically relevant number of devices to ensure reliability.

    The data provenance is from Flexicare Medical Limited, United Kingdom. The study is retrospective in the sense that it compares the newly developed Flexicare device against existing, legally marketed predicate devices (Ambu SPUR II Adult, Pediatric, and Infant Single Patient Use Resuscitators) that have already been cleared by the FDA. The performance testing itself would be prospective for the Flexicare device.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document does not mention the use of experts to establish a "ground truth" in the traditional sense for the performance testing. Instead, the ground truth is established by objective measurements against recognized international standards and direct comparison of performance characteristics with legally marketed predicate devices.

    4. Adjudication Method for the Test Set

    There is no mention of an adjudication method involving human reviewers for the test set. The validation relies on objective laboratory testing against predefined criteria and standards, and direct comparison of measured characteristics with those of predicate devices.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is typically associated with AI/CAD systems for image interpretation, which is not applicable to a manual resuscitator bag. The study is a non-clinical performance and biocompatibility evaluation.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study Was Done

    No, a standalone (algorithm-only) study was not done. This device is a manual medical device and does not involve any algorithms or AI for its operation.

    7. The Type of Ground Truth Used

    The ground truth used for this device's validation is based on:

    • Recognized International Standards: Compliance with standards such as BS EN ISO 10651-4:2009, ISO 10993 series, ISO 5356-1:2004, BS EN 13544-2:2002+A1:2009, BS ISO 18562-2 2017, and ASTM F1980.
    • Performance Specifications of Legally Marketed Predicate Devices: The performance characteristics (e.g., volume, resistance, force, dead space, stroke volume, pressure relief) of the Flexicare devices are compared directly to those of the Ambu SPUR II predicate devices. The goal is to demonstrate "substantial equivalence," meaning the new device performs as safely and effectively as the predicate.
    • Objective Laboratory Measurements: The "Non-clinical Test Results" section explicitly states that "Verification tests were performed to establish the safety and efficacy" including visual inspection, dimensional inspection, internal volume, valve function, drop testing, immersion in water, oxygen concentration, resistances, valve malfunction, tidal volume calculation, O2 tube testing, conical connector compliance, biocompatibility, and particulate emission testing.

    8. The Sample Size for the Training Set

    There is no training set as this is a physical medical device, not an AI or machine learning algorithm.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this type of medical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K170663

    Validate with FDA (Live)

    Date Cleared
    2018-05-11

    (431 days)

    Product Code
    Regulation Number
    868.5915
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Single patient use manual resuscitator for use hospital, transport, emergency, and post hospital care to temporary ventilate a patient for the given body mass ranges of: Infant: less than or equal to 10Kg. Child: less than or equal to 23 Kg. Adult: greater than 23 Kg

    This manual resuscitator may be supplied with a single patient use positive end expiratory pressure (PEEP) valve and / or disposable Airway Pressure Manometer.

    The PEEP Valve is a single patient use positive end expiratory pressure (PEEP) valve for use hospital. transport, emergency, and post hospital care to evaluate end lung pressure above atmospheric at the end of exhalation in constant and intermittent gas flow conditions. Intended for patients that the clinician has determined need PEEP.

    The Disposable Manometer is a single patient use manometer intended to be used for monitoring the patient's airway pressure during ventilation. The manometer is to be used with resuscitation systems.

    Device Description

    Foremount Disposable PVC Resuscitators are portable medical devices used to temporarily augment ventilation in patients during ventilatory insufficiency or ventilatory failure. They consist of Ventilation Bag, Patient Valve, Intake Valves, Reservoir Bag, Oxygen Tubing, Cushion Mask, and optional Diverter Ring, PEEP Valve and Manometer. Foremount Disposable PVC Resuscitators come in three sizes along with a ventilation bag:

    • . Infant - Less than or equal to 10 kg
    • Child less than or equal to 23 kg ●
    • Adult Greater than 23 kg. ●

    The ventilation bags are available in three sizes based upon the intended patient population. They are provided with masks in three sizes (#1 - Infant, #3 - Child, and #5 - Adult). The patient valve includes a duck-bill valve to prevent rebreathing and incorporates a 40 cmH2O pop off valve for Child and Infant models. The patient valve includes a duck-bill valve to prevent rebreathing and incorporates a 60 cmH2O pop off valve for Adult models.

