    Intended Use

    The CEREGLIDE 92 Catheter System is indicated for use in facilitating the insertion and guidance of appropriately sized interventional devices into a selected blood vessel in the neurovascular system.

    Device Description

    The CEREGLIDE™ 92 Catheter System consists of the CEREGLIDE™ 92 Intermediate Catheter and associated accessories, including the INNERGLIDE™ 9 Delivery Aid, Rotating Hemostasis Valve (RHV), and Slit Introducers.

    Intermediate Catheter: The CEREGLIDE™ 92 Intermediate Catheter is a single lumen, variable stiffness catheter designed to be introduced over a steerable guidewire along with a microcatheter and/or compatible support device into the neuro vasculature. The catheter consists of a lubricious PTFE lined inner lumen to facilitate movement of the guidewires and other devices, variable pitch stainless steel and tungsten braid, and various durometer polymer jackets. These jackets provide distal flexibility and gradually transition to a stiffer proximal shaft to facilitate the advancement of the catheter in the anatomy. The outer surface of the catheter is hydrophilic coated in order to reduce friction during manipulation in the vessel. A radiopaque marker at the distal end of the catheter provides fluoroscopic visualization of the catheter tip. The proximal end of the catheter incorporates a standard Luer adapter to facilitate the attachment of accessories, a hub, and an ID band.

    The CEREGLIDE™ 92 Catheter System is packaged with an INNERGLIDE™ 9 Delivery Aid, an RHV with a side port, and two Slit Introducer accessories. The INNERGLIDE™ 9 Delivery Aid is a compatible support device comprising of a single lumen catheter with a tapered tip, hydrophilic coating for increased lubricity, radiopaque tip for fluoroscopic visualization, and a proximal luer hub that is designed to be used as part of the CEREGLIDE™ 92 Catheter System to facilitate delivery of the CEREGLIDE™ 92 Intermediate Catheter to select blood vessels in the neurovasculature. The RHV with side port is used for flushing and insertion of catheters. The Slit Introducers are designed to introduce the CEREGLIDE™ 92 Intermediate Catheter into the base catheter and protect the distal tip of the CEREGLIDE™ 92 Intermediate Catheter during insertion into the hemostasis valve of the base catheter.

    AI/ML Overview

    The provided text describes information about the CEREGLIDE 92 Catheter System (K233982). However, it does not include details about acceptance criteria or a study proving the device meets those criteria, as typically found in submissions for Artificial Intelligence/Machine Learning (AI/ML) based medical devices. The document focuses on demonstrating substantial equivalence to a predicate device (CEREGLIDE™ 92 Intermediate Catheter K230726) through non-clinical performance testing (bench, animal, and clinical were not required).

    Therefore, the following information cannot be extracted from the provided text:

    • A table of acceptance criteria and the reported device performance.
    • Sample size used for the test set and the data provenance.
    • Number of experts used to establish the ground truth for the test set and their qualifications.
    • Adjudication method for the test set.
    • Multi Reader Multi Case (MRMC) comparative effectiveness study results.
    • Standalone (algorithm only without human-in-the-loop performance) study results.
    • The type of ground truth used.
    • The sample size for the training set.
    • How the ground truth for the training set was established.

    The "acceptance criteria" mentioned in the document refer to the success metrics for the bench testing conducted on the CEREGLIDE 92 Catheter System, primarily for individual physical and functional properties, not for an AI/ML-based diagnostic or prognostic task. The reported performance for these tests is simply "PASS: Samples met the established acceptance criteria."

    Summary of what the document does provide regarding performance testing:

    The device underwent non-clinical performance testing, primarily bench testing, to demonstrate substantial equivalence to the predicate device.

