Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K243617

    Validate with FDA (Live)

    Date Cleared
    2025-05-16

    (175 days)

    Product Code
    Regulation Number
    892.1750
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    uCT ATLAS Astound is a computed tomography x-ray system, which is intended to produce cross-sectional images of the whole body by computer reconstruction of x-ray transmission data taken at different angles and planes. uCT ATLAS Astound is applicable to head, whole body, cardiac, and vascular x-ray Computed Tomography.

    uCT ATLAS Astound is intended to be used for low dose CT lung cancer screening for the early detection of lung nodules that may represent cancer. The screening must be performed within the established inclusion criteria of programs / protocols that have been approved and published by either a governmental body or professional medical society.

    *Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.

    uCT ATLAS is a computed tomography x-ray system, which is intended to produce cross-sectional images of the whole body by computer reconstruction of x-ray transmission data taken at different angles and planes. uCT ATLAS is applicable to head, whole body, cardiac, and vascular x-ray Computed Tomography.

    uCT ATLAS has the capability to image a whole organ in a single rotation. Organs include, but not limited to head, heart, liver, kidney, pancreas, joints, etc.

    uCT ATLAS is intended to be used for low dose CT lung cancer screening for the early detection of lung nodules that may represent cancer. The screening must be performed within the established inclusion criteria of programs / protocols that have been approved and published by either a governmental body or professional medical society.

    *Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.

    uWS-CT-Dual Energy Analysis is a post-processing software package that accepts UIH CT images acquired using different tube voltages and/or tube currents of the same anatomical location. The various materials of an anatomical region of interest have different attenuation coefficients, which depend on the used energy. These differences provide information on the chemical composition of the scanned body materials and enable images to be generated at multiple energies within the available spectrum. uWS-CT-Dual Energy Analysis software combines images acquired with low and high energy spectra to visualize this information.

    Device Description

    The uCT ATLAS Astound with uWS-CT Dual Energy Analysis and uCT ATLAS with uWS-CT Dual Energy Analysis includes the same intended use and same indications for use as their recent cleared versions (K231482). The reason for this submission is to support the following additional functions:

    • CardioXphase (optimized)
    • CardioBoost
    • CardioCapture (optimized)
    • AIIR
    • Motion Freeze
    • Ultra EFOV
    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a Computed Tomography X-ray System (uCT ATLAS Astound with uWS-CT-Dual Energy Analysis and uCT ATLAS with uWS-CT-Dual Energy Analysis). The submission focuses on additional software functions beyond what was previously cleared.

    However, the document does not contain specific acceptance criteria, detailed study designs, or quantitative performance data to establish "proof" in the typical sense of a rigorous clinical trial with defined endpoints and statistical significance. Instead, it relies on demonstrating substantial equivalence to existing predicate devices.

    The "acceptance criteria" appear to be implicit in the non-clinical and reader studies, aiming to show that the performance of the new features is "sufficient for diagnosis," "equal or better," or "can improve" compared to a baseline or predicate. No explicit numerical thresholds for metrics like sensitivity, specificity, accuracy, or effect sizes for reader improvement are provided.

    Here's an analysis based on the information provided, highlighting what is present and what is missing concerning acceptance criteria and study details:


    Overview of Device Performance and Acceptance Criteria

    The submission does not explicitly define acceptance criteria in terms of numerical thresholds for performance metrics. Instead, it describes a "bench test" and "reader study" approach to demonstrate that the new functions do not raise new safety and effectiveness concerns and provide an equivalent or improved performance compared to the predicate/reference devices or established techniques.