    AI/ML Overview

    This is a 510(k) premarket notification for the Foremount Disposable PVC Resuscitator Model A1, A2, B1, B2 and Accessories, including a PEEP valve and a disposable manometer. The submission aims to demonstrate substantial equivalence to previously cleared devices.

    Here's an analysis of the acceptance criteria and study information provided:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't explicitly define "acceptance criteria" in a dedicated table for each device component with specific numerical targets. Instead, it demonstrates performance by comparing the Foremount devices to their respective predicates or by stating that the device meets relevant ISO standards.

    We can infer the acceptance criteria are met if the Foremount device's specifications and performance fall within acceptable limits defined by the predicate devices and applicable ISO standards.

    Inferred Acceptance Criteria and Reported Device Performance (Resuscitator)

    CharacteristicAcceptance Criteria (Inferred from Predicate/Standard)Reported Device Performance (Foremount)
    General
    Indications for UseSingle patient use manual resuscitator for hospital, transport, emergency, and post-hospital care to temporarily ventilate in given body mass ranges.Same as predicate: Infant ≤ 10Kg, Child ≤ 23 Kg, Adult > 23 Kg (for Models A1, B1); Adult > 23 Kg (for Models A2, B2)
    Patient PopulationInfant ≤ 10Kg, Child ≤ 23 Kg, Adult > 23 KgSame as predicate
    Environment of UseHospital, transport, emergency, post-hospital careSame as predicate
    Duration of UseSingle patient, disposable < 24 hoursSingle patient, disposable < 24 hours
    Operational Temperature-18°C ~ 50°C-18°C ~ 50°C
    Storage Temperature-40°C ~ 60°C-40°C ~ 60°C
    Functional Performance (ISO 10651-4 and others)
    Dimensions (Size)Adult: 570x190 mm, Child: 510x190 mm, Infant: 430x180 mm (Predicate)Adult: 445x190 mm, Child: 350x177 mm, Infant: 325x165 mm. Note: Differences in dimensions were assessed and did not raise new safety/effectiveness concerns.
    Ventilation Bag VolumeAdult 1500 ml, Child 600ml, Infant 280 ml (Predicate)Adult 1700 ml, Child 500 ml, Infant 320 ml. Note: Differences in volume were assessed and did not raise new safety/effectiveness concerns, as ISO 10651-4 specifies minimum requirements.
    Oxygen Collection Bag VolumeAdult 1000 ml, Child 1000 ml, Infant 500 ml (Predicate)Adult 1000 ml, Child 1000 ml, Infant 600 ml
    Max Stroke Volume (single hand)Adult 700 ml, Child 360 ml, Infant 150 ml (Predicate)Adult 650 ml, Child 370 ml, Infant 180 ml. Note: Differences in volume were assessed and did not raise new safety/effectiveness concerns as ISO 10651-4 specifies minimum requirements.
    Dead Space4.5 ml (measured) / 6.8 ml (labeled) for all sizes (Predicate)~ 3.8 ml for all sizes
    Inspiratory ResistanceInfant: ≤ 0.5cm H2O, Child: ≤ 0.7cm H2O, Adult: ≤ 3.3cm H2O (Predicate)Infant: 0.5cm H2O, Child: 0.5cm H2O, Adult: 3cm H2O
    Expiratory ResistanceInfant: ≤ 0.8cm H2O, Child: ≤ 0.8cm H2O, Adult: ≤ 2.6cm H2O (Predicate)Infant: 0.5cm H2O, Child: 0.5cm H2O, Adult: 2.8 cmH2O
    Supplemental Oxygen %Comparable to predicate at various flow rates and tidal volumes (e.g., Infant Vt-70ml x 20bpm: 91/97/99% at 2/5/10 lpm)Infant Vt-70ml x 20bpm: 90/98/98% at 2/5/10 lpm (and other comparable values).
    Pop-Off or Pressure Limiting40 cm H2O (Infant/Child), 60 cm H2O (Adult)40 cm H2O (Infant/Child), 60 cm H2O (Adult)
    Ability to add PEEP valvesYes, 0-20 cm H2OYes, 0-20 cm H2O
    Patient connectors15 / 22 mm (standard)15 / 22 mm
    PEEP valve fittings22 / 30 mm (standard)22 / 30 mm
    BiocompatibilityNon-cytotoxic, non-sensitizers, non-irritants; meets ISO 10993-5, -10, -11; ISO 18562-2 for PM2.5 and inorganic gases.Found to be non-cytotoxic, non-sensitizers, non-irritants. Tests for Cytotoxicity, Sensitization, Irritation, Acute Systemic Toxicity, Gas emission (VOC), PM2.5, Inorganic gases (Ozone, CO, CO2) were performed.
    MaterialsPVC, Polycarbonate, Silicone (Predicate)PVC, Polycarbonate, Silicone
    Conical connectorsMeets ISO 5356-1Meets ISO 5356-1
    Shelf-life3 years (Predicate)5 years. Note: This is a difference, but the document concludes no new safety or effectiveness concerns are raised.

