    Intended Use

    NobelProcera® Titanium ASC Abutments:

    NobelProcera® Abutment Titanium is a patient-matched CAD/CAM prosthetic component directly connected to endosseous dental implants and is indicated for use as an aid in prosthetic rehabilitation.

    Omnigrip Clinical Screw Titanium:

    Clinical and Abutment Screws are indicated for use to secure a dental abutment or framework to a dental implant in the maxilla or mandible for supporting tooth replacements and are indicated as an aid in prosthetic rehabilitation.

    Device Description

    The Subject Device NobelProcera® Titanium ASC Abutment is composed of two device lines: NobelProcera® Titanium ASC Abutment and the Clinical Screw Omnigrip Titanium.

    NobelProcera® Titanium ASC Abutment is a patient matched CAD/CAM dental prosthesis which is connected to the Nobel Biocare implants featuring an internal conical and/or internal tri-channel connection, is available in the platform sizes NP, RP WP and 6.0 (for internal tri-channel connection only) and is intended for use as an aid in prosthetic rehabilitation to restore chewing function and esthetic appearance.

    NobelProcera® Titanium ASC Abutments undergo patient matched customization at a Nobel Biocare production facility for the final, finished abutment device manufacturing.

    NobelProcera® Titanium ASC Abutment is connected to the implant with a clinical screw and features an angulated screw channel which can be defined by the customer in an angulation (to the implant's axis) between 0° and 25°, in addition the abutment can be angulated to a maximum of 30°. The clinical screw features the Omnigrip Interface which allows tightening up to 25°.

    NobelProcera® Titanium ASC Abutment and Omnigrip Clinical Screw Titanium are composed of titanium vanadium alloy Ti6Al4V ELI (ISO 5832-3, ASTM F136) and the surface of the abutments are provided with and without anodization and the Omnigrip Clinical Screw Titanium are provided with and without DLC coating.

    The finished NobelProcera® Titanium ASC Abutment supports the placement of a cement-retained dental prosthesis.

    Omnigrip Clinical Screw Titanium is available for the NP, RP, WP and 6.0 (for internal tri-channel connection only) platform, the devices connect the NobelProcera Titanium ASC Abutments to the dental implants. The devices feature an Omnigrip interface.

    AI/ML Overview

    Here's an analysis of the provided text regarding acceptance criteria and device performance.

    Important Note: The provided document is an FDA 510(k) Premarket Notification letter. This type of document is a submission seeking regulatory clearance, not a standalone study report. As such, it outlines the basis for demonstrating substantial equivalence to a predicate device, which includes non-clinical testing. It does not typically contain detailed acceptance criteria and performance data for a study proving a device meets those criteria in the way a clinical trial report or a comprehensive validation study would.

    The document primarily focuses on demonstrating that the new device is substantially equivalent to legally marketed predicate devices, meaning it is as safe and effective. It refers to non-clinical tests that were performed, but does not present the specific acceptance criteria or the numerical results of those tests in a "performance table" format.

    Therefore, many of the requested points below cannot be fully extracted from this document, as it's not the type of report that contains that level of detail for acceptance criteria and specific study outcomes. I will highlight what can be inferred or directly stated from the provided text, and what cannot.


    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly present a table of "acceptance criteria" alongside "reported device performance" in a quantitative manner as one might find in a detailed engineering validation report or a clinical study. Instead, it states that "non-clinical tests demonstrate that the device is substantial equivalent."

    The document mentions several tests and standards the device was subjected to:

    • Packaging system performance testing: per ASTM D4169. (Specific acceptance criteria and reported performance not detailed.)
    • Dynamic loading testing: conducted according to ISO 14801 and FDA Guidance Document ("Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Abutments" (May 12, 2004)). (Specific acceptance criteria and reported performance not detailed.)
    • Magnetic Resonance compatibility testing: according to ASTM F2052, ASTM F2213, ASTM F2119, and ASTM F2182. Status: "MR Conditional." (This is a performance outcome, but the specific metrics and acceptance thresholds for each ASTM standard are not detailed.)
    • Verification of biocompatibility: in accordance with ISO 10993-1. (The conclusion is that "no new issues regarding biocompatibility were raised," implying acceptance, but specific criteria and test results not detailed.)
    • End user cleaning and sterilization validation: in accordance with ISO 17665-1. (Implied acceptance, but specific criteria and results not detailed.)

    The "acceptance criteria" here are generally referred to as meeting the requirements of the standards and demonstrating substantial equivalence to the predicate device. The "reported device performance" is summarized as favorable results sufficient for substantial equivalence determination.

    2. Sample sizes used for the test set and the data provenance

    • Sample sizes: Not specified in the provided text for any of the non-clinical tests mentioned (packaging, dynamic loading, MR compatibility, biocompatibility, sterilization validation).
    • Data provenance: Not explicitly stated (e.g., country of origin). The testing is non-clinical/bench testing, not patient data. It is implied to be internal testing by the manufacturer, Nobel Biocare AB (Sweden).
    • Retrospective or prospective: N/A, as these are non-clinical/bench tests, not clinical studies involving patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Applicability: This question is not directly applicable to the type of non-clinical, bench testing described for the NobelProcera® Titanium ASC Abutment and Omnigrip Clinical Screw Titanium. "Ground truth" established by experts (e.g., radiologists for image interpretation) is relevant for AI/ML device evaluations, particularly in diagnostics. This document refers to physical device testing (mechanical, biocompatibility, etc.).

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Applicability: Not applicable. Adjudication methods are typically used in clinical studies or expert reviews to resolve disagreements in interpretations or diagnoses for establishing "ground truth," which is not the nature of the non-clinical tests described here.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Applicability: Not applicable. This device is a dental abutment and screw, not an AI/ML-enabled diagnostic device for image interpretation. Therefore, no MRMC study or AI assistance evaluation would have been performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Applicability: Not applicable. See point 5. This is a physical medical device, not a standalone algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Applicability: The concept of "ground truth" as typically applied in AI/ML validation (e.g., for diagnostic accuracy) does not directly apply to the non-clinical device testing described. For these tests, the "ground truth" is typically defined by the specified parameters and performance requirements of the relevant international standards (e.g., ISO 14801 for dynamic loading, ISO 10993-1 for biocompatibility). The devices are expected to meet the performance criteria defined by these standards.

    8. The sample size for the training set

    • Applicability: Not applicable. This refers to the training of an AI/ML algorithm. The document describes a physical medical device.

    9. How the ground truth for the training set was established

    • Applicability: Not applicable. See point 8.