    Intended Use

    SureFine Pen Needle is intended for use with a pen injector for the subcutaneous injection of insulin.

    Device Description

    SureFine Pen Needle is designed for use with a pen injector for the subcutaneous injection of insulin. The pen needle consists of a needle, hub, and shield assembly. Blister paper covers primary container. The primary container maintains sterility of the needle because primary container covers the hub and needle cap with blister paper sealed on the opening hole of primary container.

    The needle hub can be connected screwed onto the pen. The needle shield is intended to provide physical protection to the needle tube.

    SureFine Pen Needle is offered in various gauges sizes (29G, 30G, 31G, 32G) and Lengths (4mm, 6mm, 8mm, 12.7mm). These are sterile (EO gas sterilization), non-toxic, and non-pyrogenic. The pen needles are disposable, single use devices.

    AI/ML Overview

    The SureFine Pen Needle device, K250658, is a Class II hypodermic single lumen needle intended for use with a pen injector for the subcutaneous injection of insulin. The information provided is from an FDA 510(k) Clearance Letter.

    1. Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the SureFine Pen Needle are based on compliance with international standards ISO 9626, ISO 11608-2, and ISO 10993-1, 10993-4, 10993-5, 10993-7, 10993-10, and 10993-11 for biocompatibility and sterilization validation according to ISO 11135.

    Test ItemAcceptance Criteria (from referenced ISO standards)Reported Device Performance
    ISO 9626 (Needle Characteristics)All criteria specified in ISO 9626 for hypodermic needlesPass
    MaterialsConformance to specified material propertiesPass
    CleanlinessConformance to cleanliness standardsPass
    Limits for acidity and alkalinityConformance to pH limitsPass
    Size designationAccurate representation of gauge and lengthPass
    DimensionConformance to specified dimensional tolerancesPass
    StiffnessConformance to stiffness requirementsPass
    Resistance to breakageResistance to breakage under specified conditionsPass
    Resistance to corrosionResistance to corrosion under specified conditionsPass
    ISO 11608-2 (Needle-Based Injection Systems)All criteria specified in ISO 11608-2 for pen needlesPass
    DimensionConformance to specified dimensional tolerancesPass
    Flow rateConformance to prescribed flow ratePass
    Bond between hub and needle tubeSecure bond between componentsPass
    Needle point freedom from defects lubricationAbsence of defects and proper lubricationPass
    Dislocation of measuring point at patientMinimal dislocation to ensure accurate deliveryPass
    Compatibility of needles and injector systemFunctional compatibility with pen injectorsPass
    Biocompatibility (ISO 10993 Series)All evaluation acceptance criteriaMet
    Cytotoxicity testSatisfy requirements of ISO 10993-5 (in vitro cytotoxicity)Pass
    Hemolysis testSatisfy requirements of ISO 10993-4 (hemolytic properties)Pass
    Intracutaneous reactivity testSatisfy requirements of ISO 10993-10 (intracutaneous reactivity)Pass
    Skin sensitization testSatisfy requirements of ISO 10993-10 (skin sensitization)Pass
    Acute systemic toxicity testSatisfy requirements of ISO 10993-11 (acute systemic toxicity)Pass
    Pyrogen TestSatisfy requirements of ISO 10993-11 (material-mediated pyrogens)Pass
    Sterilization & Shelf-lifeMax levels of EO/ECH within ISO 10993-7 limits; adequate mitigation of bacterial endotoxins; 5-year shelf-life stability.Conforms to ISO 11135 validated half-cycle method. EOx/ECH levels meet ISO 10993-7. LAL testing performed. Supports 5-year shelf-life.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the specific sample sizes for each test mentioned (ISO 9626, ISO 11608-2, Biocompatibility). However, it indicates that "Bench tests for the device's performance were conducted" and "Biocompatibility of the SureFine Pen Needle was evaluated". The testing was conducted by Shina Med Corporation, a company based in the Republic of Korea. The nature of these tests (bench testing, lab-based biocompatibility) suggests they are prospective tests performed on manufactured samples of the device. The data provenance is primarily from internal testing and validation by the manufacturer, following international standards.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not specify the number or qualifications of experts used to establish ground truth for the "test set" in the context of device performance. The testing performed is against established international standards (ISO 9626, ISO 11608-2, ISO 10993 series), which themselves represent a form of "ground truth" for device safety and performance criteria. The evaluation results are reported as "Pass" or "Met," implying adherence to these predefined standards rather than an expert panel establishing a de novo ground truth for each specific test.

    4. Adjudication Method for the Test Set

    No information is provided regarding an adjudication method (such as 2+1, 3+1, none). The evaluation is based on objective measurements and established criteria set forth by the ISO standards. For instance, a flow rate test would have a pass/fail threshold, not require expert adjudication in the traditional sense.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. The document describes bench tests and biocompatibility evaluations for a medical device (pen needle), not a diagnostic or AI-assisted system that would typically undergo an MRMC comparative effectiveness study involving human readers. Therefore, there is no mention of an effect size of human readers improving with or without AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    No. This device is a physical medical instrument (a pen needle), not a software algorithm or AI system. Therefore, the concept of "standalone performance" in the context of algorithms is not applicable here.

    7. The Type of Ground Truth Used

    The ground truth used for demonstrating the device's performance is adherence to established international standards and regulatory requirements:

    • ISO 9626: Specifies requirements for stainless steel needle tubing for the manufacture of hypodermic needles.
    • ISO 11608-2: Specifies requirements and test methods for single-use needles used with pen-injectors.
    • ISO 10993 series: Biological evaluation of medical devices, covering various aspects of biocompatibility (cytotoxicity, sensitization, irritation, systemic toxicity, hemocompatibility, pyrogenicity).
    • ISO 11135: Specifies requirements for the development, validation, and routine control of an ethylene oxide sterilization process for medical devices.
    • ISO 10993-7: EtO residual limits.
    • LAL testing: To ensure absence of bacterial endotoxins.

    8. The Sample Size for the Training Set

    Not applicable. This is a physical medical device, not a machine learning or AI algorithm, so there is no "training set."

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for a physical medical device.