    Inferred Acceptance Criteria and Reported Device Performance (PEEP Valve)

    CharacteristicAcceptance Criteria (Inferred from Predicate/Standard)Reported Device Performance (Foremount)
    Indications for UseSingle patient use PEEP valve for hospital, transport, emergency, and post hospital care to evaluate end lung pressure above atmospheric.Same as predicate
    Pressure range2.5 to 10 cmH2O; 5 to 20 cmH2O (Predicate)0 to 10 cmH2O; 5 to 20 cmH2O. Note: Broader lower range, but within similar operational function.
    AdjustableYes (Predicate also had fixed options)Yes
    Accuracy+/- 2 cmH2O+/- 2 cmH2O
    Connectors22 mm / 30 mm (standard)22 mm / 30 mm
    BiocompatibilityMeets biocompatibility standards for externally communicating, limited duration use.Externally communicating, limited duration of use (<24 hours). Materials are Polycarbonate, Silicone, Stainless Steel spring; passed biocompatibility testing.
    MaterialsPolycarbonate, Silicone, Stainless Steel spring (Predicate)Polycarbonate, Silicone, Stainless Steel spring
    Performance TestingAccuracy, Repeatability, Effects of Aging, Drop Test, Conical fittingsAll listed tests were performed. Results implicitly met 'acceptance criteria' by demonstrating substantial equivalence, though specific numerical results beyond accuracy were not detailed in the comparison table.

    Inferred Acceptance Criteria and Reported Device Performance (Disposable Manometer)

    CharacteristicAcceptance Criteria (Inferred from Predicate/Standard)Reported Device Performance (Foremount)
    Indications for UseSingle patient use manometer for monitoring patient's airway pressure during ventilation, used with resuscitation systems.Same as predicate
    Pressure range0 to 60 cmH2O0 to 60 cmH2O
    MarkingsIncrements of 10 cmH2OIncrements of 10 cmH2O
    Accuracy±2 cmH2O @ 0-10 cmH2O, ±4 cmH2O @ 20 cmH2O, ±5 cmH2O @ 30-40 cmH2O, ±7 cmH2O @ 50-60 cmH2O (Predicate)±1 cmH2O @ 0-10 cmH2O, ±2 cmH2O @ 20-30 cmH2O, ±3 cmH2O @ 40-60 cmH2O. Note: The Foremount device reports better accuracy than the predicate. This difference is considered acceptable as it is a performance improvement.
    Connectors15 / 22 mm (standard)15 / 22 mm
    BiocompatibilityMeets biocompatibility standards for externally communicating, limited duration use.Externally communicating, limited duration of use (<24 hours). Materials are Polycarbonate, Silicone, Stainless Steel Springs; passed biocompatibility testing.
    MaterialsPolycarbonate, Silicone, Stainless Steel Springs (Predicate)Polycarbonate, Silicone, Stainless Steel Springs
    Performance TestingAccuracy, Repeatability, Effects of Aging, Drop Test, Conical fittingsAll listed tests were performed. Results implicitly met 'acceptance criteria' by demonstrating substantial equivalence, though specific numerical results beyond accuracy were not detailed in the comparison table.

