Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K232509

    Validate with FDA (Live)

    Date Cleared
    2023-11-13

    (87 days)

    Product Code
    Regulation Number
    878.4850
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Accu-Chek Safe-T-Pro Uno lancing device is intended to produce a capillary blood sample for testing utilizing small amounts of blood.

    Device Description

    The Accu-Chek Safe-T-Pro Uno lancing device is a sterile, single-use, disposable lancing device intended to be used by non-professional users 18 years and older and healthcare professionals. It is designed for capillary blood sampling from the fingertip of adults and children 1 year and older or, if the patient is a child under 1 year, from the heel.

    AI/ML Overview

    The provided text is an FDA 510(k) premarket notification for the Accu-Chek Safe-T-Pro Uno Lancing Device. It does not describe a study with acceptance criteria in the manner typically found in clinical trials or AI/software validation studies. Instead, it details a "non-clinical bench testing" approach to demonstrate substantial equivalence to a predicate device.

    Here's an analysis based on your request, highlighting what is and isn't available in the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly present a table of acceptance criteria alongside reported device performance for specific metrics. Instead, it refers to "design verification and validation testing" performed per "applicable FDA Guidance documents (Sharps Injury Prevention Features) and special controls (878.4850)." The conclusion states the device "performs as well or better than the legally marketed predicate device and legally marketed reference devices."

    The closest approximation to "device performance" mentioned is:

    • The Accu-Chek Safe-T-Pro Uno lancing device is designed for a single use only and has a sharps injury prevention feature where the lancet is retracted and concealed before and after use, and is rendered inoperative after use.
    • It uses a 28 Gauge needle with a 1.5 mm depth and is spring-driven.
    • Loading/priming is not required; activation is via a press release button.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not specify a "sample size used for the test set" or "data provenance" in terms of subject populations or data collection methods (retrospective/prospective). This is because the described testing is "non-clinical bench testing," meaning it was likely conducted in a laboratory setting on the device itself, rather than on human subjects or clinical data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This is not applicable as no "ground truth" established by experts for a test set is mentioned. The testing involves mechanical and design verification, not expert evaluation of results from human or image data.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. No adjudication method is mentioned as there's no expert review of outcomes.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. A MRMC comparative effectiveness study is relevant for AI or diagnostic imaging devices evaluating human performance. This document is for a medical device (lancing device) and does not involve AI or human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI or algorithm-based device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    Not applicable in the conventional sense of clinical ground truth. The "ground truth" for the non-clinical bench testing would be engineering specifications, regulatory standards (like ISO or FDA special controls for sharps injury prevention), and the performance characteristics of the legally marketed predicate and reference devices.

    8. The sample size for the training set

    Not applicable. This is not an AI/machine learning device that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable. This is not an AI/machine learning device.

    Summary of what is present:

    • Type of Study: Non-clinical bench testing (design verification and validation testing).
    • Purpose: To demonstrate the mechanical functions, safety (sharps injury prevention), and performance are suitable for use and are substantially equivalent to legally marketed predicate and reference devices.
    • Applicable Standards: FDA Guidance documents (Sharps Injury Prevention Features) and special controls (878.4850).
    • Conclusion: The device is "safe and effective for its intended use, and performs as well or better than the legally marketed predicate device and legally marketed reference devices."
    • Clinical Testing: "Not applicable; risk analysis confirmed that all identified risks were addressed and mitigated appropriately."
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1