Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K251520

    Validate with FDA (Live)

    Date Cleared
    2025-07-09

    (54 days)

    Product Code
    Regulation Number
    892.1650
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Cios Alpha is a mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications. Clinical applications may include, but are not limited to: interventional fluoroscopic, gastro-intestinal, endoscopic, urologic, pain management, orthopedic, neurologic, vascular, cardiac, critical care, and emergency room procedures. The patient population may include pediatric patients.

    The Cios Flow is a mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications. Clinical applications may include, but are not limited to: interventional fluoroscopic, gastro-intestinal, endoscopic, urologic, pain management, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures. The patient population may include pediatric patients.

    Device Description

    The Cios Alpha and Cios Flow (VA31A) mobile fluoroscopic C-arm X-ray System is designed for the surgical environment. The Cios Alpha and Cios Flow provide comprehensive image acquisition modes to support orthopedic and vascular procedures. The system consists of two major components:

    a) The C-arm with X-ray source on one side and the flat panel detector on the opposite side. The C-arm can be angulated in both planes and lifted vertically, shifted to the side, and moved forward/backward by an operator.

    b) The second unit is the image display station with a movable trolley for the image processing and storage system, image display, and documentation. Both units are connected with a cable.

    The main unit is connected to the main power outlet, and the trolley is connected to a data network.

    The following modifications were made to the predicate device Cios Alpha and Cios Flow. Siemens Medical Solutions USA, Inc. submits this Bundled Traditional 510k to request clearance for Subject Devices Cios Alpha and Cios Flow (VA31A) for the following device modifications made to the Predicates Device Cios Alpha and Cios Flow (VA30).

    This 510k submission, Subject Devices "Cios Alpha" and "Cios Flow" with software version VA31A, will support the following categories of modifications made to the Subject Devices in comparison to the Predicate Devices:

    1. Software updated from VA30 to VA31A to support the following software features: A. Updated InstantLink with Extended NXS Interface
    2. Updated Collimator
    3. New optional flat detector Trixell Pixium 3131SOD with IGZO (Indium Gallium Zinc Oxide) technology
    4. Updated FLC imaging system with new PC hardware Updated the High Performance Graphic Card on the Apphost PC
    5. Updated Eaton UPS 5P 850i G2 as successor of UPS 5P 850i due to obsolescense
    6. The Cios Alpha is also known as "Cios Alpha.neo" The Cios Flow is also known as Cios Flow.neo
    AI/ML Overview

    The provided 510(k) clearance letter details modifications to an existing fluoroscopic X-ray system, Cios Alpha and Cios Flow, specifically focusing on software updates and hardware changes (e.g., a new flat detector).

    However, the provided text does not contain explicit acceptance criteria tables for performance metrics (such as image quality, diagnostic accuracy, sensitivity, specificity, or AUC) or the results of a statistically powered, pre-specified study proving the device meets these criteria in a comparative effectiveness setting (e.g., MRMC study).

    The document primarily focuses on bench testing, software validation, and compliance with recognized standards to demonstrate the substantial equivalence of the modified device to its predicate. It states that "All test results met all acceptance criteria" for software modifications and that a "Clinical Cadaver Report" was conducted to assess the non-inferiority of a new flat panel detector's subjective image quality. This suggests acceptance criteria were established internally for these tests, but they are not detailed in the provided document.

    Therefore, many of the requested details about acceptance criteria, study design, and performance metrics for clinical effectiveness are not present in this 510(k) clearance letter summary. The document's purpose is to justify substantial equivalence based on safety, hardware/software changes, and compliance with standards, rather than proving enhanced clinical effectiveness through a comparative study.

    Here's an attempt to answer based on the available information, noting what is not provided:


    Acceptance Criteria and Device Performance

    No explicit quantitative acceptance criteria table for clinical performance (e.g., diagnostic accuracy metrics like sensitivity, specificity, AUC) is provided in the document. The document discusses "acceptance criteria" in the context of:

    • Software Validation: "The testing results show that all the software specifications have met the acceptance criteria." (Page 14)
    • Non-clinical Testing: "All test results met all acceptance criteria." (Page 10)
    • Clinical Cadaver Report (Subjective Image Quality): The IGZO detector was considered "non-inferior (equal or better) concerning the subjective image quality for four anatomical regions that have been investigated in the ortho-trauma setting." (Page 14) This implies a qualitative acceptance criterion of non-inferiority for subjective image quality, but no numerical thresholds are given.

    Since no specific performance metrics with numerical acceptance criteria are provided for clinical use, a table demonstrating reported device performance against such criteria cannot be created from this text. The document refers broadly to testing results meeting "acceptance criteria" but does not define them publicly.

    Study Details Proving Device Meets Acceptance Criteria

    The primary "study" mentioned for clinical relevance is a Clinical Cadaver Report.

    1. Sample Size and Data Provenance:

      • Test Set Sample Size: Not specified for the Clinical Cadaver Report.
      • Data Provenance: The study was a "Clinical Cadaver Report." This implies an experimental, non-human, pre-clinical study. The country of origin is not specified but given the manufacturing site in Germany, it's possible the testing was conducted there or at Siemens facilities elsewhere. It is inherently prospective as it's a pre-market development activity.
    2. Number of Experts and Qualifications:

      • Number of Experts: Not specified.
      • Qualifications: Not specified.
    3. Adjudication Method:

      • Adjudication Method: Not specified. Given it was a "subjective image quality" assessment, it would likely involve multiple readers, but the method (e.g., 2+1, 3+1) is not disclosed.
    4. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

      • Was an MRMC study done? The document does not indicate that a formal MRMC comparative effectiveness study was done to show human readers improve with AI vs. without AI assistance. The "Clinical Cadaver Report" focused on the subjective image quality of the new detector, not human performance with AI. The device described primarily appears to be an imaging system, not an AI-assisted diagnostic tool that would typically undergo MRMC studies for improved human interpretation.
    5. Standalone Performance:

      • Was a standalone (algorithm only without human-in-the-loop performance) done? Not explicitly stated in the context of clinical performance. The "software functional, verification, and System validation testing" (Page 11) and "software validation data" (Page 14) refer to the algorithm's internal performance against specifications, not its standalone diagnostic accuracy on clinical images.
    6. Type of Ground Truth Used:

      • Ground Truth for Clinical Cadaver Report: In the context of "subjective image quality," the "ground truth" would be the consensus assessment of the evaluating experts regarding the quality of the images generated by the new IGZO detector compared to the a-Si detector. It is not pathology or outcomes data.
    7. Training Set (if applicable for AI/Software components):

      • Sample Size for Training Set: The document does not mention an AI component that would require a distinct "training set" in the common understanding of machine learning. The "software" referred to is control software for the X-ray system, not a diagnostic AI algorithm.
    8. Ground Truth for Training Set:

      • How ground truth was established for training set: Not applicable, as there's no indication of machine learning model training. The software modifications are described as updates to system control, interfaces, and hardware support.

    In summary: The provided 510(k) clearance letter demonstrates that the modified Cios Alpha and Cios Flow systems meet regulatory requirements for substantial equivalence, primarily through non-clinical testing, compliance with safety standards, and software validation against internal acceptance criteria. A "Clinical Cadaver Report" assessed the subjective image quality of a new detector, finding it non-inferior. However, the document does not contain the specific details of clinical performance acceptance criteria, sample sizes for such studies, or a multi-reader comparative effectiveness study as would be seen for AI-enabled diagnostic tools.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1