    1. A table of acceptance criteria and the reported device performance:

    TestAcceptance CriteriaReported Device Performance
    Visual InspectionMeet visual requirements described in ISO 10555-1 Section 4.4PASS: Samples met the established acceptance criteria
    Catheter IDMeet design requirements (internal diameters)PASS: Samples met the established acceptance criteria
    Catheter ODMeet design requirements (outer diameters)PASS: Samples met the established acceptance criteria
    Catheter Working LengthMeet working length definition in ISO 10555-1 Section 3.6PASS: Samples met the established acceptance criteria
    Catheter Tip LengthMeet design requirements (tip length and navigation length)PASS: Samples met the established acceptance criteria
    Hub LuerMeet dimensional, interconnectability, and performance requirements defined in ISO 80369-7PASS: Samples met the established acceptance criteria
    System Air Leak ResistanceNo air leakPASS: Samples met the established acceptance criteria
    System Liquid Leak ResistanceMeet freedom from leakage (liquid during pressurization) requirements of ISO 10555-1:2013, section 4.7PASS: Samples met the established acceptance criteria
    Hub Pull TestingMeet acceptance criteria (hub to joint strength)PASS: Samples met the established acceptance criteria
    Shaft Tensile StrengthMeet acceptance criteria (joint strength)PASS: Samples met the established acceptance criteria
    Particulate CountComparison to predicate devicesPASS: Samples met the established acceptance criteria
    Coating Durability and FrictionMeet design requirements (friction force and durability of hydrophilic coating)PASS: Samples met the established acceptance criteria
    Coating LengthMeet design requirements (hydrophilic coating length)PASS: Samples met the established acceptance criteria
    Coating ThicknessMeet design requirements (hydrophilic coating thickness)PASS: Samples met the established acceptance criteria
    Coating IntegrityFree of uncoated individual voids (Congo Red Dye Coverage Test)PASS: Samples met the established acceptance criteria
    Kink ResistanceRemain stable and not kink during usePASS: Samples met the established acceptance criteria
    Tip FlexibilityRelative to cleared devices of similar designPASS: Samples met the established acceptance criteria
    Torque PerformanceMeet torque strength requirementPASS: Samples met the established acceptance criteria
    Trackability (Delivery and Withdrawal Force)Meet design requirementsPASS: Samples met the established acceptance criteria
    Burst PressureMeet minimum static burst pressure requirementsPASS: Samples met the established acceptance criteria
    In Vitro Usability StudiesEvaluate compatibility, trackability, and durability in a neurovascular modelPASS: Samples met the established acceptance criteria

    Biocompatibility Testing:

    Test MethodAcceptance CriteriaResult
    Cytotoxicity - MEM ElutionGrade ≤ 2PASS
    Sensitization - Guinea Pig MaximizationGrade < 1PASS
    Irritation - Rabbit Intracutaneous ReactivityDifference of ≤ 1 from control mean scorePASS
    Acute Systemic Toxicity in MiceNo biologically significant signs of systemic toxicityPASS
    Pyrogenicity - Materials Mediated Rabbit PyrogenTemperature change (△T) < 0.5°CPASS
    ASTM Hemolysis Study - Direct and Extraction MethodsHemolytic Index < 2%PASS
    SC5b-9 Complement Activation AssayResults evaluated based on comparison to negative controlsPASS
    In Vivo ThromboresistanceResults evaluated based on comparison to negative controlsPASS

    2. Sample sized used for the test set and the data provenance:
    "All testing was conducted using sampling methods as required by internal design control procedures." Specific sample sizes for each test are not provided. The data provenance is internal laboratory testing ("bench testing").

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
    Not applicable; this device is not an AI/ML diagnostic. Performance was established through engineering tests.

    4. Adjudication method for the test set:
    Not applicable; this device is not an AI/ML diagnostic.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:
    Not applicable; this device is not an AI-assisted diagnostic.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
    Not applicable; this device does not involve an algorithm.

    7. The type of ground truth used:
    For the non-clinical performance testing, the "ground truth" was established by predefined engineering specifications, industry standards (e.g., ISO 10555-1, ISO 80369-7, ISO 11135, EN ISO 10993-7, ISO 10993-1), and internal design requirements.

    8. The sample size for the training set:
    Not applicable; this device is not an AI/ML device and does not have a training set.

    9. How the ground truth for the training set was established:
    Not applicable.