Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K243993

    Validate with FDA (Live)

    Date Cleared
    2025-03-13

    (77 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Photon therapy apparatus use of the red, blue, Yellow and infrared regions of the spectrum is intended to emit energy to treat dermatological conditions.

    The red light (633±8nm wavelength) is generally indicated to treatment of superficial, benign vascular, and pigmented lesions.

    The blue light (417±8nm wavelength) indicated to treat dermatological conditions and specifically indicated to treat moderate inflammatory acne vulgaris.

    The Yellow light (590±8nm wavelength) is generally indicated to treat dermatological conditions and specifically indicated for treatment of periorbital wrinkles and rhytides.

    The infrared light (830±8mm wavelength) is generally use for the temporary relief of minor muscle and joint pain, atthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation where applied.

    Device Description

    The photon therapy apparatus (BL-10) use specific wavelengths of light, produced by LEDs (Light emitting diodes), to manage aesthetic conditions. The device produces light in the red light region of the spectrum (633±8nm), in the blue light regions of the light spectrum (417±8nm), yellow light area (590±8nm) and infrared light region of light spectrum (835±8nm).

    The photon therapy apparatus (BL-10) are equipped with six type of optional irradiators, each of which emits three different colors of light sources.

    AI/ML Overview

    The provided document is a 510(k) summary for the Xemis Medical Technology (Shenzhen) Co., Ltd. "Photon Therapy Apparatus (BL-10)". It describes the device, its intended use, and compares it to predicate devices to demonstrate substantial equivalence.

    However, the document does not contain specific acceptance criteria and detailed study results as would be typically found for a study proving a device meets acceptance criteria, especially in the context of an AI/ML-driven device's performance. The provided text primarily focuses on the device's technical specifications and a comparison to predicate devices, stating that "Non-clinical tests" were performed for electrical safety, EMC, performance, and software verification/validation. It explicitly states "Clinical study: Not applicable."

    Therefore, based on the provided text, I cannot fulfill most of your requested information as it pertains to clinical performance studies, acceptance criteria, test set details (sample size, data provenance, ground truth experts, adjudication), MRMC studies, or standalone algorithm performance, and training set details for an AI/ML device.

    The document describes a photon therapy apparatus, which is a physical device emitting light for dermatological conditions, not an AI/ML-driven medical device that would involve a test set, ground truth interpretation by experts, or MRMC studies.

    If this was a request for information on an AI/ML device, the provided text would be a "reject" or "missing information" response. As it is not an AI/ML device, the traditional performance acceptance criteria for a medical device of this type would typically involve measurements of parameters like wavelength, power output, irradiance, and safety adherence to relevant electrical and photobiological safety standards, which are mentioned generally.

    Here's what I can extract from the document regarding "acceptance criteria" in a general sense, and "study" information as provided:

    1. Table of acceptance criteria and the reported device performance:

    The document doesn't present a formal table of acceptance criteria with corresponding performance results in a standalone "acceptance criteria" section. Instead, the basis for "acceptance" (or rather, substantial equivalence for FDA clearance) is demonstrated through comparison of the proposed device's specifications and performance to predicate devices. The "performance" here refers to meeting technical specifications and safety standards, rather than clinical efficacy outcome measures from a trial.

    Feature/Criterion (Implied Acceptance)Proposed Device Performance (BL-10)Basis for "Acceptance" (Substantial Equivalence)
    WavelengthsRed light: 633±8nm, Blue light: 417±8nm, Yellow light: 590±8nm, Infrared: 830±8nmComparable to predicate devices (e.g., Red: 633±10nm, Blue: 417±10nm, Yellow: 590±10nm, Infrared: 835±15nm for K222751).
    Maximum Power Density (mW/cm²)Red: 80mW/cm², Blue: 100mW/cm², Yellow: 35W/cm² (note discrepancy in unit for Yellow vs. others), Infrared: 50mW/cm²Comparable to predicate devices (e.g., K222751: Red 2096mw/cm², Blue 10120mw/cm², Yellow 5~35mw/cm², Infrared ≤ 70mw/cm²; K190938: Red 115mW/cm², Blue 120mW/cm², IR 70mW/cm²; K200104: Red 100mW/cm², Blue 45mW/cm², Yellow 35W/cm²).
    Standard Doses (Joules)Red: 96J/cm², Blue: 120J/cm², Yellow: 42J/cm², Infrared: 60J/cm²Comparable to predicate devices (e.g., K222751: Red 155J/cm², Blue 144J/cm², Yellow 42J/cm², IR 84J/cm²; K190938: Red 138J/cm², Blue 144J/cm², IR 84J/cm²; K200104: Red 120J/cm², Blue 54J/cm², Yellow 42J/cm²).
    Adjustable Dose RangeRed: 1-144J/cm², Blue: 1-180J/cm², Yellow: 1-63J/cm², Infrared: 3-90J/cm²Comparable to predicate devices' adjustable ranges.
    Effective Irradiation Area (CM²)825cm² and 1375cm²Comparable to predicate devices (e.g., K222751: 900cm² ±10%; K190938: 756cm² and 1008cm²; K200104: 500cm² and 860cm²).
    Safety and Performance StandardsPassed tests according to: ANSI AAMI ES60601-1, IEC 60601-1, IEC 60601-1-2, IEC 60601-2-57, IEC 62471.Adherence to recognized national and international safety standards is an acceptance criterion for medical devices.
    Software Verification & ValidationSystem verification testing presented in this 510(k) demonstrated that all software requirement specifications are met. (Basic Documentation Level of concern).Software functionality and safety are confirmed.
    Intended UsePhoton therapy apparatus use of the red, blue, Yellow and infrared regions of the spectrum is intended to emit energy to treat dermatological conditions (with specific indications for each light color)."The indications for use of the Photon Therapy Apparatus are the same as those for the predicate devices." (Page 10)
    Technological DifferencesNone identified that "raise new or different questions of safety or effectiveness." (Page 10)This is a key criterion for substantial equivalence.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not applicable. This document describes a physical medical device (photon therapy apparatus), not an AI/ML diagnostic or prognostic algorithm that would have a "test set" of patient data. The "tests" mentioned are non-clinical (electrical safety, EMC, performance, software V&V).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable. As above, no clinical "test set" and thus no ground truth established by experts is described for this type of device submission.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. No clinical "test set" and thus no adjudication method is mentioned.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This is not an AI-assisted diagnostic device, so no MRMC study was performed. The document explicitly states "Clinical study: Not applicable."

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is not an algorithm.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    • Not applicable. The "ground truth" for this device's performance would be the physical measurements of its light output and adherence to electrical safety standards, not clinical ground truth from patient data.

    8. The sample size for the training set

    • Not applicable. This is not an AI/ML device that requires a training set.

    9. How the ground truth for the training set was established

    • Not applicable. This is not an AI/ML device that requires a training set.

    In summary, the provided document is a 510(k) notification for a photon therapy device, focusing on demonstrating substantial equivalence to predicate devices based on technical specifications and non-clinical safety/performance testing. It does not involve AI/ML components or clinical performance studies typically associated with the detailed criteria you've outlined.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1