Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K240545

    Validate with FDA (Live)

    Date Cleared
    2024-08-28

    (183 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The GenaCheck™ Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs and Fentanyl Citrate is a specialty medical glove which is a disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between examiner and patient. The glove was tested for use with Chemotherapy Drugs and Fentanyl Citrate as per ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

    Device Description

    The subject device is single use, disposable gloves intended for medical purposes to be worn on the examiner's hands to prevent contamination between patient and examiner. The gloves are powderfree, ambidextrous with beaded cuff, purple colored, nitrile, and tested for use with chemotherapy drugs and Fentanyl Citrate. The gloves are offered in five sizes: XS, S, M, L, XL. The subject device is non-sterile.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the GenaCheck™ Nitrile Powder Free Examination Glove, based on the provided FDA document:

    1. Table of Acceptance Criteria and Reported Device Performance

    Test Method / PropertyPurposeAcceptance CriteriaReported Device Performance
    Physical Performance (ASTM D6319-19)
    LengthPhysical Dimensions TestXS/S: ≥220 mm; M/L/XL: ≥230 mmXS/S: ≥220 mm (Pass); M/L/XL: ≥230 mm (Pass)
    WidthPhysical Dimensions TestXS: 70 ± 10 mm; S: 80 ± 10 mm; M: 95 ± 10 mm; L: 110 ± 10 mm; XL: 120 ± 10 mmXS: 73-78 mm (Pass); S: 84-88 mm (Pass); M: 93-98 mm (Pass); L: 103-108 mm (Pass); XL: 113-118 mm (Pass)
    ThicknessPhysical Dimensions TestFinger: ≥0.05 mm; Palm: ≥0.05 mmFinger: 0.069-0.109 mm (Pass); Palm: 0.052-0.079 mm (Pass)
    Watertightness (ASTM D5151-19)
    Freedom from HolesWatertightness TestMet the requirements of ASTM D5151 AQL 2.50/125 (Pass)
    Powder Content (ASTM D6124-06)
    Powder ContentPowder ContentMeet the requirements of ASTM D6124 < 2.0 mg0.27-0.41 mg (Pass)
    Physical Properties (ASTM D412-16)
    Before Aging - Tensile StrengthPhysical properties≥14 MPa14-27 MPa (Pass)
    Before Aging - Ultimate ElongationPhysical properties≥500%516-769 % (Pass)
    After Aging - Tensile StrengthPhysical properties≥14 MPa14-20 MPa (Pass)
    After Aging - Ultimate ElongationPhysical properties≥400%452-662 % (Pass)
    Biocompatibility
    Cytotoxicity (ISO 10993-5)CytotoxicityUnder conditions of the study, device extract is cytotoxic. (This criteria seems to indicate the expected outcome of the test, rather than a desired performance threshold; however, the intent is likely to confirm that it is not cytotoxic, as shown in the predicate comparison for the predicate device, which states "device extract is cytotoxic" and is marked as "Same." This implies the predicate itself also had this result, and it's being compared for similarity, not necessarily for a "pass/fail" against a non-cytotoxic ideal in this context for substantial equivalence.)Under conditions of the study, device extract is cytotoxic (Result confirms similarity to predicate's reported extract cytotoxicity). Note: Given the context of a 510(k) for an examination glove, this result, while listed as "cytotoxic", aligns with the predicate and might be interpreted as acceptable within the framework of demonstrating substantial equivalence, meaning the new device is no worse than the predicate in terms of cytotoxicity. It's unusual for "cytotoxic" to be an acceptance criteria, more commonly "non-cytotoxic". This phrasing might indicate a specific interpretation or comparison to the predicate's known characteristics. More commonly, "non-cytotoxic" is the desired outcome.
    Acute Systemic Toxicity (ISO 10993-11)Acute Systemic ToxicityUnder conditions of the study, device extract did not show acute systemic toxicity in vivoUnder conditions of the study, device extract did not show acute systemic toxicity in vivo (Pass)
    Irritation (ISO 10993-23)IrritationUnder conditions of the study, the device was non-irritatingUnder conditions of the study, the device was non-irritating (Pass)
    Sensitization (ISO 10993-10)SensitizationUnder conditions of the study, the device is non-sensitizingUnder conditions of the study, the device is non-sensitizing (Pass)
    Chemopermeation (ASTM D6978-05)
    Carmustine (3.3 mg/ml)Breakthrough Detection TimeNo explicit acceptance criterion defined in the table, but the indication highlights "low permeation time"22.8 Minutes
    Cisplatin (1.0 mg/ml)Breakthrough Detection TimeNo explicit acceptance criterion defined in the table, but the indication is for comparison to similar predicate results> 240 Minutes
    Cyclophosphamide (20.0 mg/ml)Breakthrough Detection TimeNo explicit acceptance criterion defined in the table> 240 Minutes
    Dacarbazine (10.0 mg/ml)Breakthrough Detection TimeNo explicit acceptance criterion defined in the table> 240 Minutes
    Doxorubicin HCI (2.0 mg/ml)Breakthrough Detection TimeNo explicit acceptance criterion defined in the table> 240 Minutes
    Etoposide (20.0 mg/ml)Breakthrough Detection TimeNo explicit acceptance criterion defined in the table> 240 Minutes
    Fluorouracil (50.0 mg/ml)Breakthrough Detection TimeNo explicit acceptance criterion defined in the table> 240 Minutes
    Methotrexate (25.0 mg/ml)Breakthrough Detection TimeNo explicit acceptance criterion defined in the table> 240 Minutes
    Mitomycin C (0.5 mg/ml)Breakthrough Detection TimeNo explicit acceptance criterion defined in the table> 240 Minutes
    Paclitaxel (6.0 mg/ml)Breakthrough Detection TimeNo explicit acceptance criterion defined in the table> 240 Minutes
    Thiotepa (10.0 mg/ml)Breakthrough Detection TimeNo explicit acceptance criterion defined in the table, but the indication highlights "low permeation time"37.9 Minutes
    Vincristine Sulfate (1.0 mg/ml)Breakthrough Detection TimeNo explicit acceptance criterion defined in the table> 240 Minutes
    Fentanyl Citrate Inj. (100mcg/2ml)Breakthrough Detection TimeNo explicit acceptance criterion defined in the table> 240 Minutes

