Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K232020

    Validate with FDA (Live)

    Date Cleared
    2023-10-16

    (101 days)

    Product Code
    Regulation Number
    890.5525
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Iontophoresis Electrodes are intended to be used to introduce soluble salts and other drugs into the body as an alternative to hypodermic injection.

    Device Description

    The proposed device is indicated for the administration of soluble salts or other drugs into the body for medical purposes. The Iontophoresis Electrodes utilize adhesive electrodes have both anode and cathode electrodes that adhere to the skin. The electrodes may either be connected by leads (wires) to a separate controller Forming a current loop. The portion of the active electrode that contacts the skin also has a drug/solution reservoir. End users saturate the reservoir with the ionic solution to be delivered. The reservoir contacts the electrode and the drug/solution is delivered to the patient transdermally.

    Five models (Small, Medium, Large, Butterfly and Return Electrode) are included in this series products, and they have the same intended purposes, working theories and specifications.

    AI/ML Overview

    The provided document describes the acceptance criteria and the study that proves the device meets those criteria for the Iontophoresis Electrodes (K232020).

    Here's the breakdown of the requested information:

    1. A table of acceptance criteria and the reported device performance

    The document presents a comparison to a predicate device (Iontophoresis Drug Delivery Electrodes, K040495) to demonstrate substantial equivalence, rather than setting explicit numerical acceptance criteria for a new device's performance. The "acceptance criteria" are implied by the predicate device's specifications and the non-clinical test conclusions.

    Acceptance Criteria (Implied by Predicate & Standards Compliance)Reported Device Performance (Proposed Device)Remark
    Product Code: EGJEGJSame
    Regulation Number: 21 CFR 890.552521 CFR 890.5525Same
    Indications for Use: Introduce soluble salts and other drugs into the body as an alternative to hypodermic injection.Introduce soluble salts and other drugs into the body as an alternative to hypodermic injection.Same
    Patient Population: Patients requiring iontophoresis treatmentPatients requiring iontophoresis treatmentSame
    Usage Environment: Hospitals, medical clinicsHospitals, medical clinicsSame
    Design (Electrode Size): Comparable to predicate sizes (e.g., Small, Medium, Large, Butterfly, Return Electrode)ION-T01 (69x69 mm), ION-T02 (82x82 mm), ION-T03 (88x88 mm), ION-T04 (84x94 mm), ION-T05 (64x64 mm)Same (Note 1 - differences in design requirements, not affecting safety/performance)
    Main Materials: Silver/Silver Chloride (Ag/AgCl), Buffering Agent, PU Conductive carbon filmSilver/Silver Chloride (Ag/AgCl), Buffering Agent, PU Conductive carbon filmSame
    Anatomical Sites: For body surface skinFor body surface skinSame
    Working mode: Must be used as a complete set, consisting of an active drug delivery electrode and a passive return electrode.Must be used as a complete set, consisting of an active drug delivery electrode and a passive return electrode.Same
    Compatibility with other devices: Designed for use with iontophoresis devices only.Designed for use with iontophoresis devices only.Same
    Impedance: 2MΩ ~ 10MΩ2MΩ ~ 10MΩSame
    System current: 4.0mA4.0mASame
    Maximum Dosage: 80 mA-min80 mA-minSame
    Fill Volume: Comparable to predicate fill volumes (e.g., 1.5cc, 2.5cc, 4.0cc, 2.0cc for specific models)1.5cc (ION-T01), 2.5cc (ION-T02), 4.0cc (ION-T03), 2.0cc (ION-T04)Same
    Biocompatibility (Cytotoxicity): No CytotoxicityNo CytotoxicitySame
    Biocompatibility (Skin sensitization): No evidence of sensitizationNo evidence of sensitizationSame
    Biocompatibility (Irritation): No evidence of irritationNo evidence of irritationSame
    EMC, Electrical Safety (Electrical safety): Comply with IEC 60601-1Comply with IEC 60601-1Same
    EMC, Electrical Safety (EMC): Comply with IEC 60601-1-2Comply with IEC 60601-1-2Same

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    The document explicitly states: "No clinical study is included in this submission."
    Therefore, there is no sample size for a clinical test set or information on its provenance. The testing was non-clinical.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable, as no clinical study was conducted that would require expert ground truth. The evaluation focused on non-clinical performance and safety based on recognized standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable, as no clinical study was conducted.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is an Iontophoresis Electrode, which is a physical device for drug delivery, not an AI or imaging diagnostic tool that would involve human readers or AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a physical medical device, not an algorithm or software.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Not applicable, as no clinical study was conducted. The "ground truth" for the non-clinical tests was established by compliance with international standards (IEC 60601-1, IEC 60601-1-2, ISO 10993-5, ISO 10993-10, ISO 10993-23) and comparison to a legally marketed predicate device.

    8. The sample size for the training set

    Not applicable. This is a physical medical device, not a machine learning model.

    9. How the ground truth for the training set was established

    Not applicable. This is a physical medical device, not a machine learning model.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1