    Validate with FDA (Live)

    Date Cleared
    2000-02-03

    (71 days)

    Product Code
    Regulation Number
    864.7750
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AcuSure System for Patient Self-Testing is intended for the quantitative prothombin time testing of fresh, capillary whole blood by selected and suitably trained patients or their caregivers on the prescription of the treating physician. The AcuSure System for Patient Self-Testing is for the monitoring of patients on oral anticoagulant therapy and not intended to be used for screening purposes.

    The AcuSure Pro System for Professional Use is intended for the quantitative prothrombin time testing of fresh, capillary whole blood by healthcare professionals. The AcuSure Pro System is for the monitoring of patients on oral anticoagulant therapy and not intended to be used for screening purposes.

    Device Description

    The AcuSure and AcuSure Pro test strips and meter measure capillary blood PT levels. When the sample is applied to the target area on the test strip, the meter detects the electrical drop between two electrodes on the test strip and begins the test. Inside the membrane of the test strip the red blood cells are separated from the plasma. The membrane also contains the reagents necessary to cause the coagulation reaction (clot formation) to occur. The thrombin, formed as part of the coagulation reaction, reacts with a reagent, which causes a fluorescent molecule to be released The time from the initial resistance drop to the onset of fluorescence is proportional to the prothrombin time of the sample.

    AI/ML Overview

    The provided text is a 510(k) summary for the Avocet AcuSure System. It describes the device, its intended use, and its substantial equivalence to a predicate device. However, it does not contain specific details about acceptance criteria, device performance studies, sample sizes, ground truth establishment, or any AI-related information.

    Therefore, I cannot fulfill your request for the detailed table and study information as the input text does not include that data. The document focuses on regulatory approval and substantial equivalence rather than detailed performance metrics.

    To answer your request, I would need a document that describes the clinical studies, their methodology, and results, including acceptance criteria and performance data.