    Study Details:

    For medical gloves, the "study" demonstrating that the device meets acceptance criteria typically refers to a series of non-clinical, laboratory bench tests conducted according to recognized consensus standards.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the specific sample sizes for each test. However, it references the standards used, which often specify sample sizes.

    • For Watertightness Test (ASTM D5151-19), the result "0/125/Pass" indicates that 125 samples were tested, with 0 failures.
    • For other tests like physical dimensions, powder content, and physical properties, the sample sizes are described by the respective ASTM standards but not explicitly listed in this summary.
    • Data Provenance: The document does not specify the country of origin of the data. The tests are "non-clinical testing" (bench tests) rather than human subject studies, so there's no "retrospective or prospective" status in the same sense as clinical data.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This information is not applicable for this type of device and study. The "ground truth" for non-clinical performance and material properties standards is established by the specifications defined within the referenced ASTM and ISO standards themselves (e.g., a tensile strength of ≥14 MPa, or a specific AQL for holes). There are no human experts classifying outcomes in the way one would for diagnostic imaging.

    4. Adjudication Method for the Test Set

    This is not applicable. Adjudication methods like "2+1" are used for human-reviewed data sets (e.g., medical image interpretation consensus). For bench testing against pre-defined engineering and material specifications, the results are quantitative measurements compared directly to the acceptance criteria.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, an MRMC comparative effectiveness study was not done. This type of study is relevant for comparing the performance of diagnostic tools (e.g., AI algorithms for interpreting medical images) with and without human readers' involvement. The device here is a medical glove, for which MRMC studies are not applicable.

    6. If a Standalone Study (i.e., algorithm only without human-in-the-loop performance) was Done

    No, a standalone study in the context of an "algorithm only" performance was not done. This concept is specific to software medical devices or AI algorithms. The GenaCheck™ glove is a physical medical device.

    7. The Type of Ground Truth Used

    The "ground truth" for the performance tests relies on established industry and regulatory standards.

    • For physical properties, it's defined quantities like minimum tensile strength (e.g., ≥14 MPa).
    • For resistance to permeation by chemotherapy drugs, it's the measured "breakthrough detection time" according to the specified method (ASTM D6978-05).
    • For biocompatibility, it's the absence of adverse biological reactions (non-irritating, non-sensitizing, no acute systemic toxicity, and comparison to predicate regarding cytotoxicity) as defined by ISO 10993 series.

    8. The Sample Size for the Training Set

    There is no training set for this device. Training sets are applicable to machine learning algorithms. The GenaCheck™ glove is a physical product, not an AI/ML algorithm.

    9. How the Ground Truth for the Training Set Was Established

    This is not applicable as there is no training set.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1