    2. Sample Size Used for the Test Set and Data Provenance

    The document describes non-clinical bench testing. It does not specify a "test set" in terms of patient data. The testing involves physical samples of the devices.

    • Sample Size: Not explicitly stated for each individual bench test. The nature of the tests (e.g., stroke volume, resistance, oxygen concentration, drop test, aging) implies a sufficient number of devices were tested to draw statistically valid conclusions for engineering verification.
    • Data Provenance: This is bench testing data, originating from the manufacturer's testing facilities (Foremount Enterprise Co., Ltd. in Taichung City, Taiwan). It is prospective in that the testing was performed specifically for this 510(k) submission. No patient data (retrospective or prospective) is mentioned as this is a physical device rather than an AI/software device.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This information is not applicable as this submission concerns the hardware and functional performance of medical devices (resuscitators, PEEP valves, manometers) and involves bench testing against engineering standards and predicate device performance, not clinical studies involving expert interpretation of data or images.

    4. Adjudication Method for the Test Set

    This information is not applicable. Adjudication typically refers to the process of resolving disagreements between experts in clinical studies or image review. Bench testing results are objective measurements against specified standards or comparative measurements against predicate devices.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a MRMC comparative effectiveness study was not done. This type of study is relevant for AI or imaging software where human readers' performance with and without AI assistance is evaluated on a set of cases. This submission is for a physical medical device.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable. This submission is for a physical medical device. There is no "algorithm" or standalone AI component. Performance is assessed through comprehensive bench testing of the device's physical and functional attributes.

    7. The Type of Ground Truth Used

    The "ground truth" for this submission is established by:

    • Engineering Standards: Primarily ISO 10651-4:2002 (Lung ventilators - Particular requirements for operator-powered resuscitators) and ISO 5356-1:2004 (Anaesthetic and respiratory equipment - Conical connectors).
    • Predicate Device Performance: The performance characteristics (e.g., stroke volume, resistance, oxygen concentration, accuracy for PEEP and manometer) of the legally marketed predicate devices (K082092 - GaleMed Dispo-Bag Manual Resuscitator and K122077 - Intersurgical - Airway Pressure Manometer) serve as a comparative benchmark.
    • Biocompatibility Standards: ISO 10993 series and ISO 18562-2 for material safety.

    8. The Sample Size for the Training Set

    Not applicable. This submission is for hardware medical devices. There is no concept of a "training set" as there is no AI algorithm being developed or submitted. The devices are manufactured based on design specifications and then tested.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. Since there is no "training set," the concept of establishing ground truth for it does not apply.

    Ask a Question

    Ask a specific question about this device

    K Number
    K152931

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-08-29

    (329 days)

    Product Code
    Regulation Number
    868.5915
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Ambu SPUR II resuscitator is a single patient use resuscitator intended for pulmonary resuscitation.

    The range of application for each version is:

    -Adult: Adults and children with a body weight more than 30 kg (66lbs).

    -Paediatric: infants and Children with a body weights up to 30 kg (66lbs).

    -Infant: Neonates and infant with a body weight up to 10kg (22lbs).

    Device Description

    Ambu SPUR II Resuscitator is intended for manual pulmonary resuscitation. It is a disposable device intended for Single Patient Use only.

    Ambu SPUR II Resuscitator is available in three sizes: Adult, Pediatric and Infant.

    The main components of the product are a patient valve, a bag, an inlet valve, a reservoir bag/tube and an oxygen tube.

    The patient valve directs the ventilation air through a patient connector into the patient airway and directs the patient expiration air through an expiration connector. The patient connector is constructed to turn around its own axis enabling the resuscitator to turn in relation to connected masks or endotracheal tubes.

