Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K981950

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1998-11-03

    (153 days)

    Product Code
    Regulation Number
    864.5200
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The intended use of the Sysmex R-3500 is as a fully automated reticulocyte analyzer for in vitro diagnostic use in clinical laboratories.

    Device Description

    The Sysmex™ R-3500 is an automated reticulocyte analyzer intended for in vitro use in clinical laboratories. The R-3500 provides accurate and precise test results for 8 analysis parameters in whole blood. These include RET%, RET#, RBC, IRF, LFR, MFR, HFR, and PLT. The R-3500 processes approximately 120 samples per hour and displays and prints the data for Reticulocyte number, Reticulocyte percent, Red blood cell count, Immature reticulocyte fraction, fluorescent ratios, and platelets along with representative scattergrams. Sample abnormalities are indicated by abnormal marks, flags, and error messages which appear on the DMS display screen and on the printout. This is an indication that the sample is not within the acceptable range and requires further review and investigation. The R-3500 uses the principle of flow cytometry for reticulocyte analysis. In the instrument, a whole blood sample is automatically aspirated, diluted and stained with a fluorescent dye (Auromine-O). The sample is hydrodynamically focused into a narrow path and passed through a flow cell, where it is illuminated by an Argon laser beam. The cells present in the sample will fluoresce and scatter light to varying degrees. It is the analysis of the intensity of emitted fluorescent light and intensity of scattered light which allows the R-3500 analyzer to detect and enumerate reticulocytes.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study details for the Sysmex™ Automated Reticulocyte Analyzer R-3500 based on the provided text:

    1. Acceptance Criteria and Reported Device Performance

    The document does not explicitly state pre-defined quantitative acceptance criteria (e.g., "RET% must have an r-value > 0.990"). Instead, it demonstrates the device's performance through correlation studies comparing it to a predicate device. The implicit acceptance criterion is "substantial equivalence" to the predicate device, which is supported by high correlation coefficients (r and r²) and regression equations.

    ParameternrRegression EquationReported Device Performance (Correlation with Predicate Device)
    RET#4870.9940.988y = 0.965x + 0.001Very strong positive correlation
    RET%4870.9970.994y = 0.964x + 0.051Very strong positive correlation
    RBC4860.9980.997y = 1.009x - 0.072Very strong positive correlation
    IRF4860.9560.913y = 0.948x + 1.409Strong positive correlation
    LFR4860.9560.913y = 0.948x + 3.819Strong positive correlation
    MFR4860.9230.852y = 0.917x + 1.433Strong positive correlation
    HFR4860.9540.910y = 0.940x + 0.490Strong positive correlation
    Platelet4820.9940.989y = 0.937x + 10.619Very strong positive correlation

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The sample size for the test set varied slightly by parameter:
      • RET# & RET%: 487 samples
      • RBC, IRF, LFR, MFR, HFR: 486 samples
      • Platelet: 482 samples
    • Data Provenance:
      • Country of Origin: Not explicitly stated, but the "research center" where the studies were performed is mentioned. The manufacturer is TOA Medical Electronics Co. in Kobe, Japan, and the importer/distributor is Sysmex™ Corporation in Long Grove, IL, USA. Given the context of seeking FDA clearance, it's likely the studies were conducted to satisfy US regulatory requirements, but the specific geographic origin of the patient samples is not provided.
      • Retrospective or Prospective: The text states, "In these studies, the following comparative performance evaluations were conducted using the proposed device and the predicate device to evaluate specimens from apparently healthy individuals and from patients with different pathological conditions which are expected to affect the results for particular parameters." This suggests the samples were collected and then tested on both devices for comparison, which aligns with typical prospective or concurrent comparison study methodology. It doesn't indicate purely retrospective analysis of existing data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not mention the use of experts to establish a "ground truth" for the test set. The study compares the performance of the new device (R-3500) against an already cleared predicate device (RAM-1). The predicate device's measurements are effectively treated as the reference for comparison.

    4. Adjudication Method for the Test Set

    Not applicable. Since the comparison is primarily device-to-device measurements, there is no mention of human adjudication for the test set results. Each device generated its own results, which were then statistically compared.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI/algorithm-assisted human reading device. It's a fully automated analyzer. The study focuses on the comparison between two automated instruments.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes, the study described is a standalone performance evaluation. The Sysmex™ R-3500 is an automated reticulocyte analyzer. The performance data presented (correlation coefficients and regression equations) represent the device's measurements compared directly to those of a predicate automated device. There is no human intervention in the generation of the results being compared.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" in this context is established by the measurements from the predicate device, Sysmex™ SE/RAM-1. The study's objective is to demonstrate substantial equivalence, meaning the new device performs comparably to a legally marketed device.

    8. The Sample Size for the Training Set

    The document does not provide information about a "training set" or its sample size. This type of 510(k) submission for a diagnostic analyzer typically focuses on demonstrating the performance of the final, released device compared to a predicate, rather than detailing the development and training phases of an algorithm.

    9. How the Ground Truth for the Training Set was Established

    Not applicable, as no training set information is provided in the document.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1