510(k) Data Aggregation

    K Number
    K232639

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-05-24

    (268 days)

    Product Code
    Regulation Number
    882.5805
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Apollo TMS Therapy System is indicated for the treatment of Major Depressive Disorder in adult patients who have failed to achieve satisfactory improvement from prior antidepressant medication in the current episode.

    Device Description

    The Apollo TMS Therapy System is an electromagnetic device that non-invasively delivers a rapidly pulsed magnetic field to the cerebral cortex in order to activate neurons within a limited volume without inducing a seizure. This method of cortical stimulation by application of brief magnetic pulses to the head is known as Transcranial Magnetic Stimulation (TMS).

    The Apollo TMS Therapy System is comprised of the following principal components:

    • User Interface
    • Main Unit (with or without housing)
    • Stimulation Coil
    • Coil Positioning System

    The operator controls the Apollo TMS Therapy System via the user interface (application software "Stimware"). "Stimware" is a treatment and data management software that administrates treatment protocols and the patient´s individual stimulation dose determined by the patient´s individual motor threshold. Stimulation is applied via the stimulation coil which is positioned to the left dorsolateral prefrontal cortex (DLPFC) by means of a coil positioning system. The observed and documented increase in cortical excitability after high frequency (10Hz) rTMS has been shown to persist beyond the duration of the train of stimulation.

    AI/ML Overview

    This document does not contain information about a study proving the device meets acceptance criteria. Instead, it is a 510(k) summary for the Apollo TMS Therapy System, demonstrating its substantial equivalence to previously cleared predicate devices for the treatment of Major Depressive Disorder.

    Therefore, I cannot provide the requested information regarding:

    • A table of acceptance criteria and reported device performance.
    • Sample size used for the test set and data provenance.
    • Number of experts and their qualifications for ground truth establishment.
    • Adjudication method for the test set.
    • MRMC comparative effectiveness study results.
    • Standalone performance.
    • Type of ground truth used.
    • Sample size for the training set.
    • How ground truth for the training set was established.

    The document primarily focuses on:

    • Intended Use and Device Description: The Apollo TMS Therapy System is an electromagnetic device for non-invasive stimulation of the cerebral cortex to treat Major Depressive Disorder.
    • Compliance with Standards: Conforms to various ISO and IEC standards for medical devices, electrical safety, usability, software life cycle processes, and risk management.
    • Non-Clinical and Clinical Performance Data: States that non-clinical testing was performed according to standards, and comparative testing for additional coils demonstrated substantial equivalence to predicate devices. It specifically mentions that clinical data on the treatment efficacy of the iTBS protocol was already demonstrated in a previous summary (K173620) and that verification testing confirmed constant intensity of individual stimuli.
    • Software Verification and Validation: Conducted in accordance with IEC 62304, considered "basic documentation" as software failures are not likely to present high-risk situations.
    • Risk Management: Applied throughout the product development life cycle, concluding no new hazards compared to predicate devices.
    • Substantial Equivalence: The core of the submission, asserting that the Apollo TMS Therapy System is substantially equivalent to the HORIZON® TMS Therapy System (K182853) and an earlier version of the Apollo TMS Therapy System (K180313). This equivalence is based on identical intended use, indications for use, system components, operational procedures, and treatment protocols (especially iTBS).
    • Comparative Table: Provides a detailed comparison of characteristics between the subject device and its predicates, highlighting similarities in indications for use, product code, classification, standard treatment stimulation parameters (e.g., area of brain stimulated, stimulation intensity, frequency, pulse train duration), output stimulation parameters, and coil parameters. It explicitly notes that the performance and clinical effectiveness are substantially equivalent.

    In summary, this document is a regulatory submission demonstrating substantial equivalence rather than a clinical study report with specific acceptance criteria and performance data from a clinical trial.