Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K990092

    Validate with FDA (Live)

    Date Cleared
    1999-01-29

    (18 days)

    Product Code
    Regulation Number
    862.1120
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AVL OMNI™ Analyzer is intended to be used for the measurement of pH , pCO2, pCO2 , sodium, potassium, ionized calcium, chloride, hematocrit and total hemoglobin and the hemoglobin derivatives: OzHb, COHb, MetHb, HHb, SulfHb and metabolites; glucose, lactate and urea nitrogen in samples of whole blood, serum, plasma, aqueous solutions as appropriate, in a clinical laboratory setting by personnel minimally qualified to perform and to report these results.

    Device Description

    The AVL OMNI™ Analyzer is a fully-automatic, microprocessor-controlled system that can perform up to 17 tests per sample.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the AVL OMNI™ Modular Analyzer, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state quantitative "acceptance criteria" in a typical pass/fail format with specific thresholds. Instead, it implicitly defines acceptance through the concept of "substantial equivalence" to predicate devices and the demonstration of "no significant difference in mean values" and "correlation" during clinical trials. The "reported device performance" is largely described qualitatively.

    Parameter/TestAcceptance Criteria (Implicit)Reported Device Performance (Qualitative)
    Substantial EquivalenceThe device should be substantially equivalent in function, safety, and efficacy to specific legally marketed predicate devices.Stated as substantially equivalent to "Combi Analyzers" and specifically for Urea Nitrogen to NOVA Biomedical ULTRA H [K960120] and i-STAT 200 Portable Clinical Analyzer (K940918).
    PrecisionDemonstrated acceptable "Within-Run (S_wr), Between-Day (S_dd) and Total (S_T) precision." (No numerical thresholds provided)."Typical Within-Run (S_wr), Between-Day (S_dd) and Total (S_T) precision were evaluated" for all assays. The conclusion statement implies the precision was found acceptable.
    Linearity in Aqueous SolutionsDemonstrated linearity across a known range of values. (No numerical thresholds provided).Aqueous standard solutions measured on three AVL OMNI™ units demonstrated linearity. The conclusion statement implies the linearity was found acceptable.
    Linearity in SerumDemonstrated linearity by diluting patient-sample pools. (No numerical thresholds provided).Patient-sample pools diluted and measured on two AVL OMNI™ instruments demonstrated linearity. The conclusion statement implies the linearity was found acceptable.
    InterferencesNo significant effects on measurement from various analytes or drugs. (No numerical thresholds provided)."No significant effects on measurement were demonstrated at the concentrations evaluated" for urea nitrogen.
    Clinical CorrelationNo significant difference in mean values (P<0.05) when compared to legally marketed predicate devices in a clinical setting."In all evaluations, there was no significant difference in mean values (P<0.05) obtained on measurement by the AVL OMNI™." Additionally, for Urea Nitrogen, a direct comparison showed equivalence to a photometric method on the Boehringer Mannheim/Hitachi 917 Analyzer (K953239).
    Safety and EffectivenessDevice should be safe and effective. (General regulatory requirement and implicit goal of all testing).Demonstrated safe and effective, and equivalent to predicate devices based on clinical and non-clinical trials.

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Sample Size (Clinical Tests): Specimens were "remnant from patient specimens of both whole blood and serum or plasma collected for routine analysis on existing instrumentation." Patient populations included "extreme ranges of clinical values." The total number of patient specimens or individual data points is not specified.
    • Sample Size (Non-Clinical Tests):
      • Precision: "samples of each of the specimen types suitable for measurement on the AVL OMNI™." The number of runs or samples is not specified.
      • Linearity in Aqueous Solutions: "Each of three AVL OMNI™ units" were used, with serially diluted solutions. The number of measurements or dilution series is not specified.
      • Linearity in Serum: "two AVL OMNI™ instruments" were used, with patient-sample pools mixed in varying ratios. The number of samples or ratios is not specified.
    • Data Provenance: The data is retrospective, as it used "remnant from patient specimens... collected for routine analysis on existing instrumentation." The country of origin is not specified, but given the submitter's address in Roswell, GA, USA, it's likely primarily US-based, though this is an assumption.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document states that "personnel trained to perform and report these analyses" operated the devices in the clinical setting. It does not mention a specific number of experts used for ground truth establishment or their qualifications (e.g., radiologist with X years of experience). The ground truth for the clinical correlation studies appears to be the results obtained from the legally marketed predicate devices.

    4. Adjudication Method for the Test Set

    No explicit adjudication method (e.g., 2+1, 3+1) is mentioned. The comparison was directly between the AVL OMNI™ and the predicate devices, with statistical analysis ("no significant difference in mean values (P<0.05)").

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This section is not applicable. The device described is an automated analyzer, not an AI-assisted diagnostic tool that would typically involve human readers interpreting images or data with and without AI assistance. Therefore, an MRMC study related to AI assistance for human readers was not performed.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    This section is applicable. The AVL OMNI™ Modular Analyzer is described as a "fully-automatic, microprocessor-controlled system" that performs measurements. The clinical and non-clinical tests described are evaluations of this standalone algorithm/device performance against predicate devices and established testing methodologies (precision, linearity, interferences).

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    The primary ground truth used in the clinical correlation studies was the measurements obtained from legally marketed predicate devices. For non-clinical tests like linearity in aqueous solutions, the "ground truth" was the gravimetrically prepared standards with known values.

    8. The Sample Size for the Training Set

    The document is a 510(k) summary for an in vitro diagnostic (IVD) device (a laboratory analyzer), not a machine learning/AI device that would typically have a distinct "training set" in the common sense. Therefore, no sample size for a training set is specified. The device is based on established principles of measurement (potentiometric, amperometric, photometric, spectrophotometric, conductance) rather than being trained on a large dataset.

    9. How the Ground Truth for the Training Set Was Established

    As explained in point 8, the concept of a "training set" for this type of device is not directly applicable. The device's principles of operation are based on validated chemical and physical measurement techniques, not on learning from a labeled dataset. Therefore, this information is not provided nor relevant in the context of this device's submission.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1