    The implied "acceptance criteria" are qualitative, such as:

    • "passed the basic general IQ test which satisfied the requirement of IEC 61223-3-5."
    • "showed better LCD comparing with FBP..."
    • "showed better noise comparing with FBP."
    • "showed better spatial resolution comparing with FBP..."
    • "all indicators have met the verification criteria and have passed the verification." (for CardioXphase)
    • "can reduce head motion artifacts." (for Motion Freeze)
    • "can improve the CT number..." (for Ultra EFOV)
    • "images are sufficient for diagnosis and the image quality... is equal or better than..." (for various reader studies)
    • "is helpful for both artifact suppression and clinical diagnosis." (for Motion Freeze reader study)
    • "can improve the accuracy of image CT numbers..." (for Ultra EFOV reader study)
    • "conclude the effectiveness of CardioCapture function for reducing cardiac motion artifacts as expected."

    Table of Acceptance Criteria (Implied) and Reported Device Performance

    Since explicit, quantitative acceptance criteria are not provided, this table will rephrase the reported performance as the observed outcome against the implied objective.

    Software FunctionImplied Acceptance Criteria (Objective)Reported Device Performance
    CardioBoostBench Test: Meet IEC 61223-3-5 requirements; show better LCD, noise, and spatial resolution than FBP; maintain basic general IQ. Reader Study: Images sufficient for diagnosis; image quality equal or better than KARL 3D.Bench Test: Passed basic general IQ test (IEC 61223-3-5 satisfied). Showed better LCD, noise, and spatial resolution compared to FBP at same scanning dose. Reader Study: Confirmed CardioBoost images are sufficient for diagnosis and image quality is equal or better than KARL 3D over all evaluation aspects.
    AIIRBench Test: Meet IEC 61223-3-5 requirements; show better LCD, noise, and spatial resolution than FBP; maintain basic general IQ. Reader Study: Images sufficient for diagnosis; image quality equal or better than FBP.Bench Test: Passed basic general IQ test (IEC 61223-3-5 satisfied). Showed better LCD, noise, and spatial resolution compared to FBP at same scanning dose. Reader Study: Confirmed AIIR images are sufficient for diagnosis and image quality is equal or better than FBP over all evaluation aspects.
    CardioXphaseBench Test (AI module): Quantitative assessment metrics (DICE, Precision, Recall) for heart mask and coronary artery mask extraction meet verification criteria.Bench Test (AI module): All quantitative indicators (DICE, Precision, Recall) for heart mask and coronary artery mask extracted by the new AI module have met the verification criteria and passed verification.
    Motion FreezeBench Test: Demonstrate effectiveness in reducing head motion artifacts. Reader Study: Images helpful for artifact suppression and clinical diagnosis.Bench Test: Showed that Motion Freeze can reduce head motion artifacts. Reader Study: Confirmed Motion Freeze is helpful for both artifact suppression and clinical diagnosis.
    Ultra EFOVBench Test: Demonstrate effectiveness in improving CT value accuracy when scanned object exceeds scan-FOV compared to EFOV. Reader Study: Images confirm improved accuracy of image CT numbers and homogeneity of same tissue when scanned object exceeds scan-FOV.Bench Test: Showed that Ultra EFOV can improve the CT number in cases where the scanned object exceeds the CT field of scan-FOV, compared to EFOV. Reader Study: Confirmed that images with Ultra EFOV can improve the accuracy of image CT numbers and homogeneity of same tissue, in cases where the scanned object exceeds the CT field of view.
    CardioCaptureReader Study: Effectiveness in reducing cardiac motion artifacts as expected, with clear/continuous contours, tolerable motion artifacts, and sufficient diagnostic (>=50%) coronary segments.Reader Study: Concluded the effectiveness of CardioCapture function for reducing cardiac motion artifacts as expected, based on evaluation of clear/continuous contours, tolerable motion artifacts, and number of diagnostic coronary segments (reaching at least 50% of total coronary artery segments). (Specific to AI motion correction in uCT ATLAS).

