Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K994415

    Validate with FDA (Live)

    Date Cleared
    2000-02-03

    (36 days)

    Product Code
    Regulation Number
    888.3358
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DePuy Marathon Cross-linked Polyethylene Acetabular Cup Liners are intended to be used with the DePuy Duraloc metal acetabular shells to resurface the acetabular socket in cementless total hip replacement.

    The DePuy Marathon Cross-linked Polyethylene Acetabular Cup Liners are intended to be used with the Duraloc System Acetabular Cups to resurface the acetabular socket in cemented or cementless total hip arthroplasty.

    Device Description

    The DePuy Marathon Cross-linked Polyethylene Acetabular Cup Liners are UHMWPE acetabular cup liners that are available with 28mm or 32mm inner diameters and with a neutral or 10° lip. The 28mm liners are available in sizes to fit Duraloc metal acetabular shells with outer diameters of 48-74mm. The 32mm liners are available in sizes to fit Duraloc metal acetabular shells with outer diameters of 52-80mm. The polyethylene liners are locked into the Duraloc metal shells with a metal wire locking ring which is supplied with the metal shells.

    The Marathon Acetabular Cup Liners are manufactured from UHMWPE that has been cross-linked by exposure to radiation in a very low oxygen environment and then heat treated prior to machining of the liners. The cross-linked polyethylene has physical and mechanical properties that are similar to those of standard UHMWPE but has increased resistance to oxidation and wear. Marathon cross-linked UHMWPE meets all of the specifications of ASTM F648.

    AI/ML Overview

    This document describes the DePuy Marathon Cross-linked Polyethylene Acetabular Cup Liners. The provided text refers to a 510(k) premarket notification for a medical device and doesn't contain information about an AI/ML device study. Therefore, I cannot fulfill your request for an AI-specific study breakdown. However, I can extract information related to the device's performance claims and the study backing those claims based on the provided text.

    Here's a breakdown of the information that is available in the provided document, adapted to your requested format where possible, and explicitly stating why certain AI/ML-specific fields cannot be filled:

    Acceptance Criteria and Device Performance

    The device's acceptance criteria, in terms of proving "reduced wear," are implicitly defined by the comparative wear data against a predicate device (conventional UHMWPE cups). The study aims to demonstrate a significantly lower wear rate for the Marathon cross-linked UHMWPE liners.

    Acceptance Criteria CategorySpecific Criteria/MetricReported Device Performance (Marathon Cross-linked UHMWPE)
    Wear ReductionAgainst highly polished femoral balls85-86% less wear than conventional UHMWPE
    Wear ReductionAgainst moderately roughened femoral heads74-79% less wear than conventional UHMWPE
    Wear ReductionAgainst severely roughened femoral heads17-56% less wear than conventional UHMWPE

    Note: The document explicitly states: "These in vitro results have not been correlated to clinical experience." This is a crucial disclaimer indicating that while the device performs well in a lab setting, its long-term clinical performance (which would involve human patients and outcomes) is not covered by this study.


    Study Details (Non-AI/ML Device)

    Here's a breakdown of the study attributes based on the provided text. Many fields are not applicable or cannot be determined for this type of medical device assessment.

    Information RequestedDetails from the Document
    2. Sample size for test set and data provenanceThe study involved "10 million cycle hip simulator wear data." The "sample size" refers to the specific number of cups tested, but this is not explicitly stated. The data provenance is in vitro (laboratory-based simulation), not human patient data. No country of origin for the data is mentioned, as it's a lab test rather than clinical data from a specific region.
    3. Number of experts and their qualifications for ground truthNot applicable. This is an in vitro mechanical wear study, not an AI/ML study requiring expert ground truth for image or data interpretation.
    4. Adjudication method for the test setNot applicable. There is no expert adjudication in a mechanical wear test. The measurement of wear is typically done quantitatively and objectively.
    5. Multi-reader multi-case (MRMC) comparative effectiveness studyNo. This is a mechanical wear test comparing material performance, not a clinical study involving human readers or AI assistance.
    6. Standalone (algorithm only without human-in-the-loop) performanceNot applicable. This is a medical implant, not an algorithm. The "standalone" performance refers to the device's mechanical wear characteristics as measured in the simulator.
    7. Type of ground truth usedMechanical Wear Measurement. The "ground truth" for this study is the quantified amount of material loss (wear) from the acetabular cups, measured directly following the hip simulator testing. This is a scientific, objective measurement.
    8. Sample size for the training setNot applicable. This is not an AI/ML study requiring a training set.
    9. How ground truth for the training set was establishedNot applicable. This is not an AI/ML study requiring a training set.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1