    A pressure limiting valve (40 cmHzO) with an override mechanism is placed in the patient valve housing of all Pediatric and Infant resuscitators. For the Adult resuscitators a pressure limiting valve is optional.

    M-Port is mounted in the patient valve housing and provides access to the inspiratory and expiratory gas flow allowing connection of a syringe for drug delivery or a qas sampling line for measuring side stream EtCO2. When the M-Port is not used it is sealed with a cap and the resuscitator operates as a resuscitator without the M-Port. The M-Port is an optional feature.

    The bag of the resuscitator is made of a flexible and elastic material. which ensures the bag to have sufficient compression and recoiling properties. When the bag is compressed, air is delivered via the patient valve to the patient.

    The inlet valve allows ambient air to flow into the bag and prevents air flowing backwards from the bag through the inlet valve during ventilation.

    The reservoir consists of either a reservoir bag (closed reservoir) or a reservoir tube (open reservoir) attached to the inlet valve.

    Supplementary oxygen can be supplied through the oxygen tube attached to the oxygen connector (reservoir bag) or mounted inside the inlet valve assembly (reservoir tube). Supplementary oxygen flows into the reservoir attached to the inlet valve assembly.

    A Face mask, Endotracheal Tube or Larynqeal Mask/Combitube can be connected to the patient connector of the resuscitator to ensure contact to the patient's airway.

    The Ambu SPUR II Resuscitator complies with ISO 10651-4: Particular requirements for operator powered resuscitators.

    AI/ML Overview

    The provided text is a 510(k) Premarket Notification from the FDA for the Ambu SPUR II Resuscitator. This document is a regulatory submission for a medical device and describes the device's technical specifications and how it demonstrates substantial equivalence to existing devices. It does not describe an AI/ML-driven medical device, nor does it detail a study that proves an AI/ML device meets specific acceptance criteria in the context of clinical performance or human reader improvement.

    Therefore, the requested information, specifically regarding acceptance criteria and studies related to AI/ML device performance (like sample size, data provenance, expert ground truth, adjudication, MRMC studies, standalone algorithm performance, training set details, etc.), cannot be extracted from this document, as it pertains to a manual medical device (resuscitator) and not an AI-powered diagnostic or assistive tool.

    The document primarily focuses on bench testing and compliance with recognized consensus standards to demonstrate the safety and effectiveness of the manual resuscitator in comparison to predicate devices, rather than AI-related performance metrics.

    However, I can extract the acceptance criteria and performance data for the manual resuscitator itself, as presented in the document, though it won't be AI-related.

    Here's an attempt to answer the request based only on the provided document, interpreting "acceptance criteria" as the performance parameters tested for this manual device:

    Device: Ambu SPUR II Adult, Pediatric, and Infant Resuscitator (a manual medical device, not an AI/ML device)

    The document describes the device's substantial equivalence to predicate devices based on its intended use, operation, technological characteristics, and performance testing. The "acceptance criteria" here refer to the successful passing of various performance tests as per recognized consensus standards for manual resuscitators.


    Table of Acceptance Criteria and Reported Device Performance (for the Manual Resuscitator)

