510(k) Data Aggregation

    K Number
    K250861

    Validate with FDA (Live)

    Date Cleared
    2025-10-23

    (216 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs, Fentanyl Citrate, Simulated Gastric Acid and Fentanyl in Simulated Gastric Acid (Black) is a non-sterile disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. It is also tested to be used against Chemotherapy Drugs and Fentanyl Citrate, Simulated Gastric Acid and Fentanyl in Simulated Gastric Acid. The gloves were tested for use with chemotherapy drugs and fentanyl citrate as per ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs.

    Chemotherapy DrugConcentrationMinimum Breakthrough Detection Time in Minutes
    Azacytidine(25.0 mg/ml)>240
    Carmustine(3.3 mg/ml)25.0
    Cisplatin(1.0 mg/ml)>240
    Cyclophosphamide(20.0 mg/ml)>240
    Carboplatin(10.0 mg/ml)>240
    Dacarbazine(10.0 mg/ml)>240
    Docetaxel(10.0 mg/ml)>240
    Doxorubicin HCl(2.0 mg/ml)>240
    Epirubicin(2.0 mg/ml)>240
    Etoposide(20.0 mg/ml)>240
    Fluorouracil(50.0 mg/ml)>240
    Gemcitabine(38.0 mg/ml)>240
    Ifosfamide(50 mg/ml)>240
    Irinotecan(20.0 mg/ml)>240
    Methotrexate(25.0 mg/ml)>240
    Mitomycin C(0.5 mg/ml)>240
    Mitoxantrone(2.0 mg/ml)>240
    Oncovin(1.0mg/ml)>240
    Oxaliplatin(5.0 mg/ml)>240
    Paclitaxel(6.0 mg/ml)>240
    Thiotepa(10.0 mg/ml)55.7
    Vinorelbine(10.0 mg/ml)>240
    Vincristine Sulfate(1.0 mg/ml)>240

    Testing showed a minimum breakthrough time of 25.0 minutes with Carmustine and 55.7 minutes with Thiotepa
    Warning: Do not use with Carmustine

    Fentanyl CitrateConcentrationMinimum Breakthrough Detection Time in Minutes
    Fentanyl Citrate Injection100 mcg/2ml>240
    Fentanyl Citrate mixed with Simulated Gastric Acid at a 50:50 ratio>240
    Device Description

    Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs, Fentanyl Citrate, Simulated Gastric Acid and Fentanyl in Simulated Gastric Acid (Black) is a disposable single-use, non-sterile, Black and powder-free ambidextrous glove made from nitrile butadiene rubber.