Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K242184

    Validate with FDA (Live)

    Date Cleared
    2024-12-04

    (132 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Alcon 27GA Chandelier
    The 27GA Chandelier is indicated for use during posterior segment ophthalmic surgery.

    Alcon 27+ DS Wide Angle Endoilluminator
    The 27+DS Wide-Angle Endoilluminator is indicated for use during posterior segment ophthalmic surgery.

    Alcon 27+ DS Endoilluminator
    The 27 + DS Endoilluminator is indicated for use during posterior segment ophthalmic surgery.

    Device Description

    As part of this premarket notification Alcon presents three new endoilluminators, compatible with the UNITY VCS device cleared in K233876 (clearance date: June 21, 2024).

    Alcon 27GA Chandelier: The Alcon 27GA Chandelier provides a hands-free wide field of illumination for vitreoretinal procedures. The distal end of the device incorporates a short, tapered fiber optic to provide wide angle illumination projected from a pars plana position of the eye. The distal end consists of a tapered fiber, which is inserted into the eye through the Alcon 27GA entry system. The fiber is malleable so the physician can shape it to form a stable service loop, position the tip, and minimize interference with the lid speculum.

    Alcon 27+ DS Wide Angle Endoilluminator: The Alcon 27+ DS WA Endoilluminator offers wide-angle illumination, featuring a sapphire lens at the distal end of the probe. This design ensures consistent, low-glare illumination across a broad field of view. Additionally, it incorporates a Dynamic Stiffener (DS) element-a retractable stiffener designed to minimize needle deflection at the proximal end of the stainless steel cannula. as opposed to the fixed stiffener used in previous models.

    Alcon 27+ DS Endoilluminator: The Alcon 27+ DS Endoilluminator transmits light energy into the eye and incorporates a DS element at the proximal end of the stainless-steel cannula.

    AI/ML Overview

    This FDA 510(k) summary does not contain the specific acceptance criteria or performance data in a format that directly matches the requested table. This document is a "Substantial Equivalence" submission, which focuses on demonstrating that new devices (Alcon 27GA Chandelier, Alcon 27+ DS Wide Angle Endoilluminator, Alcon 27+ DS Endoilluminator) are as safe and effective as previously cleared predicate devices, rather than establishing specific performance metrics against pre-defined acceptance criteria for a novel device.

    The document primarily relies on non-clinical testing to demonstrate substantial equivalence. Here's a breakdown of the information available:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document does not provide a table of acceptance criteria with corresponding performance results in a quantitative manner for specific clinical metrics. Instead, it states that the devices "met ISO 10993" for biocompatibility, "passed the predetermined acceptance criteria" for sterilization and shelf life, and that "safety tests demonstrated the compliance of the subject devices to the appropriate standards" for performance testing. For light hazard testing, it states the devices "meet aphakic and thermal safety standards comparable to the predicate devices."

    The comparisons in Tables 1-3 highlight similarities in technological characteristics between the subject devices and their predicates, rather than presenting numeric acceptance criteria.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Non-Clinical Testing: The document refers to "tested devices" for sterilization, "tested units" for packaging integrity, and "three subject devices" for light hazard testing. However, specific sample sizes (e.g., number of units tested for biocompatibility, number of sterilization cycles validated) are not provided.
    • Data Provenance: All testing described is non-clinical. There is no mention of human subject data, therefore no information on country of origin or retrospective/prospective nature.

    3. Number of Experts Used to Establish Ground Truth and Qualifications:

    Not applicable, as this submission concerns non-clinical testing of medical devices (endoilluminators and a chandelier) for surgical use, not AI/diagnostic devices requiring expert adjudication for ground truth.

    4. Adjudication Method for the Test Set:

    Not applicable, as this submission concerns non-clinical testing of medical devices (endoilluminators and a chandelier), not AI/diagnostic devices requiring adjudication.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done:

    No, an MRMC comparative effectiveness study was not done. This submission is for physical medical devices, not AI software where such studies are typically performed.

    6. If a Standalone Performance (Algorithm Only Without Human-in-the-Loop Performance) was Done:

    Not applicable. This is not an AI device. The "performance testing" referenced is for the physical attributes and safety of the endoilluminators (e.g., light transmission, mechanical integrity, biocompatibility), which are inherently "standalone" in their assessment as device components.

    7. The Type of Ground Truth Used:

    For the non-clinical tests, the "ground truth" is defined by:

    • Biocompatibility: ISO 10993 standards.
    • Sterilization: ISO 11135 and ISO 10993-7 standards; achieving a Sterility Assurance Level (SAL) of 10⁻⁶.
    • Shelf Life/Packaging: ASTM D4169, ASTM F1980, and ISO 11607-1 standards.
    • Performance (General): ISO 14971, ISO 15004-1, ISO 15752, and ANSI Z80.36 standards.
    • Light Hazard: ANSI Z80.36 and ISO 15752 standards, and comparability to predicate devices.

    8. The Sample Size for the Training Set:

    Not applicable. This is not an AI/machine learning device; therefore, there is no training set in the AI sense.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable. There is no training set.

    Summary of Device Acceptance:

    The acceptance of these devices is based on demonstrating substantial equivalence to existing predicate devices (cleared under K875005 and K110951) through a battery of non-clinical tests. The tests confirm they meet established safety and performance standards (ISO, ASTM, ANSI) and have comparable technological characteristics and risk profiles to their predicates, making them safe and effective for their intended use in posterior segment ophthalmic surgery.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1