    Acceptance Criteria Category/ParameterReported Device Performance/Result
    Compliance with Standards:
    - ISO 10651-4: Particular requirements for operator-powered resuscitatorsPassed. All tests described in the standard were performed, and the device demonstrated conformity.
    - ISO 5356-1: Anaesthetic and respiratory equipment Conical connectors - Part 1: Cones and socketsPassed. All tests described in the standard were performed, and the device demonstrated conformity.
    Connectors Functionality:
    - Patient connection port connectorPassed. Tested for functionality and compliance.
    - Expiratory port connector for breathing gasesPassed. Tested for functionality and compliance.
    - Face mask connectorsPassed. Tested for functionality and compliance.
    - Oxygen tube connector and pressure gauge connectorPassed. Tested for functionality and compliance.
    Operation Requirements:
    - Patient valve function after contamination with vomitsPassed.
    - Mechanical shockPassed.
    - Drop testPassed.
    - Immersion in waterPassed.
    Ventilatory Requirements:
    - Supplementary oxygen and delivered oxygen concentrationPassed.
    - Expiratory resistancePassed. (Adjusted to fulfill ISO 10651-4:2009 for Pediatric and Infant models).
    - Inspiratory resistancePassed.
    - Patient valve malfunctionPassed.
    - Patient valve leakage Forward leakagePassed.
    - Resuscitator dead-space and rebreathingPassed.
    Ventilation Performance:
    - Minimum delivered volumePassed. (Stroke volume specification updated for Adult model to show compliance with ISO 10651-4:2009).
    - Pressure limitationPassed.
    Storage and Operation (tested at time zero and after accelerated aging):
    - Inspiratory/Expiratory resistance testPassed.
    - Stroke volume testPassed.
    - Patient valve malfunction testPassed.
    - Supplementary oxygen and delivery oxygen concentrationPassed.
    - Pressure limitation testPassed.
    Additional Performance Data:
    - CO2 monitoring (M-port suitability)Passed with at least 0.4 liter of tidal volume. Acceptance criteria: Maximum deviation of 5% at a confidence level of 95% in comparison to the Philips Side Stream Adapter.
    - MR Conditional properties (Magnetic Field Interaction Testing per ASTM F2503-05)Passed. Ambu SPUR II is MR Conditional and can be used in a MRI environment under specified conditions.
    Biocompatibility (compliance with ISO 10993-1):
    - Cytotoxicity (ISO 10993-5)Passed.
    - Sensitization (ISO 10993-10)Passed.
    - Intracutaneous reactivity test (ISO 10993-10)Passed.
    - Extractable and Leachable (E&L) testingPassed.

    AI/ML Specific Questions (Cannot be answered from the provided document as it is for a manual medical device)

    Since this document describes a manual medical device (resuscitator) and not an AI/ML diagnostic or assistive device, the following points regarding AI/ML studies are not applicable and thus, no information can be extracted:

    1. Sample sizes used for the test set and the data provenance: Not applicable. The document refers to engineering bench tests, not clinical data sets for AI.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for an AI model is not relevant here.
    3. Adjudication method for the test set: Not applicable.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: No, this was not done as it's not an AI device. The document explicitly states "No Clinical tests are performed."
    5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable. The "ground truth" for this device's performance is objective measurements against engineering standards.
    7. The sample size for the training set: Not applicable. This device is not an AI/ML model that requires training data.
    8. How the ground truth for the training set was established: Not applicable.

    In summary: The provided document is a regulatory submission for a traditional, manual medical device. It thoroughly outlines the bench testing and compliance with established standards to demonstrate safety and effectiveness. However, it does not contain any information related to AI/ML device performance, human-AI interaction studies, or the methodologies typically employed for AI/ML model validation.

    Ask a Question

    Ask a specific question about this device

    K Number
    K132172

    Validate with FDA (Live)

    Date Cleared
    2015-01-15

    (549 days)

    Product Code
    Regulation Number
    868.5915
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NeoNatalie Resuscitator is a self-inflating, manual resuscitator intended for newborns and infants up to 5 kg body mass who require respiratory support.

    Device Description

    The NeoNatalie Resuscitator is a self-inflating, manual resuscitator intended for newborns and infants up to 5 kg body mass who require respiratory support. A manual resuscitator is a resuscitation device in which ventilation of the lungs is produced by the operator compressing the compressible unit of the device, the ventilation bag. The ventilation bag is self-inflating: the compressed bag will refill with ambient air via the bag inlet valve. The resuscitator provides positive pressure ventilation of the lungs (when used with a face mask). The resuscitator can be used to provide supplemental oxygen when used with the oxygen kit, the NeoNatalie Resuscitator Oxygen Kit. When a resuscitator fitted with an oxygen reservoir is used to provide supplemental oxygen, the ventilation bag will refill with oxygen from the oxygen reservoir. The NeoNatalie Resuscitator is made of polysulfone, silicone rubber and stainless steel. The NeoNatalie Resuscitator is reusable resuscitator which may be sterilized by autoclaving.

