Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K964626

    Validate with FDA (Live)

    Date Cleared
    1997-02-10

    (83 days)

    Product Code
    Regulation Number
    892.1000
    Age Range
    All
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Picker International EDGE and VISTA systems are indicated for use as a NMR device that produces images that: (1) correspond to the distribution of protons exhibiting NMR, (2) depend upon the NMR parameters (proton density, flow velocity, spin-lattice relaxation time (T1), and spin-spin relaxation time (T2)) and (3) display the soft tissue structure of the head and whole body. When interpreted by a trained physician, these images yield information that can be useful in the determination of a diagnosis.

    Device Description

    The EDGE/VISTA system enhancement package includes changes in the computer subsystem, the gradient subsystem, the RF subsystem, the magnet subsystem, the patient handling subsystem, and the magnet enclosure. The main features provided with this enhancement are a 61 cm body coil, three gradient performance levels, water cooled gradients, an increase in the SAR first controlled operating mode limit, and optional increased reconstruction rates.

    AI/ML Overview

    The provided document is a 510(k) summary for a GE Healthcare product, specifically an enhancement package for their EDGE/VISTA MR systems. This type of submission is for demonstrating substantial equivalence to a predicate device, not for proving that a device meets specific clinical acceptance criteria through a study.

    Therefore, the document does not contain the information requested for acceptance criteria and a study proving the device meets those criteria in the way a clinical trial or performance study would. It focuses on comparing the technical specifications and safety parameters of the new device to existing, legally marketed predicate devices.

    Here's why the requested information isn't present and what the document does provide:

    • No Acceptance Criteria in the context of clinical performance: The document implicitly demonstrates "acceptance" by showing the device's technical specifications and safety profiles are comparable to, or improved upon, predicate devices, indicating it meets general regulatory safety and effectiveness expectations for MR systems. There are no explicit clinical performance metrics (e.g., sensitivity, specificity for a diagnostic task) being tested.
    • No "Study" as a traditional clinical trial: The document describes modifications and enhancements to an existing MR system. The "proof" is through a comparative table highlighting substantial equivalence in engineering, safety, and operational features. It's not a study in the sense of a clinical investigation with a test set, ground truth established by experts, etc.
    • No details on sample size, data provenance, expert ground truth, adjudication, MRMC, or standalone performance: These concepts are relevant to clinical performance studies, which are not detailed in this 510(k) summary for a system enhancement.

    However, I can extract information related to the technical comparisons made for demonstrating "substantial equivalence," which serves as the "proof" in this context:

    1. Table of "Acceptance Criteria" (Implicit Safety/Performance Parameters) and Reported Device Performance

    Since there are no explicit "acceptance criteria" for clinical performance, I will present the key technical and safety specifications that are being compared to demonstrate substantial equivalence, considering these as the "criteria" for regulatory acceptance in this context.

    Item/Criterion (Implicit)Predicate Devices (EDGE/VISTA K931544, K954646) PerformanceEDGE/VISTA with System Enhancement Package PerformanceComparison Outcome (Proving "Acceptance")
    Computer SubsystemDedicated Ethernet, single monitor, second Ethernet for external network.Similar display/database computer, monitor, Ethernet. DICOM v3.0 compliant. Optional increased reconstruction rates.Substantially Equivalent; added modern communication (DICOM) and optional performance improvements.
    Image Storage (Short Term/Archival)Magnetic Disk/Optical DiskSameSubstantially Equivalent.
    Gradient CoilsAir-cooled (Standard EDGE) or Water-cooled (High Performance Gradients)Water-cooled self-shielded for all performance levels.Improved/Equivalent; standardizes to the higher performance, water-cooled model.
    Max. Gradient Strength16 or 27 mT/m16, 20 or 27 mT/mImproved/Equivalent; adds an intermediate gradient strength option.
    Max. Slew Rate25 or 72 mT/m/msec25, 40, 72 mT/m/msecImproved/Equivalent; adds an intermediate slew rate option.
    Gradient Amps135 A RMS, 220 A peak, 350 V OR 270 A RMS, 440 A peak, 600 V150 A RMS, 300 A peak, 400 V OR 270 A RMS, 440 A peak, 600 VImproved/Equivalent; slight increase in the lower-tier amp values.
    RF AmplifiersSingle frequency RF amplifiers.Single frequency RF amplifiers with increased output power (for larger body coil).Improved/Equivalent; adjusted for new 61cm body coil maintain equivalent B1 field.
    Body Coil58 cm quadrature multi-conductor transmit/receive coil.61 cm quadrature multi-conductor transmit/receive coil.Improved; larger body coil for potentially greater patient comfort/versatility.
    Patient HandlingComputer controlled patient transport.Computer controlled patient transport with increased weight capacity. New patient illumination.Improved; increased capacity and comfort features.
    Time Varying Magnetic Field (dB/dt)Standard EDGE: ≤ 20 T/s. High Performance: Normal ≤ 40 T/s, Controlled 40-60 T/s.Normal Operating Mode: ≤ 40 T/s. First Controlled Operating Mode: 40 T/s < dB/dt ≤ 60 T/s.Equivalent to High-Performance Gradient predicate; acceptable within safety limits.
    Radiofrequency Absorption (SAR)Normal: ≤ 1.2 W/kg. Controlled: ≤ 2.4 W/kg.Normal: ≤ 1.2 W/kg. First Level Controlled Operating Mode: ≤ 3.2 W/kg. Change in RF scale factor acquisition.Adjusted; Controlled Operating Mode SAR limit increased to 3.2 W/kg, requires demonstration of safety within this limit, presumed to be handled by the change in RF scale factor acquisition method.
    Acoustic Noise (Typical)90.7 dBA (avg), 100.0 dB (peak)80.6 dBA (avg), 93.3 dB (peak)Improved; lower average and peak acoustic noise.
    Acoustic Noise (Worst Case)118.7 dBA (avg), 125.6 dB (peak)115.1 dBA (avg), 123.8 dB (peak)Improved; lower average and peak acoustic noise.

    The remaining requested information is not available in this 510(k) summary document:

    • 2. Sample size used for the test set and the data provenance: Not applicable to this type of regulatory submission (substantial equivalence based on technical comparison).
    • 3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable.
    • 4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
    • 5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable; this is an MR system enhancement, not an AI diagnostic tool.
    • 6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
    • 7. The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable.
    • 8. The sample size for the training set: Not applicable.
    • 9. How the ground truth for the training set was established: Not applicable.

    This document clearly focuses on technical and safety comparisons to predicate devices rather than clinical performance studies. The "study" here is the engineering and safety analysis performed by the manufacturer to demonstrate that the enhanced system maintains or improves upon the safety and effectiveness of its predecessors.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1