Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K233308

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-04-25

    (209 days)

    Product Code
    Regulation Number
    872.4200
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Air Motor is intended for the following application(s): Driving the handpiece for the tooth restoration, prophylaxis, and root canal preparation.

    Device Description

    The air motor converts the compressed air supplied from the dental unit into rotary motion. The rotation is transmitted to instruments for tooth restoration, prophylaxis, and root canal preparation connected to straight and geared angle handpieces.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a dental air motor. The core of this document is a comparison of the subject devices (Air Motor FX204 M4 and Air Motor M205) against predicate devices to demonstrate substantial equivalence, rather than a clinical study evaluating the device against acceptance criteria in the typical sense of AI/ML performance.

    Therefore, many of the requested elements for describing "acceptance criteria and the study that proves the device meets the acceptance criteria" for AI/ML devices are not applicable in this context. The document focuses on performance testing to demonstrate safety and effectiveness for a mechanical device.

    However, I can extract the information relevant to how the device's technical specifications and performance were evaluated against established standards and predicate devices.

    Here's a breakdown of the relevant information from the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are not explicitly stated as numerical thresholds in a table format for each specific performance metric in the way you might see for an AI algorithm (e.g., AUC > 0.90). Instead, the acceptance criterion for this 510(k) submission is to demonstrate "substantial equivalence" to legally marketed predicate devices by showing that the subject devices perform similarly and meet relevant safety and performance standards.

    The closest we have to "reported device performance" and "acceptance criteria" are the comparison tables (Table 1 for FX204 M4 and Table 2 for M205) and the "Judgement" column, which indicates if the subject device is "Same," "Similar," or "Different" to the predicate. For "Different" or "Similar" aspects, further testing or justification is provided to demonstrate that the difference does not raise new questions of safety or effectiveness.

    Below are tables summarizing the key performance and safety aspects evaluated and how the subject device's performance compared to the predicates, along with the implicit "acceptance criteria" through this comparison.

    For Air Motor, FX204 M4 (model without spray) - (Based on "DEVICE AND PREDICATE COMPARISON TABLE-1")

    Characteristic / Performance AspectAcceptance Criteria (Implicit from Predicate Comparison & Standards)Reported Device Performance (Subject Device: Air Motor, FX204 M4)Judgment (vs. Predicate/Reference)
    Indications for UseSame as predicate devices for general dental applications.Driving handpiece for tooth restoration, prophylaxis, root canal preparation.Same
    Device Description/TechnologyPneumatically driven air motor for dental handpieces.Converts compressed air into rotary motion for dental instruments.Same
    Supply air pressureWithin acceptable range for dental applications (compared to predicates).0.20 – 0.25 MPaDifferent (lower pressure than primary predicate, but considered sufficient)
    Optic lightingSimilar to some predicate devices or not raising new safety/effectiveness concerns.NoSame with Reference Device (Air-Powered Handpieces)
    Speed rangeSimilar upper speed range to predicate devices.19,800 - 24,200 min⁻¹ (0.25 MPa)Similar
    Max. air consumptionSimilar to predicate devices, not raising safety/effectiveness concerns.< 66 NL/minSimilar with Reference Device (Air-Powered Handpieces)
    WeightSimilar to predicate devices, not raising safety/effectiveness concerns.61.7gSimilar with Reference Device (Air-Powered Handpieces)
    Performance TestingConformity to ISO 14457 (dental handpieces) and other relevant standards (sterilization, cleaning).Passed ISO 14457:2017; AAMI/ANSI/ISO 17665-1:2006 (Sterilization); FDA Guidance for Cleaning.Similar (All devices passed ISO 14457, relevant testing to predicates)
    BiocompatibilityConformity to ISO 7405 and AAMI/ANSI/ISO 10993-1.Conformity to ISO 7405:2018; AAMI/ANSI/ISO 10993-1:2018.Similar
    Clinical TestingNot required for this type of device based on its nature and predicate comparison.NOSame (as predicates)
    Air/water portNo spray design, similar to predicate.No spraySame
    DimensionsSimilar to predicate devices, not raising new safety/effectiveness concerns.78.3 x φ20 mmSimilar
    Patient-contacting portionsIndirect contact limited, conformity to biocompatibility standards.Direct contact: No; Indirect contact: NoSimilar
    Handpiece ConnectionISO 3964 standard connection.ISO 3964Same with Reference Device (Air-Powered Handpieces)
    Hose ConnectionISO 9168 standard connection.ISO 9168Same with Reference Device (Air-Powered Handpieces)
    LubricantPredetermined lubricant (PANA SPRAY Plus).PANA SPRAY Plus (K163483)Similar with Reference Device (Air-Powered Handpieces)

    For Air Motor, M205 (Internal Spray equipped) - (Based on "DEVICE AND PREDICATE COMPARISON TABLE-2")

