510(k) Data Aggregation

    K Number
    K250578

    Validate with FDA (Live)

    Date Cleared
    2025-04-01

    (33 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. These gloves were tested for use with Chemotherapy Drugs and Fentanyl Citrate as per ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

    Device Description

    Non-Sterile Nitrile Patient Examination Glove is made from synthetic rubber latex. It is single use and powder-free class I Nitrile Patient Examination Gloves which coated by on-line polymer with mild on-line chlorination process. These processes modify the surface characteristics and causes it to remain tack-free without the use of any dusting or donning powder. The available sizes are X-Small, Small, Medium, Large and X-Large. The Nitrile Examination Gloves Powder Free (Blue & Black) have been tested for Chemotherapy Drugs. The subject devices are identical in design and formulation to the predicate gloves of K230121.

    AI/ML Overview

    The provided FDA 510(k) clearance letter pertains to Nitrile Examination Gloves Powder Free Tested for Use With Chemotherapy Drugs & Fentanyl Citrate (Blue & Black), a medical device. This type of device is classified as a Class I medical device and does not involve AI/ML technology. Therefore, many of the requested categories related to AI/ML device studies (e.g., sample sizes for test/training sets, ground truth establishment, MRMC studies, standalone performance, expert qualifications, adjudication methods, effect sizes) are not applicable.

    The acceptance criteria and device performance are primarily evaluated based on material properties, chemical permeation resistance, and biocompatibility, using established ASTM and ISO standards for examination gloves.

    1. Table of Acceptance Criteria and Reported Device Performance

    The device, "NITRILE EXAMINATION GLOVES POWDER FREE TESTED FOR USE WITH CHEMOTHERAPY DRUGS & FENTANYL CITRATE (BLUE & BLACK)", was tested against various physical, chemical, and biological criteria.

    Test CategoryCharacteristic / StandardAcceptance CriteriaReported Device Performance (Result)
    PreambleProduct CodeLZA, LZC, QDO, OPJLZA, LZC, QDO, OPJ
    Intended UseTo prevent contamination between patient and examinerMeets stated intended use
    DesignPowder Free, Non-Sterile, Ambidextrous, Beaded CuffPowder Free, Non-Sterile, Ambidextrous, Beaded Cuff
    ConstructionAmbidextrous, Polymer Coated or Chlorinated, Powder Free NitrileAmbidextrous, Polymer Coated or Chlorinated, Powder Free Nitrile
    Color DescriptionBlue & BlackBlue & Black
    MaterialNitrileNitrile
    Single UseYesYes
    PackagingPacked in Dispenser BoxesPacked in Dispenser Boxes
    SterilityNon-SterileNon-Sterile
    Freedom From HolesASTM D6319-19, ASTM D5151-19Meet requirement inspection level G-1, AQL 2.5Pass
    DimensionASTM D6319-19Length (XS), min. 220mmMeet 220mm min
    Length (S, M, L, XL), min. 230mmMeet 230mm min
    Thickness (palm/finger), min. 0.05mmMeet 0.05mm min
    Width (XS), 70 ± 10 mmMeet 70 ± 10 mm
    Width (S), 80 ± 10 mmMeet 80 ± 10 mm
    Width (M), 95 ± 10 mmMeet 95 ± 10 mm
    Width (L), 110 ± 10 mmMeet 110 ± 10 mm
    Width (XL), 120 ± 10 mmMeet 120 ± 10 mm
    Physical Properties (Before Ageing)ASTM D6319-19Tensile Strength, 14 MPa min.Meet 14MPa min.
    Ultimate Elongation, 500% min.Meet 500% min.
    Physical Properties (After Accelerated Aging)ASTM D6319-19Tensile Strength, 14 MPa min.Meet 14MPa min.
    Ultimate Elongation, 400% min.Meet 400% min.
    Residual Powder ContentASTM D6319-19, ASTM D6124-06Not more than 2 mg per gloveMeet 2mg/glove max.
    Biocompatibility - Primary Skin IrritationISO 10993-10Not a primary skin irritantPass (did not cause an irritant response)
    Biocompatibility - Skin SensitizationISO 10993-10Not a contact sensitizerPass (did not produce a skin sensitization effect)
    Biocompatibility - In Vitro CytotoxicityISO 10993-5:2009No adverse biological reactionModerate cytotoxicity reactivity at 6.0 cm²/mL extract concentrations and no cytotoxicity reactivity at 3.0 cm²/mL extract concentrations.
    Biocompatibility - Acute Systemic ToxicityISO 10993-11No adverse biological reactionPass (no adverse biological reaction)
    Chemotherapy Drug PermeationASTM D6978-05 (Reapproved 2019)Varies by drug (Breakthrough Detection Time)See detailed table below

