Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K240761

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-03-21

    (366 days)

    Product Code
    Regulation Number
    880.5440
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Arisure® Closed Male Luer with Spike Adapter serves as a connector between an IV container and a standard IV set.

    Device Description

    The Arisure Closed Male Luer with Spike Adapter, also referred to as Arisure CML with Spike Adapter or CML with Spike Adapter, is an extension of the predicate device, i.e., Closed Male Luer. The primary components of the CML with Spike Adapter include: a Closed Male Luer, a Male Luer Lock Connector, and an Extended Spike Adapter.

    The subject device allows for administration set preservation when administering multiple drug therapies to the same patient. It provides a needle-free, drip-free connection between an IV bag or a rigid container and an IV administration set. The CML side provides a luer lock connection to a Bag Spike with Neutral Valve. It also gives the user a drip-free disconnect when needing to use multiple bags or bottles for the infusion. The Spike Adapter side allows for a secure connection to any ISO standard IV spike. An IV administration set can be connected safely and securely to the device through a pierceable septum.

    The CML with Spike Adapter is designed to comply with ISO 8536-4:2019 as appropriate.

    AI/ML Overview

    The Arisure® Closed Male Luer with Spike Adapter (YM060) is an extravascular administration set.
    The device's performance was evaluated through a series of non-clinical tests.

    Acceptance Criteria and Device Performance:

    Test NameAcceptance CriteriaReported Device Performance
    Device Flow RateNot explicitly stated, but implies meeting functional flow ratesPassed
    LeakageNot explicitly stated, but implies no leakagePassed
    Separation ForceNot explicitly stated, but implies secure connectionPassed
    Detachment ForceNot explicitly stated, but implies appropriate releasePassed
    Assembly StrengthNot explicitly stated, but implies secure assemblyPassed
    Multiple ActivationsNot explicitly stated, but implies proper function over multiple usesPassed
    Extended ActivationsNot explicitly stated, but implies proper function over prolonged usePassed
    Dry DisconnectionNot explicitly stated, but implies proper function upon disconnectionPassed
    Microbial IngressNot explicitly stated, but implies prevention of microbial entryPassed
    ParticulatesNot explicitly stated, but implies acceptable particulate levelsPassed
    Chemical CompatibilityCompatibility with chemicalsPassed
    Shelf-Life TestingThree (3) yearsAchieved three (3) years
    Distribution TestingNot explicitly stated, but implies integrity during distributionPassed
    Package Performance TestingNot explicitly stated, but implies maintaining sterility and integrityPassed
    Package Stability TestingNot explicitly stated, but implies maintaining sterility and integrity over timePassed
    Penetration ability of the septum of access portNot explicitly stated, but implies ease of penetration and integrityPassed
    Impermeability of the septum of access portNot explicitly stated, but implies no fluid passage through septumPassed
    Adhesion strength of the infusion deviceNot explicitly stated, but implies secure adhesionPassed
    Biocompatibility (Cytotoxicity, Sensitization, Irritation, Material Mediated Pyrogenicity, Acute Systemic Toxicity, Sub-Acute Toxicity, Hemocompatibility)Meeting ISO 10993-1 requirements for external communicating devices, blood path, indirect contact, prolonged duration (>24 hours to 30 days)Passed all specified tests
    Sterilization Validation (e-beam irradiation)Sterility Assurance Level (SAL) of 10^-6Achieved SAL of 10^-6

    Study Details:

    1. Sample Size and Data Provenance for Test Set:

      • The document does not explicitly state the specific sample sizes used for each individual performance test (e.g., leakage, flow rate). It generally indicates that "Aged, packaged parts underwent the appropriate testing" for shelf-life, and "The following tests were completed with passing results."
      • The data provenance is not specified (e.g., country of origin). The studies appear to be non-clinical performance and laboratory tests conducted by the manufacturer, Yukon Medical, LLC. These are retrospective data in the sense that they were collected and analyzed to support the 510(k) submission.
    2. Number of Experts and Qualifications for Ground Truth:

      • This information is not applicable. The device is a medical accessory, and its performance criteria are objective, based on engineering, material, and biological standards (e.g., ISO, ASTM). The "ground truth" for these tests is established by the specifications of these standards, not expert consensus in diagnostic interpretation.
    3. Adjudication Method for Test Set:

      • This information is not applicable as the evaluation involves objective performance metrics against established standards, not interpretation by human experts.
    4. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

      • No MRMC comparative effectiveness study was done. This type of study is typically performed for diagnostic or imaging devices where human interpretation is involved. The Arisure® Closed Male Luer with Spike Adapter is a medical accessory with objective performance characteristics.
    5. Standalone Performance Study:

      • Yes, a standalone performance study was done in the sense that the device's performance was evaluated independently against engineering, material, and biological standards. This involved testing various mechanical, chemical, and biological attributes of the device without human-in-the-loop performance measurement related to its primary function as a connector.
    6. Type of Ground Truth Used:

      • The ground truth used for performance evaluation was based on established international and national standards for medical devices, infusion equipment, and material biocompatibility. These include, but are not limited to, ISO 8536-4, ISO 80369-7, IEC 62366-1, ISO 22413, ISO 15747, ISO 10993 series, ASTM standards (e.g., D4169, F1980), and EN/SS-EN standards for sterilization. The "ground truth" is defined by the measurable parameters and acceptance limits specified within these standards.
    7. Sample Size for Training Set:

      • The document describes performance testing for a medical device accessory and does not involve AI or machine learning algorithms that would typically require a "training set." Therefore, this information is not applicable.
    8. How Ground Truth for Training Set Was Established:

      • As there is no AI/machine learning component described for this device, a training set and its associated ground truth establishment are not relevant to this submission.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1