    Characteristic / Performance AspectAcceptance Criteria (Implicit from Predicate Comparison & Standards)Reported Device Performance (Subject Device: Air Motor, M205)Judgment (vs. Predicate/Reference)
    Indications for UseSame as predicate devices for general dental applications.Driving handpiece for tooth restoration, prophylaxis, root canal preparation.Similar
    Device Description/TechnologyPneumatically driven air motor for dental handpieces, with water injection.Converts compressed air into rotary motion; has water injection mechanism.Same (pneumatically driven, includes spray feature like reference)
    Supply air pressureWithin acceptable range for dental applications (compared to predicates).0.20 – 0.25 MPaDifferent (lower pressure than primary predicate, but considered sufficient)
    Optic lightingSimilar to some predicate devices or not raising new safety/effectiveness concerns.NoSame with Reference Device (Air-Powered Handpieces)
    Speed rangeSimilar upper speed range to predicate devices.19,800 - 24,200 min⁻¹ (0.25 MPa)Similar
    Max. air consumptionSimilar to predicate devices, not raising safety/effectiveness concerns.< 66 NL/minSimilar with Reference Device (Air-Powered Handpieces)
    WeightSimilar to predicate devices, not raising safety/effectiveness concerns.78.2gSimilar with Reference Device (Air-Powered Handpieces)
    Spray water flow acc. to ISO 14457> 50 (ml/min) based on predicate.> 50 (ml/min)Same
    Water pressureSimilar range to predicate/reference.1 – 2.5 (bar)Similar range
    Chip air pressureSimilar range to predicate/reference.1.5 – 2.5 (bar)Similar range
    Performance TestingConformity to ISO 14457 (dental handpieces) and other relevant standards (sterilization, cleaning).Passed ISO 14457:2017; AAMI/ANSI/ISO 17665-1:2006 (Sterilization); FDA Guidance for Cleaning. Also check valve effect verification.Similar (All devices passed ISO 14457, relevant testing to predicates)
    BiocompatibilityConformity to ISO 7405 and AAMI/ANSI/ISO 10993-1.Conformity to ISO 7405:2018; AAMI/ANSI/ISO 10993-1:2018.Similar
    Clinical TestingNot required for this type of device based on its nature and predicate comparison.NOSame (as predicates)
    Air/water portSpray equipped.SpraySame with Reference Device (Air-Powered Handpieces)
    DimensionsSimilar to predicate devices, not raising new safety/effectiveness concerns.78.3 x φ20 mmSimilar
    Patient-contacting portionsIndirect contact (waterlines) defined, conformity to biocompatibility standards.Direct contact: No; Indirect contact: WaterlinesSimilar with Reference Device (Air-Powered Handpieces)
    Composition of the waterlinesMaterials compatible with biocompatibility standards.Fluor rubber, Stainless steel, Aluminum based alloyDifferent (but conformity to ISO 10993-1 supports substantial equivalence)
    Handpiece ConnectionISO 3964 standard connection.ISO 3964Same with Reference Device (Air-Powered Handpieces)
    Hose ConnectionISO 9168 standard connection.ISO 9168Same with Reference Device (Air-Powered Handpieces)
    LubricantPredetermined lubricant (PANA SPRAY Plus).PANA SPRAY Plus (K163483)Similar with Reference Device (Air-Powered Handpieces)

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: The document mentions "Statistical calculation of sample size" under performance testing, but does not provide specific numbers for units tested for each type of test. The nature of these tests (e.g., drop test, reprocessing resistance, water leakage, vibration, sterilization validation) typically involves testing a representative sample of devices, but the exact number isn't quantified in the provided text.
    • Data Provenance: Not explicitly stated, however, the applicant is "NAKANISHI INC." based in "700 Shimohinata Kanuma Tochigi, 322-8666, Japan". It's highly probable the testing was conducted in Japan or through certified testing facilities used by the manufacturer. The data would be prospective in the sense that these tests are performed on the manufactured devices to ensure they meet specifications.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    • This concept typically applies to clinical studies or studies involving human interpretation (e.g., radiology expert review for AI algorithms). This submission is for a mechanical dental device.
    • The "ground truth" here is compliance with engineering standards (e.g., ISO 14457 for dental handpieces), safety standards (e.g., ISO 10993 for biocompatibility), and the performance characteristics of the predicate devices.
    • No "experts" in the sense of clinical reviewers establishing ground truth for individual cases are mentioned. Instead, the evaluations are based on technical testing by qualified personnel in accredited laboratories.

    4. Adjudication Method for the Test Set

    • Adjudication methods (like 2+1, 3+1) are used in studies involving human interpretation or subjective assessments, often in AI/ML performance evaluation where ground truth is based on expert consensus.
    • This is not applicable here as the evaluation is based on objective measurements against engineering standards and comparison with predicate device specifications.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • Not applicable. This is a submission for a mechanical dental motor, not an AI/ML diagnostic or assistive device that would involve human readers interpreting cases.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    • Not applicable. This device does not involve a software algorithm in the context of an AI/ML device requiring this type of performance evaluation.

    7. Type of Ground Truth Used

    • The "ground truth" for this submission is primarily technical specifications/performance derived from engineering standards (e.g., ISO 14457, ISO 10993) and comparison to the known and established safety and effectiveness of legally marketed predicate devices.
    • For example, passing ISO 14457 test methods indicates the device performs as expected for a dental handpiece. Biocompatibility is "ground-truthed" by adherence to ISO 10993.

    8. Sample Size for the Training Set

    • Not applicable. This is a mechanical device, not an AI/ML algorithm that requires a "training set."

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable, as there is no "training set" for this type of device.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1