510(k) Data Aggregation

    K Number
    K232372

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-07-12

    (339 days)

    Product Code
    Regulation Number
    878.4493
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ADVANTIME® synthetic absorbable surgical sutures are intended for use in general soft tissue approximation and/or ligation, but not for use cardiovascular surgery, microsurgery, ophthalmic surgery, and neurological tissue.

    Device Description

    ADVANTIME® sutures are intended to be used in suturing and/or tissue ligation in open or laparoscopic surgery. These sutures are available with a disposable needle attachment.

    ADVANTIME® is a single use, sterile, single-stranded (monofilament), synthetic bioabsorbable thread made of a poliglecaprone (a copolymer of glycolide and E- caprolactone of low tissue reactivity). Sutures are commonly used in digestive, urological, aesthetic, gynecological and intradermal surgery. To perform suturing on soft tissue wounds involves a series of precise steps. The surgeon grasps the needle, pushes it through tissue, pulls it through and repeats for continuous suturing. The same applies to double-armed needles. Further, knots are tied, excess thread is cut, and used needles are single use and safely disposed in accordance with hospital procedure.

    AI/ML Overview

    The provided document is a 510(k) summary for a surgical suture device (ADVANTIME Absorbable Suture) and does not describe an AI/ML powered device. Therefore, the questions regarding acceptance criteria, study details, and performance measures (such as sample size for test/training sets, expert ground truth, adjudication methods, MRMC studies, or standalone performance of an algorithm) related to AI/ML devices are not applicable.

    The document discusses the substantial equivalence of the ADVANTIME Absorbable Suture to a predicate device based on physical and biological performance testing for a medical device that is not an AI/ML device.

    Here's the information that is relevant from the provided text regarding this medical device:

    1. Acceptance Criteria and Reported Device Performance (Non-AI/ML Device)

    The device was subjected to the requirements of the United States Pharmacopeia (U.S.P.) monograph for Synthetic Absorbable Sutures.

    Acceptance Criteria (USP 43)Reported Device Performance
    Diameter (<861>)Meets USP requirements with minor variations (within FDA guidance allowance)
    Tensile strength (<881>)Meets USP requirements
    Needle attachment (<871>)Meets USP requirements
    Length (Absorbable surgical suture)Meets USP requirements
    Extractable color (Absorbable surgical suture)Meets USP requirements
    Resorption profilesEvaluated for each material; data provided in-vitro
    StabilityTesting completed
    Biocompatibility (ISO 10993-1)All tested aspects met (Cytotoxicity, Sensitization, Irritation, Material mediated pyrogenicity, Acute Systemic Toxicity, Sub-acute/sub-chronic/chronic toxicity, Genotoxicity, Carcinogenicity, Chemical characterization and toxicological risk assessment, Implantation, Degradation, Sterilization residue testing)

    2. Sample size used for the test set and the data provenance: Not applicable as this is not an AI/ML device study. Performance was evaluated based on physical and biological tests according to USP standards and ISO 10993-1.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable as this is not an AI/ML device. Ground truth for physical properties is defined by USP standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable as this is not an AI/ML device. Performance is objectively measured against USP standards.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable as this is not an AI/ML device.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable as this is not an AI/ML device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): For the physical characteristics, the ground truth is defined by the United States Pharmacopeia (USP) standards and test methodologies. For biological evaluation, the ground truth is established by standardized testing protocols detailed in ISO 10993-1.

    8. The sample size for the training set: Not applicable as this is not an AI/ML device.

    9. How the ground truth for the training set was established: Not applicable as this is not an AI/ML device.