    Study Details

    1. Sample Size Used for the Test Set and Data Provenance:

      • Test Set Sample Size: The document does not specify the sample sizes (number of cases/studies) used for either the bench tests or the reader studies.
      • Data Provenance: Not explicitly stated (e.g., country of origin, whether retrospective or prospective). The use of "clinical images" implies real patient data, but details are missing.
    2. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

      • Number of Experts: Not specified. The document mentions "readers" (plural) for the reader studies but does not state how many participated.
      • Qualifications of Experts: Not specified. No details are given about their specialty (e.g., cardiologist, radiologist), experience level, or board certification.
    3. Adjudication Method for the Test Set:

      • Adjudication Method: Not specified. For the reader studies, it only states that images "were shown to the readers to perform a five-point scale evaluation" or "5-point scale evaluation." There's no mention of how discrepancies or disagreements among readers were handled or if a consensus ground truth was established by independent experts (e.g., 2+1, 3+1).
    4. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, What was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance:

      • MRMC Study: Reader studies were conducted comparing images reconstructed with the new AI functions (e.g., CardioBoost) to those reconstructed with traditional methods (e.g., KARL 3D, FBP). These appear to be MRMC studies in structure, as multiple readers evaluate multiple cases.
      • Effect Size: No quantitative effect sizes are provided. The results are qualitative: "equal or better," "sufficient for diagnosis," "helpful." There are no reported metrics like AUC improvement, sensitivity/specificity gains, or statistical significance of differences in reader performance with and without AI assistance.
    5. If a Standalone (i.e., Algorithm Only Without Human-in-the-Loop Performance) Was Done:

      • Standalone Performance: The "bench tests" for CardioBoost, AIIR, Motion Freeze, and Ultra EFOV evaluate the algorithms' image quality metrics (IQ, LCD, noise, spatial resolution, CT value accuracy, artifact reduction) independently of human interpretation. For CardioXphase, the evaluation of the AI module's extraction accuracy (DICE, Precision, Recall) is also a standalone assessment. These can be considered standalone performance evaluations for the image reconstruction/processing algorithms.
    6. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.):

      • Ground Truth: For the image quality bench tests (CardioBoost, AIIR, Motion Freeze, Ultra EFOV), the "ground truth" is likely defined by physical phantom measurements and adherence to engineering specifications/standards (e.g., IEC 61223-3-5, CTIQ White Paper, AAPM's report).
      • For CardioXphase, the ground truth for image segmentation accuracy (heart and coronary artery masks) was "annotated results," which typically implies expert manual annotation on imaging data.
      • For the reader studies, the "ground truth" is based on the subjective evaluation of "image quality aspects" by the readers, rather than an objective, clinically validated ground truth for a diagnostic endpoint (e.g., presence/absence of disease confirmed by biopsy or follow-up). The goal was to demonstrate that the image quality generated by the new features is non-inferior or improved for diagnostic purposes.
    7. The Sample Size for the Training Set:

      • Training Set Sample Size: The document mentions "datasets augmentation and deep learning network optimization" for CardioBoost and AIIR, and "introduction of a new deep learning based coronaries detection algorithm" for CardioXphase, and "introduces a deep learning network" for Ultra EFOV. However, the specific size of the training datasets (number of images/cases) is not provided.
    8. How the Ground Truth for the Training Set Was Established:

      • Training Set Ground Truth: Not explicitly stated. For deep learning models, training data ground truth is typically established by expert annotation or labels derived from existing clinical reports or imaging features. Given the context of image reconstruction and enhancement, it likely involves high-quality, potentially expert-annotated, imaging data. For instance, for CardioXphase, the ground truth for training the coronary artery detection algorithm would involve expert-labeled coronary anatomy. For features like CardioBoost and AIIR, which optimize image reconstruction, the ground truth for training might involve pairs of raw data and ideal reconstructed images, or image quality metrics derived from expert evaluations on initial datasets.

    In summary, the 510(k) submission successfully demonstrates "substantial equivalence" based on qualitative assessments and performance relative to known methods. However, for a detailed "proof" with explicit acceptance criteria and quantitative performance metrics, further information beyond what is presented in this FDA clearance letter summary would be needed. This is characteristic of many 510(k) submissions, which often rely on demonstrating safety and effectiveness relative to existing predicates rather than establishing novel clinical efficacy through large-scale, quantitatively defined trials.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1