    AI/ML Overview

    The provided text describes the regulatory clearance for the NeoNatalie Resuscitator and includes a comparison to a predicate device, as well as general information about testing performed. However, it does not contain the detailed clinical study information typically provided when discussing "acceptance criteria and the study that proves the device meets the acceptance criteria" in the context of an AI/ML device.

    The NeoNatalie Resuscitator is a mechanical medical device (manual resuscitator), not an AI/ML device, and therefore the specific questions regarding test sets, training sets, ground truth, expert adjudication, and MRMC studies are not directly applicable in the way they would be for an AI/ML diagnostic or predictive tool.

    Based on the provided text, here's an attempt to answer the questions, highlighting where the information is not directly available or applicable:


    1. A table of acceptance criteria and the reported device performance

    The document frames the "acceptance criteria" and "device performance" in terms of compliance with international standards and functional requirements for a manual resuscitator, rather than traditional clinical performance metrics like sensitivity or specificity for an AI algorithm.

    Acceptance Criteria (Defined by Standards/Functional Requirements)Reported Device Performance (NeoNatalie Resuscitator)Predicate Device Performance (Ambu® Mark IV Baby Resuscitator)
    Expiratory Resistance< 2.5 cmH2O at 5 LPM0.6 cmH2O at 5 LPM
    Inspiratory Resistance< 0.5 cmH2O at 5 LPM0.5 cmH2O at 5 LPM
    Dead Space4 ml< 6 ml
    BiocompatibilityMeets FDA guidance and ISO-10993 requirementsNot explicitly stated for predicate in this context
    Compliance with ISO 10651-4:2002 (Operator-powered resuscitators)Demonstrates compliance with only minor deviationsNot explicitly stated for predicate in this context
    Compliance with ISO 5356-1:2004 (Conical connectors)Demonstrates compliance with only minor deviationsNot explicitly stated for predicate in this context
    Reprocessing (High-level disinfection & performance after repeated reprocessing)Meets reprocessing instructions and performance criteriaNot explicitly stated for predicate in this context

    Note: The document states "Design verification and design validation testing demonstrates that the NeoNatalie Resuscitator meets its functional requirements and performance specifications." The table above reflects specific quantitative metrics provided and general statements about compliance with standards.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    This information is not provided in the document. For a mechanical device, "test set" typically refers to physical units or simulated environments rather than patient data. The testing mentioned appears to be laboratory-based design verification and validation.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    This information is not applicable and therefore not provided. The testing relates to engineering and material performance, not expert-derived ground truth for an interpretive task.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    This information is not applicable and therefore not provided. Adjudication methods are relevant for subjective interpretations, often in clinical studies or AI ground truth establishment.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable and therefore not provided. MRMC studies are used to evaluate the performance of diagnostic systems, often AI-assisted, and are not relevant for a manual mechanical resuscitator.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable and therefore not provided. This device is a manual resuscitator and does not involve an algorithm. Its operation inherently involves a "human-in-the-loop" as it is "operator-powered."

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for this device's performance is compliance with established engineering standards (e.g., ISO 10651-4:2002 for lung ventilators, ISO 5356-1:2004 for conical connectors) and functional specifications (e.g., specific resistance values, dead space, biocompatibility). It is determined by objective physical measurements and established test protocols, not clinical outcomes data, pathology, or expert consensus in an interpretive sense.

    8. The sample size for the training set

    This information is not applicable and therefore not provided. There is no "training set" as this is not an AI/ML device.

    9. How the ground truth for the training set was established

    This information is not applicable and therefore not provided. There is no "training set" for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K112852

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-01-13

    (87 days)

    Product Code
    Regulation Number
    868.5915
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Single patient use manual resuscitation device to temporary ventilate neonate, newborn or infant with a body mass of less than or equal to 3.3 Kg in hospital, transport, emergency and post hospital care environments.

