Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K252550
    Date Cleared
    2025-11-18

    (97 days)

    Product Code
    Regulation Number
    862.3100
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    SAFElife™ T-Cup Multi-Drug Urine Test Cup:

    SAFElife™ T-Cup Multi-Drug Urine Test Cup is a competitive binding, lateral flow immunochromatographic assay for qualitative and simultaneous detection of 6-Monoacetylmorphine, Amphetamine, Buprenorphine, Secobarbital, Oxazepam, Cocaine, 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP), Fentanyl, Methamphetamine, Methylenedioxymethamphetamine, Morphine, Opiates, Methadone, Norfentanyl, Oxycodone, Phencyclidine, Propoxyphene, Nortriptyline, Cannabinoids and Tramadol in human urine at the cutoff concentrations of:

    Drug (Identifier)Cutoff Level
    6-Monoacetylmorphine (6-MAM)10 ng/mL
    Amphetamine (AMP)1000 ng/mL or 500 ng/mL
    Buprenorphine (BUP)10 ng/mL
    Secobarbital (BAR)300 ng/mL
    Oxazepam (BZO)300 ng/mL
    Cocaine (COC)300 ng/mL or 150 ng/mL
    2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP)300 ng/mL
    Fentanyl (FTY)1 ng/mL
    Methamphetamine (MET/mAMP)1000 ng/mL or 500 ng/mL
    Methylenedioxymethamphetamine (MDMA)500 ng/mL
    Morphine (MOP)300 ng/mL
    Opiates (OPI)2000 ng/mL
    Methadone (MTD)300 ng/mL
    Norfentanyl (NFTY)5 ng/mL
    Oxycodone (OXY)100 ng/mL
    Phencyclidine (PCP)25 ng/mL
    Propoxyphene (PPX)300 ng/mL
    Nortriptyline (TCA)1000 ng/mL
    Cannabinoids (THC)50 ng/mL or 20 ng/mL
    Tramadol (TRA)100 ng/mL

    SAFElife™ T-Cup Multi-Drug Urine Test Cup offers any combinations from 1 to 19 drugs of abuse tests but only one cutoff concentration under same drug condition will be included per device. It is for in vitro diagnostic use.

    The tests may yield positive results for the prescription drugs Buprenorphine, Fentanyl, Nortriptyline, Oxazepam, Secobarbital, Oxycodone and Tramadol when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result.

    The tests provide only preliminary results. To obtain a confirmed analytical result, a more specific alternate chemical method must be used. Gas Chromatography/Mass Spectrometry (GC/MS) or Liquid Chromatography/Mass Spectrometry (LC/MS) is the recommended confirmatory method.

    SAFElife™ T-Cup Multi-Drug Urine Test Cup Dx:

    SAFElife™ T-Cup Multi-Drug Urine Test Cup Dx is a competitive binding, lateral flow immunochromatographic assay for qualitative and simultaneous detection of 6-Monoacetylmorphine, Amphetamine, Secobarbital, Buprenorphine, Oxazepam, Cocaine, 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP), Fentanyl, Methylenedioxymethamphetamine, Methamphetamine, Morphine, Opiates, Methadone, Norfentanyl, Oxycodone, Phencyclidine, Propoxyphene, Nortriptyline, Cannabinoids and Tramadol in human urine with below cutoff
    concentrations and approximate detection time:

    Drug (Identifier)CalibratorCut-off Level
    6-Monoacetylmorphine (6-MAM)6-Monoacetylmorphine10 ng/mL
    Amphetamine (AMP500)d-Amphetamine500 ng/mL
    Amphetamine (AMP1000)d-Amphetamine1000 ng/mL
    Secobarbital (BAR)Secobarbital300 ng/mL
    Buprenorphine (BUP)Buprenorphine10 ng/mL
    Oxazepam (BZO)Oxazepam300 ng/mL
    Cocaine (COC150)Benzoylecgonine150 ng/mL
    Cocaine (COC300)Benzoylecgonine300 ng/mL
    2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP)2-ethylidene-1,5-dimethyl-3,3-diphenyl-pyrrolidine300 ng/mL
    Fentanyl (FTY)Fentanyl1 ng/mL
    Methylenedioxymethamphetamine (MDMA)3,4-Methylenedioxymethamphetamine500 ng/mL
    Methamphetamine (MET500/mAMP500)D(+)-Methamphetamine500 ng/mL
    Methamphetamine (MET1000/mAMP1000)D(+)-Methamphetamine1000 ng/mL
    Morphine (MOP300)Morphine300 ng/mL
    Opiates (OPI2000)Morphine2000 ng/mL
    Methadone (MTD)Methadone300 ng/mL
    Norfentanyl (NFTY)Norfentanyl5 ng/mL
    Oxycodone (OXY)Oxycodone100 ng/mL
    Phencyclidine (PCP)Phencyclidine25 ng/mL
    Propoxyphene (PPX)d-Propoxyphene300 ng/mL
    Nortriptyline (TCA)Nortriptyline1000 ng/mL
    Cannabinoids (THC20)11-nor-Δ9-THC-9-COOH20 ng/mL
    Cannabinoids (THC50)11-nor-Δ9-THC-9-COOH50 ng/mL
    Tramadol (TRA)Tramadol100 ng/mL

    SAFElife™ T-Cup Multi-Drug Urine Test Cup Dx offers any combinations from 1 to 19 drugs of abuse tests with or without on-board adulteration/specimen validity test (SVT) but only one cutoff concentration under same drug condition will be included per device. It is for in vitro diagnostic use.

    The tests may yield positive results for the prescription drugs Buprenorphine, Fentanyl, Nortriptyline, Oxazepam, Secobarbital, Oxycodone and Tramadol when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result.

    The tests provide only preliminary results. To obtain a confirmed analytical result, a more specific alternate chemical method must be used. Gas Chromatography/Mass Spectrometry (GC/MS) or Liquid Chromatography/Mass Spectrometry (LC/MS) is the recommended confirmatory method.

    Device Description

    SAFElife™ T-Cup Multi-Drug Urine Test Cup and SAFElife™ T-Cup Multi-Drug Urine Test Cup Dx are immunochromatographic assays that use a lateral flow system for the qualitative detection of Amphetamine, Buprenorphine, Secobarbital, Oxazepam, Cocaine, 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine, Methamphetamine, Methylenedioxymethamphetamine, Morphine, Opiates, Methadone, Oxycodone, Phencyclidine, Propoxyphene, Nortriptyline, Cannabinoids, 6-Monoacetylmorphine, Fentanyl, Norfentanyl, and Tramadol in human urine. Each SAFElife™ T-Cup Multi-Drug Urine Test Cup or SAFElife™ T-Cup Multi-Drug Urine Test Cup Dx device consists of a test cup and a package insert. Each test cup is sealed with one sachet of desiccant in an aluminum pouch. SAFElife™ T-Cup Multi-Drug Urine Test Cup and SAFElife™ T-Cup Multi-Drug Urine Test Cup Dx have a general identical design. The SAFElife™ T-Cup Multi-Drug Urine Test Cup is for over-the-counter (OTC) use and the SAFElife™ T-Cup Multi-Drug Urine Test Cup Dx is for professional use.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1