510(k) Data Aggregation

    K Number
    K252867

    Validate with FDA (Live)

    Date Cleared
    2025-10-09

    (30 days)

    Product Code
    Regulation Number
    862.3100
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VINScreen Urine Drug Test Cup is lateral flow immunoassays for rapid detection of multiple commonly abused drugs in human urine. The detectable drugs and their cutoff concentrations are listed below:

    AbbreviationAnalyteCalibratorCut-off (ng/mL)
    6AM6-Acetylmorphine6-Monoacetylmorphine10
    AMPAmphetamined-Amphetamine500 or 1000
    BARSecobarbitalSecobarbital300
    BUPBuprenorphineBuprenorphine10
    BZOOxazepamOxazepam300
    COCCocaineBenzoylecgonine150 or 300
    EDDPEDDP2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine300
    FENFentanylFentanyl1
    FENNorfentanylNorfentanyl5
    MDMAEcstasyd,l-Methylenedioxymethamphetamine500
    METMethamphetamined-Methamphetamine500 or 1000
    MORMorphineMorphine300
    MTDMethadoned/l-Methadone300
    OPIOpiatesMorphine2000
    OXYOxycodoneOxycodone100
    PCPPhencyclidinePhencyclidine25
    PPXPropoxyphenePropoxyphene300
    TCANortriptylineNortriptyline1000
    THCMarijuana11-nor-Δ9-THC-9 COOH20 or 50

    The single or multi-test cup can include any combination of the analytes listed above, with and without on-board adulteration tests. However, only one cut-off concentration can be included per analyte per device.

    This in vitro diagnostic device provides only a preliminary test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical test result. GC/MS or LC/MS is the preferred confirmatory method.

    The VINScreen Urine Drug Home Test Cup is lateral flow immunoassays for rapid detection of multiple commonly abused drugs in human urine. The detectable drugs and their cutoff concentrations are listed below:

    DrugCut-off (ng/mL)Abbreviation
    6-monoacetylmorphine106AM
    Amphetamine500 / 1000AMP
    Secobarbital300BAR
    Buprenorphine10BUP
    Oxazepam300BZO
    Cocaine150 / 300COC
    EDDP300EDDP
    Fentanyl1FEN
    Norfentanyl5FEN
    Ecstasy500MDMA
    Methamphetamine500 / 1000MET
    Methadone300MTD
    Opiates300 / 2000MOR/OPI
    Oxycodone100OXY
    Phencyclidine25PCP
    Propoxyphene300PPX
    Nortriptyline1000TCA
    Marijuana20 / 50THC

    The single or multi-test cup can include any combination of the analytes listed above, with and without on-board adulteration tests. However, only one cut-off concentration can be included per analyte per device.

    This device provides only a preliminary test result. A more specific alternate method must be used in order to obtain a confirmed analytical test result. GC/MS or LC/MS is the preferred confirmatory method.

    Device Description

    VINScreen Urine Drug Test Cup and VINScreen Urine Drug Home Test Cup are immunochromatographic assays that use a lateral flow system for the qualitative detection of single or multiple drugs in human urine.

    The devices are a cup format. Each test device is sealed with sachets of desiccant in an aluminum pouch. The device is in a ready-to-use format and no longer requires assembly before use.