    Device Description

    Single patient use medical device, which temporarily augment ventilation insufficiency or ventilatory failure.

    AI/ML Overview

    The document provided outlines the A Plus Medical Babi.Plus™ Neonatal Resuscitation Bag and its substantial equivalence to a predicate device. The primary study presented relates to comparative technological characteristics rather than a clinical study involving human patients or complex AI algorithms.

    Here's a breakdown of the requested information based on the provided text, noting where information is not applicable or not present:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implicitly defined by the ISO 10651-4:2002 standard ("Lung ventilators - Part 4: Particular requirements for opered resuscitators") and the predicate device's performance. The "Babi.Plus™" column shows the reported device performance.

    ItemAcceptance Criteria (Implied by Predicate/ISO)Reported Device Performance (Babi.Plus™)
    Patient Connector:15 mm I.D. / 22 mm O.D. (Predicate)15 mm I.D.
    Face Mask:Option offered, 15 mm O.D. (Predicate)Option offered, 15 mm O.D.
    Pressure Gauge Connector:None (Predicate)4.0 mm O.D. tapered
    Supplemental oxygen delivery:(Predicate values given)VT-20 mls, Rate – 60 BPM:@ 2 LPM – 91%, @ 10 LPM – 99+%VT-20 mls, Rate – 120 BPM:@ 2 LPM – 91%, @ 10 LPM – 99+ %VT-50 mls, Rate – 30 BPM:@ 2 LPM – 75%, @ 10 LPM – 99+%VT-50 mls, Rate – 60 BPM:@ 2 LPM – 66%, @ 10 LPM – 99+%
    Expiratory resistance:< 2.8 cm H2O @ 5 LPM (Predicate)< 1.7 cm H2O @ 5 LPM
    Inspiratory resistance:< 2.8 cm H2O @ 5 LPM (Predicate)< 1.3 cm H2O @ 5 LPM
    Patient Valve malfunction:< 6.0 cm H2O @ 30 LPM (Predicate)< 6.0 cm H2O @ 30 LPM
    Dead space:6.8 mls (Predicate)6.35 mls
    Minimum / Maximum Volume:20 mls / 150 mls (Predicate)20 mls / 50 mls
    Body mass range:≤ 10.0 Kg (Predicate)≤3.3 Kg
    Pressure limitation:40 cm H2O ± 5.0 cm H2O with over-ride, unlimited pressure (Predicate)25 cm H2O ± 3.0 cm H2O with 40 cmH2O ± 5.0 cm H2O over-ride

    2. Sample Size Used for the Test Set and Data Provenance

    This document describes a pre-market notification (510(k)) based on substantial equivalence, primarily through a comparison of technical characteristics to a known predicate device and conformity to an international standard (ISO 10651-4:2002).

    • Test Set Sample Size: Not applicable. The "study" presented here is a summary of comparative technical specifications, not a clinical trial or performance evaluation using a "test set" in the context of AI/diagnostic device validation. The performance values listed (e.g., oxygen delivery, resistance) would have been determined through engineering tests on a sample of the device, but the sample size for these specific tests is not provided.
    • Data Provenance: Not applicable in the sense of clinical data. The performance data is generated through laboratory testing of the device.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. This is not a study that involves expert ground truth for image interpretation or diagnosis. The "ground truth" for the technical specifications is based on established engineering principles, metrology, and compliance with the ISO standard.

    4. Adjudication Method for the Test Set

    Not applicable. There is no expert adjudication process described or implied for the generation of these technical performance metrics.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI-enabled diagnostic device, nor is it a multi-reader study. It is a manual resuscitation bag.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a medical device (manual resuscitation bag), not an algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" for the device's performance characteristics (e.g., flow rates, volumes, resistances) is established through physical measurement and testing against the requirements of ISO 10651-4:2002 and comparison with the specifications of the predicate device. It is an engineering "ground truth."

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI device, so there is no "training set."

    9. How the Ground Truth for the Training Set was Established

    Not applicable. There is no "training set" or corresponding ground truth establishment described.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 8