Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K252490

    Validate with FDA (Live)

    Date Cleared
    2026-02-12

    (188 days)

    Product Code
    Regulation Number
    878.4850
    Age Range
    0 - 3
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Heel Stick Safety Lancet is intended for the collection of capillary blood from the heel of newborn, preemie, and toddler. The lancet has equipped with safety protection features.

    Device Description

    The heel stick safety lancet is offered in various depth size (0.65mm, 0.85mm, 1.00mm, 1.50mm,1.8mm) and width size (1.50 mm, 1.75mm, 2.50 mm, 2.80 mm, 3.0 mm).

    It consists of blade base, blade, which is welded with the blade base, spring, top head, bottom, and pushing button. The top head and bottom are intended to provide physical protection to the blade, and PET blister along with the tyvek paper cover the primary sterile barrier system for the device. The product is individually primarily packaged and sterilized to a SAL of 10-6 by radiation sterilization. It is intended for single use only. The shelf-life of the product is 5 years.

    AI/ML Overview

    N/A