    K250847

    Validate with FDA (Live)

    Device Name
    VaporShield
    Date Cleared
    2025-12-12

    (267 days)

    Product Code
    Regulation Number
    880.5860
    Age Range
    12 - 120
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VaporShield™ is a sterile, single-use, closed system transfer device (CSTD), incorporating a preassembled Drug Vial holder (DVH) and Syringe (1mL, 22GA). The preassembled device is intended for medication draw from a drug vial only. After medication draw, the DVH is removed from the syringe, and the syringe is used for direct injection of hazardous and non-hazardous medication via subcutaneous or intramuscular administration. The syringe incorporates a safety mechanism which automatically retracts the needle inside the device for drug containment, and prevention of accidental needlesticks and device reuse. The VaporShield mechanically prevents drug transfer external to the device to minimize individual and environmental drug exposure. The VaporShield is intended for use by medical professionals who prepare/administer injection(s) to adults and adolescents in healthcare facilities. The device is not intended for neonates (<28 days old), infants (< 2 yrs old), or children (> 2 yrs. old – 12 yrs. old). The VaporShield Device is not intended for Compounding, Reconstitution, Phlebotomy, Intravenous injection, Intraperitoneal injection, or Intrathecal injection.

    Device Description

    The VaporShield closed system transfer device (CSTD) is sterile, for single use, and is intended for direct injection of hazardous or nonhazardous drugs. A single configuration includes a 1 mL syringe, 22G cannula, and is pre-assembled with a Drug Vial Holder (DVH). Preassembly of the syringe and drug vial holder components are interlocked via a flex tab snap-fit connection. During drug draw, the DVH ensures a stable connection between the drug vial and syringe. The VaporShield has three (3) settings, two for needle length, and one (1) for safety. The DVH can be removed from the syringe when the syringe is placed in locked setting.

    The VaporShield includes a safety mechanism that is manually activated by full plunger depression. When activated, the safety mechanism retracts the needle inside the plunger rod for containment, preventing accidental needlestick, and disallowing device reuse.

    Syringe is intended for parenteral injection (subcutaneous / intramuscular).