    Detailed Chemotherapy Drug and Fentanyl Citrate Permeation Performance:

    Chemotherapy Drugs and Fentanyl CitrateConcentrationMinimum Breakthrough Detection Time (minutes) - Blue GloveMinimum Breakthrough Detection Time (minutes) - Black Glove
    *Carmustine (BCNU)3.3 mg/ml22.621.8
    Cisplatin1.0 mg/ml>240 min>240 min
    Cyclophosphamide (Cytoxan)20.0 mg/ml>240 min>240 min
    Dacarbazine10.0 mg/ml>240 min>240 min
    Doxorubicin HCL2.0 mg/ml>240 min>240 min
    Etoposide20.0 mg/ml>240 min>240 min
    Fluorouracil50.0 mg/ml>240 min>240 min
    Ifosfamide50.0 mg/ml>240 min>240 min
    Mitoxantrone2 mg/ml>240 min>240 min
    Paclitaxel6.0 mg/ml>240 min>240 min
    *Thio Tepa10.0 mg/ml43.917.7
    Vincristine Sulfate1.0 mg/ml>240 min>240 min
    Methotrexate25.0 mg/ml>240 min>240 min
    Mitomycin C.0.5 mg/ml>240 min>240 min
    Fentanyl Citrate100mcg/2ml>240 min>240 min
    *Warning: Do not use with Carmustine and Thiotepa due to extremely low permeation times.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the exact sample sizes for each test (e.g., number of gloves tested for permeation, holes, dimensions, etc.) but refers to adherence to the methodologies outlined in the specified ASTM and ISO standards. These standards typically define minimum sample sizes for material testing.

    • Data Provenance: The document states the applicant is from Malaysia (Careglove Global Sdn. Bhd.). The testing was conducted according to international standards (ASTM, ISO), implying the tests were performed in a controlled laboratory environment, likely by the manufacturer or a contracted lab. The data is non-clinical and derived from prospective testing of manufactured glove samples.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. This device is not an AI/ML device that requires expert review for "ground truth" establishment in the context of diagnostic or interpretive tasks. The "ground truth" for material properties is established by the physical and chemical measurements themselves, according to standardized protocols and certified laboratory equipment.

    4. Adjudication Method for the Test Set

    Not applicable. This device is not an AI/ML device that requires adjudication for interpretations. Test results are objective measurements against predefined criteria.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    Not applicable. MRMC studies are relevant for diagnostic or interpretive AI/ML devices where human readers' performance is evaluated with and without AI assistance. This is a physical and chemical barrier device.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study Was Done

    Not applicable. This device is a passive physical barrier and does not involve any algorithms or software.

    7. The Type of Ground Truth Used

    The "ground truth" for this device is based on objective measurements of physical properties (e.g., length, thickness, tensile strength, elongation), chemical resistance (breakthrough detection time), and biological reactivity (biocompatibility tests), as defined by internationally recognized standards (ASTM, ISO). There is no subjective human "ground truth" in this context.

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI/ML device that undergoes "training."

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. This is not an AI/ML device. The manufacturing process of the gloves is designed to meet the specified standards, and quality control ensures consistency, but this is not "training" in the context of AI/ML.