    Intended Use

    Lancet: The lancet is intended for capillary blood sampling.
    Lancing device: The Lancing Device is used with lancets to draw capillary blood from the fingertip, for testing utilizing small amounts of blood. The Lancing Device is intended to be used by a single patient and should not be shared. (HH-X-T, HH-XV-T, HH-XVI-T, HH-XVII-T, HH-XVIII-T, HH-XIX, HH-XXI-T, HH-XXII- T, HH-XXIII-T, HH-XXIV-T, HH-XXV-T, HH-XXVI-T, HH-XXVIII-T, HH-XXIX-T, HH-XXX-T, HH-XIII-T)
    The Lancing Device is used with lancets to draw capillary blood from the fingertip, palm (at the base of the thumb) or forearm, for testing utilizing small amounts of blood. The Lancing Device is intended to be used by a single patient and should not be shared. (HH-XXVII-T)

    Device Description

    Lancet: The Lancet (use with lancing device) is a single use, sterile medical device, which is designed for use of micro blood sampling puncture to obtain capillary blood samples from the fingertip. The Lancet is composed three components: needle, main body and protective cap. Main body and protective cap is plastic part that enclosed the needle. The Lancet is intended to be single use and the needle tip is sterilized by Radiation. The shelf-life of the product is 5 years.
    Lancing Device: Along with a lancet, the lancing device is used to obtain a capillary blood sample. The body and the active parts of the lancing device are made of ABS, POM, PC and PS Resin. And the spring is made of carbon steel. The service life of the Lancing Device is 5 years or 5000 times of normal use, whichever comes first. For Model HH-XIII-T, the service life of the Lancing Device is 10 years or 5000 times of normal use, whichever comes first. The Lancing Device is provided non-sterile.

    AI/ML Overview

    The provided text is a 510(k) summary for a medical device (Lancet and Lancing Device). This type of submission focuses on demonstrating substantial equivalence to a predicate device through non-clinical testing, rather than extensive clinical studies or AI/software performance evaluations. Therefore, many of the requested details related to "AI performance," "expert ground truth," "MRMC studies," and "training sets" are not applicable to this document.

    However, I can extract the relevant acceptance criteria and study information provided for this medical device based on the document's content.

    Here's a breakdown of the requested information based on the provided FDA 510(k) summary:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't present a formal table of quantitative acceptance criteria and corresponding performance metrics for the device in the format common for AI/software devices. Instead, it generally states that the device "met the performance criteria outlined" and that "all verification and validation tests passed without deviations." The performance evaluation for this device category (lancets and lancing devices) would typically involve mechanical, sterilization, biocompatibility, and functional tests.

    Here's an interpretation of the implied acceptance criteria and reported performance based on the "Non-Clinical Testing" section:

    Acceptance Criteria CategorySpecific Criteria (Implied)Reported Device Performance
    PerformanceDevice meets all established performance specifications (e.g., proper piercing, blood collection, retraction, durability)."The testing confirmed that the lancet and lancing device meet the performance criteria outlined."
    SterilizationSterility Assurance Level (SAL) of $10^{-6}$ for lancets sterilized by radiation."No Change" from predicate, implying adherence to this SAL.
    BiocompatibilityNew material (PE) is biocompatible and non-toxic (as per ISO 10993-1)."The results confirmed that the material is nontoxic and safe for use in its intended application."
    Simulated Clinical UseSafety mechanism meets pre-established criteria (as per FDA Guidance and ISO 23908)."The results demonstrated that the proposed device met the pre-established criteria."
    Durability (Lancing Device)Service life of 5 years or 5000 uses (or 10 years/5000 uses for model HH-XIII-T)."The service life of the Lancing Device is 5 years or 5000 times of normal use, whichever comes first." (Specifically mentions HH-XIII-T).
    Sharp Injury PreventionDevice design prevents sharp injuries after use."Lancet retracted after use to prevent sharp injure" (Comparison table), confirming this feature.

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: For the Simulated Clinical Use study, 640 device samples were used for both the Lancet and the Lancing Device. Other performance and biocompatibility tests would have their own sample sizes, but these are not explicitly stated in this summary for each specific test (e.g., mechanical tests, needle integrity, etc.).
    • Data Provenance: The studies were conducted by Tianjin Huahong Technology Co., Ltd. (China). The data origin is thus China. The studies described are prospective tests performed on the new device, not retrospective data analysis.

    3. Number of Experts Used to Establish Ground Truth and Their Qualifications

    Not applicable. This device is a mechanical medical device, not an AI/software device that requires expert radiological or clinical interpretation to establish ground truth for performance evaluation in the way AI algorithms do. The "ground truth" for this device's acceptance is based on engineering specifications, physical measurements, biological safety tests (biocompatibility), and functional performance tests against established standards.

    4. Adjudication Method for the Test Set

    Not applicable. As this is not an AI/software performance study relying on human interpretation, there is no "adjudication" in the sense of reconciling expert opinions. Performance is assessed against quantitative engineering specifications and standard test methods.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This document is for a mechanical medical device (lancet and lancing device), not an Artificial Intelligence (AI) or software device. Therefore, no MRMC study, human reader improvement, or AI assistance is relevant or discussed.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an algorithmic device. Performance is evaluated through physical and functional testing of the device itself.

    7. The Type of Ground Truth Used

    The "ground truth" for showing the device meets acceptance criteria is based on:

    • Engineering Specifications/Standards: Performance criteria outlined in internal specifications and relevant international standards (e.g., ISO 23908 for sharps injury prevention, ISO 10993-1 for biocompatibility).
    • Physical Measurements: E.g., needle length range, gauge range.
    • Functional Testing: Demonstrating the device performs its intended action (capillary blood sampling, proper retraction).
    • Biocompatibility Testing: Laboratory results confirming non-toxicity and safety of materials.
    • Safety Mechanism Testing: Verification that sharps injury prevention features work as intended.

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI/machine learning device, so there is no concept of a "training set" for an algorithm.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As there is no training set, there is no ground truth establishment for a training set.